©UCR Office of Sustainability Ultra-Low Temperature Freezer Performance and Energy Use Tests Stirling Ultracold SU78UE Eppendorf Cryocube 570h Thermo Fisher TSX600 Delphine Faugeroux, MS Green Lab Coordinator Office of Sustainability - University of California, Riverside University of California Office of the President June, 2016

Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

  • Upload
    lehuong

  • View
    214

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

©UCR Office of Sustainability

Ultra-Low Temperature Freezer Performance and Energy Use Tests

Stirling Ultracold SU78UE Eppendorf Cryocube 570h

Thermo Fisher TSX600

Delphine Faugeroux, MS Green Lab Coordinator

Office of Sustainability - University of California, Riverside University of California Office of the President

June, 2016

Page 2: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

©UCR Office of Sustainability

Disclaimer

NEITHER UC RIVERSIDE, OR ANY OF ITS EMPLOYEES MAKES ANY WARRANTY, EXPRESSED OR IMPLIED, OR ASSUMES ANY LEGAL

LIABILITY OF RESPONSIBILITY FOR THE ACCURACY, COMPLETENESS, OR USEFULNESS OF ANY DATA, INFORMATION, METHOD,

PRODUCT OR PROCESS DISCLOSES IN THIS DOCUMENT, OR REPRESENTS THAT ITS USE WILL NOT INFRINGE ANY PRIVATELY-

OWNED RIGHTS, INCLUDING BUT NOT LIMITED TO, PATENTS, TRADEMARKS, OR COPYRIGHTS.

Acknowledgments

University of California, Riverside (UC Riverside) acknowledges the support provided by the manufacturers of this equipment who loaned, shipped and advised us on correct operation. We, the Office of Sustainability, especially wishes to thank the Botany and Plant Science Department, whose staff l has been outstanding in its help and support with providing access to valuable laboratory space. UC Riverside Students played an integral role in collecting data sets, assembling and installing thermocouples. Any errors in reporting are not to be attributed to their work, as it was reviewed and supervised by permanent staff personnel. Allen Doyle, Sustainability Manager, from UC Davis, has been very generous in loaning measurement equipment, as well as serving as advisor on their calibration and installation. He also participated in the review of this report and contributed to the interpretation of some of the results. Allison Paradise, executive director of the non-profit organization my green lab, for reviewing this report and providing valuable comments. UC Riverside would like to acknowledge the support provided by Environmental Health and Safety and the Geology Department for loaning testing equipment.

Use of UC Riverside Test Results and Other Related Information

UC Riverside does not endorse particular products or services from any specific manufacturer or service provider.

UC Riverside is committed to testing laboratory equipment using the customary and standard procedures as recommended by scientists,

manufacturers, and will report levels of accuracy and precision when quantified.

UC Riverside does not, in any way, endorse any of the equipment tested on campus.

UC Riverside test results are made available to the general public through the internet and occasional technical research reports and

publications and are protected under U.S. and international copyright laws.

In the event that UC Riverside data are to be reported, quoted, or referred to in any way in publications, papers, brochures, advertising,

or any other publicly available documents, the rules of copyright must be strictly followed, including written permission from UC Riverside

in advance and proper attribution to UC Riverside and the Office of Sustainability. In any such publication, sufficient text must be

excerpted or quoted so as to give full and fair representation of findings as reported in the original documentation from UC Riverside.

Legal Notice

Reference to specific products or manufacturers is not an endorsement of that product or manufacturer by UC Riverside.

Gratitude

This project would not have been possible without the generous support of the following persons. Any oversights are the responsibility of the author. Stirling Ultracold Neill Lane, CEO Tim Kehrli, Strategic Account Manager

Thermo Fisher Doug Holdhusen, Regional Sales Manager Rob Law, Territory Sales Representative

Eppendorf Daniela Marino, Product Manager Loreline G. Lee, Territory Manager

UC Riverside Dr. Mike Roose, Botany and Plant Science Department Chair Dr. John Cook, Director of Sustainability Students from the Green Lab team, interns and volunteers

Page 3: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

3 8/22/2016

Table of contents

Acknowledgement

I. Introduction

II. ULTs at UCR: audit and cost analysis

III. Testing Methods

1. Energy tests 2. Temperature Uniformity tests 3. Thermocouple installation 4. Noise test 5. Heat test

IV. Results

1. Energy 1.1 Energy consumption at different temperatures 1.2 Energy Consumption as a function of storage capacity 1.3 Cost of operation

2. Footprint 3. Performance: average internal freezer temperature 4. Performance: pull down and warm-up time 5. Temperature uniformity

5.1 General comparison 5.2 Temperature uniformity: Stirling at -80, freezer at capacity 5.3 Temperature uniformity: Thermo at -80, freezer at capacity 5.4 Temperature uniformity: Eppendorf at -80, freezer at capacity 5.5 Temperature variability at -80 versus -70 5.6 Air circulation and capacity 5.7 Temperature Uniformity: Sample versus air

6. Noise 7. Heat production 8. Amperage used

V. Conclusion

Appendix

Page 4: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

4 8/22/2016

I. Introduction

The energy intensity of laboratories has come into focus over recent years due to increased costs of electricity and an awareness of the large carbon footprint associated with their operations. Building designers are looking for new ways to reduce building plug loads and as well as cooling needs from equipment reject heat. Most cascade (compressor-based) Ultra Low Temperature (ULT) freezers consume as much power as typical household. They also generate heat and noise, need a significant amount of electricity, and require frequent maintenance. Because space in university laboratories is at a premium, the footprint of such equipment will be discussed as well. New energy efficient units are now available on the market and provide 50% energy reduction, as well as less heat production and less noise. During fall 2015 and winter 2016, staff and students from the Office of Sustainability tested three Ultra Low Temperature Freezers (Table 1). A Stirling, with assorted racks for sample storage, was purchased for testing purposes through funding from the University of California Office of the President as part as the University’s Carbon Neutrality Initiative. The Thermo Fisher and Eppendorf freezers, and racks for both units, were loaned to the campus for three months. The Thermo Fisher was sold to a campus lab after the testing was completed and the Eppendorf was returned to the manufacturer at the end of the loan period. The characteristics of each unit are described below.

Table 1: Tested freezer’s dimensions and capacities

Distributor Stirling Ultracold Thermo Fisher Eppendorf

Manufacturer Stirling Ultracold1 Thermo2 Eppendorf

Model SU78UE TSX600 Cryocube 570h

Cubic Feet 27.5 28.8 20.1

Storage Capacity: number of 2" Boxes

600 600 400

Voltage 120V

(can be plugged in 208V but tested with a 120V outlet)

208V 120V

Amperage 15 Amps 15 Amps 20 Amps

Picture

Page 5: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

5 8/22/2016

1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine previously used by NASA. This engine does not require oil lubrication and works continuously, opposed to the stop-start operation of a typical cascade compressor. The engine is located at the top of the unit and it has the size of can. By taking less space than 2 compressors, it provide a large storage capacity without occupying a large floor space. 2 The Thermo Fisher TSX600 can operate in two different modes, as described below, and both modes were tested:

- Standard Mode – by default. This setting can be used in most applications and is more energy efficient. Standard Mode is for researchers who do not require maximum temperature uniformity.

- High Performance Mode optimizes temperature uniformity in the freezer and should be used when validation protocols require tight temperature uniformity. In addition, this mode should be used when operating in ambient temperatures higher than 80°F (27°C), humidity levels above 50% RH, or dew points of 65 or higher.

Page 6: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

6 8/22/2016

II. ULTs at UCR: audit and cost analysis

In 2014, UCR began documenting the number and condition of Ultra Low Temperature Freezers (ULT) on campus. An inventory listed 175 ULTs in research labs across campus. The office of Sustainability audited ~70% of them. The ULTs ranged in age from less than one year to more than twenty years old. As part of the audit, 11 ULTs were monitored to gather data on energy consumption (Table 1). The energy consumption data collected varies from 15kWh/day to 32kWh/day per ULT. Energy used by all ULTs on campus represents 1.3% of the total electricity use for the campus.

Figure 1: Energy usage of monitored Freezer during the freezer audit in 2014-2015

Because energy savings can be significant for ULTs, UCR decided to investigate and compare ULTs marketed as being energy efficient. These models included the Stirling, Thermo TSX and the Eppendorf Cryocube (Table 2). From vendors’ data and onsite monitoring, we have approximated the energy use for regular new ULT and half-life regular ULT.

Table 2: Energy usage and operational cost comparison between regular ULTs and EE ULTs

Average consumption

(kWh)/day

Average consumption (kWh) /year

cost $/kWh in 2015

Yearly operational

cost ($)

kWh difference for 1 year

$ difference for 1 year

Energy Efficient freezer3 8.86 3,233

0.10

$ 323

Regular half-life freezer4 24 8,760 $ 876 5,527 $ 552

regular NEW freezer5 17 6,205 $ 620 2,972 $ 297

3 The energy consumption for Energy Efficient ULT comes from the average of onsite monitoring presented in Table 3. 4 The energy consumption for half-life ULT comes from the average of freezer monitored on site. 5 The energy consumption for regular new ULT comes from vendors’ data.

17

20

18

20

2829

22

31

22

15

23

15

17

19

21

23

25

27

29

31

33

1995 1997 1999 2001 2003 2005 2007 2009 2011 2013 2015

Ener

gy c

on

sum

pti

on

(kW

h/d

ay)

Year of purchase

Energy usage in function of the year of purchase

Page 7: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

7 8/22/2016

III. Testing methods

1. Energy tests

A WattsUp Pro Meter was used to measure ULT energy consumption. Wattage and voltage were measured simultaneously. The data were reported in Watts/day. The freezers were given 24 hours once they reached the set temperature to stabilize before and between each tests, then energy measurements were logged at 10 minute intervals and added up for a 24-hour period. For each test, all freezers were full. Racks were used first and Styrofoam coolers were used to complete empty space when the number of racks available did not suffice to occupy the full freezer. Rack contains cardboard boxes that were empty.

2. Temperature Uniformity tests

On the recommendation of Allen Doyle, Sustainability Manager at UC Davis, type J thermocouple (TC) wire for temperature sensing were selected. Fifteen TCs were placed inside the freezer, which were inter-calibrated for accuracy and precision in a methanol bath with dry ice chunks. The measurements were logged at 10 second intervals continuously for 2 weeks. All 15 TC’s were logged simultaneously using two Omega TC-08 panels. For consistency, uniformity is often measured on empty freezers, yet this is not like field conditions and it does not challenge the cooling system or air circulation. Previous work (Todd Davis and Kathi Shea citation) has shown marked difference between completely and partially full freezers, and at different temperatures. Thus, for each freezer, temperature uniformity was tested with 4 different conditions described below (see also Figure 7 and Figure 8): - 80°C, freezer filled to capacity - 80°C, freezer with one rack removed per shelf to increase air circulation - 70°C, freezer filled to capacity - 70°C, freezer with one rack removed per shelf to increase air circulation Doors opening tests were conducted as well. For each condition, three replicates of 30 second door openings and three replicates of 90 second door openings were conducted. For all temperature uniformity tests, the Thermo Fisher was set to High Performance mode. Due to lack of time, the Standard mode has not been tested.

3. Thermocouple installation

The drawing representing the TC (red circles) distribution for the Thermo Fisher freezer is shown in Figure 2. The three freezers were tested following the same distribution model. One TC was installed next to the temperature probe, one on each corner at the top and the bottom of the freezer, and one on each side, and under each shelf. TC’s were bent so that the tips did not touch any surfaces and were about 1 cm) from the walls. On the top front right corner and bottom front right corner, the TCs were placed inside a plastic vial (micro-centrifuge tubes), which was located inside a 2” cardboard box (Figure 4 and Figure 6). These locations were identified as sensitive to temperature change during door openings. Measuring air

Page 8: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

8 8/22/2016

temperature with a bare TC would not realistically indicate sample conditions, so enclosing the TCs in tubes and standard box simulated what samples experience. The blue squares represent the freezer probes installed by the manufacturer. They are present in all

units.

Figure 2: Thermocouple repartition inside the Thermo freezer

Page 9: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

9 8/22/2016

Figure 3: TC installation 1

Figure 4: TC inside a plastic tube, inside a box

Figure 6: Front top and bottom corner with TC in real sample condition - inside a tube, inside a box.

Figure 7: Freezer full of racks and coolers

Figure 8: One rack removed per shelf to increase air

flow movement

Figure 5: Putty around TCs at the exit

of the freezer

Page 10: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

10 8/22/2016

4. Noise test

To measure the noise produced by the freezer, a Brunel & Kjaer type 2230 sound meter was used. It was placed 1 meter from different part of the freezer (3.3 feet). The noise was measured from the front, back and sides of the freezer whenever possible.

5. Heat test

To estimate the heat produced by the freezer, an U8865 Infrared Thermometer Caterpillar was used. The temperature was measured from the front, back and sides of the freezer whenever possible.

Page 11: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

11 8/22/2016

IV. Results

1. Energy

1.1 Energy consumption at different temperatures

The energy consumption for each freezer was measured at -80°C with no door openings and with the freezers full of racks, as shown in Table 3. The electricity cost per year was calculated at a cost per space for storing one 2” box at -80°C.

Table 3: Energy consumption at the set point -80 °C. Measurements made using a WattsUp Pro meter.

The energy consumption results for each freezer at three different temperatures: -80°C, -70°C, and -60°C are shown below in Figure 9 and Table 4.

Figure 9: Energy consumption comparison between freezers

All freezers used less than 10 kWh per day at all temperature and performance settings. The Standard mode of the Thermo freezer used less energy than the High Performance, as advertised by the company. At -80°C and -70°C, the Stirling consumes the least amount of energy as compared to the Thermo and the Eppendorf, and they all converged on the same value at -60 °C.

9.82

7.92

6.39

8.73

7.42

5.55

9.35

6.69

5.34

7.55

6.12

5.54

5

6

7

8

9

10

-80 -70 -60

Ener

gy u

se in

kW

h/d

ay

Temperature in °C

Energy consumption comparison between freezer

Thermo TSX - high perf mode Thermo TSX - economy mode Eppendorf 570h Stirling

Stirling Eppendorf Thermo TSX – High Performance mode

Thermo TSX – Standard mode

Energy Use kWh/d at -80°C 7.55 9.35 9.82 8.74

Electricity Cost/year (0.1 $/kWh) $ 275 $ 341 $ 358 $ 318

Electricity Cost/Box/y $ 0.46 $ 0.85 $ 0.60 $ 0.53

Page 12: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

12 8/22/2016

The savings between the different temperatures are presented in Table 4. There is a clear savings potential when adjusting the temperature set point by 10°C or 20°C.

Table 4: Saving by increasing the temperature

1.2 Energy Consumption as a function of storage capacity

Two measures of volume were used to compare volumetric energy consumption, the listed cubic feet, and the number of standard 2” boxes they could hold when full (Figure 10 and Figure 11).

Figure 10: Energy used in Watt/cu.ft at -80°C

Figure 11: Energy used in Watt/2"box at -80°C

Because the interior of ULT freezers has so few obstructions or compartments, the volumetric comparisons gave the same results, and all freezers were efficient compared to values reported in a rebate calculator published in 2012 (Doyle, EERE ULT guide). The Stirling used the least amount of energy when compared on the basis of storage capacity, about 15% less than the Thermo at standard (economy) mode, and 22% less at high performance mode. Because of its smaller size, the Eppendorf volumetric energy consumption was much worse than the Stirling and Thermo.

1.3 Cost of operation

Based on their energy consumption, the cost of operating these ULTs per year is compared in Figure 12. The electricity price used for calculation was $0.1/kWh.

11

19

1413

0

5

10

15

20

25

Stirling Eppendorf Thermo TSX -high perf

mode

Thermo TSX -standard

mode

Energy used in Watt/cu.ft at -80°C

0.52

0.97

0.680.61

0.00

0.20

0.40

0.60

0.80

1.00

1.20

Stirling Eppendorf Thermo TSX -high perf

mode

Thermo TSX -standard

mode

Energy used in Watt/h/2"box

Stirling Eppendorf Thermo TSX –

High Performance mode Thermo TSX –

Standard mode Savings from -80°C to -70°C 19% 29% 19% 15%

Savings from -80°C to -60°C 27% 43% 35% 48%

Page 13: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

13 8/22/2016

Figure 12: Electricity cost/year at $0.1/kWh

The electricity cost is lower when operating the Stirling freezer, between 14% and 23%.

2. Footprint

The three tested freezers have different dimensions as well as different storage capacities. Scientists are motivated to purchase unitary freezers to maintain control and minimize their individual first costs, and buildings quickly fill up with individual freezers without an overall strategy. Because space in laboratories is at a premium, a space management approach to storage and energy efficiency would be indicated, and the footprint of such equipment should be part of a decision process (Figure 13). It is also interesting to link this information to the storage capacity (Figure 14). While square footage is often compared, freezer depth is rarely limiting as corridors are relatively wide, so horizontal wall length is the limiting dimension. [For a given room, 20-40% more storage capacity can be achieved with larger freezers (Allen Doyle, UC Davis)].

Figure 13: Square footage of each freezer

Figure 14: Number of 2” boxes per square feet of floor occupied

$275

$341$359

$319

0

100

200

300

400

Stirling Eppendorf Thermo TSX - high perfmode

Thermo TSX - standardmode

Electricity cost/year at $0.1/kWh

8.8

9.3

11.4

6

7

8

9

10

11

12

Stirling Eppendorf Thermo TSX

flo

or

spac

e u

sed

in s

q.f

t

Footprint (sq. ft)

68.6

42.9

52.7

0

20

40

60

80

Stirling Eppendorf Thermo TSX

Nu

mb

er o

f b

oxe

s

Number of 2" boxes/sq.ft of floor occupied

Page 14: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

14 8/22/2016

The Stirling has a smallest footprint and allows for the most number of boxes to be stored per 1 square foot of floor space occupied.

3. Performance: average internal freezer temperature

Several TCs – or temperature probes – were installed inside the three freezers. The average temperature for set points at -80°C and -70°C is presented in Figure 15 . The temperature set point is fixed in the unit in set point menu, which used the internal temperature probe. UCR did not calibrate the freezers’ probes, so the freezer set points are relative value. This figure shows the comparison between a full freezer and a freezer with one rack removed (see Figures 7 and 8 above for details).

Figure 15: Temperature Uniformity comparison between freezers

At -80 °C and when full of racks, all three freezers were comparable—about 3-4 degrees above set point. At -70 °C or when one rack was removed, the Thermo freezer was closest to set point. This is seen through the average temperature reading values measured inside compared to the relative temperature set point. Also, most of the freezers perform slightly better when one rack per shelf was removed (denoted “Less Rack” in Figure 15 above). This can be explained by the improvement of air circulation inside due to more open space. More details can be found in the section Temperature Uniformity, which presents the variability and range of temperature for each condition in more details. Also, the detailed temperature profile for each freezer can be seen in Appendix 2.

4. Performance: pull down and warm-up time

The pull down time is the time needed for the freezer to reach set point from room temperature.

-80 -80

-70 -70

-76.8

-78.8

-68.7-69.7

-75.9 -76.2

-66.1-66.8

-75.7 -76.0

-66.6-65.8

-82

-78

-74

-70

-66

-62

-58

-80 Full Rack -80 Less Rack -70 Full Rack -70 Less Rack

Temperature Uniformity Performance Comparison

set point - relative value Thermo High Perf Mode Eppendorf Stirling

Page 15: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

15 8/22/2016

The results are presented in Figure 16 below and calculated using the average of all the TCs installed inside the freezer, or an average of 14 points. The freezers were all full of racks and boxes. While pull down from room temperature only happens rarely in the lifetime of a freezer, the rate of pull down at the coldest temperatures is the most relevant comparison, as that is the condition after door openings.

Figure 16: Pull down time in hours

It takes 9 hours for the Eppendorf to reach its -80 set point, 15 hours for the Stirling and close to 26 for the Thermo Fisher in Standard Mode. This information will be valuable when you set up your freezer and need it ready in a short period of time. The warm up time is the time needed to reach ambient temperature from -80°C (Figure 17). All three freezers were full of racks/coolers when tested. This is particularly valuable data in the case of a power outage or freezer malfunction.

-80

-70

-60

-50

-40

-30

-20

-10

0

10

20

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29

Tem

per

atu

re in

°C

Time in hours

Pull Down time comparison in hours

Average Thermo Average Eppendorf Average Stirling

Page 16: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

16 8/22/2016

Figure 17: Average Warm up time from -80 to ambient

The three freezers behave similarly in case of power outage or freezer failure. The literature states that most biological samples can be stored for a few hours at -50°C. In case of power outage or freezer failure, if the door is maintained closed at all time, the scientists have ̴9hours (Table 5) to find a backup freezer before their samples reach -50°C and 2 days before they reach 0°C. This time should allow researchers to save their samples if the issue happen when people are present in the lab.

Table 5: Average warm up time from -80 to -50.

Stirling Thermo Eppendorf

Time to reach - 50°C in hours 8.9 9.2 9.1

Zooming in on the graph above (Figure 18), it can be seen that when the freezers are set to -80°C, the average temperature inside the freezer is higher, between -76°C and -74°C. The warm-up time difference between a freezer set to -80°C versus -70°C was found to be between 1.5 to 2.8 hours. A common argument is made when changing the temperature set up from -80°C to -70°C, which is the time lost in case of power/freezer failure. The answer to that question can now be discussed and the researcher can make their decision knowing that they might have an additional 1.5 to 3 hours with no door opening to move their samples in case of failure.

-80

-75

-70

-65

-60

-55

-50

-45

-40

-35

-30

-25

-20

-15

-10

-5

0

5

0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44 46 48 50 52 54 56 58 60 62 64

Tem

per

atu

re in

°C

time in hours

Warm up time AVERAGE AVERAGE Thermo AVERAGE Eppendorf AVERAGE Stirling

Page 17: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

17 8/22/2016

Figure 18: Zoom on Warm up time

5. Temperature Uniformity

5.1 General Comparison

For all temperature uniformity tests, the Thermo Fisher freezer was set up in High Performance mode. The totality of the TC functioning was averaged for each test condition. The standard deviation was calculated and is represented in the figure below. The Thermo and Eppendorf freezers had excellent uniformity and were about the same at 2-3 degrees, while the Stirling freezer had a standard deviation of 5-6 degrees.

Figure 19: Temperature Uniformity Performance Comparison with standard deviation

-80-78-76-74-72-70-68-66-64-62-60

0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00

time in hours

ZOOM - Warm up time AVERAGE

AVERAGE thermo AVERAGE Eppendorf AVERAGE Stirling

-76.8

-78.8

-68.7

-69.7

2.8

2.1

2.6

2.4

-75.9

-76.2

-66.1

-66.8

3.2

2.5

3.2

2.1

-76

-76

-67

-66

5.5

5.6

5.6

6.3

-80 -75 -70 -65 -60 -55 -50 -45 -40 -35 -30 -25 -20 -15 -10 -5 0 5

-80 Full Rack

-80 Less Rack

-70 Full Rack

-70 Less Rack

Temperature Uniformity Performance Comparison

STD Stirling Average Stirling STD Eppendorf Average Eppendorf STD Thermo Average Thermo

Page 18: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

18 8/22/2016

The range of temperature uniformity at -80 is presented in section 5.2 below. More temperature profiles at both -80 and -70 are compared in Appendix 2.

5.2 Temperature uniformity: Stirling at -80, freezer at capacity

The color code follows the rule explained below:

The Stirling freezer presents some variability in temperature. Five points are below -78°C, 3 are situated between -75°C and -78°C and 5 are above -75°C. The temperature repartition shows a pattern with the cold air coming from the middle between shelves 2 and 3, which then disperses up and down in the freezer. The warmer temperatures are at the top and bottom. The top seems likely to be warmest.

Figure 20: Stirling Temperature Uniformity

5.3 Temperature uniformity: Thermo at -80, freezer at capacity

The Thermo freezer, set up in High Perf mode, presents some variability in temperature. Six points are below -78°C, 6 are situated between -75°C and -78°C and 2 are above -75°C. The warmer points are located at the top along the door.

Figure 21: Thermo Temperature Uniformity

5.4 Temperature uniformity: Eppendorf at -80, freezer at capacity

Page 19: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

19 8/22/2016

The Eppendorf freezer presents some variability in temperature. Six points are below -78°C, 3 are situated between -75°C and -78°C and 5 are above -75°C. The cold air seems to come from the bottom and back of the unit. The warmer temperatures are found at the top part. Thermocouples located along the door measure warmer temperatures.

Figure 22: Eppendorf Temperature Uniformity

5.5 Temperature variability at -80 versus -70

Temperature variability is a measure of consistency over time. To better understand the differences between -80°C and -70°C, Figure 23 compares the average inside temperature for the three freezers when they are at capacity (noted FR for Full of Rack in the figure below) at both temperatures.

Figure 23: Inside temperature comparison at -70 and -80

-79

-78

-77

-76

-75

-74

-73

-72

-71

-70

-69

-68

-67

-66

-65

-64

10000 12000 14000 16000 18000 20000 22000 24000 26000 28000 30000

Tem

p. (

C)

Time (s)

-80 versus -70: 3 freezers comparison Full of Racks

Stirling FR no DO at -80 Stirling FR no Do at -70 Thermo FR no DO -80

Thermo FR no DO -70 Eppend FR no Do @ -80 air Eppend FR no DO @-70 air

Page 20: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

20 8/22/2016

The Stirling has a constant profile because the Stirling engine runs constantly while cascade compressors cycle. As the cascade compressors cycle, temperature fluctuation over time is more important than the Stirling Engine. Thermo and Eppendorf have a similar profile at -80°C but the Thermo unit is slightly more variable at -70°C.

5.6 Air circulation and capacity

Freezer capacity can also impact the temperature repartition. Indeed, a freezer full will allow less air circulation and might impact the uniformity. Figure 24 presents the results of these tests for the three freezers at -80°C when not filled to capacity (note LR for Less Rack inside, at the opposite of FR for Full of Racks).

Figure 24: Inside temperature comparison between a unit full of rack and not full of racks

For both Stirling and Thermo, the average temperature is closer to the set point when the freezer is not full. The Eppendorf seems to perform the same way whether it is full or not.

5.7 Temperature Uniformity: Sample versus air

Two TCs were placed inside plastic centrifuge tubes, which were then placed inside 2” boxes. Figure 25 shows the average temperature for the “air” –a bare TC in contact with the air in the freezer- and the “sample” -the TC placed in micro-centrifuge tube. These conditions were not tested with the Thermo Fisher.

-81

-80

-79

-78

-77

-76

-75

-74

-73

0 5000 10000 15000 20000 25000 30000

Tem

p (

C)

Time (s)

Comparison at -80°C: Full of racks versus Less racks

Stirling @ -80 FR stirling @ -80 LR Thermo @ -80 FR

Thermo @ -80 LR Eppend @ -80 FR Eppend @ - 80 LR

Page 21: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

21 8/22/2016

Figure 25: Comparison between air and sample temperature at -70 and -80

In the Stirling, the samples are warmer than the air 3°C - 4°C. In the Eppendorf, the variation of temperature due the compressors going on and off is largely diminished in the sample TC and a difference of 2°C - 3°C is noticeable.

6. Noise

Figure 26 shows the noise produced by each freezer. The background noise in the laboratory was 54 decibels, and may have compromised noise measurements from the freezers. The lab has two fumehoods, and general ventilation are responsible for this value. The noise produced by the freezer comes from the compressors, located at the bottom on the Thermo and Eppendorf, and the engine at the top on the Stirling.

-78-77-76-75-74-73-72-71-70-69-68-67-66-65-64-63-62-61

0 10000 20000 30000

Tem

p. (

C)

Time (s)

Stirling FR air vs sample at -80°C and -70°C

- 80 air FR -80 sample FR

-70 air FR -70 sample FR

-78-77-76-75-74-73-72-71-70-69-68-67-66-65-64-63-62-61

0 20000 40000 60000 80000Te

mp

(C

)Time (s)

Eppendorf FR air vs sample at -80°C and -70°C

-80 Avg Air -80 Avg Sample- 70 avg air -70 avg sample

Page 22: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

22 8/22/2016

Thermo Fisher

Stirling

Eppendorf

Figure 26: Noise production by each freezer

The noise produces by the three freezer is similar, even though it doesn’t come from the same location

(top or bottom of the unit).

7. Heat production

Figure 27 shows the heat produced by each freezer. The temperature in the laboratory was 64°F. The heat produced by the freezer comes from the engine, located at the bottom on the Thermo an Eppendorf, and at the top on the Stirling. This explains why the values are higher in these locations.

Page 23: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

23 8/22/2016

Thermo Fisher

Stirling

Eppendorf

Figure 27: Heat production by each freezer

The Thermo and Eppendorf show a lot of variation at the back bottom. This is due to the tool used to measure the heat exhausted; the heat gun was pointed at hot surfaces of the freezer, and some areas were hotter than others behind the protective grill.

8. Amperage used

For circuit load purposes, the amperage was measured when the freezer were turn on (Figure 28). Their different needs in volt and amps is indicated in the legend of the figure below.

Page 24: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

24 8/22/2016

Figure 28: Amperage drawn when the freezers start during the first 20 minutes

The manufacturers generally recommend one freezer per circuit. A 20 Amp, 208 V circuit could potentially support two freezers made by either Thermo or Stirling. A 20A, 120V circuit could support two Thermo, but might be near capacity with the two Stirlings.

0

2

4

6

8

10

12

14

16

0 2 4 6 8 10 12 14 16 18 20 22

Amperage used when ULTs start - in minutes

Eppendorf Cryocube 120 V / 20 A Thermo Fisher TSX - HP mode 208 V / 15 A Stirling 120 V / 15 A

Page 25: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

25 8/22/2016

V. Conclusion

First of all, these tests validated the vendor’s data that the three units tested are energy efficient, using less than 10kWh/day. The three freezers have similar levels of performance, but a few points can be considered when making a procurement decision:

Operational cost, thanks to the energy efficiency, you can save between $297and $552 (48% to 63%) at the minimum per year, solely from plug load.

Energy consumption: all freezers used less than 10kWh/day, but the Stirling used the least amount of electricity, costs the least to operate and offers the smallest operational cost per cu.ft.

Space: the Stirling offers the larger storage capacity while occupying the least floor space.

Temperature uniformity: the Thermo remained closest to the temperature set point and showed the best internal temperature uniformity.

By design, the Stirling engine offers less temperature variability than cascade compressors.

Heat exhaust: located at the bottom for the Thermo and Eppendorf, versus the top of the unit for the Stirling.

Finally, if you already have racks, the large internal dimensions of the Stirling and Thermo might allow you to reuse your existing racks.

Once the freezer is purchased, here are some considerations:

Chilling up your freezer can reduce between 15 and 30% of energy consumption.

In case of power outage, you will have ̴2hours less to move your samples if your freezer is set up at -70°C.

Page 26: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

26 8/22/2016

APPENDIX

Appendix table of contents: APPENDIX 1: Freezer profiles at -80°C with no door openings and one rack removed per shelve

APPENDIX 2: Temperature uniformity profiles at -80°C and -70°C

Page 27: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

27 8/22/2016

APPENDIX 1: Freezer profiles at -80°C with no door openings and one rack removed per shelf

-84

-80

-76

-72

-68

-64

-60

2 3 4 5 6 7 8 9

Stirling -80°C Less Rack

-84

-80

-76

-72

-68

-64

-60

0 1 2 3 4 5 6 7 8

Eppendorf -80°C Less Rack

-84

-80

-76

-72

-68

-64

-60

0 1 2 3 4 5 6 7 8 9 10

Thermo -80°C Less Rack

Page 28: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

28 8/22/2016

APPENDIX 2: Temperature uniformity profiles at -80°C and -70°C

Legend:

: Indoor freezer probe

Freezer set point at -80°C:

Freezer set point at -70°C:

Page 29: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

29 8/22/2016

-80°C Full of Racks One rack removed per shelf

Stirling

Thermo

Eppendorf

Page 30: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

30 8/22/2016

-70 Full of Racks One rack removed per shelf

Stirling

Thermo

Eppendorf

Page 31: Delphine Faugeroux, MS Office of Sustainability ... · 1 The Stirling freezer does not use typical compressor but a free-piston Stirling engine ... TC’s were bent so that the tips

31 8/22/2016