    DePuy Synthes Mitek Sports Medicine Visualization System Setup & Troubleshooting Guide

  • 2

    1 : Visualization Front Side Setup

    1 POWER ON TOWER

    Turn on the cart’s main power switch. Turn on the front side power switch of each individual component.

    2 ENTER PATIENT INFORMATION Select “New Case” on the Image Capture Unit.

    A surgical team member will then enter relevant patient data using a Keyboard or other external device.

    3 STERILE PROCEDURE

    A surgical team member working within the sterile field will pass the equipment-end of the sterilized Camera Head Cable and Light Cable to a surgical team member outside the sterile field for attachment to their respective units.

    4 ATTACH CAMERA HEAD TO CCU

    Remove protective cap from Camera Head Cable. Attach Camera Head Cable to CCU (Camera Control Unit) by aligning the red dots and pushing into the CCU.

    5 ATTACH LIGHT CABLE TO LIGHT SOURCE

    Make sure that the light turret is turned so the correct position is on top and that it matches the style of the Light Cable. DePuy Synthes Mitek Sports Medicine Light Cables use the Storz connection here.

    CAUTION Ensure the backside wiring setup is complete and that all backside power switches are turned on.

    1

    2

    3

    4

    5

  • 3

    1 : Visualization Front Side Setup

    6 ATTACH COUPLER TO CAMERA HEAD Attach the desired Coupler to the Camera Head

    by screwing them together. All surfaces should be dried thoroughly before attaching to ensure no fogging occurs.

    NOTE: It is recommended to use 2 different gauze pads with a 1 minute wait time between wiping down all surfaces.

    7 ATTACH SCOPE TO COUPLER

    Attach the desired Scope to the Coupler. All surfaces should be dried thoroughly with gauze pad before attaching.

    NOTE: Press locking mechanism to accept desired Scope. Once inserted, push button in opposite direction to lock Scope into place.

    8 ATTACH LIGHT CABLE TO SCOPE

    The DePuy Synthes Mitek Sports Medicine Light Cable uses an ACMI connection here. All Scope rings should be removed from the Scope’s light-post to access this connection.

    NOTE: DePuy Synthes Mitek Sports Medicine Light Cable Connections

    Proximal – Storz / Distal – ACMI

    9 WHITE BALANCE THE CAMERA

    Make sure the Light Source is turned on and not in Standby Mode. Place several gauze pads approximately 1 inch from the Scope tip and wait 1-2 seconds for shutter to adjust.

    Then press-and-hold the “+ / WB” button on the Camera Head for approximately 3 seconds. When the White Balance is complete, a green check mark appears over the WB symbol on the Display.

    10 ATTACH SHEATH TO SCOPE

    Attach the Sheath to the Scope and the Sheath’s Stopcocks to the appropriate FMS® Fluid Management System inflow and outflow connection tubes.

    6

    7

    8

    9

    10

    9

  • 4

    2 : Visualization Backside Setup

    DVI > DVI

    HDMI > DVI

    Remote > GPIO

    Connect Camera Control Unit to Display

    DVI 1 out > DVI 1 in

    Connect Camera Control Unit to

    Image Management System

    DVI 2 out > DVI in

    Connect Image Management System to

    Display

    HDMI out > DVI 2 in

    Connect Camera Control Unit to Image

    Management System (connect cable with red band to Remote 2)

    Remote 1&2 out > MedX EVO GPIO Port

    Connect Image Management System

    to Printer

    USB out > USB in

    Connect Image Management System to

    Wireless Imaging Control

    USB out > USB in

    Ethernet NIC > Ethernet in

    Connect Display to Wireless Transmitter (Tx) (see next page for secondary display setup)

    DVI out > DVI in

    Power to all units from power

    strip inside tower

    1

    2

    3

    4

    5

    6

    7

    8

    D IG

    IT A

    L PR

    IN TE

    R IM

    A G

    E M

    A N

    A G

    EM EN

    T SY

    ST EM

    LI G

    H T

    SO U

    RC E

    C A

    M ER

    A C

    O N

    TR O

    L U

    N IT

    D IS

    PL A

    Y W

    IR EL

    ES S

    TR A

    N SM

    IT TE

    R (T

    x)

    Display Power

    Printer Power

    Power Splitter (for use w/ Wireless Imaging)

    Image Management System Power

    USB & Ethernet (wireless control power & network)

  • 5

    2 : Visualization Backside Setup – Wireless Imaging/Secondary Display

    Wireless Setup – Bonding Wireless Units

    VISUALIZATION TOWER

    WIRELESS TRANSMITTER (Tx)

    SECONDARY DISPLAY

    WIRELESS RECIEVER (Rx)

    Connect Wireless Receiver (Rx) to Secondary Display

    DVI 1 out > DVI 1 in

    Power to wireless units utilizing power splitter

    off display power

    1

    2

    DVI > DVI Power Splitter (for use w/ Wireless Imaging)

    Display Power

    1 Press and hold the “bond” button (see image 1 for location) on the transmitter (Tx) unit until its LED begins flashing orange rapidly, then release.

    NOTE: Ensure both displays are powered on prior to performing the pairing procedure.

    Next press and hold the “bond” button (see image 2 for location) on the receiver (Rx) unit until its LED begins flashing orange rapidly, then release.

    3 When the bonding process is complete each unit’s LED will be steady blue.

    3

    2

    1 2

  • 6

    3 : Troubleshooting

    PROBLEM CAUSES POSSIBLE SOLUTION

    Device will not power on

    Power cable unplugged Ensure power cable is plugged in

    Supply Mains switch OFF Power ON Supply Mains switch

    Power switch OFF Power ON Power switch

    Fuse is blown Replace fuse

    Video cart is not powered ON Power ON video cart

    Live video does not display

    Camera Head (TCKI) not connected Connect Camera Head (TCKI)

    Video cable not connected to the device Connect video cable to device output

    Video cable not connected to the Display Connect video cable to Display input

    Display is not powered ON Power ON Display

    Display cannot accept video resolution Use a display which complies with the

    technical specifications

    Supply Mains switch powered OFF Power ON Supply Mains switch

    Power switch OFF Power ON Power switch

    Display not connected to power source Connect Display to power source

    Display set to wrong input Set Display to correct input

    LCD screen does not function

    Firmware fault Reset Camera Control Unit (OVBI) by powering OFF and the ON the Supply

    Mains Power switch

    Supply Mains switch OFF Power ON Supply Mains switch

    Power switch OFF Power ON Power switch

    Camera indicator light does not turn on

    Camera Head (TCKI) not connected Connect Camera Head Camera Head

    (TCKI) ) to device

    Camera Head (TCKI) not recognized by device

    Ensure the Camera Head (TCKI) is properly connected – Reset Camera

    Control Unit (OVBI)

    Supply Mains switch OFF Power ON Supply Mains switch

    Power switch OFF Power ON Power switch

  • 7

    3 : Troubleshooting

    PROBLEM CAUSES POSSIBLE SOLUTION

    Loss of video

    Video output cable disconnected from device

    Connect video cable to device output

    Video output cable disconnected from Display

    Connect video cable to Display input

    Display disconnected from power or turned off

    Ensure Display is connected to power and turned on

    Display settings changed Ensure Display settings are correctly set

    Device disconnected from power or powered off

    Ensure device is connected to power with Supply Mains switch powered ON –

    Power ON Power switch

    Video resolution changed Video resolution changes cause

    temporary video loss If video loss persists, reset device

    Video timing changed Video timing changes cause

    temporary video loss If video loss persists, reset device

    Light Source power off Turn Light Source on

    Color bars display Camera Head (TCKI) disconnected Reconnect Camera Head (TCKI)

    Incorrect color representation

    Improper White Balance Ensure screen is white and contains detail

    Bright white without detail will not white balance correctly

    Device is in Fiberscope Specialty Fiberscope Specialty may cause

    poor color quality

    Debris on Endoscope tip Clean Endoscope tip

    Debris on Device window or Coupler optics

    Remove Coupler from device and clean glass surfaces on device and Coupler

    Improper video settings Adjust video settings or reset settings

    to factory defaults

    Video too dark

    Exposure level too low Increase exposure until any area in

    the image is too bright

    Light Source too low Increase Light Source level

    Obstruction in light path Check Light Cable and Scope for broken

    fibers or light obstruction at Scope tip

    Brightness level too low Increase the brightness level

    Gamma level too low Increase the gamma level

    Video too bright

    Brightness leve