Designed to amplify your talent:
Introducing R-Evolution®
2
Presenting R-Evolution®
• Agile Fluidics™ supporting Anterior Chamber stability via:
– IOP Management with Advanced Microprocessor-controlled Active Irrigation System and the proprietary algorithm “Dynamic IOP Control” acting synergistically to mitigate surge via adjusting the intraocular pressure and responding to varying aspiration demand
– Selectable Aspiration System between Flow or Vacuum based pumps designed for surgeons who who like to be free of constraints
– Unique EasySys Cassette System with integrated dual pump fluidics interface engineered with the dual purpose of reducing the surge and boost performance
– Integrated and Motorized IV Pole for optional Gravity Irrigation
• Optimized Energy Management fostering U/S efficiency via:
– Minimal Stress™ Technology, a patented technology to optimise U/S energy delivery, it always ensures consistency between programmed and actual phaco tip stroke regardless of nucleus hardness
– Balanced U/S Cutting Protocols designed to boost efficiency
– 40 KHz Phaco Handpiece, choice between Slim4 (traditional 4 crystals) or SIX (enhanced 6 crystals)
– Wide Phaco Tip range supporting incision size from 1.8 to 2.8mm, available with straight, flared or flared bent configuration
R-Evolution is the system that gives a wide range of options for each step of the cataract, vitreo-retinal or combined procedure, enhancing performances and functionalities
3
Presenting R-Evolution®
Vitreo-Retinal Surgery providing everything needed via:
• Uninterrupted vitreous removal flow with the TWEDGE™ vitreoretinal cutters, available in 23, 25 and 27G, for a paradigm shift of efficiency during Vitrectomy procedures either in the anterior or posterior segment
• LED LIGHTING SYSTEM with 3 independent LED light sources with selectable color filters for enhanced visualization, free of harmful UV and IR emissions
• Integrated Green Laser Module 532 nm for endo-photocoagulation (optional)
• Ancillary functions: Fluid/air exchange, Silicon oil Injection/Removal, Eso/Endo Diathermy, Pars Plana phaco option.
• ANGel™ Perfusion Pressure Monitoring of ocular perfusion pressure
User Interface assisting human-machine interaction via:
– Large High Visibility 19” touch screen
– Wireless Foot-Pedal Programmable as Single Linear or Dual Linear
– Same EasySys Cassette for Phaco/Retina/Combined procedures
– Several programmable Voice and Audio Feedbacks
– IR Remote Control (optional)
Professional solutions sustaining trustable partnership via:
• Highly professionally trained clinical application support and technical service
• BVI wide products portfolio with flexible commercial solutions tailored to your needs
R-Evolution is the system that gives a wide range of options for each step of the cataract, vitreo-retinal or combined procedure, enhancing performances and functionalities
4
Agile Fluidics™: IOP Management
The core components of Agile Fluidics™ include:
• The Advanced Microprocessor-controlled Active Infusion System delivering enhanced Irrigation IN-FLOW capabilities and supporting faster reaction time to compensate for IOP fluctuations
• The Dynamic IOP Control System is a proprietary sw. algorithm that links the IOP to demanded vacuum smoothing risks associated with IOP fluctuations at varying aspiration flow rates.
WITHOUT Dynamic IOP Control System WITH Dynamic IOP Control System
5
Agile Fluidics™: Pump System & Surge Reduction
❑ High performance Flow based or Vacuum pump selectable on fly❑ Selectable Integrated Motorized IV pole for optional gravity irrigation❑ Further Surge reduction through synergic work of the small-bore aspiration tubing combined with Flared tip design
I.D.Ø 1.3mm
small-bore aspiration tubing U/S Flared tip design
Surge = IOP drop x duration
6
Optimized U/S Energy: Minimal Stress™ Technology
Normal phaco System without minimal stress
Contact between phaco tip and cataract material
Phaco System with minimal stress technology
Contact between phaco tip and cataract material
Minimal Stress™ Technology works in the background to maintain consistency between the programmable phaco tip stroke and the actual tip elongation, when the phaco tip faces the resistance opposed by cataract material.
7
Optimized U/S Energy: Minimal Stress™ Technology
Minimal Stress™ Technology is characterized by monitoring and sensing the dumping of the phaco tip stroke under the resistance of the lens material, it transmits continuous feedback related to the actual tip movement and continuously adjust energy supply to the piezoelectric components of the phaco handpiece in order to align actual tip elongation to the value the surgeon has programmed.
8
NO Minimal Stress™ Technology WITH cataract impact
Normal phaco System without minimal stress
Contact between phaco tip and cataract material
Programmed tip elongation (stroke µm)
Delivered tip elongation (stroke µm)
NO consistency between the programmed phaco tip stroke and the actual tip elongation, when the phaco tip faces the resistance opposed by cataract material.
Cataract particle
9
WITH Minimal Stress™ Technology WITH cataract impact
Programmed tip elongation (stroke µm)
Delivered tip elongation (stroke µm)
Phaco System with minimal stress technology
Contact between phaco tip and cataract material
Minimal Stress Technology works in the background to maintain consistency between the programmed phaco tip stroke and the actual tip elongation, when the phaco tip faces the resistance opposed by cataract material.
Cataract particle
10
CUTTINGEMULSIFY
HEAT
REPULSION
FLUIDICSCOOLINGATTRACT
HOLD
REMOVE
ON ONOFF OFF
1 sec.
Balanced U/S cutting protocols
• ON period supports cutting/emulsification without repulsion• OFF period allows the fluidics to solely act and facilitate
attraction, holding and/or pulling-removal of lens material
ON OFF
CUTTINGEMULSIFY
HEAT
REPULSION
FLUIDICSCOOLINGATTRACT
HOLD
REMOVE
CUTTINGEMULSIFY
HEAT
REPULSION
FLUIDICSCOOLINGATTRACT
HOLD
REMOVE
11
Frequency range from 1Hz to 100Hz
3 efficiency tested prefixed duty cycles protocols:
➢ STD (Standard): 50% Time ON - 50% Time OFF
➢ Cold: 25% Time ON - 75% Time OFF
➢ Ice Cold: 10% Time ON - 90% Time OFF
Optimized U/S Energy: boost of efficiency
Cold: Ton=25ms, Toff=75ms
12
Entering POS3 start the phaco pulse is delivered as a single burst of ultrasound power with a duration of 120 ms.To obtain another single burst must move footswitch from position 3 to 2 and enter again POS3 start
Entering POS3 start the phaco pulse is delivered as single burst with a duration of 80 ms, the nominal interval between bursts is approx. 1 second, as the footswitch is depressed the interval between burst decrease (frequency of bursts increases) up to a maximal rate of 4 bursts per second at the end of POS 3
Entering POS3 start, phaco pulse is delivered with a duration of 80 ms, the interval is approx. 1 second, as the footswitch is depressed the frequency of bursts increases up to become a continuous mode at the end of POS 3
Optimized U/S Energy: delivery protocols
POS 3 start POS 3 end
µm
POS 3 endPOS 3 start
POS 3 start POS 3 end
POS 3 start POS 3 end
T
T
T
µm
µm
In burst mode the U/S energy is modulated by regulating intervals
rather than the stroke which remain panel
Appropriate use of foot-pedal with Burst enhance efficiency of the lens
removal
13
Streamlined phaco tip options
Straight or bent depending on personal preferenceSame Color code sleeve/tip
Incision 3.0-2.8mm
Incision 2.4-2.2mm
Incision 2.0-1.8mm
Port OD Port ID Shaft OD Shaft ID Port area Vacuum Holdability Cutting edge Stroke µm Cut-ability
phaco tip 20G 30° straight 0,9 0,7 0,9 0,7 1,78 563 1000 3,05 33 100,00
phaco tip 20G 30° flared 1,04 0,94 0,8 0,6 3,20 312 1000 3,53 28 100,00
phaco tip 21G 30° straight 0,8 0,6 0,83 0,6 1,31 766 1000 2,71 37 100,01
phaco tip 21G 30° Flared 0,94 0,84 0,7 0,45 2,56 391 1000 3,19 31 100,01
Vac x Port Area Stroke X Port edge
14
User Interface
• Optikon has simplified man-machine interaction by directing the surgeon's gaze towards a single 19" touchscreen.
• The user-friendly graphic interface is conceived to ensure easy handling of a highly complex instrument Wireless footswitch with
programmable buttons
15
Posterior Segment Surgery: GUI & connectivity
16
33 mm long
28 mm long
Posterior Segment Surgery: TWEDGE™ 20K
The same vitrectomy probe is also applicable in the anterior vitrectomy
• The chart shows the differences in aspiration flow(balanced salt solution, Vacuum 650 mmHg, Venturipump, that are achieved using the Twedge™ vitreous cutter (solid line) compared to a single blade vitreous cutter (dashed line), as the cutting rate changes.
• In particular, the solid line shows how the flow remains constant up to 20,000 cuts/min with the Twedge™ vitreous cutter (10,000 cuts/min with
• R-EVO smart); in comparison, the flow decreases as cutting rate increases using the single blade vitreous cutter.
0,25 mm
Same tip-to-port distance for all gaugesa
28 mm long
17
Posterior Segment Surgery: LED LIGHTING SYSTEM
• The R-Evolutio® CR is equipped with three independent, high-efficiency LED light sources, free of harmful UV and IR emissions.
• The LED lighting system allows surgeons to perfect tissue visualization (white light close to sunlight) and guarantees the maximum protection against phototoxicity.
18
The R-Evolution®
R-Evolution® combines functionality and