Determination of Sulfur in Biodieselby X-Ray Fluorescence Spectroscopy

Bruce S. MacDonald, Lydia R. Barker, W. Robert Kelly, William F. GuthrieNational Institute of Standards and Technology

Page 2: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

ASTM D 6751 – 06aStandard Specification forBiodiesel Fuel Blend Stock (B100) for Middle DistillateFuels

Page 4

X1.5 Sulfur

NOTE X1.1 – “Test Method D 5453 (Ultraviolet Fluorescence) should be used withbiodiesel.”

“Biodiesel sulfur analysis … using Test Method D 2622 (Wavelength Dispersive XRF)yielded falsely high results due to the presence of oxygen in the biodiesel.”

This preliminary study will show that D 5453 and WD-XRFcan give consistent results for sulfur in biodiesel.

Page 3: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

Sulfur Determinations in Biodiesel by XRF

Instrument: XOS SINDIE 4000

Problems with Biodiesel determinations using calibration curves:

1. Unknown sample and calibration samples must be matrix matched.2. The C/H problem3. The oxygen problem: biodiesel contains ≈11 % oxygen; petroleum dieselis oxygen free.4. If biodiesel is compared to a petroleum diesel calibration curve, thenthe biodiesel sulfur value will be low by ≈ 20 %.

Page 4: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

? biodiesel is approximately 11% oxygen by weight

? regular diesel standards contain negligible oxygen

X

X

X

o

• Calibration curve yields a bias result because of differential absorptionof sulfur x-rays (interference or matrix mismatch).

Graphical illustration of the oxygen problem in XRF:

Page 5: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

The Solutions:? Matrix effects can be overcome by:

1. Math – make corrections

2. Experimental design – match matrices

3. Gravimetric standard addition method

systematic addition of analyte to samples; linear regression: x intercept = concentration of analyte in unspiked sample

====

<<<<

( m s i C s )/m x i

- 2 0 0 2 0 4 0 6 0 8 0 1 0 0 1 2 0

[(m

x i + m

s i)/mx i]R

i

3 0 0 0

7 0 0 0

1 1 0 0 0

1 5 0 0 0

[ S ] = - b / m = 1 2 . 3 9

====

Page 6: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

Volumetric Standard Addition Method (SAM)

Graphics from David L. ZellmerDepartment of ChemistryCalifornia State University, Fresno

T.J. Chow & T.G. ThompsonAnalytical Chemistry 1951, 27, 18-21.

Page 7: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

Experimental Approach

? NIST diesel SRM 2723a (11 ± 1.1 ppm sulfur) and CRM LGC10 (11 ± 0.9 ppm sulfur) as control “unknowns”

? calibration curve and standard addition with diesel SRM 2770

? various biodiesels

? designer diesel SRM calibration curve and standard addition with high concentration sulfur (SRM 1624d)

? Are there significant differences?

Page 8: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

Petroleum Diesel Samples

Biodiesel Samples

-- 18 %18 %7.51 ± 0.376.16 ± 0.10SRM 2773

-- 20 %20 %12.29 ± 0.139.78 ± 0.29NREL 52533

3.59 ± 0.26

11.69 ± 0.35

10.94 ± 0.35

Standard AdditionMethod (SAM)

(ppm)

-- 20 %20 %2.86 ± 0.22 NREL 52537

- 2.0 %11.46 ± 0.27LGC 10

- 2.4 %10.68 ± 0.48SRM 2723a

RelativeDifference

%

Calibration Curve(ppm)

Sample

Results by the Two Methods

Page 9: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

Petroleum Diesel Samples

Biodiesel Sample

7.49 7.49 ±± 0.500.507.51 ± 0.376.16 ± 0.10SRM 2773

11.69 ± 0.35

10.94 ± 0.35

Standard AdditionMethod (SAM)

(ppm)

11.0 ± 0.911.46 ± 0.27LGC 10

11.0 ± 1.110.68 ± 0.48SRM 2723a

CertifiedValues(ppm)

Calibration Curve(ppm)

Sample

Results by the Two Methods compared toCertified Values

Round Robin Data D 5453 April 2007N = 44; 95 % CI

Page 10: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

1 2 3

Sulfurµg/g

(ppm)

5

6

7

8

9

XRFCalibration

Curve

XRFGravimetric

Standard AdditionMethod(SAM)

RR D5453April 2007

n = 44

Sulfur in BiodieselSRM 2773

Page 11: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

Sulfurµg/g

(ppm)

9

10

11

12

13

Calibration Curve

G-SAMNIST Certified Value

SRM 2723aERM - EF 673a

LGC 10

Calibration Curve

G-SAMERM Certified Value

Sulfur in Two DieselCertified Reference Materials

Page 12: Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy · 2013-08-28 · Determination of Sulfur in Biodiesel by X-Ray Fluorescence Spectroscopy Bruce S. MacDonald,

Summary? Biodiesel XRF problem: matrix effects – Oxygen and Carbon

? A Solution: gravimetric standard addition method (G-SAM)

? Statistical analysis of uncertainties is complex.

? The XRF method is very precise; uncertainty is less than 1 ppm in the 6 to 10 ppm range.

? The XRF method using G-SAM is accurate - uncertainties overlap certified values.

? For B5 and B20 it may be possible to use blending to construct calibration curves.