Development of Picosecond- Resolution Large-Area Time-of- Flight Systems C. Ertley 2 , J. Anderson 1 , K.Byrum 1 , G.Drake 1 , H.Frisch 2 , J. Genat 2 , H. Sanders 2 , F.Tang 2 , J. Va’vra 3 PSEC is a collaborative effort between groups at Argonne National Laboratory 1 , University of Chicago 2 , and the Stanford Linear Accelerator Center 3 . Mirrors to delay light. 50/50 Beam Splitter Lens to focus beam on MCP MCP 1 MCP 2 X-Y Stager Laser Test Stand at Argonne A dark box was constructed at Argonne National Laboratory to test the MCP’s. It uses a Hamamatsu laser to illuminate the MCP’s and commercial electronics to measure the timing resolution and collected charge. The intrinsic jitter of the system is ~4ps and it has a resolution of 3.13ps. Results Abstract The measurement of time-of-flight (TOF) of relativistic particles in high-energy colliders with psec resolution would qualitatively change the ability to identify underlying parton-level processes at future colliders or upgrades of existing detectors. We are exploring a psec-resolution TOF system using micro-channel plates (MCP's) incorporating: a source of light with sub-psec jitter, in this case Cherenkov light generated at the MCP face (i.e. no bounces), short paths for charge drift and multiplication, a low- inductance return path for the high-frequency component of the signal, optimization of the anode for charge-collection over small transverse distances, and the development of multi-channel psec-resolution custom readout electronics directly mounted on the anode assembly. TDR Results: 256 = 11pF The second collection scheme we tested read out groups of 16, 9, 4, and 1 anode. The board was attached to the Mark-1 and the capacitance was again measured using time domain reflectometry (TDR). The board was not correctly aligned causing loss in gain. We are currently working on a new way to attach the collection scheme to the “Mark-1”. Conclusion We have measured the timing resolution of two MCP’s using a newly assembled test-stand based on a Hamamatsu PLP-10 picosecond laser and a commercial CAMAC readout electronics system. We have developed a new collection scheme to be attached to a 1024 anode MCP. The capacitance and gain of the MCP and board were tested. Micro-Channel Plate PMT The “Mark-1” is a 1024 anode 10um pore micro- channel plate photo-multiplier tube. We are testing charge collection schemes by clamping the board to the tube and putting a conductive elastomer between them. The commercial tube is a 25um pore 64 anode tube from Photonis. It has a commercially available collection scheme. The “Mark-P” is also a 25um 64 anode tube. It has some improvements, including a new ground plane to reduce ringing. Tim ing R esolution of 408nm vs.635nm Laser 0 20 40 60 80 100 120 0.0 20.0 40.0 60.0 80.0 100.0 Npe Tim ing R esolution (ps) 635nm 408nm 635nm 0 10 20 30 40 50 60 70 80 90 100 110 0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0 Npe Tim ing R esolution (ps) 0dB 10dB 20dB 408nm 0 10 20 30 40 50 60 70 80 90 0 20 40 60 80 100 Npe Tim ing R esolution (ps) 0dB 10dB 20dB 30dB We have measured the timing properties of two MCP-PM’s from Photonis. The two 64-anode 25- micron pore tubes have a commercially available collection scheme and were used to find a limit on the timing resolution. The system was calibrated before the measurements were taken. Here we show a comparison of a red laser (635nm) and a blue laser (408nm). The discontinuity in the 635nm graph is caused by a change in attenuation. Collection Anode The first collection scheme that was tested tied 256 anodes together. After attaching it to the Mark-1 the capacitance was measured to be 11pF using time domain reflectometry (TDR). This helped explain the slow decay time in the pulse. TDR Results : 1 = 1.2pF, 4 = 1.5pF, 9 = 2.3pF, 16 = 2.8pF Ch 1: Mark-1 (5mV/div) (2ns/div) Ch 2: Commercial (5mV/div) 9 – pixels tied together (~6mm) 1 – pixel (~6mm) 2.29kV 1.67kV Ch 1: Commercial (5mV/div) (2ns/div) Ch 2: Mark-1 (5mV/div) 256 – pixels tied together (~26mm) 1 – pixel (~6mm) 2.29kV 1.67kV Histogram using the 408nm laser set at ~50pe. Histogram using the 635nm laser set at ~55pe. The “Mark-1” in the clamp with a new collection scheme. The anode side of the “Mark-1”. The commercial tube with a small board tying 4 anodes together. All of the anodes on the “Mark-P” are grounded except for the one being read out. Ch 1: Yellow Ch 2: Blue Ch 1: Yellow Ch 2: Blue

Development of Picosecond-Resolution Large-Area Time-of-Flight Systems

  • Upload
    camdyn

  • View
    33

  • Download
    1

Embed Size (px)

DESCRIPTION

Development of Picosecond-Resolution Large-Area Time-of-Flight Systems C. Ertley 2 , J. Anderson 1 , K.Byrum 1 , G.Drake 1 , H.Frisch 2 , J. Genat 2 , H. Sanders 2 , F.Tang 2 , J. Va’vra 3. Abstract - PowerPoint PPT Presentation

Citation preview

Page 1: Development of Picosecond-Resolution Large-Area Time-of-Flight Systems

Development of Picosecond-Resolution Large-Area Time-of-

Flight SystemsC. Ertley2, J. Anderson1, K.Byrum1, G.Drake1, H.Frisch2, J. Genat2,

H. Sanders2, F.Tang2, J. Va’vra3

PSEC is a collaborative effort between groups at Argonne National Laboratory1, University of Chicago2, and the Stanford Linear Accelerator Center3.

Mirrors to delay light.

50/50 Beam Splitter

Lens to focus beam on MCP

MCP 1

MCP 2

X-Y Stager

Laser Test Stand at ArgonneA dark box was constructed at Argonne National Laboratory to test the MCP’s. It uses a Hamamatsu laser to illuminate the MCP’s and commercial electronics to measure the timing resolution and collected charge. The intrinsic jitter of the system is ~4ps and it has a resolution of 3.13ps.

ResultsAbstractThe measurement of time-of-flight (TOF) of relativistic particles in high-energy colliders with psec resolution would qualitatively change the ability to identify underlying parton-level processes at future colliders or upgrades of existing detectors. We are exploring a psec-resolution TOF system using micro-channel plates (MCP's) incorporating: a source of light with sub-psec jitter, in this case Cherenkov light generated at the MCP face (i.e. no bounces), short paths for charge drift and multiplication, a low-inductance return path for the high-frequency component of the signal, optimization of the anode for charge-collection over small transverse distances, and the development of multi-channel psec-resolution custom readout electronics directly mounted on the anode assembly.

TDR Results: 256

= 11pF

The second collection scheme we tested read out groups of 16, 9, 4, and 1 anode. The board was attached to the Mark-1 and the capacitance was again measured using time domain reflectometry (TDR). The board was not correctly aligned causing loss in gain. We are currently working on a new way to attach the collection scheme to the “Mark-1”.

ConclusionWe have measured the timing resolution of two MCP’s using a newly assembled test-stand based on a Hamamatsu PLP-10 picosecond laser and a commercial CAMAC readout electronics system. We have developed a new collection scheme to be attached to a 1024 anode MCP. The capacitance and gain of the MCP and board were tested.

Micro-Channel Plate PMTThe “Mark-1” is a 1024 anode 10um pore micro-channel plate photo-multiplier tube. We are testing charge collection schemes by clamping the board to the tube and putting a conductive elastomer between them. The commercial tube is a 25um pore 64 anode tube from Photonis. It has a commercially available collection scheme. The “Mark-P” is also a 25um 64 anode tube. It has some improvements, including a new ground plane to reduce ringing.

Timing Resolution of408nm vs. 635nm Laser

0

20

40

60

80

100

120

0.0 20.0 40.0 60.0 80.0 100.0

Npe

Tim

ing

Re

so

luti

on

(p

s)

635nm

408nm

635nm

0

10

20

30

40

50

60

70

80

90

100

110

0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0

Npe

Tim

ing

Re

so

luti

on

(p

s)

0dB

10dB

20dB

408nm

0

10

20

30

40

50

60

70

80

90

0 20 40 60 80 100

Npe

Tim

ing

Re

so

luti

on

(p

s)

0dB

10dB

20dB

30dB

We have measured the timing properties of two MCP-PM’s from Photonis. The two 64-anode 25-micron pore tubes have a commercially available collection scheme and were used to find a limit on the timing resolution. The system was calibrated before the measurements were taken. Here we show a comparison of a red laser (635nm) and a blue laser (408nm). The discontinuity in the 635nm graph is caused by a change in attenuation.

Collection AnodeThe first collection scheme that was tested tied 256 anodes together. After attaching it to the Mark-1 the capacitance was measured to be 11pF using time domain reflectometry (TDR). This helped explain the slow decay time in the pulse.

TDR Results : 1 = 1.2pF, 4 =

1.5pF, 9 = 2.3pF, 16 = 2.8pF

Ch 1: Mark-1 (5mV/div) (2ns/div) Ch 2: Commercial (5mV/div) 9 – pixels tied together (~6mm) 1 – pixel (~6mm) 2.29kV 1.67kV

Ch 1: Commercial (5mV/div) (2ns/div) Ch 2: Mark-1 (5mV/div) 256 – pixels tied together (~26mm) 1 – pixel (~6mm) 2.29kV 1.67kV

Histogram using the 408nm laser set at ~50pe.

Histogram using the 635nm laser set at ~55pe.

The “Mark-1” in the clamp with a new collection scheme.

The anode side of the “Mark-1”.

The commercial tube with a small board tying 4 anodes together.

All of the anodes on the “Mark-P” are grounded except for the one being read out.

Ch 1: Yellow Ch 2: Blue

Ch 1: Yellow Ch 2: Blue