Developments in Process Gas Analysis
Robert Wright, Thermo ONIX
APACT 03 Page 2 of 25
Introduction – Developments in Gas Analysis
• Sampling from variable pressure – VOD process analysis
• Sampling using low flow – respiratory gas analysis with mammalian cell cultures
• Sampling from liquids – for dissolved toxic gases
• Integrating several techniques – to provide complete solution
APACT 03 Page 3 of 25
Sampling from Variable Pressure
VOD process analysis
sample pressure varies 1000 mbar – 0.3mbar
APACT 03 Page 4 of 25
VOD/RH Process Gas Analysis
APACT 03 Page 5 of 25
How a Mass Spectrometer Works
Sample gas
Inlet (filtering and pressure adjustment)
Ion Source
Mass Filter
Detector
MS Controlle
rVacuum
Chamber
APACT 03 Page 6 of 25
VG Prima Process Mass Spectrometer
APACT 03 Page 7 of 25
VG Prima VOD/RH Inlet
APACT 03 Page 8 of 25
RH Process
APACT 03 Page 9 of 25
Hit Ratio
APACT 03 Page 10 of 25
Sampling using Low Flow
Respiratory gas analysis with mammalian cell cultures or anaerobic microbial cultures
APACT 03 Page 11 of 25
Mammalian Cell Cultures & Anaerobic Microbial Cultures
Low Sparge Flow In
Low Sample Flow Out
Animal cells are very ‘shear-sensitive’ and cannot tolerate high impeller speeds or large gas bubbles in the growth media.
Anaerobic processes may require very low flow or zero sparge gas.
APACT 03 Page 12 of 25
Capillary Inlet With Molecular Leak and Bypass
APACT 03 Page 13 of 25
Very Low Flow Modification
APACT 03 Page 14 of 25
Sampling from Liquids
For environmental monitoring of organic chlorides in process water streams
APACT 03 Page 15 of 25
Membrane Inlet - 100x Enhancement of VOCs - for Low Detection
APACT 03 Page 16 of 25
Supported Capillary Membrane Probe
Probe is designed to work in a process line monitoring trace organics
About 2m of silicone rubber tube is in direct contact with process water
Organics permeate through the silicone
Carrier gas in
Carrier gas out
5” (127mm)
Hollow –Fiber Pervaporation Membrane
Dow Chemical US Patent 5,317,932 Licensed to FIA Solutions
APACT 03 Page 17 of 25
Installation Arrangement
Air flow 20 ml/min at 10 psi
4m of 1/8” stainless steel tube, trace heated to 60OC
Cooling water out
Cooling water in
VG RMS Rapid Multistream
Sampler
RMSExhaust
APACT 03 Page 18 of 25
Typical Multi-Probe Installation
Sample Water In
Sample Water Out
Carrier Gas In
Carrier Gas Out
APACT 03 Page 19 of 25
EDC in Water
APACT 03 Page 20 of 25
Analysis of Impurities in Beverage Grade CO2
Measurements of the following:-
• Total Sulphur
• Total Hydrocarbons
• Total Aromatics
• Moisture
APACT 03 Page 21 of 25
On-line Analyser with PUVF, NDIR and FID detectors
APACT 03 Page 22 of 25
Multiple Detector Results
APACT 03 Page 23 of 25
Conclusions
New applications in Process Analysis
demand special adaptations of existing techniques
Examples shown include:-
• Variable pressure VOD process
• Low sample flows in certain fermentations
• Sampling of organic pollutants in water
• Analysis of impurities in CO2
APACT 03 Page 24 of 25
Acknowledgements
Graham Josland
Stuart Mackenzie
Mark Wilson
APACT 03 Page 25 of 25
Thank you for listening!
Any Questions?