Developments in Process Gas Analysis

Robert Wright, Thermo ONIX

Page 2: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 2 of 25

Introduction – Developments in Gas Analysis

• Sampling from variable pressure – VOD process analysis

• Sampling using low flow – respiratory gas analysis with mammalian cell cultures

• Sampling from liquids – for dissolved toxic gases

• Integrating several techniques – to provide complete solution

Page 3: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 3 of 25

Sampling from Variable Pressure

VOD process analysis

sample pressure varies 1000 mbar – 0.3mbar

Page 4: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 4 of 25

VOD/RH Process Gas Analysis

Page 5: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 5 of 25

How a Mass Spectrometer Works

Sample gas

Inlet (filtering and pressure adjustment)

Ion Source

Mass Filter

Detector

MS Controlle

rVacuum

Chamber

Page 6: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 6 of 25

VG Prima Process Mass Spectrometer

Page 7: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 7 of 25

VG Prima VOD/RH Inlet

Page 8: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 8 of 25

RH Process

Page 9: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 9 of 25

Hit Ratio

Page 10: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 10 of 25

Sampling using Low Flow

Respiratory gas analysis with mammalian cell cultures or anaerobic microbial cultures

Page 11: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 11 of 25

Mammalian Cell Cultures & Anaerobic Microbial Cultures

Low Sparge Flow In

Low Sample Flow Out

Animal cells are very ‘shear-sensitive’ and cannot tolerate high impeller speeds or large gas bubbles in the growth media.

Anaerobic processes may require very low flow or zero sparge gas.

Page 12: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 12 of 25

Capillary Inlet With Molecular Leak and Bypass

Page 13: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 13 of 25

Very Low Flow Modification

Page 14: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 14 of 25

Sampling from Liquids

For environmental monitoring of organic chlorides in process water streams

Page 15: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 15 of 25

Membrane Inlet - 100x Enhancement of VOCs - for Low Detection

Page 16: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 16 of 25

Supported Capillary Membrane Probe

Probe is designed to work in a process line monitoring trace organics

About 2m of silicone rubber tube is in direct contact with process water

Organics permeate through the silicone

Carrier gas in

Carrier gas out

5” (127mm)

Hollow –Fiber Pervaporation Membrane

Dow Chemical US Patent 5,317,932 Licensed to FIA Solutions

Page 17: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 17 of 25

Installation Arrangement

Air flow 20 ml/min at 10 psi

4m of 1/8” stainless steel tube, trace heated to 60OC

Cooling water out

Cooling water in

VG RMS Rapid Multistream

Sampler

RMSExhaust

Page 18: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 18 of 25

Typical Multi-Probe Installation

Sample Water In

Sample Water Out

Carrier Gas In

Carrier Gas Out

Page 19: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 19 of 25

EDC in Water

Page 20: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 20 of 25

Analysis of Impurities in Beverage Grade CO2

Measurements of the following:-

• Total Sulphur

• Total Hydrocarbons

• Total Aromatics

• Moisture

Page 21: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 21 of 25

On-line Analyser with PUVF, NDIR and FID detectors

Page 22: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 22 of 25

Multiple Detector Results

Page 23: Developments in Process Gas Analysis Robert Wright, Thermo ONIX

APACT 03 Page 23 of 25

Conclusions

New applications in Process Analysis

demand special adaptations of existing techniques

Examples shown include:-

• Variable pressure VOD process

• Low sample flows in certain fermentations

• Sampling of organic pollutants in water

• Analysis of impurities in CO2