Upload
doannhi
View
220
Download
0
Embed Size (px)
Citation preview
1. L. Rogule. Bachelor work “ Optiskā metode ādas rekapilarizācijas laika noteikšanai” , Riga, 73p
(2011).
2. E. Kviesis-Kipge, E. Čurkste, J. Spigulis, L. Eihvalde. Real-time analysis of skin capillary-refill
processes using blue LED. Proc. SPIE, Vol.7715, 7715231-5 (2010).
REFERENCES
This prototype device was made for contact (in vitro) use on human
to measure capillary blood refill time. The main idea is to make pressure on
finger tip or nail (it’s not excluded to use other human body parts) with this
device moving tip part, then to take pressure off and after some time to
finish measurement. The pressure value is displayed during measurement.
At the end of measurement user gets capillary refill time value. This time
could be used to detect dehydration or hypothermia in tissues.
Abstract
Financial supports from European Social Fund, project
#2009/0211/1DP/1.1.1.2.0/09/APIA/VIAA/077 is highly appreciated.
ACKNOWLEDGEMENTS
CONCLUSIONS
Blue polarized light (450 nm) is used to irradiate skin, then
backwards coming light signal is detected through orthogonally orientated
polarizer (signal from the skin capillaries layer comes depolarized to the
photodiode, but still polarized light from the skin surface is filtered). Blood
amount change in capillaries is the reason for more or less absorbed light on
tissue, so without pressure finger tip is red and blue light absorption is high,
but when blood is out (under pressure), finger becomes white and blue light
absorption is low (but reflection high). The signal intensity changes are
evaluated before and after applied pressure. So it’s possible to measure, how
fast blood returns into capillaries.
How it works
Device construction
The device consists of mechanical and electronic parts: body,
transparent moving tip with blue LEDs, photodiode and amplification
circuit, pressure sensor that is connected to the tip with damping element,
and main signal processing electronic circuit with LCD display, battery
and buttons.
Figure 1. Blue light penetrates in capillaries depth of skin,
but not deeper, so there isn’t any needless information
about bigger blood vessels.
Figure 2. Device for capillary blood refill time
assessment.
Capillary blood refill time curve is showed in fig.4., as we see it in
computer DataScope program (device is absolutely autonomic, but we use
DataScope for checking the signal shape). The yellow curve is the signal,
but red is the derivation of yellow signal. Red signal 2 keen pikes are found
start and finish points of refill time process. Signal is strongly filtered to
reduce the pulse noise.
Signal curve
Fig.4 shows light intensity (yellow curve) changes during the time
at measurement process. It is allowed to prepare finger for measurement
before point one, then at point 1 device gives sound signal to make
pressure on finger, at point 2 pressure is appied and is kept for few seconds
until next sound signal at point 3. After taking pressure away at point 3, the
blood refill begins with blood overshoot utill point 4. Then signal starts to
normalize. When point 4 is found, capillary blood refill time is showed on
display and measurement is finished.
Fig.4. Signal of capillary blood refill time in DataScope.
1,8 1,9
1,3 1,2
1,6
2,7
1,9
2,3 2,2
1,3
2,0 2,0 2,1
3,5
4,1 4,0
3,5 3,3
4,3
2,9
0,0
1,0
2,0
3,0
4,0
5,0
6,0
7,0
Vecums
tim
e, s
2 3 3 3 3 5 5 5 5 12 14 14 15 3 6 10 11 13 15 17 age,years
Patients without dehydr. symptoms
Snmid ≈ 0,5s
Patients with dehydr. symptoms Snmid ≈ 0,9s
Measurement results
Clinical measurements were made for kids with and without
dehydratation symptoms (2 fingers of both hands were used for
measurements) (fig.5). Most kids from first group show refill time under 2
s, but most from second group above 3 s. There are also some kids from
first group in risk category (blue columns). High standart deviations could
mean not precise measurements because of moving patient or really
different refill time at various human fingers. Usually, cold fingers gives
higher refill time values, so it’s recommended to measure warm fingers at
room temperature.
Figure 3. Capillary blood refill time measurement
process.
Fig.5. Capillary blood refill time of patients with and without dehydratation symptoms.
Working capillary blood refill time device was made and clinically
tested. Measurement results show that patients with water lack in body
have higher refill time than patients without such problems.
.
Device for Measurements of Capillary Blood Refill Time
E.Laksa, J. Zaharans, J. Spigulis and L. Eihvalde
University of Latvia, Institute of Atomic Physics and Spectroscopy, Riga, Latvia