12
Different Detectors for T ime C orrelated S ingle P hoton C ounting G. Rehm DEELS workshop ESRF, 12-13 May 2014

Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

  • Upload
    others

  • View
    11

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

Different Detectors for Time Correlated Single Photon Counting

G. Rehm

DEELS workshop

ESRF, 12-13 May 2014

Page 2: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

Overview

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 2

• So far, we’ve been using TCSPC to measure fill pattern (precise charge in filled bunches) more than single bunch purity.

• More beamlines are starting to use timing and hybrid bunch, anticipate better bunch purity is required

• We are using a PicoHarp 300 with 65536 buckets of 32ps width, 65536 counts depth

• We have acquired various single photon detectors over the years – Hamamatsu MCP-PMT R3809U-50 – id-quantique SPAD id-100-50 – PicoQuant Hybrid PMA – X-ray APD: Hamamatsu S2381 with Picosecond Pulse Labs

5530B bias tee and Femto HSA-X-1-40

Page 3: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

MCP-PMT IRF short scale

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 3

-1 -0.5 0 0.5 1 1.510

0

101

102

103

104

105

106

107

time [ns]

counts

Datasheet Our measurement with pulsed laser

Page 4: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

MCP-PMT IRF long scale

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 4

0 50 100 150 200 250 30010

0

101

102

103

104

105

106

107

108

time [ns]

counts

Delayed pulses, physics not understood

Afterpulsing Counting deadtime 88ns

Page 5: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

SPAD v MCP-PMT

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 5

Delayed pulses: ‘diffusion tail’ from red photons

Afterpulsing

Page 6: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

SPAD white v blue

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 6

Diffusion tail can be suppressed. No impact on afterpulsing.

Page 7: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

SPAD v PMA

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 7

Afterpulsing after deadtime

Page 8: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

SPAD v PMA v X-ray APD

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 8

Page 9: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

Detector Comparison

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 9

Detector IRF FW 10e-1 [ps]

IRF FW 10e-5 [ps]

Dark Counts [1/s]

After-pulsing [%]

Cost (Year)

Remarks

MCP-PMT R3809U-50

100 1250 1 0.5 GBP12000 (2005)

External HV

SPAD id-100-50

170 19000 white 800 blue?

60 1.2 CHF 4200 (2010)

Fully integrated, TTL pulse out, 5V supply

Hybrid PMA

150 1100 24 0.05 €10000 (2013)

Peltier cooled, 12V supply

X-ray APD S2381

950 2500 0! 0? GBP55+700+1550 (2013)

120V supply

Page 10: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

Injected v ‘Purified’ Single Bunch

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 10

Page 11: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

Pile up correction

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 11

Page 12: Different Detectors for Time Correlated Single Photon Counting · Time Correlated Single Photon Counting G. Rehm DEELS workshop ESRF, 12-13 May 2014 . Overview DEELS, 12-13 May 2014,

Summary/Open Issues

DEELS, 12-13 May 2014, Different Detectors for TCSPC, G.Rehm 12

• Hybrid PMA best visible photon detector, but X-ray photon APD is best of all (no news).

• Visible light also easy at lowest beam currents, just take out ND filters.

• For X-ray APD, currently, need to chose threshold for counting high, otherwise dark counts. Then only 0.1 counts per turn at full current. Sees only scatter from Al window.