Digital Radiography

Page 2: Digital Radiography

• Computed radiography, CR

• Direct radiography or direct digital radiography, DR

• Digital subtraction angiography, DSA

Page 3: Digital Radiography

Computed Radiography (CR)

is the generic term applied to an imaging system comprised of:

★Photostimulable Storage Phosphor

to acquire the x-ray projection image

★ CR Reader

to extract the electronic latent image

★ Digital electronics

to convert the signals to digital form

Page 4: Digital Radiography

• Computed Radiography (CR) uses very similar equipment to conventional radiography except that in place of a film to create the image, an imaging plate (IP) is used. The imaging plate is placed under the object to be examined and the x-ray exposure is made.

Page 5: Digital Radiography

X-rays

Conventional radiographyScreen-film combination

IP only

Cassette

Page 6: Digital Radiography

CR Detector

• Photostimulable Storage Phosphor (PSP)

• BaFBr compound, Eu activated

Phosphor Plate Or imaging plate”IP” Cassette Holder

Barium fluorin bromide Eu: europium

Page 7: Digital Radiography

• instead of taking an exposed film into a darkroom for developing in chemical tanks or an automatic film processor, the imaging plate is run through a special laser scanner to read and digitize the image.

Page 8: Digital Radiography
Page 9: Digital Radiography

Computed Radiography “reader”

Plate stacker

Information panel

Page 10: Digital Radiography
Page 11: Digital Radiography
Page 12: Digital Radiography
Page 13: Digital Radiography
Page 14: Digital Radiography
Page 15: Digital Radiography
Page 16: Digital Radiography

• The CR imaging plate contains photostimulable storage phosphors, which store the radiation level received at each point in local electron energies. When the plate is put through the scanner, the scanning laser beam causes the electrons to relax to lower energy levels, emitting light that is detected by a photo-multiplier tube (PMT), which is then converted to an electronic signal.

Page 17: Digital Radiography

photo-multiplier tube

Page 18: Digital Radiography
Page 19: Digital Radiography
Page 20: Digital Radiography

• The electronic signal is then converted to discrete (digital) values and placed into the image processor pixel map.

• Imaging plates can theoretically be re-used thousands of times if they are handled carefully, although handling under industrial conditions may result in damage after a few hundred uses. An image can be erased by simply exposing the plate to a room-level fluorescent light.

Page 21: Digital Radiography
Page 22: Digital Radiography

POST-PROCESSING

Page 23: Digital Radiography
Page 24: Digital Radiography

• The digital image can then be viewed and enhanced using software that has functions very similar to other conventional digital image-processing software, such as contrast, brightness, filtration and zoom.

Page 25: Digital Radiography
Page 26: Digital Radiography
Page 27: Digital Radiography
Page 28: Digital Radiography
Page 29: Digital Radiography
Page 30: Digital Radiography
Page 31: Digital Radiography

Change the contrast

Curve 1

Curve 2

Page 32: Digital Radiography
Page 33: Digital Radiography
Page 34: Digital Radiography
Page 35: Digital Radiography

GA GT GC GS RN RT RE

影像A 1.2 E 1.6 -0.14 4 R 0.5

影像B 0.9 A 1.6 -0.42 4 R 4.0

Page 36: Digital Radiography

GA GT GC GS RN RT RE

影像A 1.2 D 0.7 0.0 3 Q 1.0

影像B 1.1 A 0.8 -0.01 3 Q 5.0

Page 37: Digital Radiography

GA GT GC GS RN RT RE

影像A 1.0 F 0.7 0.38 4 R 0.5

影像B 1.0 A 0.7 0.10 4 R 5.0

Page 38: Digital Radiography

GA GT GC GS RN RT RE

影像A 1.4 0 1.6 0.7 4 Q 1.1

影像B 0.9 A 0.8 0.0 4 Q 5.0

Page 39: Digital Radiography

GA GT GC GS RN RT RE

影像A 0.6 0 0.6 0.63 5 T 0.5

影像B 0.9 A 0.6 0.20 5 T 5.0

Page 40: Digital Radiography

GA GT GC GS RN RT RE

影像A 0.9 E 1.6 -0.16 4 R 0.5

影像B 0.7 A 1.5 -0.58 4 R 5.0

影像C -1.3 A 1.5 0.66 4 R 5.0

Page 41: Digital Radiography

Subtraction

Page 42: Digital Radiography
Page 43: Digital Radiography
Page 44: Digital Radiography
Page 45: Digital Radiography
Page 46: Digital Radiography
Page 47: Digital Radiography
Page 48: Digital Radiography
Page 49: Digital Radiography
Page 50: Digital Radiography
Page 51: Digital Radiography

CR Networking

• DICOM

– Digital Imaging Communications in Medicine

– Provides open architecture solutions for modality

interfaces, storage/retrieval, and print functions• Technologist QC Workstation• Modality Worklist Input• Processed image output

Page 52: Digital Radiography

Medical applications

Page 53: Digital Radiography

• Computed Radiography systems are the most common in medical applications because of their low cost.

• CR IP's can be used in multiple X-ray sites and the IP's processes through a single reader (scanner), similar to a film processor.

Page 54: Digital Radiography

• No silver based film or chemicals; instead a computer is used.

• Exposure times are a fraction of typical film exposures, which reduces occupational exposure to radiographers in field applications and is more economical.

• Image acquisition is much faster; commonly one minute instead of seven minutes, as with conventional radiography.

Page 55: Digital Radiography

• By adjusting image brightness and/or contrast, a wide range of thicknesses may be examined in one exposure, unlike conventional film based radiography, which may require a different exposure or multiple film speeds in one exposure to cover wide thickness range in a component. Computed radiography often requires fewer re-shots due to under- or over-exposure.

• Images can be enhanced digitally to aid in interpretation.

Page 56: Digital Radiography

• Images can be stored on disk or transmitted for off-site review.

Page 57: Digital Radiography

Disadvantages

• CR is still not an approved method for most higher quality radiological applications (aerospace), due to the possibility of digital manipulation to the captured image, the inherent geometric unsharpness and resultant lower spatial resolution as compared to film (radiographic) images, SNR (signal vs. noise) issues and sensitivity to scattered radiation

Page 58: Digital Radiography

• Imaging plates (IP's) are quite expensive and can be easily damaged, if the system being used requires manual handling of the IP's. Theoretically, IP's may be reused thousands of times, but constant use will always result in damage to the IP and image artifacts.

Page 59: Digital Radiography

CR Trends

• Lower system costs• Smaller footprint• High throughput systems• Low throughput systems “Table-top” units• Integrated QC workstations for images• DICOM output

Page 60: Digital Radiography