2
Date Revised: 2018.11.06 © 2018 PDC Facilies, Inc. All rights reserved. For informaonal purposes only. Content is subject to change. 700 Walnut Ridge Drive Hartland, WI 53029 Phone: 262-367-7700 Web: www.pdcbiz.com Email: [email protected] Performance Summary Inial Delivered Lumens: 2900lm Input Power: 24 Was CCT: 3000K - 4000K Limited Warranty: 3 years Lifeme: Designed to last 50,000 hours Dimming: Linear 1-07-000158 - 30K Ethernet Wiring 1-07-000159 - 30K DC Wiring 1-07-000160 - 35K Ethernet Wiring 1-07-000161 - 35K DC Wiring 1-07-000162 - 40K Ethernet Wiring 1-07-000163 - 40K DC Wiring DL6 Recessed Luminaire MRI Compatible Mechanical Minimum Typical Maximum Units Package - Self-Contained - ANSI Std. Electrical Connecon - Juncon Box - ANSI Std. Fixture Weight - < 2 - lbs Electrical Minimum Typical Maximum Units Input Voltage - 34 - Vdc Input Current - 700 - mA Input Power - 24 - W BTU - 81 - BTU/hr Compliance Minimum Typical Maximum Units IP - 20 - Dry Locaon Regulatory - Listed - UL 2108 CSA 22.2#250.0 RoHS Pending 2011/65/EU Environmental Minimum Typical Maximum Units Storage Temperature -40 25 85 C Operang Temperature -20 25 35 C Humidity 5 - 95 % Lifespan 50000 - - Hours Warranty 3 - - Years Installaon Notes A PDC RF Filter is required for proper operaon in the MRI suite A PDC Power Supply is required for proper operaon in the MRI suite *PWM is known to cause image arfacts, audible noise and light flicker Product Description The DL6 Recessed Luminaire is designed for MRI suite applicaons and has a non-ferrous assembly which can be mounted within the MRI room. The DL6 is available in color temperatures of 30, 35 and 40K and comes equipped with either DC or Ethernet wiring. The DL6 is supplied with a juncon box and requires no addional housing or can for low-profile recessed applicaons. Contains no on-board electronics therefore will not interfere with MRI imaging. Cerfied to meet MRI vendor EMI requirements. MRI applicaons require EMI free emissions for specific frequency bands. All drive circuitry employs linear regulaon and no pulse width modulaon (PWM*). 24.0” 3.9”

DL6 Recessed Luminaire MRI Compatible - pdcbiz.comThe DL6 Recessed Luminaire is designed for MRI suite applications and has a non-ferrous assembly which can be mounted within the MRI

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: DL6 Recessed Luminaire MRI Compatible - pdcbiz.comThe DL6 Recessed Luminaire is designed for MRI suite applications and has a non-ferrous assembly which can be mounted within the MRI

Date Revised: 2018.11.06

© 2018 PDC Facilities, Inc. All rights reserved. For informational purposes only. Content is subject to change.

700 Walnut Ridge DriveHartland, WI 53029Phone: 262-367-7700Web: www.pdcbiz.comEmail: [email protected]

Performance SummaryInitial Delivered Lumens: 2900lm

Input Power: 24 Watts

CCT: 3000K - 4000K

Limited Warranty: 3 years

Lifetime: Designed to last 50,000 hours

Dimming: Linear

1-07-000158 - 30K Ethernet Wiring1-07-000159 - 30K DC Wiring1-07-000160 - 35K Ethernet Wiring1-07-000161 - 35K DC Wiring1-07-000162 - 40K Ethernet Wiring1-07-000163 - 40K DC Wiring

DL6 Recessed Luminaire MRI Compatible

MechanicalMinimum Typical Maximum Units

Package - Self-Contained - ANSI Std.Electrical Connection - Junction Box - ANSI Std.Fixture Weight - < 2 - lbs

ElectricalMinimum Typical Maximum Units

Input Voltage - 34 - VdcInput Current - 700 - mAInput Power - 24 - WBTU - 81 - BTU/hr

ComplianceMinimum Typical Maximum Units

IP - 20 - Dry Location

Regulatory - Listed - UL 2108CSA 22.2#250.0

RoHS Pending 2011/65/EU

EnvironmentalMinimum Typical Maximum Units

Storage Temperature -40 25 85 COperating Temperature -20 25 35 CHumidity 5 - 95 %Lifespan 50000 - - HoursWarranty 3 - - Years

Installation NotesA PDC RF Filter is required for proper operation in the MRI suiteA PDC Power Supply is required for proper operation in the MRI suite*PWM is known to cause image artifacts, audible noise and light flicker

Product DescriptionThe DL6 Recessed Luminaire is designed for MRI suite applications and has a non-ferrous assembly which can be mounted within the MRI room. The DL6 is available in color temperatures of 30, 35 and 40K and comes equipped with either DC or Ethernet wiring. The DL6 is supplied with a junction box and requires no additional housing or can for low-profile recessed applications. Contains no on-board electronics therefore will not interfere with MRI imaging. Certified to meet MRI vendor EMI requirements. MRI applications require EMI free emissions for specific frequency bands. All drive circuitry employs linear regulation and no pulse width modulation (PWM*).

24.0”

10.0”

3.9”

11.0”

Page 2: DL6 Recessed Luminaire MRI Compatible - pdcbiz.comThe DL6 Recessed Luminaire is designed for MRI suite applications and has a non-ferrous assembly which can be mounted within the MRI

Date Revised: 2018.11.06

© 2018 PDC Facilities, Inc. All rights reserved. For informational purposes only. Content is subject to change.

700 Walnut Ridge DriveHartland, WI 53029Phone: 262-367-7700Web: www.pdcbiz.comEmail: [email protected]

0

10

20

30

40

50

60

400 450 500 550 600 650 700

mW

/nm

Wavelength

Photometric Data and IES Distribution

PhotometricMinimum Typical Maximum Units

Luminous Flux - 2900 - lmCCT - 3000 - kCRI 80 - - RaChromaticity (u’) - 0.248 -Chromaticity (v’) - 0.522 -Efficacy - 119.9 - lm/W*3500K and 4000K options available

Spectral Data Over Visible Wavelength Polar Plot

Mechanical Dimensions

24.0”

10.0”

3.9”

11.0”