DMTA-10006-01EN [U8778382] Revision A March 2011
EPOCH 600Users Manual
Olympus NDT, 48 Woerd Avenue, Waltham, MA 02453, USA
2011 Olympus NDT, Inc. All rights reserved. No part of this publication may be reproduced, translated or distributed without the express written permission of Olympus NDT, Inc.
This document was prepared with particular attention to usage to ensure the accuracy of the information contained therein. It corresponds to the version of the product manufactured prior to the date appearing on the title page. There could, however, be some differences between the manual and the product if the product has been modified thereafter.
The information contained in this document is subject to change without notice.Part number: DMTA-10006-01EN [U8778382]Revision AMarch 2011
Printed in the United States of America
All brands are trademarks or registered trademarks of their respective owners and third party entities.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
Table of Contents
List of A
Labels an
ImportanIntendInstrucInstrumRepairSafetySafetyNotes SafetyWarninWEEEChina EMC DWarraTechni
IntroducProduEPOCAboutAudieTypog
1. Instru1.1 Table of Contents iii
bbreviations ....................................................................................... xi
d Symbols ........................................................................................... 1
t Information Please Read Before Use ..................................... 5ed Use .......................................................................................................................... 5tion Manual ................................................................................................................ 5ent Compatibility ..................................................................................................... 5
and Modification ....................................................................................................... 6 Symbols ....................................................................................................................... 6 Signal Words ............................................................................................................... 7Signal Words ............................................................................................................... 8 ....................................................................................................................................... 8
gs ................................................................................................................................. 8 Directive ...................................................................................................................... 9RoHS .......................................................................................................................... 10irective Compliance ................................................................................................ 10
nty Information ......................................................................................................... 11cal Support ................................................................................................................ 11
tion ...................................................................................................... 13ct Description ............................................................................................................ 13H 600 ........................................................................................................................... 14 This Document ......................................................................................................... 14nce ............................................................................................................................... 14raphical Conventions ............................................................................................... 15
ment Overview ................................................................................ 17Whats In the Box ...................................................................................................... 17
DMTA-10006-01EN [U8778382], Rev. A, March 2011
iv Table o
1.2 EPOCH 600 Instrument Configurations ................................................................ 181.3 Connectors ................................................................................................................. 191.4 Power Requirements ................................................................................................. 22
1.4.1 Lithium-Ion Battery ....................................................................................... 221.4.2 AC Charger/Adaptor ..................................................................................... 231.4.3 Alkaline Batteries ........................................................................................... 26
1.5 Installing the MicroSD Card .................................................................................... 28
2. Basic2.1
2.12.12.12.12.12.1
2.2 2.22.22.22.2
2.3 2.32.32.32.3
2.4 2.42.4
2.5 2.52.5
3. EPOC3.1 3.2
3.23.23.23.23.2f Contents
Operation .......................................................................................... 31User Interface ............................................................................................................. 31.1 Menus and Parameters .................................................................................. 32.2 Parameter Adjustment Knob Configuration ........................................... 33.3 Parameter Adjustment Navigation Pad Configuration ......................... 34.4 Direct-Access Keys ......................................................................................... 34.5 Special Functions ............................................................................................ 36.6 Submenus ........................................................................................................ 37
Pulser and Receiver Settings ................................................................................... 37.1 Sensitivity ........................................................................................................ 37.2 Reference Gain ................................................................................................ 38.3 Pulser ............................................................................................................... 38.4 Receiver ............................................................................................................ 39
Gates ............................................................................................................................ 40.1 Quickly Adjusting Basic Gate Parameters .................................................. 40.2 Gate 1 and Gate 2 ........................................................................................... 40.3 Gate Setup ....................................................................................................... 41.4 Alarm Indications ........................................................................................... 42
Calibration .................................................................................................................. 43.1 Measurement calibration .............................................................................. 43.2 Angle Beam Calibration ................................................................................ 47
Data Logger ................................................................................................................ 47.1 Calibration Files .............................................................................................. 48.2 Other Create Functions ................................................................................. 50
H 600 Hardware Features ............................................................... 51Hardware Overview ................................................................................................. 52Front Panel User Interface ....................................................................................... 53.1 Knob Configuration ....................................................................................... 53.2 Navigation Pad Configuration ..................................................................... 54.3 General Purpose Keys ................................................................................... 55.4 Function and Parameter Keys ...................................................................... 57.5 Parameter Adjustment ................................................................................... 58
DMTA-10006-01EN [U8778382], Rev. A, March 2011
3.2.6 About the Direct-Access Keypad ................................................................. 583.2.7 About Indicators ............................................................................................ 62
3.3 Connectors ................................................................................................................. 623.3.1 Transducer Connections ................................................................................ 623.3.2 Input/Output Connectors ............................................................................. 633.3.3 Battery Compartment .................................................................................... 643.3.4 MicroSD and PC Communication ............................................................... 65
3.4 Various Hardware Features ..................................................................................... 663.43.43.4
3.5
4. Powe4.1 4.2 4.3 4.4 4.5
5. EPOC5.1
5.15.15.15.15.15.15.15.15.1
5.2 5.3
5.35.35.35.35.35.3
5.4 5.4Table of Contents v
.1 Instrument Stand ............................................................................................ 67
.2 O-Ring Gasket and Membrane Seals .......................................................... 67
.3 Display Protection .......................................................................................... 68Environmental Ratings ............................................................................................ 68
ring the EPOCH 600 ......................................................................... 69Starting the EPOCH 600 ........................................................................................... 70Using the AC Line Power ........................................................................................ 71Using Battery Power ................................................................................................. 72Charging the Battery ................................................................................................ 74Replacing the Battery ............................................................................................... 76
H 600 Software Features ................................................................ 79Software Main Display ............................................................................................. 80.1 About the Menu System ............................................................................... 80.2 About the Convention to Identify Menu Elements ................................... 82.3 About the Focus ............................................................................................. 84.4 About Button Types ....................................................................................... 85.5 About the File Identifier and Message Bars ............................................... 86.6 About Direct-Access Parameters ................................................................. 86.7 About Measurement Reading Boxes ........................................................... 87.8 About the Live-Scan Area ............................................................................. 88.9 About Flags ..................................................................................................... 89
Menu Contents .......................................................................................................... 91About Setup Pages .................................................................................................... 95.1 Display Setup Page ........................................................................................ 96.2 Reading Setup Page ....................................................................................... 98.3 General Setup Page ...................................................................................... 102.4 Status Setup Page ......................................................................................... 103.5 Software Options Page ................................................................................ 105.6 Clock Setup Page ........................................................................................ 105
Basic Procedures ..................................................................................................... 107.1 Navigating in the Menu Structure ............................................................. 107
DMTA-10006-01EN [U8778382], Rev. A, March 2011
vi Table o
5.4.2 Changing a Parameter Value ...................................................................... 1075.4.3 Navigating in a Setup Page ......................................................................... 1085.4.4 Entering an Alphanumeric Value Using the Virtual Keyboard ............ 108
5.5 Resets Menu ............................................................................................................. 1105.6 Software Diagnostic ................................................................................................ 110
6. Adjusting the Pulser/Receiver ............................................................... 1116.1 Adjusting the System Sensitivity (Gain) .............................................................. 1116.2 6.3 6.4
6.46.46.46.46.46.4
6.5 6.56.5
6.6
7. Mana7.1 7.2 7.3 7.4 7.5
8. Gates8.1 8.2 8.3 8.4 8.5 8.6 8.7
8.78.7
8.8 8.8f Contents
Using the AUTO XX% Feature .............................................................................. 112Setting Reference Gain and Scanning Gain ......................................................... 113Pulser Adjustments ................................................................................................. 114.1 Pulse Repetition Frequency (PRF) ............................................................. 114.2 Pulse Energy (Voltage) ................................................................................ 115.3 Damping ........................................................................................................ 115.4 Test Mode ...................................................................................................... 116.5 Pulser Waveform .......................................................................................... 117.6 Pulser Frequency Selection (Pulse Width) ................................................ 117
Receiver Adjustments ............................................................................................. 118.1 Digital Receiver Filters ................................................................................ 118.2 Waveform Rectification ............................................................................... 119
Custom Filter Sets ................................................................................................... 119
ging Special Waveform Functions .............................................. 121Reject ......................................................................................................................... 121Peak Memory ........................................................................................................... 122Peak Hold ................................................................................................................. 124Freeze ........................................................................................................................ 124Grid Modes .............................................................................................................. 125
........................................................................................................... 129Measurement Gates 1 and 2 .................................................................................. 130Quickly Adjusting Basic Gate Parameters ........................................................... 132Gate Measurement Modes ..................................................................................... 133Viewing Measurement Readings .......................................................................... 136Gate Tracking and Echo-to-Echo Measurements ............................................... 136Operating in Time-of-Flight Mode ....................................................................... 137Using the Zoom ....................................................................................................... 138.1 Activating the Zoom .................................................................................... 138.2 Zoom Applications ...................................................................................... 139
Gate Alarms ............................................................................................................. 139.1 Threshold Alarms ........................................................................................ 140
DMTA-10006-01EN [U8778382], Rev. A, March 2011
8.8.2 Minimum Depth Alarm .............................................................................. 1418.8.3 Minimum Depth Alarm with a Single Gate ............................................. 1418.8.4 Minimum Depth Alarm with Gate Tracking ........................................... 141
9. Input and Output Features ..................................................................... 1439.1 VGA Output ............................................................................................................ 1439.2 Analog Output ........................................................................................................ 1449.3 Serial Communication (RS-232) ............................................................................ 1469.4
9.49.4
9.5
10. Calib10.1 10.2
1010
10.3 10.4 10.5 10.6 10.7
10101010
10.8 10.9 10.10
11. Mana11.1 11.2 11.3
1111
11.4 11Table of Contents vii
USB Communication .............................................................................................. 146.1 USB Client ..................................................................................................... 146.2 USB Host ....................................................................................................... 147
Serial/USB Command Protocol ............................................................................. 147
rating the EPOCH 600 ................................................................... 149Getting Started ........................................................................................................ 150Calibration Modes .................................................................................................. 151.2.1 Straight Beam Modes ................................................................................... 151.2.2 Angle Beam Modes ...................................................................................... 152Calibrating with a Straight Beam Transducer .................................................... 152Calibrating with a Delay Line Transducer .......................................................... 157Calibrating with a Dual Element Transducer ..................................................... 163Calibrating in Echo-to-Echo Mode ....................................................................... 168Calibrating to Known Sound Path Values with an Angle Beam Transducer . 172.7.1 Locating the Beam Index Point .................................................................. 172.7.2 Verifying the Refracted Angle .................................................................... 174.7.3 Calibrating for Distance .............................................................................. 176.7.4 Calibrating for Sensitivity ........................................................................... 181Calibrating to Known Depth Values with an Angle Beam Transducer .......... 183Curved Surface Correction .................................................................................... 188Common Angle Beam Calibration Block Diagrams .......................................... 189
ging the Data Logger ..................................................................... 195Data Logger Overview ........................................................................................... 195Data Logger Storage Capacity .............................................................................. 196Data Logger Menu Functions ............................................................................... 196.3.1 File Menu ....................................................................................................... 196.3.2 Manage Menu ............................................................................................... 198Data Logger Parameter Functions ........................................................................ 199.4.1 Open Function .............................................................................................. 199
11.4.1.1 Selecting a File as the Active Storage Location. ............................ 19911.4.1.2 Viewing details about a specific file ............................................... 200
DMTA-10006-01EN [U8778382], Rev. A, March 2011
viii Table
11.4.1.3 Viewing Setup and Waveform Data for Saved IDs in a File ....... 20111.4.1.4 Recalling a File ID to Bring Saved Data to the Live Screen ......... 20311.4.1.5 Viewing a Summary of all Saved Data in a File ............................ 20411.4.1.6 Exporting File Data to the Micro SD Card ..................................... 205
11.4.2 Create Function ............................................................................................ 20611.4.2.1 File Types ............................................................................................ 20611.4.2.2 File Creation ....................................................................................... 20611.4.2.3 Saving Data to Files ........................................................................... 208
11111111
11.5
12. Softw12.1 12.2
12121212
1212
12.3 121212121212
12.4 12121212
12.5 1212 of Contents
.4.3 Quick Recall Function .................................................................................. 209
.4.4 First ID, Last ID, and Select ID Functions ................................................. 210
.4.5 Reset Function .............................................................................................. 211
.4.6 Edit, Copy, and Delete Functions ............................................................... 212Saving Screen Captures .......................................................................................... 215
are Features and Options ............................................................. 217Defining Licensed and Unlicensed Software Features ...................................... 217Dynamic DAC/TVG ................................................................................................ 219.2.1 Feature Activation and Reference Correct ................................................ 220.2.2 ASME/ASME III DAC/TVG ........................................................................ 221.2.3 ASME III DAC Setup Example .................................................................. 221.2.4 Gain Adjustment Options ........................................................................... 227
12.2.4.1 Scanning Gain .................................................................................... 22712.2.4.2 Curve Adjustment Gain (DAC Gain or TVG Gain) ..................... 22912.2.4.3 Transfer Correction ........................................................................... 230
.2.5 JIS DAC .......................................................................................................... 231
.2.6 Custom DAC Curves ................................................................................... 231DGS/AVG ................................................................................................................. 233.3.1 Option Activation and Setup ...................................................................... 234.3.2 Curve Adjustment Options ......................................................................... 239.3.3 Transfer Correction ...................................................................................... 239.3.4 DGS/AVG Curve Gain ................................................................................. 240.3.5 Registration Level Adjustment ................................................................... 241.3.6 Relative Attenuation Measurement ........................................................... 242AWS D1.1/D1.5 Weld Rating Software ................................................................ 243.4.1 Description .................................................................................................... 243.4.2 Option Activation ......................................................................................... 244.4.3 Scanning Gain ............................................................................................... 247.4.4 Calculating A and C Values ........................................................................ 248API 5UE .................................................................................................................... 249.5.1 Option Activation and Setup ...................................................................... 250.5.2 Envelope Mode ............................................................................................. 252
12.5.2.1 Calibration .......................................................................................... 252
DMTA-10006-01EN [U8778382], Rev. A, March 2011
12.5.2.2 Crack Sizing ....................................................................................... 25312.5.3 Manual Mode ............................................................................................... 255
12.5.3.1 Calibration .......................................................................................... 25512.5.3.2 Crack Sizing ....................................................................................... 258
12.6 Waveform Averaging ............................................................................................. 25912.6.1 Option Setup ................................................................................................. 26012.6.2 Using the Averaging Option ...................................................................... 260
13. Main13.1 13.2 13.3 13.4 13.5
14. Speci14.1 14.2 14.3
Appendi
Appendi
Appendi
List of Fi
List of Ta
Index ....Table of Contents ix
tenance & Troubleshooting .......................................................... 263Instrument Cleaning ............................................................................................... 263Verifying O-Ring Gaskets and Seals .................................................................... 263Protecting the Display ............................................................................................ 264Annual Calibration ................................................................................................. 264Troubleshooting ...................................................................................................... 264
fications ............................................................................................ 267General and Environmental Specifications ......................................................... 267Channel Specifications ........................................................................................... 269Input/Output Specifications .................................................................................. 271
x A: Sound Velocities .................................................................... 273
x B: Glossary ................................................................................... 275
x C: Parts List .................................................................................. 283
gures ................................................................................................. 287
bles ................................................................................................... 293
............................................................................................................. 295
DMTA-10006-01EN [U8778382], Rev. A, March 2011
x Table of Contents
DMTA-10006-01EN [U8778382], Rev. A, March 2011
List of Abbreviations
ACACTADDT
AVGAWSBIPCSCDACDCDGSEFUP
EMCFSHIDIFIIWIPList of Abbreviations xi
alternating currentamplitude comparison techniqueamplitude-distance differential techniqueabstand verstrkung greAmerican Welding Societybeam index pointcurved surface correctiondistance amplitude correctiondirect currentdistance gain sizeenvironmental friendly usage pe-riodelectromagnetic compatibilityfull-screen heightidentifierinterface gateInternational Institute of Weldingingressprotection
LCD liquid crystal displayMEM memoryNDT non-destructive testingOS overshootPK peakPRF pulse repetition frequencyR receiveRoHS restriction of hazardous substan-
cesSDH side-drilled holeT/R transmit/receiveTrig trigonometryTVG time-varied gainUSB universal serial busVAC volts alternating currentVGA video graphics adapterWEEE waste electrical and electronic
equipment
DMTA-10006-01EN [U8778382], Rev. A, March 2011
xii List of Abbreviations
DMTA-10006-01EN [U8778382], Rev. A, March 2011
Labels and Symbols
A ratinshowninstrumLabels and Symbols 1
g label with safety-related symbols is attached to the instrument at the location in the following figure. The serial number label is located at the bottom of the
ent. If labels are missing or illegible, please contact Olympus.
Location of rating label(see Table 1 on page 2)
DMTA-10006-01EN [U8778382], Rev. A, March 2011
2 Labels
Table 1 Contents of the rating and the serial number labels
Rating plate label:
Contand Symbols
ains:
The CE marking is a declaration that this product conforms to all the applicable directives of the European Community. See the Declaration of Conformity for details.
The WEEE symbol indicates that the product must not be disposed of as unsorted municipal waste, but should be collected separately.
The C-Tick label indicates that the product complies with the applicable standard and establishes a traceable link between the equipment and the manufacturer, importer or their agent responsible for compliance and for placing it on the Australian market.
The China RoHS mark indicates the products Environmental Friendly Usage Period (EFUP). The EFUP is defined as the number of years for which listed controlled substances will not leak or chemically deteriorate while in the product. The EFUP for the EPOCH 600 has been determined to be 15 years. Note: The Environmental Friendly Usage Period (EFUP) is not meant to be interpreted as the period assuring functionality and product performance.
The direct current symbol
DMTA-10006-01EN [U8778382], Rev. A, March 2011
SERIAL The serial number is a 9-digit number in the following format: yynnnnnmm
where:yy Production yearnnnnn Unit number manufactured that month.
Table 1 Contents of the rating and the serial number labels (continued)Labels and Symbols 3
mm Production monthFor example, the 100000504 serial number indicates that the fifth unit (00005) was produced in April 2010.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
4 Labels
Do not touch the inner conductor of the BNC (or LEMO) connectors to avoid risks of an electric shock. Up to 400 V can be present on the inner conductor. The warning symbol between the Transmit/Receive (T/R) and the Receive (R) BNC connectors shown in the figure below indicates this electric shock risk.
BN(LEand Symbols
Warning symbol
C connector inner conductorMO 01 series also available)
DMTA-10006-01EN [U8778382], Rev. A, March 2011
Important Information Please Read Before Use
Intended
The EPindust
Do notuse.
Instructi
This inproduand us
Keep t
Instrume
Only u
ReImportant Information Please Read Before Use 5
Use
OCH 600 instrument is designed to perform nondestructive inspections of rial and commercial materials.
use the EPOCH 600 instrument for any purpose other than its intended
on Manual
struction manual contains essential information on using this Olympus ct safely and effectively. Before use, thoroughly review this instruction manual e the product as instructed.
his instruction manual in a safe, accessible location.
nt Compatibility
se the EPOCH 600 instrument with the following ancillary equipment:
chargeable lithium-ion (Li-ion) battery pack (P/N: 600-BAT-L [U8760056])
DMTA-10006-01EN [U8778382], Rev. A, March 2011
6 Import
Optional stand-alone external battery charger (P/N: EPXT-EC-X) [varies by configuration; must select power cord]
Charger/adaptor (P/N: EP-MCA-X) [varies by configuration; must select power cord]
Using damag
Repair a
The EP
Do nothuman
Safety S
The fomanua
Thfolant Information Please Read Before Use
incompatible equipment can result in a malfunction and/or in equipment e.
nd Modification
OCH 600 instrument does not contain any user-serviceable parts.
disassemble, modify, or attempt to repair the instrument to prevent injury and/or equipment damage.
ymbols
llowing safety symbols may appear on the instrument and in the instruction l:
General warning symbol:is symbol is used to alert you to potential hazards. All safety messages that low this symbol shall be obeyed to avoid possible harm.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
High voltage warning symbol:This symbol is used to alert you to potential electric shock hazards higher than 1,000 volts. All safety messages that follow this symbol shall be obeyed to avoid possible harm.
Safety Signal Words
The fo
The DAattentiadhereDANGE
The WattentiadhereWARNI
The CAattentiperfordamagdata. Dfully uImportant Information Please Read Before Use 7
llowing safety symbols may appear in the documentation of the instrument:
NGER signal word indicates an imminently hazardous situation. It calls on to a procedure, practice, or the like, which, if not correctly performed or d to, could result in death or serious personal injury. Do not proceed beyond a R signal word until the indicated conditions are fully understood and met.
ARNING signal word indicates a potentially hazardous situation. It calls on to a procedure, practice, or the like, which, if not correctly performed or d to, could result in death or serious personal injury. Do not proceed beyond a NG signal word until the indicated conditions are fully understood and met.
UTION signal word indicates a potentially hazardous situation. It calls on to an operating procedure, practice, or the like, which, if not correctly med or adhered to, could result in minor or moderate personal injury, material e, particularly to the product, destruction of part or all of the product, or loss of o not proceed beyond a CAUTION signal word until the indicated conditions are
nderstood and met.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
8 Import
Notes Signal Words
The following safety symbols may appear in the documentation of the instrument:
The IMinform
The Nthat rethat is
The TItechniprovid
Safety
Beforehave bon theBefore
Warning
Caon
Ke Foant Information Please Read Before Use
PORTANT signal word calls attention to a note that provides important ation or information essential to the completion of a task.
OTE signal word calls attention to an operating procedure, practice, or the like quires special attention. A note also denotes related, parenthetical information useful but not imperative.
P signal word calls attention to a type of note that helps you apply the ques and procedures described in the manual to your specific needs, or that es hints on how to use effectively the capabilities of the product.
applying power to the EPOCH 600, verify that the correct safety precautions een taken (see the following warnings). In addition, note the external markings instrument that are described in the Important Information Please Read Use section.
s
General Warningsrefully read the instructions contained in the users manual prior to powering the instrument.ep the users manual in a safe place for further reference.llow the installation and operation procedures.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
It is imperative to respect the safety warnings on the instrument and in the users manual.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment could be impaired.
Do not install substitute parts or perform any unauthorized modification to the instrument.
Service instructions, when applicable, are for trained service personnel. To avoid a dFoau
Beof Thprco
Wins
Thtyp
WEEE DImportant Information Please Read Before Use 9
angerous electric shock, do not perform any service unless qualified to do so. r any problem or question regarding this apparatus, contact Olympus or an thorized Olympus representative.
fore turning on the instrument, you must connect the protective earth terminal the charger/adaptor to the protective conductor of the (mains) power cord.e mains plug shall only be inserted in a socket outlet provided with a otective earth contact. Never negate the protective action by using an extension rd (power cable) without a protective conductor (grounding).henever it is likely that the ground protection is impaired, you must turn off the trument and secure it against any unintended operation.e instrument must be connected only to a power source corresponding to the e indicated on the rating plate.
irective
In accordance with European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE), this symbol indicates that the product must not be disposed of as unsorted municipal waste, but should be collected separately. Refer to your local Olympus distributor for return and/or collection systems available in your country.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
10 Impor
China RoHS
China RoHS is the term used by industry generally to describe legislation implemented by the Ministry of Information Industry (MII) in the Peoples Republic of China for the control of pollution by electronic information products (EIP).
EMC Dire
FCC (UThis eqdigitalprovidoperatradiateinstrucOperainterfeexpens
ICES-0This C
Cet ap
The China RoHS mark indicates the products Environmental tant Information Please Read Before Use
ctive Compliance
SA) Complianceuipment has been tested and found to comply with the limits for a Class A
device, pursuant to Part 15 of the FCC Rules. These limits are designed to e reasonable protection against harmful interference when the equipment is ed in a commercial environment. This equipment generates, uses, and can radio frequency energy and, if not installed and used in accordance with the tion manual, may cause harmful interference to radio communications.
tion of this equipment in a residential area is likely to cause harmful rence in which case you will be required to correct the interference at your own e.
03 (Canada) Compliancelass A digital apparatus complies with Canadian ICES-003.
pareil numrique de la classe A est conforme la norme NMB-003 du Canada.
Friendly Usage Period (EFUP). The EFUP is defined as the number of years for which listed controlled substances will not leak or chemically deteriorate while in the product. The EFUP for the EPOCH 600 has been determined to be 15 years. Note: The Environmental Friendly Usage Period (EFUP) is not meant to be interpreted as the period assuring functionality and product performance.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
Warranty Information
Olympus guarantees Olympus products to be free from defects in materials and workmanship for a period and conditions specified in the Olympus terms and conditions, available at http://www.olympus-ims.com/en/terms/.
The Olympus warranty only covers equipment that has been used in a proper manner as described in this instruction manual and has not been subjected to excessive abuse, attemp
Inspecthat mimmedshipmorder tassista
This inprodushall nverificprocedthis reexampthat th
Olymprespon
Technica
Olympprodufails tothen, inearesImportant Information Please Read Before Use 11
ted unauthorized repair, or modification.
t materials thoroughly on reception for evidence of external or internal damage ight have occurred during shipment. Notify the carrier making the delivery iately of any damage, since the carrier is normally liable for damage in
ent. Preserve packing materials, waybills, and other shipping documentation in o establish a damage claim. After notifying the carrier, contact Olympus for nce with the damage claim and equipment replacement, if necessary.
struction manual attempts to teach the proper operation of your Olympus ct. The information contained herein is intended solely as a teaching aid and ot be used in any particular application without independent testing and/or ation by the operator or the supervisor. Such independent verification of ures becomes more important as the criticality of the application increases. For
ason, Olympus makes no warranty, expressed or implied, that the techniques, les, or procedures described herein are consistent with industry standards nor ey meet the requirements of any particular application.
us reserves the right to modify all products without incurring the sibility for modifying previously manufactured products.
l Support
us is firmly committed to providing the highest level of customer service and ct support. If you experience any difficulties when using our product, or if it operate as described in the documentation, first consult the users manual, and f you are still in need of assistance, contact our after-sales service. To locate the t service center, visit the Service Centers page at: www.olympus-ims.com.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
12 Important Information Please Read Before Use
DMTA-10006-01EN [U8778382], Rev. A, March 2011
Introduction
The pr
P E A A T
Product
The EPto detemateriflaw dfeatureresolufor sup
The EPenhanenhan
Ca(n
Cotec
Co A Introduction 13
eface contains the following topics:
roduct Description on page 13POCH 600 on page 14bout This Document on page 14udience on page 14ypographical Conventions on page 15
Description
OCH 600 is a portable ultrasonic non destructive test (NDT) instrument used ct flaw conditions in welds, pipes, and many other structural and/or industrial als. The instrument may be used in indoor and outdoor environments. This etector offers advanced conventional ultrasonic performance. The instrument s a large dynamic range, superior measurement resolution, a full VGA
tion (640 x 480 pixels) color-liquid crystal display with transflective technology erior visibility, and an intuitive user interface.
OCH 600 features many performance, durability, and operational cements when compared to previous EPOCH flaw detector instruments. These cements include:
se sealed to the requirements of IP66 (adjustment knob configuration) or IP67 avigation pad configuration) environmental ratinglor liquid crystal display (LCD) with full VGA resolution and transflective hnologympliance with EN12668-1100 % digital high dynamic range receiver design
DMTA-10006-01EN [U8778382], Rev. A, March 2011
14 Introd
8 digital receiver filters Maximum 2000 Hz pulse repetition frequency (PRF) Dynamic DAC/TVG sizing software Onboard DGS/AVG sizing software Optional analog output Digital alarm outputs USB and RS-232 connectivity Ad 2 G VG
Read tso thatthe ins
EPOCH 6
In ordavailabwith apad areither adjust
About Th
This droutin
Audienc
This dthat alultrasooperatoperatuction
justment knob or navigational arrow keysB MicroSD memory cardA Output Capability
hrough this information completely at least once with your EPOCH 600 in hand you can combine reading the descriptions and examples with the actual use of trument.
00
er to accommodate different user needs and preferences, the EPOCH 600 is le in two hardware configurations: one with an adjustment knob and the other
navigation pad. The adjustment knob and the arrow keys on the navigation e responsible for parameter adjustment and value slewing. You can choose the adjustment knob or the navigation pad when ordering depending on which ment method you are more comfortable with.
is Document
ocument is the users manual for the EPOCH 600. This manual describes e tasks for operating the EPOCH 600.
e
ocument is intended for operators of the EPOCH 600. Olympus recommends l operators have a thorough understanding of the principles and limitations of nic non destructive testing. Olympus assumes no responsibility for incorrect ional procedure or interpretation of test results. We recommend that all ors seek adequate training prior to using this equipment.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
While the EPOCH 600 is a continuously self-calibrating instrument, you must determine regulatory requirements. Olympus offers calibration and documentation services. Contact Olympus or your local representative with any special requests.
Typographical Conventions
Table 2 on page 15 presents the typographical conventions employed in this docum
Conv
Bold
[BOL
[2ND
ALL
SMAL
Italics
Introduction 15
ent.
Table 2 Typographic conventions
ention Description
Used when referring to the label of a user interface element, including menu items, buttons, toolbar names, options, and tabs.
D] Used when referring to an instrument front panel key.
F], (BOLD) Used when referring to the secondary function of an instrument front panel key. (BOLD) is the secondary function of the key, appearing above the direct-access keypad key.
UPPERCASE Used to refer to computer keyboard keys
L CAPS Used to refer to marking on the instrument for connector names for example.
Used to refer to document titles.
Used to represent a variable.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
16 Introduction
DMTA-10006-01EN [U8778382], Rev. A, March 2011
1. Instrument Overview
This chEPOC
W E C P In
1.1 W
The EPFigure
2-G[U
ACpo
Po Al In Ge EP
co EP
MInstrument Overview 17
apter provides a brief overview of all common operational requirements of the H 600 instrument. This chapter is organized in the following manner:
hats In the Box on page 17POCH 600 Instrument Configurations on page 18onnectors on page 19ower Requirements on page 22stalling the MicroSD Card on page 28
hats In the Box
OCH 600 instrument comes standard with several key accessories (see 1-1 on page 18):
B MicroSD removable memory card and adaptors (P/N: MICROSD-ADP-2GB 8779307]) charger/adaptor (P/N: EP-MCA-X). Varies by configuration; must select
wer cord.wer cordkaline battery holder (P/N: 600-BAT-AA [U8780295])strument transport case (P/N: 600-TC [U8780294])tting Started leaflet (P/N: DMTA-10008-01EN [U8778373])OCH 600 Basic Operation Manual (P/N: DMTA-10007-01EN [U8778365]) on hard pyOCH 600 Users Manual (P/N: DMTA-10006-01EN) on CD-ROM (P/N: EP600-
ANUAL-CD [U8778381])
DMTA-10006-01EN [U8778382], Rev. A, March 2011
18 Chapt
For a l
1.2 EP
The EPsevera
Basic Operation Manual
AC
Getting Started leaflet
Transp
Users Manual on CD-ROMer 1
Figure 1-1 Transport case contents
ist of optional accessories, please refer to parts list on page 283.
OCH 600 Instrument Configurations
OCH 600 instrument is configured according to the customers preferences in l ways. Available configurations include:
Lock
Alkaline battery holder
charger/adaptor
Power cord Memory card and adaptors
EPOCH 600 unit
ort case
DMTA-10006-01EN [U8778382], Rev. A, March 2011
Adjustment knob or navigation pad Keypad overlay (English, Japanese, Chinese, or Symbolic) BNC or LEMO 01 series transducer connectors Optional analog output port
Configuration selections must be made when ordering the instrument. The EPOCH 600 also comes standard with a rechargeable lithium-ion battery, an adjustable pipe stand, and a flexible display protector.
1.3 Co
FigureAC ch
Use oncord w
T
chaInstrument Overview 19
nnectors
1-2 on page 19 illustrates the connections of the EPOCH 600 with the arger/adaptor, the MicroSD card, and a PC.
Figure 1-2 The EPOCH 600 connections
ly the AC power cord supplied with the EPOCH 600. Do not use this AC power ith other products.
o poweroutlet
DC power plug
EP-MCA-Xrger/adaptor
AC power cord
To PC
MicroSD card
DMTA-10006-01EN [U8778382], Rev. A, March 2011
20 Chapt
The DC power, the Receive probe connector, and the Transmit/Receive probe connector are located on the top end of the EPOCH 600 (see Figure 1-3 on page 20).
The USthe rigpage 2
Transmit/Receive transducer connector
Receive transducer connectorer 1
Figure 1-3 The top end connectors
B On-The-Go port and the external MicroSD memory card slot are located on ht side of the instrument, hidden behind the side door (see Figure 1-4 on 1).
DC power connector
DMTA-10006-01EN [U8778382], Rev. A, March 2011
The RSinstrumeach co
MicroSD card slot
Side doorInstrument Overview 21
Figure 1-4 The connectors behind the side door
-232/Alarms and the VGA Out connectors are located at the back of the ent, in the upper section (see Figure 1-5 on page 21). A rubber cover protects
nnector.
Figure 1-5 The RS-232/Alarms and VGA Out connectors
USB port
The VGA Out connector
The RS-232/Alarms connector
DMTA-10006-01EN [U8778382], Rev. A, March 2011
22 Chapt
1.4 Power Requirements
Press the [ON/OFF] key to turn on the EPOCH 600 (see Figure 1-6 on page 22). Pressing this key once causes an initial beep, followed by the instrument startup screen and a second beep approximately 5 seconds later.
The EPOCH 600 is designed to operate using three power supply methods:
In Di In
1.4.1The litThis bwhen battery
To install o
1. Un2. At
co3. Re4. Reer 1
ternal lithium-ion batteryrectly from the EPOCH charger/adaptorternal alkaline batteries
Figure 1-6 Location of the EPOCH 600 power key and indicator
Lithium-Ion Battery hium-ion (Li-ion) battery is the primary method for powering the EPOCH 600. attery comes installed in every instrument. When properly maintained, and the instrument is operated under typical inspection conditions, the lithium-ion should provide between 12 and 13 hours of continuous operation.
r replace the lithium-ion battery
fold the instrument pipe stand. the back of the instrument, loosen the two thumb screws securing the battery mpartment cover (see Figure 1-7 on page 23).move the battery compartment cover (see Figure 1-7 on page 23).move the battery and/or install the battery in the battery compartment.
[ON/OFF]power key
AC charger/adaptor power indicator
DMTA-10006-01EN [U8778382], Rev. A, March 2011
5. Ensure that the gasket of the battery compartment cover is clean and in good condition.
6. Install the battery compartment cover at the back of the instrument, and then tighten the two thumb screws to complete the installation (see Figure 1-7 on page 23).
1.4.2The EPchargeinstallthe insstatus
To connect
1. CoouInstrument Overview 23
Figure 1-7 Removing the lithium-ion battery
AC Charger/Adaptor OCH 600 AC charger/adaptor is provided with every instrument. This r/adaptor allows you to power the EPOCH 600 with or without a battery ed, as well as to charge the lithium-ion rechargeable battery when installed in trument. A power indicator on the front panel of the unit displays the current of the AC charger/adaptor.
the AC charger/adaptor
nnect the power cord to the charger/adaptor unit and to an appropriate power tlet.
Battery compartment cover
Thumb screws (2)
Battery
Pipe stand
DMTA-10006-01EN [U8778382], Rev. A, March 2011
24 Chapt
Use only the AC power cord supplied with the EPOCH 600. Do not use this AC power cord with other products.
2. LiEP
3. Cocoer 1
Figure 1-8 Connecting the charger/adaptor
ft the rubber seal that covers the AC adaptor connector on top of the OCH 600 instrument.nnect the DC output power cable from the AC charger/adaptor to the adaptor
nnector (see Figure 1-9 on page 25).
To power outlet
DC power plug
EP-MCA-X
AC power cord
DMTA-10006-01EN [U8778382], Rev. A, March 2011
The poindica(see Ta
Pind
s
G
DC power plug
DC adaptor connector (rubber seal not shown)Instrument Overview 25
Figure 1-9 Connecting the DC power plug
wer status of the AC charger/adaptor and the battery charge condition are ted both on the front panel of the EPOCH 600, as well as in the user interface ble 3 on page 25).
Table 3 AC charger/adaptor power indicator status
ower icator
tatus
AC line power connected Indicator meaning
Battery indicator
reen Yes Internal battery fully charged
Red Yes Internal battery is charging
Off No AC charger/adaptor not connected
DMTA-10006-01EN [U8778382], Rev. A, March 2011
26 Chapt
1.4.3 AThe EP[U8780wheredischawill pr
To install th
1. Un2. Lo
of Fig
3. Re
Green Yes AC charger/adaptor connectedNo battery installed
Table 3 AC charger/adaptor power indicator status (continued)
Power indicator
status
AC line power connected Indicator meaning
Battery indicatorer 1
lkaline BatteriesOCH 600 comes standard with an alkaline battery holder (P/N: 600-BAT-AA 295]). This holder accommodates 8 AA size alkaline batteries in situations
an AC power source is not available and the internal Li-ion battery is rged. When operated under typical inspection conditions, the alkaline batteries ovide 3+ hours of continuous operation.
e alkaline battery holder
fold the instrument pipe stand.osen the two thumb screws securing the battery compartment cover at the back the instrument, and then remove the battery compartment cover (see ure 1-10 on page 27).
move the lithium-ion battery (if installed) [see Figure 1-10 on page 27].
DMTA-10006-01EN [U8778382], Rev. A, March 2011
Figur
4. In5. Co6. Po
on
Thumb screws (2)
BattInstrument Overview 27
e 1-10 Removing the battery compartment cover and the lithium-ion battery
stall 8 AA size alkaline batteries into the alkaline battery holder.nnect the alkaline battery holder connector into the instrument.sition the alkaline battery holder in the battery compartment (see Figure 1-11 page 27).
Figure 1-11 The alkaline battery holder
BatteryBattery compartment cover
Battery holder with alkaline batteries (8)
ery holderconnector
DMTA-10006-01EN [U8778382], Rev. A, March 2011
28 Chapt
7. Install the battery compartment cover at the back of the instrument, and then tighten the two thumb screws.
When alkaline batteries are installed in the instrument, the battery indicator in the user interface displays ALK. The AC charger/adaptor does not recharge the batteries installed in the alkaline battery holder.
1.5 In
The 2-
To install th
1. Re2. Lo
Fig
Thumer 1
stalling the MicroSD Card
GB MicroSD card is included with every EPOCH 600 shipment.
e MicroSD removable memory card
move the card from its packaging.osen the two thumb screws, and then open the EPOCH 600 side door (see ure 1-12 on page 28).
Figure 1-12 The side door
USB port
b screws(2)
MicroSD card slot
DMTA-10006-01EN [U8778382], Rev. A, March 2011
3. Hold the card so that the MicroSD label faces toward the rear of the instrument.4. Carefully slide the card into the MicroSD slot until it clicks (see Figure 1-12 on
page 28).
To remove the MicroSD card, carefully push the card into the instrument and release. A spring mechanism will partially eject the card, and then you can grasp and remove it fromInstrument Overview 29
the instrument.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
30 Chapter 1
DMTA-10006-01EN [U8778382], Rev. A, March 2011
2. Basic Operation
This chultrasoof thescoveremanne
U P G C D
2.1 Us
The EPcontrofunctioaccessautomcapabi
AdjustadjustmdetermmethoBasic Operation 31
apter is intended to quickly provide guidance for users who are familiar with nic flaw detection, but are new to the EPOCH 600. More detailed descriptions e topics, along with a comprehensive guide to more advanced functions, are d in the later chapters of this manual. This chapter is organized in the following r:
ser Interface on page 31ulser and Receiver Settings on page 37ates on page 40alibration on page 43ata Logger on page 47
er Interface
OCH 600 uses a blend of direct-access keys and software menus to achieve full l over the instrument. The direct-access keypad provides instant control over ns that are commonly used during inspections. The software menus provide
to the majority of instrument functions, such as pulser/receiver settings, ated calibration, measurement setups, software features, data logging lities, and more.
ing the value of a selected parameter is achieved using either the EPOCH 600 ent knob or the navigation keypad. The parameter adjustment method is
ined by the configuration of the instrument selected at time of ordering. Both ds are outlined below.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
32 Chapt
2.1.1 Menus and ParametersThe majority of functions on the EPOCH 600 can be accessed and adjusted using the horizontal and vertical keys in the software user interface. The vertical keys, located on the right side of the display, are called menus, and the horizontal keys along the bottom of the display are called parameters, functions, or submenus (see Figure 2-1 on page 32). Each menu or parameter can be selected for adjustment by pressing its corresponding [F] or [P] key on the instrument keypad (see Figure 2-1 on page 3
There with a[NEXT
Live A
Mearead
Parame
GaGer 2
2).
Figure 2-1 Software main display elements
are five groups of menus on the EPOCH 600. Each menu group is identified number (1/5, 2/5, 3/5, 4/5, and 5/5). To scroll through all menu groups, use the GROUP] key (see Figure 2-2 on page 33).
-scan view
surementing boxes
ter buttons
Flag
Menu
Basic parameters
te 2 (blue)ate 1 (red)
Menu group level number
DMTA-10006-01EN [U8778382], Rev. A, March 2011
2.1.2Once aadjustand fin
CoarseFigure
[NEXT GROUP] key
1Basic Operation 33
Figure 2-2 The menu groups and their level numbers
Parameter Adjustment Knob Configuration parameter is selected, you can modify it using the adjustment knob. You can
most parameters in either coarse or fine increments. To toggle between coarse e adjustment, press the [CHECK] key.
adjustment selection is indicated by brackets around the parameter key (see 2-3 on page 33). Fine adjustment is selected when no brackets are displayed.
Figure 2-3 Coarse and fine adjustment selection
/5 2/5 3/5 4/5 5/5
DMTA-10006-01EN [U8778382], Rev. A, March 2011
34 Chapt
2.1.3 Parameter Adjustment Navigation Pad ConfigurationOnce a parameter is selected, you can modify it using the arrow keys on the navigation pad (see Figure 2-4 on page 34). You can adjust most parameters in either coarse or fine increments. The up and down arrows adjust in coarse increments, and the left and right arrows adjust in fine increments.
2.1.4 DTo facidirect-jumpser 2
Figure 2-4 Arrow keys on the navigation pad
irect-Access Keyslitate fast adjustment of common parameters, the EPOCH 600 features a set of access keys (see Figure 2-5 on page 35). When pressed, the software interface directly to the associated parameter or activates the correct function.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
The EPpage 3
Figu
Navigation pad configuration
Adjustment knobconfigurationBasic Operation 35
Figure 2-5 The direct-access keys Both configurations (English)
OCH 600 is also offered in an international-symbol version (see Figure 2-6 on 5).
re 2-6 The direct-access keys Both configurations (international symbols)
Direct-access keysDirect-accesskeys
Navigation pad configuration
Adjustment knobconfiguration
Direct-access keysDirect-accesskeys
DMTA-10006-01EN [U8778382], Rev. A, March 2011
36 Chapt
When the [RANGE], (DELAY), [dB], or (REF dB) direct-access keys are pressed, common preset values appear above the [P] parameter keys. To select a preset value, press the corresponding [P] parameter key.
2.1.5 Special FunctionsThere are several special functions to be aware of while using the EPOCH 600:
Afpr
Prme
Thpaac
Praufuer 2
ter adjusting a parameter using direct-access keys, you can get back to the evious menu group using the [ESCAPE] key or the [NEXT GROUP] key.essing the [ESCAPE] key from any menu group will return the user to the Basic nu, which is the default menu at start-up.e [LOCK] key (knob configuration only) automatically locks adjustment of all rameters to prevent any unwanted parameter modifications due to an cidental movement of the knob (see Figure 2-7 on page 36).
Figure 2-7 Adjustment knob configuration [LOCK] key
essing [2ND F], (AUTO XX%) activates the AUTO XX% feature, which tomatically adjusts the gain in order to set the gated echo amplitude to XX% of ll screen height (the default XX value is 80 %). (See Figure 2-8 on page 37.)
[LOCK] key
DMTA-10006-01EN [U8778382], Rev. A, March 2011
2.1.6 SWhen througused toview.
2.2 Pu
The EPPulsercontro
2.2.1The sy
To adjust th
1. Pr2. Ad
ORBasic Operation 37
Figure 2-8 The AUTO XX% feature
ubmenusselecting some items, such as Display Setup, the [NEXT GROUP] key scrolls h rows within the submenu, the adjustment knob or navigation pad arrows are adjust the selected value, and the [ESCAPE] key returns to the live A-scan
lser and Receiver Settings
OCH 600 allows access to most of its pulser and receiver settings through the and the Rcvr menus. System sensitivity (gain) and reference gain are lled exclusively using the direct-access keys.
Sensitivitystem sensitivity (gain) is adjusted using the [dB] direct-access key.
e system sensitivity
ess [dB].just the value:Using the navigation pad arrow keys or by rotating the adjustment knob in either coarse or fine increments
Pressing one of the [P] parameter keys to select the corresponding preset value
DMTA-10006-01EN [U8778382], Rev. A, March 2011
38 Chapt
Gain can also be automatically adjusted using the AUTO XX% feature. See Special Functions on page 36.
2.2.2 Reference GainYou can define a reference gain by pressing [2ND F], (REF dB). This sets the current gain as the reference gain and activates a scanning gain for further adjustments (see Figure 2-9 on page 38).
When follow
Adres
Sc Of +6 6
2.2.3 PThe EPEach inadjust
The fo
PRthof
PR10
En Daer 2
the reference gain is set, the [P] parameter keys allow access to the ing functions:
d: Combines current scanning gain with current reference gain and sets the ultant as the new reference gain.an dB: Toggles between current scanning gain and 0.0 dB scanning gain.f: Turns off reference gain feature (scanning gain is lost).dB: Increases the scanning gain by 6 dB.dB: Decreases the scanning gain by 6 dB.
Figure 2-9 The reference and scanning gains
ulser OCH 600 main pulser functions are accessible by selecting the Pulser menu. dividual pulser parameter is displayed above the parameter keys, and can be
ed by pressing the corresponding [P] parameter key.
llowing pulser functions are available for live adjustment in the Pulser menu:
F Mode: Selects either Auto or Manual PRF adjustment mode. Auto changes e PRF setting based on the screen range, and Manual allows manual adjustment PRF in 10-Hz increments.F: Pulse Repetition Frequency (PRF) values range: from 10 Hz to 2000 Hz in
-Hz incrementsergy: Pulse voltage values available: 0 V, 100 V, 200 V, 300 V, or 400 Vmp: Pulse damping values available: 50 , 100 , 200 , or 400
Scanning gainReference gain
DMTA-10006-01EN [U8778382], Rev. A, March 2011
Mode: pulse modes available: P/E (pulse-echo), Dual, and Thru (through-transmission)
Pulser: Pulser waveforms available: Spike or Square (tunable square wave) Freq: Pulse frequency (square wave pulse width) ranges from 0.1 MHz to
20.00 MHz
The Sp
2.2.4The EPmenu.can be
The fomenu:
Fil Re
[h Re
The eigsettingset wit
0.2 2.0 8.0 0.5 0.2 1.5 5.0 DCBasic Operation 39
ike pulse is equivalent to a 20-MHz square wave pulse.
ReceiverOCH 600 standard receiver functions are accessible by selecting the Rcvr
Each individual receiver parameter is displayed above the parameter keys, and adjusted by pressing the corresponding [P] parameter key.
llowing receiver parameters are available for live adjustments in the Rcvr
ter: Receiver filter setting ct: Waveform rectification (Full [fullwave], Half+ [halfwave positive], Half
alfwave negative], RF [no rectification])ject: Percent reject (from 0 % to 80 %)
ht filters available on the EPOCH 600 allow broadband or narrowband s to match the requirements of a given application. Each is a fully digital filter h the following low-pass and high-pass cutoffs:
MHz10 MHzMHz21.5 MHzMHz26.5 MHzMHz4.0 MHzMHz1.2 MHzMHz8.5 MHzMHz15 MHz10 MHz
DMTA-10006-01EN [U8778382], Rev. A, March 2011
40 Chapt
2.3 Gates
The EPOCH 600 features two standard, independent measurement gates: gate 1 and gate 2. Gate 1 is displayed as a red, solid horizontal line. Gate 2 is displayed as a blue, hollow horizontal line. These gates independently define digital measurements regions for amplitude, time-of-flight measurements, and other specialized readings. Each gate also features alarm and zoom functions.
2.3.1 QThe [Gadjustmmost c
When start pincreas
Pressinsettingto the gates w
For momanag
2.3.2Both Gfunctioabove er 2
uickly Adjusting Basic Gate ParametersATES] direct-access key allows instant access to gate start, width, and level ents without requiring the user to enter any gate-related menus. This is the
ommon method for gate adjustments.
the [GATES] key is pressed, the box above the first menu displays the gate 1 osition (see Figure 2-10 on page 40). Once selected, the knob or arrows can e or decrease this value, in coarse or fine increments.
Figure 2-10 The Gate 1 start position adjustment
g the [GATES] key repeatedly will scroll through the start, width, and level of every active gate. Pressing [ESCAPE] or [NEXT GROUP] returns the user
menu group used prior to gate adjustment, allowing efficient adjustment of the ith minimal interference for the operator.
re comprehensive gate setup and adjustment, there are three menus that e gate settings: Gate1, Gate2, and Gate Setup.
Gate 1 and Gate 2ate 1 and Gate 2 menus allow access to specific gate positioning and alarm ns. When either menu is selected, the following parameters become visible
the [P] parameter keys.
Gate 1 start position adjustment
DMTA-10006-01EN [U8778382], Rev. A, March 2011
Zoom: Sets the display range to begin at the start position of the selected gate, and to stop at the end position of the selected gate (gate start + gate width).
Pressing the Zoom [P] parameter key repeatedly toggles the zoom on and off.
St W Le Al
De M
th St
ga
2.3.3 GThe Gprior tfollow
G1 G1
rec G1
%Pe
G2 G2
Tr
Gate mecho o
PeregBasic Operation 41
art: Adjusts the start position of the selected gate.idth: Adjusts the width of the selected gate.vel: Adjusts the screen height of the selected gate (from 3 % to 95 %).arm: Sets the alarm condition of the selected gate (Off, Positive, Negative, Min pth)
in Depth: Visible only when the Alarm parameter is set to Min Depth. Adjusts e threshold in time-of-flight units that triggers a Min Depth alarm condition.atus: Turns the gate on or off (effects measurements, alarms, and visibility of te on screen).
ate Setupate Setup menu allows more advanced settings for each gate to be adjusted o inspection. These settings are visible above the [P] parameter keys. The ing gate settings are available in the Gate Setup menu:
Mode: Sets the measurement trigger mode of gate 1 (Peak, 1stPeak, Edge) RF: Sets the polarity of the gate 1 when the instrument is operating in RF tification (Dual, Positive, Negative) %Amp: Used only in Edge mode. Sets the measurement trigger for the amplitude digital measurement of gate 1 when in Edge detection mode (High ak, 1stPeak). Mode/G2 RF/G2 %Amp: Same settings as above, but for gate 2 only. Tracks: Toggles tracking mode of gate 2 with respect to gate 1 on or off.
acking ON is considered true echo-to-echo measurement mode.
easurement modes, selected by G1 Mode or G2 Mode, determine which gated r echo parameters triggers a digital measurement.
ak: Acquires measurement readings based on the highest peak within the gated ion (does not have to cross the gate threshold) [see Figure 2-11 on page 42].
DMTA-10006-01EN [U8778382], Rev. A, March 2011
42 Chapt
1stPeak: Acquires measurement readings based on the first peak to cross the gate threshold (level) [see Figure 2-11 on page 42].
Edge: Acquires measurement readings based on the position of the first crossing point of a gated signal (see Figure 2-11 on page 42).
2.3.4 AWhenetwo di
An On
EPer 2
Figure 2-11 Measurement trigger in Edge, Peak, and 1stPeak modes
larm Indicationsver an alarm is triggered on either measurement gate, the operator is alerted in
fferent ways:
audible tone is emitted from the EPOCH 600e of two alarm indicators (LEDs) is illuminated on the front panel of the OCH 600 (see Figure 2-12 on page 42)
Figure 2-12 Gate 1 and gate 2 alarm indicator lights
Edge Peak 1stPeak
Gate 1 alarm indicator
Gate 2 alarm indicator
DMTA-10006-01EN [U8778382], Rev. A, March 2011
The EPOCH 600 is also capable of outputting alarm signals through the 9-pin D-sub connector on the back panel of the instrument.
2.4 Ca
The EPprovidEPOCsectiondetaile
2.4.1CalibrknownTHICKcalibra
To begin ca
1. Co2. Po3. Br4. SeBasic Operation 43
libration
OCH 600 can be easily calibrated for both zero offset and velocity in order to e accurate thickness (soundpath) measurements from indications. The
H 600 utilizes an auto-calibration system for a simple, two-step approach. The below outlines the most standard and basic calibration example. More d information on calibration can be found in chapter 10 on page 149.
Measurement calibration ation for accurate digital measurements is typically accomplished using two thicknesses of a representative material. In this section, the terms THIN and will be used to refer to either two thicknesses of a block or plate (straight beam tions) or two lengths of an angled soundpath (angle beam calibration).
libration
uple the transducer to the thin step of the material.sition gate 1 around the resulting indication on screen.ing the indication to 80 % full screen height using the (AUTO XX%) function.lect the Auto Cal menu (see Figure 2-13 on page 44).
DMTA-10006-01EN [U8778382], Rev. A, March 2011
44 Chapt
5. Usfor
6. Prer 2
Figure 2-13 The Auto Cal menu
e the Mode parameter to select the appropriate calibration mode (Thickness straight beam or Soundpath for angle beam are the most common).
ess Cal-Zero (see Figure 2-14 on page 44).
Figure 2-14 The Cal-Zero value
DMTA-10006-01EN [U8778382], Rev. A, March 2011
7. Use the knob or arrow keys to adjust the displayed value to the appropriate thickness. For this example, the transducer is coupled to a 5 mm step.
8. Press Continue to accept the adjusted value, and then move on to the second calibration step.
9. Couple the transducer to the thick step of material.10. Position gate 1 around the resulting indication on screen.11. Bring the indication to 80 % full screen height using the (AUTO XX%) function
(se12. Pr
13. Pr14. Us
thFigBasic Operation 45
e Figure 2-15 on page 45).ess [ESCAPE] to return to the Auto Cal menu.
Figure 2-15 The gate 1 start
ess Cal-Vel.e the knob or arrow keys to adjust the displayed value to the appropriate
ickness. For this example, the transducer is coupled to a 12.5 mm step (see ure 2-16 on page 46).
DMTA-10006-01EN [U8778382], Rev. A, March 2011
46 Chapt
15. Pr16. Pr
Figer 2
Figure 2-16 The Velocity Cal value
ess Done to accept the adjusted value and complete the calibration process.ess [RANGE], and then adjust the screen range to the desired setting (see ure 2-17 on page 46).
Figure 2-17 The Range value
DMTA-10006-01EN [U8778382], Rev. A, March 2011
2.4.2 Angle Beam CalibrationAngle beam calibrations typically require four steps. Each step is outlined below:
To perform angle beam calibration
1. Verify the Beam Index Point (B.I.P.).2. Verify the refracted angle of wedge.3. Co
(C4. Se
indref
2.5 Da
The EPstoragoptionRegardBasic Operation 47
mplete the distance calibration using steps of section 2.4.1 on page 43 alibration mode set to SoundPath or Depth depending on the reflector type).t the sensitivity by capturing sensitivity hole or notch in gate 1, bringing the ication to 80 % screen height using (AUTO XX%), and then setting the
erence gain by pressing [2ND F], (REF dB) [see Figure 2-18 on page 47].
Figure 2-18 Reference gain setting
ta Logger
OCH 600 features an onboard data logging system that allows standard e of inspection files (Inc) and dedicated calibration files (Cal), as well as al inspection files in corrosion style configurations (2D, 3D, Boiler, etc). less of file type, each discrete data point saved on the EPOCH 600 stores all
DMTA-10006-01EN [U8778382], Rev. A, March 2011
48 Chapt
active digital measurements, compressed A-scans, calibration data, alarm conditions, and active software features. The onboard memory is stored on a 2-GB MicroSD card, allowing storage of over 500,000 individual data points.
The section below outlines the procedure for setup and storage of the most basic and common file type, the calibration file. For full details and procedures on standard and optional file types, see chapter 11 on page 195.
2.5.1 CTo setuand so
To complet
1. Se2. Pr
Th3. In4. Us5. In
ther 2
alibration Filesp and save a calibration file on the EPOCH 600, first complete all ultrasonic
ftware setups that will be saved within the calibration.
e all ultrasonic and software setups
lect the File menu.ess the Create key to enter the file creation screen.e Create screen appears (see Figure 2-19 on page 48).
the File Type box, use the knob or arrow keys to select Cal.e the [NEXT GROUP] key to advance to the Filename box.
the Filename box, create a file name (cannot exceed 32 characters) by pressing e Edit parameter key.
Figure 2-19 The Create screen
DMTA-10006-01EN [U8778382], Rev. A, March 2011
6. Use the knob or arrows to navigate through the virtual keypad (see Figure 2-20 on page 49).
7. Insert a character by pressing INS (see Figure 2-20 on page 49).
8. Us
For cre
9. Pr10. PrBasic Operation 49
Figure 2-20 The virtual keypad
e the [NEXT GROUP] key to complete the file naming process.
ation of an incremental (Inc) file, you must first create a Start Point.
ess the [NEXT GROUP] key until the Create button is highlighted.ess Save (see Figure 2-21 on page 50).
DMTA-10006-01EN [U8778382], Rev. A, March 2011
50 Chapt
11. PrTh
2.5.2 OTwo o
Crloc
Oploc(Sstaer 2
Figure 2-21 The Save dialog box
ess [P1] to accept the function. e user is then returned to the live screen.
ther Create Functionsther file creation functions exist, and are described below:
eate: Creates the file in memory but does not open this file as the active storage ation. Typically used when creating several files at once without saving data.en: Creates the file in memory, and then opens it as the active storage ations, but does not save any file parameters to the file until you press [2ND F],
AVE). Typically used for inspection files, where the file is created prior to the rt of the inspection.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
3. EPOCH 600 Hardware Features
The EPimprofor you
This se
H F C V EEPOCH 600 Hardware Features 51
OCH 600 has many physical features that are either completely new or ved compared to previous EPOCH flaw detector instruments. It is important to be familiar with the use and maintenance of these items.
ction covers the following topics:
ardware Overview on page 52ront Panel User Interface on page 53onnectors on page 62arious Hardware Features on page 66nvironmental Ratings on page 68
DMTA-10006-01EN [U8778382], Rev. A, March 2011
52 Chapt
3.1 Hardware Overview
Figure 3-1 on page 53 shows the EPOCH 600 instrument and identifies its main components.
D-rinoptier 3
Protective rubber bumpers
gs (4) to attach theonal chest harness
Front panel user interface
Key pad(navigation pad
configurationshown)
DMTA-10006-01EN [U8778382], Rev. A, March 2011
3.2 Fr
In ordavailabwith apad areither adjustm
3.2.1 KThe ad[ESCAFigureparamvalue s
AC power connector
The VGA Outconnector
Alarms connector
Conventional transducer connectorsEPOCH 600 Hardware Features 53
Figure 3-1 Overview of the EPOCH 600 hardware
ont Panel User Interface
er to accommodate different user needs and preferences, the EPOCH 600 is le in two hardware configurations: one with an adjustment knob and the other
navigation pad. The adjustment knob and the arrow keys on the navigation e responsible for parameter adjustment and value slewing. Users can choose the adjustment knob or the navigation pad when ordering depending on which
ent method they are more comfortable with.
nob Configurationjustment knob on the EPOCH 600 is used along with the [CHECK] and PE] keys to adjust parameter values in either coarse or fine increments (see 3-2 on page 54). The user has the ability to lock the knob to prevent accidental eter value changes during an inspection. This configuration provides smooth lewing for customers who prefer adjusting parameters using the knob.
Pipe stand
Battery compartment cover
Membrane ventSide door
DMTA-10006-01EN [U8778382], Rev. A, March 2011
54 Chapt
3.2.2The nadown the lefadditio[CHEC(1000 S
The EPdirect-keys topanel of valudispla
Adjustmentknober 3
Figure 3-2 The EPOCH 600 Adjustment knob configuration
Navigation Pad Configurationvigation pad is a hallmark feature of the EPOCH flaw detectors. The up and arrows on the navigation pad are used for coarse parameter adjustment, and t and right arrows for fine adjustment. The navigation pad also contains nal functions and frequently used parameters such as gain, save, and the K] and [ESCAPE] keys. This configuration is similar to previous EPOCH units eries, XT, LTC) to provide easy transitions within the EPOCH product line.
OCH 600 front panel shown in Figure 3-3 on page 55 features a combination of access keys, navigation arrows, and dynamic function and parameter access optimize the usability of the instrument in any mode. The layout of the front
provides direct access to common inspection parameters, and easy adjustment es from either side of the instrument without obstructing the view of the
y.
DMTA-10006-01EN [U8778382], Rev. A, March 2011
The diaccesssection
3.2.3 GThe EPthe conpurpoand Fi
Direct-ak
Indicators[ON/OFF]
power key
Display windowNavigation
padEPOCH 600 Hardware Features 55
Figure 3-3 The EPOCH 600 Navigation pad configuration
rect-access keypad, on the left side of the front panel, provides keys for direct to software parameters that are commonly used during inspection (see
3.2.6 on page 58 for details).
eneral Purpose KeysOCH 600 provides navigation arrows or an adjustment knob (depending on figuration), but also the [CHECK] and [ESCAPE] keys, which are general
se keys, regardless of instrument mode or function (see Figure 3-4 on page 56 gure 3-5 on page 56).
ccesseypad
Function [F] keys
Parameter [P] keys
DMTA-10006-01EN [U8778382], Rev. A, March 2011
56 Chapt
The [C
Wad
[CHECK] key
[ESCAPE] key
[RIGHT][LEFT]
[UP]
[ESCAPE][DOWN]
[CHECK]
[er 3
Figure 3-4 General purpose keys English version
Figure 3-5 General purpose keys International-symbol version
HECK] key has two primary functions:
hen an adjustable parameter is highlighted, [CHECK] toggles the parameter justment between a coarse and a fine adjustment (knob configuration).
[CHECK] key
[ESCAPE] key
[RIGHT][LEFT]
[UP]
[ESCAPE]DOWN]
[CHECK]
DMTA-10006-01EN [U8778382], Rev. A, March 2011
Coarse adjustment is indicated by brackets around the parameter key. Fine adjustment does not have brackets.
The [CHECK] key is used to scroll through the menus in numerical order (navigation pad configuration).
The [ESCAPE] key has three primary functions:
When in a setup page, [ESCAPE] returns to the live inspection screen. W W
[E
3.2.4 FThe EPusing verticakeys ([aroundEPOCH 600 Hardware Features 57
hichever the selected menu, the [ESCAPE] key returns to the Basic menu.hen a direct access parameter (gain, range, gates, etc.) is selected, pressing the SCAPE] key will return to the previous menu.
unction and Parameter KeysOCH 600 handles adjustments and/or the activation of most software functions
an intuitive menu system. Menu system software buttons appear at all times, lly on the right side and horizontally at the bottom of the display. Five function F1] through [F5]) and seven parameter keys ([P1] through [P7]) are located the display, allowing you to individually activate a software button.
Figure 3-6 [F] and [P] keys pointing to software buttons
Parameter keys
Function keys
DMTA-10006-01EN [U8778382], Rev. A, March 2011
58 Chapt
3.2.5 Parameter AdjustmentSoftware parameter value adjustments, such as gain or range, can be accomplished using two primary methods, depending on the instrument configuration:
Use the [UP], [DOWN], [LEFT], and [RIGHT] arrow keys to increase and decrease a parameter value in either coarse or fine steps (navigation pad configuration).
Tuco(kn
The [Cfocus o
3.2.6This seFigureer 3
rn the adjustment knob clockwise to increase parameter values, and unterclockwise to decrease a parameter value in either coarse or fine steps ob configuration).
HECK] key sets the coarse/fine mode and the [ESCAPE] key returns the n the Basic menu.
About the Direct-Access Keypadction describes the direct-access keypad for both the navigation pad (see 3-7 on page 59) and the knob configurations (see Figure 3-8 on page 60).
DMTA-10006-01EN [U8778382], Rev. A, March 2011
FigEPOCH 600 Hardware Features 59
ure 3-7 Navigation pad configuration (English and international-symbol versions)
DMTA-10006-01EN [U8778382], Rev. A, March 2011
60 Chapt
Fig
Table 4the key
Nav.coner 3
ure 3-8 Knob configuration (English and international-symbol versions)
on page 60 presents a description of each of the key for the English version of pad.
Table 4 English direct-access keypad key description
pad fig.
Knob config.
Function
[dB]Adjusts system sensitivity. [2ND F], (REF dB)Locks the reference gain level and allows the scanning gain to be used.
DMTA-10006-01EN [U8778382], Rev. A, March