Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) version REV. REVISION DESCRIPTION ECO NO. DATE NAME 7 8 Added the need that cable of sensor to be connected from back of the bed. – for safety purposes Changed pictured and changed labels, class A EMC ECO-11-12 Feb. 2, 2011 April 2011 Dalia Argaman Dalia 9 Software ver. 3.0.13 ECO-11-21 June 2011 Shiraz 10 Clarifying “children” definition, updated cleaning and disinfection procedure and changing “EverOn” to “EarlySense System” ECO-12-01 Dec 2011 Shiraz / Dalia / Aya 11 Change name of device to EarlySense (Labels with the new “EarlySense” logo) ECO-12-13 April 2012 Shiraz NAME DATE SIGNATURE Written by: Shiraz Levkovich Appr. Marketing Maayan Wenderow Appr. Product manager Roman Karasik Appr. Clinic., Regul. Dalia Argaman Appr. QA Asaf Grosgold EarlySense Ltd. 12 Tzvi St., Ramat Gan, Israel Proprietary and Confidential CONFIDENTIALITY NOTICE This document contains valuable trade secrets and confidential information of EarlySense Ltd. Nothing herein may be copied, reproduced or distributed in any form or any medium, or disclosed to any third party in any manner, without prior written authorization of EarlySense.

Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

  • Upload
    ngonhan

  • View
    230

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

Doc. No. MK600

SUBJECT: EarlySense User Guide – USA (FDA) version

REV. REVISION DESCRIPTION ECO

NO.

DATE NAME

7

8

Added the need that cable of sensor to be

connected from back of the bed. – for

safety purposes

Changed pictured and changed labels,

class A EMC

ECO-11-12

Feb. 2, 2011

April 2011

Dalia Argaman

Dalia

9 Software ver. 3.0.13 ECO-11-21 June 2011 Shiraz

10 Clarifying “children” definition, updated

cleaning and disinfection procedure and

changing “EverOn” to “EarlySense

System”

ECO-12-01 Dec 2011 Shiraz / Dalia / Aya

11 Change name of device to EarlySense

(Labels with the new “EarlySense” logo)

ECO-12-13 April 2012 Shiraz

NAME DATE SIGNATURE

Written by: Shiraz Levkovich

Appr. Marketing Maayan Wenderow

Appr. Product

manager

Roman Karasik

Appr. Clinic., Regul. Dalia Argaman

Appr. QA Asaf Grosgold

EarlySense Ltd.

12 Tzvi St., Ramat Gan, Israel

Proprietary and Confidential

CONFIDENTIALITY NOTICE

This document contains valuable trade secrets and confidential information of EarlySense Ltd.

Nothing herein may be copied, reproduced or distributed in any form or any medium, or

disclosed to any third party in any manner, without prior written authorization of EarlySense.

Page 2: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

The EarlySense System

(EarlySense Model 1.0 - Equivalent to model EverOn 1.0)

User Guide

Page 3: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

i

EarlySense Ltd.

EarlySense 1.0

User Guide Revised April, 2012

SW Version: 3.0.13

Document Num. MK-600 Revision Ver. 11

COPYRIGHT 2011 EARLYSENSE LTD. ALL RIGHTS RESERVED WORLDWIDE THIS MANUAL IS

PROPRIETARY TO EARLYSENSE LTD. UNAUTHORIZED USE, DISCLOSURE OR REPRODUCTION IS

PROHIBITED

EARLYSENSE TECHNOLOGIES ASSUMES NO RESPONSIBILITY FOR ERRORS THAT MAY OCCUR IN

THIS MANUAL. EARLYSENSE LTD RESERVES THE RIGHT TO MAKE CHANGES WITHOUT NOTICE

TO THIS EVERON MANUAL, ALSO REFERRED TO AS “THE EARLYSENSE SYSTEM”,AND THE

PRODUCTS DESCRIBED HEREIN.

Contact EarlySense at:

EarlySense Ltd.

European Representative MedNet GmbH Borkstrasse 10 48163 Münster, Germany Phone +49 251 322660 Fax +49 251 3226622

Israel

EarlySense Ltd.

12 Tzvi St.,

Ramat Gan 52504, Israel

Web-site: www.earlysense.com

USA EarlySense Inc. 135 Beaver Street,

Suite 211, Waltham, MA 02452

Phone: (781) 373 3228 E Mail: [email protected] Support Telephone:

East Coast - (617) 800-6668

West Coast - (818) 370-7748

Web-site: www.earlysense.com

Page 4: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

ii

Table of Contents

1 Introduction ....................................................................................................................... 1

The EarlySense System Bedside Monitor ........................................................................................... 1

About This Manual ............................................................................................................................... 2

Manual Conventions ............................................................................................................................ 2

Menus Screens and dialog boxes ............................................................................................................. 2

Warnings, Cautions and Notes ................................................................................................................. 3

Terms ................................................................................................................................................... 3

2 Safety ................................................................................................................................. 4

General Safety Guidelines ................................................................................................................... 4

General Hazards .................................................................................................................................. 4

Electrical Shock ........................................................................................................................................ 5

Defibrillation .............................................................................................................................................. 5

Electric Magnetic Interference .................................................................................................................. 5

WiFi Communication ................................................................................................................................. 5

Electrical Fire ............................................................................................................................................ 5

Classification ........................................................................................................................................ 5

Intended Use ........................................................................................................................................ 6

Contraindications for Use ..................................................................................................................... 6

Warnings and Cautions ........................................................................................................................ 6

System Labels...................................................................................................................................... 9

Bedside Unit Label.................................................................................................................................... 9

Sensing Unit Label.................................................................................................................................. 10

System Packaging Label ........................................................................................................................ 10

Explanation of System Labels ................................................................................................................ 11

Compliance with Standards ............................................................................................................... 12

How the Bedside Unit is supplied ...................................................................................................... 12

3 System Description ........................................................................................................ 13

System Components .......................................................................................................................... 13

Sensor .................................................................................................................................................... 13

Sensor Extension Cable ......................................................................................................................... 13

Bedside Unit Controls and Indicators ...................................................................................................... 14

Battery ............................................................................................................................................. 16

The EarlySense System’s User Interface and Screens ..................................................................... 17

User Interface ......................................................................................................................................... 17

System Messages and Alerts ........................................................................................................... 18

Screens .................................................................................................................................................. 18

Setting the volume level of the bedside unit (page 32).............................................................................. 21

Page 5: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

iii

4 System Operation ........................................................................................................... 22

Overview ............................................................................................................................................ 22

Positioning the Bedside and Sensing Units ....................................................................................... 23

Turning the System On/Off ................................................................................................................ 24

Electrical Failure................................................................................................................................. 24

After Electrical Power is Restored .......................................................................................................... 26

Patient Information (Admit/Discharge) ............................................................................................... 26

Admitting a Patient.................................................................................................................................. 26

Discharging a Patient / Ending a Patient Monitoring Session ("Discharge") ........................................... 31

Pausing a Patient Monitoring Session ............................................................................................... 32

Setting the General Volume ............................................................................................................... 33

Alerts .................................................................................................................................................. 34

Alerts Table ............................................................................................................................................ 35

Multiple Alerts ................................................................................................................................ 36

Alert Triggers .......................................................................................................................................... 37

Alerting Parameters......................................................................................................................... 37

Modifying Alert Thresholds ..................................................................................................................... 38

Audible and Visual Alert Notification ....................................................................................................... 39

Blue Alerts .............................................................................................................................................. 39

Suspending Red and Yellow Alerts......................................................................................................... 40

Disabling and Enabling Alerts ................................................................................................................. 40

Alerts Setup ............................................................................................................................................ 42

Setting Individual Alert Volume and Tone ..................................................................................... 42

Silencing Alerts ............................................................................................................................... 43

Bed Exit .............................................................................................................................................. 43

Configuring the Bed Exit Alert ................................................................................................................. 44

Setting Bed Exit Sensitivity ............................................................................................................ 44

Patient Turn ........................................................................................................................................ 45

Setting the Patient Turn Interval ............................................................................................................. 45

Starting the Patient Turn Counter ........................................................................................................... 45

Documenting Patient Turn and Resetting Counter ................................................................................. 46

Air Mattress ............................................................................................................................................ 47

Patient Movement .............................................................................................................................. 48

Low Patient Movement Alert ................................................................................................................... 48

Patient Motion Detection ......................................................................................................................... 49

Recording and Viewing Events .......................................................................................................... 50

Manually Recording an Event ................................................................................................................. 50

Trends ................................................................................................................................................ 51

Trends Graph ......................................................................................................................................... 51

Alert Symbols ................................................................................................................................. 54

Event Symbols ................................................................................................................................ 55

Setting the Time Scale .................................................................................................................... 56

Analysis .................................................................................................................................................. 56

Page 6: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

iv

Settings .............................................................................................................................................. 57

Brightness ............................................................................................................................................... 57

Choosing a Language ............................................................................................................................. 57

Maintenance ........................................................................................................................................... 58

Reports ............................................................................................................................................... 59

Generating Reports (Data Retrieval) ...................................................................................................... 59

Downloading Patient Data and Reports .................................................................................................. 59

5 Specifications, Maintenance and Troubleshooting ...................................................... 63

Specifications ..................................................................................................................................... 63

Physical Characteristics .......................................................................................................................... 63

Performance ........................................................................................................................................... 64

Electrical ................................................................................................................................................. 64

Operating Conditions .............................................................................................................................. 64

Storage and Transportation Conditions .................................................................................................. 64

Maintenance and Cleaning ................................................................................................................ 65

Cleaning and Disinfection of the Bedside Unit ........................................................................................ 65

Cleaning and Disinfection of the Sensing Unit ........................................................................................ 65

Troubleshooting ................................................................................................................................. 66

6 European Representative and Contact Information ..................................................... 67

Page 7: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

1

1 Introduction

The EarlySense System Bedside Monitor The EarlySense System (also known as EverOn) is designed for continuous and contact-

free measurement of respiration rate and heart rate when the patient is lying in bed. In

addition the system tracks body motion and monitors patient movement and can notify

users upon patient bed exit .The system displays numerical heart rate, respiration rate and

movement level and graphical data of trends of these parameters.

The system's design allows the operator to adjust the settings of the threshold parameters

and notifies the caregiver when heart or respiration rate averaged over time, passes above

or below predefined limits. The system also allows the health practitioner to document

the changes in the patient's position in bed by recording the event. The system then

verifies the change in position and logs the information. The system also provides an

option to notify the health care practitioner if a specific amount of time has passed since

the nurse last changed the patient's position. The users can decide to activate or deactivate

these notifications.

Providing contact-free, passive monitoring capabilities, with no need for patient

activation or involvement, the EarlySense System enables continuous monitoring of

patients in home, hospital or clinical settings. The data acquired by the System is

continuously logged in a Bedside Unit, thus, when using the System at home, the data

may be later presented, in a time stamped format, for off-line health care practitioner

analysis.

The system includes a Sensing Unit that is placed beneath the bed mattress or between

the mattress and a mattress pad or mattress cover, and a Bedside Unit which sets the

system options and displays the collected data. Monitoring of the patient begins

automatically as soon as the patient enters the bed. The data acquired by the system is

continuously logged in the Bedside unit, thus allowing the data to be presented, in a time

stamped format. The data provided by the system is intended to aid in the evaluation

process of a patient's clinical status and should be interpreted by a health care practitioner

only.

Page 8: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

2

Note

The EarlySense System has not been studied on any specific patient

group, neither has it been studied as a diagnostic tool of any

specific disease or medical condition. It is meant as an adjunctive

tool only for measuring respiratory rate, heart rate and movement

rate.

About This Manual This manual provides the information necessary to operate the EarlySense system in a

safe and efficient manner. Please read the manual thoroughly and understand its contents

before operating the system. If any part of this manual is not clear, contact EarlySense

Customer Support for clarifications.

The manual is intended to serve as an accompanying document to the EarlySense system.

It is not intended to replace the users training course.

This manual should always accompany the unit, and all personnel operating the unit must

be aware of its location.

Manual Conventions The following manual conventions are used throughout the manual.

Menus Screens and dialog boxes

The names of menus, screens and dialog boxes are presented in bold

Page 9: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

3

Warnings, Cautions and Notes

Warnings, Cautions and Notes are used throughout this manual:

Warning

A warning indicates precautions and instructions which, if not

followed may result in serious bodily injury or death

Caution

A caution indicates instructions, or cautionary notes which, if not

followed, may result in a damage to the equipment or to the quality

of measurements

Note

Notes contain helpful information and tips

Terms

HR Heart Rate

RR Respiratory Rate

Page 10: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

4

2 Safety

General Safety Guidelines

Warning

US Federal Law restricts this device to sale by or on

the order of a physician.

The data acquired by the EarlySense System should

be interpreted by a health care professional only.

• Handle the Bedside Unit with care. Do not drop, knock, or shake the Bedside

Unit. Rough handling can damage the internal circuit boards.

• The System is intended for indoor operation only.

• A damaged system should not be disposed of as unsorted municipal waste.

Contact your local distributor for unit disposal.

• Changes or modifications not expressly approved by EarlySense Ltd. could affect

the safety or effectiveness of the EarlySense System and void the system's

warranty.

• The EarlySense Sensing Unit should be used only with the EarlySense Bedside

Unit, and the EarlySense Bedside Unit should be connected only to the

EarlySense Sensing Unit.

• The EarlySense Bedside Unit is continuously operated equipment with water-

proof splash sensor (IPX4).

General Hazards • Do not use a damaged System. Use of damaged components might result in

malfunctioning of the System.

• The EarlySense System should be installed and serviced only by qualified service

personnel, authorized by EarlySense Ltd.

Page 11: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

5

Electrical Shock

The system contains no user serviceable parts. Do not open the system covers.

The Bedside Unit is not waterproof. Keep the Bedside Unit dry to avoid electrical shock

or malfunction.

Defibrillation

The system is Defibrillation proof Type BF applied part.

Electric Magnetic Interference

This equipment complies with IEC EN 60601-1-2:2001 for electromagnetic compatibility

for medical electrical equipment and/or systems. This standard is designed to provide

reasonable protection against harmful interference in a typical medical installation.

However, because of the proliferation of radio-frequency transmitting equipment and

other sources of electrical noise in healthcare and other environments, it is possible that

high levels of such interference due to close proximity or strength of a source might

disrupt the performance of this device. Medical electrical equipment needs special

precautions regarding EMC, and all equipment must be installed by qualified service

personnel.

WiFi Communication

The system incorporates an off-the-shelf certified Wifi communication card (complies

with FCC part 15)

Electrical Fire

Avoid placing liquids or food on any part of the System. Do not allow conductive fluids

to leak into the active circuit components of the System as this may cause a short circuit,

which could result in an electrical fire. In such an event, only fire extinguishers approved

for use on electrical fires should be used. The device is not intended for use in the

presence of flammable mixtures.

Classification Mode 1: The unit is classified as Class I

Page 12: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

6

Mode 2: The unit is internally powered, continuously powered, ordinary portable

equipment with applied part (IPX4).

Only equipment specified in this manual and complying with requirements of EN60601

should be connected to the system

The system complies with IEC 60601-1, IEC 60601-1-2 and IEC60601-1-4 EMC class A

device.

Intended Use The EarlySense system is intended for continuous measurement of respiration rate, heart

rate and movement, in an automatic contact-less manner, at home, hospital or clinic

setting. The system is indicated for use in children, adolescents and adults. The operation

of the EarlySense System has been studied in children (weight ≥ 10 Kg) and adults

(weight ≤111 Kg) during sleep and resting condition.

Note

Definition of Age for Children: Children aged 2 and above

Contraindications for Use The EarlySense System is contraindicated for use in:

• Patients in whom proper positioning cannot be achieved or maintained.

• Patients who do not meet the weight limits tested or specified.

• Situations where a dry environment cannot be ensured.

• An MR environment

• An explosive atmosphere or in the presence of flammable anesthetics or gases

Warnings and Cautions • The EarlySense System should be installed and serviced only by qualified service

personnel, authorized by EarlySense Ltd.

• Mounting bed-side unit to the wall should be performed while exercising out-

most caution. DO NOT place the bed-side unit over patient’s head, to avoid

safety related conditions.

Page 13: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

7

• Mounting of bed-side unit to the wall should be performed by mechanical experts

of the institution (e.g., Biomed/ engineering), to make sure safe attachment.

EarlySense is not responsible for any harm or damage related to wrongful

placement of the bed-side unit

• Implementation of the nurse call connector option provided in The EarlySense

Bedside Monitor requires coordination between the hospital and EarlySense. Do

not implement this option without consulting with, and receiving approval from,

the hospital administration.

• The data acquired by The EarlySense System should be interpreted only by a

health care practitioner.

• In the event that the EarlySense System does not operate properly, contact

EarlySense Ltd. customer support: (617) 800-6668 (for the east coast) or (818)

370-7748 (for the west coast).

• Never open the Bedside Unit housing as this may damage the System. Refer all

servicing to an authorized technician.

• Only equipment specified in this manual and complying with requirements of

EN60601 might be connected to the system.

• Measurements may be affected by cable lengths. Shorten or extend the lengths

only according to the manual’s direction.

• Changes or modifications not expressly approved by EarlySense Ltd. could affect

the safety or effectiveness of the EarlySense System and void the System's

warranty.

• The System should be operated within a temperature range of 10-35°C (50-95 °F)

for the Bedside Unit, and 5-40°C (41-104 °F) for the Sensing Unit, and within a

relative humidity of 30-80%, non condensing.

• Do not use a damaged System. Use of damaged components might result in

malfunctioning of the System.

• Avoid placing liquids or food on any part of the System. Do not allow conductive

fluids to leak into the active circuit components of the System as this may cause a

short circuit, which could result in an electrical fire. In this event, only fire

extinguishers approved for use on electrical fires should be used. Care should be

taken for patients with poor bladder functioning or control, including small

children, when an EarlySense sensor is placed under their mattress.

Page 14: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

8

• Do not share the bed with another person or pet during EarlySense System

recording session. Sharing the bed could affect the effectiveness of the system

and the accuracy of the measurements.

• Avoid using heating blankets. Use of heating blankets could affect the safety or

effectiveness of the EarlySense System and void the System's warranty.

• Do not use the EarlySense System for patients who weigh more than 200 kg (440

Pounds). Usage of the System for such patients might result in malfunction of the

Sensing Unit.

• The patient should not have a direct contact with the Sensing Unit. A mattress,

mattress pad or mattress cover should be always placed as a barrier between the

Sensing Unit of the EarlySense System and the patient. Patients should be

frequently checked to insure direct contact does not occur.

• Careful oversight should be provided when the EarlySense System is used with

children.

• As with all medical equipment, carefully route cables and connections to reduce

the possibility of tripping, entanglement or strangulation.

• Do not create sharp bends in the cable, as this may tear or break the shielding.

Warning

The EarlySense System is not intended for monitoring high risk

situations where ECG monitoring is required. The most reliable

method of patient monitoring combines close personal surveillance

with correct operation of the monitoring equipment.

! Caution: US Federal Law restricts this device to sale by or on the order of a

physician.

Page 15: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

9

System Labels

Bedside Unit Label

Figure 1: Bedside Unit Label

Incorporates WiFi module

FCC ID: PD9WM3B2200BG

Page 16: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

10

Sensing Unit Label

Figure 2: Sensing Unit Label

System Packaging Label

Figure 3: Packaging Label

Page 17: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

11

Explanation of System Labels

The following provides a description of the graphical symbols that appear on the

EarlySense System components and package.

!

Caution/Warning

Consult Accompanying documents

Defibrillation proof, type BF, IPX4 applied part

Fragile, handle with care

Keep dry

Indoor operation only

Sorted disposal

Page 18: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

12

Compliance with Standards The EarlySense System was tested and found to be in compliance with the following

standards:

How the Bedside Unit is supplied The EarlySense System is shipped in a protective package containing the following

components:

• 1 Control Unit (Bedside unit)

• 1 Sensor

The System should be unpacked and installed only by an authorized EarlySense

technician.

STANDARD #

1. Medical electrical equipment- general requirements

for safety. Part 1: General Requirements for Safety IEC 60601-8

(2006)

2. Medical Electrical Equipment - Part 1: General

Requirements for Safety; Electromagnetic

Compatibility -- Requirements and Tests

EN/IEC 60601-1-2

(2005)

Page 19: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

13

3 System Description

System Components The system consists of a Sensing Unit and a Bedside Unit connected by a cable.

1.Bedside Unit

2. Sensor

Figure 4: The EarlySense System Components

Sensor

The Sensor is placed under the mattress (See Positioning the Bedside and Sensing Unit,

page 23) and detects a signal generated by the patient's breathing, heartbeat and motion.

The sensor’s cable, which should be connected to the bedside unit, is three-meter long.

Sensor Extension Cable

Only if required, an optional two-meter extension cable for the sensor is available.

Note

Use only one extension cable per sensor.

1

2

Page 20: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

14

Bedside Unit

The Bedside Unit shows system status, displays the heart rate, respiration rate, movement

level, patient turns and bed exit data, and allows system configuration.

1. Alert indicator key that can be

pressed as a response to an active

alert

2. On/Off

3. Suspend alerts for 15 minutes

4. Download reports

5. Speaker/ buzzer

Figure 5: Bedside Unit Indicators and Keys

Bedside Unit Controls and Indicators

The Bedside Unit contains the following controls and indicators:

1. Alert indicator key - the light will flash red in case of red alerts (e.g. heart rate,

respiration rate and bed exit), yellow in case of yellow alerts (movement and patient turn

reminder) and blue in case of technical alerts. While an alert is active, alert light key can

be pressed to respond to the alert.

2. ON-OFF button. Use the button to switch the bedside unit ON or OFF.

3. 15 min Suspend Key – Pressing it will disable the red and yellow alerts for 15 minutes.

No red or yellow alert will be given by the EarlySense System during the next 15

1

2

3

4

5

Page 21: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

15

minutes after this action! Alert symbols on the screen will turn into a Timer (see Figure

6) to notify of this temporary suspension mode. Pressing the Suspend key button again

turns the alerts back on. Blue alerts cannot be suspended under any circumstances.

Figure 6: Suspension Timer

4. Download Reports Key – password will be required

1. USB Connector

2. Nurse Call Connector

3. Sensor Connector

4. Network Connector

Other connectors (for future use) are

clearly labeled as not in use.

Figure 7: Bedside Unit - Side View

1

2

3

4

Page 22: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

16

Note

The nurse call in the EarlySense System is a UL1069 compliant

nurse call relay closure connector – Supplementary Device

The EarlySense bedside unit can interface with the existing nurse

call systems found in the hospital., though the actual connection to

the nurse call system requires the support of the Hospital

Biomedical Engineering department

Battery

The system is internally powered by a lithium polymer rechargeable battery that provides

power in case of a power failure (See Electrical Failure, page 24). The battery is not

intended to be replaced by the operator. In standard operating mode, the battery is

designed to allow approximately 30 minutes of operation while the unit is unplugged

from the Electrical Outlet, or in case of power failure.

The battery is automatically recharged by the system. In case the battery no longer

maintains the necessary charge, it must be replaced by an EarlySense authorized

technician. Contact EarlySense technical support.

Note

The system batteries will function when the system is unplugged. In

order to maintain the batteries adequately charged, make sure that

the system is always plugged in to the Electrical outlet during

routine operation.

Sorted Disposal

Do not discard this product. Contact your local authorized representative for additional information for collection and recovery programs available for this product and for appropriate facilities for recovery and recycling

Page 23: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

17

The EarlySense System’s User Interface and Screens

User Interface

Interaction with the user interface is via touch screen. Touching some functions (Such as

the Discharge and Trends buttons) will open additional screens, while other buttons

will open additional adjustment functions in the Home screen (Such as Patient Turn q

and Bed Exit sensitivity level). For any interaction, the screen must be unlocked by

touching the lock image in the upper left corner of the Home screen. If the user attempts

to press the touch screen while the system is locked, the lock in the upper left corner will

show a red flash.

Page 24: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

18

System Messages and Alerts

There are four types of system messages and alerts:

List of messages/alerts in this category Example

High/Low HR

High/Low RR

Patient Signal Lost

Bed Exit

Patient Out of Bed

Extremely High Motion

Patient Turn is OFF – Low Patient

Movement

Patient Turn is OFF – Air Mattress

Motion Detected

Turn Counter Exceeded

Technical Alert – indicates that there is a

technical problem with the system:

Check Sensor

No AC Power

Very Low Battery

System Malfunction

Functional Message - Informative

message.

Screens

The Home Screen displays the monitoring data and provides access to additional menus.

Choices are selected by touching the screen.

Page 25: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

19

1. Lock/Unlock status

2. Nurses details ( Name, Ext.

Number and assigned color

3. Patient details (ID, Room and

Bed numbers)

4. Date/Time

5. Battery status/AC indicator

6. Pause

7. Admit /Discharge patient

8. Enable/Disable heart rate alerts

9. Settable "High" HR threshold

10. Current heart rate

11. Settable "Low" HR threshold

12. Enable/Disable RR alerts

13. Settable "High" RR threshold

14. Current respiratory rate

15.Settable "Low" RR threshold

16. Enable/Disable movement alerts

17. Current movement level (0, Low, Medium, High, Extremely High)

18. Enable/Disable out of bed alert

19. Bed Exit Settings

20. Enable/Disable Patient Turns Counter

21. Patient Turn Counter: Time passed since last turn documented by nurse

22. Patient Turn Protocol Settings (q# - the counter time setting: 1 Hour, 2 Hours, 4 Hours, 6

Hours) indicating the number of hours between each reminder.

Page 26: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

20

23. Clock time of last self movement

24.Operational buttons - to open additional screens

Figure 8: Home Screen

The following menus are accessed from the bottom of the Home Screen.

Menu Use For

Adding an event (See Recording

and Viewing Events, page 50).

Page 27: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

21

Viewing trends (See Trends,

page 51)

Setting or deactivating Alerts

(See Alerts, page 34).

Setting the volume level of the

bedside unit (page 32)

Access advanced functions

Page 28: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

22

4 System Operation

Overview The EarlySense System is designed to display and log respiration rate, heart rate and

movement parameters in an automatic contact-free manner when the patient is lying in

bed. The System is indicated for use in indoor environments. Monitoring of the patient

begins automatically as soon as the patient gets into bed.

The system design allows the user to set high and low threshold parameters that will alert

the user if either heart or respiration rate have crossed predefined thresholds. The system

can alert the user if patient leaves the bed (bed-exit) and allows the healthcare practitioner

to document that s/he changed patients’ position in bed. The health-care practitioner can

insert the event of patient's change of position by documenting the event on the

EarlySense bedside screen. The system then verifies the change in position and logs the

information. The system also provides an option to notify the user if a certain amount of

time (as set by the user) has passed since last position changing (timer expired). The users

can decide to activate or deactivate these notifications

Data is continuously presented on the Bedside Unit main screen. The data obtained

during the monitoring period is recorded by the system in order to enable off-line

presentation and data printing.

The Sensing Unit should be placed under the mattress, underneath the patient’s chest area

and the system should be turned on. After approximately 60-120 sec of the patient lying

in bed in resting position, the system will begin displaying the monitored values on the

bedside Unit Home screen.

By entering the patient’s basic information (MRN, Initials) into the system the data will

be transferred to the Central Display Station (if being used). Heart and respiration rates

and patient movement level will be continuously displayed on screen. Heart and lung

icons will blink and the values will appear in white on the screen as long as

measurements of HR and RR are current.

Page 29: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

23

If the system cannot detect (or measure) the momentary heart or respiration rate for more

than 1 minute, the last measured result of HR or RR will be displayed on screen in grey

and the blinking heart or lung icon will stop blinking.

Positioning the Bedside and Sensing Units The Bedside Unit should be mounted to the wall beside the patient’s bed. The power

cable should be connected to the power outlet and the short “telephone” cable should be

connected to the monitor from one side and to the sensor cable on the other side. Please

make sure the cable is not trailing on the floor.

Place the Sensing Unit under the patient’s mattress, and locate it underneath the chest

area, horizontally. Position the sensor so the top surface is facing upwards (as marked)

(See Figure 9). Position the cord that connects between the Sensor and the Bedside Unit

so that it comes out from under the mattress from the head of the bed, in order to not

interfere with the movement of the patient in and out of bed. If the sensor is placed on a

metal framework that does not allow solid support for the sensor, please consult your

authorized EarlySense representative.

Warning

If you mount the Monitor to the wall, ensure that the wall-mount is not

placed directly above the patient's bed.

!

Caution

The patient should not be in direct contact with the Sensing

Unit. A mattress, mattress pad or mattress cover should

always be placed as a barrier between the Sensing Unit of the

EarlySense System and the patient.

Page 30: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

24

!

Caution

Handle the Bedside Unit with care.

Do not drop, bang or shake the Bedside Unit. Rough handling

can break the internal circuit boards.

Figure 9: Positioning the Sensor and Bedside Units

Turning the System On/Off Turn the EarlySense System "ON" by pressing the On/Off button on the front of the

Bedside Unit (See Figure 5). The Admit Window will automatically appear. After

admitting the patient, the Home Screen will be displayed.

To turn the system off, press the On/Off switch on the front of the Bedside Unit. You will

be asked if you are sure you want to turn the system off. Confirm turning off by hitting

Yes on the touch screen.

Electrical Failure In case of electrical failure, the internal system battery allows the system to function for

approximately 30 minutes.

Check that the problem is not with the system itself by verifying that the power cable is

not damaged and is connected both to the Electrical Outlet and to the Bedside Unit.

Page 31: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

25

The battery icon displayed on the main screen indicates the status of the battery and if the

battery is being recharged.

Figure 10: No AC Power Warning

The battery icon displayed on the main screen indicates the status of the battery as

follows:

Green with a lightning

Battery is charged and

connected to a power outlet

Green

Battery is charged but

disconnected from a power

outlet

Orange

System is disconnected from

power outlet, battery is not

charged.

Page 32: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

26

Red

Battery is almost empty.

System is about to shut

down.

The lightning symbol on the battery indicates that the battery is charging.

After Electrical Power is Restored

1. Make sure that the system is plugged into the Electrical Outlet.

2. Turn the System On by pressing the On/Off switch on the front of the Bedside Unit

(See Figure 5). The System automatically starts operating, as detailed above (see

Turning the System On/Off – page 24).

The internal battery automatically recharges when the electrical power is restored.

If System operation is not restored, contact EarlySense Ltd. customer support.

Patient Information (Admit/Discharge)

Admitting a Patient

The Admit Patient screen appears every time the system is switched on automatically

and can also be found by touching the Admit button in the upper right corner of the

Home screen.

Monitoring can take place without entering a patient ID / MRN; however it is advisable

to enter a new patient ID before starting the monitoring, otherwise the monitored

information will not be saved and the information will not be transferred to the main

nurses’ station.

1. Choose Admit in the Home Screen. The first Admit Patient screen will open (Figure

11).

Note

Only letters can be entered in the first and last name boxes and only

alphanumeric characters can be entered in the ID box.

Page 33: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

27

Figure 11: Patient Admit Screen

2. Insert the patient's name (Letters only), MRN (ID) (Alphanumeric only), room

number (digits only) and bed number (Alphanumeric only). The MRN (medical

record number), Room No. and Bed No. must be entered for the patient to be

admitted. Choose a color to assign the patient to a nurse, in case you are using the

EarlySense Central Display Station (also referred to as “Patient Care Manager”)

3. To advance to the Patient Demographics screen (Figure 12) press Next.

4. To return to the Home screen without admitting the patient press Cancel or Home.

To admit the patient and return to the Home screen press Done. After pressing Done,

two questions regarding the required settings will show (Figure 14: Bed Exit Alert).

Note

The Patient Demographics screen will not open unless a patient

ID / MRN has been entered.

Page 34: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

28

Figure 12: Patient Demographics Screen

The following additional fields can also be entered but are not mandatory:

Gender: Choose M or F.

Date of Birth: Enter the patient’s date of birth (Figure 13).

Figure 13: Date of Birth

Page 35: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

29

Choose Done to finish admitting the patient, or Cancel to return to the Home screen

without admitting the patient. If after 10 minutes you have not pressed a button the

screen will close automatically without admitting the patient.

5. After pressing Done, you will be asked two setting questions. The first is whether

you want to set a Bed Exit alert for this patient (Figure 14). If you choose Yes, the

Bed Exit alert will be enabled automatically with the default setting of sensitivity

level. If you choose No, the Bed Exit alert will be disabled. After answering the first

question, you will be asked whether you want a Patient Turn reminder for this patient

(Figure 15). If you choose Yes, the Patient Turn Counter will be enabled

automatically with the default setting of a reminder every 2 hours, and if you choose

No, the Patient Turn Counter will be disabled. If after 1 minute you have not pressed

a button the screen will close automatically, the patient will be admitted and the Bed

Exit alert and the Patient Turn Reminder will remain disabled.

Page 36: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

30

Figure 14: Bed Exit Alert

Figure 15: Patient Turn Reminder

Page 37: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

31

Discharging a Patient / Ending a Patient Monitoring Session ("Discharge")

Warning

To avoid recording patient data under an erroneous

name you must perform the discharge procedure

before beginning monitoring of a new patient.

When the recording session is completed perform the "Discharge" procedure:

1. In the Home screen choose Discharge. The Confirm Discharge screen will open

(Figure 16)

2. Choose Yes.

Figure 16: Confirm Discharge Screen

3. You will be asked if you want to admit a new patient (Figure 17)

Page 38: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

32

Figure 17: Admit Confirmation Screen

4. Touch Yes to admit a new patient and No to return to the Home screen without any

patient being admitted.

Pausing a Patient Monitoring Session The Pause button stops the recording session temporarily, while the patient’s

identification and trends are saved. At any time the patient’s recording can be resumed,

and all the previous trends will be displayed as usual. Pause should be used when the

sensor is temporarily disconnected from the bedside unit to avoid a blue ‘Check Sensor’

technical alert from sounding, this function should be used for example when the patient

needs to be transported with his/her bed and the sensor needs to be disconnected from the

Bedside unit.

To pause the recording session:

1. Press the Pause button at the Home screen (See Figure 8). A Pause screen will be

displayed (Figure 18).

Page 39: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

33

Figure 18: Pause Screen

2. To continue the recording session, press Resume at the Pause Screen (Figure 18).

Setting the General Volume The General Volume sets the volume of all alert tones. In addition volumes and tones for

individual alerts can also be set (See Alerts Setup

Setting Individual Alert Volume and Tone, page 42). Note that the alert volumes are

determined by which of the two parameters (General or Alert) was set last and that the

default setting for the alerts in the bedside unit is mute. At startup and when admitting a

new patient following discharge, the system returns to default alert settings.

To set the General Volume:

1. Choose the Volume button from the Home screen. The General Volume box

opens (Figure 19)

Figure 19: General Volume Box

Page 40: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

34

2. Push the Plus button to raise the volume or the Minus button to lower the volume.

3. Click OK to confirm

The Volume icon is marked with an X if the volume is set to zero.

Figure 20: Alerts Volume Off Status

Alerts Alerts are used to notify personnel of situations that may require attention. The

EarlySense System enables the user to define upper and lower thresholds for the heart

rate, respiration rate, time out of bed and patient turn. If any of the parameters, as set by

the user, cross these specified thresholds, an audible and visual alert notification is

triggered.

Page 41: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

35

Alerts Table

The system uses the following alerts:

Message Color Description

High/Low Heart Rate:___ Red Heart Rate exceeds thresholds.

High/Low Resp. Rate:___ Red Respiration Rate exceeds thresholds.

Bed Exit

Sensitivity Settings:

1 (Low) – 6 (High)

Red Bed Exit

Patient Out of Bed

For ____ Min

Red System detected that patient is out of bed for a time

interval exceeding a preset threshold.

Patient Signal Lost Red System cannot detect patient’s signal.

Check patient and sensor location.

Extremely High Motion Yellow Patient Movement Level has been extremely high

during the last minute.

Patient Turn is OFF

Low Patient Movement

Yellow Patient Movement Level has been very low during

nighttime (00:00-6:00). Consider setting Patient Turn

reminder ON.

Patient Turn is OFF

Air Mattress Motion

Detected

Yellow Air Mattress was detected on the bed. Consider setting

Patient Turn reminder ON.

Page 42: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

36

Turn Counter Exceeded

____ Hours

Yellow Threshold time interval for turning the patient expired.

Check Sensor

Reconnect/Replace Sensor

or Restart

Blue Technical alert indicating that the sensor connection is

malfunctioning.

Try to reconnect the sensor and if problem persists

replace the sensor (See Troubleshooting page 66).

No AC Power

Operating on Battery

Backup

Blue An alert indicating that the EarlySense Bedside unit is

disconnected from the electrical power outlet.

Reconnect the EarlySense bedside unit to the

electricity Mains – otherwise the bed-side will

shutdown (See Troubleshooting page 66).

Very Low Battery

System Shutting Down

Blue An alert indicating that the battery is low, and system

is about to shut down.

Reconnect the control unit to the electricity Mains (See

Troubleshooting page 66).

System Malfunction

Please Contact Service

Engineer

Blue

Technical alert indicating that the system is

malfunctioning. User should contact a service engineer

(See Troubleshooting page 66).

Multiple Alerts

Multiple alerts occur when more than one alert is triggered at the same time. For instance

if a blue system malfunction occurs while an elevated HR is in progress, this situation is

defined as a multiple alert.

There can only be one active alert at a time. If an additional alert occurs while an alert is

already active, the system will then determine which alert will be active according to the

following priority:

• Blue

• Red

Page 43: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

37

• Yellow

A new alert will become active only if the currently active alert has a lower priority.

Alert Triggers

Default alert status is:

Alert Status

Heart Rate ON

Respiratory Rate ON

Movement ON

Patient Turn counter OFF

Bed Exit OFF

Default alert threshold values are:

Low limit threshold High limit threshold

Heart Rate 40 BPM 130 BPM

Respiratory Rate 8 Br/min 32 Br/min

Patient Turn counter

(default)

2 hours

Bed Exit (default) Immediate – sensitivity 3

Alerting Parameters

Low Heart Rate High Heart Rate

Time to alert for change

in HR

Max. 60 seconds Max 60 seconds

Page 44: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

38

Low Respiration Rate High Respiration Rate

Time to alert for

change in RR

Max. 150 seconds Max. 120 seconds

Modifying Alert Thresholds

When beginning to monitor a new patient, it is advisable to set patient specific alert

thresholds. If relevant for a specific patient, Bed Exit alert sensitivity or duration and

Patient Turn interval should be set from the relevant section in the home screen next to

each function setting (see pages. 44 and 45). To change the default HR and RR

thresholds:

1. Touch the upper or lower thresholds in the HR or RR boxes in the Home screen

2. Touch the down-pointing arrow to lower the threshold or the up-pointing arrow

to raise the threshold (Figure 21)

3. When you reach the desired threshold touch the number displayed, or wait.

Figure 21: Adjusting the HR Upper Threshold

Page 45: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

39

Thresholds can be set in the following ranges:

Lowest settable limit

threshold

Highest settable limit

threshold

Heart Rate 35 BPM 150 BPM

Respiratory Rate 8 Br/min 44 Br/min

Audible and Visual Alert Notification

When an alert is triggered an audible alert sounds and the corresponding alert parameter

on the monitor main screen will be shown with its background color corresponding to the

alert color, either red or yellow (Blue alerts are not connected to a specific monitoring

parameter). A message will appear on the bottom of the window detailing the nature of

the Alert (Figure 22) and the time passed since alert was raised

Figure 22: Alert Screen

Blue Alerts

Blue alerts are managed differently than yellow or red alerts. Blue alerts cannot be

suspended; their volume is set at maximum and is non-adjustable. Blue alerts appear only

Page 46: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

40

with an “OK” button and remain active as long as the alert condition exists. They

disappear only after the technical issue is resolved. The rest of the information in this

section relates to red and yellow alerts only.

Suspending Red and Yellow Alerts

Press the “OK” button to close an active alert (Figure 22). Notice that as long as the alert

condition exists, the alert will show up again. To close an active alert and temporarily

disable for 15 minutes - all red and yellow alerts, press the 15 min Suspend Alert button

(Figure 22). To cancel the 15 min suspension, press twice on the external “Suspend alerts

for 15 min” button (See Figure 5).

Disabling and Enabling Alerts

When beginning to monitor a new patient, it is advisable to set the specific alerts that

should be enabled for this specific patient. By default all alerts apart from the Bed Exit

and Patient Turn alerts are enabled. Per each patient’s condition, the user is advised to

enable Bed Exit or Patient Turn alert as required.

To disable alerts for specific parameters touch the bell symbol next to the alert in either

the Home screen or the Alerts Setup screen. To enable alerts touch the bell symbol

again.

If an alert is disabled while the alert is active, the alert will remain active until it is

suspended by the user.

When the Bed Exit alert is being disabled, the question “Are you sure you want to set bed

exit alert off?”(Figure 23) To disable the alert, choose Yes.

Page 47: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

41

Figure 23: Bed Exit Off

When the Patient Turn Counter alert is being disabled, the question “Are you sure you

want to set patient turn alert off?” (Figure 24) To disable the alert, choose Yes.

Figure 24: Patient Turn Off

In case the alerts are disabled for a specific parameter the bell icon for the corresponding

parameter in the main and alerts setup screens will be marked with an X as follows:

Page 48: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

42

To suspend all red and yellow alerts for 15 minutes, press the hardware Suspend alerts for

15 min button (Figure 5). Press again the same button to enable alerts.

Alerts Setup

Setting Individual Alert Volume and Tone

Press the Alerts button on the bottom of the Home screen. The Alerts Setup screen will

open (Figure 25). When entering the Alerts Setup screen the current status of all alerts,

as shown in the following list, is displayed:

• Alert volume

• Tone

• Enabled/disabled

Figure 25: Alerts Setup Screen

• Click the plus or minus button to raise each alert volume

• Press buttons 1, 2 or 3 to choose a Tone.

Page 49: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

43

• After setting up the alerts, touch Submit to activate these settings or Cancel

to preserve the existing settings.

Note

In addition to changing the individual alert volumes as described

above, you can also change the general volume (See page 32) or by

setting volume and tone of all in the alerts setup screen. The alert

volumes are determined by which of the two parameters (General

or Alert) was set last.

Note

When a patient is admitted or discharged all alert settings will

return to default.

Warning

The Audible alert in the EarlySense System is not intended for

monitoring high risk situations where ECG monitoring is required.

The most reliable method of patient monitoring combines close

personal surveillance with correct operation of the monitoring

equipment

Silencing Alerts

A green bell icon indicates that the alert is active.

Silence alerts volume by setting the alerts volume to zero in either the general alerts

volume icon in the Home screen or in the Alerts Setup screen.

In case the alerts are silenced for a specific parameter the bell icon for the corresponding

parameter in the main screen will be displayed with a mute sign as follows:

Bed Exit The EarlySense System tracks if there is a Bed Exit. This alert is disabled by default.

Page 50: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

44

Note

If the patient is allowed to get out of bed without the nurse being

notified, you can disable the Bed Exit alert by touching the bell in

the In Bed section of the Home Screen.

Configuring the Bed Exit Alert

Bed Exit alert can be configured in terms of alert volume and tone (See Setting Individual

Alert Volume and Tone, page 42) and sensitivity.

Setting Bed Exit Sensitivity

Press the Bed Exit Settings button on the lower right of the Home screen. The Bed Exit

Settings Area will open (Figure 26).

Figure 26: the Bed Exit Settings Area

The Bed Exit Alert can be set either to provide an immediate alert when the patient exits

the bed (varying in 6 levels of bed exit sensitivity levels) or to provide an alert when the

patient has been out of bed for a period ranging from 3-20 minutes.

Use the arrows to choose the desired delay or sensitivity setting. After 3 sec the arrows

will disappear and the change will be applied.

• If you want an immediate alert when a patient exits the bed, click the arrows

to choose a sensitivity ranging from 1-6, with 1 being the least sensitive and

6 being the most sensitive. Note that sensitivities 5 and 6 are highly sensitive

and might be triggered due to excessive movements of a patient while in bed.

• If you want the system to alert after a period of time that the patient is out of

bed, choose the time duration you would like between when the patient

leaves the bed, and when the alert sounds. Use the same set of arrows that

you use to set the sensitivity to select the number of minutes for the delayed

Bed Exit alert.

The Bed Exit alert is disabled by default. Enable the Bed Exit alert by pressing the bell

icon.

Page 51: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

45

The active mode of the Bed Exit alert is indicated by the bell icon at the Bed Exit section

of the Home screen (Figure 27).

Figure 27: Bed Exit Box

Patient Turn Patient Turn is used to document change of the patient’s positioning in bed and to

produce an alert if the interval set for changing the position has been exceeded (timer

expired). The Patient Turn alert is also disabled by default.

Setting the Patient Turn Interval

The turn interval is referred to as "q". For instance if you would like the system to

generate an alert as a reminder to turn the patient every 2 hours, set q to q2.

To change the time interval between the turn reminders touch the q button in the Patient

Turn box (Figure 28) in the Home screen (Figure 8). Using the arrows, choose the

relevant q number (number of hours). After 3 sec the arrows will disappear and the

change will be applied.

Figure 28: Setting Patient Turn Interval

Starting the Patient Turn Counter

To enable and disable Patient Turn counter touch the Bell Icon in the Patient Turn

section that is in the Home screen. The counter will become enabled (Figure 30). To

disable the counter press the Bell icon again (Figure 29).

Page 52: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

46

Figure 29: Patient Turn Box – Disabled Counter

Figure 30: Patient Turn Box – Enabled Counter

Documenting Patient Turn and Resetting Counter

After turning the patient, press Patient Turned within 1 minute from the time that the

patient turn was performed. The system logs the event (Figure 31). In addition, the

system detects the corresponding change in body movement and verifies the event. The

verification will appear in the printed report of the same day (See Reports, page 59)

Figure 31: Recording Patient Turned Event

If the patient turn interval passes without the Patient Turned button being pressed by the

user, a Patient Turn alert will occur (Figure 32) indicating that the counter has exceeded

its time threshold. The alert message will disappear by pressing Ok or by documenting a

patient turn by pressing Patient Turned. If Ok was pressed, but the Patient Turned

button was not pressed within 15 minutes, the alert will go off again.

Page 53: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

47

Figure 32: Patient Turn Alert Screen

Air Mattress

If the Patient Turn alert is turned off and the system detects an air mattress in operating

condition, the following alert will appear (Figure 33).

Figure 33: Air Mattress Alert

After you press “OK” you will be asked if you would like a patient turn reminder for the

patient (Figure 34Error! Reference source not found.).

Figure 34: Patient Turn Reminder

In addition, if an air mattress is not detected and a Turn Counter Exceeded alert is

generated, you will be suggested to consider using an air mattress.

Page 54: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

48

Patient Movement Movement rates are determined in terms of percentage of movement time within a

defined period of time as described in the following table. The accuracy of the system in

displaying different rates is also presented below.

Rate Percentage of movement

during defined period

Accuracy of rates

displayed

(For adults)

Accuracy of rates

displayed (For children)

0 0 100% 100%

L Up to 40% 100% 100%

M 40-60% 81% 81%

H 60-80% 100% 86%

EH 80-100% 96% 94%

Patient movement over time can be viewed in Trends (See Trends, page 51).

Low Patient Movement Alert

Alerts for movement are displayed in cases of low and extremely high movement. An

alert for Low movement (Figure 35) may come up once a day at 6:00 AM if during the

night (00:00-6:00) the amount of patient movement or activity level was below a certain

threshold, which might imply the patient is in risk for developing pressure ulcers.

Following dismissing the alert of low movement, you will be asked if you want a Patient

Turn reminder for this patient (Figure 34).

Page 55: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

49

Figure 35: Low Patient Movement Alert

Patient Motion Detection

In case that the Patient Turn Reminder is being used, and the system detects additional

patient movement during the night – following the first Turn Reminder in the morning

you will be suggested to consider re-evaluating Braden score (Figure 36).

Page 56: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

50

Figure 36: Excess Patient Movement Detected

Recording and Viewing Events Occurring events can be noted by the user and entered into the system for documentation

purposes. These events will then appear in the Trends display.

Manually Recording an Event

1. To access the Events menu click Events in the Home Screen. The Events

menu appears (

2. Figure 37).

Figure 37: Events Menu

Page 57: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

51

3. Touch one of the four primary events on the left or secondary events on

the right. The button changes color showing that it has been successfully

chosen. The following message appears: Recording Event: <Event

Name>

4. Touch the Home button to confirm your choice and close the screen or

Cancel to cancel your choice.

The following Events are available:

Primary:

• Emerg. Med. – Emergency medication given when patient is in distress

• CPAP/BPAP

• Intubation

• Patient Fall

Secondary:

• Continuous Med. – routine medication

• Pain Relief Med. – pain relief medication

• Patient Agitation

• Skin Breakdown – a situation that may lead to pressure ulcer development

• General Flag 1 – can be used in order to mark any desired event of type 1 which

is not included in the list of events

• General Flag 2 – can be used in order to mark any desired event of type 2 which

is not included in the list of events

Trends Trends are displayed in a graph format.

Trends Graph

To view the Trends Data click the “Trends” button in the lower section of the Home

Screen.

To advance or go back in the time frame press the forward or back arrows.

Page 58: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

52

Figure 38: Trends Data Screen

Number Name Description

1 Current heart rate Current heart rate in beats per minute

2 Events Displays events that users have manually entered (for

documented) during the time displayed

3 Respiratory rate Current respiratory rate in breaths per minute

4 Heart rate trends Displays heart rate in beats (graphs) per minute during the

time displayed.

5 Respiration rate

trends

Displays respiration rate (graphs) in breaths per minute

during the time displayed.

Page 59: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

53

6 Scroll back Touch here to scroll back.

7 Alerts Displays alerts that occurred during the time displayed

(See Alerts, page 34).

8 Time scale Time scale for data presented

9 Scroll forward Touch here to scroll forward.

10 Movement trends Shows patients movement level while in bed during the

timescale presented (See Patient Movement, page 48).

11 Out of bed Shows the time intervals when patient has been out of bed

(Bed Exit, page 43)

12 In bed Shows the time intervals patient spent in bed

13 Time period buttons

(10 min, 1 hour, 3

hours, 8 hours, 1 day)

Touch one of these buttons to change the time period

displayed in the graph.

14 Analysis screen Open the Analysis screen (Figure 39).

Page 60: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

54

Alert Symbols

The following symbols are used in the Trends graph:

Icon Meaning

Blue Alerts —Technical Alerts: System Malfunction, Sensor Malfunction, Low

Battery.

High/Low Heart Rate – Red Alert

High/Low Respiration Rate – Red Alert

Movement Level High/Low – Yellow Alert

Patient Turn Counter – Yellow Alert

Out of Bed – Red Alert

Signal Lost – Red Alert

Page 61: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

55

Event Symbols

When entered by the user, the following Event symbols are used in the Trends Graph:

Intubation

Patient Fall

Emergency Medication

Continuous Medication – Routine Medication

Pain Relief Medication

Patient Agitation

Skin Breakdown

General Flag 1 - Can be used in order to mark any desired event of type 1

which is not included in the list of events

General Flag 2 - Can be used in order to mark any desired event of type 2

which is not included in the list of events

Patient Turn – Inserted automatically if a staff member pressed the RESET

on patient turn counter and the system detected that patient turn occurred.

Page 62: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

56

Setting the Time Scale

Upon entering the Trends screen, the default time scale displayed is 3 hours. You can

return to this time scale at any time by touching the Default button. Use the buttons

under the graph to determine the time scale displayed. If you choose a larger time scale

the current time that is in the center of the time scale will remain the center. If you choose

a shorter time scale you will be prompted to touch the screen to determine where you

would like to center the displayed timescale.

Analysis

The Analysis function allows you to see the average respiration, heart rate and movement

time percentage for a specified interval each day. Touch the Analysis in the trends screen

to open the Analysis screen (Figure 39).

Figure 39: Analysis Screen

Use the arrows to set the time slot you want, and then click Show Data. The data will

appear in the table.

Page 63: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

57

Settings To access the settings click Settings in the Home Screen. The Settings screen has three

tabs: Brightness, Setup and Maintenance.

Brightness

In the Brightness tab (Figure 40) Press the “-” to darken the screen, and press the “+” to

brighten it.

Please note that if the brightness level is set to the lowest level, the screen might be seen

as black in daylight. Touching the screen while brightness is set to the lowest level will

change the brightness to a higher level, in order to be able seeing buttons and labels and

enabling adjustment of brightness to higher level manually.

Figure 40: Brightness

Choosing a Language

In the Setup tab (Figure 41) choose the desired language to be used in the system

interface and click Done.

Page 64: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

58

Figure 41: Setup Tab

Maintenance

This mode should be used only by a technician authorized by EarlySense Ltd.

Entering the technician mode requires a password. To access the Technician mode, press

the Maintenance tab (Figure 42).

Figure 42: Maintenance Tab

Page 65: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

59

Reports

Generating Reports (Data Retrieval)

Data analysis may be performed on-line during the monitoring session based on the

bedside unit display or reports.

Reports to be printed may be downloaded at the bedside unit or at the Central Display

Station (See EarlySense Central Display Station manual). To ensure the privacy of the

patient's medical data a PIN must be entered before reports can be downloaded.

The data can be downloaded as CSV file and as a complete report in Acrobat PDF

format. In case of CSV file the data can be viewed by the physician using standard

available software e.g., Microsoft Excel. The data is presented numerically; any other

desired graph can be generated by the physician, based on the data. The PDF report

includes both numerical and trend vs. time graph that can also be printed and includes

information about Heart and Respiratory rates as well as patient's movement, in/out of

bed status, alerts history and patient turns (See Figure 45 for an example).

Warning

The data acquired by the EarlySense System should be

interpreted by a healthcare practitioner only

Downloading Patient Data and Reports

1. Insert a flash drive with at least 0.5 GB of free space into the USB slot. (If the

flash drive is already inserted it must be removed and re-inserted before

commencing the data download procedure).

2. Press the Download Report button (See Figure 5) on the front panel.

3. You will be prompted to insert a PIN (Figure 43) and click Submit. A message

will appear on the bottom of the screen informing you that downloading has

commenced (Figure 44)

Page 66: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

60

Figure 43: PIN Window

Figure 44: Downloading Message

4. The system will automatically download the data (CSV file) and the report for

(.PDF) file of the last 2 admitted patients.

5. A message is displayed when the download is completed.

Page 67: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

61

Page 68: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

62

Figure 45: Example of Printed Report (PDF File)

Page 69: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

63

5 Specifications, Maintenance and Troubleshooting

Specifications

Physical Characteristics

Sensing Unit Bedside Unit

Dimensions

300 x 210 x 2.5 mm 250 x 260 x 110 mm

Weight

160 g 3.0 kg

Materials

ABS + Poly carbonate ABS + Poly carbonate

Page 70: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

64

Performance

Respiration Rate Heart Rate Movement

Range 6 – 45 Br./min 30 - 170 BPM* 0, L, M, H, EH

Averaging Period

1 Min. 1 Min. 15 seconds

Accuracy

±4% or ±1.5 Br/min, whichever is

greater

±4% or ±5 BPM,

whichever is greater

• The System detects heart rate that is > 1.8 times the respiration rate

• Total system accuracy including undetected signals 90% for RR and for HR. The total system

accuracy was measured as +/- 10% of the predicate device

Alert Thresholds

Default: Low=8 Br./min ; High=32

Br./min

Default Alert: Low=40

BPM ; High=130 BPM

Default: Low = 0%

High = "Extremely-High"

movement > 1 minute

Min- Max settable Alert Thresholds

Low: 8 Br./min ;

High: 44 Br. / min

Low: 35BPM

High: 150 BPM

Electrical

Voltage Input Range 100-240 VAC, 50-60 Hz, 0.9A max Battery Rechargeable battery Fuses 2 x 1 Amp Isolation Medical grade isolation between the Bedside Unit and the power supply

Operating Conditions

Relative Humidity 30% - 80% non condensing Ambient Operating

Temperature Bedside Unit: 10- 35 °C (50-95 °F) Sensing Unit: 5- 40°C (41-104 °F)

Storage and Transportation Conditions

Temperature Range 0-50 °C (32-122 °F) Relative Humidity 10%-90% non condensing

Page 71: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

65

Maintenance and Cleaning The EarlySense System is designed to provide a trouble-free and maintenance-free

operation. The only maintenance required is to keep the System clean and dry, and to

verify that there is no physical damage to the Bedside Unit and to the Sensing Unit.

The sensing unit should be replaced on an annual basis to ensure proper functionality of

the system.

Cleaning and Disinfection of the Bedside Unit

Gently wipe the exterior of the Bedside Unit using soft, slightly damp cloth / wipes

containing anti-septic substances. Wipes containing alcohols, Chlorohexidine, and bleach

material with concentration up to 5% can be used. Avoid excessive liquids.

Cleaning and Disinfection of the Sensing Unit

Gently wipe the Sensing Unit using a soft, slightly damp cloth / wipes containing anti-

septic substances. Wipes containing alcohols, Chlorohexidine, and bleach material with

concentration up to 5% can be used). Please ensure that the sensor is dry before re-use.

Avoid excessive liquids.

!

Caution

Avoid placing liquids or food on any part of the System. Do not allow

conductive fluids to leak into the active circuit components of the System

as this may cause a short circuit, which could result in an electrical fire. In

such an event, only fire extinguishers approved for use on electrical fires

should be used.

!

Caution

• Never open the Bedside Unit housing as this may damage the system.

• Handle the Bedside Unit with care. Do not drop, knock, or shake the

Bedside Unit. Rough handling can damage internal circuit boards.

Page 72: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

66

Troubleshooting

Problem Possible Cause Solution

The system is

functioning but the AC

icon is not displayed on

the battery icon

System is unplugged

from the Electrical

Outlet

Reconnect the unit to the Electrical Outlet.

Screen is dark but LEDs

are on

The screen was set to

be darkened

or

Technical problem

Touch the screen once, it will become brighter. To

changes the brightness press the button Settings (see

page no. 57).

Turn off the system by pressing the On/Off switch for

10 seconds, and then turn the system back ON by

briefly pressing the On/Off switch. If problem is not

resolved call your service engineer

Blue Alert is On

"System Malfunction”

Message

System Malfunction Contact a service engineer.

Blue Alert is On

"Check Sensor"

Message

Bedside Unit has

detected that the

sensor is not

connected

• Check if the sensor is connected to bedside Unit.

Reconnect if necessary.

• If problem persists, replace sensor

• If problem persists, restart system

• If problem persists, call your service engineer

Blue Alert is On

"Low Battery” Message

System is unplugged

and battery is

discharging

• Reconnect Bedside Unit to the Electrical Outlet.

Bedside Unit is not

responding (ON /OFF is

not possible)

No electrical power

as system is

unplugged and

battery is discharged

or,

System error.

• Check connection to the electrical Power

• Reconnect if necessary.

• If problem persists, call the service engineer.

The Bedside Unit is

operating but no HR or

RR is displayed.

Difficulty in

detecting a signal

• Check sensor location. The sensor should be

placed under the mattress below the patient's chest

area, horizontally.

• Reposition the sensor if it moved from the

specified location. If the problem persists, replace

the sensor.

Page 73: Doc. No. MK600 SUBJECT: EarlySense User Guide – USA … · Doc. No. MK600 SUBJECT: EarlySense User Guide – USA (FDA) ... (Admit/Discharge) ... The system incorporates an off-the-shelf

67

6 European Representative and Contact Information

Contact EarlySense at:

European Representative MedNet GmbH Borkstrasse 10 48163 Münster, Germany Phone +49 251 322660 Fax +49 251 3226622

Israel Headquarters EarlySense Ltd. 12 Tzvi St., Ramat Gan 52504, Israel Website: www.earlysense.com

USA Headquarters EarlySense Inc. 135 Beaver Street,

Suite 211, Waltham, MA 02452

Phone: (781) 373 3228 E Mail: [email protected] Support Telephone:

East Coast - (617) 800-6668

West Coast - (818) 370-7748