FUJIFILM DR-ID 300 Console : DR-ID 300CL Service Manual FUJIFILM DR Document No. 010-305-07E First Edition : Aug. 20, 2009 7th Edition : Sep. 12, 2012 Printed in Japan

DR-ID_300CL-07E pg 494

Embed Size (px)

Citation preview

Service Manual
FUJIFILM DR
Document No. 010-305-07E First Edition : Aug. 20, 2009 7th Edition : Sep. 12, 2012
Printed in Japan
010-305-06E 03.31.2012 FM6000DR-ID 300CL Service Manual
Export Restrictions Applying to this Equipment Japanese export restrictions and foreign exchange regulations apply to this equipment. Therefore, government approval obtained through due procedure is required in order to export or otherwise remove such equipment from Japan.
Note that the same restrictions and regulations apply not only to the main unit but also to after-sale parts, service manuals and service training (for exporting or transferring technology) when such items are dispatched or shipped to foreign countries and when service training is given for trainees from foreign countries.
Caution : RX only in the United States. (Federal law restricts this device to sale by or on the order of a physician.)
Precautions in Using the Service Manual This manual should not be used by anyone other than the servicing staff members qualified through the training specified by our company. Since this manual contains confidential information including engineering know- how related to product development, it is prohibited to distribute the manual to users or to allow them free access to the manual without obtaining prior permission from FUJIFILM Corporation.
This manual is intended to assist the service person in the safe and efficient servicing of the equipment described. The equipment must be used in accordance with the procedures contained in this manual and must not be used for purposes other than those which are described herein.
The equipment should only be used by persons having recognized qualifications and, if relevant, having adequate training on this equipment, especially regarding protective measure such as laser radiation protection.
It is the responsibility of the user to ensure that existing legal regulations regarding installation of the building with respect to the equipment are observed. Moreover the legal regulations regarding the operation of the equipment are to be observed.
Incorrect operation, or failure of the user to maintain the equipment in accordance with the schedule of maintenance, relieves the manufacturer or his agent from all responsibility for consequent non-compliance, damage, injury, defects and/or other malfunction.
FCR is a trademark or registered trademark of FUJIFILM Corporation.
Copyright © 2009-2012 by FUJIFILM Corporation. All rights reserved.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by means, electronic, mechanical, photocopying, recording or otherwise, without the prior written permission of FUJIFILM Corporation.
010-305-07E 09.12.2012 FM6055DR-ID 300CL Service Manual
WARNING
This manual is written for International distribution (including distribution in the U.S.) and therefore may contain references to mammography related options. Mammography functionality for this product is not commercially available in the US, pending FDA Approval. This functionality must not be activated on US models, the only breast related functionality available in the U.S. is the Specimen Imaging Application.
Im portant N
anual
BLANK PAGE
How to Use This Manual
Notes on the Use of the Service Manual
1. This document is intended to be read by the service personnel of FUJIFILM Corporation or authorized agent who is commissioned by FUJIFILM Corporation to perform servicing tasks.
Under these circumstances, no one is allowed to perform the following actions without the prior written permission from FUJIFILM Corporation Duplicating or reproducing this document in whole or in part
Disclosing this document, in whole or in part, to a person other than the authorized service personnel
Lending or transferring this document, in whole or in part, to a third party
2. Due care must be exercised when handling this document because it contains product internal structure descriptions and other confidential information. * When the contents of this document need be changed, the
information about such changes will be conveyed using “Engineering Change Notice (ECN)”.
The “Engineering Change Notice (ECN)” must also be handled using the same care as for this document.
3. Due to a product development schedule, the product specifications and available options/variations prevailing at the time of the release of this document may differ from the descriptions given in this document. In such an instance, a Service Engineering Information will be issued as needed to furnish latest information about inapplicable specifications and components.
4. If you have any doubt about the contents of this document or find errors, contact your local FUJIFILM representative.
Warning and Cautionary Information Precautions during the service work are indicated by the following terms and symbols.
WARNING
Serious injuries or death may result if this precaution is not observed. Damage of the unit and customer database may also occur.
CAUTION Mild or moderate injuries may result if this precaution is not observed. Serious malfunction (which may be unrepairable or difficult to repair) may also occur.
IMPORTANT Abnormal operations (easy to repair) may result if this precaution is not observed.
NOTE Indicates items which require particular attention.
REFERENCE Explanation of related items and terms.
DR-ID 300CL Service Manual010-305-03E 02.28.2011 FM5818 2
Servicing Instruments and Tools That Require Inspection/Calibration
The machine should be installed and serviced by use of servicing instruments and tools that have been regularly inspected and calibrated as appropriate. If the machine were serviced using servicing instruments and tools that have not been inspected and calibrated, proper performance of the machine could not be guaranteed. Servicing instruments and tools that require inspection/calibration are as listed below. The inspection/calibration should be planned and performed in accordance with the specifications and instruction manuals of the applicable servicing instruments and tools.
Instruments and tools that require inspection/calibration
Name Inspection Calibration
Digital multimeter
Screen Displays According to Different OSs This manual uses either of the Windows Vista or Windows 7 screen displays. When setting procedures to be performed and displayed windows are different between Windows Vista and Windows 7, follow the procedure specific to the used OS to perform necessary settings.
DR-ID 300CL Service Manual
010-305-02E 04.15.2010 FM5668 3
Safety Precautions
3. Caution for Handling Personal Information .............. 10
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 4
Product Specifications
1. Equipment Specifications .......................................... 1 1.1 Product Code ......................................................... 1 1.2 Optional Components ............................................. 1 1.3 Dimensions/Weight ................................................. 4 1.4 Environmental Conditions ....................................... 5 1.5 Electrical Specifications .......................................... 5 1.6 Applicable Standards .............................................. 6 1.7 Network Cable ........................................................ 6 1.8 Equipment Disposal ................................................ 6
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 5
1. System Configuration ................................................. MD-1 1.1 Desktop PC ............................................................ MD-1
1.1.1 System Configuration ...................................... MD-1 1.1.2 Operating Environment .................................... MD-2
1.2 Laptop PC .............................................................. MD-3 1.2.1 System Configuration ...................................... MD-3 1.2.2 Operating Environment .................................... MD-3
2. Connectors and Slots ................................................. MD-4 2.1 Dell OptiPlex 760 MT .............................................. MD-5 2.2 Dell OptiPlex 780 MT .............................................. MD-6 2.3 Dell OptiPlex 780 DT .............................................. MD-7 2.4 Dell OptiPlex 790 DT .............................................. MD-8 2.5 Dell OptiPlex 960 MT (for Mobile Model) .................. MD-9 2.6 HP Compaq 6200 Pro SF ........................................ MD-10 2.7 HP Compaq 8200 Elite MT ...................................... MD-11
3. System Connection Example ..................................... MD-12 3.1 System Connection Example (for Desktop PC) ......... MD-12 3.2 System Connection Example (for Laptop PC) ........... MD-13
4. Software Function Overview ...................................... MD-14
5. Relationship between FPD and DR-ID 300CL ........... MD-20 5.1 Software (AP/OS) and Data Flow ............................ MD-20
6. Relationship between a Cassette-type Image Reader and the DR-ID 300CL ... MD-21 6.1 Software (AP/OS) and Data Flow ........................... MD-21
7. Cluster Connection Function Overview .................... MD-22
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 6
MT: Machine Troubleshooting
1. Error Code Table ......................................................... MT-1 1.1 Referring to Error Messages .................................... MT-1 1.2 Precautions ............................................................ MT-4 1.3 Error Message ........................................................ MT-4
2. Using the Event Viewer .............................................. MT-5 2.1 Overview of the Event Viewer .................................. MT-5 2.2 Starting the Event Viewer ........................................ MT-5 2.3 Configuration and How to View the Event Viewer ..... MT-6 2.4 Details of Events ..................................................... MT-7 2.5 Saving the Event Log .............................................. MT-8 2.6 Deleting the Event Log ............................................ MT-9 2.7 Quitting the Event Viewer ........................................ MT-9
3. Information Useful for Error Recovery ...................... MT-10 3.1 Precautions to be Observed During
the Use of Options .................................................. MT-10
4. Troubleshooting (for DELL PC).................................. MT-11 4.1 Trouble Diagnosis Tool (Dell Diagnostics) ................ MT-12
4.1.1 Starting the Trouble Diagnosis Tool (Dell Diagnostics) ............................................ MT-12
4.1.2 Running the Trouble Diagnosis Tool (Dell Diagnostics) ............................................ MT-13
4.2 Checking the Hard Disk .......................................... MT-17 4.3 Resetting the Refresh Frequency ............................ MT-18 4.4 Resetting the Touch Panel Sound ........................... MT-20
4.5 Changing the Barcode Reader/Magnetic Card Reader Driver Software ....................................................... MT-21
4.6 Confirmation Required When Beeps Sound at PC Startup (For Dell OptiPlex 760/780/790) .............................. MT-25
5. Troubleshooting (for HP PC) ...................................... MT-26 5.1 Checking the Memory ............................................. MT-27 5.2 Setting the BIOS ..................................................... MT-31
Appendix A DR-ID 300CL Error Message Table (V4.0 or Earlier) .............................. Appx MT A-1
Appendix B DR-ID 300CL Error Message Table (V5.0 or Later) ................................ Appx MT B-1
Appendix C Cooper Application Error Message Table...................... Appx MT C-1
Appendix D Precise Enlargement Function Error Message Table...................... Appx MT D-1
Appendix E Free Layout Print Error Message Table...................... Appx MT E-1
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 7
1. Reinstalling the OS and Application ....................... MC-1
2. Installing and Setting Up the OS ............................. MC-2 2.1 Installing and Setting Windows Vista ...................... MC-3
2.1.1 Installing and Setting Windows Vista ................ MC-3 2.1.2 Installing the Driver Software .......................... MC-18 2.1.3 Installing Windows Vista SP2 .......................... MC-24 2.1.4 Setting the OS .............................................. MC-25
2.2 Installing and Setting Windows 7 ........................... MC-48 2.2.1 Installing and Setting Windows 7 ..................... MC-48 2.2.2 Installing the Driver Software .......................... MC-64 2.2.3 Setting the OS .............................................. MC-72
3. Setting Up the Monitor .............................................. MC-99 3.1 17" Monitor with Touch Panel ................................ MC-99 3.2 2M Color LCD Monitor
Radiforce RX211/RX220/RX240 ............................ MC-99
4. Installing the Database Management System ........ MC-100 4.1 For Software V4.0 or Earlier .................................. MC-100 4.2 For Software V5.0 or Later .................................... MC-101
4.2.1 Installing .NET Framework ............................. MC-101 4.2.2 Applying Windows Installer ............................. MC-102 4.2.3 Installing SQL Server ..................................... MC-102
5. Installing the DR-ID 300CL-AP ................................. MC-104 5.1 Installing the AP .................................................... MC-104 5.2 Replacing the Menu Database ............................... MC-106 5.3 Installing Correction Parameters (V4.0 Only) .......... MC-108
5.4 Installing the RAID Status Display Tool (Only for RAID Configuration in V5.0 or Later) ........ MC-109
6. Installing Various Components ............................... MC-111 6.1 Built_in Console (V4.0 or Earlier) ........................... MC-111 6.2 FRIS Modules ....................................................... MC-112
7. Installing the DR-ID 300CL Standard Key ............... MC-114 7.1 Installing the DR-ID 300CL Standard Key
(V4.0 or Earlier) .................................................... MC-114 7.2 Installing the DR-ID 300CL Standard Kit
(V5.0 or Later) ...................................................... MC-115
9. Option Key Installation ............................................. MC-118
10. AP Key Installation .................................................... MC-120
11. Initializing the Image Database ................................ MC-121
12. Display Optimization (LUT file settings) ................. MC-123
13. Restoring the Config ................................................. MC-124
14. Settings for the FPD/Cassette-type Image Readers ... MC-128
15. Firewall Settings ........................................................ MC-129
17. Setting the Event Log ............................................... MC-138
18. Checking the Virtual Memory Size (Windows Vista Only) ................................................ MC-140
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 8
Appendix A Replacement Procedure of Damaged RAID Configuration HDD ... Appx MC A-1
1. Confirming Which HDD is Damaged ........... Appx MC A-1 2. Replacing the HDD ...................................... Appx MC A-3 3. Rebuilding RAID .......................................... Appx MC A-3 4. Verification ................................................... Appx MC A-5
Appendix B Setting the RAID Status Display Tool ....... Appx MC B-1
1. Setting the Operation of RAID Status Display Tool ............................ Appx MC B-1
2 Disabling RAID Status Display Tool ............ Appx MC B-3
Appendix C Reinstalling the OS and Application on Laptop PC ....................................... Appx MC C-1
1. Installation Workflow .................................... Appx MC C-1 2 Installing and Setting Up the OS .................. Appx MC C-2
2.1 Installing and Setting Windows 7 ............ Appx MC C-3 2.2 Setting the OS ....................................... Appx MC C-3
3. Installing the Database Management System ............ Appx MC C-4
4. Installing the DR-ID 300CL-AP .................... Appx MC C-4 4.1 Installing the DR-ID 300CL-AP ............... Appx MC C-4 4.2 Replacing the Menu Database ............... Appx MC C-4
5. Installing the FRIS Modules ......................... Appx MC C-4 6. Installing the SKB Launcher ........................ Appx MC C-4 7. Installing the Standard Kit ............................ Appx MC C-5
8. Installing the Electronic Manual ................... Appx MC C-5 9. Installing the Option Key .............................. Appx MC C-5 10. Installing the AP Key .................................... Appx MC C-5 11. Initializing the Image Database .................... Appx MC C-5 12. Display Optimization (LUT file settings) ....... Appx MC C-5 13. Restoring the Config .................................... Appx MC C-5 14. Settings for
the FPD/Cassette-type Image Readers ....... Appx MC C-5 15. Firewall Settings .......................................... Appx MC C-5 16. Canceling AutoPlay Mode ........................... Appx MC C-5 17. Setting the Event Log .................................. Appx MC C-5 18. Other Settings .............................................. Appx MC C-6
DR-ID 300CL Service Manual
010-305-06E 03.31.2012 FM6000 9
Appendix D Reinstalling the OS and Application on PPU ... Appx MC D-1
1. For PPU-AP V1.1 ......................................... Appx MC D-1 1.1 Installation Workflow .............................. Appx MC D-1 1.2 Installing and Setting Up the OS ............ Appx MC D-2
1.2.1 Installing and Setting Windows 7 ..... Appx MC D-3 1.2.2 Installing the Driver Software .......... Appx MC D-12
1.3 Setting the OS (1/4) ............................... Appx MC D-14 1.4. Installing the SQL Server ....................... Appx MC D-18
1.4.1 Installing the Microsoft SQL Server 2008 SP1 ... Appx MC D-18
1.4.2 Installing the Microsoft SQL Server Management Studio ...................... Appx MC D-19
1.5 Creating the DB/Table ........................... Appx MC D-21 1.6 Setting the OS (2/4) ............................... Appx MC D-22 1.7 Installing the PPU-AP ............................ Appx MC D-27 1.8 Setting the OS (3/4) ............................... Appx MC D-28 1.9 Installing the MFC ................................. Appx MC D-35 1.10 Setting the Update Interval for
Time Synchronization ............................ Appx MC D-35 1.11 Setting the PPU Startup ......................... Appx MC D-36 1.12 Registering/Executing the JOB ............... Appx MC D-36 1.13 Setting the OS (4/4) ............................... Appx MC D-37 1.14 Procedure to be Performed If Necessary ... Appx MC D-39
2. For PPU-AP V1.2 ......................................... Appx MC D-40 2.1 Installation Workflow .............................. Appx MC D-40 2.2 Installing and Setting Up the OS ............ Appx MC D-40
2.2.1 Installing and Setting Windows 7 ..... Appx MC D-41 2.2.2 Installing the Driver Software .......... Appx MC D-57
2.3 Setting the OS ....................................... Appx MC D-61 2.4 Installing the PPU-AP (1/2) .................... Appx MC D-72 2.5 Changing Server Authentication Mode ... Appx MC D-79 2.6 Creating DSN ........................................ Appx MC D-82 2.7 Installing the PPU-AP (2/2) .................... Appx MC D-85 2.8 Setting up the FTP Server ..................... Appx MC D-87 2.9 Verifying the Operation of Application ..... Appx MC D-91
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 10
Appendix E Reinstalling the OS and Application (HP PC) ........................................... Appx MC E-1
1. Installation Workflow .................................... Appx MC E-1 2. Installing and Setting Up the OS .................. Appx MC E-2
2.1 Installing and Setting Windows 7 ............ Appx MC E-3 2.2 Installing the Driver Software ................. Appx MC E-9 2.3 Setting the OS ....................................... Appx MC E-10
3. Setting Up the Monitor ................................. Appx MC E-15 4. Installing Database Management System ... Appx MC E-15 5. Installing the DR-ID 300CL-AP .................... Appx MC E-15
5.1 Installing the AP .................................... Appx MC E-15 5.2 Replacing the Menu Database ............... Appx MC E-15 5.3 Installing the RAID Status Display Tool
(Only for RAID Configuration) ................ Appx MC E-15 6. Installing FRIS Modules ............................... Appx MC E-15 7. Installing the DR-ID 300CL Standard Kit ..... Appx MC E-15 8. Installing the Electronic Manual ................... Appx MC E-15 9. Option Key Installation ................................. Appx MC E-15 10. AP Key Installation ....................................... Appx MC E-15 11. Initializing the Image Database .................... Appx MC E-16 12. Display Optimization (LUT file settings) ....... Appx MC E-16 13. Restoring the Config .................................... Appx MC E-16 14. Settings for
the FPD/Cassette-type Image Readers ....... Appx MC E-16 15. Firewall Settings .......................................... Appx MC E-16
16. Canceling the AutoPlay ............................... Appx MC E-16 17. Setting the Event Log .................................. Appx MC E-16
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 11
2. Overview of Service Utility Functions ..................... MU-4
3. Various Configuration Settings — Setup Configuration Item ......................................... MU-10 3.1 Meanings of Individual Tree Items ......................... MU-11 3.2 Editing the Listed Configuration Items .................... MU-13 3.3 Saving the Configuration Data ............................... MU-14 3.4 Configuration Details — 1. IMAGE MODALITY ....... MU-15 3.5 Configuration Details — 3. QA FUNCTION ............ MU-29 3.6 Configuration Details — 4. PRINTER ..................... MU-33 3.7 Configuration Details — 5. CSL/IDT FUNCTION .... MU-52 3.8 Configuration Details — 7. CONFIG OPTION ......... MU-73 3.9 Configuration Details — 8. CONFIG QC ................ MU-75 3.10 Configuration Details — 10. CONFIG QR ............... MU-76 3.11 Configuration Details — 99. CONFIG FIX .............. MU-77 3.12 Setting the Name Elements for DICOM Connection ... MU-78 3.13 Setting Up the DICOM Print Film Sorting Function ... MU-79 3.14 Patient Information Sharing Function Setup ........... MU-82 3.15 MAC Address Manual Input ................................... MU-84 3.16 Log Information Sharing Function Setup ................ MU-85
4. Display Optimization — LUT .................................... MU-87
5. Restoring/Backing Up the Configuration File — Configuration Restore/Backup ................................ MU-88
6. Retrieving EDR Data — Retrieve EDR Backup Data ... MU-89
7. Verifying the Connection to Other Connected Equipment — Verify Connection ... MU-91
8. Print Output Function ............................................... MU-93 8.1 Film Annotation Character Setup
(Direct Editing of FilmStrFmt) ................................ MU-94 8.1.1 Procedure for
Directly Editing the FilmStrFmt File .................. MU-94 8.1.2 FilmStrFmt File Description Format .................. MU-95 8.1.3 Changing the Film Annotation Characters ......... MU-97 8.1.4 Format Descriptors ........................................ MU-101 8.1.5 Attribute Descriptors ...................................... MU-105 8.1.6 Film Annotation Character Verification Procedure ... MU-110
9. Management of Various Databases — DataBase Utility ......................................................... MU-111 9.1 Initializing the Image Database .............................. MU-111 9.2 Clearing the Contents of the Image Output Queue ... MU-112 9.3 Initializing the Patient Database ............................. MU-113 9.4 Initializing the Calibration Database
(V4.0 or Earlier) .................................................... MU-113
10. Setting the Magnetic Card — Magnetic Card Setting .............................................. MU-114 10.1 Setting the Hospital Card Data Format ................... MU-114 10.2 Defining the ISO Card Character Read Code ......... MU-114
11. Setting the Barcode — Barcode Setting ................. MU-115
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 12
12. Setting the Selectors — Selector Setting ............... MU-117 12.1 Setting the Selectors (V4.0 or Earlier) .................... MU-117 12.2 Setting the Selectors (V5.0 or Later) ...................... MU-118
13. Collecting the Log and Configuration Files — Log/Configuration Files Acquisition Tool ............... MU-120
14. Setting the DICOM MWL C-FIND Tags — DICOM MWL C-FIND Tag Setting (V4.0 or Earlier) ... MU-123
15. Registering the Image Detectors — Device Setting (V5.0 or Later) .................................. MU-124
16. DR Console Setup Tool............................................. MU-125 16.1 Setting the Wake-on-LAN Utility ............................ MU-125 16.2 Setup for Addition of Descriptive Characters to Icons
[Applicable only in Japan] ..................................... MU-128 16.3 Setting the Grid Pattern Removal Processing ......... MU-129 16.4 Setup of Auto Shuttering Processing ...................... MU-130
17. Menu Import Export Tool .......................................... MU-131 17.1 Functions and Overview of
the Menu Import Export Tool ................................. MU-132 17.2 Setting Up the Menu Import Export Tool ................ MU-134 17.3 Starting/Shutting Down
the Menu Import Export Tool ................................. MU-135 17.4 Exporting Exposure Menus ................................... MU-135
17.5 Cautions Observed Before Menu Import ................ MU-136 17.5.1 Cautions Observed When
Editing Exposure Menus and Study Menus ....... MU-136 17.5.2 Setting the Positioning ID and
the Tube/Technique Code (Applicable only other than in Japan) ............... MU-137
17.6 Importing Exposure Menus and Study Menus ........ MU-139 17.7 Checking APR Information .................................... MU-140 17.8 Setting the 180° Image Rotation Information .......... MU-141 17.9 Automatic Technique Settings ............................... MU-142
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 13
Appendix G Filter Information Setup (Applicable only other than in Japan) ...................... Appx MU G-1
Appendix H Setting the Exposure Index Function (V5.0 or later) .................................. Appx MU H-1
Appendix I Setting the Technique Code ......... Appx MU I-1
Appendix J DICOMTagEditTool (V5.0 or Later) ... Appx MU J-1 1. Functions and Overview of
DICOMTagEditTool ..................................... Appx MU J-1 2. Setting up the DICOMTagEditTool .............. Appx MU J-5 3. Creating Tag Conversion Files .................... Appx MU J-6
3.1 Creating the Operation Definition Selection List ........................................ Appx MU J-6
3.2 Configuring Operation Definitions ........... Appx MU J-6 3.3 Saving Tag Conversion Files ................. Appx MU J-8 3.4 Testing the Saved Tag Conversion Files ... Appx MU J-9 3.5 Applying Tag Conversion Files ............... Appx MU J-10
4. Details of Each Command ........................... Appx MU J-11 4.1 Details of Operation Commands ............ Appx MU J-11 4.2 Details of Judgment Command .............. Appx MU J-13 4.3 Details on Description of Tags ............... Appx MU J-14
Appendix A DICOM Tag Mapping Function (V4.0 or Earlier) .............................. Appx MU A-1
Appendix B Creating/Registering/Deleting Markers ........................................... Appx MU B-1
1. Creating Markers ......................................... Appx MU B-1 2. Marker Creation Example [Reference] ........ Appx MU B-4 3. Preparing Markers for Mammography
(V7.0 or Later) .............................................. Appx MU B-7 4. Registering Markers (V5.0 or later) .............. Appx MU B-7 5. Deleting Markers .......................................... Appx MU B-8
5.1 Deleting Markers (V4.0 or Earlier) .......... Appx MU B-8 5.2 Deleting Markers (V5.0 or Later) ............ Appx MU B-8
Appendix C Using the DICOM Log Mode ........ Appx MU C-1
Appendix D DICOM Tag Editing Function (V4.0 or Earlier) .............................. Appx MU D-1
1. MPM Code Related Settings ....................... Appx MU D-2 2. Settings Enabling Differentiation Between
a DR Image and a CR Image ...................... Appx MU D-3
Appendix E Setup of Log Output Items of the Exposure Result Log Function ... Appx MU E-1
Appendix F Setup of Rules Applied for Automatic Patient ID Issuance (V4.0 or Earlier) .............................. Appx MU F-1
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 14
5. Taking over the Tag Conversion Function in V4.0 or Earlier .............................................. Appx MU J-15 5.1 Taking over the Conversion Setting of
the “TagReplace.ini” File ........................ Appx MU J-16 5.2 Taking over the Conversion Setting of
the “TagLookUp.ini” File ........................ Appx MU J-17 5.3 Taking over the DICOM MWM C-FIND Tag
Customization ....................................... Appx MU J-18
Appendix K Editing the AEC Conversion File (V6.1 or Later) ................................ Appx MU K-1
1. Format of the AEC Conversion File ............. Appx MU K-2 2. Procedures of Editing AEC Conversion File ... Appx MU K-3
Appendix L CS Edit Tool for Storage ............... Appx MU L-1 1. Functions and Overview of
CS Edit Tool for Storage .............................. Appx MU L-1 2. Setting up the CS Edit Tool for Storage ....... Appx MU L-4
2.1 Operating Environment .......................... Appx MU L-4 2.2 Tool Installation ..................................... Appx MU L-4 2.3 Starting/Shutting Down
the CS Edit Tool for Storage .................. Appx MU L-5 3. Operation list of Editing Functions ............... Appx MU L-6 4. Editing Samples ........................................... Appx MU L-7
Appendix M Setting the Default Positioning ID (Only for Connecting to the DR-ID 200) ................................ Appx MU M-1
Appendix N Extended Parameter Convert Tool (V7.0 or Later) ................................ Appx MU N-1
1. Overview of the Extended Parameter Convert Tool ........ Appx MU N-1
2. Setting up the Extended Parameter Convert Tool ........ Appx MU N-1
3. Starting/Operating the Extended Parameter Convert Tool ........ Appx MU N-2
4. Errors and Remedies ................................... Appx MU N-4 5. List of Update Files ...................................... Appx MU N-5 6. List of Additional Menus ............................... Appx MU N-6
6.1 Mammography Menu ............................. Appx MU N-6 6.2 Tomosynthesis Menu ............................ Appx MU N-9
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 15
SP: Service Parts
1. Service Parts List ........................................................ SP-1 1.1 How to Use Service Parts List ................................. SP-1 1.2 Service Parts List .................................................... SP-3
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 16
1. Preventive Maintenance Items ................................... PM-1
2. Checking for a Fatal Error .......................................... PM-2 2.1 Confirming the Event Log ........................................ PM-2
3. Safety ........................................................................... PM-2 3.1 Service Voltage ...................................................... PM-2 3.2 Power Plug ............................................................. PM-2
4. Hardware Cleaning ...................................................... PM-3 4.1 PC Internal Cleaning ............................................... PM-3 4.2 Keyboard and Mouse Cleaning ................................ PM-3 4.3 LCD Cleaning ......................................................... PM-4 4.4 DVD/CD-RW Combo Drive Cleaning ....................... PM-4
5. Executing the Error Analysis Tool ............................. PM-5 5.1 Executing Dell Diagnostics ...................................... PM-5
6. Functional Checks ...................................................... PM-5 6.1 Confirmation of Image Input to Storage onto HDD .... PM-5 6.2 Image Output to Film .............................................. PM-5 6.3 Image Transfer ....................................................... PM-5
7. Backup of Setup Files and Verification of Date and Time ..................................... PM-6 7.1 Backup of Setup Files and Log Files ........................ PM-6 7.2 Verification of Date and Time .................................. PM-6
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 17
IN: Installation
1. DR-ID 300CL Installation Conditions ......................... IN-1 1.1 Dimensions and Weight .......................................... IN-1 1.2 Temperature and Humidity ...................................... IN-1 1.3 Places of Installation ............................................... IN-1 1.4 Power Requirements .............................................. IN-1
2. Installation Procedure Overview ............................... IN-2
3. DR-ID 300CL Hardware Setup .................................... IN-3 3.1 Unloading and Unpacking ....................................... IN-3 3.2 Checking the Component Items ............................... IN-4 3.3 Installing the Accessories,
Optional Items, and Available Items ......................... IN-6 3.3.1 Installing the Accessories and Optional Items
(for Desktop PC) ............................................. IN-6 3.3.2 Installing the Accessories and Available Items
(for Laptop PC) ............................................... IN-7 3.4 Making a Cable Connection to
the FPD/Cassette-type Image Reader ..................... IN-8 3.5 Connecting to the Other Equipment ......................... IN-8
4. Software Setup ............................................................ IN-9 4.1 DR-ID 300CL-AP Startup ........................................ IN-9
4.1.1 OS Startup ..................................................... IN-10 4.1.2 OS Setups ..................................................... IN-11 4.1.3 Touch Panel Driver Installation (Only on Desktop PC)
[Skip this procedure if the 17" monitor with a touch panel is provided.] ................................ IN-12
4.1.4 Touch Panel Monitor Setups/ Operational Verification (Only on Laptop PC) ...... IN-19
4.1.5 Setup for Addition of Descriptive Characters to Icons [Applicable only in Japan] ................................. IN-20
4.1.6 Installing the DR-ID 300CL Standard Kit (V5.0 or Later) ................................................ IN-21
4.1.7 Option Key Installation ..................................... IN-25 4.1.8 AP Key Installation .......................................... IN-26 4.1.9 DR-ID 300CL Startup Verification ...................... IN-27
4.2 Settings on the DR-ID 300CL for Connection with the FPD/Cassette-type Image Readers .................... IN-28 4.2.1 Starting the Service Utility ................................ IN-28 4.2.2 DR-ID 300CL System Information Setup ............ IN-29 4.2.3 Registering the Image Detector (V5.0 or Later) .... IN-60 4.2.4 Setting the Selectors ....................................... IN-61 4.2.5 Optimizing the Monitor Display .......................... IN-64 4.2.6 Changing the Menu Database ........................... IN-65 4.2.7 Exiting the Service Utility .................................. IN-66 4.2.8 Installing the MC Software ................................ IN-66
4.3 Setting Network Connection in Laptop PC ................ IN-67 4.3.1 Setting the IP Address in
Each Network Configuration ............................. IN-67 4.3.2 Enabling the Built-in Wireless Network Adaptor of
a Laptop PC ................................................... IN-71 4.3.3 Setting the Connection Between
DR-ID 300CL (Laptop PC) and the Image Local Network AP ............................ IN-72
4.4 Setting Network Connection for the Desktop PC with Additional NIC ......................... IN-73
4.5 Setting the Priority Order of the Network Binding ...... IN-74 4.5.1 For the Laptop PC ........................................... IN-74 4.5.2 For the Desktop PC with Additional NIC ............. IN-75
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 18
4.6 Procedures for Changing the DR-ID 300CL and RU (FPD and Cassette-type Image Reader) IP Addresses .......................................................... IN-75
4.7 Procedures for Changing the DR-ID 300CL Host Name .................................. IN-75
4.8 DR-ID 300CL Setup for Connection to Other Equipment IN-76
4.9 Barcode Reader/Magnetic Card Reader Connection ... IN-76 4.10 Setting the DX Parameter ....................................... IN-76 4.11 Setting the VIP Account Password ........................... IN-77
5. Checkout Procedures ................................................. IN-78 5.1 Checking the Connection between
the DR-ID 300CL and FPD/Cassette-type Image Reader ........................... IN-78
5.2 Checking the Connection between the DR-ID 300CL and Other Equipment ................... IN-78
5.3 Checking the Output Image ..................................... IN-78
6. Setup File Backup, Image Database Initialization and Error Log Verification/Deletion .................................. IN-79 6.1 Backing Up the DR-ID 300CL Setup Files ................ IN-80 6.2 Initializing Various Image Databases ....................... IN-81
6.2.1 Initializing Image Databases ............................. IN-81 6.2.2 Clearing the Image Output Queue ..................... IN-82 6.2.3 Initializing the Patient Database ........................ IN-82 6.2.4 Initializing the Calibration Database
(V4.0 or Earlier) .............................................. IN-83 6.3 Checking and Deleting the DR-ID 300CL Error Log ... IN-84 6.4 Turning OFF the DR-ID 300CL ................................ IN-84
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 19
Appx IN: Installation – Appendix
Appendix 1 Connecting the Barcode Reader [Option] ........................................... Appx IN1-1
1. Connecting the Barcode Reader to the DR-ID 300CL .... Appx IN1-1
2. Settings Required When Using the Study List Search Function Based on the Hospital Barcode ................................... Appx IN1-5 2.1 Setting the Search Key .......................... Appx IN1-5 2.2 Setting the Barcode Type ...................... Appx IN1-6
Appendix 2 Setting Up the Magnetic Card [Option] ........................................... Appx IN2-1
1. Installing the Bracket ................................... Appx IN2-1 2. Connecting the DR-ID 300CL to the Reader ... Appx IN2-2 3. Setting the Reader and Magnetic Card ....... Appx IN2-4 4. Hospital Card Data Format Setup Example
[Reference] .................................................. Appx IN2-8 5. ID Card Data Format [Reference] ................ Appx IN2-12
Appendix 3 Procedures for Changing the DR-ID 300CL and RU IP Addresses ............................ Appx IN3-1
1. Incorporating into the Network ..................... Appx IN3-1 2. Changing the Software Settings .................. Appx IN3-2
Appendix 4 Changing the RU Host Name ........ Appx IN4-1
Appendix 5 Changing the DR-ID 300CL Host Name ......... Appx IN5-1
Appendix 6 Changing the DR-ID 300CL IP Address ......... Appx IN6-1
Appendix 7 Uninstalling the AP ........................ Appx IN7-1 1. DR-ID 300CL-AP ......................................... Appx IN7-1 2. SQL Server .................................................. Appx IN7-2
2.1 SQL Server 2005 .................................. Appx IN7-2 2.2 SQL Server 2008 .................................. Appx IN7-4
3. Option Keys ................................................. Appx IN7-7 4. Console Software (Built_inConsole)
(V4.0 or Earlier) ........................................... Appx IN7-8 5. FRIS Modules [Applicable only in Japan] .... Appx IN7-9 6. Synapse Client (Synapse Workstation) ....... Appx IN7-10 7. Precise Enlargement Software .................... Appx IN7-11 8. Free Layout Print Software .......................... Appx IN7-13
Appendix 8 Upgrading the AP .......................... Appx IN8-1 1. Upgrading the DR-ID 300CL ........................ Appx IN8-1 2. System Migration to
the DR-ID 300CL V5.0 or Later ................... Appx IN8-15 3. Upgrading the PPU-AP ................................ Appx IN8-28
Appendix 9 Setup for Security Enhancement ... Appx IN9-1
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 20
Appendix 10 DICOM CR Storage ........................ Appx IN10-1 1. Hardware Setup ........................................... Appx IN10-3 2. Software Setup ............................................ Appx IN10-4
2.1 DR-ID 300CL Software Setup ................ Appx IN10-4 2.1.1 Starting the Service Utility .............. Appx IN10-4 2.1.2 Setup for the Study Modality ........... Appx IN10-5 2.1.3 Setup of Information on
This Equipment for the DICOM CR Storage Function .... Appx IN10-6
2.1.4 Registration of Information on Other Connected Equipment for the DICOM CR Storage Function .... Appx IN10-9
2.1.5 Settings of Automatic Image Distribution (DISTRIBUTION CODE) ................ Appx IN10-13
2.1.6 Setup for CONNECTING EQUIPMENT.......... Appx IN10-14
2.1.7 Settings for UID Issuance ............... Appx IN10-14 2.1.8 Changing the Name Elements ........ Appx IN10-16 2.1.9 Display Gradation Correction Setup ... Appx IN10-16 2.1.10 Service Utility Shutdown ................ Appx IN10-18
2.2 Software Setup on Other Connected Equipment .................. Appx IN10-18
3. Connection Checkout .................................. Appx IN10-19 3.1 DR-ID 300CL-AP Startup ....................... Appx IN10-19 3.2 Image Transfer to
Other Connected Equipment .................. Appx IN10-19 4. Setup File Backup ........................................ Appx IN10-20
Appendix 11 DICOM DX Storage: Transfer of Processed Images ..... Appx IN11-1
1. Hardware Setup ........................................... Appx IN11-3 2. Software Setup ............................................ Appx IN11-4
2.1 DR-ID 300CL Software Setup ................ Appx IN11-4 2.1.1 Starting the Service Utility .............. Appx IN11-4 2.1.2 Setup for the Study Modality ........... Appx IN11-5 2.1.3 Setup of Information on
This Equipment for the DICOM DX Storage Function .... Appx IN11-6
2.1.4 Registration of Information on Other Connected Equipment for the DICOM DX Storage Function .... Appx IN11-9
2.1.5 Settings of Automatic Image Distribution (DISTRIBUTION CODE) ................ Appx IN11-13
2.1.6 Setup for CONNECTING EQUIPMENT.......... Appx IN11-14
2.1.7 Settings for UID Issuance ............... Appx IN11-14 2.1.8 Changing the Name Elements ........ Appx IN11-16 2.1.9 Display Gradation Correction Setup ... Appx IN11-16 2.1.10 Service Utility Shutdown ................ Appx IN11-18
2.2 Software Setup on Other Connected Equipment .................. Appx IN11-18
3. Connection Checkout .................................. Appx IN11-19 3.1 DR-ID 300CL-AP Startup ....................... Appx IN11-19 3.2 Image Transfer to
Other Connected Equipment .................. Appx IN11-19 4. Setup File Backup ........................................ Appx IN11-20
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 21
Appendix 12 DICOM Print ................................... Appx IN12-1 1. Hardware Setup ........................................... Appx IN12-2 2. Software Setup ............................................ Appx IN12-3
2.1 DR-ID 300CL Software Setup ................ Appx IN12-3 2.1.1 Installation of
the DICOM Print Option Key ........... Appx IN12-3 2.1.2 AP Key Reinstallation .................... Appx IN12-4 2.1.3 Service Utility Startup .................... Appx IN12-4 2.1.4 Setup of Information on
This Equipment for the DICOM Print Function .............. Appx IN12-5
2.1.5 Registration of Information on Other Connected Equipment for the DICOM Print Function .............. Appx IN12-8
2.1.6 Registration of Printer That Automatically Outputs Print Films Upon Completion of Checkout – CONNECTING EQUIPMENT.......... Appx IN12-11
2.1.7 Film Annotation Character Format Setup ............................... Appx IN12-12
2.1.8 Changes in Name Elements ........... Appx IN12-13 2.1.9 Setup for Film Sorting Function ....... Appx IN12-13 2.1.10 Setting the DICOM Print
Output Conditions ......................... Appx IN12-13 2.1.11 Service Utility Shutdown ................ Appx IN12-14
2.2 Software Setup for Other Connected Equipment .................. Appx IN12-14
3. Verifying the Connection .............................. Appx IN12-15 3.1 Starting the DR-ID 300CL-AP ................ Appx IN12-15 3.2 Verifying
the Film Annotation Character Format .... Appx IN12-15 4. Backing Up the Setup File ........................... Appx IN12-16
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 22
Appendix 13 RIS: DICOM MWM/DICOM PPS ..... Appx IN13-1 1. Hardware Setup ........................................... Appx IN13-4 2. Software Setup ............................................ Appx IN13-5
2.1 DICOM MWM or DICOM PPS Option Key Installation ........................... Appx IN13-5
2.2 AP Key Reinstallation ............................ Appx IN13-6 2.3 Setup for the DICOM MWM Function ..... Appx IN13-6
2.3.1 Setup of DR-ID 300CL Information for the DICOM MWM Function ............ Appx IN13-6
2.3.2 Registration of RIS Information ....... Appx IN13-9 2.3.3 Registration of RIS Used for MWM ... Appx IN13-10
2.4 Setup for the DICOM PPS Function ....... Appx IN13-11 2.4.1 Setup of DR-ID 300CL Information for
the DICOM PPS Function .............. Appx IN13-11 2.4.2 Registration of RIS Information ....... Appx IN13-13 2.4.3 Registration of RIS Used with PPS ... Appx IN13-14
2.5 MWM/PPS Configuration Setup ............. Appx IN13-14 2.5.1 General Settings for MWM/PPS ...... Appx IN13-15 2.5.2 Setting
the Method of Inquiring at the RIS ... Appx IN13-16 2.5.3 Setting
Additional MWM Inquiry Items ........ Appx IN13-17 2.5.4 Settings for the Compliance with
the JJ1017 Guideline (Applicable only in Japan) .............. Appx IN13-17
2.5.5 Setting the Additional PPS Items to be Returned to the RIS .................. Appx IN13-18
2.5.6 Setting the Timing for Transmission of PPS Study Termination Notice (V4.0 or Earlier) ............................ Appx IN13-19
2.5.7 Embedding the Exposure Information into a PPS Standard Tag ..................... Appx IN13-19
2.5.8 Settings for Editing the Exposure Result (V4.0 or Earlier) ............................ Appx IN13-20
2.5.9 Setting the MWM Code Meaning as a Menu Name .............................. Appx IN13-21
2.5.10 Settings for Regenerating StudyInstanceUID for a Study Acquired Using the MWM Function ........................ Appx IN13-21
2.6 Changing Name Elements ..................... Appx IN13-24 2.7 Setup for
Exposure Menu/Study Menu Mapping .... Appx IN13-24 3. Verifying the Connection .............................. Appx IN13-26
3.1 Starting the DR-ID 300CL-AP ................ Appx IN13-26 3.2 Verifying MWM ...................................... Appx IN13-26 3.3 Verifying PPS ........................................ Appx IN13-26
4. Backing Up the Setup File ........................... Appx IN13-27
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 23
1. Hardware Setup ........................................... Appx IN14-3 2. Software Setup ............................................ Appx IN14-4
2.1 Installing the DICOM Commitment Push Option Key ............................................ Appx IN14-4
2.2 Reinstalling the AP Key ......................... Appx IN14-5 2.3 Performing DICOM CR Storage Setup ... Appx IN14-5 2.4 Setting Information on This Equipment for
the Commitment Push Function ............. Appx IN14-6 2.5 Registering the Archiver Information for
the Commitment Push Function ............. Appx IN14-7 2.6 Registering the SC AE Name ................. Appx IN14-9
3. Verifying the Function .................................. Appx IN14-10 3.1 Starting the DR-ID 300CL-AP ................ Appx IN14-10 3.2 Transferring an Image to the Archiver ..... Appx IN14-10
4. Backing Up the Setup File ........................... Appx IN14-11
Appendix 15 F-RIS: Ordering .............................. Appx IN15-1
Appendix 16 Connecting the High-resolution Monitor ......... Appx IN16-1
1. Changing to the High-resolution Monitor (1600 × 1200 monitor) ................................. Appx IN16-1
2. Reinstalling the Application and Performing Necessary Settings ................... Appx IN16-3 2.1 Reinstalling the Application
(V4.0 or Earlier) ..................................... Appx IN16-3 2.2 Setting the High-Resolution Monitor ....... Appx IN16-4
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 24
Appendix 17 DICOM Server : Image Reprint (V4.0 or Earlier) ..... Appx IN17-1
1. Hardware Setup ........................................... Appx IN17-3 2. Software Setup ............................................ Appx IN17-4
2.1 DR-ID 300CL Software Setup ................ Appx IN17-4 2.1.1 Installing the Image Search/Acquisition
Option Key ................................... Appx IN17-4 2.1.2 Reinstalling the AP Key ................. Appx IN17-5 2.1.3 Setting the Information on
This Equipment for Image Search/Acquisition ............... Appx IN17-5
2.1.4 Registering the Information on Other Connected Equipment for Image Search/Acquisition ............... Appx IN17-8
2.1.5 Setting the Information on This Equipment for the DICOM CR Storage Function .... Appx IN17-12
2.1.6 Registering the Information on Other Connected Equipment for the DICOM CR Storage Function .... Appx IN17-14
2.1.7 Setting CONNECTING EQUIPMENT ... Appx IN17-16 2.1.8 Setting CONFIG QR ...................... Appx IN17-17 2.1.9 Setting the LUT Reverse-conversion
Processing ................................... Appx IN17-19 2.1.10 Exiting the Service Utility ................ Appx IN17-20
2.2 Software Setup on Other Connected Equipment .................. Appx IN17-20
3. Verification ................................................... Appx IN17-21 3.1 Starting the DR-ID 300CL-AP ................ Appx IN17-21 3.2 Searching/Acquiring Images from
the DICOM Server ................................. Appx IN17-22 3.3 Checking Film Output Images ................ Appx IN17-22
4. Setup File Backup ........................................ Appx IN17-23
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 25
Appendix 18 Setup of Date and Time Display ... Appx IN18-1 1. Displaying the Date and Time
(V4.0 or Earlier) ........................................... Appx IN18-1 2. Setting the Date and Time ........................... Appx IN18-3 3. Setting the Display Format of Date and Time ... Appx IN18-4 4. Setting the Display Font of Date and Time ... Appx IN18-6 5. Setting the Display Font Size of
Date and Time (V4.0 or Earlier) ................... Appx IN18-8
Appendix 19 X-Ray Controller ............................ Appx IN19-1 1. Hardware Setup ........................................... Appx IN19-2 2. Software Setup ............................................ Appx IN19-3
2.1 Installing the Software for Directly Connecting the X-CON .............. Appx IN19-3
2.2 Reinstalling the AP Key ......................... Appx IN19-5 2.3 Configuration setting for
Direct Connection with the X-CON ......... Appx IN19-6 2.4 Setting the OS COM Ports ..................... Appx IN19-11
3. Verifying the Connection/Data Exchange .... Appx IN19-12 3.1 Verifying the Serial Connection .............. Appx IN19-12 3.2 Checking the Data Exchange ................. Appx IN19-13
4. Backing Up the Setup File ........................... Appx IN19-14 5. Switching to DR Connection Environment
(V4.0.0001) .................................................. Appx IN19-15
010-305-07E 09.12.2012 FM6055 26
1. DR-ID 300CL Software Setup ...................... Appx IN20-4 1.1 Installation of
the Media Storage Option Key ............... Appx IN20-4 1.2 Reinstallation of the AP Key ................... Appx IN20-5 1.3 Setup of Removable Media Information on
the DICOM Media Storage Function ....... Appx IN20-6 1.4 CONNECTING EQUIPMENT Settings .... Appx IN20-10 1.5 Configuration Settings ........................... Appx IN20-11 1.6 Setting When DVD Media are Used ....... Appx IN20-13
1.6.1 For DR-ID 300CL V4.0 .................. Appx IN20-13 1.6.2 For DR-ID 300CL V7.0 or Later ....... Appx IN20-14
2. Connection Checkout .................................. Appx IN20-15 2.1 DR-ID 300CL-AP Startup ....................... Appx IN20-15 2.2 Image Storage onto
the Removable Media ............................ Appx IN20-16 3. Setup File Backup ........................................ Appx IN20-17
Appendix 21 Synapse Server (Web Query) ....... Appx IN21-1 1. Hardware Setup ........................................... Appx IN21-2 2. DR-ID 300CL Software Setup ...................... Appx IN21-3
2.1 Installing the Synapse Client Software .... Appx IN21-3 2.2 Installing the Synapse Web Query
Application (V4.0 Only) .......................... Appx IN21-4 2.3 Setting System Environment Variables ... Appx IN21-5 2.4 Web Query Related
Configuration Settings ........................... Appx IN21-6 2.5 Setting the User Utility ........................... Appx IN21-8
3. Connection Checkout .................................. Appx IN21-9 3.1 DR-ID 300CL-AP Startup ....................... Appx IN21-9 3.2 Web Query Checkout ............................ Appx IN21-9
4. Setup File Backup ........................................ Appx IN21-11
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 27
Appendix 22 Auto Convert Patient ID ................ Appx IN22-1
Appendix 23 Settings Necessary for Browsing the Operation Manuals (PDF) ....... Appx IN23-1
1. Software Setup ............................................ Appx IN23-2 1.1 Tool Installation ..................................... Appx IN23-2 1.2 AP Key Installation ................................ Appx IN23-3 1.3 Setting the Adobe Reader
Operating Environment .......................... Appx IN23-4 1.4 Correcting and Moving the Setup Files ... Appx IN23-6
2. Verification ................................................... Appx IN23-7 3. Setup File Backup/Restore .......................... Appx IN23-8 4. Browsing Operation Manuals (PDF)
Other Than for the Console ......................... Appx IN23-9
Appendix 24 Connecting 2nd Monitor ............... Appx IN24-1 1. Uninstalling the Video Driver ....................... Appx IN24-2 2. Mounting the Video Board ........................... Appx IN24-4 3. Installing and Setting the Video Driver ......... Appx IN24-6
3.1 2M Monitor (Video Board Not Added) ..... Appx IN24-7 3.1.1 Connecting the Monitor .................. Appx IN24-7 3.1.2 Setting the Monitor ........................ Appx IN24-9
3.2 3M Monitor (Video Board: FireGL V3600) ... Appx IN24-11 3.2.1 Setting the Monitor
(Before installing the driver) ............ Appx IN24-11 3.2.2 Installing the Video Driver .............. Appx IN24-13 3.2.3 Setting the Monitor ........................ Appx IN24-15
3.3 3M Monitor (Video Board: FirePro V3800) ................ Appx IN24-19 3.3.1 Setting the Monitor
(Before Installing the Driver) ........... Appx IN24-19 3.3.2 Installing the Video Driver .............. Appx IN24-20 3.3.3 Setting the Monitor ........................ Appx IN24-23
4. Changing the 2nd Monitor ........................... Appx IN24-27 4.1 Changing the 2nd Monitor ...................... Appx IN24-27
4.1.1 Changing the 2M Monitor to the 3M Monitor ............................. Appx IN24-27
4.1.2 Changing the 3M Monitor to the 2M Monitor ............................. Appx IN24-27
4.2 Having the 2nd Monitor Recognize the Linking Application ........................... Appx IN24-27
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 28
Appendix 25 Precise Enlargement Function ..... Appx IN25-1 1. Software Setup ............................................ Appx IN25-2
1.1 Installing the Precise Enlargement Function Software ............................................... Appx IN25-2
1.2 Setting the Firewall (V4.0 Only) .............. Appx IN25-5 1.3 Installing the Precise Enlargement Function
Option Key ............................................ Appx IN25-6 1.4 Reinstalling the AP Key ......................... Appx IN25-6 1.5 Setting the System ................................ Appx IN25-7
1.5.1 Setting the OS .............................. Appx IN25-7 1.5.2 Setting the Service Utility ............... Appx IN25-7 1.5.3 Setting the User Utility ................... Appx IN25-8 1.5.4 Setting the Overlay Items ............... Appx IN25-9
2. Confirming the Settings ............................... Appx IN25-11 2.1 Starting the DR-ID 300CL-AP ................ Appx IN25-11 2.2 Confirming the Image Display ................ Appx IN25-11
3. Backup/Restore ........................................... Appx IN25-12 3.1 Backup ................................................. Appx IN25-12 3.2 Restore ................................................. Appx IN25-14
4. Disabling the Precise Enlargement Function ............... Appx IN25-15
5. Enabling the Precise Enlargement Function ............... Appx IN25-16
Appendix 26 Adding and Setting the Serial Connector ..................... Appx IN26-1
1. Adding a Serial Connector ........................... Appx IN26-1 1.1 Adding a Serial Connector from
the Mother Board .................................. Appx IN26-2 1.2 Mounting the Serial Connector
Add-in Board ......................................... Appx IN26-5 1.3 Mounting the MOXA Board .................... Appx IN26-6
1.3.1 Setting the DIP Switch ................... Appx IN26-6 1.3.2 Mounting the MOXA Board ............. Appx IN26-6
2. Installing the Driver ...................................... Appx IN26-7 2.1 Installing the Serial Connector
Add-in Board Driver ............................... Appx IN26-7 2.1.1 Installing the Serial Connector
Add-in Board Driver ....................... Appx IN26-7 2.1.2 Confirming Installation of
the Serial Connector Add-in Board Driver ....................... Appx IN26-8
2.2 Installing the MOXA Board Driver ........... Appx IN26-9 2.2.1 Installing the MOXA Board Driver
(for Windows Vista) ....................... Appx IN26-9 2.2.2 Installing the MOXA Board Driver
(for Windows 7) ............................ Appx IN26-11 2.2.3 Confirming Installation of
the MOXA Board Driver ................. Appx IN26-14 3. Port and Service Utility Settings .................. Appx IN26-15
3.1 The Case Where the PC-standard Serial Connector and an Add-in Serial Connector are Used ..... Appx IN26-15
3.2 The Case Where a MOXA Board is Used ... Appx IN26-17
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 29
Appendix 27 DAP (Dose-area Product Meter) ... Appx IN27-1 1. Hardware Setup ........................................... Appx IN27-2 2. Software Setup ............................................ Appx IN27-3
2.1 Starting the Service Utility ...................... Appx IN27-3 2.2 Setting the Configuration Items .............. Appx IN27-5 2.3 Exiting the Service Utility ....................... Appx IN27-7
3. Verification ................................................... Appx IN27-8 3.1 Starting the DR-ID 300CL-AP ................ Appx IN27-8 3.2 Verifying Connection to a DAP ............... Appx IN27-8 3.3 Verifying Reception of
Dose-area Product Information .............. Appx IN27-9 3.4 Verifying Transfer of
Dose-area Product Information .............. Appx IN27-10 3.4.1 When DICOM PPS Connection is
Enabled ....................................... Appx IN27-10 3.4.2 When DICOM Storage Connection is
Enabled ....................................... Appx IN27-10 3.4.3 When DICOM Print Connection is
Enabled ....................................... Appx IN27-10 4. Setup File Backup ........................................ Appx IN27-11
Appendix 28 Simple Ordering or RIS Order USB ............................... Appx IN28-1
1. Hardware Setup ........................................... Appx IN28-2 2. Software Setup ............................................ Appx IN28-3
2.1 Installing the Simple Ordering Option Key ............................................ Appx IN28-3
2.2 Reinstalling the AP Key ......................... Appx IN28-4 2.3 Setting the Configuration Related to
the Simple Ordering or RIS Order USB ... Appx IN28-4 2.4 Creating and Setting the Shared Folder ... Appx IN28-6
3. Verification ................................................... Appx IN28-8 3.1 Starting the DR-ID 300CL-AP ................ Appx IN28-8 3.2 Verifying Receiving of
the Order Information ............................ Appx IN28-8 3.2.1 For Simple Ordering ...................... Appx IN28-8 3.2.2 For RIS Order USB ....................... Appx IN28-9
4. Setup File Backup ........................................ Appx IN28-10
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 30
Appendix 29 PDI (Portable Data of Imaging) ..... Appx IN29-1 1. Software Setup ............................................ Appx IN29-2
1.1 Installing the PDI Application .................. Appx IN29-2 1.2 Installing the PDI Option Key ................. Appx IN29-4 1.3 Reinstalling the AP Key ......................... Appx IN29-4 1.4 Set up configurations related to
PDI function .......................................... Appx IN29-5 1.5 Editing the Institution Information ........... Appx IN29-7
2. Verification ................................................... Appx IN29-8 2.1 Starting the DR-ID 300CL-AP ................ Appx IN29-8 2.2 Writing Study Information Using
the PDI Function ................................... Appx IN29-9 2.3 Verifying the Created CD-R ................... Appx IN29-11
3. Setup File Backup ........................................ Appx IN29-13
Appendix 30 Free Layout Print ........................... Appx IN30-1 1. Software Setup ............................................ Appx IN30-2
1.1 Installing the Free Layout Print Application ............ Appx IN30-3
1.2 Installing the Free Layout Print Option Key ............................................ Appx IN30-6
1.3 Reinstalling the AP Key ......................... Appx IN30-6 1.4 Set up Configurations Related to
Free Layout Print Function ..................... Appx IN30-7 1.5 Setting the Cooper Application ............... Appx IN30-8
2. Verification ................................................... Appx IN30-9 2.1 Starting the DR-ID 300CL-AP ................ Appx IN30-9 2.2 Verifying the Print Output ....................... Appx IN30-10
3. Setup File Backup ........................................ Appx IN30-11 4. Backing up/Restoring
Cooper Application Settings ........................ Appx IN30-12 4.1 Backup ................................................. Appx IN30-12 4.2 Restore ................................................. Appx IN30-14
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 31
Appendix 31 Performing Study Information Sharing Setup (Cluster Connection) ..................... Appx IN31-1
1. Hardware Setup ........................................... Appx IN31-3 2. Software Setup ............................................ Appx IN31-3
2.1 Software Setup for All DR-ID 300CLs ..... Appx IN31-4 2.1.1 Verifying the Host Name ................ Appx IN31-4 2.1.2 Initializing the Output Queue and
Image Database ........................... Appx IN31-5 2.1.3 Equalizing the Configuration Settings ... Appx IN31-6 2.1.4 Verifying the Settings of
Connected Devices and Selectors ... Appx IN31-7 2.1.5 Equalizing the Environment/Table/Menu
Settings ....................................... Appx IN31-8 2.1.6 Performing Setup for Drive Sharing ... Appx IN31-14 2.1.7 Performing
Study Information Sharing Setup ..... Appx IN31-17 2.1.8 Defining the Share Range .............. Appx IN31-18 2.1.9 Backing Up/Restoring
the Share Settings ........................ Appx IN31-19 2.2 Software Setup for
Individual DR-ID 300CLs ....................... Appx IN31-21 2.2.1 Performing Integrated
Output Processing Setup ............... Appx IN31-21 2.2.2 Performing Study List Setup ........... Appx IN31-23 2.2.3 Setting User Utility in
All DR-ID 300CLs ......................... Appx IN31-23
3. Connection Verification ................................ Appx IN31-24 3.1 Confirming Consistency ......................... Appx IN31-24 3.2 Study Registration ................................. Appx IN31-25 3.3 Image Reading ...................................... Appx IN31-26 3.4 Study Information and Image Display ..... Appx IN31-26 3.5 Register IP Addresses and
Host Names of Other DR-ID 300CLs to Share Study Information with ................. Appx IN31-27
4. Setup File Backup ........................................ Appx IN31-28 5. Adding or Deleting
the Cluster-connected DR-ID 300CL ........... Appx IN31-29 5.1 Adding
the Cluster-connected DR-ID 300CL ...... Appx IN31-29 5.1.1 Setting the Existing DR-ID 300CL .... Appx IN31-29 5.1.2 Settings the DR-ID 300CL to
be Added ..................................... Appx IN31-30 5.1.3 Performing Study List Setup ........... Appx IN31-30
5.2 Deleting the Cluster-connected DR-ID 300CL ...... Appx IN31-31
5.3 Replacing the Cluster-connected DR-ID 300CL ...... Appx IN31-32
Appendix 32 Installing and Setting up the Control Software as Software Product ........................... Appx IN32-1
Appendix 33 Laptop PC ....................................... Appx IN33-1
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 32
Appendix 34 Setting up the USB Flash Drive .... Appx IN34-1 1. Setting up the Security Software ................. Appx IN34-1
1.1 Setting up the Security Area .................. Appx IN34-2 1.2 Setting the Password ............................. Appx IN34-3 1.3 Confirming the Login to the Security Area ... Appx IN34-4 1.4 Creating a Folder .................................. Appx IN34-5
2. Changing the Drive Letter of the USB Flash Drive .................................... Appx IN34-6
Appendix 35 Patient Information Display Linkage (for V6.1 or Later) ........................... Appx IN35-1
1. Hardware Setup ........................................... Appx IN35-2 2. Software Setup ............................................ Appx IN35-3
2.1 Setting in the DR-ID 300CL .................... Appx IN35-3 2.1.1 Installing the Application for
the Remote Display ....................... Appx IN35-3 2.1.2 Installing the Option Key ................ Appx IN35-6 2.1.3 Reinstalling the AP Key ................. Appx IN35-6 2.1.4 Checking the IIS ........................... Appx IN35-7 2.1.5 Setting the Wake-on-LAN Utility ...... Appx IN35-8 2.1.6 Setting the Automatic Shutdown of
the Remote Terminal ..................... Appx IN35-9 2.1.7 Setting the Configuration Related to
the Patient Information Display Linkage ............................ Appx IN35-12
2.2 Setting in the Remote Terminal .............. Appx IN35-14 3. Verification ................................................... Appx IN35-17
3.1 Verifying the Patient Information Display ... Appx IN35-17 3.2 Verifying Wake-on-LAN ......................... Appx IN35-18
4. Setup File Backup ........................................ Appx IN35-19 5. Changing
the Patient Information Display Items .......... Appx IN35-20 5.1 Display/Nondisplay of
the Patient Information Items ................. Appx IN35-21 5.2 Changing the Display Order of
Patient Information Items ...................... Appx IN35-22 5.3 Changing the Font Size/Color of
Patient Information Items ....................... Appx IN35-23
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 33
Appendix 36 Remote Image Display .................. Appx IN36-1 1. Hardware Setup ........................................... Appx IN36-2 2. Software Setup ............................................ Appx IN36-3
2.1 Setting in the DR-ID 300CL .................... Appx IN36-3 2.1.1 Installing the Application for
the Remote Display ....................... Appx IN36-3 2.1.2 Installing the Option Key ................ Appx IN36-6 2.1.3 Reinstalling the AP Key ................. Appx IN36-6 2.1.4 Checking the IIS ........................... Appx IN36-7 2.1.5 Setting the Wake-on-LAN Utility ...... Appx IN36-8 2.1.6 Setting the Automatic Shutdown of
the Remote Terminal ..................... Appx IN36-9 2.2 Setting in the Remote Terminal .............. Appx IN36-12
2.2.1 PC Terminal ................................. Appx IN36-12 2.2.2 Mobile Terminal ............................ Appx IN36-15
3. Verification ................................................... Appx IN36-16 3.1 Verifying the Image Display ................... Appx IN36-16 3.2 Verifying Wake-on-LAN ......................... Appx IN36-17
4. Setup File Backup ........................................ Appx IN36-18
5. Changing Display Layout ............................. Appx IN36-19 5.1 Changing the Items Displayed in
the Patient Information Display Area ...... Appx IN36-20 5.2 Changing the Items Displayed in
the Image Processing Button Display Area ... Appx IN36-21 5.3 Changing the Items Displayed in
the Exposure Information Area (Displaying Overlay Items) ..................... Appx IN36-22
5.4 Changing the Items Displayed in the Exposure Information Area (Displaying an Exposed Image) .............. Appx IN36-24
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 34
Appendix 37 Energy Subtraction (ES) ............... Appx IN37-1 1. Hardware Setup ........................................... Appx IN37-2 2. Software Setup ............................................ Appx IN37-3
2.1 Installing the ES Option Key .................. Appx IN37-4 2.2 Reinstalling the AP Key ......................... Appx IN37-4 2.3 Setting the Position ID ........................... Appx IN37-5 2.4 Set up Configurations Related to
Energy Subtraction ................................ Appx IN37-6 2.5 Setting up the User Utility Related to
Energy Subtraction ................................ Appx IN37-8 2.5.1 Starting User Utility ....................... Appx IN37-8 2.5.2 Displaying the ES Exposure Menu ... Appx IN37-8 2.5.3 Setting up the ES Exposure Menu ... Appx IN37-9 2.5.4 Exiting User Utility ......................... Appx IN37-10
3. Verification ................................................... Appx IN37-11 3.1 DR-ID 300CL-AP Startup ....................... Appx IN37-11 3.2 Verifying the ES Exposure ..................... Appx IN37-11
4. Setup File Backup ........................................ Appx IN37-12
Appendix 38 Tomosynthesis............................... Appx IN38-1 1. Hardware Setup ........................................... Appx IN38-2 2. Software Setup ............................................ Appx IN38-3
2.1 DR-ID 300CL Software Setup ................ Appx IN38-3 2.1.1 Installing
the Tomosynthesis Option Key ....... Appx IN38-3 2.1.2 Reinstalling the AP Key ................. Appx IN38-4 2.1.3 Setting the Position ID
(PositionIDDefault.ini) .................... Appx IN38-4 2.1.4 Setting a Module for Tomosynthesis ... Appx IN38-5 2.1.5 Setting the Service Utility ............... Appx IN38-8 2.1.6 Setting the User Utility ................... Appx IN38-12
2.2 PPU Software Setup .............................. Appx IN38-13 2.2.1 Setting the IP Address ................... Appx IN38-13 2.2.2 Setting
the SQL Server Remote Connection ... Appx IN38-13 3. Verification ................................................... Appx IN38-14
3.1 Starting the DR-ID 300CL-AP ................ Appx IN38-14 3.2 Checking the Tomosynthesis ................. Appx IN38-14
4. Setup File Backup ........................................ Appx IN38-15 5. PPU Data Initialization ................................. Appx IN38-16 6. Time Synchronization for
the DR-ID 300CL and PPU .......................... Appx IN38-17 6.1 Settings on the DR-ID 300CL Side ......... Appx IN38-17 6.2 Settings on the PPU Side ...................... Appx IN38-19
7. Setting of the PPU at the Time of Changing the IP Address of DR-ID 300CL ... Appx IN38-20
8. Tomosynthesis Menu List ............................ Appx IN38-21
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 35
Appendix 39 Remote Desktop ............................ Appx IN39-1 1. Hardware Setup ........................................... Appx IN39-2 2. Software Setup ............................................ Appx IN39-3
2.1 Setting in the DR-ID 300CL .................... Appx IN39-3 2.1.1 Installing the Remote Desktop
Option Key ................................... Appx IN39-3 2.1.2 Reinstalling the AP Key ................. Appx IN39-4 2.1.3 Setting the UltraVNC ..................... Appx IN39-4
2.2 Setting in the Mobile Terminal ................ Appx IN39-10 2.2.1 Setting the Wyse PocketCloud ........ Appx IN39-10 2.2.2 Setting the Connection of
Wyse PocketCloud ........................ Appx IN39-10 3. Verification ................................................... Appx IN39-12
3.1 Starting the DR-ID 300CL-AP ................ Appx IN39-12 3.2 Checking the Operation of
Remote Desktop ................................... Appx IN39-12 4. Solving Connection Problems ...................... Appx IN39-13
4.1 When the VNC Connection Fails ............ Appx IN39-13 4.2 When the Performance of
the Remote Desktop is Unstable ............ Appx IN39-14 4.3 When the Remote Operation from
the Mobile Terminal to DR-ID 300CL Cannot be Performed ............................ Appx IN39-14
Appendix 40 DoseSR (Output for Exposure Dose Management) ...... Appx IN40-1
1. Software Setup ............................................ Appx IN40-2 1.1 Setup of Information on This Equipment ... Appx IN40-2 1.2 Registering
the Exposure Dose Management System... Appx IN40-4 1.3 Setting up the Exposure Dose
Management System Output .................. Appx IN40-7 2. Verification ................................................... Appx IN40-8
2.1 Starting the DR-ID 300CL-AP ................ Appx IN40-8 2.2 Verifying Sending of
Exposure Dose Information When Finishing QA ......................................... Appx IN40-8
2.3 Verifying Sending of Exposure Dose Information When Redistributing Study .............................. Appx IN40-8
3. Setup File Backup ........................................ Appx IN40-9
Appendix 41 Screening Examination Order Issuance .............................. Appx IN41-1
DR-ID 300CL Service Manual
010-305-07E 09.12.2012 FM6055 36
Appendix 42 DICOM MG Storage: Digital Mammogram Transfer ....... Appx IN42-1
1. Hardware Setup ........................................... Appx IN42-3 2. DR-ID 300CL Software Setup ...................... Appx IN42-4
2.1 Setting the Configuration Related to the DICOM MG Storage Function ........... Appx IN42-4
2.2 Setting Up on the DR-ID 300CL for DICOM MG Storage Function ................ Appx IN42-7
2.3 Registering on the Other Connected Equipment for DICOM MG Storage Function ................ Appx IN42-10
2.4 Setting Up the Automatic Image Distribution Function (DISTRIBUTION CODE) ........................ Appx IN42-13
2.5 Setting Up CONNECTING EQUIPMENT ... Appx IN42-14 2.6 Setting Up the UID Issuance Method ...... Appx IN42-14 2.7 Changing the Name Elements ............... Appx IN42-15
3. Connection Checkout .................................. Appx IN42-16 3.1 DR-ID 300CL-AP Startup ....................... Appx IN42-16 3.2 Image Transfer to
Other Connected Equipment .................. Appx IN42-16 4. Setup File Backup ........................................ Appx IN42-17
Appendix 43 Adding the NIC ............................... Appx IN43-1 1. Adding a Serial Connector ........................... Appx IN43-1
1.1 Mounting PWLA8391GTL ...................... Appx IN43-1 1.2 Mounting EXPI9301CT .......................... Appx IN43-2
2. Updating Network Driver ............................. Appx IN43-3 2.1 EXPI9301CT ......................................... Appx IN43-3
2.1.1 Updating EXPI9301CT network driver ........... Appx IN43-3
2.1.2 Confirming the update EXPI9301CT network driver .............................. Appx IN43-4
DR-ID 300CL Service Manual010-305-07E 09.12.2012 FM6055 37
PC: Performance Check
DR-ID 300CL Performance Checklist ............................... PC-1
DR-ID 300CL Config List ................................................... PC-2 1. IMAGE MODALITY ................................................. PC-2 2. QA FUNCTION ....................................................... PC-4 3. PRINTER ............................................................... PC-4 4. CSL/IDT FUNCTION ............................................... PC-6 5. CONFIG OPTION ................................................... PC-8 6. CONFIG QC ........................................................... PC-8 7. CONFIG QR ........................................................... PC-8 8. CONFIG FIX ........................................................... PC-8
DR-ID 300CL System Configuration List ......................... PC-9
Contents (DR-ID 300CL Service Manual)
DR-ID 300CL Service Manual010-305-07E 09.12.2012 FM6055 38
BLANK PAGE
BLANK PAGE
BLANK PAGE
Safety Precautions-I
Safety Precautions
Safety Precautions-II
Control Sheet Issue date Revision number Reason Pages affected
08/20/2009 01 New release (FM5581) All pages 04/15/2010 02 Revision for V4.0 (FM5668) I, II, 1–10 02/28/2011 03 Revision for V5.0 (FM5818) I, II, 2 12/15/2011 05 Revision for V6.1 (FM5944) I, II, 2
DR-ID 300CL Service Manual010-305-02E 04.15.2010 FM5668
Safety Precautions-1
Safety Precautions-1
1. Safety Precautions
Precautions to be observed during operations To prevent danger to humans and serious accidents, observe the following.
WARNING
Unless indicated otherwise, be sure to turn off the power switch before maintenance work, and disconnect the power plug from the outlet. Even if the power switch is OFF, 100 to 240VAC (differs according to machine specifications) will still be supplied to the primary input terminal of the unit if the power cable is connected, and this may result in possible electric hazards. As neckties, scarves, and necklaces may get caught in the unit
during work, remove these items or take precautions against such accidents. Never modify the unit unless otherwise specified. Modification of the safety unit (interlock, breaker, temperature
fuse, etc.) will cause serious problems as it is not only dangerous but violates the related laws and regulations.
CAUTIONS The grounding wire must be connected. To avoid damages to the parts/units due to static discharge, use
a wrist strap putting one end on your wrist and the other end on the grounded section of the equipment, as illustrated below. When doing so, make sure to unplug the power connector from
the socket to avoid electrical hazards.
00010003.EPS
When performing settings for several software programs, ensure that applications other than those of specified procedures must have been shut down appropriately.
DR-ID 300CL Service Manual010-305-05E 12.15.2011 FM5944
Safety Precautions-2
Safety Precautions-2
This equipment is affixed with a label. Follow the instructions provided on the label when performing necessary works.
DXL00010002.ai
Shock Hazard Precautions
Control Unit When installing, using, or servicing the Control Unit (hereinafter referred to as the PC), observe the following precautions. For details, refer to the User’s Guide, which is supplied with the PC. Prevention of electric shock
For the PC main body and monitor, use a power cord provided with a 3-prong plug. Connect the PC main body and monitor to a power outlet that is properly grounded. If the use of an extension cord is required, use a three-wire type having a grounding pin.
LCD WARNING
A high voltage is applied to the inside of the LCD monitor. To avoid safety hazards such as electric shock hazard, never access the interior of the LCD. The LCD monitor has no parts that can be serviced by a service engineer.
Power supply WARNING
When supplying power to system components, do not use a multi- outlet power strip or extension cable because the capacity of such a power strip or extension cable may be exceeded to create a fire hazard or shock hazard.
DR-ID 300CL Service Manual010-305-02E 04.15.2010 FM5668
Safety Precautions-3
Safety Precautions-3
Connection Precautions WARNING
Be sure that only specified external devices are connected to the apparatus.
Isolation Precautions WARNING
To assure complete system isolation, do not install unspecified accessories or options.
Disinfection Precautions WARNING
Make sure that a legal limit is not exceeded by the respiratory concentration of an employed disinfectant containing a solvent. Some disinfectants may pose a hazard to human health. Before using any disinfectant, be sure to thoroughly read its documentation.
WARNING
Do not use the following disinfectants. If you use them, the apparatus becomes damaged so that the quality, performance, and safety of the apparatus cannot be guaranteed. Chlorine disinfectants that readily corrode metals and rubber
members. Disinfectants that are described in their documentation as being
inapplicable to metals, plastics, rubber members, and paints. Formalin gases or spray-type disinfectants that may enter the
apparatus. It is recommended that ethanol for disinfection be used as a disinfectant. Before using a disinfectant, thoroughly read its documentation.
Installation Precautions When installing, using, or servicing the PC, observe the following precautions. For details, refer to the User’s Guide, which is supplied with the PC.
Installation site Exercise care so that no dust or foreign matter enters an opening in the PC or the interior of the PC. Do not install the PC near a heating device or other heat source. When mounting the PC on a floor, ensure that it is positioned at least 5 cm above the surface of the floor.
Precautions to observe when servicing the interior of the PC (1) Turn off the PC, monitor, and other currently used peripheral devices. (2) Disconnect the power cord from the power outlet (to avoid electric
shock hazard). (3) Open the I/O panel on the rear of the PC main body, and touch a
metal surface to discharge any static buildup on your body before accessing the interior of the PC.
(4) If it is necessary to remove, for instance, a circuit board or a memory module, handle it without directly touching its connector or IC pin and place it in an antistatic bag.
Environmental conditions Avoid installing the machine at the following places: Places where the temperature changes drastically Places near heaters or other heat sources Places where the machine is likely to be wet or submerged Places where the machine is likely to come into contact with any
corrosive gas Dusty or dirty places. Places where the machine is subjected to
constant vibration or shock. Places where the machine is subjected to excessive vibration or
shock. Places exposed to direct sunlight Also, be sure that the monitor is not installed at the following places: Places where the monitor is surrounded, such as in a rack case Places whether the display surface is exposed to direct sunlight Places near a transformer, another monitor, a power line, or a 0.02
Gauss or more of AC magnetic field
DR-ID 300CL Service Manual010-305-02E 04.15.2010 FM5668
Safety Precautions-4
Safety Precautions-4
Cleaning Precautions CAUTION
For exterior cleaning, use a cloth moistened with neutral detergent or ethanol. Do not use thinner, benzine, or other similar solvent because the exterior will corrode.
DR-ID 300CL Service Manual010-305-02E 04.15.2010 FM5668
Safety Precautions-5
Safety Precautions-5
2.1 Electromagnetic Compatibility (EMC) Related Standard
As required by IEC 60601-1-2:2001+A1:2004/EN 60601-1-2:2001+A1: 2006, this equipment complies with the following internationally recognized standard regarding Electromagnetic Compatibility (EMC). EMC standard : IEC 60601-1-2:2001+A1:2004 This does not guarantee that there is no harmful electromagnetic interference under any installation environment. This equipment can generate, use and radiate radio frequency energy. If the equipment is not installed and used in accordance with the instructions, or if peripheral devices that are not complied with the EMC standard, harmful interference may be generated under a particular environment causing malfunction of the equipment and other devices. If this equipment causes harmful interference to other devices, or if this equipment is affected by interference from other devices, the user is encouraged to try to correct the interference by one or more of the following measures: Reorient or relocate the device. Increase the separation between the equipment. Connect the equipment into an outlet on a circuit different from that to
which the other device(s) are connected. If the problem cannot be solved with the above measures, stop using this equipment and consult the manufacturer or our official dealer for help.
WARNING
Do not place devices generating electromagnetic wave near this equipment.
This Service Manual contains the names of standards applied in IEC 60601-1-2. The names of standards applied in EN 60601-1-2 are replaced as follows.
IEC 60601-1-2 (descriptions in this Service Manual) Standards applied in EN 60601-1-2
CISPR 11 EN 55011
IEC 61000-3-2 EN 61000-3-2
IEC 61000-3-3 EN 61000-3-3
IEC 61000-4-2 EN 61000-4-2
IEC 61000-4-3 EN 61000-4-3
IEC 61000-4-4 EN 61000-4-4
IEC 61000-4-5 EN 61000-4-5
IEC 61000-4-6 EN 61000-4-6
IEC 61000-4-8 EN 61000-4-8
IEC 61000-4-11 EN 61000-4-11
Safety Precautions-6
Safety Precautions-6
2.2 F