Upload
others
View
6
Download
0
Embed Size (px)
Citation preview
DROPLET FORMATION AND ENTRAINMENT IN LIQUID-GAS
MICROFLUIDIC SYSTEMS
A Thesis Presented
By
Pooyan Tirandazi
to
The Department of Mechanical and Industrial Engineering
in partial fulfillment of the requirements
for the degree of
Master of Science
in the field of
Mechanical Engineering
Northeastern University Boston, Massachusetts
August 2017
ii
TABLE OF CONTENTS
LIST OF FIGURES……………………………………………………………………………………………………………...iii
LIST OF TABLES………………………………………………………………………………………………………………. vi
ABSTRACT……………………………………………………………………………………………………………………… vii
Chapter 1. Introduction …………………………………………………………………………………………………….. 1
1.1 Microfluidics and Droplet Microfluidics ......................................................................... 1
1.2 Droplet Generation in Microfluidic Networks ................................................................. 3
1.3 A Guide Through the Thesis ............................................................................................ 7
Chapter 2. Gas-Liquid Droplet Microfluidics …………………………………………………………………….. 9
Chapter 3. Experimental ………………………………………………………………………………………………….. 12
3.1 Fabrication of Microfluidic Devices .............................................................................. 12
3.2 Control and Modification of Microchannel Surfaces..................................................... 15
3.3 Experimental Setup ........................................................................................................ 19
Chapter 4. Flow Regime Mapping …………………………………………………………………………...………. 22
Chapter 5. Dripping Regime Characterization ……………………………………………………………………27
5.1 Generation Frequency .................................................................................................... 27
5.2 Droplet Morphology ...................................................................................................... 29
5.3 Droplet Monodispersity ................................................................................................. 35
Chapter 6. Conclusions ………………………………………………………………………………………………........ 38
REFERENCES…………………………………………………………………………………………………………………... 40
iii
LIST OF FIGURES
Figure 1. A comparison of mixing reaction in a standard pressure-driven continuous-flow
microfluidic platform (a) and using a droplet-based microfluidic system (b). In a continuous-flow
scheme due to wall effects and different residence times mixing happens slower. However, by
digitizing the mixing reaction inside distinct droplets, each droplet is mixed much faster and the
total mixing time is reduced significantly as a result [13]. .............................................................. 2
Figure 2. Schematic representation of the most frequently used geometries in microfluidics for
droplet formation. (A) T-junction or cross-flow geometry, (B) co-flow geometry, and (C) flow-
focusing geometry. ........................................................................................................................... 5
Figure 3. Representation of dripping and jetting inside microfluidic capillary tubes [23]. (A)
Dripping regime; here droplet formation occurs at the tip of the inner capillary in the dripping
regime. (B) Jetting regime; in this case instabilities of an extended liquid jet creates droplets
which are commensurate to the jet diameter and are usually smaller in comparison to the dripping
regime. ............................................................................................................................................. 5
Figure 4. Axisymmetric flow-focusing architecture for controlled production of droplets using a
focusing air in a non-microfluidic format [52]. Here, the applied air pressure across the circular
orifice results in breakup of the liquid jet into individual droplets. ............................................... 10
Figure 5. Schematic representation of the fabrication procedure for production and assembly of
microfluidic devices. (1) Required channel geometry is designed and printed on transparency
sheets. (2) UV photolithography process is performed using the printed mask to define the
patterns on a negative photoresist that is coated on a silicon substrate. (3) After mold fabrication
soft lithography using PDMS prepolymer is performed to replicate the channels from the mold
into the elastic polymer. (4) After PDMS being solid, each device is cut and punched for the
inlets/outlets. (5) PDMS chips are bonded to a clean glass slide after being plasma cleaned. (6)
Final microfluidic device for water droplet formation in air. (7) Image of the flow-focusing
junction under microscope. Liquid water is injected through the middle channel and meets the
two side air streams at a flow-focusing junction. The microchannels’ depth is 40μm. ................. 14
Figure 6. Schematic of the method used treating the microchannel walls with fluorosilane
chemical. All the setup are placed inside a vacuum chamber in order to prevent the deposition of
toxic fluorosilane vapor into the environment. .............................................................................. 16
iv
Figure 7. Interaction of liquid and gas in different channel conditions. (A) After plasma
treatment; in this situation channels are hydrophilic and as a result the liquid water tends to
adhere to the side walls of the microchannel once it enters the junction. (B) After modifying the
channel walls with vaporized fluorosilane. In this case, the droplets could form in the channel.
However, due to the challenges in the coating process the generation was not reproducible. (C)
After post-baking step; here the PDMS has regained its hydrophobic nature and droplets maintain
their morphology after generation.................................................................................................. 17
Figure 8. Comparison of the surface tension forces between a glass-PDMS microchannel (left)
and a PDMS-PDMS microchannel (right). Using the provided contact angles in the table, the net
surface tension force for a glass-PDMS combination is less. Therefore, it provides less resistance
during the course of droplet generation and movement inside the channel which results in a more
reproducible formation process. ..................................................................................................... 19
Figure 9. Schematic of the flow circuit for liquid and gas control. ............................................... 20
Figure 10. Actual image of different parts of the experimental setup. (1) Air desiccator, (2)
manual coarse pressure regulator, (3) voltage-controlled valves, (4) mass flow sensors, (5) needle
valves, (6) liquid syringe pump, (7) white light source for the microscope, (8) high-speed camera
connected to the microscope, and (9) microfluidic chip placed on the microscope with all the
ports connected to their corresponding tubing. .............................................................................. 21
Figure 11. Flow regime map of water-in-air droplet formation in a planar flow-focusing
microfluidic device. Different flow regions are distinguished with different colors. The transition
between the flow regimes is also represented as a finite shaded region rather than a solid line to
account for the uncertainties associated with the experiments and calculations of the We values.
....................................................................................................................................................... 25
Figure 12. Comparison of the flow map for chips with different liquid channel sizes. The area for
the Dripping region (the green region) is reduced as the liquid channel size increases. However,
the Threading and Co-flow regions have expanded as a result of increase in the surface tensions
forces that holds the thread and prevents the detachment at lower gas flow rates. ........................ 26
Figure 13. Experimental data of droplet generation frequency as a function of air flow rate for
different liquid flow rates. We can see that there is a relatively linear correlation between the
frequency value and the air flow rate. ............................................................................................ 28
Figure 14. Experimental data of droplet generation frequency at increasing liquid flow rate for
two different gas flow rates. We can see that there is a relatively linear correlation between the
frequency value and the liquid flow rate. ....................................................................................... 28
Figure 15. Experimental data of the non-dimensioalized droplet generation frequency.
Generation frequency in this system can be scaled as ƒ∝QGQL. ................................................... 29
Figure 16. Experimental data of droplet width inside the microchannel for different liquid and
gas flows. ....................................................................................................................................... 30
v
Figure 17. Experimental data of droplet length inside the microchannel for different liquid and
gas flows. ....................................................................................................................................... 30
Figure 18. Snapshots of droplets moving inside microchannel under different gas flow rates. ... 31
Figure 19. Corrected length used in non-dimensionalizing droplet morphology data In high liquid
flows, droplet exhibit a tail-shape which deviate from their common circular morpholy. ............ 31
Figure 20. Schematic of the Flow-Focusing geometry and the nomenclature used in scaling
analysis for droplet morphology. ................................................................................................... 34
Figure 21. Droplet length normalized with the minimum width of the droplets in the Dripping
regime as a function of gas Weber number (WeG). Minimum droplet width and length are
assumed to be equal values that are obtained based on the minimum volume of the prefilled
droplet during the formation process at high gas flow rates where droplet tends to have a more
circular morphology. This minimum is dependent on the geometry of the flow-focusing
microchannel and is calculated to be about 86μm for the geometry used in out experiments. ...... 34
Figure 22. Droplet length normalized with the minimum length of the droplets in the Dripping
regime as a function of gas Weber number (WeG). Minimum droplet width and length are
assumed to be equal values that are obtained based on the minimum volume of the prefilled
droplet during the formation process at high gas flow rates where droplet tends to have a more
circular morphology. This minimum is dependent on the geometry of the flow-focusing
microchannel and is calculated to be about 86μm for the geometry used in out experiments. ...... 35
Figure 23. (Left) Experimental images of the collected droplets in the immiscible liquid solution.
(Right) detected droplets using image analysis with MATLAB. ................................................... 36
Figure 24. Actual images and size distributions of samples of 500 collected droplets within the
Dripping and Multi-Satellite Formation regimes. Liquid flow rate for both cases is 5µL/min.
Multi-Satellite Formation is obtained at higher WeG according to the flow map. In the Dripping
region by having a constant liquid and gas flow rate the generated droplets are uniform in size. In
Multi-Satellite Formation regime, however, small daughter droplets are generated beside the main
droplet which results in polydisperse distribution of the droplets. ................................................ 37
vi
LIST OF TABLES
Table 1. Measured contact angles for 10 samples of PDMS and glass substrates with the same
preparation conditions of plasma treatment followed by a prebake process. The prebake process
results in regaining the hydrophobic nature of PDMS. The glass surface also becomes less
hydrophilic after the prebake process. Smaller contact angle hysteresis of the glass substrate
results in easier droplet movement during formation. Therefore, it will help to obtain a wider
Dripping regime and higher monodispersity. ................................................................................ 17
vii
ABSTRACT
Controlled generation of droplets in microfluidic networks is a promising tool for numerous
applications in biochemistry and material sciences. Droplet-based microfluidic systems are
typically utilized for creating dispersions of a liquid or gas within a second continuous liquid
phase, such as creating uniform liquid emulsions or gas bubbles (foams). In this work, we have
studied some of the fundamental aspects of liquid droplet formation within a continuous gas flow
for creating uniform aerosols inside microfluidic channels. We have experimentally investigated
interactions of liquid water within a high-speed air flow inside confined flow-focusing
microchannels in which we identify six distinct flow regimes: Co-flowing, Threading, Plugging,
Dripping, Multi-Satellite Formation, and Jetting. These flow regimes and their transitions are
plotted and characterized based on the Weber number (We) of the two phases. Generation
frequency, morphology, and monodispersity of the droplets are characterized in more detail for
the Dripping regime and scaling laws are provided to better elucidate the mechanism of droplet
formation for this system. Results of this work establish a relationship between generation
frequency and the product of the liquid and gas flow rates. However, droplet morphology (length
and width) is exclusively dependent on the air flow rate. Finally, we demonstrate the production
of monodisperse droplets (d<100μm and σ/d<0.05) at kHz formation rates in liquid-gas
microfluidic systems. The results of this work provide practical and useful guidelines for precise
oil-free delivery of ultra-small volumes of fluid which can be integrated in Lab-on-a-Chip (LOC)
and micro-Total-Analysis Systems (μTAS) for a variety of applications in biochemical research
and material synthesis.
1
Chapter 1. Introduction
1.1 Microfluidics and Droplet Microfluidics
Microfluidics has been increasingly developed over the past two decades. A consequence of this
galloping growth is transforming the way in which many chemical and biological experimentations
are performed [1]–[4]. Thus, it has become an essential tool in a wide range of medical researches
and technical applications. Despite the wide spectrum of microfluidic devices and schemes, they
can be categorized into two major groups, continuous-flow and droplet-based systems [5].
Continuous-flow scheme is characterized by manipulation of a continuous fluid stream inside
usually-confined microchannels. In a droplet-based system, however, one fluid is broken up within
a second phase to create discrete droplets. The development of microfluidic platforms in different
science and technologies has been driven by a number of fundamental characteristics related to
system miniaturization [6]. As such, many advantages are associated with these miniaturized
platform, including the small sample volumes consumption, low-cost fabrication methods,
enhanced operational flexibility, and the ability to integrate different components within more
complex networks for different analytical schemes [7]. Droplet-based microfluidics enables even
more capabilities while overcoming some of the inherent limitations of the continuous-flow
systems. In a continuous-flow scheme, intimate contact of the continuous fluid stream with
microchannel walls is associated with parabolic velocity profile and therefore, different residence
times [8]. Moreover, near-zero velocities at the wall result in fouling and contamination of the flow
and the whole process as a consequence [9]. Creating microdroplets has emerged as a potential and
appealing approach to perform a wide range of chemical and biological processes in a peerless
fashion [10], [11]. The use of picoliter to nanoliter scale droplets further reduces sample
consumption and increases surface-are-to-volume ratio which corresponds to much faster diffusion
rates [5]. More importantly, discrete droplets are ideally suited to isolate and compartmentalize a
2
sample or reaction which obviate issues of a continuous-flow scheme (Figure 1). Therefore, unlike
continuous-flow systems, this approach allows precise definition of the components of a process
inside distinct droplets. In addition, droplet-based systems are intrinsically associated with high
throughputs and parallel processing since typical droplet-based platforms can generate hundreds to
thousands of drops per second allowing for any post process to be performed in parallel on
individual droplets [12].
Figure 1. A comparison of mixing reaction in a standard pressure-driven continuous-flow microfluidic
platform (a) and using a droplet-based microfluidic system (b). In a continuous-flow scheme due to wall
effects and different residence times mixing happens slower. However, by digitizing the mixing reaction
inside distinct droplets, each droplet is mixed much faster and the total mixing time is reduced significantly
as a result [13].
Microfluidic technologies provide tremendous advantages for precise and rapid manipulation of
fluids at sub-microliter scale inside micrometer channels. All the advantages of microfluidic
systems come at the cost of understanding physics of the flow at microscale [14], [15]. These
physics can be better understood in terms of a few dimensionless numbers. Typically, gravity
3
effects are neglected in the microfluidic domain. The importance of gravity to surface tension forces
is expressed in term of Bond number (Bo);
Bo= ∆ρgL
2
σ (1.1)
where Δρ is the density difference between the dispersed and continuous phase, g is the
gravitational acceleration, D is the characteristic length scale, and σ is the interfacial tension. Due
to micrometer dimensions of the channels in microfluidic systems the effect of interfacial tension
is more pronounced and as a result Bo<<1. The Reynolds number (Re) compares the relative
importance of inertial and viscous forces and is defined as;
Re= ρUD
μ (1.2)
where U a characteristic velocity, and μ is the fluid viscosity. Similarly, the Weber number defined
as;
We= ρU2D
σ (1.3)
compares inertia to surface tension. Capillary number (Ca) which relates the ratio of viscous to
surface tension forces is the most commonly used dimensionless parameter in droplet microfluidic
systems [15] and is defined as;
Ca= μUσ
(1.4)
Expressing the flow in terms of the above dimensionless numbers facilitates better identification of
the key parameters in droplet generation and breakup, thereby unifying the behavior of a wide
variety of microfluidic systems.
1.2 Droplet Generation in Microfluidic Networks
Droplet formation and entrainment processes are the essential steps for using any droplet-based
system in applications which rely on discretizing a fluid sample. Many techniques for droplet
formation have been presented which can be classified in various ways [16]. Droplets in
4
microfluidic networks can be produced using either active or passive techniques. Active droplet
generation is achieved with the aid of additional energy inputs. Some of the commonly used
methods in this category utilize electrical, magnetic, mechanical, centrifugal, and thermal
mechanisms to actuate and control droplet breakup [17]. In passive methods, on the other hand,
there are no external actuation mechanisms and droplet formation solely relies on the interaction
of fluids interfaces inside special microgeometries. In comparison to active methods, passive
techniques are usually less complex and easier to implement.
Geometry of the microchannel, especially in passive formats, is of great importance for controlling
droplet generation. Various microfluidic device geometries have been demonstrated for this
purpose among which T-junction (cross-flow), co-flow, and flow-focusing configurations are the
most prevalent formats (Figure 2) [18]. In a T-junction geometry, one channel intersects a main
channel perpendicularly where one phase disperses within a continuous flow of a different phase
[19]. In a co-flow configuration, the continuous and dispersed phase flow in parallel channels
whereas the continuous phase surrounds the dispersed fluid [20]. This geometry is usually
fabricated using inserted capillary tubes and is challenging to fabricate and implement in quasi-
two-dimensional microfluidic formats. Flow-focusing geometries utilize a constriction in their
junction where both phases are hydrodynamically focused through a small orifice [21]. Since the
number of geometrical parameters in a flow-focusing geometry is higher in comparison to a T-
junction and co-flow configurations, flow-focusing systems offer a more versatile generation
scheme for controlling the droplet size and generation frequency. Several works have studied these
geometries and other variations to obtain small droplets with high degree of monodispersity.
5
Figure 2. Schematic representation of the most frequently used geometries in microfluidics for droplet
formation. (A) T-junction or cross-flow geometry, (B) co-flow geometry, and (C) flow-focusing geometry.
Despite the different configurations for droplet formation, breakup process in microfluidic systems
can be divided into a few fundamental modes. Dripping and jetting formation of droplets are usually
the two most desired modes of breakup inside microfluidic channels to produce drops [22], [23].
Dripping mode is characterized by formation and breakup of droplets very close to the dispersed
channel inlet, whereas in the Jetting regime, a liquid jet elongates in the microchannel outlet where
the breakup occurs due to growing instabilities in the liquid column (Figure 3). Dripping regime is
associated with higher monodispersity yet bigger droplets at slower frequencies comparing to the
jetting regime. By increasing either the continuous phase or the dispersed phase flow rates,
transition from dripping to jetting regime could occur. The region associated with these regimes
and their transitions to other ones is usually mapped in a plot that usually consist of capillary
numbers of the two phases in the system [24].
Figure 3. Representation of dripping and jetting inside microfluidic capillary tubes [23]. (A) Dripping
regime; here droplet formation occurs at the tip of the inner capillary in the dripping regime. (B) Jetting
regime; in this case instabilities of an extended liquid jet creates droplets which are commensurate to the jet
diameter and are usually smaller in comparison to the dripping regime.
Another major classification in microfluidic production of drops is by considering the phases
involved in the generation process (liquid or gas). In this sense, three combinations are possible i.e.
liquid-in-liquid or emulsion generation, gas-in-liquid or bubble generation, and liquid-in-gas or
aerosol generation. Creating uniform microemulsions is the most widely used class in
microfluidics. In this case, a continuous liquid phase meets the dispersed liquid in one of the
6
aforementioned geometries where individual droplets are pinched off as a result of interfacial
tension and shear forces induced by the continuous phase. Droplet sizes can be accurately
controlled in this scheme and a variety of techniques have been proposed for subsequent
manipulation of the drops in other system processes such as fusion [25]–[27], fission [28], sorting
[29], [30], and mixing [31], [32] in liquid-liquid environments [33]. Therefore, numerous
applications in microbiology, drug discovery, and particle synthesis exist that highly rely on
controlled formation on liquid droplets within a continuous liquid stream [34]–[36]. For instance,
these microdroplets are ideal for encapsulating single biological cells where specific dynamics and
metabolites can be conveniently probed within each droplet [37], [38]. Another huge area of liquid-
liquid digital microfluidic droplets is performing Polymerase Chain Reaction (PCR) in a digitized
format [39]–[42]. This process which is known as ddPCR (digital droplet PCR) has attracted
tremendous attentions during recent years as a highly-sensitive approach for DNA amplification.
In this technique, by partitioning the sample into millions of droplets, PCR can be performed on
individual drops and the number of positive compartments in which a target sequence is detected
can be counted and subsequently related to the initial number of targets in the original sample.
Research on incorporating a gaseous phase in microfluidic systems has mostly focused on
generation of uniform microbubbles and segmented gas slugs within a liquid phase. This process
has been driven by a different set of applications mainly in food industry, pharmaceutical sciences,
and medicine. For example, microfluidic formation of bubbles is proven in creation of polymer-
lipid microbubbles (PLBs) as a template for fabrication of 3D porous materials and scaffolds [43].
Moreover, using encapsulated microbubbles as ultrasound contrast agents (UCA) in medical
imaging has gained a lot of attention for their potential in early detection and characterization of
diseases [44]. Well-established microfluidic techniques have also been developed in this class for
production of highly-monodisperse bubbles in liquid environments using similar device geometries
as liquid-liquid systems [45], [46]. However, there is a paucity of studies regarding microfluidic
7
techniques for formation of liquid droplets inside a gaseous flow. Previous works related to liquid
droplet formation in a gaseous flow is provided in the next chapter. The major challenge addressed
in this thesis concerns investigation of droplet formation in gaseous microflow inside similar planar
microfluidic networks as the previous classes. The objective of this study is to gain a better
understanding of the fundamental physics behind this process with the ultimate goal of creating
uniform micron-sized droplets completely in air in microfluidic devices.
1.3 A Guide Through the Thesis
In Chapter 2, an overview of droplet formation in gaseous environments is presented. Previous
researches in this regard are mostly geared towards droplet generation using an axisymmetric
focusing air in unconfined nozzle-like geometries.
In Chapter 3, we discuss the experimental aspects of this work. The methodology for microfluidic
chip fabrication is described. We examine droplet formation in a gas flow inside a microfluidic
flow-focusing device (MFFD) with mixed hydrophobicity surface conditions. We discuss
preparation of the microchannels for obtaining monodisperse droplets in confined microchannels.
Unlike previous studies that used different surface chemical coatings to render the microchannel
fully hydrophobic, we show that incorporating a mixed hydrophobic/hydrophilic PDMS-glass
hybrid structure without chemical coatings results in the most robust droplet generation in these
systems.
In Chapter 4, we provide the experimental results of different breakup scenarios that occur in the
liquid-gas domain inside the microchannel. We identify six distinct regimes which are: Co-flowing,
Threading, Plugging, Dripping, Jetting, and Multi-Satellite Formation [47]. The effect that
microchannel size has on droplet formation is investigated towards the development of these richer
flow maps.
8
In Chapter 5, particular attention is paid to the Dripping regime for the generation of uniform
droplets. Droplet morphology, monodispersity and generation frequency are characterized as
functions of gas and liquid flow rates in this regime for the first time. Scaling laws are also provided
for the droplet morphology and generation frequency to non-dimensionalize the results.
In Chapter 6, we provide conclusions of this work and the future avenues that can be pursued for
implementation of this system in next generation Lab-on-a-Chip (LOC) technologies.
9
Chapter 2. Gas-Liquid Droplet Microfluidics
Spontaneous generation of liquid droplets within a gaseous flow in microfluidic networks is a
newer approach to the conventional liquid-in-liquid systems. One of the main advantages of this
approach for droplet formation is the possibility of creating uniform particles purely in air
(monodisperse aerosols) and without the presence of a second liquid carrier. Therefore, the final
product can be readily attained obviating extra washing steps to remove the carrier oil [35].
Moreover, generation of droplets in a liquid medium is usually facilitated by adding surfactants to
the continuous liquid carrier. However, it is believed that using certain surfactants may impose
usage limitations mainly due to cross-contamination of the droplet contents [48], [49]. Surfactants
are clearly not required for droplet generation in a gaseous phase, ensuring preservation of the
droplets’ contents.
Some early works have investigated the breakup of a liquid jet into droplets using a focusing air in
unconfined nozzle-like geometries as in Figure 4 [50]–[54]. Here, a flowing liquid through a
capillary tube is forced through a coaxial round orifice located downstream of the tube. The liquid
stream is drawn by a focusing gas stream discharging through the nozzle into an infinitely large
chamber. In this scheme, drop breakup occurs as a result of the instability of a capillary jet and it
has been demonstrated that under certain conditions [54]–[56], monodisperse droplets can be
achieved from an unstable liquid jet.
10
Figure 4. Axisymmetric flow-focusing architecture for controlled production of droplets using a focusing air
in a non-microfluidic format [52]. Here, the applied air pressure across the circular orifice results in breakup
of the liquid jet into individual droplets.
Generation of monodisperse drops within a microfluidic system is an entirely different process due
to confinement induced effects [57]. Interaction of liquid and gas inside confined microchannel
geometries has also been widely studied in the operation of Proton Exchange Membrane Fuel Cells
(PEMFC) [58]–[67]. However, droplet generation in gaseous microfluidic systems that employ
similar architectures as those used in conventional oil-based systems has been the subject of a very
few studies during recent years. The gas flow velocities required for droplet generation are at least
an order-of-magnitude higher than those employed when using a highly viscous oil phase. As such,
inertial effects become more prominent and relevant in the fluid-fluid interactions of these systems.
Droplet detachment has been numerically and experimentally investigated in a T-shaped junction
under the introduction of a high-speed gaseous flow [68], [69]. It was demonstrated that the
mechanism responsible for the breakup of the drops transitions from hydrodynamic pressure
difference (arising from the microchannel confinement) at lower Re to inertial drag at higher Re as
the gaseous flow increases. Confined generation of aqueous droplets has also been shown in a
11
circular capillary co-flow system using gas as the continuous phase. Under relatively low liquid
flow rates (less than 1 μl/min) and gas velocities (below 3 m/s) uniform droplets between 250μm
and 320μm were obtained within concentric hydrophobic glass capillaries [70]. More recently,
confluence of liquid in a gaseous stream in various planar confined flow-focusing microgeometries
was studied. It was shown that over a wide range of liquid and gas flow rates three main regions
(i.e. Dripping, Jetting, and Stratified Flow) are observed within confined PDMS microchannels
[71]. While this study provided useful insights regarding the dynamics of gas-liquid interactions in
planar flow-focusing geometries, it did not address several key aspects of this system. For example,
the monodispersity and characteristics of the droplets generated within the Dripping region was not
presented. This is very relevant information towards the design of these systems since the Dripping
regime is one of the two desirable modes of operation for droplet generation. Furthermore, neither
study looked at the effect that flow conditions has on the droplet morphology or generation
frequency in this flow regime.
12
Chapter 3. Experimental
3.1 Fabrication of Microfluidic Devices
We fabricated the microfluidic devices using standard photolithography process with SU-8 and soft
lithography with polydimethylsiloxane (PDMS) [72]. First, the flow-focusing configuration was
designed in a CAD software (SolidWorks 2014) and then printed out on high-resolution
transparencies (CAD/Art services). A 4-inch single side polished silicon (Si) substrate is used as
the master mold. The mold is prepared using standard photolithography process with a negative
near-UV photoresist (SU-8 2050 Microchem Corp.). Initially, the substrate is cleaned with acetone
and isopropyl alcohol, and dried with a nitrogen gun. The clean wafer is spin-coated (Laurell
Technologies) with the photoresist. The spinning speed is initially set to 500rpm for complete
spreading of the photoresist followed by a final speed of 3000 rpm for 1 min to obtain 40μm
thickness. Residues of the photoresist also known as edge beads were removed from the sides of
the substrate using a razor blade to ensure a flat coated surface for better mask attachment and
exposure process. The coated wafer is then baked for solvent removal on a hot plate at 65°C and
95°C respectively. The prebaked mold is left at room temperature for about 5 min to reduce thermal
stresses in the photoresist layer. The transparency mask sheet which contains the chip geometries
is attached to a borosilicate glass sheet to provide a rigid support for the mask when covering the
wafer. The mask and the wafer are loaded in the mask alignment equipment (Quintel 4000 Mask
Aligner) and exposed to UV light through the printed mask for about 45s. The exposed wafer is
then post-baked with similar temperatures as the prebake step for 5 min and 10 min respectively.
Once the mold reached the room temperature after the post-bake, it was soaked in a solution of SU-
8 developer for defining the micropatterns on the photoresist. A hard-bake process at 120°C is
performed for improving the adhesion of the photoresist to the wafer as well as balancing the
thermal stresses in the photoresist film. Once the mold fabrication is complete, microfluidic chips
13
were casted out of the SU-8 master molds by performing standard soft lithography method.
Polydimethylsiloxane (PDMS) prepolymer mixture (Sylgard 184, Dow Corning) is prepared from
the elastomer base and curing agent at a 10:1 weight ratio. The solution is then placed onto the
Silicon mold and is degassed in a vacuum chamber. The assembly is cured at 80°C for
approximately 2 hours on a leveled hotplate. Hardened PDMS is carefully peeled from the mold.
The available area on a 4-inch silicon mold allows a number of devices to be replicated in a single
lithography process. Each microfluidic chip is cut out of the PDMS layer. After punching the
required flow inlets/outlets, each chip is cleaned with DI water and dried with a nitrogen gun. Each
piece is then bonded to a pre-cleaned standard microscope glass slide (Thermo Scientific) using an
oxygen plasma cleaner (Harrick Plasma). Finally, the chips are post-baked for 4 hours at 200°C
and can be used for experiments after being cooled down to room temperature. The fabrication
process and final device and the microchannel configuration is depicted in Figure 5.
14
Figure 5. Schematic representation of the fabrication procedure for production and assembly of microfluidic
devices. (1) Required channel geometry is designed and printed on transparency sheets. (2) UV
photolithography process is performed using the printed mask to define the patterns on a negative photoresist
that is coated on a silicon substrate. (3) After mold fabrication soft lithography using PDMS prepolymer is
performed to replicate the channels from the mold into the elastic polymer. (4) After PDMS being solid, each
device is cut and punched for the inlets/outlets. (5) PDMS chips are bonded to a clean glass slide after being
plasma cleaned. (6) Final microfluidic device for water droplet formation in air. (7) Image of the flow-
focusing junction under microscope. Liquid water is injected through the middle channel and meets the two
side air streams at a flow-focusing junction. The microchannels’ depth is 40μm.
15
3.2 Control and Modification of Microchannel Surfaces
The wetting behavior of the microchannel walls plays a critical role in the performance of droplet-
based microfluidic systems that use a planar architecture. In conventional liquid-in-liquid systems
the surface of the microchannels is modified to reduce droplet interactions with the walls. In other
words, stable water-in-oil emulsions can be formed in hydrophobic microchannels and on reverse
emulsions (oil-in-water) the microchannels should be hydrophilic (oleophobic). Although PDMS
is intrinsically hydrophobic, the use of oxygen plasma for bonding renders them hydrophilic which
makes liquid handling difficult within microchannels. Moreover, compatibility issues of PDMS
surfaces with organic solvents [73], including a variety of the oils used in droplet-based systems,
pose additional challenges for droplet generation in liquid-liquid systems. Obtaining the conditions
that suit the generation process is usually achieved by manipulating the microchannel surfaces
through chemical coatings. A variety of techniques and chemicals are employed to achieve the
desired hydrophobicity as well as solvent compatibility [74]. However, the chemicals used for this
purpose are typically hazardous to work with. Moreover, applying a reproducible and uniform
coating especially in confined microchannels is a challenge from the preparation perspective.
Having a gaseous phase as the carrier fluid entirely eliminates the compatibility issue of PDMS in
our system. Figure 7A represents the behavior of droplet formation after plasma treatment in this
liquid-gas system. Droplets in this case show an affinity to ride on the side walls due to their
hydrophilic condition after exposure to plasma. As a result, droplets are not capable of keeping
their morphology during formation and tend to spread and form a liquid film on the side walls. To
overcome this issue, we implemented two methods for modify the properties of the microchannel
walls. In the first approach, we used a fluorosilane to coat the microchannels. Because of the silanol
groups existing on the surface of the PDMS, a wide spectrum of silanes can be used for surface
modification[74], [75]. We tested Trichloro(1H,1H,2H,2H-perfluorooctyl)silane (FDTS 97%,
Sigma-Aldrich) to modify the surface of the PDMS using vapor deposition [76]. Since coating
16
process prior to bonding caused some issues and prevented strong adhesion of the PDMS chips to
the glass substrate, we performed the deposition after the bonding step. The process for treating the
microchannels inner surfaces involves passing the vaporized FDTS into the bonded PDMS
microchannel as shown in Figure 6 [77]. For this purpose, a microdroplet of FDTS is placed on a
slide next to the previously-bonded chip, while the outlet of the chip is connected to the vacuum
port of a micro diaphragm pump (Parker Corp.) All the parts, including pump, chip, and slide are
placed inside a vacuum chamber. By turning on the pump the vaporized silane is directed towards
the inlets of the chip and is forced to pass through the microchannels and being deposited on the
channel walls. After 1 hour of coating process, the chip was disconnected from the vacuum pump,
and used for the droplet generation experiments. Results of droplet formation after applying silane
on the walls is shown in Figure 7B. Although, droplet generation after silane coating showed
improvements, the results were not consistent due to the challenges associated with the
experimental procedure of applying the chemical on the confined structure. Therefore, we tried a
different and more robust approach to tackle this issue.
Figure 6. Schematic of the method used treating the microchannel walls with fluorosilane chemical. All the
setup are placed inside a vacuum chamber in order to prevent the deposition of toxic fluorosilane vapor into
the environment.
In the second approach, a high-temperature post-baking step was used instead of chemical
treatment to achieve the surface conditions suitable for droplet generation in our system. Table 1
shows the measured contact angles for PDMS and glass substrates after the post-bake process. This
17
technique ensures that all the PDMS walls regain their hydrophobic nature in a fast and uniform
manner, minimizing the tendency of the droplets to adhere to the walls as shown in Figure 7C. This
process is very reproducible and does not involve the challenges of the previous technique.
Figure 7. Interaction of liquid and gas in different channel conditions. (A) After plasma treatment; in this
situation channels are hydrophilic and as a result the liquid water tends to adhere to the side walls of the
microchannel once it enters the junction. (B) After modifying the channel walls with vaporized fluorosilane.
In this case, the droplets could form in the channel. However, due to the challenges in the coating process the
generation was not reproducible. (C) After post-baking step; here the PDMS has regained its hydrophobic
nature and droplets maintain their morphology after generation.
Table 1. Measured contact angles for 10 samples of PDMS and glass substrates with the same preparation
conditions of plasma treatment followed by a prebake process. The prebake process results in regaining the
hydrophobic nature of PDMS. The glass surface also becomes less hydrophilic after the prebake process.
Smaller contact angle hysteresis of the glass substrate results in easier droplet movement during formation.
Therefore, it will help to obtain a wider Dripping regime and higher monodispersity.
Static Angle
(θS)
Advancing Angle
(θA)
Receding Angle
(θR)
Glass 60±4 69±5 52±5
PDMS 102±4 116±4 75±5
In confined microfluidic channels droplet morphology is bounded by the microchannel dimensions.
Thus, if the diameter of the generated droplet exceeds the height of the microchannel the droplet
will be in contact with the top and bottom walls which result in additional interactions that influence
18
the formation process. This condition occurred throughout the Dripping regime of our system. It
has been observed that using a non-hydrophobic glass substrate as the bottom wall enables a more
robust and reproducible dripping over a wider range of the flow map with higher monodispersity.
We believe this behavior can be related to the smaller contact angle hysteresis of a glass slide in
comparison to a PDMS surface. Based on the measurements of the static, advancing, and receding
contact angles for the glass and PDMS (see Table 1) surfaces we can schematically represent the
droplet inside the microchannel from a side view as shown in Figure 8 for each case. From these
measurement two things become apparent. First, post-bake treatment of the glass substrate
substantially reduces its hydrophilicity, rendering it partially hydrophilic. Secondly, the droplet
holding or restraining force from surface tension effects is smaller for the hybrid (glass substrate-
PDMS) structure than for the PDMS-PDMS one, due to the smaller overall contact angle hysteresis
effect. This can be quantified by comparing the added top and bottom hysteresis of the hybrid
structure ((γcosθRP - γcosθA
P) + (γcosθRG - γcosθA
G)) to the hysteresis from the PDMS-PDMS
structure (2(γcosθRP-γcosθA
P)). This results in a surface tension force for the hybrid structure that
is less than half that of the PDMS-PDMS structure. As a result, especially in lower flow rates of
the gaseous phase, we speculate that the higher surface tension force from the PDMS-PDMS
structure creates an added hindrance to the motion of the droplet while being formed and generated.
Therefore, obtaining a reproducible formation process within the flow conditions of the Dripping
regime in these structures becomes a challenging task, with the delineation of the boundary between
the Dripping regime and its neighboring regimes becoming more blurred.
19
Figure 8. Comparison of the surface tension forces between a glass-PDMS microchannel (left) and a PDMS-
PDMS microchannel (right). Using the provided contact angles in the table, the net surface tension force for
a glass-PDMS combination is less. Therefore, it provides less resistance during the course of droplet
generation and movement inside the channel which results in a more reproducible formation process.
3.3 Experimental Setup
A custom microfluidic test setup has been developed for the experimental investigation of the gas-
liquid droplet generation. The gas used as the continuous phase in all the experiments is air at room
temperature which is supplied from the facility compressed air source. The air is first dried by
passing through a desiccator and cleaned using 0.2μm filters (Wilkerson Corp.). Air is regulated in
multiple steps before being routed into the microfluidic chip inlets. A manual coarse pressure
regulator (SMC corp, up to 50 psi) reduces the pressure of the supply line. A voltage-controlled
pressure regulator (Proportion-Air, up to 18±0.04 psi) is used to deliver a precise air pressure into
the system. For real-time monitoring of the air flow rate, an in-line mass flow meter (Sierra Smart-
Trak2) is placed in the setup. All pressure regulators and flow sensors are powered using dedicated
24V DC power supplies. A data logger/switch unit (Agilent 34970A) is also used for data
acquisition which is controlled by a customized LabVIEW interface for controlling the valve and
collecting data from the flow meter. A needle valve (Swagelok) is also placed upstream of the chip
for a finer flow regulation prior to injection. The air stream bifurcates after passing through the
needle valve and is connected to the corresponding inlets of the chip through 1/16-inch PEEK tubes.
Liquid flow control is provided by a glass syringe using a high precision syringe pump (PHD2000
20
Harvard Apparatus). The accuracy of the injected flow rates are within 0.35% of the specified value
for the flow. Imaging and visualization of the formation process is facilitated using a high-speed
CMOS camera (Photron SA5) which is capable of providing a 7500 frames per second (fps) at the
resolution of 1024×1024. The camera is mounted to an inverted microscope (Nikon Ti-U) and
illumination is provided by a metal halide white light (Prior Scientific). Schematic and actual
images of the experimental setup are shown in Figure 9 and Figure 10 respectively.
Figure 9. Schematic of the flow circuit for liquid and gas control.
21
Figure 10. Actual image of different parts of the experimental setup. (1) Air desiccator, (2) manual coarse
pressure regulator, (3) voltage-controlled valves, (4) mass flow sensors, (5) needle valves, (6) liquid syringe
pump, (7) white light source for the microscope, (8) high-speed camera connected to the microscope, and (9)
microfluidic chip placed on the microscope with all the ports connected to their corresponding tubing.
In order to analyze the population of the generated drops, they are collected off the chip after
generation in air, inside a different liquid phase. For this purpose, n-Hexadecane (Sigma-Aldrich
Corp.) which is almost insoluble in water, is used as the collection medium. In order to prevent
droplets from merging inside the solution 2% of the nonionic surfactant span 80 (Sigma-Aldrich
Corp.) is mixed with the oil. The outlet of the microfluidic chip is cut and submerged in the oil bath
[78]. As a result, the generated droplets are collected downstream of the device outlet once they
transition into the oil medium.
22
Chapter 4. Flow Regime Mapping
Unlike liquid-liquid droplet microfluidic systems that are usually characterized by the Capillary
number (Ca) of the two fluids [24], [79], Ca in our experiments always remained below 0.02. In
contrast, the Reynolds number (Re) for the continuous gaseous phase was in the range between 10
and 600 indicating the relevance of inertia in this system. However, such high Re and low Ca for
all the flow conditions of the system do not clearly illuminate the fundamental mechanisms behind
the different flow regimes. Therefore, we employed the Weber number (We) to parametrize the
droplet breakup processes as it changed in a moderate range in our experiments (We ≡ Re.Ca).
Utilizing We led to a better characterization of the flow regimes presented in this system. The We
number for both phases is defined as:
We =ρU2
DH
σ
(4.1)
where ρ and σ are the fluid density and interfacial tension respectively. Air is used as the gaseous
phase whose density throughout the experiments were calculated considering its compressible
behavior inside the microchannels. Distilled water serves as the liquid phase. U is the average fluid
velocity inside the channel which is calculated by dividing the flow rate by the cross section area,
and DH represents the hydraulic diameter of the rectangular cross section of the microchannel. The
We for the gaseous phase (WeG) is defined based on the throat cross section at the flow-focusing
junction, and for the liquid phase WeL is defined based on the liquid channel dimensions. Using the
aforementioned definitions, WeL is in the range 0.0001-0.3 and WeG in the range 0.008-25. These
wide range of flow conditions result in various breakup modes which are explained in the following
section.
Experiments on each chip were performed by initiating a gaseous flow inside the device to prevent
the liquid from flooding the microchannels. Liquid water is subsequently injected into the flow-
23
focusing junction. Mapping experiments were conducted by setting a specified flow rate and
increasing the gas flow rate. As such, the maps consist of multiple horizontal lines of a fixed flow
rate (WeL), with transitions occurring as the flow rate (WeG) is changed (increased). We conducted
several experiments using multiple chips to account for wettability changes associated with the
chips being exposed to large amount of water for a long period of time. Over a wide range of flow
conditions we observed six distinct regimes which are defined as: Co-Flowing (■), Threading (♦),
Plugging (▲), Dripping (●), Jetting (►), and Multi-Satellite Formation (▼). A typical flow regime
map depicting these different breakup modes is shown in Figure 11 based on WeG and WeL. In this
map the region associated with each flow regime is distinguished with a different color. The gray
areas in the plot represent the transition regions across which the transition between two regimes
takes places. We used these finite transition regions rather than infinitesimally thin solid lines to
depict the change in the flow regime more realistically. This is more consistent with the
observations from experiments since flow regime transition could occur anywhere within the
shaded regions from a statistical point of view, which is determined by the microfluidic chips and
flow rate measurements in the experiments. However, dashed lines are also included in the flow
map as a reference and guideline to distinguish between the actual flow regimes and intermediate
transition regions. Furthermore, these lines can be considered as approximate boundaries between
each two flow regime regions.
At low values of WeG, the flow is characterized by the formation of liquid threads that extend along
the microchannel outlet. We observed that at low values of liquid flow (for WeL<10-2), the formed
thread does not remain stable along the outlet. In this region which is referred to as the Threading
regime, an unstable thread extends downstream of the junction where it breaks up intermittently at
different locations and results in multiple plugs and droplets. In the Co-flowing regime, which
occurs at higher values of WeL, the liquid thread is stable and extends throughout the microchannel
outlet. As the value of WeL increases, the thickness of this continuous liquid thread increases until
24
it obstructs the microchannel outlet. A periodic breakup takes place at higher WeG where each
formed thread is broken up into a single plug whose length is bigger than its width
(LDroplet>2WDroplet). This corresponds to the Plugging region, with threads that break up very close
to the flow-focusing junction at a distance Lbreakup<10Woutlet measured from the junction. The
Dripping regime is observed beyond the Plugging regime when the inertia forces start balancing
the surface tension forces at the junction (around WeG≈0.3). In this regime, circular drops are
formed at the junction which are highly-reproducible and uniform over a wide range of frequencies
and sizes. Increasing WeG, however, creates a chaotic condition that involves unsteady and random
droplet breakup dynamics. It can be seen from the flow map that a transition from the Dripping to
the Multi-Satellite Formation regime occurs around WeG≈6. Beyond this point a polydisperse spray
of the liquid is generated inside the microchannel. The Multi-Satellite Formation regime has not
been previously reported in liquid-liquid systems since its appearance is mostly due to the highly
inertial nature of the continuous phase which is usually negligible in oil-water systems. We
observed that at higher liquid flows a long liquid jet is formed after the junction which breaks into
individual droplets due to Rayleigh capillary instability [80]. In this Jetting regime, characterized
by droplet tip streaming of an extended liquid jet, droplet generation is less controlled. However,
generated droplets are in the order of the liquid jet diameter and much smaller than the droplets
from the Dripping regime.
25
Co-Flowing
Dripping
Threading Multi-Satellite
Formation
Plugging Jetting
Figure 11. Flow regime map of water-in-air droplet formation in a planar flow-focusing microfluidic device.
Different flow regions are distinguished with different colors. The transition between the flow regimes is also
represented as a finite shaded region rather than a solid line to account for the uncertainties associated with
the experiments and calculations of the We values.
In addition to the flow conditions, the geometry of the microchannel flow-focusing section plays
an important role in the flow map characteristics of this system. We were interested on the influence
that liquid microchannel size has on the Dripping regime in particular. Three different chips with
liquid channel widths of 25μm, 50μm, and 100μm were considered for this part. Flow mapping
experiments were conducted for the wider liquid channels (50μm and 100μm) and the results are
compared against the original flow map with the narrower liquid channel (25μm) as shown in
26
Figure 12. For these maps we have only included the approximate dashed lines to better compare
the flow regimes layout together. It can be seen that increasing the liquid channel size results in a
smaller Dripping region, which is highlighted with the green color in the flow map. Increasing the
size of the liquid channel results in larger surface tension forces that hold the droplet during
formation. Therefore, higher detaching forces are required to pinch-off the thread at the junction to
create discrete droplets. This will shift the transition from the Threading and Plugging regimes to
the Dripping regime to higher values of WeG. Transition to the Satellite Formation regime is less
dependent on the liquid channel size since it is mostly due to the inertial effects of the gaseous
phase. According to the flow maps, the Satellite Formation transition typically occurs around
WeG≈6-7 for all the chips. Consequently, increasing the size of the liquid channel would shrink the
Dripping region by expanding the Threading and Co-flow region.
Figure 12. Comparison of the flow map for chips with different liquid channel sizes. The area for the
Dripping region (the green region) is reduced as the liquid channel size increases. However, the Threading
and Co-flow regions have expanded as a result of increase in the surface tensions forces that holds the thread
and prevents the detachment at lower gas flow rates.
0.001
0.01
0.1
0.1 1 10
0.001
0.01
0.1
0.1 1 10
0.001
0.01
0.1
0.1 1 10
25μm 50μm
100μm
WeG
We L
27
Chapter 5. Dripping Regime Characterization
5.1 Generation Frequency
In this chapter, three important aspects of the Dripping regime are investigated by characterizing
the generation frequency, morphology of the droplets inside the microchannel, and monodispersity
of the generated droplets within this region. The frequency of droplet formation (ƒGeneration) is
obtained by manually counting the number of droplets generated over a given time interval from
the high-speed videos. To do this, the sequential images taken from the high speed camera were
analyzed in a lower speed to count the generated droplets with naked eye. Figure 13 and Figure 14
show the frequency data for the Dripping regime at different liquid and gas flow rates. It can be
seen that the frequency varies between less than 30 up to 1000 Hz (drops per second) over the entire
Dripping regime. Results show that frequency of the generation increases as either the liquid or gas
flows increase. Moreover, experimental plots show that frequency values are proportionate with
both flow rates. Therefore, we scaled the frequency in this system as the product of the two flow
rates (i.e. ƒ∝QGQL). In order to non-dimensionalize the frequency results, flow rates are non-
dimensionalized by characteristic inertial velocities (U*=(σ/ρDH)1/2) obtained from a Weber number
of unity for both phases, and a characteristic timescale (τ*=DH/U*). Therefore, generation frequency
can be non-dimensionalized as ƒ(τ*Gτ*
L)1/2. The results of this scaling are shown in Figure 15.
28
Figure 13. Experimental data of droplet generation frequency as a function of air flow rate for different liquid
flow rates. We can see that there is a relatively linear correlation between the frequency value and the air
flow rate.
Figure 14. Experimental data of droplet generation frequency at increasing liquid flow rate for two different
gas flow rates. We can see that there is a relatively linear correlation between the frequency value and the
liquid flow rate.
0
400
800
1200
0 10 20
ƒG
ener
ati
on
(Hz)
QG (mL/min)
A
0
400
800
1200
0 20 40
ƒG
ener
ati
on
(Hz)
QL (μL/min)
B
29
Figure 15. Experimental data of the non-dimensioalized droplet generation frequency. Generation frequency
in this system can be scaled as ƒ∝QGQL.
5.2 Droplet Morphology
Images of the droplets moving inside the microchannel are captured and analyzed to determine
their corresponding length and width. Experiments are conducted multiple times for each flow
condition to verify the reproducibility of the measured values. In the presence of a continuous gas
flow we always find a finite gap between the droplets and the walls within the Dripping regime.
Therefore, both the width and length of the droplets change as a function of flow rate. Figure 5
shows the droplet length and width as a function of gas flow rate (QG), for different liquid flow
rates (QL). Snapshots of the droplets inside the microchannel for different flow conditions shown
in Figure 18 clearly demonstrate these changes in length and width. It can be seen that a higher air
flow rate results in decreasing both length and width. However, liquid flow rate does not have an
apparent effect on the droplet width. Moreover, at high liquid flow rates, droplets exhibit a long tail
0.0001
0.001
0.01
0.01 0.1 1
ƒ(τ
* Gτ* L
)1/2
QG/(U*GD2
H,G) QL/(U*LD
2H,L)
30
which deviated from their common circular morphology. In order to take into account this length
for later scaling analysis, we used a correct length for this purpose as shown in Figure 19.
Figure 16. Experimental data of droplet width inside the microchannel for different liquid and gas flows.
Figure 17. Experimental data of droplet length inside the microchannel for different liquid and gas flows.
80
130
180
5 10 15 20
WD
rop
let(μ
m)
QG(mL/min)
80
100
120
140
160
180
200
5 10 15 20
LD
rop
let(μ
m)
QG (mL/min)
QL=1 μL/min
QL=5 μL/min
QL=10 μL/min
QL=20 μL/min
QL=30 μL/min
QL= 40 μL/min
31
QL=5μL/min
QG=8 mL/min
QG=10 mL/min
QG=12 mL/min
QG=14 mL/min
QG=16 mL/min
Figure 18. Snapshots of droplets moving inside microchannel under different gas flow rates.
Figure 19. Corrected length used in non-dimensionalizing droplet morphology data In high liquid flows,
droplet exhibit a tail-shape which deviate from their common circular morpholy.
100μm
1.
32
During the course of the generation process, a droplet starts forming from the liquid microchannel
in the flow-focusing junction. As the liquid is injected into the droplet volume, the droplet grows
in the plane at the junction. The droplet is attached to the liquid microchannel by a surface tension
force that can be scaled as,
Fσ ~ σDH,L (5.1)
where DH,L is the hydraulic diameter of the liquid microchannel. As the droplet protrudes into the
junction, it experiences inertial forces of the high speed gaseous medium that co-flows with the
liquid droplet. The inertial force of the gaseous phase can be scaled as;
FI ~ρQ
G2
DH,G4 h(W*-WL) cos θ (5.2)
where DH,G is the hydraulic diameter of the gas microchannel, h is the droplet height which is equal
to the channel height, θ is the angle at which the air enters the junction, WL is the liquid
microchannel width, and W* is the droplet width during the formation stage (see Figure 20). When
the aforementioned forces balance, the droplet width can be approximated as
W*~σDH,G4 /ρQ
G2 hcosθ+WL. However, the morphology of the droplets was assessed from the images
of the droplets moving inside the microchannel outlet after the droplets are detached. As such, a
similar scaling dependence can be used for the measured width with respect to the inertial force but
with a different proportionality factor and minimum width. To calculate this minimum width (W),
we consider the limiting case of W*≈WL which occurs experimentally at high gas flow rates and
also from the scaling equation. In this case, the prefilled droplet volume in the junction forms a
rectangular liquid ligament whose width is in the order of the liquid microchannel width and the
length is stretched to the throat of the junction (∆Z). This volume results in minimum droplet width
of 86μm after detachment, assuming a circular morphology for these droplets at high gas flow rates,
which is a reasonable assumption based on the experimental data. Since the complete generation
process always starts with the prefilling process, this obtained droplet size can be considered the
33
minimum width and length that can be achieved within the Dripping regime. Therefore, droplet
width can be calculated as W=C1 σ ρQG2⁄ +C2, where C1 takes into account the geometrical
parameters and a fitting factor, and C2 is the minimum droplet width that can be achieved in the
Dripping regime of this system.
The volume of the droplet in the dripping regime can be related to the generation frequency and
the liquid flow rate via conservation of mass of the liquid flow,
Vdroplet = Q
L
ƒ (5.3)
If the droplet is assumed to have a relatively uniform width over its entire length, the volume of the
droplet is proportional to the product of its dimensions. Therefore, the volume can be approximated
as Vdroplet ∝ LWh, where L, W, and h are the droplet length, width, and height, respectively. In order
to be consistent with the assumed (and mostly observed) droplet volume, experimental data for the
measured length were corrected in some cases where the droplet contained a morphology with non-
uniform width. By substituting the aforementioned scaling for the frequency (ƒ∝QGQL), droplet
length can be scaled as L~1/hWQG. Figure 24 shows the normalized droplet morphology using the
minimum values obtained in the Dripping regime as a function of the WeG. It can been seen that
both droplet length and width decrease with increasing gas flow. However, the effect of liquid flow
rate is less discernable in the morphology and more pronounced in the formation frequency.
34
Figure 20. Schematic of the Flow-Focusing geometry and the nomenclature used in scaling analysis for
droplet morphology.
Figure 21. Droplet length normalized with the minimum width of the droplets in the Dripping regime as a
function of gas Weber number (WeG). Minimum droplet width and length are assumed to be equal values that
are obtained based on the minimum volume of the prefilled droplet during the formation process at high gas
flow rates where droplet tends to have a more circular morphology. This minimum is dependent on the
geometry of the flow-focusing microchannel and is calculated to be about 86μm for the geometry used in out
experiments.
0
1
2
3
0 5 10
W/W
min
WeG
Scaling
QL=1
QL=5
QL=10
QL=20
QL=30
QL=40
35
Figure 22. Droplet length normalized with the minimum length of the droplets in the Dripping regime as a
function of gas Weber number (WeG). Minimum droplet width and length are assumed to be equal values that
are obtained based on the minimum volume of the prefilled droplet during the formation process at high gas
flow rates where droplet tends to have a more circular morphology. This minimum is dependent on the
geometry of the flow-focusing microchannel and is calculated to be about 86μm for the geometry used in out
experiments.
5.3 Droplet Monodispersity
Generated droplets are collected off the chip in a glass vial filled with a different liquid which is
immiscible with the water droplets. Captured images from the collected droplets were processed
using a customized MATLAB program to analyze the size of the drops and their distribution as
shown in Figure 23. Since the droplets are spherical after transitioning into the liquid medium, we
use the droplets’ diameter as the characteristic length scale in this part. Comparison between
droplets in the Dripping and Multi-Satellite Formation regions is shown in Figure 24. Size
uniformity of the collected droplets within the Dripping regime demonstrates the controlled
generation of the droplets within this region. The generated droplets have a polydispersity index
0
1
2
3
0 5 10
L/L
min
WeG
Scaling
QL=1
QL=5
QL=10
QL=20
QL=30
QL=40
36
(PDI) defined as σ/d (where σ represents the standard deviation of the droplet diameters and d is
the average diameter of the generated drops), of less than 5%. Droplets generated in the Dripping
regime can form a uniform closely-packed structure inside the liquid medium as shown in the
figure. Satellite Formation regime, however, results in polydisperse droplets which are distributed
over a wide range of diameters even for a fixed flow condition.
Figure 23. (Left) Experimental images of the collected droplets in the immiscible liquid solution. (Right)
detected droplets using image analysis with MATLAB.
37
Figure 24. Actual images and size distributions of samples of 500 collected droplets within the Dripping
and Multi-Satellite Formation regimes. Liquid flow rate for both cases is 5µL/min. Multi-Satellite Formation
is obtained at higher WeG according to the flow map. In the Dripping region by having a constant liquid and
gas flow rate the generated droplets are uniform in size. In Multi-Satellite Formation regime, however, small
daughter droplets are generated beside the main droplet which results in polydisperse distribution of the
droplets.
Dripping regime
Multi- Satellite Formation regime
38
Chapter 6. Conclusions
Combination of fluid behavior at microscale and advances in microengineering methods have
provided unique capabilities for controlling fluid-fluid interfaces. Controlled generation of liquid
droplets in microfluidic networks is an area of huge promise for a variety of applications in
biochemistry and material science. Droplet-based microfluidic devices typically employ
dispersions of one liquid within a continuous immiscible liquid phase, such as the creation of
uniform water-in-oil emulsions. Thanks to recent technological advancements in microfabrication
techniques, droplets can be produced in a planar chip format. In addition to the ease of fabrication,
planarization provides considerable benefits in terms of size and cost of the droplet system and,
more importantly, it enables the integration of the device in more complex LOC platforms.
Generation of aqueous liquid droplets using a continuous gaseous phase has been recently
demonstrated within microfluidic devices. This approach provides great potentials in modern Lab-
on-a-Chip (LOC) systems for high-throughput production of oil-free particles and airborne
microreactors. In this work we investigated different aspects of controlled formation of uniform
micrometer-sized water droplets in air using a microfluidic flow-focusing device (MFFD). The
droplet generation module is fabricated following standard soft lithography techniques with PDMS.
A standard microscope glass slide acts as the bottom surface on which each device is bonded using
a lab-scale plasma cleaner. Microfluidic devices are finally post baked at high temperatures for
several hours. The hydrophilic bottom wall (glass) creates a mixed surface condition inside the
microchannel since the post bake step renders the PDMS walls highly hydrophobic. By using a
PDMS-glass hybrid structure, we found six different flow regimes in this system: Co-flowing,
Threading, Plugging, Dripping, Jetting, and Multi-Satellite Formation. We further investigated the
Dripping regime which corresponds to controlled generation of monodisperse droplets at the flow-
focusing junction inside the microchannel. It was shown that increasing the liquid channel size
39
results in a smaller Dripping regime. Generation rate, droplet morphology, and monodispersity of
the droplets were investigated for different liquid and air flows within the Dripping regime. Results
of this work demonstrate smaller droplets (less than 100μm) at higher generation rates (up to 1
kHz) that are highly uniform in size (PDI<%5) which shows a significant improvement for liquid-
in-gas microfluidic systems. The contributions presented in this article provide practical guidelines
towards creating oil-free uniform droplets that can be used as templates for microparticle and
microfiber synthesis with control on the size and frequency. Moreover, many biochemically
relevant applications can also benefit from uniform digitization of sample within a gaseous
medium, namely for aerosol drug delivery and high-throughput screening of targets from gaseous
media. It is worth mentioning that in addition to the principal advantages of using droplet-based
systems for enhancing reaction rates and detection precision, the high-speed droplets generated in
this method, in contrast to conventional liquid-in-liquid generation, represent the basis for a new
approach based on droplet pair collisions for fast efficient micromixing [81], [82]. This
incomparable feature would indeed provide a significant development towards manufacturing fast
and efficient LOC platforms in which airborne microreactors can be constructed and different
reagents can be rapidly mixed afterwards for detection purposes.
40
REFERENCES
[1] E. K. Sackmann, A. L. Fulton, and D. J. Beebe, “The present and future role of
microfluidics in biomedical research,” Nature, vol. 507, no. 7491, pp. 181–189, 2014.
[2] J. Atencia and D. J. Beebe, “Controlled microfluidic interfaces,” Nature, vol. 437, no.
September, pp. 648–655, 2005.
[3] A. J. DeMello, “Control and detection of chemical reactions in microfluidic systems.,”
Nature, vol. 442, no. July, pp. 394–402, 2006.
[4] G. M. Whitesides, “The origins and the future of microfluidics.,” Nature, vol. 442, no.
7101, pp. 368–373, 2006.
[5] S.-Y. Teh, R. Lin, L.-H. Hung, and A. P. Lee, “Droplet microfluidics.,” Lab Chip, vol. 8,
pp. 198–220, 2008.
[6] A. Huebner, S. Sharma, M. Srisa-Art, F. Hollfelder, J. B. Edel, and A. J. deMello,
“Microdroplets: A sea of applications?,” Lab Chip, vol. 8, no. 8, p. 1244, 2008.
[7] R. Seemann, M. Brinkmann, T. Pfohl, and S. Herminghaus, “Droplet based
microfluidics,” Reports Prog. Phys., vol. 75, no. 1, p. 16601, 2012.
[8] X. Casadevall i Solvas and A. deMello, “Droplet microfluidics: recent developments and
future applications.,” Chem. Commun. (Camb)., vol. 47, no. 7, pp. 1936–1942, 2011.
[9] O. J. Dressler, R. M. Maceiczyk, S.-I. Chang, and A. J. DeMello, “Droplet-Based
Microfluidics: Enabling Impact on Drug Discovery,” J. Biomol. Screen., vol. 19, no. 4, pp.
483–496, 2014.
[10] T. S. Kaminski, O. Scheler, and P. Garstecki, “Droplet microfluidics for microbiology:
techniques, applications and challenges.,” Lab Chip, 2016.
41
[11] T. P. Lagus and J. F. Edd, “A review of the theory, methods and recent applications of
high-throughput single-cell droplet microfluidics,” J. Phys. D. Appl. Phys., vol. 46, no. 11,
p. 114005, 2013.
[12] B. Kintses, L. D. van Vliet, S. R. a Devenish, and F. Hollfelder, “Microfluidic droplets:
New integrated workflows for biological experiments,” Curr. Opin. Chem. Biol., vol. 14,
no. 5, pp. 548–555, 2010.
[13] H. Song, J. D. Tice, and R. F. Ismagilov, “A microfluidic system for controlling reaction
networks in time,” Angew. Chemie - Int. Ed., vol. 42, pp. 768–772, 2003.
[14] T. M. Squires and S. R. Quake, “Microfluidics: Fluid physics at the nanoliter scale,” Rev.
Mod. Phys., vol. 77, no. 3, pp. 977–1026, 2005.
[15] C. N. Baroud, F. Gallaire, and R. Dangla, “Dynamics of microfluidic droplets.,” Lab Chip,
vol. 10, pp. 2032–2045, 2010.
[16] P. Zhu and L. Wang, “Passive and active droplet generation with microfluidics: a review,”
Lab Chip, vol. 17, no. 1, pp. 34–75, 2017.
[17] Z. Z. Chong, S. H. Tan, A. M. Ganan-Calvo, S. B. Tor, N. H. Loh, and N.-T. Nguyen,
“Active droplet generation in microfluidics,” Lab Chip, vol. 16, no. 1, pp. 35–58, 2016.
[18] G. F. Christopher and S. L. Anna, “Microfluidic methods for generating continuous
droplet streams,” J. Phys. D. Appl. Phys., vol. 40, no. 19, pp. R319–R336, 2007.
[19] T. Thorsen, R. W. Roberts, F. H. Arnold, and S. R. Quake, “Dynamic pattern formation in
a vesicle-generating microfluidic device,” Phys. Rev. Lett., vol. 86, pp. 4163–4166, 2001.
[20] P. B. Umbanhowar, V. Prasad, and D. A. Weitz, “Monodisperse emulsuion generation via
drop break-off in a co-flowing stream,” Langmuir, no. 4, pp. 1–22, 1999.
42
[21] S. L. Anna, N. Bontoux, and H. a. Stone, “Formation of dispersions using ‘flow focusing’
in microchannels,” Appl. Phys. Lett., vol. 82, no. 3, pp. 364–366, 2003.
[22] J. K. Nunes, S. S. H. Tsai, J. Wan, and H. a Stone, “Dripping and jetting in microfluidic
multiphase flows applied to particle and fiber synthesis.,” J. Phys. D. Appl. Phys., vol. 46,
no. 11, p. 114002, 2013.
[23] A. S. Utada, A. Fernandez-Nieves, H. A. Stone, and D. A. Weitz, “Dripping to jetting
transitions in coflowing liquid streams,” Phys. Rev. Lett., vol. 99, no. 9, pp. 1–4, 2007.
[24] T. Cubaud and T. G. Mason, “Capillary threads and viscous droplets in square
microchannels,” Phys. Fluids, vol. 20, no. 5, p. 53302, 2008.
[25] L. M. Fidalgo, C. Abell, and W. T. S. Huck, “Surface-induced droplet fusion in
microfluidic devices.,” Lab Chip, vol. 7, no. c, pp. 984–986, 2007.
[26] L.-H. Hung, K. M. Choi, W.-Y. Tseng, Y.-C. Tan, K. J. Shea, and A. P. Lee, “Alternating
droplet generation and controlled dynamic droplet fusion in microfluidic device for CdS
nanoparticle synthesis,” Lab Chip, vol. 6, no. 2, p. 174, 2006.
[27] Y.-C. Tan, Y. L. Ho, and A. P. Lee, “Droplet coalescence by geometrically mediated flow
in microfluidic channels,” Microfluid. Nanofluidics, vol. 3, no. 4, pp. 495–499, 2007.
[28] D. R. Link, S. L. Anna, D. A. Weitz, and H. A. Stone, “Geometrically Mediated Breakup
of Drops in Microfluidic Devices,” Phys. Rev. Lett., vol. 92, no. 5, p. 54503, 2004.
[29] Y.-C. Tan, J. S. Fisher, A. I. Lee, V. Cristini, and A. P. Lee, “Design of microfluidic
channel geometries for the control of droplet volume, chemical concentration, and
sorting.,” Lab Chip, vol. 4, pp. 292–298, 2004.
[30] Y.-C. Tan, Y. L. Ho, and A. P. Lee, “Microfluidic sorting of droplets by size,” Microfluid.
Nanofluidics, vol. 4, no. 4, pp. 343–348, 2008.
43
[31] F. Sarrazin, L. Prat, N. Di Miceli, G. Cristobal, D. R. Link, and D. a. Weitz, “Mixing
characterization inside microdroplets engineered on a microcoalescer,” Chem. Eng. Sci.,
vol. 62, pp. 1042–1048, 2007.
[32] M. R. Bringer, C. J. Gerdts, H. Song, J. D. Tice, and R. F. Ismagilov, “Microfluidic
systems for chemical kinetics that rely on chaotic mixing in droplets,” Phil. Trans. R. Soc.
Lond. A, no. i, pp. 1087–1104, 2004.
[33] M. G. Simon and A. P. Lee, Microdroplet Technology. New York, NY: Springer New
York, 2012.
[34] N. Shembekar, C. Chaipan, R. Utharala, and C. A. Merten, “Droplet-based microfluidics
in drug discovery, transcriptomics and high-throughput molecular genetics,” Lab Chip,
vol. 16, pp. 1314–1331, 2016.
[35] L. Martín-Banderas, M. Flores-Masquera, P. Riesco-Chueca, A. Rodríguez-Gil, Á.
Cebolla, S. Chávez, and A. M. Gañán-Calvo, “Flow focusing: A versatile technology to
produce size-controlled and specific-morphology microparticles,” Small, vol. 1, no. 7, pp.
688–692, 2005.
[36] J. Wang, Y. Li, X. Wang, J. Wang, H. Tian, P. Zhao, Y. Tian, Y. Gu, L. Wang, and C.
Wang, “Droplet microfluidics for the production of microparticles and nanoparticles,”
Micromachines, vol. 8, no. 1, pp. 1–23, 2017.
[37] R. Zilionis, J. Nainys, A. Veres, V. Savova, D. Zemmour, A. M. Klein, and L. Mazutis,
“Single-cell barcoding and sequencing using droplet microfluidics.,” Nat. Protoc., vol. 12,
no. 1, pp. 44–73, 2017.
44
[38] S. S. Terekhov, I. V. Smirnov, A. V. Stepanova, T. V. Bobik, Y. A. Mokrushina, N. A.
Ponomarenko, A. A. Belogurov, M. P. Rubtsova, O. V. Kartseva, M. O. Gomzikova, A.
A. Moskovtsev, A. S. Bukatin, M. V. Dubina, E. S. Kostryukova, V. V. Babenko, M. T.
Vakhitova, A. I. Manolov, M. V. Malakhova, M. A. Kornienko, A. V. Tyakht, A. A.
Vanyushkina, E. N. Ilina, P. Masson, A. G. Gabibov, and S. Altman, “Microfluidic droplet
platform for ultrahigh-throughput single-cell screening of biodiversity,” Proc. Natl. Acad.
Sci., p. 201621226, 2017.
[39] M. Baker, “Digital PCR hits its stride,” Nat. Methods, vol. 9, no. 6, pp. 541–544, 2012.
[40] H. Tanaka, S. Yamamoto, A. Nakamura, Y. Nakashoji, N. Okura, N. Nakamoto, K.
Tsukagoshi, and M. Hashimoto, “Hands-off preparation of monodisperse emulsion
droplets using a poly(dimethylsiloxane) microfluidic chip for droplet digital PCR,” Anal.
Chem., vol. 87, no. 8, pp. 4134–4143, 2015.
[41] M. M. Kiss, L. Ortoleva-donnelly, N. R. Beer, J. Warner, C. G. Bailey, B. W. Colston, J.
M. Rothberg, D. R. Link, H. Leamon, R. Technologies, and H. Ave, “High-Throughput
Quantitative PCR in Picoliter Droplets,” Anal. Chem., vol. 80, no. 23, pp. 8975–8981,
2008.
[42] P. Kumaresan, C. J. Yang, S. A. Cronier, R. G. Blazej, and R. A. Mathies, “High-
throughput single copy DNA amplification and cell analysis in engineered nanoliter
droplets,” Anal. Chem., vol. 80, no. 10, pp. 3522–3529, 2008.
[43] K. Hettiarachchi and A. P. Lee, “Polymer-lipid microbubbles for biosensing and the
formation of porous structures,” J. Colloid Interface Sci., vol. 344, no. 2, pp. 521–527,
2010.
45
[44] K. Hettiarachchi, E. Talu, M. L. Longo, P. a Dayton, and A. P. Lee, “On-chip generation
of microbubbles as a practical technology for manufacturing contrast agents for ultrasonic
imaging.,” Lab Chip, vol. 7, no. 4, pp. 463–468, 2007.
[45] P. Garstecki, I. Gitlin, W. Diluzio, G. M. Whitesides, E. Kumacheva, and H. A. Stone,
“Formation of monodisperse bubbles in a microfluidic flow-focusing device,” Appl. Phys.
Lett., vol. 85, no. 13, pp. 2649–2651, 2004.
[46] P. Garstecki, a M. Gañán-Calvo, and G. M. Whitesides, “Formation of bubbles and
droplets in microfluidic systems,” Bull. Polish Acad. Sci., vol. 53, no. 4, pp. 361–372,
2005.
[47] P. Tirandazi and C. H. Hidrovo, “Liquid-in-gas droplet microfluidics ; experimental
characterization of droplet morphology , generation frequency , and monodispersity in a
flow-focusing microfluidic device,” J. Microelectromechanical Syst., 2017.
[48] L. S. Roach, H. Song, and R. F. Ismagilov, “Controlling nonspecific protein adsorption in
a plug-based microfluidic system by controlling interfacial chemistry using fluorous-phase
surfactants,” Anal. Chem., vol. 77, no. 3, pp. 785–796, 2005.
[49] Y. Skhiri, P. Gruner, B. Semin, Q. Brosseau, D. Pekin, L. Mazutis, V. Goust, F.
Kleinschmidt, A. El Harrak, J. B. Hutchison, E. Mayot, J.-F. Bartolo, A. D. Griffiths, V.
Taly, and J.-C. Baret, “Dynamics of molecular transport by surfactants in emulsions,” Soft
Matter, vol. 8, no. 41, p. 10618, 2012.
[50] a J. Acero, N. Rebollo-Muñoz, J. M. Montanero, a M. Gañán-Calvo, and E. J. Vega, “A
new flow focusing technique to produce very thin jets,” J. Micromechanics
Microengineering, vol. 23, no. JUNE, p. 65009, 2013.
46
[51] a J. Acero, J. M. Montanero, C. Ferrera, M. a Herrada, and a M. Gañán-Calvo,
“Enhancement of the stability of the flow focusing technique for low-viscosity liquids,” J.
Micromechanics Microengineering, vol. 22, no. 11, p. 115039, 2012.
[52] A. Luque, F. Perdigones, J. Estevé, J. Montserrat, A. Gañán-Calvo, and J. M. Quero,
“Reduction of droplet-size dispersion in parallel flow-focusing microdevices using a
passive method,” J. Micromechanics Microengineering, vol. 19, no. 4, p. 45029, 2009.
[53] A. Gañán-Calvo, “Generation of Steady Liquid Microthreads and Micron-Sized
Monodisperse Sprays in Gas Streams,” Phys. Rev. Lett., vol. 80, no. 2, pp. 285–288, 1998.
[54] J. M. GORDILLO, M. PÉREZ-SABORID, and A. M. GAÑÁN-CALVO, “Linear stability
of co-flowing liquid–gas jets,” J. Fluid Mech., vol. 448, pp. 23–51, 2001.
[55] M. A. Herrada, A. M. Gañán-Calvo, A. Ojeda-Monge, B. Bluth, and P. Riesco-Chueca,
“Liquid flow focused by a gas: Jetting, dripping, and recirculation,” Phys. Rev. E - Stat.
Nonlinear, Soft Matter Phys., vol. 78, no. 3, pp. 1–16, 2008.
[56] E. J. Vega, J. M. Montanero, M. A. Herrada, and A. M. Gañán-Calvo, “Global and local
instability of flow focusing: The influence of the geometry,” Phys. Fluids, vol. 22, no. 6,
pp. 1–10, 2010.
[57] S. L. Anna, “Droplets and Bubbles in Microfluidic Devices,” Annu. Rev. Fluid Mech., vol.
48, no. 1, pp. 285–309, 2016.
[58] C. Fang, C. Hidrovo, F. Wang, J. Eaton, and K. Goodson, “3-D numerical simulation of
contact angle hysteresis for microscale two phase flow,” Int. J. Multiph. Flow, vol. 34, no.
7, pp. 690–705, 2008.
[59] M. J. Cheah, I. G. Kevrekidis, and J. B. Benziger, “Water slug to drop and film transitions
in gas-flow channels,” Langmuir, vol. 29, no. 48, pp. 15122–15136, 2013.
47
[60] M. J. Cheah, I. G. Kevrekidis, and J. B. Benziger, “Water slug formation and motion in
gas flow channels: The effects of geometry, surface wettability, and gravity,” Langmuir,
vol. 29, no. 31, pp. 9918–9934, 2013.
[61] D. R. Kim, J. M. Koo, C. Fang, J. E. Steinbrenner, E. S. Lee, F. M. Wang, C. H. Hidrovo,
J. K. Eaton, K. E. Goodson, and Asme, “Compact model of slug flow in microchannels,”
in Proceedings of the Fifth International Conference on Nanochannels, Microchannels,
and Minichannels ICNMM2007, 2007, pp. 321–329.
[62] C. H. Schillberg and S. G. Kandlikar, “A Review of Models For Droplet Detachment
From The Gas Diffusion Layer-Gas Flow Channel Interface In PEMFCs,” in Proceedings
of the Fifth International Conference on Nanochannels, Microchannels and Minichannels
ICNMM2007, 2007, pp. 1–12.
[63] A. Mukherjee and S. G. Kandlikar, “A numerical analysis of growing water droplet inside
an air supply channel of a PEM fuel cell,” in Proceedings of IMECE2006 2006 ASME
International Mechanical Engineering Congress and Exposition, 2006, pp. 1–6.
[64] E. S. Lee, C. H. Hidrovo, J. E. Steinbrenner, F.-M. Wang, S. Vigneron, K. E. Goodson,
and J. K. Eaton, “Flow Structures and Frictional Characteristics on Two-Phase Flow in
Microchannels in PEM Fuel Cells,” in Proceedings of FEDSM2005 2005 ASME Fluids
Engineering Summer Conference, 2005, pp. 1–8.
[65] M. Koz and S. G. Kandlikar, “Numerical investigation of interfacial transport resistance
due to water droplets in proton exchange membrane fuel cell air channels,” J. Power
Sources, vol. 243, pp. 946–957, 2013.
[66] J. E. Steinbrenner, E. S. Lee, C. H. Hidrovo, J. K. Eaton, and K. E. Goodson, “Impact of
channel geometry on two-phase flow in fuel cell microchannels,” J. Power Sources, vol.
196, no. 11, pp. 5012–5020, 2011.
48
[67] X. Zhu, P. C. Sui, and N. Djilali, “Numerical simulation of emergence of a water droplet
from a pore into a microchannel gas stream,” Microfluid. Nanofluidics, vol. 4, no. 6, pp.
543–555, 2008.
[68] B. Carroll and C. Hidrovo, “Droplet Detachment Mechanism in a High-Speed Gaseous
Microflow,” J. Fluids Eng., vol. 135, no. 7, p. 71206, May 2013.
[69] C. H. Hidrovo, F.-M. Wang, J. E. Steinbrenner, E. S. Lee, S. Vigneron, C.-H. Cheng, J. K.
Eaton, and K. E. Goodson, “Water slug detachment in two-phase hydrophobic
microchannel flows,” in Proceedings of ICMM2005 3rd International Conference on
Microchannels and Minichannels, 2005, pp. 1–7.
[70] K. Jiang, A. X. Lu, P. Dimitrakopoulos, D. L. DeVoe, and S. R. Raghavan, “Microfluidic
generation of uniform water droplets using gas as the continuous phase,” J. Colloid
Interface Sci., vol. 448, pp. 275–279, 2015.
[71] A. Shahriari, M. M. Kim, S. Zamani, N. Phillip, B. Nasouri, and C. H. Hidrovo, “Flow
regime mapping of high inertial gas–liquid droplet microflows in flow-focusing
geometries,” Microfluid. Nanofluidics, vol. 20, no. 1, p. 20, 2016.
[72] Y. Xia and G. M. Whitesides, “Soft lithography,” Annu. Rev. Mater. Sci., vol. 28, no. 12,
pp. 153–184, 1998.
[73] J. N. Lee, C. Park, and G. M. Whitesides, “Solvent Compatibility of
Poly(dimethylsiloxane)-Based Microfluidic Devices,” Anal. Chem., vol. 75, no. 23, pp.
6544–6554, 2003.
[74] I. Wong and C.-M. Ho, “Surface molecular property modifications for
poly(dimethylsiloxane) (PDMS) based microfluidic devices.,” Microfluid. Nanofluidics,
vol. 7, no. 3, pp. 291–306, Sep. 2009.
49
[75] N. R. Glass, R. Tjeung, P. Chan, L. Y. Yeo, and J. R. Friend, “Organosilane deposition for
microfluidic applications,” Biomicrofluidics, vol. 5, no. 3, pp. 1–7, 2011.
[76] Z. Pan, H. Shahsavan, W. Zhang, F. K. Yang, and B. Zhao, “Superhydro-oleophobic bio-
inspired polydimethylsiloxane micropillared surface via FDTS coating/blending
approaches,” Appl. Surf. Sci., vol. 324, pp. 612–620, Jan. 2015.
[77] P. Tirandazi and C. H. Hidrovo, “Experimental Investigation of Geometrical Parameters
for Gas-Liquid Droplet Generation In Flow-Focusing Configurations,” in ASME 2015
13th International Conference on Nanochannels, Microchannels, and Minichannels
ICNMM15, 2015, pp. 1–6.
[78] P. Tirandazi and C. H. Hidrovo, “Generation of Uniform Liquid Droplets in A
Microfluidic Chip Using A High-speed Gaseous Microflow,” in Proceedings of the ASME
2016 14th Conference on Nanochannels, Microchannels, and Minichannels ICNMM2016,
2016, pp. 1–6.
[79] J. Carneiro, E. Doutel, J. B. L. M. Campos, and J. M. Miranda, “PDMS droplet formation
and characterization by hydrodynamic flow focusing technique in a PDMS square
microchannel,” J. Micromechanics Microengineering, vol. 26, no. 10, pp. 1–8, 2016.
[80] S. L. Anna and H. C. Mayer, “Microscale tipstreaming in a microfluidic flow focusing
device,” Phys. Fluids, vol. 18, no. 12, 2006.
[81] B. Carroll and C. Hidrovo, “Experimental Investigation of Inertial Mixing in Colliding
Droplets,” Heat Transf. Eng., vol. 34, no. 2–3, pp. 120–130, Jan. 2013.
[82] B. Carroll and C. Hidrovo, “Droplet collision mixing diagnostics using single fluorophore
LIF,” Exp. Fluids, vol. 53, no. 5, pp. 1301–1316, Aug. 2012.