DSD-110-HU0165 HOOKUP CONNECTION GUIDE Using MEDIVATORS ® DSD and SSD Series Automated Endoscope Reprocessors READ ALL INSTRUCTIONS PRIOR TO CONNECTION Locate the endoscope port and follow the instructions for attaching the corresponding Endoscope Connector. All connectors must be attached. DESIGNATED HOOKUP FOR THE FOLLOWING FUJIFILM ENDOSCOPES*: EC-530HL2 EC-530LS2 EC-580RD/L EC-600HL EC-600WL EG-600WR-V2 EC-600ZW/M EG-600WR EC-530HLV2 EC-530LSV2 EC-580RD/M EC-600LS EC-600WL-V2 EC-600ZW/L EG-580RD EG-600ZW IF THE ENDOSCOPE IS NOT LISTED HERE PLEASE SEE THE OTHER GUIDES. Basin Connector (see Basin Connection Instructions on pg 2) Air Probe Connector Channel Connection Tube Biopsy Inlet Connector Suction Connector Forward Water Jet Connector ADDITIONAL REQUIRED ENDOSCOPE CONNECTIONS - VERSION 1 Basin Connector (see Basin Connection Instructions on pg 2) Air Probe Connector Channel Connection Tube Forward Water Jet Connector Suction Connector Biopsy Inlet Connector VERSION 1 DSD HOOKUP ENDOSCOPE CONNECTORS (Figure 1) Ensure proper alignment of all connectors with the endoscope mating parts prior to attachment. Suction Connector Align suction connector with suction barb and push onto barb until fully engaged. Biopsy Inlet Connector Align stainless steel fitting with biopsy port, push and turn clockwise to connect. FUJIFILM™ CA-503S/A Cleaning Adapter Align the red and blue rings with the corresponding valve ports. Squeeze the black tab and frame while inserting the connector. Release the tab to secure the frame. Ensure that the frame is properly secured. Leak Test Connector Align the pins on the connector with the slots on the scope. Push on and turn clockwise to engage. A2 FUJIFILM CA-500J/A Cleaning Adapter Align the slots on the cap with the pins on the Air/ Water bottle connector, push on fully and twist clockwise to engage. Air Probe Connector Align with air probe, push and slide until fully engaged. FUJIFILM CA-503B/C Cleaning Adapter Place the biopsy slide over the biopsy inlet, depress the frame and slide until fully engaged. Channel Connection Tube Align each stainless steel fitting with corresponding female luer, push and turn clockwise to connect. Forward Water Jet Connector Align stainless steel fitting with female luer, push and turn clockwise to connect. ADDITIONAL REQUIRED ENDOSCOPE CONNECTIONS - VERSION 2 FUJIFILM™ CA-503S/A Cleaning Adapter Align the red and blue rings with the corresponding valve ports. Squeeze the black tab and frame while inserting the connector. Release the tab to secure the frame. Ensure that the frame is properly secured. Leak Test Connector Align the pins on the connector with the slots on the scope. Push on and turn clockwise to engage. A2 FUJIFILM CA-503T/R/C Cleaning Adapter Align the slots on the cap with the pins on the Air/Water bottle connector, push on fully twist clockwise to engage. FUJIFILM CA-503B/C Cleaning Adapter Place the biopsy slide over the biopsy inlet, depress the frame and slide until fully engaged. Suction Connector Align suction connector with suction barb and push onto barb until fully engaged. Channel Connection Tube Align each stainless steel fitting with corresponding female luer, push and turn clockwise to connect. Biopsy Inlet Connector Align stainless steel fitting with biopsy port, push and turn clockwise to connect. Forward Water Jet Connector Align stainless steel fitting with female luer, push and turn clockwise to connect. Air Probe Connector Align with air probe, push and slide until fully engaged. Leak Test Hookup PN DSD-110-HU0103 For MEDIVATORS AERs with leak test capability. (see Basin Connection Instructions on page 2) A1 A2 Leak Test Hookup PN DSD-110-HU0103 For MEDIVATORS AERs with leak test capability. (see Basin Connection Instructions on page 2) A1 A2 DESIGNATED HOOKUP FOR THE FOLLOWING FUJIFILM ENDOSCOPES*: EC-530HL EC-530LS IF THE ENDOSCOPE IS NOT LISTED HERE PLEASE SEE THE OTHER GUIDES. FUJIFILM™ CA-503 KIT which includes: FUJIFILM CA-503 S/A FUJIFILM CA-500J/A FUJIFILM CA-503 B/C FUJIFILM™ CA-503 KIT which includes: FUJIFILM CA-503 S/A FUJIFILM CA-503 T/R/C FUJIFILM CA-503 B/C DESIGNATED HOOKUP FOR THE FOLLOWING FUJIFILM ENDOSCOPES*: EC-250HL5 EC-450WI5 EC-530LT EC-590WL EC-590ZW3M EG-530CT EC-250LS5 EC-450WL EC-530MT EC-590WI4 EC-590ZWL EG-530UR EC-250WE5 EC-450WL5 EC-530WI EC-590WL4 EC-590ZWM ES-250WE5 EC-250WI5 EC-450WM5 EC-530WL EC-590WM4 EC-L590ZW ES-450WE5 EC-250WL5 EC-530FI EC-530WL3 EC-590WL-H EC-L590ZW/L ES-530WE EC-250WM5 EC-530FL EC-530WM EC-590WM EG-250CT5 EC-450HL5 EC-530FM EC-530WM EC-590WM2 EG-410CT EC-450LS5 EC-530IT EC-590WI EC-590ZW3L EG-450CT5 IF THE ENDOSCOPE IS NOT LISTED HERE PLEASE SEE THE OTHER GUIDES. ADDITIONAL REQUIRED ENDOSCOPE CONNECTIONS - VERSION 3 Basin Connector (see Basin Connection Instructions on pg 2) Air Probe Connector Channel Connection Tube Biopsy Inlet Connector Suction Connector Forward Water Jet Connector Leak Test Hookup PN DSD-110-HU0103 For MEDIVATORS AERs with leak test capability. (see Basin Connection Instructions on page 2) A1 A2 FUJIFILM™ CA-503 KIT which includes: FUJIFILM CA-503 S/A FUJIFILM CA-503 T/R/C FUJIFILM CA-503 B/C FUJIFILM CJ-500 FUJIFILM™ CA-503S/A Cleaning Adapter Align the red and blue rings with the corresponding valve ports. Squeeze the black tab and frame while inserting the connector. Release the tab to secure the frame. Ensure that the frame is properly secured. Leak Test Connector Align the pins on the connector with the slots on the scope. Push on and turn clockwise to engage. A2 FUJIFILM CA-503T/R/C Cleaning Adapter Align the slots on the cap with the pins on the Air/ Water bottle connector, push on fully and twist clockwise to engage. FUJIFILM CA-503B/C Cleaning Adapter Place the biopsy slide over the biopsy inlet, depress the frame and slide until fully engaged. Suction Connector Align suction connector with suction barb and push onto barb until fully engaged. Channel Connection Tube Align each stainless steel fitting with corresponding female luer, push and turn clockwise to connect. Biopsy Inlet Connector Align stainless steel fitting with biopsy port, push and turn clockwise to connect. Forward Water Jet Connector Align stainless steel fitting with female luer, push and turn clockwise to connect. CJ-500 Cleaning Adapter Align the CJ-500 adapter with the forward water jet channel inlet and screw clockwise to engage. Air Probe Connector Align with air probe, push and slide until fully engaged. © 2016 Medivators Inc. All Rights Reserved. 50098-349/C * This list of corresponding endoscope models is subject to change without notice. For a full list, please go to: www.medivators.com/hookuplookup VERSION 2 DSD HOOKUP ENDOSCOPE CONNECTORS (Figure 1) VERSION 3 DSD HOOKUP ENDOSCOPE CONNECTORS (Figure 1) * Representative endoscope for illustrative purposes only. For exact port locations, refer to endoscope manufacturer’s instruction guide. Ensure that the endoscope model you are connecting is listed on this IFU. NOT PICTURED ABOVE Luer Fitting behind Biopsy Port ES-200ER Luer Balloon Connection EC-450BI5, EN-450P5/20, EN-450T5, EN-450WM23 Luer Biopsy Connection EG-530UT * This list of corresponding endoscope models is subject to change without notice. For a full list, please go to: www.medivators.com/hookuplookup * This list of corresponding endoscope models is subject to change without notice. For a full list, please go to: www.medivators.com/hookuplookup

DSD-110-HU0165 HOOKUP CONNECTION GUIDE READ ALL ... REV C.pdfDSD-110-HU0165 HOOKUP CONNECTION GUIDE Using MEDIVATORS® DSD and SSD Series Automated Endoscope Reprocessors READ ALL

  • Upload
    others

  • View
    55

  • Download
    1

Embed Size (px)

Citation preview

DSD-110-HU0165 HOOKUP CONNECTION GUIDE

Using MEDIVATORS® DSD and SSD Series Automated Endoscope Reprocessors

READ ALL INSTRUCTIONS PRIOR TO CONNECTION

Locate the endoscope port and follow the instructions for attaching the corresponding Endoscope Connector. All connectors must be attached.

DESIGNATED HOOKUP FOR THE FOLLOWING FUJIFILM ENDOSCOPES*:

EC-530HL2 EC-530LS2 EC-580RD/L EC-600HL EC-600WL EG-600WR-V2 EC-600ZW/M EG-600WREC-530HLV2 EC-530LSV2 EC-580RD/M EC-600LS EC-600WL-V2 EC-600ZW/L EG-580RD EG-600ZW

IF THE ENDOSCOPE IS NOT LISTED HERE PLEASE SEE THE OTHER GUIDES.

Basin Connector (see Basin Connection Instructions on pg 2) Air Probe Connector

Channel Connection Tube Biopsy Inlet Connector

Suction Connector Forward Water Jet Connector

ADDITIONAL REQUIRED ENDOSCOPE CONNECTIONS - VERSION 1

Basin Connector (see Basin Connection Instructions on pg 2) Air Probe Connector

Channel Connection Tube Forward Water Jet Connector

Suction Connector Biopsy Inlet Connector

VERSION 1DSD HOOKUP ENDOSCOPE CONNECTORS (Figure 1)

Ensure proper alignment of all connectors with the endoscope mating parts prior to attachment.

Suction ConnectorAlign suction connector with suction barb and push onto barb until fully engaged.

Biopsy Inlet ConnectorAlign stainless steel fitting with biopsy port, push and turn clockwise to connect.

FUJIFILM™ CA-503S/A Cleaning AdapterAlign the red and blue rings with the corresponding valve ports. Squeeze the black tab and frame while inserting the connector. Release the tab to secure the frame. Ensure that the frame is properly secured.

Leak Test ConnectorAlign the pins on the connector with the slots on the scope. Push on and turn clockwise to engage.

A2

FUJIFILM CA-500J/A Cleaning AdapterAlign the slots on the cap with the pins on the Air/Water bottle connector, push on fully and twist clockwise to engage.

Air Probe ConnectorAlign with air probe, push and slide until fully engaged.

FUJIFILM CA-503B/C Cleaning AdapterPlace the biopsy slide over the biopsy inlet, depress the frame and slide until fully engaged.

Channel Connection TubeAlign each stainless steel fitting with corresponding female luer, push and turn clockwise to connect.

Forward Water Jet ConnectorAlign stainless steel fitting with female luer, push and turn clockwise to connect.

ADDITIONAL REQUIRED ENDOSCOPE CONNECTIONS - VERSION 2

FUJIFILM™ CA-503S/A Cleaning AdapterAlign the red and blue rings with the corresponding valve ports. Squeeze the black tab and frame while inserting the connector. Release the tab to secure the frame. Ensure that the frame is properly secured.

Leak Test ConnectorAlign the pins on the connector with the slots on the scope. Push on and turn clockwise to engage.

A2

FUJIFILM CA-503T/R/C Cleaning AdapterAlign the slots on the cap with the pins on the Air/Water bottle connector, push on fully twist clockwise to engage.

FUJIFILM CA-503B/C Cleaning AdapterPlace the biopsy slide over the biopsy inlet, depress the frame and slide until fully engaged.

Suction ConnectorAlign suction connector with suction barb and push onto barb until fully engaged.

Channel Connection TubeAlign each stainless steel fitting with corresponding female luer, push and turn clockwise to connect.

Biopsy Inlet ConnectorAlign stainless steel fitting with biopsy port, push and turn clockwise to connect.

Forward Water Jet ConnectorAlign stainless steel fitting with female luer, push and turn clockwise to connect.

Air Probe ConnectorAlign with air probe, push and slide until fully engaged.

Leak Test Hookup PN DSD-110-HU0103

For MEDIVATORS AERs with leak test capability. (see Basin Connection Instructions on page 2)

A1

A2

Leak Test Hookup PN DSD-110-HU0103

For MEDIVATORS AERs with leak test capability. (see Basin Connection Instructions on page 2)

A1

A2

DESIGNATED HOOKUP FOR THE FOLLOWING FUJIFILM ENDOSCOPES*:

EC-530HL

EC-530LS

IF THE ENDOSCOPE IS NOT LISTED HERE PLEASE SEE THE OTHER GUIDES.

FUJIFILM™ CA-503 KIT which includes:

FUJIFILM CA-503 S/A FUJIFILM CA-500J/AFUJIFILM CA-503 B/C

FUJIFILM™ CA-503 KIT which includes:

FUJIFILM CA-503 S/A FUJIFILM CA-503 T/R/CFUJIFILM CA-503 B/C

DESIGNATED HOOKUP FOR THE FOLLOWING FUJIFILM ENDOSCOPES*:EC-250HL5 EC-450WI5 EC-530LT EC-590WL EC-590ZW3M EG-530CT

EC-250LS5 EC-450WL EC-530MT EC-590WI4 EC-590ZWL EG-530UR

EC-250WE5 EC-450WL5 EC-530WI EC-590WL4 EC-590ZWM ES-250WE5

EC-250WI5 EC-450WM5 EC-530WL EC-590WM4 EC-L590ZW ES-450WE5

EC-250WL5 EC-530FI EC-530WL3 EC-590WL-H EC-L590ZW/L ES-530WE

EC-250WM5 EC-530FL EC-530WM EC-590WM EG-250CT5

EC-450HL5 EC-530FM EC-530WM EC-590WM2 EG-410CT

EC-450LS5 EC-530IT EC-590WI EC-590ZW3L EG-450CT5

IF THE ENDOSCOPE IS NOT LISTED HERE PLEASE SEE THE OTHER GUIDES.

ADDITIONAL REQUIRED ENDOSCOPE CONNECTIONS - VERSION 3

Basin Connector (see Basin Connection Instructions on pg 2)

Air Probe Connector

Channel Connection Tube Biopsy Inlet Connector

Suction Connector Forward Water Jet Connector

Leak Test Hookup PN DSD-110-HU0103

For MEDIVATORS AERs with leak test capability. (see Basin Connection Instructions on page 2)

A1

A2

FUJIFILM™ CA-503 KIT which includes:

FUJIFILM CA-503 S/A

FUJIFILM CA-503 T/R/C

FUJIFILM CA-503 B/C

FUJIFILM CJ-500

FUJIFILM™ CA-503S/A Cleaning AdapterAlign the red and blue rings with the corresponding valve ports. Squeeze the black tab and frame while inserting the connector. Release the tab to secure the frame. Ensure that the frame is properly secured.

Leak Test ConnectorAlign the pins on the connector with the slots on the scope. Push on and turn clockwise to engage.

A2

FUJIFILM CA-503T/R/C Cleaning AdapterAlign the slots on the cap with the pins on the Air/Water bottle connector, push on fully and twist clockwise to engage.

FUJIFILM CA-503B/C Cleaning AdapterPlace the biopsy slide over the biopsy inlet, depress the frame and slide until fully engaged.

Suction ConnectorAlign suction connector with suction barb and push onto barb until fully engaged.

Channel Connection TubeAlign each stainless steel fitting with corresponding female luer, push and turn clockwise to connect.

Biopsy Inlet ConnectorAlign stainless steel fitting with biopsy port, push and turn clockwise to connect.

Forward Water Jet ConnectorAlign stainless steel fitting with female luer, push and turn clockwise to connect.

CJ-500 Cleaning AdapterAlign the CJ-500 adapter with the forward water jet channel inlet and screw clockwise to engage.

Air Probe ConnectorAlign with air probe, push and slide until fully engaged.

© 2016 Medivators Inc. All Rights Reserved. 50098-349/C

* This list of corresponding endoscope models is subject to change without notice. For a full list, please go to:www.medivators.com/hookuplookup

VERSION 2DSD HOOKUP ENDOSCOPE CONNECTORS (Figure 1)

VERSION 3DSD HOOKUP ENDOSCOPE CONNECTORS (Figure 1)

* Representative endoscope for illustrative purposes only. For exact port locations, refer to endoscope manufacturer’s instruction guide.

Ensure that the endoscope model you are connecting is listed on this IFU.

NOT PICTURED ABOVE

Luer Fitting behind Biopsy Port

ES-200ER

Luer Balloon Connection

EC-450BI5, EN-450P5/20, EN-450T5, EN-450WM23

Luer Biopsy Connection

EG-530UT

* This list of corresponding endoscope models is subject to change without notice. For a full list, please go to:www.medivators.com/hookuplookup

* This list of corresponding endoscope models is subject to change without notice. For a full list, please go to:www.medivators.com/hookuplookup

DESIGNATED HOOKUP FOR THE FOLLOWING FUJIFILM ENDOSCOPES*:

EN-580T

IF THE ENDOSCOPE IS NOT LISTED HERE PLEASE SEE THE OTHER GUIDES.

* This list of corresponding endoscope models is subject to change without notice. For a full list, please go to:www.medivators.com/hookuplookup

ADDITIONAL REQUIRED ENDOSCOPE CONNECTIONS - VERSION 4

FUJIFILM™ CA-503S/A Cleaning AdapterAlign the red and blue rings with the corresponding valve ports. Squeeze the black tab and frame while inserting the connector. Release the tab to secure the frame. Ensure that the frame is properly secured.

Leak Test ConnectorAlign the pins on the connector with the slots on the scope. Push on and turn clockwise to engage.

A2

FUJIFILM CA-606 Cleaning AdapterAlign the CA-606 adaptor with the balloon channel inlet and screw clockwise to engage.

FUJIFILM CA-503B/C Cleaning AdapterPlace the biopsy slide over the biopsy inlet, depress the frame and slide until fully engaged.

Suction ConnectorAlign suction connector with suction barb and push onto barb until fully engaged.

Channel Connection TubeAlign each stainless steel fitting with corresponding female luer, push and turn clockwise to connect.

Biopsy Inlet ConnectorAlign stainless steel fitting with biopsy port, push and turn clockwise to connect.

Balloon ConnectorAlign stainless steel fitting with female luer, push and turn clockwise to connect.

Air Probe ConnectorAlign with air probe, push and slide until fully engaged.

Basin Connector (see Basin Connection Instructions on pg 2)

Air Probe Connector

Channel Connection Tube Biopsy Inlet Connector

Suction Connector Balloon Connector

Leak Test Hookup PN DSD-110-HU0103

For MEDIVATORS AERs with leak test capability. (see Basin Connection Instructions on page 2)

A1

A2

FUJIFILM CA-503T/R/C Cleaning AdapterAlign the slots on the cap with the pins on the Air/Water bottle connector, push on fully and twist clockwise to engage.

If you have any questions regarding this hookup, in the U.S., please contact MEDIVATORS Technical Support at 1-800-444-4729. Outside the U.S., please contact your local MEDIVATORS Distributor.

Medivators BVSourethweg 116422PC HeerlenThe NetherlandsTel: +31.45.5.471.471

Cantel Medical Asia/Paci�c Pte. Ltd.1A International Business Park#05-01 Singapore 609933Tel: +65.6227.9698

www.medivators.com

Manufactured in the USA by:

Medivators Inc. 14605 28th Avenue NorthMinneapolis, MN 55447 USAToll Free: +1.800.444.4729

Cantel Medical Devices (China) Co. Ltd.Unit 804-805, Innov Tower Block A,Hongmei Road, Xuhui 200233 ShanghaiTel: +86 21 60161380Fax: +86 21 61210913

Ensure proper alignment of all connectors with the endoscope mating parts prior to attachment.

VERSION 4DSD HOOKUP ENDOSCOPE CONNECTORS (Figure 1)

* Representative endoscope for illustrative purposes only. For exact port locations, refer to endoscope manufacturer’s instruction guide.

Ensure that the endoscope model you are connecting is listed on this IFU.

WARNINGS & CAUTIONSThis hookup is intended for connection only with the specific endoscope models identified in the appropriate MEDIVATORS online hookup guide. Use of this hookup with endoscopes other than those specified by the on line hookup guide may result in inadequate HLD.

Users operating MEDIVATORS AERs and hookups must be trained and competent in the understanding of endoscope channel systems.

Prior to reprocessing, users must verify that the hookup used contains connections for all appropriate channels that require a separate reprocessing connection according to the hookup instructions.

All connectors and adapters must remain firmly attached and unrestricted for the entire disinfection cycle to ensure adequate HLD. Users must inspect the endoscope and hookup combination to verify proper connection and flow.

Modifications or repairs to hookups which do not correspond to the manufacturer’s specifications may result in inadequate HLD and/or damage to endoscopes.

Failure to properly clean and prepare endoscope for immersion prior to hookup connection may result in inadequate HLD and/or damage to the endoscope.

Ensure proper alignment of all connectors with the endoscope mating parts prior to attachment. Misalignment may cause damage to hookup components and/or endoscope mating parts.

Endoscopes must be inspected for damage and verified to be in proper working order prior to hookup connection. Connection of hookups to damaged endoscopes or endoscopes with obstructed/restricted channels could result in inadequate HLD.

Manual leak testing must be carried out in accordance with manufacturer’s recommendations prior to reprocessing.

Install all necessary waterproof caps, plugs and cleaning adapters prior to immersion.

Do NOT autoclave MEDIVATORS hookups.

MEDIVATORS makes no claim on high-level disinfection efficacy when these reprocessing instructions are not followed, or when this hookup kit is applied to endoscopes other than those specified in the online interactive hookup guide.

Avoid biological contamination. Always wear personal protective equipment when handling endoscopes.

CLEANING ADAPTERS AND CHANNEL PLUGSFor some endoscope models, endoscope manufacturer and model-specific cleaning adapter(s) and/or channel plugs may be required to prepare the endoscope for reprocessing. MEDIVATORS manufactures some MEDIVATORS cleaning adapters and/or channel plugs specific to each endoscope manufacturer and model(s). The requirement for these additional adapter(s) will be noted in the online hookup guide. Where MEDIVATORS does not manufacture or supply these adapters, reference will be made to the use of adapters from the endoscope manufacturer’s manual cleaning kit supplied with the endoscope.

DSD HOOKUPSDSD Hookups are the interface between the MEDIVATORS® DSD and SSD Series Automated Endoscope Reprocessors (AERs) and the endoscope, to ensure correct connectivity and flow rates per manufacturer’s specifications.

LEAK TEST HOOKUPSThe Leak Test Hookup connects the endoscope’s leak test port to the DSD/SSD Basin Leak Test Connector for leakage detection in the endoscope. MEDIVATORS recommends the use of the leak test feature in AERs with the leak test capability during the reprocessing cycle.

The Leak Test Hookup will always be used in conjunction with a DSD Hookup. For endoscopes with channels, please refer to the online hookup guide for the appropriate DSD Hookup. For endoscopes without channels, hookup DSD-110-HU0122 must always be used.

Leak Test Hookups are endoscope manufacturer specific. One leak test hookup is supplied per each endoscope basin based on individual facility endoscope inventory. Leak test hookups are also sold separately.

MEDIVATORS DSD/SSD automated endoscope reprocessors with leak test capability include: DSD-201LT, DSD EDGE, SSD-102LT.

DSD/SSD BASIN CONNECTION INSTRUCTIONSDSD Hookup Insert the DSD Hookup quick connect fitting into the AER by inserting into the Basin Endoscope Hookup Connector, and push down until the fitting clicks into place.

Leak Test Hookup Insert the Leak Test Hookup quick connect fitting into the Basin Leak Test Connector and push until fitting clicks into place. See figure 2.

DSD Hookup Quick Connect

Basin Endoscope Hookup Connector

Leak Test Hookup Quick Connect

Basin Leak Test Connector

Figure 2

A1

FUJIFILM™ CA-503 KIT which includes:

FUJIFILM CA-503 S/A

FUJIFILM CA-503 T/R/C

FUJIFILM CA-503 B/C

FUJIFILM CJ-500