Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
CAPABILITIES
SPECIFICATIONS
Parameter LOT MT /MTH HT
Temperature range -48°C to 150°C (@20°C)MT Tamb to 600°C
200 to 1100°C
Stability ±0.03°C ±0.05°C @ 450 °C ±0.3°C @ 1000 °CAccuracy ±0.15°C ±0.3°C ±0.3% rdgResolution 0.01°CUniformity (well to well) ±0.1°C ±0.2°C ±0.2°CHeating time 20°C/min 20°C/min 20°C/minCooling time 22°C/min 25°C/min 18°C/minWell size 35 x 135mm 50 mm 44 mmImmersion depth 130 mm 185 mm 220Standard insert 3.5-4.5-5.5-6.5-8.5-10.5mm3.5--5-6.5-8.5-12.5 mm 7-9-11-13.5 mm
phisical dimension 16 x 33 x 34 cm 14 x 37 x 30 cm 17 x 45 x 33 cmPower supply 110 or 220 Vac, 50/60 Hz 110 or 220 Vac, 50/60 HzWeight 10 Kg 8 Kg 12 KgInterfaces RS-232 standard RS-232 standard RS-232 standard
Certificate included included included
Features LOT MT HTRange -48°C to 150°C Tamb 600°C T200 to 1100°CRemovable inserts • • •Customised inserts optional optional optionalFast heating/Cooling • • •Switch test • • •Fully Programmable ramp • • •Electrical inout for Reference and UUT TC version TC version TC versionSerial Interface RS232 • • •ATCS compatible • • •
Electrical Input (TC version) Range Accuracy
RTD’s type PT100 (IEC), 3,4 wires -100°C to 700 °C 0.3°CT/C’s type J 0 to 1000°C 1°CT/C’s type K, N, R, S 0 to 1300°C 1°C
ModelLOT Dry Block Temperature Calibrator
MT Dry Block Temperature Calibrator HT Dry Block Temperature Calibrator
Version TS basic version TC Dual input for RTD’s and TC (Ref and UUT)
Power supply1 = 115 Vac, 50/60 Hz2 = 220 Vac, 50/60 Hz
LOT - TC-2 Typical ordering code
Standard delivery includes: Standard insert, switches test leads, instructionmanual, calibration certificate with data.
Accessories and inserts: See accessories page
Automatic Temperature Calibration System (ATCS)Save time and effort by performingon-line calibration procedure. TheATCS system is an high accuracymodular scanner for T/C’s, RTD’spressure switches and temperaturetransmitters controlled by theadvanced version softwarepackage. The system is able to handlemany temperature sensorssimultaneously. Learn more on howATCS can improve efficency in your calibration routineand saving money.
The Eigroup range of field dry blockcalibrator consist of there separatemodels for calibrating temperaturesensors, loop and switches in atemperature range between -48 and1100 °C. These self-contained units areportable, flexible and are suitable forboth field and laboratory temperaturecalibration.
The control technology ensure extremelyfast setting time whilst offering anexcellent temperature stability and wellto well uniformity. The new redesignedcontroller deliver high display readingaccuracy. For improved measurementuncertainty, Eigroup provide a complete range of “working standard” both traceable and UKAS certified.
Multi-hole insert are available to suit thevariety of probe size and the
programmable ramp and switch inputmake simple and quick the testing oftemperature switches. The MaxiCal Temp Pro series is available asas temperature source only, or TC versionwhich provide dual electrical input forThermocouple and RTD’s.
The standard RS232 and softwarepackage allow to run automatic test andcalibration procedure.
LOT , uses a special peltier element tocool from ambient temperature down to48 °C. Features:• Wide range from -48 to 150 °C• Excellent stability to 0.03 °C • well to well uniformity of 0.1 °C• Reading accuracy ±0.15 °C
MT , suitable for intermediatetemperature range offers very fastheating and cooling time. Features:• Wide range from 20 to 600 °C• Excellent stability to 0.05 C • well to well uniformity of 0.2 °C• Reading accuracy ±0.3 °C
HT , for very high temperature range,still do not compromise in technicalspecification. All the necessaryprecaution are taken to ensure maximumsafety and temperature stability. Features:• Wide range from 200 to 1100 °C• Excellent stability to 0.3 °C• Ceramic insulation for safety and
stability• Customised inconel inserts
Ordering Informations
TM-Series Compact
E InstrumentsAdvanced Fluid & DryBlock Calibrators
MAXICAL TEMP PRO
Bulletin CAL TEMP PRO - ed. 08-2011
TECHNICAL SPECIFICATIONS
ModelLOT-Fluid MicroBath Temperature CalibratorMT-Fluid MicroBath Temperature Calibrator
Version TS Basic version TC Dual input for RTD’s and TC (Ref and UUT)
Power Supply1 = 115 Vac, 50/60 Hz2 = 220 Vac, 50/60 Hz
LOT Fluid - TC - 2 Typical ordering code
Standard delivery includes: 2 x 500 cc silicon oil bottle, closing cup for transport,empty fluid system, support for glass thermometer, electrical power cord, sparesfuses, clamping-screw adapter for bushes, switch test leads, instruction manual,calibration certificate with data and soft vinyl transport case and CalPman basicsoftware package.
Accessories and inserts: See accessories page
Automatic Temperature Calibration System (ATCS)Save time and effort by performing on-line calibration procedure.The ATCS system is an high accuracy modular scanner for T/C’s,RTD’s pressure switches and temperature transmitters controlledby the advanced version software package. The system isable to handle up to 32 temperature sensors simultaneously. Learnmore on how ATCS can improve efficency in your calibrationroutine and saving money.
The LOT/MT Fluid represent one ofthe smaller portable liquid bathavailable on the market. Whereveryou go with this compact selfcontained liquid bath you can counton their performance. Both modeloffer an excellent stability to 0.02 °Cand an improved uniformity or 0.03°C for low uncertainty budget whenusing a reference thermometer. Theuniformity is maintained by amagnetic mixer which speed isadjustable ti suit the fluid viscosity.The excellent display accuracy it hasbeen greatly improved to 0.15 °C sothat you can use the bath forcalibration without a referencethermometer.
We have obtained the stability andaccuracy performance of a liquid bathinto a small self contained dry blocksize.
EXTREME FLEXIBILITY- 2 IN 1 When the use of the liquid it is notaccepted for your application, in suchcases like risk of contamination mightoccur, then the MaxiCal Fluid offerthe great flexibility to add a “DryBlock insert” turning the unit into acommon Dry Block temperaturecalibrator. These calibrator features• Two different models• Temperature range from -35 to
200 °C• Display Accuracy to 0.15 °C• Stability 0.02 °C• Uniformity to 0.03 °C• Large Capacity 400 cc• 2 in 1 Liquid Bath and Dry Block• Dual Electrical Input (TC version)• RS232 interface as standard• Switch test contact capture• Programmable Ramp
APPLICATIONThe Temp Fluid are the perfectsolution when you need laboratoryaccuracy in the field, or when youcalibrate short, square or odd-shapedsensors. The liquid bath eliminatethe source of error related toimmersion depth and fit which youcan find using common dry blockcalibrator. The fluid media offer anexcellent thermal contact anduniformity. These calibrators are theperfect solution when calibratingsolution. • Short, odd-shaped and square
sensors• Bimetal and glass thermometer
Ordering Informations
Parameter LOT Fluid MT Fluid
Temperature range -30 °C to 125 °C (@ 20 °C) Tamb to 200 °CStability ±0.02 °C @ 5 °C ±0.02 °C @ 150 °CAccuracy ±0.15 °CResolution 0.01 °CUniformity (radial) ±0.03 °C as liquid bathUniformity(axial) ±0.03 °C as liquid bathHeating Time 12 °C/min 10 °C/minCooling Time 9 °C/min 3 °C/minWell Size 50 mm (400 CC capacity)Well Depth 150 mmphisical Dimension 16 x 33 x 35 cm Weight 8 KgPower Supply 110 or 220 Vac, 50/60 Hz 110 or 220 Vac, 50/60 HzInterfaces RS-232 standard RS-232 standardCertificate included included
Electrical Input (TC version) Range Accuracy
RTD’s type PT100 (IEC), 3, 4 wires -100 °C to 700 °C 0.3 °CT/C’s type J 0 to 1000 °C 1°CT/C’s type K, N, R, S 0 to 1300 °C 1°C
Features LOT Fluid MT FluidRange -30 °C to 125 °C Tamb to 200 °CType Liquid bath Liquid bathDry Block Insert optional optionalFast heating/cooling • •Switch test • •Fully programmable ramp • •Electrical inout for reference and UUT TC version TC versionSerial Interface RS232 • •ATCS compatible • •
FEATURES
TM-Series Compact
E InstrumentsMicro Bath Fluid Calibrators
MAXICAL TEMP PRO
Accessories
PSER-ACC-VC Vinyl carrying case with shoulder strap PSER-ACC-AC Aluminium carrying case PHT-ACC-AC Alumium transit case case HT onlyPSER-ACC-USB USB cable/converter
LOT INSERTSOrdeing codePLOT-INS-A Blank insertPLOT-INS-B 6 holes: 4, 4.5, 5.5, 6.6, 8.5, 10.5 mm (included as standard)PLOT-INS-C 2 holes: 6.5 mm, 19.5 mmPLOT-INS-S3 Custom insert 3 holes maxPLOT-INS-S6 Custom insert 6 holes maxPLOT-INS-S10 Custom insert 10 holes max
MT INSERTSOrdeing codePMT-INS-A Blank insertPMT-INS-B Standard Insert 5 holes 12.5, 8.5, 6.5, 4.5 and 3.5 mmPMT-INS-S3 Custom Insert, 3 holes maxPMT-INS-S6 Custom Insert, 6 holes max
PSER-ACC- ACPHT-ACC-AC
PSER-ACC-VC
PSER-ACC-USB
6.5 mm
19.5 mm
4.5 mm5.5 mm4 mm6.6 mm10.5 mm8.5 mm
INS-A INS-B INS-C INS-S3 INS-S6 INS-S10blank max. 3 holes max. 6 holes max. 10 holes
INS-A INS-B INS-S3 INS-S6 INS-E INS-Fblank supplied as standard max 3 holes max 6 holes
HT INSERTSOrdeing codePHT-INS-A Blank insertPHT-INS-B 4 holes: 7, 9, 11, 13.5 mm, included as standard
LOT/MT FLUID INSERTS Ordeing codePFL-INS-D Blank aluminium insert 60 x 170 mmPFL-INS-E Drilled aluminium insert 60 x 170 mm with: 2 x 4 mm, 1 x 4.5 mm,
1 x 5.5 mm, 2 x 6.5 mm, 1 x 8.5 mm, 1 x 10.5 mm and 1 x 12.5 mm
7 mm11 mm
13.5 mm
9 mm
4 mm4.5 mm5.5 mm
6.5 mm
6.5 mm
8.5 mm
10.5 mm12.5 mm
INS-A INS-Bblank INS-D INS-E
blank
12 mm
• • • • • Distributed by:
3.5 mm4.5 mm6.5 mm8.5 mm12.5 mm
E Instruments Group [email protected]
TM-Series Compact
E InstrumentsAdvanced Fluid & DryBlock Calibrators
MAXICAL TEMP PRO