troubleshooting guide
SeQual® Technologies inc.
eclipse® Platform PAOS™
EclipsE Data acquisition toolversion 3.1
2
tAble oF Contents
iNTROduCTiON 3
PRePARATiONS PRiOR TO uSe 3
BegiN HeRe 4
eVeNT 1 - YeLLOW LigHT FLASHiNg 5
BATTeRY STATuS iNdiCATOR ReAdiNg NeAR ZeRO 6
eVeNT 2 - YeLLOW LigHT iS STeAdY 7
eVeNT 3 - Red LigHT iS BLiNKiNg 8
eVeNT 4 - Red LigHT iS ON 9
eVeNT 5 - LOW PuRiTY 12
eVeNT 6 - CANNOT CHANge FLOW SeTTiNgS ON FRONT PANeL 15
eVeNT 7 - THe uNiT iS NOiSY OR RATTLiNg 15
eVeNT 8 - PuLSe MOde dOeS NOT FuNCTiON 15
eVeNT 9 - geNeRAL eVeNTS LiST FAuLTS 16
APPeNdiX A 19
APPeNdiX B 20
OPeRATOR COdeS 21
SeQuAL CONTACT iNFORMATiON 22
Eclipse Data aquisition tool
eclipse data Aquisition Tool Troubleshooting Guide
3
This Troubleshooting Guide together with the EDAT software is intended to provide a qualified technician specific information that will aid in identifying potentially faulty modules or parts associated with a device malfunction.
This Guide and the associated EDAT software are not to be considered replacements for basic technician troubleshooting techniques used in medical oxygen system repair and servicing.
It is assumed that the EDAT operator/service technician has attended the SeQual Level 1 Technical Training session and/or has a thorough understanding of the methods and technical practices necessary to correctly and safely remove and replace modules and components in SeQual Eclipse oxygen concentrators.
The application of the EDAT User’s Manual, EDAT Troubleshooting Guide and associated EDAT software and hardware is limited to Eclipse1 (Model 1000), Eclipse 2 (Model 1000A), Eclipse 2 Plus and Eclipse 3 (Model 1000B) medical oxygen concentrators only.
This Troubleshooting Guide does not constitute a User Manual nor does it represent the complete documentation associated with this EDAT software set. (Refer to the EDat user Manual for details.)
PrePArAtion Prior to use
•Computerisonandfunctioning •EDATsoftwaresetisonandfunctioning(softwarekeyrequired) •EDATinterfacecableconnectedtothecomputerUSBconnectorandto the Eclipse serial connection
consult the user Manual for details on correct operation of EDat and Eclipse devices.
Before proceeding, ensure that the Eclipse is not providing oxygen to a patient nor should a patient be connected to the Eclipse during this process or any troubleshooting or diagnostic activity.
WaRninG:
eclipse data Aquisition Tool Troubleshooting Guide
4
begin here
Before the technician determines the best place to start the fault identification process the event table should be downloaded from the Eclipse device. This is an important piece of information that can be used to determine possible faults and problem resolution. Real time performance data is also displayed.
First: •TurnontheEclipse •EnsurethatthecomputerisonandEDATisrunningandthattheEclipseiscorrectlyconnected to your computer system. •PressEclipse in the EDAT drop down menu, then press Event table and EDAT will download both power manager and control board event tables
oPerAtor rePorteD FAilure inDiCAtion(s)(Determine the one that best fits the situation)
•TheEclipseisdisplayingredoryellowflashingorsteadyalarms •CannotchangeflowsettingsonFrontPanel •TheEclipseisexcessivelynoisyorrattling •PulseModedoesnotfunction •TheEclipsewillnotpoweron
eclipse data Aquisition Tool Troubleshooting Guide
5
event 1 - Yellow light FlAshing Eclipse 2 plus & Eclipse 3 error codes: 002, 008, 010, 020 Eclipse & Eclipse 2 error codes: 02, 08, 10, 20
assumptions:
•Eclipsepluggedintoac power and operating •EDATconnectedandoperatingcorrectly •Eventlogsdownloaded •Errorcodes02,08,10or20displayedonEclipsecontrolpanel
Observe value of the alarm threshold on the EDAT real time screen.
is the AlArm thresholD vAlue below 170?
Refer to the low purity, Event 5
Continuetonextstep
eclipse data Aquisition Tool Troubleshooting Guide
6
is the bAtterY stAtus inDiCAtor on the eCliPse Control PAnel reADing neAr zero (1-2 bArs For e1, 1 bAr For e2)
Review the Event table and determine if low Battery appears in the last few events of the power Manager Event table.
it is likely that the charge on the power cartridge is very low. If the battery charge indicator is not showing charging (cascading display) and the External Power indicator is oFFthenthemostlikelyfaultsareexternalorthe power, and could be:•ACwalloutletfailed•ACPowerconverterfailedorcordisunpluggedatconverterbox•PowerCartridgehasoverheatedfromchargingoruse (Let cartridge cool to normal operating range)the device may need service with possible faults in: •PowerManagerPCB•BatteryBridgePCB•Wiringharness•FailedPowerCartridge
after further investigation of the possible faults listed above, was the fault resolved?
Test and certify the device for patient application
ContactSeQualTechnicalService
eclipse data Aquisition Tool Troubleshooting Guide
7
event 2 - Yellow light is steADY Eclipse 2 plus & Eclipse 3 error codes: 001, 020 Eclipse 2 error codes: 01, 20 Eclipse error codes: 20
assumptions:
•EclipsepluggedintoACPowerandoperating •EDATconnectedandoperatingcorrectly •Eventlogsdownloaded •ErrorcodedisplayedonEclipsecontrolpanel
Was the unit turned on within the last minute or so?
The unit is still in warm-up. This is normal. Wait5-10minutes
Checkbothinput(underintakefilter)andexhaustoutlet(nearbottomofunit)forobstructions.Insureintakefilterisclean.Removeallexternaloutput items such as cannulas, humidifiers and extensions.
are inlet and outlet clear?
Proceed to next step
Install clean, dry inlet filter or remove debris from outlet as needed. Restart device and verify operation
Remove device cover, power unit on and carefully check for air flow near the bottom of the compressor housing. is air flow present?
Dovisualcheckforkinkedlines,airleaks,obstructions.ContactTechnicalServices.
CheckforanylooseelectricalconnectionstocompressororATF.IfOK,itislikelythecompressororthePowerSupply/BatteryBridgePCBcould have failed.
eclipse data Aquisition Tool Troubleshooting Guide
8
event 3 - reD light is blinking Eclipse 2 plus & Eclipse 3: 004, 040 Eclipse 2: 04, 40 Eclipse: 04
is the ac or Dc plug disconnected?
PlugtheACorDCpowersupplyintotheunit.Ifredlightcontinuestoblink,ensureexternalpowersourceisadequateforpowersupplyandthatAC/DCPowerSupplyisfunctioningcorrectly.-CheckEDATdisplaypower in Voltagereads28volts-CheckPowercartridgechargestatus.IfexternalsupplyOK proceed to next step
AttachACorDCpowersupply.VerifythattheExternal Voltage is in the acceptable range.
is the alarm threshold on EDat equal to or below 140?
Refer to low purity, Event 5
Check9VDCbatterypoweronEclipseControlPanel(E2Only).Installnew battery (or measure the battery voltage) for E1 devices. Replace as needed.
is Red light still blinking?
A careful visual inspection of tubing, filters, connections and componentsforleaks(pressureorvacuum)shouldbeperformedtodetermine possible failed component. Modules that could have failed includeATFandcompressor.ContactTechnicalServices.
Assemble device and test
eclipse data Aquisition Tool Troubleshooting Guide
9
event 4 - reD light is on (unit shutDown) Eclipse 2: Refer to fail code description for details
Before proceeding with the steps below, refer to the malfunctions code list which should assist the technician in identifying the specific cause and possible correction.
step 1.TurntheEclipsedeviceOFF.Withthedeviceoff,ensurethattheEDAT communications cable is connected to the Eclipse unit.
step 2. Select Eclipse in the EDAT drop down menu then press Event table. The event list will download from the unit. This will download the Event table including the start-up information
step 3.Withthecover,PowerCartridgeandAC/DCPowerSupplyremoved,carefullyinspectthe connectionbetweentheBatteryBridgePCBandthePowerManagerPCB.Iftheboardisnot connected properly, reconnect the boards.
eclipse data Aquisition Tool Troubleshooting Guide
10
step 4.RemovetheBatteryBridgePCBfromthePowerManagerPCB.Inspecttheconnectoron thePowerManagerOCBthatconnectstotheBatteryBridgeOCBfordamageorcorrosion. ReplacePCBsasneeded.
NOTE:PictureshowsBatteryBridgePCBwithJ1connectorhousingremoved.
broken PCb Pin
DAmAgeD PCb Pins
eclipse data Aquisition Tool Troubleshooting Guide
11
step 5. The compressor could be incorrectly connected to the Power Manager Board. Verifytheconnectionsarecorrect.
a close-up of the connection between the compressor and the battery bridge board
eclipse data Aquisition Tool Troubleshooting Guide
12
event 5 - low PuritY Eclipse 2 plus & Eclipse 3 error codes: 004, 008 Eclipse & Eclipse 2 error codes: 04, 08 step 1. General Tubing.Checkforleaksorkinks(vacuumandpressure)onallplastictubing, andtherubberexhausttube(indicatedbyarrowsinthephotos).Kinkedtubingshouldalwaysbe replaced and NOT reused.
eclipse data Aquisition Tool Troubleshooting Guide
13
step 2. HEPAFilter. Inspect the HEPA filter (indicated by arrow). Replacefilterifitisdirtyorleaking.
eclipse data Aquisition Tool Troubleshooting Guide
14
step 3. ProductTanks,ATF.WhiletheEclipseisrunningandtheiscoverremoved.Useanapproved leakdetectionsolution(Snooporequivalent).Spraylightlyontubeconnectionsandother assembledmodulecomponents.WhentestingtheATFuseextremecarenottowettheoutputend oftheATFassembly.Whenbubblesareseenatornearacomponentormoduleitisanindication that item is faulty. Pay particular attention to the areas indicated in the pictures below:
TheProducttankrubbercapsandendconnectors
ThehardwarefittingsoftheATF
eclipse data Aquisition Tool Troubleshooting Guide
15
event 6 - CAnnot ChAnge Flow settings on Front PAnel
Remove the connector between the membrane panel and the control panel. Replace the membrane panel. UseextremecautionwhendisconnectingthemembranepanelfromtheControlPCB.DamagetotheControlPCBcaneasilyresult.
event 7 - the unit is noisY or rAttling
Listenforaclackingorunusualsoundnearthecompressor.Ifthereisaheavyclackingsound,replacethecompressor module assembly.
event 8 - Pulse moDe Does not FunCtion
•PressthePulseFlowModeButtontotoggletheoxygendeliverymodeonthefrontoftheEclipseunit.•Iftheflowdeliverymodedoesnotchange,connecttheEclipseunittoEDAT.•ClickontheControlPanelmenuitemintheEclipsemenuofEDAT.•Toggletheflowdeliverymode.•IftheEclipseunittogglestheoxygendeliverymodecorrectly,replacethemembranepanel.•IftheEclipseunitdoesnottoggletheoxygendeliverymodecorrectly,contactTechnicalServices.
eclipse data Aquisition Tool Troubleshooting Guide
16
event 9 - generAl events list FAults
step 1. Review both event lists from EDAT (a summary of these are available in Appendix A) and determine if one of these events can be identified in the most recent ten events listed.
•Battery communication Failure •Breath Detect sensor Failure •ultrasonic Failure (shutdown) •compressor Motor too Hot (shutdown) •pcB too Hot (shutdown) •low nine Volt Battery (9VDc)
If any of these 6 events have occurred refer to the list below to determine the probable cause and resolution to the failure mode.
Battery communication Failure
Possible faults could be: •ConnectionbetweenbatterybridgeboardandPowerManagerPCBisfaultyorcorroded •BatteryBridgePCBneedstobereplaced •PowerCartridgeneedstobereplaced •PowerManagerPCBneedstobereplaced(lesslikely)
eclipse data Aquisition Tool Troubleshooting Guide
17
Breath Detect sensor Failure
ThereisahighprobabilitythattheControlPCBneedstobereplaced.
ultrasonic Failure (unit has shutdown)
ThereisahighprobabilitythattheControlPCBneedstobereplaced.
compressor Motor too Hot (unit has shutdown)
Themostlikelyfailedmoduleisthecompressor Module. Before beginning this replacement, however,otheritemsthatcouldbecheckedinclude: •Blockedinlet(filter)and/orairpassagesintheEclipsecase •PossibleexternalblockagesofnormalairflowinandoutoftheEclipse(animalhair, dirt, carry bags or other restrictions) •Kinkedtubingfromreplacingcompressorandcuttingtubingtooshort •Badelectricalconnection
pcB too Hot, refers to the power Manager pcB (unit has shutdown)
•RemovethefrontcoveroftheEclipseunit •CarefullyconnectACpowertothedeviceandobserveifthesmallelectronicscoolingfan locatedonthePowerManagerPCBisrunningornoisy(theEclipsedoesnothaveto be turned on) •Ifthefanisnotrunning,verifyitiscorrectlypluggedintothePowerManagerPCB •Usinganotherfanwilldetermineifthefanhasfailed •Removepowerbeforecontinuingrepair
eclipse data Aquisition Tool Troubleshooting Guide
18
low 9 Volt Battery (three short beeps on post)
Replacethe9VoltBattery.
FAn ConneCts here
eclipse data Aquisition Tool Troubleshooting Guide
19
APPenDiX AEclipse 2 plus & Eclipse 3 only
DISPLAYEDONLYAFTERACCESSINGDIAGNOSTICCODESONUNIT
alarm priority Green Yellow Red Horn code
AllOK,warm-up done
14/None on OFF OFF OFF None
Warm-up: %O2<85
13/Low on on OFF OFF 008
Warm-up:%O2 < 70
12/Low on on on OFF 004
OneWire 11/Low N/A (Flashing) oFF 1 beep 100
CannotchargePowerCartridge
10/Low N/A on OFF OFF 001
No breath detection
9/Medium on on OFF OFF 200
70% < %O2<85%
8/Medium OFF (1s/1s) OFF OFF 008
Flowrateerror 7/Medium OFF (1s/1s) OFF (0.5s/119.5s) 020
Warmbattery 6/Medium on (1s/1s) OFF (0.5s/119.5s) 002
Low battery 6/Medium on (1s/1s) OFF2 Beeps
0.5s/0.5s every 30s010
DCcabledetected
4/Medium N/A (1s/1s) OFF (0.5s/1.5s) None
%O2 < 70 3/High OFF OFF (0.25s/0.25s)3 Beeps
(0.5s/0.5 every 120s)
004
Loss of power 2/High OFF OFF (0.25s/0.25s) on 040
Malfunction 1/High OFF OFF on 10s then silence 080
eclipse data Aquisition Tool Troubleshooting Guide
20
APPenDiX bEclipse 2 only–pleaser refer to user Manual for other versions.
DISPLAYEDONLYAFTERACCESSINGDIAGNOSTICCODESONUNIT
alarm priority Green Yellow Red Horn code
AllOK,warm-up done
13/None on OFF OFF OFF None
Warm-up: %O2<85
12/Low on on OFF OFF 08
Warm-up:%O2 < 70
11/Low on on on OFF 04
CannotchargePowerCartridge
10/Low N/A on OFF OFF 01
No breath for45seconds
9/Medium on on OFF OFF 20
70% < %O2<85%
8/Medium OFF (1s/1s) OFF OFF 08
Flowrateerror 7/Medium OFF (1s/1s) OFF (0.5s/119.5s) 20
Warmbattery 6/Medium on (1s/1s) OFF (0.5s/119.5s) 02
Low battery 6/Medium on (1s/1s) OFF2 Beeps
0.5s/0.5s every 30s10
DCcabledetected
4/Medium N/A (1s/1s) OFF (0.5s/1.5s) None
%O2 < 70 3/High OFF OFF (0.25s/0.25s)3 Beeps
(0.5s/0.5 every 120s)
04
Loss of power 2/High OFF OFF (0.25s/0.25s) on 40
Malfunction 1/High OFF OFF on 10s then silence
eclipse data Aquisition Tool Troubleshooting Guide
21
oPerAtor mAlFunCtion CoDes
if a malfunction occurs in the eclipse, the device will stop, the Red Led on the front panel will light and the buzzer will sound for 10 seconds and then silence. The LCd will display one of the following Malfunction Codes:
malfunction Code recommended Action
Invalid RESET 80 -VerifyfunctionalityofCSPCB
I/O port failure 81 -VerifyfunctionalityofCSPCB
RAM failure 82 -VerifyfunctionalityofCSPCB
FLASHfailure 83 -VerifyfunctionalityofCSPCB
EEPROM failure 84 -VerifyfunctionalityofCSPCB
iPC watchdog timeout 90-Checkforsecurecommunicationcables
-VerifyfunctionalityofPMandCSPCBs
Compressor motor too hot 91-Checkcompressorfanfunctionality
-Checktoensureventsarenotobstructed-VerifyfunctionalityofCompressorbox
PCB too hot 92-VerifyfunctionalityofPMfan
-Checktoensureventsarenotobstructed
Compressor motor stalled 94-Checkcompressorwiresfordamage
-VerifyfunctionalityofPMPCB-VerifyfunctionalityofCompressorbox
Battery communication failure 95
-VerifyfunctionalityofBatteryBridgePCB-CheckBatteryandBatteryBridgePCBforpindamage
-VerifyfunctionalityofBattery-VerifyfunctionalityofPMPCB
iPC failure A0 -VerifyfunctionalityofPMPCB
Product pressure sensor failure A1 -VerifyfunctionalityofCSPCB
Breath pressure sensor failure A2 -VerifyfunctionalityofCSPCB
ultrasonic failure A3-VerifyfunctionalityofCSPCB
-Checkforwaterintrusionfromhumidifier
Product temperature sensor failure A4-VerifyfunctionalityofDB9cableharness
-VerifyfunctionalityofCSPCB
Loss of product tank pressure failure A5-Checkforleaking,crackedHEPAfilterordamagedtubing
-VerifyATFmotoristurning
Purity calibration data failure A6-Checkforleaking,crackedHEPAfilterordamagedtubing
-VerifyfunctionalityofCSPCB
Flow calibration data failure A7-VerifyfunctionalityofProportionalValve
-VerifyfunctionalityofCSPCB
Breath sensitivity data failure A8-Checkfordamagedtubing
-VerifyfunctionalityofCSPCB
Hour of operation data failure A9 -VerifyfunctionalityofCSPCB
invalid assembly option AA - Remove power to reset
Ambient pressure sensor failure Ab -VerifyfunctionalityofCSPCB
Pn: 5467
Rev. C
0909
seQuAl Customer serviCe ContACt inFormAtion