20
Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part No. 2420-EX For 2 NTC sensors; external: -50°C..140°C / internal: -35°C..55°C External Start / Alarm Reset / InPos detection (with special connector) ECOLOG TN3-P Part No. 2420-P ECOLOG TN3-P for EX-Zone 1 Part No. 2420-PEX For 3 NTC sensors; external: -50°C..140°C / internal: -35°C..55°C Alarm output External Start / Alarm Reset / InPos detection (with special connector) Automatic printout for data and alarm protocols ECOLOG TN4 Part No. 2421 ECOLOG TN4 for EX-Zone 1 Part No. 2421-EX For 4 NTC sensors -50°C..140°C Alarm output; 2 digital inputs 4 button keypad; External Start, Alarm Reset, Measurement and Alarm Scroll Direct connection to printer for data and alarm printout ECOLOG TN4-L Part No. 2422 ECOLOG TN4-L for EX-Zone 1 Part No. 2422-EX Direct connection to printer for data and alarm printout Alarm output; 2 digital inputs 4 button keypad; External Start, Alarm Reset, Measurement and Alarm Scroll Direct connection to printer for data and alarm printout NTC Sensors page 5 ECOLOG TH1 with standard Sensor Part No. 2423 ECOLOG TH1 for EX-Zone 1 with standard sensor Part No. 2423-EX ECOLOG TH1-M with whit casing and standard sensor Part No. 2423-M Connection for 1 integrated (-35°C ..55°C; 0 ..100%rH) or up to 2 external humidity and temperature sensors -35°C ..55/110°C; 0..100%rH or a second temperature sensor -50°C ..140°C Calibrated, interchangeable humidity and temperature sensors (Part No. 3087; 3087-A; 3087-B) Alarm output; 1 digital inputs Calibrated, interchangeable humidity and temperature sensors (Part. No: 3087) Direct connection to printer for data and alarm printout TH1 Sensor Configuration page 7 ECOLOG TH2 Part No. 2426 ECOLOG TH2 for EX-Zone 1 Part No. 2426-EX For 2 external, calibrated and interchangeable humidity and temperature sensors -35°C ..70°C; -35°C..55/110°C; 0%..100%rH (Part No. 3087; 3087-A; 3087-B) Alarm output; 1 digital inputs 4 button keypad, External Start, Alarm Reset, Measurement and Alarm Scroll Direct connection to printer for data and alarm printout TH2 Sensor Configuration page 9 rH/T Sensor 3087-B page 9 ECOLOG TP2 Part No. 2425-2T ECOLOG TP2 for EX-Zone 1 Part No. 2425-2TEX ECOLOG TP4-L Part No. 2425 ECOLOG TP4-L for EX-Zone 1 Part No. 2425-EX TP2: For 2 PT100 sensors -200°C..550°C - 4 wire system with DB15 connector TP4-L: For 4 PT100 sensors -200°C..550°C - 4 wire system with LEMO connector Alarm output; 1 digital inputs 4 button keypad; External Start, Alarm Reset, Measurement and Alarm Scroll Direct connection to printer for data and alarm printout PT100 Sensors page 11 Accessories; Intrinsically safe page 12 Mounting Fixtures page 13 Alarm page 14 Accuracy; Traceability page 15 Calibration page 17 elproLOG WIN page 19

ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

  • Upload
    others

  • View
    19

  • Download
    0

Embed Size (px)

Citation preview

Page 1: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 1

ECOLOG Data SheetsStatus July 2002

Price list on demandSubject to alterations

ECOLOG TN2 Part No. 2420ECOLOG TN2 for EX-Zone 1 Part No. 2420-EX• For 2 NTC sensors; external: -50°C..140°C / internal: -35°C..55°C• External Start / Alarm Reset / InPos detection (with special connector)

ECOLOG TN3-P Part No. 2420-PECOLOG TN3-P for EX-Zone 1 Part No. 2420-PEX• For 3 NTC sensors; external: -50°C..140°C / internal: -35°C..55°C• Alarm output• External Start / Alarm Reset / InPos detection (with special connector)• Automatic printout for data and alarm protocols

ECOLOG TN4 Part No. 2421ECOLOG TN4 for EX-Zone 1 Part No. 2421-EX• For 4 NTC sensors -50°C..140°C• Alarm output; 2 digital inputs• 4 button keypad; External Start, Alarm Reset, Measurement and Alarm Scroll• Direct connection to printer for data and alarm printout

ECOLOG TN4-L Part No. 2422ECOLOG TN4-L for EX-Zone 1 Part No. 2422-EX• Direct connection to printer for data and alarm printout• Alarm output; 2 digital inputs• 4 button keypad; External Start, Alarm Reset, Measurement and Alarm Scroll• Direct connection to printer for data and alarm printout

NTC Sensors page 5

ECOLOG TH1 with standard Sensor Part No. 2423ECOLOG TH1 for EX-Zone 1 with standard sensor Part No. 2423-EXECOLOG TH1-M with whit casing and standard sensor Part No. 2423-M• Connection for 1 integrated (-35°C ..55°C; 0 ..100%rH) or up to 2 external humidity and

temperature sensors -35°C ..55/110°C; 0..100%rH or a second temperature sensor -50°C ..140°C• Calibrated, interchangeable humidity and temperature sensors (Part No. 3087; 3087-A; 3087-B)• Alarm output; 1 digital inputs• Calibrated, interchangeable humidity and temperature sensors (Part. No: 3087)• Direct connection to printer for data and alarm printout

TH1 Sensor Configuration page 7

ECOLOG TH2 Part No. 2426ECOLOG TH2 for EX-Zone 1 Part No. 2426-EX• For 2 external, calibrated and interchangeable humidity and temperature sensors -35°C ..70°C;

-35°C..55/110°C; 0%..100%rH (Part No. 3087; 3087-A; 3087-B)• Alarm output; 1 digital inputs• 4 button keypad, External Start, Alarm Reset, Measurement and Alarm Scroll• Direct connection to printer for data and alarm printout

TH2 Sensor Configuration page 9rH/T Sensor 3087-B page 9

ECOLOG TP2 Part No. 2425-2TECOLOG TP2 for EX-Zone 1 Part No. 2425-2TEXECOLOG TP4-L Part No. 2425ECOLOG TP4-L for EX-Zone 1 Part No. 2425-EX• TP2: For 2 PT100 sensors -200°C..550°C - 4 wire system with DB15 connector

TP4-L: For 4 PT100 sensors -200°C..550°C - 4 wire system with LEMO connector• Alarm output; 1 digital inputs• 4 button keypad; External Start, Alarm Reset, Measurement and Alarm Scroll• Direct connection to printer for data and alarm printout

PT100 Sensors page 11Accessories; Intrinsically safe page 12Mounting Fixtures page 13Alarm page 14Accuracy; Traceability page 15Calibration page 17elproLOG WIN page 19

Page 2: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 2

Technical DataGeneral: TN2: 2 channel datalogger with display and alarm indication on display

TN3-P: 3 channel datalogger with display and alarm functionsCase: Thermoplastic ABS, IP54 with internal sensor and cover on DB15, suitable for

foodstuff applications, 110x85x35mmDisplay: Large LCD display, visible down to -20°C, with alarm indicationMemory: 64’000 data points (128 kB)

Loop memory or start - stop mode with ext. start optionInterval: Programmable, 1 second to 3 hours,Log Period: Days, months, yearsAlarm: TN2: No external, but alarm display on LCD screen (programmable)

TN3-P: External on DB15, and alarm display on LCD screen (programmable)Operating: -35°C..55°C, display readable down to -20°CMeasuring: 1 built-in NTC sensor -35°C..55°C and/or up to 2/3 external NTC sensors -50°C..140°CBattery: 1x Lithium 3.6V, user-replaceable, life-span approx. 2 years

Low-battery warning for the remaining 6 months of operationEvaluation: PC software elproLOG WIN for all communication, reprogramming,

display, statistics and printout (fast data transmission RS232 with 38 400 Baud)Features: Start extern and InPos with DB15 start socket, display alarm reset with DB15 reset socket

TN2: No print functionTN3-P: Direct printout of short protocol (serial printer RS232 w. 9600 Baud)

Accessories: Part No.PC evaluation software 2338-CDVData cable PC 2318Simple fixation bracket 2804-AMounting bracket for DB15 2804-BBracket with screw terminals 2804-C..Seiko DPU414 protocol printer 2319Data cable for Seiko DPU414 2309-FScriptos protocol printer 2319-ST

NTC temperature sensors Page 5DB15 socket for sensor etc. 3032DB15 with screw terminals 3034

DB15 socket Start / Inpos 3032-ISDB15 socket Alarm reset 3032-EA

Simple Fixation Bracket 2804-A

NTC+Ref

UbatStart/Inpos

Wiring DiagrammeDB15 Connector

ECOLOG

DB15connectorforsensors,RS232and Alarmon TN3-P

8 GND 15 RxD7 TxD 14 Busy6 NTC2 13 Res.5 +Ref 12 Start/Inpos4 Res. 11 Ubat.3 Res. / NTC3 10 Alarm reset2 +Ref. 9 Res. / Alarm1 NTC1 Ubat

Alarm reset

Internal NTC Sensor

1212

123123

ECOLOG TN2ECOLOG TN3-PDatalogger Systemfor 1-3 Temperatures

Part No. 2420Part No. 2420-EXPart No. 2420-PPart No. 2420-PEX

Page 3: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 3

Technical DataGeneral: 4 channel datalogger with display and alarm functionsCase: Thermoplastic ABS, IP52 with ext. sensor, suitable for foodstuff applications, 110x85x35mmDisplay: Large LCD display, visible down to -20°C, with alarm indicationKey pad: 4-keys: reset alarm, step by step data or alarm display, printout data / alarmMemory: 64’000 data points (128 kB)

Loop memory or start - stop mode with ext. start by using the key padInterval: Programmable, 1 second to 3 hours,Log Period: Days, months, yearsAlarm: External on DB15, and alarm display on LCD screen (programmable)Operating: -35°C..55°C, display readable down to -20°CMeasuring: 4x NTC sensors -50°C..140°C, connected on DB15 connectorBattery: 1x Lithium 3.6V, user-replaceable, life-span approx. 2 years

Low-battery warning for the remaining 6 months of operationEvaluation: PC software elproLOG WIN for all communication, reprogramming,

display, statistics and printout (fast data transmission RS232 with 38 400 Baud)Printer: Direct printout of data, alarm information and status (serial printer RS232 with 9600 Baud)

ECOLOG

Accessories: Part No.PC evaluation software 2338-CDVData cable PC 2318Simple fixation bracket 2804-AMounting bracket for DB15 2804-BBracket with screw terminals 2804-C..Seiko DPU414 protocol printer 2319Data cable for Seiko DPU414 2309-FScriptos protocol printer 2319-ST

NTC temperature sensors Page 5DB15 socket for sensor etc. 3032DB15 with screw terminals 3034DB15 socket with built-in NTC 3032-A

DB15connectorforsensors,RS232,alarm

Key pad

Bracket 2804-B with Sensor/Alarm Cable

8 GND 15 RxD7 TxD 14 Busy6 NTC2 13 Res.5 +Ref 12 DigitIn14 NTC4 11 Ubat.3 NTC3 10 DigitIn22 +Ref. 9 Alarm1 NTC1

NTC+Ref

UbatDigitIn

Alarm

GND

Wiring DiagrammeDB15 Connector

+0..50VDCmax 200mA

ECOLOG TN4

Datalogger Systemfor 1-4 Temperatures

Part No. 2421Part No. 2421-EX

Page 4: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 4

Technical DataGeneral: 4 channel datalogger with display and alarm functionsCase: Thermoplastic ABS, IP52 with ext. sensor, suitable for foodstuff applications, 110x85x35mmDisplay: Large LCD display, visible down to -20°C, with alarm indicationKey pad: 4-key: reset alarm, step by step data or alarm display, printout data / alarmMemory: 64’000 data points (128 kB)

Loop memory or start - stop mode with ext. start by using the key padInterval: Programmable, 1 second to 3 hours,Log Period: Days, months, yearsAlarm: External on DB15, and alarm display on LCD screen (programmable)Operating: -35°C..55°C, display readable down to -20°CMeasuring: 4 connectors for NTC sensors -50°C..140°C, w. LEMO connectors 2 PinBattery: 1x Lithium 3.6V, user-replaceable, life-span approx. 2 years

Low-battery warning for the remaining 6 months of operationEvaluation: PC software elproLOG WIN for all communication, reprogramming,

display, statistics and printout (fast data transmission RS232 with 38 400 Baud)Printer: Direct print out of data, alarm information and status (serial printer RS232 w. 9600 Baud)

Accessories: Part No.PC evaluation software 2338-CDVData cable PC 2318Simple fixation bracket 2804-AMounting bracket for DB15 2804-BBracket with screw terminals 2804-C..Seiko DPU414 protocol printer 2319Data cable for Seiko DPU414 2309-FScriptos protocol printer 2319-ST

NTC temperature sensors Page 5DB15 socket for alarm etc. 3032DB15 with screw terminals 3034

DB15connectorforsensors,RS232,alarm

Bracket 2804-B with Alarm Cable

ECOLOG

4x LEMO connector

Key pad

8 GND 15 RxD7 TxD 14 Busy6 NTC2 13 Res.5 +Ref 12 DigitIn14 NTC4 11 Ubat.3 NTC3 10 DigitIn22 +Ref. 9 Alarm1 NTC1

NTC+Ref

UbatDigitIn

Alarm

GND

Wiring DiagrammeDB15 Connector

+0..50VDCmax 200mA

ECOLOG TN4-LDatalogger Systemfor 1-4 Temperatureswith LEMO plug 2 pin

Part No. 2422Part No. 2422-EX

Page 5: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 5

NTC Temperature SensorsOrder Example

Part No. Length Diameter T.-Range Cable material; T maxmm mm

3010-.. LS=100 D=4 -50°C..140°C Silicon; 180°C3011-.. LS=25 D=4 -50°C..140°C Silicon; 180°C3090-.. LS=50 D=6 -50°C..140°C Silicon; 180°C3091-.. LS=150 D=6 -50°C..140°C Silicon; 180°C

3095-.. LS=70 D=3.4 -50°C..140°C Silicon; 180°C

3013-.. LS=15 D=2 -50°C..140°C Silicon; 180°C3020-.. LS=100 D=4 -50°C..140°C Silicon; 180°C

3021-.. LS=100 D=4 -50°C..140°C Silicon; 180°C

3024-.. LS=150 D=5 -50°C..140°C Silicon; 180°C

3041-.. LS=30 D=6 -50°C..140°C Silicon; 180°C

3050-.. LS=22 D=12 -50°C..140°C Silicon; 180°C

3055-.. D=2.5 -50°C..140°C Silicon; 180°C3056-.. D=1 -50°C..140°C Silicon; 180°C

3060 LS=40 D=10 -50°C..140°C Protection tube

3093-.. LS=20 D=5 -40°C..105°C Thermopl. elastomerflat cable; 105°C

3094-.. LS=20 D=5 -40°C..105°C Thermopl. elastomerround; 105°C

Ordering example - 4 transporters: bracket each with 2 sensors; printer and PC evaluation4x 2420-P ECOLOG TN3-P dataloggers4x 3090-L05 Temperature sensors; cable 5 meters4x 3090-L10 Temperature sensors; cable 10 meters4x 2804-B Mounting brackets for loggers and DB154x 3032-B DB15 connectors - already mounted1x 2319-ST Scriptos printer1x 2318 PC data cable1x 2338-CDV PC evaluation software

The sensors can be mounted at the following socket connectors:DB15 connector for sensor / alarm sensor connection by customer Part No. 3032

Part No. 3034

DB15 for sensors, with 1xNTC internal sensor -35°C..55°C Part No. 3032-ADB15 for sensors, with up to 4 connection cables already mounted by ELPRO Part No. 3032-B

All sensors are available with the 2 pol LEMO connector also. The modified part number is as following: yyyy-SxOrdering information: Connector type, ECOLOG type and state the sensor with its relevant channel

Page 6: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 6

Technical DataGeneral: 4 channel (2 x Humidity and 2 x Temperature) datalogger with display and alarm functionsCase: Thermoplastic ABS, IP50, suitable for foodstuff applications, 110x85x35mmDisplay: Large LCD display, visible down to -20°C, with alarm indicationKey pad: 4-key: reset alarm, step by step data or alarm display, printout data / alarmMemory: 64’000 data points (128 kB)

Loop memory or start - stop mode with ext. start by using the key padInterval: Programmable, 1 second to 3 hours,Log Period: Days, months, yearsAlarm: External on DB15, and alarm display on LCD screen (programmable)Operating: -35°C..55°C, display readable down to -20°C; 0%..100%rH, with condensationReaction Constant: Temperature: 110s; Humidity: 110s

Logger with sensor, standard dust filter, air speed: 1m/sMeasuring: - 3087 or 3087-A: Integrated or up to 2 external temperature and humidity sensors:

T: -35°C..70°C, H: 0%..100%rH- 3087-B: 1 or 2 external temperature and humidity sensors: T: -35°C..110°C#, H: 0%..100%rH

# 110°C for temperature peeks, permanent operation 100°C only- Up to 2 external temperature sensors: T: -50°C..140°C

Battery: 1x Lithium 3.6V, user-replaceable, life-span approx. 2 yearsLow-battery warning for the remaining 6 months of operation

Evaluation: PC software elproLOG WIN for all communication, reprogramming,display, statistics and printout (fast data transmission RS232 with 38 400 Baud)

Printer: Direct printout of data, alarm information and status (serial printer RS232 with 9600 Baud)

DB15 Connector on ECOLOG

8 GND 15 RxD7 TxD 14 Busy6 +Ref. 13 *)5 *) 12 NTC24 *) 11 *)3 *) 10 DigitIn2 *) 9 Alarm1 *)

Interchangeablehumidity andtemperaturesensor

NTC2+Ref

+RefDigitIn

0..50VDCmax 200mA

Wiring Diagramme

+

*) For a second rH/T sensor usebracket type 2805-CR

Alarm

GND

ECOLOG TH1Datalogger Systemfor Temperature andHumidity

Part No. 2423Part No. 2423-EX

Accessories: Part No.PC evaluation software 2338-CDVData cable PC 2318Simple fixation bracket 2804-AMounting bracket for 3215-Sx 2804-BBracket with 3 x DB15 sockets 2805-CRSeiko DPU414 protocol printer 2319Data cable for Seiko DPU414 2309-FScriptos protocol printer 2319-STDB15 socket for alarm etc. 3032DB15 with screw terminals 3034Humidity-temperature sensor 3087Humidity-temperature sensor (replacement) 3087-AHumidity-temperature sensor (high temp.) 3087-BHumidity calibration set 2812-BExtension cable 1, 2, 5, 10m 3215-SxxConnection lead for two rH/T sensors 2, 5m 3215-DxxAdapter sensor 1/2 to 3/4 3215-DXAdapter for 2 temperature sensors 3215-VNNTC temperature sensors Page 5

DB15connectorforsensors,RS232,alarm

Key pad

ECOLOG

Page 7: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 7

Sensor ConfigurationECOLOG TH1

Important Note:Max. added cable length for rH/T sensor1 and 2 is 20m(e.g. 1x3215-S20 or 2x3215-S10)

3215-S1

3087

3215-Sxxxx = 1; 2; 5; 10; 15; 20m

3215-DX

3087 3215-Dxxxx = 2m; 5m

3215-B

2804-B

2805-CR

3087

3215-Sxxxx = 1; 2; 5; 10; 15; 20m

3215-VN

any NTCtemperaturesensor withLEMO plug

3034

any NTCtemperaturesensor

any NTCtemperature sensor

ECOLOG TH1 with 1internal rH/T sensor

ECOLOG TH1Calibration

ECOLOG TH1 with asecond temperaturesensor and bracket2804-B

ECOLOG TH1 with asecond temperaturesensor and connector3034

Configuration Function

Internal:Humidity andtemperature

Internal:Humidity andtemperatureExternal:Temperature

External:Humidity andtemperature

Internal:Humidity andtemperatureExternal:Temperature

ECOLOG TH1 with asecond rH/T sensorand bracket 2805-CR

Internal:Humidity andtemperatureExternal:Humidity andtemperature

ECOLOG TH1 with asecond rH/T sensorand adapter 3215-DX

Internal:Humidity andtemperatureExternal:Humidity andtemperature

External:Two times humidityand temperature

External:Two timestemperature

ECOLOG TH1 withtwo rH/T sensors andconnection lead 3215-Dxx

ECOLOG TH1 withtwo temperaturesensor and adapter3215-VN

Page 8: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 8

Technical DataGeneral: 4 channel (2 x Humidity and 2 x Temperature) datalogger with display and alarm functionsCase: Thermoplastic ABS, IP52 with ext. sensor, suitable for foodstuff applications, 110x85x35mmDisplay: Large LCD display, visible down to -20°C, with alarm indicationKey pad: 4-key: reset alarm, step by step data or alarm display, printout data / alarmMemory: 64’000 data points (128kB)

Loop memory or start - stop mode with ext. start by using the key padInterval: Programmable, 1 second to 3 hours,Log Period: Days, months, yearsAlarm: External on DB15, and alarm display on LCD screen (programmable)Operating: -35°C..55°C, display readable down to -20°C

0%..100%rH, with condensationMeasuring: - 3087 or 3087-A: 1 or 2 external temperature and humidity sensors: T: -35°C..70°C , H: 0%..100%rH

- 3087-B: 1 or 2 external temperature and humidity sensors: T: -35°C..110°C#, H: 0%..100%rH # 110°C for temperature peeks, permanent operation 100°C only

- Up to 2 external temperature sensors: T: -50°C..140°CBattery: 1x Lithium 3.6V, user-replaceable, life-span approx. 2 years

Low-battery warning for the remaining 6 months of operationEvaluation: PC software elproLOG WIN for all communication, reprogramming,

display, statistics and printout (fast data transmission RS232 with 38 400 Baud)Printer: Direct printout of data, alarm information and status (serial printer RS232 w. 9600 Baud)

DB15connectorforsensors,RS232,alarm

+RefDigitIn

Alarm

GND

Wiring DiagrammeDB15 Connector

Key pad

8 GND 15 RXD7 TXD 14 Busy6 +Ref. 13 NTC15 A1 12 NTC24 A2 11 D13 B1,2 10 DigitIn2 D2 9 Alarm1 C1,2

0..50VDCmax 200mA

+

For a second rH/T sensor, alarm etc. theuse of bracket 2805-CR is recommended

ECOLOG TH2Datalogger Systemfor Temperature andHumidity

Part No. 2426Part No. 2426-EX

Accessories: Part No.PC evaluation software 2338-CDVData cable PC 2318Simple fixation bracket 2804-AMounting bracket for 3215-Sx 2804-BBracket with 3 x DB15 sockets 2805-CRSeiko DPU414 protocol printer 2319Data cable for Seiko DPU414 2309-FScriptos protocol printer 2319-STDB15 socket for alarm etc. 3032DB15 with screw terminals 3034Humidity-temperature sensor 3087Humidity-temperature sensor (replacement) 3087-AHumidity-temperature sensor (high temp.) 3087-BHumidity calibration set 2812-BExtension cable 1, 2, 5, 10m 3215-SxxConnection lead for two rH/T sensors 2, 5m 3215-DxxAdapter sensor 1/2 to 3/4 3215-DXAdapter for 2 temperature sensors 3215-VNNTC temperature sensors Page 5D 15mm

30873215-Sx

Page 9: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 9

Sensor ConfigurationECOLOG TH2rH/T Sensor 3087-B

Important Note:Max. added cable length for rH/T sensor1 and 2 is 20m (e.g. 1x3215-S20 or 2x3215-S10)

Function

ECOLOG TH2 asportable rH/T datalogger

ECOLOG TH2 with 1rH/T sensor andbracket 2804-B

ECOLOG TH2 with arH/T sensor, secondtemperature sensor andbracket 2805-CR

3215-B

Configuration

3215-Sxxxx = 1; 2; 5; 10; 15; 20m

30872804-B

2805-CR

3087 3215-Dxxxx = 2m; 5m

External:Temperature3034

any NTC temperaturesensor

3215-VN

any NTCtemperaturesensor withLEMO plug

ECOLOG TH2 withtemperature sensorand connector 3034

External:Two times humidityand temperature

External:Two timestemperature

ECOLOG TH2 withtwo rH/T sensors andconnection lead3215-Dxx

ECOLOG TH2 withtwo temperaturesensor and adapter3215-VN

ECOLOG TH2 withtwo rH/T sensors andbracket 2805-CR

External:Two times humidityand temperature

External:Humidity andtemperature

External:Humidity andtemperature andsecond temperature

External:Humidity andtemperature

Save Operation Area of rH/T sensor 3087-B; Sensor with extended temperatur range up to 100°C

Save operation area of the rH/T sensor 3087-B for extended temperatureapplication corresponds to the plot shown.110°C for temperature peeks, permanent operation 100°C only.

AttentionTemperature range for the sensor cable 3215-Sxx is -35°C ..80°CThe cable should not be moved at such high temperatures!

Sensor 3087-B temperature dependence of the measurement errorat the time of shipment.For sensor 3087 and 3087-A reduced temperature range:-35°C..70°C only

Page 10: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 10

Bracket 2804-B with Alarm Cable

Technical DataGeneral: TP2: 2 channel datalogger with display and alarm functions

TP4-L: 4 channel datalogger with display and alarm functionsCase: Thermoplastic ABS, IP52 with ext. sensor, suitable for foodstuff applications, 110x85x35mmDisplay: Large LCD display, visible down to -20°C, with alarm indicationKey pad: 4-key: reset alarm, step by step data or alarm display, printout data / alarmMemory: 64’000 data points (128 kB)

Loop memory or start - stop mode with ext. start by using the key padInterval: Programmable, 1 second to 3 hoursResolution: High and low resolution selectable. Attention: Display in low resolution mode always!Log Period: Days, months, yearsAlarm: External on DB15, and alarm display on LCD screen (programmable)Operating: -35°C..55°C, display readable down to -20°CMeasuring: 2/4 x PT100 sensors -200°C..550°C, 4 wire systemSensor connection:TP2: DB15 connector

TP4-L: 4 LEMO connectors 4 Pin or sensor 1 & 2 on DB15 connectorBattery: 1x Lithium 3.6V, user-replaceable, life-span approx. 1.5 years, depending on measurement

interval and resolution. Low-battery warning for the remaining 6 months of operationEvaluation: PC software elproLOG WIN for all communication, reprogramming,

display, statistics and printout (fast data transmission RS232 with 38 400 Baud)Printer: Direct printout of data, alarm information and status (serial printer RS232 with 9600 Baud)

Accessories: Part No.PC evaluation software 2338-CDVData cable PC 2318Simple fixation bracket 2804-AMounting bracket for DB15 2804-BSeiko DPU414 protocol printer 2319Data cable for Seiko DPU414 2309-FScriptos protocol printer 2319-ST

PT100 temperature sensors see separate listDB15 socket for alarm etc. 3032DB15 with screw terminals 3034

DB15connectorforsensors,RS232,alarm ECOLOG

4x LEMO connector

Key pad

8 GND 15 RxD7 TxD 14 Busy6 Ubat 13 A25 A1 12 A24 A1 11 B23 B1 10 DigitIn2 B1 9 Alarm1 B2

Alarm

GND

Wiring Diagramme

UbatDigitIn

PTAPTAPTBPTB

LEMO Connector

A

A B

B

Plug solderingside seen

0..50VDCmax 200mA

+

DB15 Connector

ECOLOG TP2ECOLOG TP4-LDatalogger Systemfor 1-4 PT100 Sensors

Part No. 2425-2TPart No. 2425-2TEXPart No. 2425Part No. 2425-EX

Page 11: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 11

PT100 Temperature SensorsOrder Example

Part No. Length Diameter T.-Range Cable material; T max.mm mm

3159-.. 75x20 0°C..180°C Teflon; 220°C

3160 LS=30 D=10 -35°C..55°C Protection tube

3161-.. LS=200 D=6 -80°C..260°C Teflon; 220°C3163-.. LS=100 D=4 -90°C..260°C Teflon; 220°C

3162-.. LS=200 D=6 -80°C..500°C Glasfibre; 400°C3169-.. LS=50 D=6 -80°C..500°C Glasfibre; 400°C

3164-.. LS=100 D=4 -50°C..260°C Teflon; 220°C

3165-.. LS=350 D=3 -200°C..500°C Silicon; 180/90°C#

3166-.. LS=250 D=3 -80°C..500°C Silicon; 180/90°C#

3172-.. LS=750 D=3 -200°C..500°C Silicon; 180/90°C#

3167-.. LS=20 D=5 -80°C..105°C Silicon; 180°C

3168-.. LS=120 D=4 -50°C..180°C Silicon; 180°C

3171-.. LS=35 D=6 -50°C..250°C Teflon; 220°C

3173-.. LS=24 D=3 -50°..120°C Silicon; 180°C

3174-.. LS=50 D=2 -80°C..350°C Teflon; 220°C

3181-.. LS=100 D=5 -200°..180°C Teflon; 220°C

# Temperature maximum for sensor connector is 90°C

If not other specified the accuracy of all PT100 sensors from ELPRO-BUCHS AG comply to DIN 1/3-B.

The sensors can be mounted at the following socket connectors:DB15 connector for sensor / alarm sensor connection by customer Part No. 3032

Part No. 3034

DB15 for sensors, with up to 2 connection cables already mounted by ELPRO Part No. 3032-B

All sensors are available with the 4 pol LEMO connector also. The modified part number is as following: yyyy-SxOrdering information: Connector type, ECOLOG type and state the sensor with its relevant channel

LS

D

Ordering example - 2 fridges at -45°C: bracket; 2 sensors; printer and PC evaluation2x 2425 ECOLOG TP4-L dataloggers4x 3167-S02 Temperature sensors; cable 2 meters; LEMO connector2x 2804-E Mounting brackets for loggers, secured with pad lock1x 2319 Seiko DPU414 printer1x 2309-F Data cable for printer1x 2318 PC data cable1x 2338-CDV PC evaluation software

Page 12: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 12

Seiko DPU414 printer Part No. 2319- For 220V and battery operation, 110mm thermopaper- Printer cable Part No. 2309-F- Printer paper (3 rolls) Part No. 2311

AccessoriesIntrinsically SafeECOLOG Dataloggers

Intrinsically Safe Logger; Part No. xxxx-EX

II 2 G I I Equipment group II intended for use in all potentially explosiveatmospheres apart from mines

2 Category 2, suitable for use in zone 1 (occasional explosionhazard) as well as in zone 2 (rare explosion hazard)

G Atmosphere with explosion hazard arising from gases and vaporsbut not from dust

EEx ib IIB T4 EEx Explosion protection type according to European directives:EN 50014: 1997 +A1 +A2 and the directives for specialtypes of protection against ignition

ib Type of protection for intrinsic safety against ignition: category ibwith 1 failure according to EN 50020:1994

IIB Use in all potentially explosive atmospheres apart from mines:group II sub-clause B

T4 Temperature class T4: max. surface temperature 135°C with asafety margin of 5 kelvin for permanently hot surfacesT4 applies for compound materials with an ignition temperature oft > 135°C, essentially ethyl ether and ethanal, which are usedfor industrial production of synthetics and solvents.

Anschlussdaten des eigensicheren Steckers siehe Manual EZ-7003BSecurité intrinsèque interface : voir manual EZ-7003B

For el. data of the intrinsically safe interface see manual EZ-7003B

SEE 00 ATEX 3115II 2 GEEx ib IIB T4

0499

Scriptos printer Part No. 2319-ST- Multivolt for operation in transporters 12V and 24V- Inclusive printer cable- Printer paper (3 rolls) Part No. 2311

Replacement battery for ECOLOG data loggers Part No. 2820Set of 2 batteries, minimum storage time is 5 years

USB - RS232 Adapter Part No. 2317-USB- To connect any ECOLOG data loggers to a PC by using the USB port of the PC.

Page 13: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 13

Mounting FixturesECOLOG Dataloggers

DB15 connector Part No. 3032Sensor with customized mounting Part No. 3032-B2-3 cables can be attachedMetal housing with connector head - solderhook terminated for sensor, alarm output, etc.

DB15 connector with sensor Part No. 3032-AFor ECOLOG TN4 as internal sensor, analog 3032Operating: -35°C..55°C

Fixation bracket ECOLOG Part No. 2804-Bfor DB15Ideal for ECOLOG TN2, TN4 and TH2Made of stainless steel for wall-mountingWith mounting bracket to attach DB15 connectorWith 2-3 cables, without DB15 connectorThe logger can be protected and secured with a padlock (not part of delivery)

Protective Housing ECOLOG Part No. 2350-xxAs accessories ELPRO provides a protective housing made of shock proofplastic material with IP66, and 3 differenterent brackets for simple fixation ofdataloggers.For more information see specific data sheets.

Fixation bracket ECOLOG Part No. 280x-C..with terminalsIdeal for customer applicationsMade of stainless steel for wall-mountingFor simple attachment of all sensor cables, digital inputs and alarm cablesto the connecting terminalsThe logger can be protected and secured with a padlock (not part of delivery)

Logger Type Bracket TypeTNx: 2804-C

2804-CR (with additional RS232 connector)THx: 2805-CRTPx: 2801-CR (with additional RS232 connector)

Simple fixation bracket Part No. 2804-AIdeal for ECOLOG TN2 and TH1Support plate made of stainless steelWith 2 PT screws for attachment to logger

Fixation bracket with padlock Part No. 2804-EIdeal for ECOLOG TN2 and TH1Support plate made of stainless steelThe logger can be protected and secured with a padlock (not part of delivery)

2804-C.../2801-CR

2805-CR

RS 232

Sen

sor

1/2

Sen

sor

3/4

Page 14: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 14

AlarmECOLOG Datalogger

Beacon with Sound Part.-No 2311-C- Bracket 2804C/CR for ECOLOG TNx- Bracket 2801C/CR for ECOLOG TPx

Alarm with e.g. auto dialler(contact input) for ECOLOG TNx orTPx

1234512345123451234512345123451234512345123451234512345123451234512345

1234512345123451234512345123451234512345

1234123412341234123412341234123412341234123412341234123412341234123412341234123412341234123412341234123412341234

3456789

10

n.c. 12

11121314

+ 12VGnd (-)

Alarm

+

12VDC500mAPower

Supply Unit

=

EC

OLO

Gb

row

n

Bea

con

= w

hite

Bea

con

+ s

ound

= g

reen

EC

OLO

G

1234512345123451234512345123451234512345123451234512345123451234512345

1234512345123451234512345123451234512345

12

3456

+ 12VGnd (-)

Alarm

12VDC500mAPower

Supply Unit

=+

bro

wn

Bea

con

= w

hite

Bea

con

+ s

ound

= g

reen

Beacon with Sound Part.-No 2311-CBracket 2805CR for ECOLOG THx

2

6

-

+ = -+12VDC

2

14

-

+ = -+12VDC

Alarm with e.g. auto dialler(contact input) for ECOLOG THx

Page 15: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 15

Temperature Measurement TNx and THx(Datalogger with sensor typical; logger at room temperature)Operating Range Resolution Accuracy

-50°C.. -25°C 0.1°C ± 0.4°C-25°C.. 0°C 0.1°C ± 0.3°C

0°C.. 70°C 0.1°C ± 0.2°C70°C.. 100°C 0.1°C ± 0.4°C

100°C.. 140°C 0.1°C ± 0.7°C

Temperature Measurement TPx(Datalogger only, at room temperature)Operating Range Resolution Linearity-200°C.. 100°C 0.2°C ± 0.3°C-100°C.. 400°C 0.1°C ± 0.2°C 400°C.. 500°C 0.1°C ± 0.3°C 500°C.. 550°C 0.2°C ± 0.5°C

Accuracy; TraceabilityTemperature

TraceabilityELPRO uses calibrated measuring units for factory calibration. The ELPRO certificate can be used for GLP applications.The following calibrated normals are implemented for the calibration procedure:Voltage / current calibration source GENERAL-RESISTANCE DAS-57ALResistance reference CROPICO type RBB5100 Ohm reference Vishay type VHA 414KTemperature reference 0°C ice water SCS enclosed-scale thermometer 0°CPT100 precision temperature sensor ASL Y1266, ITS-90 calibration EAMPrecision resistance bridge HART 1502ACalibration bath -20°C... 150°C ASL LR100 & TAMSON TV2000

NormsEN12830 Temperature recording instrument for transport, storage and distribution of foodstuffsGZ1480 Exceptional approval for calibration GZ1480/2000 from 10. 4. 2000, BEV AustriaEX Approval for intrinsically safe area 1; EN 50014 and EN 50020FDA Software validation for GLP applicationCE The loggers are conform to EN 50081-2 and EN 50082-2

Check / Verification of Temperature Measurement1) New devices:- All dataloggers are factory-checked using precision

resistors and subsequently receive a calibration certificate.- NTC resistor sensors (thermal resistor) are interchange-

able in terms of accuracy - see the adjacent table.- PT100 sensors are interchangeable with respect to their

class of accuracy.2) Periodical recalibration:Datalogger: every 2 years - with calibration resistor by end useror by ELPRO service center.Sensor: every 2 years or when deviations occur - in calibrationbath by end user or by ELPRO service center.

Norms passedNorms passedNorms passedNorms passedNorms passed

Accuracy of Temperature SensorsTemperature Deviation °C

NTC PT100, IEC751, A-200°C --- ±0.55-100°C --- ±0.35-50°C ±0.3 ±0.25-25°C ±0.2 ±0.20°C ±0.1 ±0.1525°C ±0.1 ±0.250°C ±0.1 ±0.2570°C ±0.1 ±0.3100°C ±0.4 ±0.35125°C ±0.6 ±0.4200°C --- ±0.55300°C --- ±0.75400°C --- ±0.95500°C --- ±1.15

Page 16: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 16

Relative Air Humidity ECOLOG THxOperating range Resolution Accuracy of measurement0% rH .. 100%rH 0.2%rH At ambient temperature, 23°C: ± 1.5%rH

Hysteresis 10-90-10%rH: <1%rHTemperature coefficient: see page 9

Check / Verificaton of the Relative Humidity Measurement1)New devices:All dataloggers are factory-calibrated with SCS calibration solutions and subsequently receive a calibration certificate. Theadjustment points are 0%rH and 80%rH (95%rH for high levels of humidity). The humidity sensors are calibrated andinterchangeable. The calibration values are read in by the logger.2) Periodical recalibration:With SCS calibration solutions and calibration device by the end user or by ELPRO service center.Alternatively there is the possibility to get a calibrated sensor as an interchangeable part from ELPRO.Interval: every 12 months in clean operating environment; in environment with high humidity, dust, smoke etc,every 6 months or in case of doubt.

NB: SCS = SWISS CALIBRATION SERVICE

Interchangeability of rH/T Sensorsa) Humidity Sensor used in rH/T Sensors

All humidity sensors are factory-calibrated with SCS calibration solutions and subsequently receive a calibrationcertificate. The adjustment points are 0%rH and 80%rH. (95%rH for high levels of humidity)The ECOLOG rH/T sensors are interchangeable in their pre-calibrated state. The calibration data are read in by thelogger.

b) Temperature Sensor used in rH/T SensorsFor the temperature sensor used in the rH/T sensor are the same conditions valid as for the NTC sensors used.Based on the strong relation between temperature and measurement value, in most of our cases a check at 0°C ice-water is sufficient.

c) Data LoggerFor the production of our data loggers we use high quality components only. The functionality of all loggers ischecked by the use of high precision resistors for the temperature measurement and with a simulated signal for thehumidity measurement. According to these checks all data loggers receive a calibration certificate.

Check / Verification of rH/T Sensorsa) Temperature measurement

According to the information about temperature measurement, see page 15.b) Humidity measurement

With SCS calibration solutions and calibration device by the end user or by ELPRO service center.Interval: In clean operating environment every 12 month, in environments with high humidity, dust, smoke etc. every 6month or in case of doubt.

Required for humidity calibration and adjustment: Part No.Extension cable for ECOLOG THx for calibration: 3215-S01Calibraton unit for humidity logger: 2902Calibration ampullae (set of 5 ) 2901-Hx, x= 0%, 35%, 50%, 80%, 95%

TraceabilityELPRO uses SCS calibrated humidity standards for calibration. The ELPRO certificate can be used for GLP applications.

Accuracy; TraceabilityHumidity

Page 17: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 17

Methods for Temperatur Calibrationa) 0°C ice-waterCalibration of modules with their sensors which uses the triplepoint of ice-water (0°C) as reference temperature. You canexpect an accuracy of approx. 0°C ±0.1°C.b) Calibration bathWhen a calibration bath is used (-20°C..200°C), pay attentionto the fact that the reference sensor should be fastened to thecomparison sensor. This will ensure the temperature at theindividual sensors is identical, i.e. that there is no temperaturedifference. All the sensors should be immersed to the samedepth in the bath and the implemented reference bath shouldbe stable. Also make sure that the sensors do reach thereference temperature. The measured values can beimproved by repeated measuring and averaging.

c) Dry calibratorYou should only calibrate sensors in the dry calibrator whenthe diameter of the aperture in the test block correspondsexactly with the diameter of the sensor and when the sensorcan be inserted deep into the test block (min. 100 mm ordeeper, depending on the sensor).

Ice water calibration procedure1. Prepare the ice water in an insulated container, e.g. a

camping cool-box:Fill the cool-box with 10 liters of ice cubes. Use ice froman ice machine (-1°C), not from a freezer (-20°C). Fill upthe cool-box with cold water to the filling height of the ice.Stir the contents of the cool-box thoroughly to mix ice andcold water.

2. Program a short recording interval (1 min).

3. a) Wrap up the logger with internal sensor in watertight packaging material, i.e. pack it in a latex glove.b) Plug in external temperature sensors at the module.c) Use the extension cable to connect the rH/T sensors for THx modules to the module and wrap up the sensor in watertight packaging material, e.g. in a latex glove.

ECOLOG TNx , THx and TP4 Temperature - Calibration or AdjustmentECOLOG TNx and THx: Modules for measuring temperature with precise NTC sensorsDataloggers belonging to the TNx and THx series are supplied with very precise temperature sensors. Consequently, it ispossible to dispense with adjustment procedures. However we recommend that you perform an operational check on themodule and its temperature sensors approx. every 12 months. If you detect a deviation from the permissible range, thereis a defect at the sensor, cable or connector. The cause of this defect must be eliminated.

ECOLOG TPx: Modules for measuring temperature with PT100 sensorsDataloggers belonging to the TPx series are factory-adjusted with precision resistors to the theoretical PT100 characteristic.The implemented 4-line measuring technique automatically compensates the measuring cable influence. In exceptionalcases, it is possible to readjust PT100 sensor deviation, i.e. when very exact measurements must be made in one specialoperating point. However, an readjusted module must be marked as follows: Only to be used with sensor XYZ. We recommendthat you perform an operational check on the module and its temperature sensors approx. every 12 months.

NOTESensor interchangeability can no longer be guaranteed if thismethod is used. A possible offset at the used sensor will betaken into account of during the calibration procedure. Forthis reason, only use the above method in exceptional cases.

4. Submerge the module / sensor completely in the ice waterand wait 2 h. If you are testing combined rH/T sensors,part (min. 0.5m) of the cable must be submerged alongwith the sensor in the ice water otherwise the requiredmeasuring accuracy will not be attained.

5. Calibration: compare setpoint and actual value

a) The precision of PT100 sensors will depend on your quality class (A or B). Measured value deviation should not exceed the following:

PT100 A : -100°C..100°C ±0.35°CPT100 B : -100°C..100°C ±0.8°C

b) Deviation for the NTC should not exceed the following:

NTC : -20°C..50°C ±0.2°CIf deviation is greater, conditions during measurementwere not stable enough or the module / sensor has adefect (cable, connector, etc.).

6. Adjustment: Only readjust the module (TPx) when the levelof deviation is too high and when the module or the sensordoes not have a defect (cable, connector, etc.).

Connect the test plug or the decade resistor to the module.Adjust the decade resistor in accordance with a PT100standards table.In menu item “Extended Setup”, select single-pointadjustment for the relevant adjustment point and the sensor.Enter the temperature for the adjustment procedure andconfirm your entries by activation the OK button.Logger adjustment requires two temperature points. Toguarantee high-quality adjustment, both of theseadjustment points should be outside the logger’s normaloperativerange. Use the status printout to check thesuccess of the adjustment.

NOTEPrior to further operation, reprogram the originaltemperature and time ranges.

Dev

iatio

n

0°C

2 hEvaluate the module dataand make a record of themeasured deviationvalues in your calibraitonlogbook.

CalibrationTemperature

Page 18: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 18

CalibrationHumidity

ECOLOG THx Humidity - Calibration or AdjustmentModules for measuring relative air humidityAt delivery, each of our humidity dataloggers is fitted with a precisely calibrated humidity sensor. We recommend that humiditysensors used in normal working environments are calibrated every 6 to 12 months. If necessary, they should also be readjusted.In particuarly contaminated environments, it is necessary to clean the rH sensor very carefully using water or a solution with max.40% alcohol.

MethodUse our SCS-certified humidity calibration solutions foroperating points with relative humidity values of 0%, 5%, 10%,20%, 35%, 50%, 65%, 80% and 95%.Verification is more important than adjustment! The calibrationsolutions and the calibration device can be used to simulatea variety of humidity levels at the sensor. Single-pointadjustment of the upper adjustment point with the 80%calibration solution is suitable for the majority of applications.The following material and equipment are required:- a calibration device (Part No. 2902)- calibration solution of 0% .. 95% (Part No. 2901-Hxx)- extension cable 1m (Part No. 3215-S01)- calibration case or thermally isolated boxor calibration set for 0% and 80% (Part No. 2812-B)and elproLOG WIN evaluation software

Procedure

WARNING- do not clean the sensor with pure benzene or acetone- do not pack the sensor in plastic bags- do not touch the sensor with your hands

Calibration procedurea) Return the sensor to ELPRO-BUCHS AG.b) Replace the sensor with a calibrated sensor.c) Calibration at customer-site performed by the

ELPRO-BUCHS AG calibration service (Switzerland only).d) Calibration in accordance with the following instructions.

1. Clean the calibration devicethoroughly and dry it in an airflow

IMPORTANTThe calibration device must be absolutely clean and drybefore it is used. Wash the disassembled device and dry itthoroughly in an air flow.

Insert the supplied fabric disc (except at0%) in the device base plate. Break to openthe ampoule containing the calibrationsolution and distribute its contents ontothe fabric

4.

Calibration: compare setpoint and actual valueUnplug the extension cable, leavethe sensor with calibration device inthe calibration case and evaluatethe module data. The module mustbe adjusted if the measured values areoutside the max. permissibledeviation range.

The temperature in the calibrtion case should be atapprox. 20 °C (ambient temperature).

6.

7.

Adjustment: Only readjust the module data when the level ofdeviation is too high.In menu item “Extended Setup”, select single-pointadjustment for the relevant adjustment point with theimplemented calibration solution and the sensor and reducethe wait time to 15 min. Return the module to the sensor inthe calibration case and plug the extension cable back intothe module.When the above wait time has elapsed, disconnect themodule and sensor from the extension cable and makeanother status printout (Status B). This status printoutindicates whether the adjustment procedure has been carriedout successfully. If the adjustment has not been a success,you can use menu item “Direct Calibration...” to reenter theinitial adjustment data contained in the original statusprintout (Status A) or to repeat the adjustment procedure.

8.

9.

Set-point

3 h

Place the logger for 3 h in an insulated container or in thecalibration case.

Dev

iatio

n

Slide the calibration device (screwed-onbase plate facing downwards) onto thesensor and tighten the sealing ring.

5.

IMPORTANTFor the adjustment of a sensor, connected one sensor to thelogger only!

NOTE- Prior to further operation, reprogram the original

temperature and time ranges.- Measurement is stable after 10 seconds.

Program a short recording interval (1 min) and make astatus printout (Status A).Turn the black ring, in the direction of the arrow, to its limitstop position and pull the sensor out of the coupling.Wait until the display shows “n.c. %rH” and then use theextension cable to connect the sensor to the module.

2.

3.

Module

Extension cable

Calibration device

Cal ibrat ioncase

Sensor

When “n.c. %rH” appears in the display, plug in the sensorat the module and secure it by turning the ring in the oppositedirection to the arrow until it reaches its limit stop position.The logger is operable again when “Ld. %rH” appears onthe display.

Page 19: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 19

Evaluation SoftwareelproLOG WIN32

Also available:Software Validation Workbook Part No. 2336Copy of Software Validation Protocol Part No. 2337

These documents comply to: DIN EN 12820, EG92/1, FDA Reg. 21CFR Part 11, 58, 820

PC Requirements Hardware PC with pentium/133Software Windows 95, 98, Me, NT4 or 2000

Part No

Items required for evaluation Software Full version 2338-CDVUpdate 2338-CDU

Data cable Used to interconnect the logger to the pc serial port 2318

Screen ShotsON/OFF Statistics withadditional Information

Automatic scaling andblock time zoom

Autoscale for time anddate

Info line with data at cursor

Filen name and date/timeof reading

Info windows, toolboxes and sub-menues- dew point- F-Value- multi curves- free colour selection- export- 1/2 page printout

Structured setup window

Alarm levels may beimplemented in Histogram

Draw marking lines andstore configurations

LOGO of customer on all printout

6 line- text editor, automatic time/date/values stamp with the use of Mark Lines

Statistics with timeperiod ‘value outside ofMin/Max’

clear graphic printout

Tabel printout with change of colour whenoutside Min/Max level

Status printout with all relevant informationabout logger settings

As easy as 123,perfect documents for your quality assurance

Page 20: ECOLOG Data Sheets - tqcsheen.com · Page - 1 ECOLOG Data Sheets Status July 2002 Price list on demand Subject to alterations ECOLOG TN2 Part No. 2420 ECOLOG TN2 for EX-Zone 1 Part

Page - 20

ELPRO-BUCHS AG ELPRO-BUCHS SA ELPRO MESSTECHNIK GmbHLangäulistrasse 62 Route de Grandvaux 26 Baumwasenstrasse 20/1CH-9471 Buchs SG CH-1096 Cully D-73614 SchorndorfSchweiz Suisse DeutschlandTel:+41 81 750 03 11 Tel: + 41 21 799 27 66 Tel: +49 7181 97 94 80Fax: +41 81 750 03 17 Fax: +41 21 799 27 46 Fax: +49 7181 97 94 85email: [email protected] email: [email protected] email: [email protected] D-EZ-2001ElInternet: http://www.elpro.com 7.2002