12
www.baumerprocess.com Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter Hygienic process connections Complete range of electrical connections 4 ... 20 mA and Voltage outputs Accuracy: 0.1%, 0.2% and 0.4% FS Quick response time < 25 ms (Option 5 ms) High long-term stability Excellent repeatability Active compensation for temperature drifts ATEX II 1G, II 2G, II 1/2G and II 1D Lloyd‘s Register Approval Compatibility with CIP/SIP Description The large choice of pressure and electrical connections makes the ED 701 series suitable for a large variety of applications. The ED 701 pressure transmitters are available in 4 … 20 mA current loop or in voltage output signal versions. This transmitters offer a very high accuracy over a wide temperature range, an excellent repeatability and long time stability as well as a short response time (down to 5 ms). The piezo-resistive silicon sensor is anodic bonded on a very stable glass base, which is attached to a stainless steel construction. This assembly guarantees an excellent thermal isolation. The sensor is isolated from the process by a stainless steel diaphragm (1.4435 / 316L) and a filling liquid. The electronics is located within the hermetically sealed transmitter housing which gives the ED 701 an excellent resistance to humidity, shock and vibration. Protection class, from IP 65 to IP 67 only depends on the choice of the electrical connection. The electronics is a state-of-the-art signal processing unit fitted with a fast micro-controller allowing the compensation of drift effects due to temperature on the sensor signal, over a wide temperature range with a fast response time. The ED 701 is available with a great variety of pressure and electrical connections to meet all requirements. The electrical connection can be replaced easily and gives access to the Autozero adjustment. The hygienic design of the ED 701 series is particularly dedicated to the food and beverage processing industry, the pharma and the biotechnology industry. According to its ATEX approval the ED 701 can be used in potentially explosive atmospheres (gases, vapors and dust). The ED 701 has got the Lloyd‘s Register approval, which makes it the ideal product for environments such as in the marine, offshore and industrial applications. ED701 for general industry applications: refer to datasheet ED701_Industrial (D6.180, ED701 – General Industry Transmitter)

ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

Embed Size (px)

Citation preview

Page 1: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com Data Sheet A 11 D6.181

Design and specifications subject to change without notice Page 1

ED 701 – Hygenic Pressure Transmitter

Hygienic process connections

Complete range of electrical connections

4 ... 20 mA and Voltage outputs

Accuracy: 0.1%, 0.2% and 0.4% FS

Quick response time < 25 ms (Option 5 ms)

High long-term stability

Excellent repeatability

Active compensation for temperature drifts

ATEX II 1G, II 2G, II 1/2G and II 1D

Lloyd‘s Register Approval

Compatibility with CIP/SIP

Description

The large choice of pressure and electrical connections makesthe ED 701 series suitable for a large variety of applications. TheED 701 pressure transmitters are available in 4… 20 mA currentloop or in voltage output signal versions. This transmitters offer avery high accuracy over a wide temperature range, an excellentrepeatability and long time stability as well as a short responsetime (down to 5 ms).

The piezo-resistive silicon sensor is anodic bonded on avery stable glass base, which is attached to a stainless steelconstruction. This assembly guarantees an excellent thermalisolation.

The sensor is isolated from the process by a stainless steeldiaphragm (1.4435 / 316L) and a filling liquid. The electronics islocated within the hermetically sealed transmitter housing whichgives the ED 701 an excellent resistance to humidity, shock andvibration. Protection class, from IP 65 to IP 67 only depends onthe choice of the electrical connection.

The electronics is a state-of-the-art signal processing unitfitted with a fast micro-controller allowing the compensation ofdrift effects due to temperature on the sensor signal, over a widetemperature range with a fast response time.

The ED 701 is available with a great variety of pressure andelectrical connections to meet all requirements. The electricalconnection can be replaced easily and gives access to theAutozero adjustment.

The hygienic design of the ED 701 series is particularlydedicated to the food and beverage processing industry, thepharma and the biotechnology industry. According to its ATEXapproval the ED 701 can be used in potentially explosiveatmospheres (gases, vapors and dust). The ED 701 has gotthe Lloyd‘s Register approval, which makes it the ideal productfor environments such as in the marine, offshore and industrialapplications.

ED701 for general industry applications: refer to datasheet ED701_Industrial (D6.180, ED701 – General Industry Transmitter)

Page 2: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page2

1)PressureconnectionL,Z,N/9412,N/9413,Y/9451andY/9452:Notavailableforpressureranges<400mbar.Availableonlywithanaccuracyof0.4%(OrderingcodePosition5)forpressureranges<1bar.Availablewitheveryaccuracycodesforpressureranges1bar.

2)PressureconnectionZ,Y/9451,8and9:Onlyavailablewithoutcoolingneck.TheaccuracycodesAandBarethennotavailable(OrderingcodePosition5).

3)PressureconnectionW:OnlyavailablewiththeaccuracycodesAundB(OrderingcodePosition5),becauseacoolingneckisalwaysintegrated.Onlyavailableforpressureranges10bar.

4)Pressureconnection8and9:Theseconnectionshaven’tgotanydiaphragm.Theyarethenonlyavailablewithoutanyoilfilling(Code9oftheorderingcodePosition11).

5)Lloyd‘sRegisterapprovedED701haveareducedsupplyvoltagerangeandtheambianttemperatureislimiteddependingonthemediumtemperature.ChecktheLloyd‘sRegisterparagraph.

6)FieldHousing:Thecableglandarecompatiblewithcablediameterof5mmto12mm.

7)AccuracycodeAandB(hightemperature):Ifnocalibrationtemperatureisspecified(Option9007/TTTT),thecompensatedtemperaturerangeis[–10°..+80°C].Themaximummediumtemperatureisdependingontheambienttemperature(Tamb):IfTamb40°Cthemax.allowedmediumtemperatureis300°C.IfTamb60°Cthemax.allowedmediumtemperatureis200°C.IfTamb80°Cthemax.allowedmediumtemperatureis150°C.Themax.mediumtemperaturemustnevergoover+200°C,whentheoilfillingistheFDAapprovedwhiteoil.

8)AccuracycodeA:Fortemperatureabove200°ConlythepressureconnectionsJandWwithsiliconorhalocarbonoilareavailable(OrderingcodePosition11).

9)ATEXcertificationisonlyavailablewith4...20mAoutputsignal

Ordering details

Orderingcodedigit:Positions 1 2 3 4 5 6 7 8 9 10 11 12

Process connectionMalethreadG1”flushdiaphragmcone J MalethreadG1”flushdiaphragmO’ring K MalethreadG1⁄2”flushdia.DIN3852(CombiConnect)1) L MalethreadG1⁄2”flushdiaphragmO’ring1)2) ZMalethreadG1⁄2”flushdiaphragmcone3) WAsepticClampDIN11864-3(Additionalcodeneeded) FClampDIN32676(Additionalcodeneeded) HNeumoBioConnect®asepticconnection(Additionalcodeneeded) I ClampISO2852(Additionalcodeneeded)1) NTri-Clamp®,sanitaryclampconnection(Additionalcodeneeded)1)2) YMalethreadG1⁄2”EN837-1forassemblywithpressureseal2)4) 8 MalethreadG1⁄4”EN837-1forassemblywithpressureseal2)4) 9Specificprocessconnection(Additionalcodeneeded) 0

Output signal4...20mA 2 0...10VDC 4 0...5VDC 5

ApprovalCEconform 0ATEX9) 1Lloyd‘sRegister5) 3Lloyd‘sRegister5)+ATEX9) 4

Electrical connection2metershieldedcable3conductors,IP65 12metershieldedcable3conductors,IP67 9 FischerPlug,IP67 2DIN41524BinderPlug,IP65 3 DIN43650Plug,IP65 4 Bendix6polePlug,IP65 5 M12,4poleIndustrialPlug,IP67 6 1⁄2”NPT,cablelength1m,IP67 7 FieldHousing,cableglandbrassnickel-plated,IP676) AFieldHousing,cableglandstainlesssteel,IP676) BFieldHousingcableglandPA,IP676) CFieldHousingM12Plug,stainlesssteel,IP67 D

AccuracyMedium temperature up to 125°C0.4%FS,compensatedtemperaturerange–10°…+125°C 10.2%FS,compensatedtemperaturerange–10°…+125°C 20.1%FS,compensatedtemperaturerange–10°…+80°C 35 ms response time0.4%FS,5ms,compensatedtemperaturerange–10°…+80°C 60.2%FS,5ms,compensatedtemperaturerange0°…+60°C 7High medium temperature0.4%FS,integratedcoolingneck(uptomax.300°C)7)8) A0.2%FS,integratedcoolingneck(uptomax.200°C)7) B

Pressure unit Gaugepressure(100mbarupto40bar) 3Absolutepressure(400mbarupto40bar) 4Negativegaugepressure(0…-100mbar,-1…40bar) 5

Pressure range 100mbarupto40bar:gaugepressure 400mbarupto40bar:absolutepressure

SeetablePressurerange

Page 3: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page3

Ordering details (continued)

Orderingcodedigit:Positions 1 2 3 4 5 6 7 8 9 10 11 12

Wetted parts Allstainlesssteel 1AllHastelloyC 2Allstainlesssteel,withelectropolishedpressureconnection(Ra0.8) 8Strengthenedstainlesssteeldiaphragm10) 3Stainlesssteel,goldplated10) 9

Oil fillingFDAapprovedwhiteoil(Standard)11) 1Siliconoil11) 0Halocarbonoil11) 2Nooilfilling12) 9

10)Strengtheneddiaphragmandgoldplated:Ifnocalibrationtemperatureisspecified(Option9007/TTTT),thecompensatedtemperaturerangeis-10°…+80°C.OnlyavailablewiththepressureconnectionsJ,K,N/9414,N/9416,N/9418.Onlyavailablefortheaccuracyclasses0.2%und0.4%(OrderingcodePosition5).Onlyavailableforpressureranges1bar.

11)Operatingtemperature:refertothetechnicaldata.12)Nooilfilling:onlyforassemblywithpressureseal.

Thefollowingorderingcodeshavetobecompletedwithanadditionalcodetobepreciselydefined:

Process connection Code F (Ordering code Position 1) – Aseptic clamp DIN 11864-3Thefollowingadditionalcodesareavailable.Othersonrequest: /9390 Clampconnectionwithexternaldiameter50.5mm,BKS,SeriesA,FormA,NominaldiameterDN25DIN. PipeaccordingtoDIN11866-A. /9391 Clampconnectionwithexternaldiameter50.5mm,BKS,SeriesB,FormA,NominaldiameterDN33.7ISO. PipeaccordingtoDIN11866-B Example:ED701Fxx.xxx.xxx.xx/9391

Process connection Code H (Ordering code Position 1) – Clamp DIN 32676 Thefollowingadditionalcodesareavailable.Othersonrequest: /9533 Clampconnection,externaldiameter50.5mm,,NominaldiameterDN25,DN32andDN40. PipeaccordingtoDIN11850 Example:ED701Hxx.xxx.xxx.xx/9533

Process connection Code I (Ordering code Position 1) – Neumo BioConnect® aseptic connectionThefollowingadditionalcodesareavailable.Othersonrequest: /9402 FlangeformV,NominalDimension48.3.Flangeexternaldiameter100mm.PipeaccordingtoDIN11866-B, DINENISO1127 Example:ED701Ixx.xxx.xxx.xx/9402

Process connection Code N (Ordering code Position 1) – Clamp ISO 2852 Thefollowingadditionalcodesareavailable.Othersonrequest: /9412 Clampconnection,externaldiameter34mm,NominalDimensionDN21.3.PipeaccordingtoISO2037 /9413 Clampconnection,externaldiameter50.5mm,NominalDimensionDN25.PipeaccordingtoISO2037 /9414 Clampconnection,externaldiameter50.5mm,NominalDimensionDN33.7andDN38. PipeaccordingtoISO2037 /9416 Clampconnection,externaldiameter64mm,NominalDimensionDN40andDN51.PipeaccordingtoISO2037 /9418 Clampconnection,externaldiameter77.5mm,NominalDimensionDN63.5.PipeaccordingtoISO2037 Example:ED701Nxx.xxx.xxx.xx/9414

Process connection Code Y (Ordering code Position 1) – Tri-Clamp®, sanitary clamp connection Thefollowingadditionalcodesareavailable.Othersonrequest: /9451 Tri-Clamp®3/4“connection,externaldiameter24.9mm. /9452 Tri-Clamp®1“connection,externaldiameter50.4mm. /9453 Tri-Clamp®11/2“connection,externaldiameter50.4mm. /9454 Tri-Clamp®2“connection,externaldiameter63.9mm. Example:ED701Yxx.xxx.xxx.xx/9451

Process connection Code 0 (Ordering code Position 1) – Specific pressure connection Thefollowingadditionalcodesareavailable.Othersonrequest: /9483 BBS-Systems®,Quick-Connectasepticclampwithnotch,ISODN25,Externalclampdiameter50.5mm /9486 Malethread1“-11.5NPT,flushdiaphragm Example:ED7010xx.xxx.xxx.xx/9483

Additional ordering codes

Page 4: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page4

Standard ED 701 – Delivery

Calibration temperature:TheED701arecalibratedat+25°C.Ifanothercalibrationtemperatureisneeded,theoption/9007/TTTThastobementioned,aspartoftheorderingcode(seeOptions).Electrical connections:Nocablesocketwillbeprovided.Suchaccessorieshavetobeorderedseparately(seeAccessories).Test Certificate:EachED701willbedeliveredwithaTestCertificateatthecalibrationtemperature(6points).Ifnotspecified,theCertificatewillbeinGerman(seeOptions).O-Ring seals:ThepressureconnectionsJ,K,L,ZandW(OrderingcodePosition1)arealwaysdeliveredwithaNBRO-Ring.OtherO-Ringsareavailable(seeAccessories).AlltheO-RingsareapprovedbytheFDA.TheED701withpressureconnectionF,I,N,Yand0aredeliveredwithoutanyseals.

126 0...100B06 0...150136 0...160A56 0...200146 0...250B26 0...300B36 0...350156 0...400A66 0...500166 0...600B46 0...700176 0...1000P46 0...1250B56 0...1400186 0...1600D06 0…–10020)D16 0…–16020)D26 0…–20020)

L25 1...10015 0...0.6025 0...1035 0...1.6A15 0...2045 0...2.5L35 0...2.8055 0...4A25 0...5065 0...6L45 0...7L05 0...8075 0…10L15 0…12085 0…16A35 0…20095 0…25

Code Range Code Range Code Range Code Range mbar mbar bar bar

L55 0...30105 0...40D85 0…–120)E05 –0.1…120)E15 –0.1…220)315 –1…020)H05 –1…120)525 –1…320)535 –1…520)545 –1…920)E65 –1…1.620)F05 –1…220)F15 –1…1020)F45 –1…12.520)F25 –1…2020)F35 –1…2420)F65 –1…4020)

D46 0…–40020)D56 0…–50020)D66 0…–60020)A76 0...2000196 0...2500206 0...4000A86 0...5000216 0...6000P76 0...7300226 0...10000P86 0...18300H66 –100…10020)H76 –200…20020)H86 –500…50020)C46 –200…020)C56 –500…020)

Description Ordering code

Specific pressure range: /SETRAnapplicationspecificpressurerangecanbedefinedhere.Therangemustbegiveninplaintext.Thesmallestavailablepressurerangesare100mbar(gaugeandnegativegaugepressure)and400mbar(absolutepressure).Example:/SETR500mbar…2500mbarSpecific calibration temperature (temperature of the medium): /9007/TTTTExample:/9007/0100.TheED701willbecalibratedforamediumtemperatureof100°C(exfactory).Thisoptionisreducingthecompensatedtemperaturerangeto±25°Caroundthespecifiedcalibrationtemperature.Ferrite content according to customer specifications /9330Ferritecontentofthepiecesincontactwiththefluidaccordingtocustomerspecifications.Example:/9330/Pressureconnection<1%,weld<3%.Free of oil and grease /9331Freeofoilandgrease(piecesincontactwiththefluid).TheED701iscleanedandpackedaccordingly.Pressure connection electropolished according to customer specifications /9344Thepressureconnectioniselectropolishedaccordingtocustomerspecifications.Onlyavailablewithcode8attheposition10oftheorderingcode.Example:/9344/Ra0.4.Language of the Test Certificate and Operating Manual:BothdocumentsareavailableinGerman,FrenchandEnglish.Thedesiredlanguagemustbespecifiedinplaintextwiththeorderingcode.StandardwillbeGerman.

Options

Pressure range

20)Negativegaugepressureranges:onlyavailablewiththepressureunitcode5(OrderingcodePosition6).

Page 5: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page5

Code Range Code Range Code Range psi psi in H2O [4°C]

20)Negativegaugepressureranges:onlyavailablewiththepressureunitcode5(OrderingcodePosition6).

17G 0...4018G 0...5019G 0...6022G 0...10030G 0...30040G 0...750

Code Range Code Range Code Range Code Range mm H2O [4°C] m H2O [4°C] m H2O [4°C] m H2O [4°C]

177 0...1000P47 0...1250187 0...1600A77 0...2000197 0...2500207 0...4000A87 0...5000217 0...6000P77 0...7300227 0...10000P87 0...18300

028 0...1038 0...1.6A18 0...2048 0...2.5L38 0...2.8058 0...4A28 0...5068 0...6L48 0...7L08 0...8078 0...10L28 1...10L18 0...12088 0...16A38 0...20098 0...25

H08 -1...120)H18 -2...220)H28 -5...520)H38 -10...1020)C08 -2...020)C18 -5...020)D88 0...-120)E08 -0.1...120)E18 -0.1...220)E68 -1...1.620)F08 -1...220)F18 -1...1020)F28 -1...2020)F38 -1...2420)F48 -1...12.520)

17D 0…30018D 0…40032D 0…50070D –1.5…1.520)71D –3…320)72D –6…620)80D –15…1520)60D –1.5…020)61D –3…020)62D –6…020)81D –15…3020)82D –15…6020)83D –15…15020)

L58 0...30L78 0...32108 0...40A48 0...50118 0...60L88 0...64L68 0...70128 0...100L98 0...128B08 0...150138 0...160A58 0...200148 0...250B28 0...300B38 0...350158 0...400

Code Range Code Range Code Range Code Range MPa KPa KPa KPa

013 0...0.6023 0...1033 0...1.6A13 0...2043 0...2.5053 0...4883 -0.1...0.920)E03 -0.1...120)E13 -0.1...220)

072 0...10L22 1...10L12 0...12082 0...16A32 0...20092 0...25L52 0...30L72 0...32102 0...40A42 0...50112 0...60L82 0...64L62 0...70122 0...100132 0...160A52 0...200142 0...250B22 0...300B32 0...350152 0...400

392 -60...020)402 -100...020)C92 0...-6020)D02 0...-10020)642 -10...1520)652 -15...1020)662 -15...2520)672 -25...1520)682 -20...4020)692 -40...2020)702 -40...6020)712 -60...4020)722 -60...10020)732 -100...6020)742 -100...15020)G02 -100...20020)922 -100...30020)932 -100...50020)

A62 0...500162 0...600B42 0...700172 0...1000P42 0...1250B52 0...1400182 0...1600A72 0...2000192 0...2500202 0...4000H32 -10...1020)H42 -20...2020)H52 -50...5020)H62 -100...10020)352 -10...020)362 -16...020)C22 -20...020)372 -25...020)382 -44...020)C32 -50...020)

05D 0...1.506D 0...2.507D 0...308D 0...409D 0...645D 0...7.510D 0...1011D 0...1512D 0…3013D 0…6051D 0…7514D 0...10015D 0...15016D 0…200

Pressure range

Page 6: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page6

Technical data

Measurement characteristics

Pressure rangeGaugepressure: from0…100mbarupto0…40bar.Absolutepressure: from0…400mbarupto0…40bar.Thestandardpressurerangesarelistedinthepressurerangetable.OtherspecificrangesareavailablewiththeoptionSETR.

Over pressure limit Pmax.Depending on the pressure range 0.. PP [bar] 0,1..0,35 0,4..1,0 1,6..5 6..20 25 30..40Maximum pressure Pmax [bar] 1 3 15 60 70 135

Accuracy0.1%FS,0.2%FSund0.4%FSclasses(seeorderingcodePosition5).Hysteresis and repeatability ±0.05%FS

Long term stability ±0.2%FS/year

Response time (10 … 90%) 25ms 5mswiththeaccuracycodes6und7 (OrderingcodePosition5).

Temperature influence (in compensated range)

Dependingontheaccuracyclass:

Orderingcode,Position5=1and6(class0.4%) Zero: TC±0.20%FS/10K. Span: TC±0.15%FS/10K.

Orderingcode,Position5=2and7(class0.2%) Zero: TC±0.15%FS/10K. Span: TC±0.10%FS/10K.

Orderingcode,Position5=3(class0.1%) Zero: TC±0.05%FS/10K. Span: TC±0.05%FS/10K.

Orderingcode,Position5=AandB(hightemperature) Zero: TC±0.20%FS/10K. Span: TC±0.15%FS/10K.

Dependingonthewettedpartsmaterial:

Orderingcode,Position10=3and9 Zero: TC±0.30%FS/10K. Span: TC±0.20%FS/10K.

Admissible temperature of the medium

ED701 without integrated cooling neck (Ordering code Position 5: Code 1, 2, 3, 6 or 7): ED701withFDAwhiteoil: -10°C…+125°C ED701withsiliconorhalocarbonoil: -30°C…+125°C

ED701 with integrated cooling neck (Ordering code Position 5: Code A or B): Forhighmediumtemperature(120°Candhigher)theED701hastohaveanintegratedcoolingneck. ED701withcoolingneckwith FDAwhiteoil: -10°C…+200°C ED701withcoolingneckwithsiliconor halocarbonoil: -30°C…+300°C

Ambient temperature (housing)

-10°C…+80°C

Storage temperature

-10°C…+80°C

Sterilisation temperature (SIP Process)

Sterilisation:max.+140°C,max.30minutes

SterilisationCompletedevice,notpowered,onlyforexecutionswiththeFischerelectricalplug.Processdurationmax.60minutes,sterilisationtemperaturemax.140°C,pressure3500mbar,notpowered.

Materials specification

Wetted partStainlessSteel1.4435(AISI316L)

Parts not in contact with mediumStainlessSteel1.4301,1.4305andelectricalconnectormaterial.

Certificate EN102043.1availableonrequest

Electrical specification

Supply voltage 9...30VDC(4…20mA) 10...28VDC(4…20mAEEX) 14...24VDC(4...20mALloyd‘sRegister) 15...30VDC(0…10and0...5VDC) 20...24VDC(0…10and0...5VDCLloyd‘sRegister)

Insulation Resistance >1GOhm,500VDC

Load (currentoutput) RL50xUB-450Load(voltageoutput) RL5kOhm

Protected against reverse signal polarity.

Switch ON time5sec.

EMC

Influence of EMC Lessthan0.15%FS

Conformity EN61000-6-3,EN61000-6-2,EN61326, EN50121-4

ATEX data

IntrinsicallysafeconformingtoEN50020andEN50281-1-1Applications:

II1GII1D Devicelocatedinzone0/zone20

II1/2GII1D Pressureconnectioninzone0/zone20 Housingandelectricalconnectionin zone1/zone20

II2GII1D Devicelocatedinzone1/zone20

IntrinsicallysafetyEExiaIIC.Thedevicecanonlybeconnectedtoanapprovedintrinsicallysafecurrentloopwiththefollowingbounderyvalues:U0=28V Ik=100mA P=1.5WThedeviceinternalcapacityandinductivityvaluesare:Ci=1nF Li=0.1mH

FortheapplicationinEExzoneyouhavetorespecttheconditionsmentionedintheATEXTypeExaminationCertificate.Youfindthecertificatesandmanualsunderhttp://www.baumerprocess.com/en/downloads/.

Environmental specification

Humidity AccordingtoEN60068-2-38Vibrations AccordingtoEN60068-2-6Shocks AccordingtoEN60068-2-31

Protection class

RatingIP65andIP67,dependingonelectricalplug.

Mounting

Positionatcalibrationtime:Vertical

Physical characteristics

Weight Between223and885gdependingoftheversion.

DimensionsSeedrawings.

Disposal of product and packingAccordingtonationallawsorbyreturningtothemanufacturer.

Page 7: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page7

Technical data

Zero adjustment and factory setting:

ThezeroadjustmentandfactorysettingaredirectlyavailableontheED701byremovingtheelectricalplug.Apushbuttononthetopoftheelectronicunitcanbeactivated.Zero adjustment:Bypressingthepushbuttontheappliedpressurewillbeconsideredasthenewzeropointandtheoutputsignalwillbereset(i.e.4mAor0VDC).Thisfunctionallowssomecorrectionsifthemountingpositioninfluencesthemeasurement,iftemperatureorlongtermdriftsoccursandifaresettinginalevelmeasurementapplicationisneeded.Factory setting(zero):Thefactorysettingcanbereloadedatanytime.Todoitthepushbuttonmustbepressedduringatleast10seconds.

Oil filling

Standard:ThestandardoilfillingistheFDAapprovedwhiteoil(Paraffinoil).Thisisthebestchoiceforalltheapplicationsinthefood,beverages,pharmaceuticalandbiotechindustries.Temperaturebelow-10°Candabove+200°Cshouldnotbeused,itcouldseverelydamagetheED701.Alternative:Thesiliconoilcanberecommendedforallgeneralindustrialapplicationsandcanbeusedinatemperaturerangefrom-30°Cupto+300°C.Foroxygenserviceorpaintingindustry,choosethehalocarbonoil.Halocarbonissiliconfreeandwithstandstemperaturesfrom-30°Cupto+300°C.Withthisparticularfilling,weguaranteethatallpartsthatmaybeincontactwiththemedium(e.g.incaseofdiaphragmdamage)aresiliconfree.Componentslocatedinsideofthesealedtransmitterhousingmaycontainsiliconinsolidform.

Pressure connection tightness with O-Ring

Themountingexamplesonthenextpagesshowtheappropriatetightnessmethodsforeachtypeofpressureconnection.AmetaltometalsealispossiblewiththepressureconnectionJandW.ThisuniqueBourdon-Haenniconeconstructionhastoberealizedusingtheadequateweldingsleeve(seeAccessories).ThepressureconnectionsKandZareusuallysealedbytheO-ringlocatedonthefrontoftheconnection.TheconnectionLiscompatiblewiththeCombiConnectsystemproposedbyBourdon-Haenni,allowingahygienicsealatthefrontoftheconnection.TheCombiConnectsystemconsistsofaseriesofhygienic,replaceableconnections.Moredetailsonrequest.AllthepressureconnectionsJ,K,L,W,ZcanalsobesealedbytheO-ringlocatedattherearofthethread(ortightnesssurface).

Lloyd‘s Register (LR)

TheED701hasbeentestedwithsatisfactoryresultsinaccordancewiththerequirementsoftheLRTypeApprovalSystem.

Certificate number: 06/20034.(Expires:29August2011).

Applications: Marine,offshoreandindustrialapplicationsforuseinenvironmentalcategoriesENV1,ENV2andENV3asdefinedinLloyd‘sRegister‘sTypeApprovalSystem,TestSpecificationNumber1-2002.

Limitations:ThesupplyvoltagerangeofED701isisreducedcomparedtothenonLRapprovedED701.(PleaserefertotheparagraphTechnicaldata-Electricalspecification).AccordingtothecurvebelowtheallowedambienttemperatureforLloyd‘sRegisterapprovedED701isderatedatmediumtemperaturesabove150°C.

80 ºC

T°ambient

AllowedZone

60 ºC

-10 ºC T° medium

150 ºC 200 ºC

Page 8: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page8

30)Onlyforpressureranges 1barandaccuracyclasses0.2%and0.4%.31)Onlyforpressureranges 400mbarandwiththeaccuracyclass0.4%.32)Onlyforpressureranges 10barandwiththeaccuracyclassesAandB.33)Onlyforpressureranges 1bar.34)Onlyavailablewithsiliconoilorhalocarbonoil.

35)UseappropriateO-Ringorsealfortemperatureupto200°C.WerecommendViton®.

Viton®isaregisteredtrademarkofDuPontPerformanceElastomers

Available Availablewithrestrictions

Process connection availability – Overview table

Code Process connection Material Oil filling Accuracy

J MalethreadG1“flushdiaphragmwithcone 30) 30) 35) 35) 34)

K MalethreadG1“flushdiaphragmwithO’ring 30) 30) 35) 35)

L MalethreadG1⁄2“flushdiaphragmDIN3852 31) 33) 33) 31) 35) 35)

Z MalethreadG1⁄2“flushdiaphragmwithO’ring 31) 33) 33) 31)

W MalethreadG1⁄2“flushdiaphragmwithcone 32) 35) 35) 34)

F /9390ClampDIN11864-3,BKS,DN25DIN 35) 35)

F /9391ClampDIN11864-3,BKS,DN33.7ISO 35) 35)

H /9533ClampDIN32676,DN25,DN32,DN40 30) 30) 35) 35)

I /9402NeumoBioConnect®,FormV,DN48.3 35) 35)

N /9412ClampISO2852,DN21.3 31) 33) 33) 31) 35) 35)

N /9413ClampISO2852,DN25 31) 33) 33) 31) 35) 35)

N /9414ClampISO2852,DN33.7,DN38 30) 30) 35) 35)

N /9416ClampISO2852,DN40,DN51 30) 30) 35) 35)

N /9418ClampISO2852,DN63.5 30) 30) 35) 35)

Y /9451Tri-Clamp®3⁄4” 31) 33) 33) 31)

Y /9452Tri-Clamp®1” 31) 33) 33) 31) 35) 35)

Y /9453Tri-Clamp®11⁄2” 35) 35)

Y /9454Tri-Clamp®2” 35) 35)

8 MalethreadG1⁄2”EN837-1forpressureseal

9 MalethreadG1⁄4”EN837-1forpressureseal

0 /9483BBS,Quick-Connect,ISODN25 35) 35)

0 /9486Malethread1“-11.5NPTflushdiaphragm 35) 35)

1.44

35

Hast

ello

y-C

Stre

ngth

ened

dia

phra

gm

0.2%

FS

0.4%

FS

0.2%

for h

igh

med

ium

te

mpe

ratu

re (u

p to

200

°C)

FDA

Silic

on

Halo

carb

on

P <

1 b

ar

P

1 b

ar

0.1%

FS

0.4%

for h

igh

med

ium

te

mpe

ratu

re (u

p to

200

°C)

0.4%

for h

igh

med

ium

te

mpe

ratu

re (u

p to

300

°C)

Gold

pla

ted

Page 9: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page9

Drawing (all dimensions in mm)

OnlyfororderingcodePosition5=AandB

Code l4

K, J 48L, W 50F /9390, F /9391 50I /9402 50N /9412, N /9413, N /9414 50N /9416, N /9418 500 /9483

DIN 43650 plugCode:4

34

17

Bendix 6 pole plugCode:5

34

17

Cable 3 conductorsIP 65 / IP 67 Code: 1(IP65) 9(IP67)

34

10.4

DIN 41524Binder plugCode:3

34

14

Fischer plugCode:2

Field housingCode:A,B,CandD

34 55

14.5

M12 industrial plugCode:6

34

10.4

55

NPT 3 conductorsCode:7

34

10

SW27

34

61.5

40

l 4

Page 10: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page10

Drawing (all dimensions in mm)

Mounting example

einbau_g1_ohne_konus_D13708.dxf

Seal

Seal

Weldingsleeve

Pressureport

Screwhole

Pressureport

PressureportSeal

Pressureport

Screwhole

Clamp

Code Threadd1 d2 l1 l2

K G1” 30 12.5 20.5Z G1⁄2” 18 10.5 20.50 /9486 1“-11.5 NPT

Code Nominalsize d1 d2 l1H /9533 DN 25/32/40 50,5 24N /9412 DN 21,3 34 17,5 15N /9413 DN25 50,5 17.5 15N /9414 DN 33,7/DN 38 50,5 24 15N /9416 DN 40/DN 51 64 24 15N /9418 DN 63,5 77,5 24 15Y /9451 3⁄4” 24,9 16 31,7Y /9452 1” 50,4 17,5Y /9453 11⁄2” 50,4 24Y /9454 2” 63,9 240 /9483 ISO DN25 50,5 24

key41(TypK,and0/9486)key27(TypZ)

d2

d1

l 1

l 212

d1

l 1

d2diaphragm

key41(TypJ)key27(TypW)

d1

d2

l 1l 2

l 3

70°

Code Threadd2 l1 l2 l3

J G1” 25.5 18 9W G1⁄2” 15.2 13.7 9 16.5

key27

d2

d1

l 116

.5

l 2

Code Threadd1 d2 l1 l2

L G1⁄2” 17.5 16 2.5

Code Nominalsize d1 d2 l1F /9330 DN 25 DIN 50.5 24 15F /9331 DN 33.7 ISO 50.5 24 15

d1

l 1

d2diaphragm

Code Nominalsize d1 d2 l1I /9402 100 24

d1

d2diaphragm

Page 11: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page11

Accessories

Connecting diagram

Electrical connection

ForVoltagesignalOutput0…max.10V For4...20mAcurrentloopversion

a

b

c

d

IA

UA RL

UB UB

RLIA

a

b

d

Pin assignment

Contact Fischer plug NPT 3 conductors M12 industrial plug Field housing

4...20mA 0...5/10VDC 4...20mA 0...5/10VDC 4...20mA 0...5/10VDC 4...20mA 0...5/10VDCa 1 1 Red Red 3 3 1 1b 2 2 Black Black 1 1 2 2c - 3 - Green - 4 - 3d Case Case Case Case 4 4

Viewsolderingside,cablesocket

Welding sleeves

Code D a H Material Ordering code J G1” 55 23 Stainlesssteel1.4435 D 11737.0002 J G1” 55 23 HastelloyC D 11737.0022 W G1⁄2” 35 20 Stainlesssteel1.4435 D 11737.0000 W G1⁄2” 35 20 HastelloyC D 11737.0020

Code D a H Material Ordering code

K G1” 60 20.5 Stainlesssteel1.4435 D 12403.1 Z G1⁄2” 60 20.5 Stainlesssteel1.4435 D 12403.0

Front piece with thread

Code D a b sw Material Ordering code J G1” 18 27 41 Stainlesssteel1.4435 D 11870.0002 J G1” 18 27 41 HastelloyC D 11870.0022 W G1⁄2” 12 23 27 Stainlesssteel1.4435 D 11870.0000 W G1⁄2” 12 23 27 HastelloyC D 11870.0020

aD

H

Pin assignment

Contact DIN 43650 plug Bendix 6 poles plug Cable 3 conductors DIN 41524 Binder plug

4...20mA 0...5/10VDC 4...20mA 0...5/10VDC 4...20mA 0...5/10VDC 4...20mA 0...5/10VDCa 1 1 B B Blue Blue 3 3b 2 2 E E Brown Brown 1 1c - 3 - D - Black - 4d GND GND Case Case Shield Shield Case Case

Viewsolderingside,cablesocket

red blackgreen

blue

blackbrown

shield

Page 12: ED 701 – Hygenic Pressure Transmitter - PCN Europe · Data Sheet A 11 D6.181 Design and specifications subject to change without notice Page 1 ED 701 – Hygenic Pressure Transmitter

www.baumerprocess.com DataSheet A11 D6.181

Designandspecificationssubjecttochangewithoutnotice Page12

Electrical sockets

Description Ordering codeCablesocketconnectorDIN43650 E 6844.0

Description Ordering codeCablesocketconnectorBinder E 13265.0

Description Ordering codeCablesocketconnectorBendix E 6586.1

Description Ordering codeCablesocketconnectorM12 E 15560.0

Cable Ordering codeCablewithM12socketconnectorforED701 E 15543.x LLL

Type ShieldedPVCKabel 0 ShieldedPURKabel 1 Length 2meters 002 5meters 005 10meters 010 20meters 020

Description Ordering codeCablesocketconnectorM12 E 15560.1

Accessories

Description Ordering codeCablesocketconnectorFischer E 13267.0

O-Rings

TheFDAapprovedO-ringsareavailableinNBR,EPDMandViton®(FPM).Theyaredeliveredinpacksof5units.

ThefollowingoperatingtemperatureoftheO-ringsareonlyindicativevalues(dryair):NBR(FDA): –25°…+125°CEPDM(FDA): –40°…+160°CFPM/Viton®(FDA): –20°…+200°C

Process connection Designation Material Ordering code

L,W,Z O-RingformalethreadG1⁄2”(attherear) NBR(FDA) A8550.9110L,W,Z O-RingformalethreadG1⁄2”(attherear) FPM/Viton®(FDA) A8550.9111L,W,Z O-RingformalethreadG1⁄2”(attherear) EPDM(FDA) A8550.9112J,K O-RingfürG1”(attherear) NBR(FDA) A8550.9120J,K O-RingfürG1”(attherear) FPM/Viton®(FDA) A8550.9121J,K O-RingfürG1”(attherear) EPDM(FDA) A8550.9122K O-RingfürG1”(onthefront) NBR(FDA) A8550.9130K O-RingfürG1”(onthefront) FPM/Viton®(FDA) A8550.9131K O-RingfürG1”(onthefront) EPDM(FDA) A8550.9132Z O-RingfürG1⁄2”(onthefront) NBR(FDA) A8550.9140Z O-RingfürG1⁄2”(onthefront) FPM/Viton®(FDA) A8550.9141Z O-RingfürG1⁄2”(onthefront) EPDM(FDA) A8550.9142

Accessories for CombiConnect

SeeDataSheet3400-3,FlexbarHRTaccessories.

Viton®isaregisteredtrademarkofDuPontPerformanceElastomers

35

9

35

9

52.5

35

2

0

M12

x1

SW

19

D6.

181/

UK

/200

8-09

-22

This

dat

ash

eet

may

onl

yb

ere

pro

duc

edin

full.