Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT. Kathleen Vunckx Johan Nuyts Nuclear Medicine, K.U.Leuven. Outline. Introduction Pinhole design evaluation method Effect overlap Conclusions and future work. Pinhole SPECT. Pinhole collimator. - PowerPoint PPT Presentation

Citation preview

Page 1: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Effect of Overlapping Projections on Reconstruction Image Quality in

Multipinhole SPECT

Kathleen Vunckx

Johan Nuyts

Nuclear Medicine, K.U.Leuven

Page 2: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Outline

• Introduction

• Pinhole design evaluation method

• Effect overlap

• Conclusions and future work

Page 3: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole SPECT

Clinical SPECT camera

Pinhole collimator

Pinhole insert

Page 4: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Parallel hole vs. pinhole SPECTParallel hole collimator

Pinhole collimator

Geometrical Magnification

Reduced field of view and

increased resolution

=> ideal for small animal imaging

Detector

Focal Point

Detector

Page 5: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Single vs. multipinhole SPECT

Page 6: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Single vs. multipinhole SPECT

• Improved sensitivity• Better sampling

Higher reconstructio

n image quality

Page 7: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Outline

• Introduction

• Pinhole design evaluation method

• Effect overlap

• Conclusions and future work

Page 8: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

• # detectors• # pinholes/detector

• Diameter of the pinholes

• Acceptance angle (A)• Inclination angle (I)• Position of the

pinholes (PA, PD)

• Distance to detector (DD)

• Distance to AOR (DA)• Insert thickness (T)• Focussing point (F)

axis of rotation (AOR)

T

DD

DAF

PD

PAA

I

Pinhole design evaluation method: design parameters

Page 9: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method: introduction

• Aim: Fast method for evaluation of various (multi)pinhole collimator designs

• Evaluation: Quantify the effect of the design on the signal and the noise of the reconstructed voxel values

Fixed target resolution!

Impulse response & covariance

Page 10: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method:impulse response

Phantom/Animal

Projection data

Recon-struction

Without impulse

With impulse

Impulse response

central value & variance

contrast-to-noise ratio (CNR)

Page 11: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method: reference method

Very slow, takes weeks to find a decent design!

impulse

impulsewithoutreconimpulsewithrecon .. central value

variance

• Impulse response:

• Covariance image:For each design:

For a set of voxels:Simulate a lot of iterative reconstructions with different noise on the projection data

Page 12: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

i i

ii AXy2

2

Pinhole design evaluation method: efficient analytical method

• Can we more efficiently calculate:– Impulse response– Covariance image

• Weighted least squares approximation:

AXYCAXY YT 1

reconstructed image

projection

measurement in detector pixel i

variance (= weight)

Assumes Gaussian distribution!

Matrix notation:

covariance matrix of Y

Page 13: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method: efficient analytical method

• Impulse response:

ACAACAX YT

YT 111

0x

AXYCAXY YT 1min

YCAACAX YT

YT 111

AY

Weighted least squares

reconstructed image

projection

covariance matrix of Y

measurement

Impulse response in voxel j:

impulse

Fisher information

jj FFX 1

Page 14: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

TT

YT

T

YYYT

TT

Y

T

YT

T

X

FFF

FACAF

FACCCAF

FACYYYYECAF

XXXXEC

11

111

1111

1111

Pinhole design evaluation method: efficient analytical method

• Covariance matrix Cx

covariance matrix of Y

YCAFX YT 11

jTjX FFFC 11

Covariance in voxel j:

Page 15: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method: efficient analytical method

• For a WLS approximation:

• For pinhole SPECT post-smoothed MLEM reconstruction:– Shift-variant

jTjX

jj

FFFC

FFX

11

1

F point 1 F point 2

12

Page 16: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method: efficient analytical method

• For a WLS approximation:

• For pinhole SPECT post-smoothed MLEM reconstruction:– Shift-variant => assume local shift-invariance

=> turn F into a circulant matrix F j

jTjX

jj

FFFC

FFX

11

1

…row j

shifted row j

shifted row j

matrix F matrix Fj

Page 17: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method: efficient analytical method

• For a WLS approximation:

• For pinhole SPECT post-smoothed MLEM reconstruction:– Shift-variant:– Local approximations

Good approximation close to voxel j, less accurate if distance from j increases

jTjX

jj

FFFC

FFX

11

1

=> multiply elements of Fj with weight that linearly descreases with increasing distance from j

jFF

Page 18: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method: efficient analytical method

• For a WLS approximation:

• For pinhole SPECT post-smoothed MLEM reconstruction:– Shift-variant:– Local approximations:– Incomplete data => not of full rank

=> does not exist

jTjX

jj

FFFC

FFX

11

1

=> Replace by approximated pseudoinverse G (calculated in Fourier domain)

jj FF~

jFF

jF~

1~ jF

Page 19: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method: efficient analytical method

• For a WLS approximation:

• For pinhole SPECT post-smoothed MLEM reconstruction:– Shift-variant:– Local approximations:– Incomplete data:– Post-smoothing to obtain fixed target resolution

=> for each voxel j, choose an isotropicalGaussian P such that is as

accurateas possible

jTjX

jj

FFFC

FFX

11

1

jFPGT~

isotropical Gaussian with FWHM = target resolution

jFF jj FF~

GF j 1~

Page 20: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method: efficient analytical method

• For a WLS approximation:

• For pinhole SPECT post-smoothed MLEM reconstruction:– Shift-variant:– Local approximations:– Incomplete data:– Post-smoothing:

jTTjjX

jjj

PGFPGC

FPGX

~

~

jTjX

jj

FFFC

FFX

11

1

jFF jj FF~

GF j 1~

jFPGT~

Page 21: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

• Good local approximations for:– Impulse response central value– Covariance image variance

• Approximates post-smoothed MLEM with fixed target resolution

Pinhole design evaluation method: efficient analytical method

Quite efficient, takes a few hours to find a decent design!

Page 22: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Pinhole design evaluation method: validation study

we

eks

hours

Page 23: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Outline

• Introduction

• Pinhole design evaluation method

• Effect overlap

• Conclusions and future work

Page 24: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Collimator

Detector

Collimator

Detector

Overlapping projections:problem statement

OVERLAP NO OVERLAP

Extra shielding

Page 25: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Collimator

Detector

Collimator

Detector

Overlapping projections:benificial?

Higher sensitivity

OR Unambiguous information

OVERLAP NO OVERLAP

Extra shielding

Page 26: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Overlapping projections:benificial?

105 15

Page 27: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Overlapping projections:conclusions

• With overlap– Do not use too many apertures– Pinhole apertures farther from each other

• Without overlap– More pinholes– Pinhole apertures closer to each other– Better than same design with overlap

REMOVE OVERLAP

Page 28: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Outline

• Introduction

• Pinhole design evaluation method

• Effect overlap

• Conclusions and future work

Page 29: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Conclusions

• Accurate and efficient method to evaluate pinhole collimator designs by predicting the image reconstruction quality.

• Useful to investigate:– effect of design parameters– effect of overlapping projections

• Removing overlap gives promissing results

Page 30: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Future work

• Optimize (dual head) multipinhole design for:– rat brain imaging– whole body mouse imaging– cardiac imaging in rabbits

• Investigate influence of multipinhole-specific artifacts

• Extend method to model collimator penetration and attenuation

Page 31: Effect of Overlapping Projections on Reconstruction Image Quality in Multipinhole SPECT

Nuclear Medicine

Effect of Overlapping Projections on Reconstruction Image Quality in

Multipinhole SPECT

Kathleen Vunckx

Johan Nuyts

Nuclear Medicine, K.U.Leuven