Upload
phungminh
View
216
Download
0
Embed Size (px)
Citation preview
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 1 of 41
DL OPEN REVOLUTION REBREATHER FAMILY OF PRODUCTS
EFFECT OF SMALL VARIATIONS IN TEST CONDITIONS ON THE CO2
SCRUBBER ENDURANCE TO EN14143:2003 AND NORSOK U-101
DOCUMENT NUMBER:[Filename]
DV_OR_ScrubberEndurance_081124.doc
CONTRIBUTORS: Dr. Bob Davidov, Dr. Oleg Zagrebelny, Mihael Soloviev, Viktor Sudakov, Dr. Alex Deas
DEPARTMENT: Engineering LAST REVISED: 24th Nov 2008
REVISION: C
APPROVALS _______________________________ Project Manager
________________________Date
_______________________________ Quality Officer
________________________Date
Controlled Document Classified Document DO NOT COPY.
Revision History Revision Date Description
A1 5th May 2008 Document assembled from other files
B 5th May 2008 Calibration information moved to DV report
C 24th Nov 2008 Horizontal cases added. Correlation with results at Micropore facility.
Copyright © 2008 Deep Life Ltd
All patents, patent applications, design and topographical rights reserved. No circuit may be reproduced without a licence for the topographical rights contained therein from Deep Life Ltd.
This document does not constitute a licence to use and patent, patent application or topographical right of Deep Life Ltd.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 2 of 41
Table of Contents1 PURPOSE AND SCOPE...........................................................................................3
2 TEST STRUCTURE AND PARAMETERS................................................................3
3 CALIBRATION AND EQUIPMENT...........................................................................8
4 TEST RESULTS........................................................................................................9
4.1 Standard conditions: Vertical, Gas temperature 32C, water 4C ................................................ 9
4.2 Colder water: Vertical, 32C exhale, but ambient water just 1C, test 3 scrubbers .................. 13 4.2.1 First scrubber sample ............................................................................................................... 13 4.2.2 Second scrubber sample ........................................................................................................... 16 4.2.3 Third scrubber sample.............................................................................................................. 20
4.3 Colder water: Vertical, 32C exhale, but ambient water just 1C, short interrupt of CO2 injection ..................................................................................................................................................... 22
4.4 Effect of using a scrubber for two dives in very cold conditions............................................... 24
4.5 Reducing the exhaled gas temperature from 32C to 21C .......................................................... 28
4.6 Horizontal Position: 32C exhale, 1C water, test three scrubbers.............................................. 31 4.6.1 First scrubber cartridge sample ................................................................................................ 32 4.6.2 Second scrubber sample ........................................................................................................... 35 4.6.3 Third scrubber sample.............................................................................................................. 37
5 MAXIMUM INHALED CO2 LIMITS .........................................................................38
6 CORRELATION WITH THIRD PARTY TEST RESULTS .......................................38
7 CONCLUSION ........................................................................................................39
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 3 of 41
1 PURPOSE AND SCOPE This document describes the test method and results of scrubber endurance trials carried out by Deep Life Limited to assess the effect of small variations in temperature on the scrubber endurance. The variations considered are:
1. Reference test under standard conditions defined in EN14143 and NORSOK U-101
2. Effect of reducing the ambient water temperature from 4C to 1C
3. Effect of reducing the exhaled gas temperature from 32C to 21C
4. Effect of performing two dives with an interval of 18 hours without changing the scrubber rotation, and assuming the dives have a low exhaled gas temperature.
5. Effect of adopting a horizontal position instead of a vertical position in very cold ambient water (1C)
All tests were done at 40msw, using the 1.6 lpm flow rate specified as one of the tests in NORSOK U-101 and the only test condition in EN14143:2003.
Tests were stopped at a PPCO2 corrected to surface pressure of at least 2kPa: the NORSOK U-101 maximum permitted CO2 level, to ensure the curve is gradual and continuous. DL generally rate scrubbers to the 2kPa limit permitted in NORSOK U-101, as this limit is free of symptoms for normal diving over the short duration this spike occurs.
The scrubbers are tested inside the single scrubber Open Revolution rebreather (Sample 1).
2 TEST STRUCTURE AND PARAMETERS The tests use the following equipment:
1. Single scrubber Open Revolution rebreather, fitted with scrubber health monitor (12 temperatures measured axially along the scrubber) and RH sensor immediately prior to scrubber.
2. Deep Life Breathing Machine A, fitted with humidifier
3. Matlab data capture system.
4. 600mm diameter chamber working pressure to 2000msw, with refrigeration
5. 160mm diameter chamber rated to 300msw for calibration purposes.
6. Chamber refrigeration cooling system, computer controlled.
7. Supply of compressed gases including air, pure carbon dioxide and nitrogen.
8. Pressure, CO2, temperature and humidity sensors
9. Mass flow control valve
10. Fresh Micropore EAC scrubbers are required for each test run. The scrubbers were the 125mm Micropore EACs, with large internal bore, of a batch marked “Use by February 2009”.
11. Scales to weigh the CO2 cylinder before and after each test to confirm that the correct amount of CO2 is injected. The smallest cylinder that can provide sufficient CO2 for the test is a 10 litre cylinder for a single cartridge: the test requires 9.1 litre calculated as (1.6lpm_of_MR x (150min_of_test+50_min_of_reserve)) / (45bar_of_start_pressure – 10bar_of_end_pressure). The weight of a 10 litre cylinder is 15 kg. The weight of the gas used for the 150 min test is 473 g (as 1.6l/m x 150
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 4 of 41
min x 1.97 g/l). This check is only a gross check: it is not a calibration check because it would require a balance with 20 kg range and 5g or 0.025% accuracy to measure the amount of CO2 weigh with 1% accuracy.
Note: the pressure of the remaining CO2 is monitored, with correction for temperature. The maximum pressure of CO2 against temperature is shown in Figure 2-1. This plot is derived from the CO2 pressure – temperature phase diagram shown in Figure 2-2
0 5 10 15 20 25
35
40
45
50
55
60
65
Pressure - Temperature phase diagram for CO2
Temperature, C
Pre
ssur
e, a
tm
formulatable data
Figure 2-1. Gas – liquid phase diagram of CO2 (detail).
Figure 2-2. Gas – liquid phase diagram of CO2.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 5 of 41
The CE EN14143:2003 and the comparable NORSOK U-101 test, specify the following test conditions:
1. Depth/ambient pressure is 40 msw / 5 bar
2. Water temperature is 4+/-1C or less
3. Diluent gas is Nitrogen
4. Carbon dioxide injection rate is 1.6 lpm
5. Breathing simulator rate is 40 lpm
6. Breathing frequency is 20 min-1
As for EN14143:2003, NORSOK U-101 the scrubber endurance is stated for the PPCO2 being less than 5 mbar shown in Figure 2-3 and less than the warning levels of 8 mbar and 20 mbar shown in Figure 2-4.
5
EnduranceTime
PPCO2,mbar
Figure 2-3. Scrubber endurance as defined in EN14143:2003. The 0.5 kPa is a very low level. DL generally uses the NORSOK limits which are compared below.
PPCO2
mbar kPa
Standard2
5
10
8
20 2
1
0.5
Warning level:5+/- 3 mbar
(0.5+/- 0.3 kPa)
EN14143:2003 NORSOK U-101
Warning level:< 20 mbar(< 2 kPa)
Figure 2-4. PPCO2 warning levels for EN14143 and NORSOK U-101.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 6 of 41
Exhale flow is separated from the breathing cycle by sampling the gas line at the diver’s lips (using a dummy mouthpiece bite), and switching it using a solenoid which is open during exhale cycles. The exhale cycle is synchronized with the breathing machine without any delay.
Before testing the mass flow control valve used for CO2 injection has been set for 1.6 lpm flow rate under the valve input pressure of 10 bar, which has been calibrated: a DV report is available. A flow metre provides a further measure to monitor flow levels.
The test fixture used for scrubber endurance tests is shown in Figure 2-5.
All operations of the test fixture other than connecting gases, are computer controlled and logged.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 7 of 41
N2
cylin
der
1.6 l/min
Sin
gle
UN
IT
CO
2 cy
linde
r
< 0.16
l/min
Bre
athi
ng
cont
rol
< 0.
16 l/
min
CO
2 se
nsor
Scr
ubbe
r ca
nist
er
Figure 2-5. Deep Life’s scrubber endurance test fixture.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 8 of 41
3 CALIBRATION AND EQUIPMENT The calibration of the response of the CO2 sensor was a complex process, as it involved confirming that it can provide a 150ms response. Due to the extensive nature of those calibrations, a detailed calibration report and equipment description has been produced for these tests, in DV_CO2_Cal_081201.pdf This report is summarised below.
This calibration process forms the bulk of the separate calibration report. It is noted that the mixing that occurs in the sampling line will reduce the effective response time.
The calibration of the instruments used in the tests are managed within the normal QA controls, except for the flow and CO2 sensors, which are critical to the scrubber endurance tests, so is quad checked using the process describe below and then further audited using a fifth method, as well as having tests carried out at an independent laboratory.
The volume of CO2 injected from the cylinder is verified by the use of four methods (that is, by using all four methods each time to ensure there is no significant error):
1. By the change in the weight of the CO2 cylinder,
2. By the cylinder pressure. As a reference the temperature of CO2 inside the cylinder is calculated using the measured pressure and the pressure-temperature phase diagram shown in Figure 2-1.
3. By oxygen trace gases. The CO2 is mixed with air, such that a bleed measurement gas can be taken from the device under test without reducing the pressure in the chamber. The bleed causes a reduction in PPO2 which should be compensated almost exactly by the injected O2, if the injection rate is correct. To check this, the PPO2 of the loop is monitored. The flow from the bleed line is monitored: this should be between 0.08 and 0.2 lpm.
4. By the period during which there is a CO2 flow at a constant rate: this is the primary method used to set the amount of CO2 dosed into the scrubber. The accuracy this method depends on the accuracy of the CO2 flow through the mass flow controller. The controller is calibrated using a 1.6 lpm test flow at STPD (273ºK, i.e. 0ºC, and 1 atm). At room temperature a temperature correction must be used as shown in Figure 3-1.
Figure 3-1. CO2 flow calibration.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 9 of 41
The flow sensor output is checked by filling a known volume with gas, by displacement of water. The flow rate is proportional to the change in volume over a fixed time interval.
The calibration of the flow sensor found it had a different reading if vertical compared to horizontal. It was calibrated in the position used for the appropriate test. Calibration data is held on SVN.
A post-test audit was carried out by requesting the reading from the CO2 sensor immediately prior to the scrubber, and confirming that the sum of the inhaled PPCO2 and the injected PPCO2 is shown on that sensor. All sensor data was stored for the entire sequence of these tests.
4 TEST RESULTS
4.1 Standard conditions: Vertical, Gas temperature 32C, water 4C The power of the breathing gas heating was set to 108W as 34V^2/11ohm ..(35V^2/11ohm)*97%, to achieve an exhaled gas temperature of 32C. During the temperature stabilisation period of 50 minutes, the power was 106W.
The test includes the following steps.
1. Run Re-Br 2. Close chamber 3. Fill chamber via the breathing loop by air up to 5 bar absolute pressure (40 msw) 4. Apply CO2 of 1.6 lpm via the mass control valve to BM 5. Open CO2 measuring channel to set the flow rate of ~0.16 lpm. Flow is checked
using the flow sensor 6. Apply the external air flow into the breathing loop to compensate the gas flows from
the breathing loop into the CO2 measuring channel. 7. Testing until measured CO2 surface corrected, exceeds 4% 8. Stop Re-Br 9. Stop CO2 injection. 10. Decrease chamber pressure down to atmospheric pressure. 11. Stop test data recording. 12. Open chamber
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 10 of 41
0 50 100 150 200 250 3001
1.5
2
2.5
3
3.5
4
4.5
5
Time, min
Pre
ssur
e, b
ar
Ambient and loop pressure against Time
Amb pressureLoop pressure
Figure 4-1. Ambient pressure against time.
0 50 100 150 200 250 300-1
0
1
2
3
4
5
Time, min
CO
2, %
CO2 against Time
Figure 4-2. . CO2 in the breathing loop against time. SCRB time life (2mbar PPCO2) = 77 min; SCRB time life (5mbar PPCO2) = 108 min or 1 hour 48 min; SCRB time life (20mbar PPCO2) = 165 min or 2 hour 45 min;
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 11 of 41
0 50 100 150 200 250 3000
5
10
15
20
25
30
35
Time, min
Tem
pera
ture
, C
Heater, BM and water temperature against Time
HeaterBMWater
Figure 4-3. Temperature water, heater, and inside BM.
0 50 100 150 200 250 3005
10
15
20
25
30
Time, min
Tem
pera
ture
, C
Temperature against Time
BL tempSCRB temp
Figure 4-4. Temperature in rebreather and scrubber stick.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 12 of 41
0 50 100 150 200 250 30035
40
45
50
55
60
65
70
75
Time, min
Hum
idity
, %Humidity against Time
Humidity in the breathing machine
0 50 100 150 200 250 300
60
65
70
75
80
85
90
95
100
Time, min
Hum
idity
, %
Humidity inside the membrane bell against Time
Humidity immediately prior to the scrubber.
Figure 4-5. Humidity in the breathing machine and the rebreather against time. The
average humidity in the breathing machine is 55%.
0 50 100 150 200 250 3000.02
0.04
0.06
0.08
0.1
0.12
0.14
0.16
0.18
0.2
Time, min
Flow
, lpm
Gas flow in measuring channel against Time
Figure 4-6. Gas flow from the breathing loop exhale into the CO2 measuring
channel against time.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 13 of 41
4.2 Colder water: Vertical, 32C exhale, but ambient water just 1C, test 3 scrubbers The exhale gas temperature was set to 32C (power to the breathing gas heating was set to 110W as 35V^2/11ohm*99). The water was cooled to 1C - there was some lag at the start of the test. Three scrubber samples were tested.
The test includes the following steps.
1. Run Re-Br 2. Close chamber 3. Fill chamber via the breathing loop by air up to 5 bar absolute pressure (approx 40
msw) 4. Apply CO2 of 1.6 lpm via the mass control valve to BM 5. Open CO2 measuring channel to set the flow rate of ~0.16 lpm. Flow is checked
using the flow sensor. 6. Apply the external air flow into the breathing loop to compensate the gas flows from
the breathing loop into the CO2 measuring channel. 7. Testing until measured CO2 surface corrected, exceeds 4% 8. Stop Re-Br 9. Stop CO2 injection. 10. Decrease chamber pressure down to atmospheric pressure. 11. Stop test data recording. 12. Open chamber
4.2.1 First scrubber sample
0 20 40 60 80 100 120 140 160 180 200
1
1.5
2
2.5
3
3.5
4
4.5
5
Time, min
Pre
ssur
e, b
ar
Ambient and loop pressure against Time
Amb pressureLoop pressure
Figure 4-7. Ambient pressure against time.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 14 of 41
0 50 100 150 200 250-0.5
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
Time, min
CO
2, %
CO2 against Time
Figure 4-8. . CO2 in the breathing loop against time. SCRB time life (2mbar PPCO2) = 58 min; SCRB time life (5mbar PPCO2) = 83 min or 1 hour 23 min; SCRB time life (20mbar PPCO2) = 144 min or 2 hour 24 min;
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 15 of 41
0 50 100 150 200 2500
5
10
15
20
25
30
35
Time, min
Tem
pera
ture
, C
Heater, BM and water temperature against Time
HeaterBMWater
Figure 4-9. Temperature water, heater, and inside B
0 50 100 150 20010
15
20
25
30
35
Time, min
Tem
pera
ture
, C
Temperature against Time
BL tempSCRB temp
Figure 4-10. Temperature in rebreather and scrubber stick.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 16 of 41
0 50 100 150 200 25050
55
60
65
70
75
80
85
Time, min
Hum
idity
, %Humidity against Time
Humidity in the breathing machine
0 20 40 60 80 100 120 140 160 180 200
88
90
92
94
96
98
100
Time, min
Hum
idity
, %
Humidity inside the membrane bell against Time
Humidity inside the rebreather immediately before the scrubber.
Figure 4-11. Humidity in the breathing machine and the rebreather against time. The average humidity in the breathing machine during the scrubber endurance is
75%.
0 20 40 60 80 100 120 140 160 180 2000.02
0.04
0.06
0.08
0.1
0.12
0.14
0.16
0.18
Time, min
Flow
, lpm
Gas flow in measuring channel against Time
Figure 4-12. Gas flow from the breathing loop exhale into the CO2 measuring
channel against time.
4.2.2 Second scrubber sample Note: The test is terminated on this run when the CO2 reaches gets 5.5 mbar.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 17 of 41
20 40 60 80 100 120 140 160
1.5
2
2.5
3
3.5
4
4.5
5
Time, min
Pre
ssur
e, b
ar
Ambient and loop pressure against Time
Amb pressureLoop pressure
Figure 4-13. Ambient pressure against time.
40 60 80 100 120 140 160
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
Time, min
CO
2, %
CO2 against Time
Figure 4-14. . CO2 in the breathing loop against time. CO2 injection starts at 25.5 min SCRB time life (2mbar PPCO2) = 63 min or 1 hour 3 min; SCRB time life (5mbar PPCO2) = 104 min or 1 hour 44 min;
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 18 of 41
40 60 80 100 120 140 160
5
10
15
20
25
30
Time, min
Tem
pera
ture
, C
Heater, BM and water temperature against Time
HeaterBMWater
Figure 4-15. Temperature water, heater, and inside BM.
40 60 80 100 120 140 1600
5
10
15
20
25
30
Time, min
Tem
pera
ture
, C
Temperature against Time
BL tempSCRB temp
Figure 4-16. Temperature in rebreather and scrubber stick.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 19 of 41
40 60 80 100 120 140 16045
50
55
60
65
70
75
Time, min
Hum
idity
, %
Humidity against Time
Humidity in the breathing machine (~72%)
40 60 80 100 120 140 16060
65
70
75
80
85
90
95
100
Time, min
Hum
idity
, %
Humidity inside the membrane bell against Time
Humidity inside the membrane bell (~91%).
Figure 4-17. Humidity in the breathing machine (~71%) and the rebreather against
time.
20 40 60 80 100 120 140 160 1800.085
0.09
0.095
0.1
0.105
0.11
0.115
0.12
0.125
0.13
Time, min
Flow
, lpm
Gas flow in measuring channel against Time
Figure 4-18. Gas flow from the breathing loop exhale into the CO2 measuring
channel against time.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 20 of 41
4.2.3 Third scrubber sample
0 20 40 60 80 100 120 140 160 180 200
1
1.5
2
2.5
3
3.5
4
4.5
5
Time, min
Pre
ssur
e, b
ar
Ambient and loop pressure against Time
Amb pressureLoop pressure
Figure 4-19. Ambient pressure against time.
0 50 100 150 200 250-0.5
0
0.5
1
1.5
2
2.5
3
Time, min
CO
2, %
CO2 against Time
Figure 4-20. . CO2 in the breathing loop against time. CO2 injection starts at 24 min SCRB time life (2mbar PPCO2) = 61 min or 1 hour 1 min; SCRB time life (5mbar PPCO2) = 96 min or 1 hour 36 min; SCRB time life (20mbar PPCO2) = 188 min or 3 hour 8 min;
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 21 of 41
0 50 100 150 200 2500
5
10
15
20
25
30
35
Time, min
Tem
pera
ture
, C
Heater, BM and water temperature against Time
HeaterBMWater
Figure 4-21. Temperature water, heater, and inside BM.
0 50 100 150 2005
10
15
20
25
30
Time, min
Tem
pera
ture
, C
Temperature against Time
BL tempSCRB temp
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 22 of 41
4.3 Colder water: Vertical, 32C exhale, but ambient water just 1C, short interrupt of CO2 injection The CO2 injection is interrupted at 167 min for 4 min to calibrate the CO2 response in measurement channel.
20 40 60 80 100 120 140 160 180 200 2201
1.5
2
2.5
3
3.5
4
4.5
5
Time, min
Pre
ssur
e, b
ar
Ambient and loop pressure against Time
Amb pressureLoop pressure
Figure 4-22. Ambient pressure against time.
0 50 100 150 200 250-0.5
0
0.5
1
1.5
2
2.5
3
3.5
Time, min
CO
2, %
CO2 against Time
Figure 4-23. . CO2 in the breathing loop against time. CO2 injection starts at 8 min SCRB time life (2mbar PPCO2) = 47 min; SCRB time life (5mbar PPCO2) = 77 min or 1 hour 17 min; SCRB time life (20mbar PPCO2) = 192 min or 3 hour 128 min;
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 23 of 41
0 50 100 150 200 2500
5
10
15
20
25
30
35
Time, min
Tem
pera
ture
, C
Heater, BM and water temperature against Time
HeaterBMWater
Figure 4-24. Temperature water, heater, and inside BM
20 40 60 80 100 120 140 160 180 200 22010
15
20
25
30
35
Time, min
Tem
pera
ture
, C
Temperature against Time
BL tempSCRB temp
Figure 4-25. Temperature in rebreather and scrubber stick.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 24 of 41
0 50 100 150 200 25050
55
60
65
70
75
80
Time, min
Hum
idity
, %
Humidity against Time
Humidity in the breathing machine (~71%)
0 50 100 150 20050
55
60
65
70
75
80
85
90
Time, min
Hum
idity
, %
Humidity inside the membrane bell against Time
Humidity inside the membrane bell (~83%).
Figure 4-26. Humidity in the breathing machine (~71%) and the rebreather against
time.
4.4 Effect of using a scrubber for two dives in very cold conditions This test considered the effects of two ice dives, in very cold conditions.
From other testing, DL know that interrupting the test seems to extend the scrubber life. This is particularly the case if the scrubber is rotated about its axis by 180 degrees. This test in cold conditions did not rotate the scrubber, but simply ran two dives with an 18 hour interval between them.
To recreate the worst possible diving conditions, the exhale gas temperature was reduced from the usual 32C to 7C for the first dive, and 13C for the second dive.
During the first stage air is used as a diluent gas, the heater of the exhale gas is off.
During the second stage the pure nitrogen fills the breathing loop to compensate the ambient pressure, the heater of the exhale gas is on. The power of the breathing gas heating is 26W as 24V^2/11ohm/2 to achieve 13C.
The first stage of the test includes the following steps.
1. Run Re-Br 2. Close chamber 3. Fill chamber via the breathing loop by air up to 5 bar absolute pressure (40 msw) 4. Apply CO2 of 1.6 lpm via the mass control valve to BM 5. Open CO2 measuring channel to set the flow rate of ~0.16 lpm. Flow is checked
using the flow sensor 6. Apply the external air flow into the breathing loop to compensate the gas flows from
the breathing loop into the CO2 measuring channel. 7. Testing in one hour 8. Stop Re-Br 9. Stop CO2 flow input into the rebreather 10. Decrease chamber pressure down to atmospheric pressure. 11. Stop test data recording.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 25 of 41
12. Open chamber
The second stage steps are
13. Run Re-Br 14. Close chamber 15. Fill chamber via the breathing loop by N2 up to 5 bar absolute pressure (40 msw) 16. Apply CO2 of 1.6 lpm via Porter flow controller to BM 17. Open CO2 measuring channel with flow rate of ~0.16 lpm. Flow is checked using the
flow sensor 18. Apply external N2 flow into the breathing loop to compensate flow into the CO2
measuring channel. 19. Testing until CO2 in the measuring channel exceeds 4%, surface corrected 20. Stop CO2 injected into the breathing loop 21. Drop the chamber pressure down to the atmospheric pressure 22. Stop Re-Br 23. Stop test data recording. 24. Open chamber
0 10 20 30 40 50 601
1.5
2
2.5
3
3.5
4
4.5
5
5.5
6
Time, min
Pre
ssur
e, b
ar
Ambient pressure against Time
First stage
40 60 80 100 120 140 160 180 2000.5
1
1.5
2
2.5
3
3.5
4
4.5
5
5.5
Time, min
Pre
ssur
e, b
ar
Ambient and loop pressure against Time
Amb pressureLoop pressure
Second stage
Figure 4-27. Ambient pressure against time.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 26 of 41
50 100 150 200-1
0
1
2
3
4
5
Time, min
CO
2, %
CO2 against Time
Figure 4-28. .Second stage of the test. CO2 in the breathing loop against time.
Reminding the scrubber endurance during the first stage of the test SCRB time life (2mbar PPCO2) = 59 min; SCRB time life (5mbar PPCO2) = 91 min or 1 hour 31
min; SCRB time life (20mbar PPCO2) = 130.5 min or 2 hour 10.5 min;
0 10 20 30 40 50 600.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
1.1
Time, min
PP
O2,
ATA
PPO2 against Time
Diluent gas is air
50 100 150 2000
0.05
0.1
0.15
0.2
0.25
Time, min
PP
O2,
ATA
PPO2 against Time
Diluent gas is pure nitrogen
Figure 4-29. PPO2 in the breathing loop
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 27 of 41
0 10 20 30 40 50 602
3
4
5
6
7
8
9
Time, min
Tem
pera
ture
, C
Heater, BM and water temperature against Time
HeaterBMWater
50 100 150 200
4
6
8
10
12
14
16
18
20
22
24
Time, min
Tem
pera
ture
, C
Heater, BM and water temperature against Time
HeaterBMWater
Figure 4-30. Temperature water, heater, and inside BM.
0 10 20 30 40 50 60
6
8
10
12
14
16
18
20
22
24
Time, min
Tem
pera
ture
, C
Temperature against Time
BL tempSCRB temp
60 80 100 120 140 160 1805
10
15
20
25
Time, min
Tem
pera
ture
, C
Temperature against Time
BL tempSCRB temp
Figure 4-31. Temperature in rebreather and scrubber stick.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 28 of 41
5 10 15 20 25 30 35 40 45 50 55
50
60
70
80
90
100
Time, min
Hum
idity
, %
Humidity inside the membrane bell against Time
Figure 4-32. Humidity in the rebreather. The humidity is below 80% for the beginning of the dive because the gas temperature is below the dew point.
4.5 Reducing the exhaled gas temperature from 32C to 21C The power of the breathing gas heating was set to 52W (24V^2/11ohm), to achieve a desired breathing gas temperature of 21C. This is to assess the effect of additional gas cooling from the diver, due to longer hoses, narrower hoses or stainless steel hoses.
The test includes the following steps.
1. Run Re-Br 2. Close chamber 3. Fill chamber via the breathing loop by air up to 5 bar absolute pressure (40 msw) 4. Apply CO2 of 1.6 lpm via the mass control valve to BM 5. Open CO2 measuring channel to set the flow rate of ~0.16 lpm. Flow is checked
using the flow sensor 6. Apply the external air flow into the breathing loop to compensate the gas flows from
the breathing loop into the CO2 measuring channel. 7. Testing until CO2 in the breathing loop exceeds 4%, surface corrected 8. Stop Re-Br 9. Stop injection of CO2. 10. Decrease chamber pressure down to atmospheric pressure. 11. Stop test data recording. 12. Open chamber
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 29 of 41
0 20 40 60 80 100 120 140 1601
1.5
2
2.5
3
3.5
4
4.5
5
Time, min
Pre
ssur
e, b
ar
Ambient and loop pressure against Time
Amb pressureLoop pressure
Figure 4-33. Ambient pressure against time.
0 50 100 150-1
0
1
2
3
4
5
Time, min
CO
2, %
CO2 against Time
Figure 4-34. . CO2 in the breathing loop against time. SCRB time life (2mbar PPCO2) = 42 min; SCRB time life (5mbar PPCO2) = 69.3 min or 1 hour 9.3 min; SCRB time life (20mbar PPCO2) = 108.7 min or 1 hour 48.7 min;
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 30 of 41
0 50 100 1500
5
10
15
20
25
30
35
Time, min
Tem
pera
ture
, C
Heater, BM and water temperature against Time
HeaterBMWater
Figure 4-35. Temperature water, heater, and inside BM.
20 40 60 80 100 120 1405
10
15
20
25
30
Time, min
Tem
pera
ture
, C
Temperature against Time
BL tempSCRB temp
Figure 4-36. Temperature in rebreather and scrubber stick.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 31 of 41
20 40 60 80 100 120 140
60
65
70
75
80
85
90
95
100
Time, min
Hum
idity
, %
Humidity inside the membrane bell against Time
Figure 4-37. Humidity in rebreather against time, measured immediately prior to the
scrubber.
4.6 Horizontal Position: 32C exhale, 1C water, test three scrubbers
The horizontal position was tested using three scrubbers, under as similar conditions as could be achieved in a reasonable time, to assess the effect of a horizontal attitude in the worst case diving conditions.
The test depth and CO2 flow rate was otherwise set to be the same as for the test of the rebreather in the vertical position in 1C ambient water, with 32C exhale temperature.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 32 of 41
4.6.1 First scrubber cartridge sample
0 50 100 150 200
1
1.5
2
2.5
3
3.5
4
4.5
5
Time, min
Pre
ssur
e, b
ar
Ambient and loop pressure against Time
Amb pressureLoop pressure
Figure 4-38. Ambient pressure against time.
0 50 100 150 200 2500
0.5
1
1.5
2
2.5
3
Time, min
CO
2, %
CO2 against Time
Figure 4-39. . CO2 in the breathing loop against time. CO2 injection starts at 16 min. SCRB time life (2mbar PPCO2) = 54 min; SCRB time life (5mbar PPCO2) = 90 min or 1 hour 30 min; SCRB time life (20mbar PPCO2) = 169 min or 2 hour 49 min;
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 33 of 41
0 50 100 150 200 2500
5
10
15
20
25
30
35
Time, min
Tem
pera
ture
, C
Heater, BM and water temperature against Time
HeaterBMWater
Figure 4-40. Temperature water, heater, and inside BM.
20 40 60 80 100 120 140 160 180 200 2205
10
15
20
25
30
35
Time, min
Tem
pera
ture
, C
Temperature against Time
BL tempSCRB temp
Figure 4-41. Temperature in rebreather and scrubber stick.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 34 of 41
20 40 60 80 100 120 140 160 180 200 22065
70
75
80
85
90
95
Time, min
Hum
idity
, %Humidity against Time
Humidity in the breathing machine (~80%)
0 50 100 150 20075
80
85
90
95
100
105
Time, min
Hum
idity
, %
Humidity inside the membrane bell against Time
Humidity inside the membrane bell (~96%).
Figure 4-42. Humidity in the breathing machine and the rebreather against time..
0 50 100 150 200 2500.06
0.08
0.1
0.12
0.14
0.16
0.18
Time, min
Flow
, lpm
Gas flow in measuring channel against Time
Figure 4-43. Gas flow from the breathing loop exhale into the CO2 measuring channel against time: this is the measurement monitoring channel, not the CO2
injection channel. This is typical of that in the other tests and is provided here by way of illustration.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 35 of 41
4.6.2 Second scrubber sample
0 50 100 150 2001
1.5
2
2.5
3
3.5
4
4.5
5
Time, min
Pre
ssur
e, b
ar
Ambient and loop pressure against Time
Amb pressureLoop pressure
Figure 4-44. Ambient pressure against time.
0 50 100 150 200 250-0.5
0
0.5
1
1.5
2
2.5
3
Time, min
CO
2, %
CO2 against Time
Figure 4-45. . CO2 in the breathing loop against time. CO2 injection starts at 11 min. SCRB time life (2mbar PPCO2) = 71 min or 1 hour and 11 min; SCRB time life (5mbar PPCO2) = 122 min or 2 hour 2 min; SCRB time life (20mbar PPCO2) = 196 min or 3 hour 16 min;
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 36 of 41
0 50 100 150 200
15
20
25
30
35
Time, min
Tem
pera
ture
, C
Temperature against Time
BL tempSCRB temp
Figure 4-46. Temperature in rebreather and scrubber stick.
0 50 100 150 200 25055
60
65
70
75
80
85
90
Time, min
Hum
idity
, %
Humidity against Time
Humidity in the breathing machine (~77%)
0 50 100 150 200
55
60
65
70
75
80
85
90
95
100
Time, min
Hum
idity
, %
Humidity inside the membrane bell against Time
Humidity inside the membrane bell (~95%).
Figure 4-47. Humidity in the breathing machine and the rebreather against time..
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 37 of 41
4.6.3 Third scrubber sample
20 40 60 80 100 120 140 160 180 200 220
1
1.5
2
2.5
3
3.5
4
4.5
5
Time, min
Pre
ssur
e, b
ar
Ambient and loop pressure against Time
Amb pressureLoop pressure
Figure 4-48. Ambient pressure against time.
0 50 100 150 200 2500
0.5
1
1.5
2
2.5
3
Time, min
CO
2, %
CO2 against Time
Figure 4-49. . CO2 in the breathing loop against time. SCRB time life (2mbar PPCO2) = 68 min or 1 hour 8 min; SCRB time life (5mbar PPCO2) = 96 min or 1 hour 36 min; SCRB time life (20mbar PPCO2) = 181 min or 3 hours 1 min;
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 38 of 41
0 50 100 150 200 25040
45
50
55
60
65
70
75
80
85
Time, min
Hum
idity
, %Humidity against Time
Humidity in the breathing machine (~65%)
20 40 60 80 100 120 140 160 180 200 220
65
70
75
80
85
90
95
100
Time, min
Hum
idity
, %
Humidity inside the membrane bell against Time
Humidity inside the membrane bell (~91%).
Figure 4-50. Humidity in the breathing machine and the rebreather against time..
5 MAXIMUM INHALED CO2 LIMITS There is a problem that some standards define CO2 breakthrough at such a low level that there are no symptoms, so divers do not treat the figure seriously and go onto exceed scrubber durations without having any good data points on what the limit actually is. For example, EN14143:2003 defines a maximum inhaled CO2 limit of 0.5kPa: the U.S. Occupational Safety and Health Administration advise that average exposure for healthy adults during an eight-hour work day should not exceed 0.5kPa. As a scrubber breakthrough figure this is just not meaningful.
For exposure up to 10 minutes, the U.S. National Institute for Occupational Safety and Health (NIOSH) and American Conference of Government Industrial Hygienists (ACGIH) limit is 3kPa.
NORSOK U101:1999 takes a more pragmatic and sensible approach, giving a limit of 1kPa generally, and for short periods 2 kPa. A scrubber breakthrough event is a short period, so the limit on scrubber endurance of a 2 kPa breakthrough should be one that divers can accept.
The 2kPa limit is used by Deep Life Ltd. For certification purposes, the 0.5 kPa and 1.0 kPa
6 CORRELATION WITH THIRD PARTY TEST RESULTS Tests were carried out by Micropore at their Newark facility using a dumb head (i.e. single scrubber Open Safety Apocalypse version of the rebreather) and these correlate to within 4 minutes of the times measured by Deep Life at the 0.5 kPa calibration point used by Micropore, with 4C +/- 1.2C water and the rebreather in the vertical position, at 130 fsw.
Temperature, C Scrubber endurance, minutes Mean and SD of group at 0 .5 kPa
Water Inhale
at DSV
Exhale at
DSV
0.5 kPa
1.0 kPa
1.5 kPa 2 kPa
4.8 29.64 20.39 93.25 134.75 189.25 224.75 104.42
4.8 29.27 20.06 106.00 135.50 161.25 182.75 10.47
5.2 30.7 20.39 114.00 158.50 193.00 222.5
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 39 of 41
7 CONCLUSION The table below tablulates the test results. Using the NORSOK U-101 2kPa limit, the scrubber life in 4C of 2 hours 45 mins seems to be a reliable figure that is lost only if the exhale temperature reduces excessively. There is no loss in performance from an interrupt of 18 hours between two dives: in fact, it is in line with other test results showing the scrubber performance improves slightly.
Table 1. Scrubber endurance test results
Temperature, C Humidity, %
CSCRB endurance, min
Water Breathing machine
Exhale Breathing loop
Scrubber Breathing machine
0.2 kPa 0.5 kPa 2kPa
Notes
4 15 32 12 25 55 77 108 165 Standard vertical test conditions. 2 hours 45
min endurance in NORSOK.
1 15 32 13 24 72 58 83 144 Standard vertical test conditions other than
water is 1C.
1 17 32 13 25 72/91 63 104 - 2nd vertical 1C test
1 16.5 32 13 25 72/93 61 96 188 3rd vertical 1C test
1 17 32 13 25 71/83 47 77 192 Vertical 1C test interrupted at 167 min
for 4 min.
5 6 10 12 18 59 91 130.5 Vertical very cold expired gas, but 18-hours-break between
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 40 of 41
two test stages
4 12 21 10 20 42 69.3 108.7 Vertical reduced exhale temperature.
2 12 32 13 25 80/96 54 90 169 Horizontal std to cold test (first test).
3 13 32 13 26 77/95 71 122 196 2nd std horizontal test
4 14 32 13 26 65/91 68 96 181 3rd std horizontal test
The NORSOK U-101 2 kPa limit that is a practical limit for divers, is 2 hours 45 minutes. This is extended by around 20 minutes when the diver is horizontal. There is a small spread of results, so for conservatism, the manufacture will state the duration of the scrubber is 2 hours 35 minutes at 40msw, with a 40lpm RMV, using air as the diluent.
The reduction of the water temperature to 1C does not cause a statistically significant change in the duration.
A reduction in the exhale temperature does reduce the scrubber duration substantially. This underlines the importance of the biased counterlungs, and large bore diameters in the Open Revolution rebreathers to prevent gas cooling from the diver to the scrubber.
From other tests, there is no difference in scrubber duration between the instrumented eCCR head used for the Apollo Incursion and the dumb head used for the Open Safety Apocalypse rebreather product. The dual scrubber rebreather in the Open Revolution series used for the Bell Diver and Cave Diver products is the subject of a separate report.
Open Publication
DV_OR_ScrubberEndurance_081124.doc Rev C Page 41 of 41