Electrical Safety - 8 (1)

Embed Size (px)

Citation preview

  • 8/9/2019 Electrical Safety - 8 (1)

    1/52

    Electrical Safety

    Medical Instrumentation Application and Design, 4th

    Edition, Chapter 14

    John G. Webster, Univ. of Wisconsin, Madison

    ISBN: 978-0-471-67600-3

  • 8/9/2019 Electrical Safety - 8 (1)

    2/52

    Taught Matter of Lectures

    Introduction

    Basic Theory of Measurements

    Beginnings of Basic Sensors

    Sensors [MEMS]

    Signals and Noise

     Amplifiers of SignalsConnection and Protection of Signals

    Data Acquisition and Data Converters

    Electric Safety in Medical Systems

  • 8/9/2019 Electrical Safety - 8 (1)

    3/52

    Electrical Safety

    Agenda

    • Introduction

    • Physiological Effects of Electricity

    • Susceptibility Parameters

    • Distribution of Electric Power 

    • Macroshock Hazards

    • Microshock Hazards• Electrical-Safety Codes and Standards

    • Approaches to Protection Against Shock

    • Power Distribution

    • Equipment Design

    • Testing the Electric System

    • Tests of Electric Appliances

    • Conclusion

  • 8/9/2019 Electrical Safety - 8 (1)

    4/52

    Introduction

    ES as elemetary protection

    • Medical technology has improved health care in ALL medical

    specialties, with rising complexity

    • More than 10,000 device-related patient injuries

    • Most patient injuries are attributable to improper use

    • Medical personnel rarely read user manuals until a problem has

    occurred

    Result: Medical instrumentation safety

    • Safe Design

    • Safe Use... Safe EVERYTHING (one of the most regulated industrial market)

    ES is one of the basic protection mechanisms for patient, operater, and

    third persons and part of this chapter 

  • 8/9/2019 Electrical Safety - 8 (1)

    5/52

    Physiological effects

    Current can heal and harm

    • For a physiological effect the body must become part of an electrical

    circuit

    • Three phenomena occur when el. current flows

    1. El. stimulation of excitable tissue (muscle, nerve)

    2. Resistive heating of tissue

    3. Electrochemical burns

    • Further consideration are based on the following parameters

    • Human body with contact to el. circuit at left and right hand

    • Body weight: 70 kg•  Applied current time: 1 s to 3 s

    • Current frequency: 60 Hz

  • 8/9/2019 Electrical Safety - 8 (1)

    6/52

    Figure 14.1 Physiological effects of electricity Threshold or estimated mean values are given

    for each effect in a 70 kg human for a 1 to 3 s exposure to 60 Hz current applied via copper

    wires grasped by the hands.

    Physiological effects

    Current can heal and harm

    6

    5

    43

    21

  • 8/9/2019 Electrical Safety - 8 (1)

    7/52

    Physiological effects

    Threshold of Perception (1)

    • Current density is just large enough to excite nerve endings in the

    skin

    • Subject feels tingling sensation

    • Lowes values with moistered hands (decreases contact resistance)

    • 0.5 mA at 60 Hz

    • 2 mA to 10 mA DC

    • The subject meight feel a slight warming

  • 8/9/2019 Electrical Safety - 8 (1)

    8/52

    Physiological effects

    Let-go Current (2)

    • The let-go current is defined as the maximal current at which the

    subject can withdraw voluntarily

    • For higher current nerves and muscles are vigorously stimulated

    • Involuntary contraction or reflex withdrawals may cause secondary

    physical injuries (falling off the ladder)

    • The minimal threshold for the let-go current is 6 mA

  • 8/9/2019 Electrical Safety - 8 (1)

    9/52

    Physiological effects

    Respiratory Paralysis, Pain, Fatigue (3)

    • Higher current causes involuntary contraction of muscles and

    stimulation of nerves what can lead to pain and cause fatigue

    Example: stimulation of respiratory muscles lead to involuntary

    contraction with the result of asphyxiation if current is not interrupted

    Of course, today’s ethics commission would never allow these

    experiments on human beings.

  • 8/9/2019 Electrical Safety - 8 (1)

    10/52

    Physiological effects

    Ventricular Fibrillation (4)

    • The heart is especially susceptible to electric current.

    • Just 75 mA to 400 mA (AC) can rapidly disorganize the cardiac

    rhythm and death occurs within minutes

    • Only a brief high-current pulse from a defibrillator can depolarize all

    the cells of the heart muscle simultaneously

    • Within the U.S. occur approximately 1,000 death per year due to cord-

    connected appliances

  • 8/9/2019 Electrical Safety - 8 (1)

    11/52

    Physiological effects

    Sustained Myocardial Contraction (5)

    • When current is high enough to stimulate the entire heart muscle, it

    stops beating

    • Usually the heart-beat ensues when the current is interrupted

    • Minimal currents range from 1 A to 6 A (AC), like used in defibrillators

  • 8/9/2019 Electrical Safety - 8 (1)

    12/52

    Physiological effects

    Burns and Physical Injury (6)

    • Resistive heating cause burns

    • Current can puncture the skin

    • Brain and nerve tissue may lose all functional excitability

    • Simultaneously stimulated muscles may contract strong enough to

    pull the attachment away from the bone or bread the bone

  • 8/9/2019 Electrical Safety - 8 (1)

    13/52

    Susceptibility Parameters

    Introduction

    The current needed to produce each effect depends on these parameters

    • Threshold of Perception and Let-Go Variability

    • Frequency

    • Duration

    • Body Weight (and gender)

    • Points of Entry• Macroshock

    • Microshock

  • 8/9/2019 Electrical Safety - 8 (1)

    14/52

    Figure 14.2 Distributions of perception thresholds and let-go currents These data depend on

    surface area of contact (moistened hand grasping AWG No. 8 copper wire). (Replotted from C. F.Dalziel, "Electric Shock," Advances in Biomedical Engineering, edited by J. H. U. Brown and J.

    F. Dickson III, 1973 3, 223-248.)

    Susceptibility Parameters

    Variability of threshold and Let-go current

  • 8/9/2019 Electrical Safety - 8 (1)

    15/52

    • Let-go current versus frequency

    Minimal let-go currents occur for commercial power-line frequencies (50 Hz to 60

    Hz)

    For frequencies below 10 Hz let-go current rises again (muscle can relax)

    Figure 14.3 Let-go current versus frequency

    Percentile values indicate variability of let-go currentamong individuals. Let-go currents for women are

    about two-thirds the values for men. (Reproduced,

    with permission, from C. F. Dalziel, "Electric

    Shock," Advances in Biomedical Engineering, edited

     by J. H. U. Brown and J. F. Dickson III, 1973, 3,223–248.)

    Susceptibility Parameters

    Frequency

  • 8/9/2019 Electrical Safety - 8 (1)

    16/52

    Geddes and Baker (1989) presented the excitation behavior of

    myocardial cells by a lumped parallel RC circuit that represents the

    resistance and capacitance of the cell membrane.This model determines the cell excitation thresholds that exceed about

    20 mV for varying rectangular pulse duration d by assigning the

    rheobase currents Ir and cell membrane time constant τ=RC.

    The strength-duration equation

    For a short duration:

    Stimulation current Id is inversely related to the pulse duration d (Figure

    14.4)

    Susceptibility Parameters

    Duration

  • 8/9/2019 Electrical Safety - 8 (1)

    17/52

    Figure 14.4 Normalized analytical strength–duration curve for current I, charge Q, and energy U.

    The x axis shows the normalized duration of d/τ (From Geddes, L. A., and L. E. Baker, Principlesof Applied Biomedical Instrumentation, 3rd ed. New York: John Wiley & Sons, 1989).

    Susceptibility Parameters

    Duration

  • 8/9/2019 Electrical Safety - 8 (1)

    18/52

    • Several studies (animals) show a clear dependency of the fibrillating

    current to the body weight (Figure 14.5)

    50 mA rms for 6kg dogs to 130 mA rms for 24 kg dogs

    Figure 14.5 Fibrillation current versus shock

    duration. Thresholds for ventricular fibrillation in

    animals for 60 Hz ac current. Duration of current

    (0.2 to 5 s) and weight of animal body were varied.

    (From L. A. Geddes, IEEE Trans. Biomed. Eng.,

    1973, 20, 465–468. Copyright 1973 by the Institute

    of Electrical and Electronics Engineers. Reproduced

    with permission.)

    Susceptibility Parameters

    Body weight

  • 8/9/2019 Electrical Safety - 8 (1)

    19/52

    Figure 14.6 Effect of

    entry points on current

    distribution (a)

    Macroshock, externally

    applied current spreads

    throughout the body, (b)

    Microshock, all the current

    applied through anintracardiac catheter flows

    through the heart. (From F.

    J. Weibell, "Electrical

    Safety in the Hospital,"

    Annals of BiomedicalEngineering, 1974, 2, 126– 

    148.)

    • Macroshock: only a small fraction of the total current flows through

    the heart. Magnitude to harm the heart is far greater

    • Microshock: all the current applied flows through the heart

    Susceptibility Parameters

    Points of Entry

  • 8/9/2019 Electrical Safety - 8 (1)

    20/52

    • 60 Hz for 5 s to a ventricular pacing catheter during implantable

    cardioverter-defibrillator implant testing in 40 patients

    • Result

    Intermittent capture with a minimum current of 20 µA

    Continuous capture with a minimum current of 49 µA

    • Resulting Regulation

    The widely accepted safety limit to prevent microshocks is 10 µA

    Susceptibility Parameters

    Points of Entry - Example

  • 8/9/2019 Electrical Safety - 8 (1)

    21/52

    Figure 14.7. Percentile plot of thresholds for continuous

    capture and VF (or sustained VT). Cumulative percent of

     patients is shown on abscissa and root-mean-square AC

    current (in µA) on ordinate. Squares denote unipolar data;circles, bipolar data. Solid symbols identify data from patients

    in whom the only clinical arrhythmia was atrial fibrillation

    (AF). Top, Thresholds for continuous capture. Current

    strength of 50 µA caused continuous capture in 5 patients

    (12%) with unipolar AC and in 9 (22%) with bipolar AC(P=0.49). Bottom, Thresholds for sustained VT/VF. These

     plots do not reach 100% because sustained-VT/VF thresholds

    exceeded maximum output of stimulator in 6 patients (15%)

    with bipolar AC and 8 (20%) with unipolar AC. From

    Swerdlow, C. D., W. H. Olson, M. E. O’Connor, D. M. Gallik,

    R. A. Malkin, M. Laks, “Cardiovascular collapse caused by

    electrocardiographically silent 60-Hz intracardiac leakage

    current – Implications for electrical safety.” Circulation.,

    1999, 99, 2559–2564.

    Susceptibility Parameters

    Points of Entry - Example

  • 8/9/2019 Electrical Safety - 8 (1)

    22/52

    Distribution of Electric Power 

    Introduction

    • Electric Power is needed in health-care facilities not only for medical

    devices but also for any other electrical equipment like lightning, air

    condition, telephone, television etc.

    • BUT

    Medical devices underlie special safety regulations as they might stay

    in special contact to and with patients, applicants and third persons1. Overvoltage protection

    2. Special ground

    • Example

    A lightning causes an overvoltage at the public power supply. The

    overvoltage is transferred directly to the patients heart by applied

    ECG-Electrodes.

    => Over voltage protection

  • 8/9/2019 Electrical Safety - 8 (1)

    23/52

    Figure 14.8 Simplified electric-power distribution for 115 V circuits. Power frequency is 60

    Hz.

    • Health-care facilities need an additional (green) ground path for all

    receptacles redundant to the metal (white) ground path

    Distribution of Electric Power 

    El. power-distribution from grid to receptacles

  • 8/9/2019 Electrical Safety - 8 (1)

    24/52

    Figure 14.9 Power-isolation-transformer system with a line-isolation monitor to detect

    ground faults.

    • Isolation transformer like this example protect systems from ground

    faults

    • The Line-isolation monitor must be used to detect the occurrence ofground faults

    Distribution of Electric Power 

    Isolated-power systems

  • 8/9/2019 Electrical Safety - 8 (1)

    25/52

    Macroshock Hazards

    Macroshock = current spreads through the body

    • Two factors reduce danger in case of an electric shock

    1. High skin resistance (15kOhm to 1 MOhm at 1 cm2)

    2. Spatial distribution

    • Many medical devices

    • Reduce the skin resistance with ionic gel (good electrode contact), or 

    • Bypass the natural protection by bypassing the skin (thermometer in the mouth,

    intravenous catheters, etc.)

    • Many fluids conduct electricity (blood, urine, intravenous solution,

    etc.)

    Result

    • Patients in medical-care facilities are much more susceptible tomacroshocks

  • 8/9/2019 Electrical Safety - 8 (1)

    26/52

    • Ground fault with short circuit to a metal chassis

    a. not grounded chassis macroshock

    b. grounded chassis safe

    Figure 14.10 Macroshock due to a

    ground fault from hot line to

    equipment cases for (a) ungrounded

    cases and (b) grounded chassis.

    Macroshock Hazards

    Protection

  • 8/9/2019 Electrical Safety - 8 (1)

    27/52

    Microshock Hazards

    Microshock = all current flows through the heart

    • Microshock accidents generally result from

    • leakage-currents in line-operated equipment

    • differences in voltage between grounded conductive surfaces due to large

    currents in the grounding system

    • Microshock currents can flow either into or out of the electric

    connection to the heart

    Result

    • Patient is only in danger of microshock if there is some electric

    connection to the heart

  • 8/9/2019 Electrical Safety - 8 (1)

    28/52

    • Leakage-current flows

    a. through the ground wire – no microshock occurs

    b. through the patient if he touches the chassis and has a grounded catheter etc.

    c. through the patient if he is touching ground and has a connected catheter etc.

    Figure 14.11 Microshock leakage-current

     pathways. Assume 100 μA of leakagecurrent from the power line to the

    instrument chassis, (a) Intact ground, and

    99.8 μA flows through the ground, (b)

    Broken ground, and 100 μA flows through

    the heart, (c) Broken ground, and 100 μAflows through the heart in the opposite

    direction.

    Microshock Hazards

    Protection

  • 8/9/2019 Electrical Safety - 8 (1)

    29/52

    Microshock Hazards

    Conductive Path to the Heart

    Specific types of electric connections to the heart can be identified

    • Epicardial or endocardial electrodes (i.e. temporary externalized

    pacemakers)< 1 Ohm

    • Electrodes for intracardiac electrogramm (ECT)

    < 1 Ohm

    • Liquid-filled catheters placed in the heart (i.e. measure bloodpreassure, withdraw blood samples, inject substances, etc.)

    usually 50 kOhm to 1 MOhm

    Internal resistance of the body is about 300 Ohm

  • 8/9/2019 Electrical Safety - 8 (1)

    30/52

    Microshock Hazards

    Conductive Path to the Heart

    Figure 14.11Microshock Assume 100

    leakage-current pathways. μA of leakage

    current from the power line to the

    instrument chassis, (a) Intact ground, and99.8 μA flows through the ground, (b)

    Broken ground, and 100 μA flows through

    the heart, (c) Broken ground, and 100 μA

    flows through the heart in the opposite

    direction.

    •• Ventricular fibrillation and pump failure thresholds vs. electrode area

  • 8/9/2019 Electrical Safety - 8 (1)

    31/52

    • Patient is connected to:

    • ECG monitor that grounds the

    right-leg electrode to reduce60 Hz interferce

    • Blood-pressure monitor that

    monitors the left-ventricular

    blood-pressure

    Figure 14.13 (a) Large ground-faultcurrent raises the potential of one

    ground connection to the patient. The

    microshock current can then flow out

    through a catheter connected to a

    different ground, (b) Equivalent circuit.Only power-system grounds are shown.

    Microshock Hazards

    Example of patient in the intensive-care unit

  • 8/9/2019 Electrical Safety - 8 (1)

    32/52

    Electrical-Safety Codes and Standards

  • 8/9/2019 Electrical Safety - 8 (1)

    33/52

    Basic Approaches to Protection against Shock

    There are two fundamental methods of protecting patients against shock

    1. Complete isolation and insulation from all grounded objects and all

    sources of electric current

    2. Same potential of all conducting surfaces within reach of the patients

    • Neither approach can be fully achieved in most practical

    environments, so some combination must usually suffice

    • Protection must include patient, applicants and third party persons

  • 8/9/2019 Electrical Safety - 8 (1)

    34/52

    • Low-resistance grounding system carry currents up to circuit-breaker

    ratings by keeping all conductive surfaces on the same potential (refer

    to Figure 14.10 and 14.11)• Patient-equipment grounding point

    • Reference grounding point

    • Connections for other patient-equipment

    Figure 14.14 Grounding system

    All the receptacle grounds and conductive surfaces inthe vicinity of the patient are connected to the

     patient-equipment grounding point. Each patient-

    equipment grounding point is connected to the

    reference grounding point that makes a single

    connection to the building ground.

    Protection: Power Distribution

    Grounding System

  • 8/9/2019 Electrical Safety - 8 (1)

    35/52

    • Ground-fault circuit interrupters disconnect the source power when a

    ground fault greater than about 6 mA occurs

    • GFCI senses differences in the in- an outgoing current

    • Most GFCI use differential transformer and solid-state circuitry

    • Most GFCI are protectors against macroshocks as they are usually

    not as sensitive as 10 µA or the medical equipment has a fault current

    greater than that

    Protection: Power Distribution

    Ground-Fault Circuit Interrupter (GFCI)

  • 8/9/2019 Electrical Safety - 8 (1)

    36/52

    Protection: Power Distribution

    Example of a GFCI

    Figure 14.15 Ground-fault circuit

    interrupters (a) Schematic diagram of a

    solid-state GFCI (three wire, two pole,

    6 mA). (b) Ground-fault current versustrip time for a GFCI. [Part (a) is from

    C. F. Dalziel, "Electric Shock," in

    Advances in Biomedical Engineering,

    edited by J. H. U. Brown and J. F.

    Dickson III, 1973, 3: 223–248.]

  • 8/9/2019 Electrical Safety - 8 (1)

    37/52

    Protection: Equipment Design

    Introduction

    • Most failures of equipment ground occur at the ground contact or in

    the plug and cable

    • Molded plugs should be avoided because of invisible breaks

    • Strain-relief devices are recommended

    • No use of three-prong-to-two-prong adapters (cheater adapters)

    • Reduction of leakage current

    • Special use of low-leakage power cords

    • Capacitance-minimized design (special layout-design and usage of insulation)

    • Maximized impedance from patient leads to hot conductors and from patient

    leads to chassis ground

    • Double-Insulated equipment

    • Interconnection of all conducting surfaces

    • Separate layer of insulation to prevent contact with conductive surfaces (e.g.

    non conductive chassis, switch levers, knobs, etc.)

    • Operation at low voltages

    • Electrical isolation

  • 8/9/2019 Electrical Safety - 8 (1)

    38/52

    Protection: Equipment Design

    Introduction

    • Operation at low voltages (< 10 V)

    • Reduction of risk of macroshock with reduced operation voltage

    • Risk of microshock still exists

    • Electrical isolation with isolation amplifiers

    • Isolation amplifiers break the ohmic continuity of electric signals

    • Isolation amplifiers use different voltage sources and different grounds

    • Isolation voltage  νiso is rated from 1 kV to 10 kV without breakdown and

    described by the isolation-mode rejection ration (IMMR) (lightning example form thebeginning)

    • Next slides describe isolation amplifiers

  • 8/9/2019 Electrical Safety - 8 (1)

    39/52

    Three fundamental design methods

    1. Transformer isolation

    • Frequency-modulated or pulse-width-modulated carrier signal with small signal

    bandwidths

    • Possibility to transmit energy and/or information

    2. Optical isolation

    • Uses LED on source-side and photodiode on output-side

    • Very fast signal transmission possible, but no energy

    3. Capacitive isolation

    Protection: Equipment Design

    Electric Isolation

  • 8/9/2019 Electrical Safety - 8 (1)

    40/52

    Figure 14.16 Electrical

    isolation of patient leads to

     biopotential amplifiers (a)General model for an isolation

    amplifier, (b) Transformer

    isolation amplifier (Courtesy of

    Analog Devices, Inc., AD202).

    (c) Simplified equivalent circuitfor an optical isolator (Copyright

    © 1989 Burr-Brown

    Corporation. Reprinted in whole

    or in part, with the permission of

    Burr-Brown Corporation. BurrBrown ISO100). (d) Capacitively

    coupled isolation amplifier

    (Horowitz and Hill, Art of

    Electronics, Cambridge Univ.

    Press, Burr Brown ISO106).

    Protection: Equipment Design

    Electric Isolation Transformer 

  • 8/9/2019 Electrical Safety - 8 (1)

    41/52

    Protection: Equipment DesignElectric Isolation optical and capacitive methods

  • 8/9/2019 Electrical Safety - 8 (1)

    42/52

    • The best way to minimize hazards of microshocks is to isolate or

    eliminate electric connections to the heart

    • Fully insulated connectors for external cardiac pacemakers powered bybatteries

    • Blood-pressure sensors with triple insulation between the column of liquid, the

    sensor case, and the electric connections

    • Catheters with conductive walls all the way inside the patient to distribute theshock throughout the body (enlarged surface)

    Protection: Equipment Design

    Good Practice for isolated Heart Connections

  • 8/9/2019 Electrical Safety - 8 (1)

    43/52

    Testing the Electric System

    Introduction

    • Test equipment: electrical-safety analyzers

    • Testing the electric system

    • Receptacles

    • Grounding system

    • Isolated power system

    • Testing the electric appliance

    • Ground-pin-to-chassis resistance

    • Chassis leakage current

    • Leakage current in patient leads

  • 8/9/2019 Electrical Safety - 8 (1)

    44/52

    Testing the Electric System

    Electrical-Safety Analyzers

    • Wide product range of el.-safety analyzers is availablee.g. http://www.electricalsafetyanalyzers.com

    • Medical-facility power systems• Medical appliances

    • Medical devices

    • Special use-cases

    • …• Products range from simple conversion box to computerized

    automatic measurement systems

  • 8/9/2019 Electrical Safety - 8 (1)

    45/52

    Testing the Electric System

    Testing Receptacles

    Figure 14.17 Three-LED receptacle tester Ordinary silicon diodes prevent damaging

    reverse-LED currents, and resistors limit current. The LEDs are ON for line voltages from

    about 20 V rms to greater than 240 V rms, so these devices should not be used to measure linevoltage.

  • 8/9/2019 Electrical Safety - 8 (1)

    46/52

    Testing the Electric System

    Ground-pin-to-Chassis Resistance

    • Resistance between the ground pin of the plug and the equipment

    chassis and exposed metal objects should not exceed 0.15 Ohm

    during the life of the appliance

    Figure 14.18 Ground-pin-to-chassis resistance test

  • 8/9/2019 Electrical Safety - 8 (1)

    47/52

    Testing the Electric System

    Chassis Leakage Current

    Figure 14.19 (a) Chassis leakage-

    current test, (b) Current-meter circuitto be used for measuring leakage

    current. It has an input impedance of 1

    k Ω and a frequency characteristic that

    is flat to 1 kHz, drops at the rate of 20

    dB/decade to 100 kHz, and thenremains flat to 1 MHz or higher.

    (Reprinted with permission from

     NFPA 99-2005, "Health Care

    Facilities," Copyright ©2005, National

    Fire Protection Association, Quincy,MA 02269. This reprinted material is

    not the complete and official position

    of the National Fire Protection

    Association, on the referenced subject,

    which is represented only by thestandard in its entirety.)

  • 8/9/2019 Electrical Safety - 8 (1)

    48/52

    Testing the Electric System

    Leakage Current in Patient Leads (1)

    Leakage current from patient leads to ground

    Figure 14.20 Test for

    leakage current from patient

    leads to ground (Reprinted

    with permission from NFPA

    99-2005, "Health Care

    Facilities," Copyright

    ©2005, National Fire

    Protection Association,

    Quincy, MA 02269. This

    reprinted material is not the

    complete and official

     position of the National Fire

    Protection Association, on

    the referenced subject,which is represented only by

    the standard in its entirety.)

  • 8/9/2019 Electrical Safety - 8 (1)

    49/52

    Testing the Electric System

    Leakage Current in Patient Leads (2)

    Leakage current between patient leads

    Figure 14.21 Test for

    leakage current between

     patient leads (Reprinted with

     permission from NFPA 99-

    2005, "Health Care

    Facilities," Copyright ©

    2005, National Fire

    Protection Association,

    Quincy, MA 02269. This

    reprinted material is not the

    complete and official position

    of the National Fire Protetion

    Association, on the

    referenced subject, which is

    represented only by the

    standard in its entirety.)

  • 8/9/2019 Electrical Safety - 8 (1)

    50/52

    Testing the Electric System

    Leakage Current in Patient Leads (3)

    AC isolation current

    Figure 14.22 Test for ac isolationcurrent (Reprinted with

     permission from NFPA 99-2005,

    "Health Care Facilities," Copyright

    © 2005, National Fire Protection

    Association, Quincy, MA 02269.

    This reprinted material is not the

    complete and official position of

    the National Fire Protection

    Association, on the referenced

    subject, which is represented only

     by the standard in its entirety.)

  • 8/9/2019 Electrical Safety - 8 (1)

    51/52

    Conclusion

    • Adequate electrical safety in health-care facilities and systems can be

    achieved at moderate costs by combining:

    1. Good power-distribution system2. Well designed equipment

    3. Periodic maintenance and testing of power systems and equipment

    4. Modest training program for medical personnel

  • 8/9/2019 Electrical Safety - 8 (1)

    52/52

    With special thanks…

    …to the author of the corresponding book: John G. Webster 

    Further reading

    Medical Instrumentation Application and Design, 4th Edition, Chapter 14

    John G. Webster, Univ. of Wisconsin, Madison

    ISBN: 978-0-471-67600-3