Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
DC voltage sources:
single values
Solid state voltage
standard, standard
cell
Direct comparison 1 10 V Fixed voltage 1 V, 1.018 V, 10 V 0.02 to 0.2 µV/V 2 95% Yes 1_1_1Approved on 06
August 20131n
DC voltage sources:
low values
Multifunction
calibrator
Comparison with
reference
standard
0 10 V 0.1 to 20.3 µV 2 95% No 1_1_23Approved on 06
August 20137n
DC voltage sources:
intermediate values
Multifunction
calibrator
Comparison with
reference
standard
10 1000 V 0.07 to 10 mV 2 95% No 1_1_23Approved on 06
August 201316n
DC voltage meters:
very low valuesNanovoltmeter
Comparison with
reference
standard
0 1 mV 100 nV 2 95% NoApproved on 06
August 201312b
DC voltage meters:
intermediate values
DC voltmeter,
multimeter
Comparison with
reference
standard
0.001 1000 V5E-05 to
10mV 2 95% No 1_2_2
Approved on 06
August 201318n
DC voltage ratios:
up to 1100 V
Resistive divider:
ratio, k
Direct comparison
with standard0.1 0.1 Input voltage 1 V to 100 V
(0.4E-06 +
U 0/U out)k ,
U 0 = 50
nV, U out
output
voltage,
values
from 9E-
08 to 4E-
08
2 95% NoApproved on 06
August 201328
DC voltage ratios:
up to 1100 V
Resistive divider:
ratio, k
Direct comparison
with standard0.01 0.01 Input voltage 1 V to 1000 V
(1E-06 +
U 0/U out)k ,
U 0 = 50
nV, U out
output
voltage,
values
from 6E-
08 to 1E-
08
2 95% NoApproved on 06
August 201329
DC resistance
standards and
sources: low values
Standard resistor
CCC bridge, DCC
bridge, range
extender
0.1 1000 mW 0.8 to 10 µW/W 2 95% Yes 2_1_1Approved on 06
August 201332n
DC resistance
standards and
sources:
intermediate values
Standard resistorCCC bridge, DCC
bridge1 1000000 W 0.02 to 2 µW/W 2 95% Yes 2_1_2
Approved on 06
August 201330n
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
The BIPM key comparison database, June 2017 1/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
DC resistance
standards and
sources: high
values
Standard resistor
DCC bridge,
modified
Wheatstone
bridge
0.001 100000 GW 3 to 1500 µW/W 2 95% Yes 2_1_3Approved on 06
January 201537n
DC resistance
standards and
sources: standards
for high current
DC shunt
Voltage and
current
measurement
0.01 10000 mΩ Current < 500 A 15 to 200 µΩ/Ω 2 95% Yes 2_1_4Approved on 07
April 201638
DC resistance
standards and
sources: multiple
ranges
Multifunction
calibrator
Direct comparison
with standard0.001 1000000 kΩ 5 to 100 µΩ/Ω 2 95% Yes 2_1_5
Approved on 07
April 201640n
DC resistance
meters: low values
Microohmmeter,
multimeter,
resistance bridge
Direct comparison
with standard0.0001 1 Ω decadic values 10 µΩ 2 95% No 2_2_2
Approved on 07
April 201653b
DC resistance
meters:
intermediate values
Multimeter,
resistance bridge
Direct comparison
with standard0.001 1000000 kΩ
decadic values
multiplied by 1 or
1.9
1.2 to 25 µΩ/Ω 2 95% Yes 2_2_2Approved on 07
April 201654n
DC resistance
meters: high values
High resistance
meter, high
resistance bridge
Direct comparison
with standard1 100000 GΩ decadic values 25 to 5000 µΩ/Ω 2 95% Yes 2_2_2
Approved on 07
April 201655
DC current sources:
low values
Low current source,
multifunction
calibrator
Electrometer 0 100 µA0.0001 to
2000pA 2 95% No 3_1_1
Approved on 06
August 2013195n
DC current sources:
intermediate values
Multifunction
calibrator
Voltage across
resistance
standard,
comparison with
digital multimeter
0.0001 20 A 5 to 20 µA/A 2 95% Yes 3_1_2Approved on 06
August 201368n
DC current sources:
high valuesCurrent source
Zero flux
transformer20 100 A 20 µA/A 2 95% Yes
Approved on 07
April 201669
DC current sources:
transconductance
ratio
Transconductance
amplifier 1 1 S
Amplifier output
current2 A to 20 A 100 µS/S 2 95% Yes
Approved on 06
August 201379
DC current meters:
low values
Picoammeter,
nanoammeter,
electrometer,
multimeter
Comparison with
reference current
source, voltage
across resistance
standard
0 100 µA0.0001 to
2000pA 2 95% No 3_2_1
Approved on 06
August 2013191n
The BIPM key comparison database, June 2017 2/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
DC current meters:
intermediate valuesMultimeter
Voltage across
resistance
standard,
comparison with
multifunction
calibrator
0.0001 20 A 5 to 20 µA/A 2 95% Yes 3_2_2Approved on 06
August 201368n
DC current meters:
high valuesCurrent clamp Multiturn coil 10 100 A 0.5 % 2 95% Yes
Approved on 06
August 201377
AC resistance:
modulusStandard resistor
Direct comparison
with standard by
digital sampling
method
0.001 1 W Frequency 45 Hz to 65 Hz 40 to 20 µW/W 2 95% YesApproved on 06
August 201380a
AC resistance:
argumentStandard resistor
Direct comparison
with standard by
digital sampling
method
-0.1 0.1 rad Resistance 1 mW to 1 W 0.1 mrad 2 95% NoApproved on 06
August 201381a
Frequency 45 Hz to 65 Hz
AC resistance:
modulusStandard resistor
Direct comparison
with standard by
digital sampling
method
1 1000 W Frequency 45 Hz to 65 Hz 20 µW/W 2 95% YesApproved on 06
August 201382
AC resistance:
argumentStandard resistor
Direct comparison
with standard by
digital sampling
method
-0.1 0.1 rad Resistance 1 W to 1 kW 0.1 mrad 2 95% NoApproved on 06
August 201383
Frequency 45 Hz to 65 Hz
Capacitance: low
loss capacitorsStandard capacitor AH2500 bridge 0.01 100 nF Frequency 1 kHz 2 to 100 µF/F 2 95% Yes 4_2_1
Approved on 07
April 201684n
Capacitance:
dielectric capacitors
Fixed capacitor:
capacitance, CAH2500 bridge 0 10 pF Frequency 1 kHz
Q[5E-06,
1E-05C ],
C in pF,
values
range from
5E-06 to
1E-04
pF 2 95% NoApproved on 06
August 201387
Capacitance:
dielectric capacitorsFixed capacitor AH2500 bridge 0.01 1000 nF Frequency 1 kHz 10 to 100 µF/F 2 95% Yes 4_2_2
Approved on 07
April 201688n
The BIPM key comparison database, June 2017 3/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
Inductance: self
inductance,
intermediate values
Fixed inductor2 DVM, series
resonance0.1 100 mH Frequency
50 Hz to 2000 Hz,
1592 Hz 20 to 100 µH/H 2 95% Yes 4_3_2
Approved on 11
May 201793n
AC voltage: AC-DC
transfer difference
at low voltages
Thermal converter,
resistive dividerDirect comparison 2 500 mV Frequency 10 Hz to 1 MHz 5 to 1200 µV/V 2 95% Yes ACDC_U_matrix
Approved on 06
August 201394b
AC voltage: AC-DC
transfer difference
at low voltages
AC-DC transfer
standardDirect comparison 2 500 mV Frequency 10 Hz to 1 MHz 20 to 1200 µV/V 2 95% Yes
ACDC_U_matrix
_2
Approved on 06
August 2013141b
AC voltage: AC-DC
transfer difference
at low voltages
AC-DC transfer
standard
Comparison to AC-
DC meter0.1 0.5 V Frequency 20 Hz to 1 MHz 35 to 450 µV/V 2 95% Yes AC-DC METER
Approved on 06
August 2013142b
AC voltage: AC-DC
transfer difference
at low voltages
Multifunction
calibratorACV meter 0.1 0.5 V Frequency 20 Hz to 50 kHz 35 to 75 µV/V 2 95% Yes
CALIBRATOR
AC-DC
Approved on 06
August 2013135b
AC voltage: AC-DC
transfer difference
at medium voltages
Thermal converter Direct comparison 0.5 5 V Frequency 10 Hz to 1 MHz 5 to 50 µV/V 2 95% Yes ACDC_U_matrixApproved on 06
August 201394c
AC voltage: AC-DC
transfer difference
at medium voltages
AC-DC transfer
standardDirect comparison 0.5 5 V Frequency 10 Hz to 1 MHz 20 to 250 µV/V 2 95% Yes
ACDC_U_matrix
_2
Approved on 06
August 2013141c
AC voltage: AC-DC
transfer difference
at medium voltages
AC-DC transfer
standard
Comparison to AC-
DC meter0.5 5 V Frequency 20 Hz to 1 MHz 17 to 400 µV/V 2 95% Yes AC-DC METER
Approved on 06
August 2013142c
AC voltage: AC-DC
transfer difference
at medium voltages
Multifunction
calibratorACV meter 0.5 5 V Frequency 20 Hz to 50 kHz 20 to 70 µV/V 2 95% Yes
CALIBRATOR
AC-DC
Approved on 06
August 2013135c
AC voltage: AC-DC
transfer difference
at higher voltages
Thermal converter,
range resistorDirect comparison 5 1000 V Frequency 10 Hz to 1 MHz 5 to 100 µV/V 2 95% Yes ACDC_U_matrix
Approved on 06
August 201394d
AC voltage: AC-DC
transfer difference
at higher voltages
AC-DC transfer
standardDirect comparison 5 1000 V Frequency 10 Hz to 1 MHz 15 to 300 µV/V 2 95% Yes
ACDC_U_matrix
_2
Approved on 06
August 2013141d
The BIPM key comparison database, June 2017 4/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
AC voltage: AC-DC
transfer difference
at higher voltages
AC-DC transfer
standard
Comparison to AC-
DC meter5 1000 V Frequency 20 Hz to 1 MHz 19 to 400 µV/V 2 95% Yes AC-DC METER
Approved on 06
August 2013142d
AC voltage: AC-DC
transfer difference
at higher voltages
Multifunction
calibratorACV meter 5 1000 V Frequency 20 Hz to 50 kHz 20 to 60 µV/V 2 95% Yes
CALIBRATOR
AC-DC
Approved on 06
August 2013135d
AC voltage up to
1000 V: sources
Multifunction
calibrator
Direct comparison
with DC sampling0.01 700 V Frequency 1 Hz to 2 kHz 20 to 100 µV/V 2 95% Yes ACV_matrix
Narrow-bandwidth
measurement
Approved on 06
August 2013
121
AC voltage up to
1000 V: sources
Multifunction
calibratorACV meter 0.001 1000 V Frequency 10 Hz to 1 MHz 15 to 5000 µV/V 2 95% Yes
CALIBRATOR
ACV RMS
Approved on 11
May 2017123b
AC voltage up to
1000 V: metersMultimeter
Comparison to
calibrator0.002 1000 V Frequency 20 Hz to 1 MHz
60 to
14000µV/V 2 95% Yes DMM ACV RMS
Approved on 11
May 2017147b
AC voltage ratio:
modulus
Voltage divider: ratio,
kDirect comparison 0.00005 0.1 Input voltage 10 V to 1000 V
(5E-04 +
U 0/U out)k ,
U 0 = 1 µV,
U out
output
voltage
2 95% NoApproved on 06
August 2013155
Frequency 45 Hz to 65 Hz
AC voltage ratio:
argumentVoltage divider Direct comparison -1.57 1.57 rad Voltage ratio 5E-05 to 0.1 1 mrad 2 95% No
Approved on 06
August 2013156
Input voltage 10 V to 1000 V
Frequency 45 Hz to 65 Hz
AC current: AC-DC
transfer difference
Multifunction
calibrator, amplifier
AC meter plus
thermocouple
shunt resistors
0.01 20 A Frequency 40 Hz to 10 kHz 45 to 250 µA/A 2 95% YesAC-DC
CURRENT
Approved on 06
August 2013163b
AC current up to
100 A: sources
Multifunction
calibrator, amplifierDirect comparison 0.0001 20 A Frequency 40 Hz to 10 kHz 35 to 260 µA/A 2 95% Yes
AC RMS
CURRENT MFC
Approved on 06
August 2013174c
AC current up to
100 A: sources
Multifunction
calibrator, AC current
source
Direct comparison 0.025 10 A Frequency 45 Hz to 65 Hz 25 µA/A 2 95% YesApproved on 06
August 2013179a
AC current up to
100 A: sources
Multifunction
calibrator, AC current
source
Direct comparison 10 100 A Frequency 45 Hz to 65 Hz 50 µA/A 2 95% YesApproved on 06
August 2013187d
AC current up to
100 A: sources:
transconductance
Transconductance
amplifier
AC/DC standard
resistor1 1 S Output current 2 A to 20 A 180 µS/S 2 95% Yes
Approved on 06
August 2013173
Frequency 40 Hz to 1 kHz
AC current up to
100 A: metersMultimeter Direct comparison 0.0001 20 A Frequency 40 Hz to 10 kHz 45 to 300 µA/A 2 95% Yes
AC RMS
CURRENT DMM
Approved on 06
August 2013174d
AC current up to
100 A: metersMultimeter Direct comparison 0.025 10 A Frequency 45 Hz to 65 Hz 25 µA/A 2 95% Yes
Approved on 06
August 2013179b
The BIPM key comparison database, June 2017 5/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
AC current up to
100 A: meters
Rogowski coil,
current transducer
Comparison with
standard10 100 A Frequency 45 Hz to 65 Hz 50 µA/A 2 95% Yes
Approved on 06
August 2013187a
AC power and
energy: single
phase (f <= 400
Hz), active power
Power meter or
converter
Comparison with
standard0 5000 W Voltage 30 V to 500 V 50 µW/VA 2 95% Yes
Approved on 06
August 2013189a
Current 0.005 A to 10 A
Power factor1 to 0, inductive or
capacitive
Frequency 45 Hz to 65 Hz
AC power and
energy: single
phase (f <= 400
Hz), reactive power
Power meterComparison with
standard0 5000 var Voltage 30 V to 500 V 100 µvar/VA 2 95% Yes
Approved on 06
August 2013190a
Current 0.005 A to 10 A
Power factor1 to 0, inductive or
capacitive
Frequency 45 Hz to 65 Hz
AC power and
energy: three
phase, active power
Power meter or
converter
Comparison with
standard0 12.6 kW Voltage 50 V to 350 V 120 µW/VA 2 95% Yes
Voltage and
current ranges are
per phase
Approved on 06
August 2013
189b
Current 0.005 A to 12 A
Power factor1 to 0, inductive or
capacitive
Frequency 45 Hz to 65 Hz
AC power and
energy: three
phase, reactive
power
Power meterComparison with
standard0 12.6 kvar Voltage 50 V to 350 V 250 µvar/VA 2 95% Yes
Voltage and
current ranges are
per phase
Approved on 06
August 2013
190b
Current 0.005 A to 12 A
Power factor1 to 0, inductive or
capacitive
Frequency 45 Hz to 65 Hz
The BIPM key comparison database, June 2017 6/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
AC power and
energy: three
phase, active
energy
Active energy meterComparison with
standard0 0.35 kWh Voltage 50 V to 350 V 120
µWh/V
Ah2 95% Yes
Voltage and
current ranges are
per phase
Approved on 06
August 2013
189c
Current 0.005 A to 12 A
Power factor1 to 0, inductive or
capacitive
Frequency 45 Hz to 65 Hz
Measurement
time1 s to 100 s
AC power and
energy: three
phase, reactive
energy
Reactive energy
meter
Comparison with
standard0 0.35 kvarh Voltage 50 V to 350 V 250
µvarh/V
Ah2 95% Yes
Voltage and
current ranges are
per phase
Approved on 06
August 2013
190c
Current 0.005 A to 12 A
Power factor1 to 0, inductive or
capacitive
Frequency 45 Hz to 65 Hz
Measurement
time1 s to 100 s
High DC voltage:
high voltage
sources
DC kilovolt source
Dedicated set-up
for high voltage
measurement
1 1000 kV 5 to 100 µV/V 2 95% Yes 8_1_12Approved on 06
January 2015HV1n
High DC voltage:
high voltage meters
DC kilovoltmeter,
dedicated set-up for
high voltage
Comparison with
dedicated set-up
for high voltage
1 1000 kV 5 to 100 µV/V 2 95% Yes 8_1_12Approved on 06
January 2015HV3n
High DC voltage:
ratios
High voltage resistive
divider, DC high
voltage probe
Comparison with
reference divider1E-06 1 Input voltage 1 kV to 1000 kV
5E-06 to
100E-062 95% Yes 8_1_3
Approved on 06
January 2015HV5n
Output voltage 100 mV to 1000 V
High voltage
impedance:
capacitance
Compressed gas
capacitor, capacitor
for high voltage
Current
comparator bridge0.01 100000 nF Frequency 45 Hz to 60 Hz 20 to 200 µF/F 2 95% Yes 8_2_1a
Approved on 07
April 2016HV7n
Voltage < 200 kV
Current 30 µA to 1 A
High voltage
impedance:
capacitance
dissipation factor
Compressed gas
capacitor, capacitor
for high voltage
Current
comparator bridge0 1E-03 Capacitance 10 pF to 1 µF 1E-05 2 95% No
Approved on 06
August 2013HV11
Frequency 45 Hz to 60 Hz
Voltage < 200 kV
Current 30 µA to 1 A
The BIPM key comparison database, June 2017 7/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
High voltage
impedance:
capacitance
dissipation factor
Compressed gas
capacitor, capacitor
for high voltage
Current
comparator bridge1E-03 2 Capacitance 10 pF to 1 µF 1 % 2 95% Yes
Approved on 06
August 2013HV43
Frequency 45 Hz to 60 Hz
Voltage < 200 kV
Current 30 µA to 1 A
High voltage
impedance:
capacitance
dissipation factor
Compressed gas
capacitor, capacitor
for high voltage
Current
comparator bridge0 1E-03 Capacitance 1 µF to 10 µF 2E-05 2 95% No
Approved on 06
August 2013HV44
Frequency 45 Hz to 60 Hz
Voltage < 1 kV
Current > 30 µA
High voltage
impedance:
capacitance
dissipation factor
Compressed gas
capacitor, capacitor
for high voltage
Current
comparator bridge1E-03 2 Capacitance 1 µF to 10 µF 2 % 2 95% Yes
Approved on 06
August 2013HV45
Frequency 45 Hz to 60 Hz
Voltage < 1 kV
Current > 30 µA
The BIPM key comparison database, June 2017 8/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
High voltage
impedance:
capacitance
dissipation factor
Compressed gas
capacitor, capacitor
for high voltage
Current
comparator bridge0 1E-03 Capacitance 10 µF to 100 µF 2E-05 2 95% No
Approved on 06
August 2013HV46
Frequency 45 Hz to 60 Hz
Voltage < 1 kV
Current > 30 µA
High voltage
impedance:
capacitance
dissipation factor
Compressed gas
capacitor, capacitor
for high voltage
Current
comparator bridge1E-03 2 Capacitance 10 µF to 100 µF 2 % 2 95% Yes
Approved on 06
August 2013HV47
Frequency 45 Hz to 60 Hz
Voltage < 1 kV
Current > 30 µA
High voltage
impedance:
inductance
High voltage reactor:
loss angle
Simultaneous
sampling of
current and
voltage
1E-06 10 H Frequency 45 Hz to 60 Hz 0.3 mH/H 2 95% YesApproved on 08
April 2009HV12
Current 50 mA to 300 A
Voltage 10 mV to 1000 V
High voltage
impedance:
inductance loss
angle
High voltage reactor
Simultaneous
sampling of
current and
voltage
0 350 mrad Frequency 45 Hz to 60 Hz 0.2 mrad 2 95% NoApproved on 08
April 2009HV13
Current 50 mA to 300 A
Voltage 10 mV to 1000 V
High voltage
impedance: burden:
modulus
Instrument
transformer burden:
absolute value of
impedance
Simultaneous
sampling of
current and
voltage
0.1 30000 W Frequency 45 Hz to 60 Hz 1 mW/W 2 95% YesApproved on 08
April 2009HV60
Voltage < 300 V
Current 10 mA to 15 A
High voltage
impedance: burden:
argument
Instrument
transformer burden:
phase angle
Simultaneous
sampling of
current and
voltage
0 p/2 rad Frequency 45 Hz to 60 Hz 2 mrad 2 95% NoApproved on 08
April 2009HV61
Voltage < 300 V
Current 10 mA to 15 A
The BIPM key comparison database, June 2017 9/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
AC high voltage:
meters
AC high voltage
meter, dedicated set-
up for high voltage
measurements: rms
value
Comparison with
dedicated set-up
for high voltage
1 200 kV Frequency 45 Hz to 60 Hz 50 µV/V 2 95% YesApproved on 11
May 2017HV15
AC high voltage:
peak values
AC high voltage
meter, dedicated set-
up for high voltage
measurements
Comparison with
dedicated set-up
for high voltage
1 200 kV Frequency 45 Hz to 60 Hz 50 µV/V 2 95% YesApproved on 11
May 2017HV16
AC high voltage:
voltage ratio non-
linearity
AC high voltage
meter, dedicated set-
up for high voltage
measurements
Linearity extension
according to IEC
60060-2
0 10 % Voltage range 100 kV to 1000 kV 0.1 % 2 95% YesApproved on 11
May 2017HV16a
Frequency 46 Hz to 60 Hz
AC high voltage:
ratio error
High voltage
transformer, high
voltage divider
Current
comparator bridge0 3 % Frequency 45 Hz to 60 Hz
0.002 to
0.005% 2 95% No 8_3_4
Approved on 11
May 2017HV17
Input voltage 20 V to 200 kV
Output voltage 2 V to 1000 V
AC high voltage
ratio: phase
displacement
High voltage
transformer, high
voltage divider
Current
comparator bridge0 2 mrad Frequency 45 Hz to 60 Hz 0.02 mrad 2 95% No
Approved on 11
May 2017HV18
Input voltage 20 V to 200 kV
Output voltage 2 V to 1000 V
AC high voltage
ratio: phase
displacement
High voltage
transformer, high
voltage divider
Current
comparator bridge2 100 mrad Frequency 45 Hz to 60 Hz 10
mrad/ra
d2 95% Yes
Approved on 11
May 2017HV18b
Input voltage 20 V to 200 kV
Output voltage 2 V to 1000 V
AC high voltage:
ratio error
High voltage
transformer, high
voltage divider
Simultaneous
sampling of input
and output
voltages
0 3 % Frequency 45 Hz to 60 Hz 0.01 % 2 95% NoApproved on 08
April 2009HV17a
Input voltage 1 kV to 200 kV
Output voltage 0.1 V to 20 V
AC high voltage
ratio: phase
displacement
High voltage
transformer, high
voltage divider
Simultaneous
sampling of input
and output
voltages
0 100 mrad Frequency 45 Hz to 60 Hz 0.02 mrad 2 95% NoApproved on 08
April 2009HV18a
Input voltage 1 kV to 200 kV
Output voltage 0.1 V to 20 V
The BIPM key comparison database, June 2017 10/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
Pulsed high voltage
and current:
lightning impulse
voltage parameters
Digital recorder, peak
voltmeter, impulse
calibrator, lightning
impulse voltage
measurement set-up,
EMC surge
generator: peak
value
Impulse voltage
calibrator or
comparison with
dedicated set-up
for high voltage
0.05 400000 V Impulse shapeaccording to
IEC60060-11 to 5 mV/V 2 95% Yes 8_4_1
Approved on 06
August 2013HV19n
Pulsed high voltage
and current:
lightning impulse
time parameters
Digital recorder,
impulse calibrator,
lightning impulse
voltage
measurement set-up,
EMC surge
generator: front time
Impulse voltage
calibrator or
comparison with
dedicated set-up
for high voltage
0.84 1.56 µs Impulse shapeaccording to
IEC60060-110 to 20 ms/s 2 95% Yes 8_4_2a
Approved on 06
August 2013HV22n
Voltage 0.05 V to 400 kV
Pulsed high voltage
and current:
lightning impulse
time parameters
Digital recorder,
impulse calibrator,
lightning impulse
voltage
measurement set-up,
EMC surge
generator: time to
half value
Impulse voltage
calibrator or
comparison with
dedicated set-up
for high voltage
40 60 µs Impulse shapeaccording to
IEC60060-15 to 10 ms/s 2 95% Yes 8_4_2b
Approved on 06
August 2013HV25n
Voltage 0.05 V to 400 kV
Pulsed high voltage
and current:
switching impulse
voltage parameters
Digital recorder, peak
voltmeter, switching
impulse voltage
measurement set-up,
impulse calibrator:
peak value
Impulse voltage
calibrator or
comparison with
dedicated set-up
for high voltage
0.05 200000 V Impulse shapeaccording to
IEC60060-11 to 2 mV/V 2 95% Yes 8_4_3
Approved on 06
August 2013HV28n
Pulsed high voltage
and current:
switching impulse
time parameters
Digital recorder, peak
voltmeter, switching
impulse voltage
measurement set-up,
impulse calibrator:
time to peak
Impulse voltage
calibrator or
comparison with
dedicated set-up
for high voltage
20 300 µs Impulse shapeaccording to
IEC60060-130 ms/s 2 95% Yes
Approved on 06
August 2013HV31n
Voltage 0.05 V to 200 kV
The BIPM key comparison database, June 2017 11/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
Pulsed high voltage
and current:
switching impulse
time parameters
Digital recorder, peak
voltmeter, switching
impulse voltage
measurement set-up,
impulse calibrator:
time to half value
Impulse voltage
calibrator or
comparison with
dedicated set-up
for high voltage
2000 4000 µs Impulse shapeaccording to
IEC60060-15 to 10 ms/s 2 95% Yes 8_4_4
Approved on 06
August 2013HV34n
Voltage 0.05 V to 200 kV
Pulsed high voltage
and current:
impulse current
parameters
Dedicated
measurement set-up,
current shunt: peak
value
Comparison with
dedicated set-up
for high voltage
1 10000 A Impulse shapeaccording to
IEC60060-130 mA/A 2 95% Yes
Approved on 08
April 2009HV37
Pulsed high voltage
and current:
impulse current time
parameters
Dedicated
measurement set-up,
current shunt
Comparison with
dedicated set-up
for high voltage
1 100 µs Impulse shapeaccording to
IEC60060-150 ms/s 2 95% Yes
Approved on 08
April 2009HV50
Electric discharge:
apparent charge
Partial discharge
calibrator or meter:
apparent charge q
Comparison with
partial discharge
calibrator
1 10 pC 0.2 pC 2 95% NoApproved on 06
August 2013HV39
Electric discharge:
apparent charge
Partial discharge
calibrator or meter:
apparent charge
Comparison with
partial discharge
calibrator
10 100000 pC 20 mC/C 2 95% YesApproved on 11
May 2017HV51
High AC current:
sources
High current AC
source
Comparison with
dedicated set-up
for high AC
current
0.1 8 kA Frequency 45 Hz to 65 Hz 50 µA/A 2 95% YesApproved on 11
May 2017HV54n
High AC current:
meters
Rogowski coil,
current transducer
Comparison with
standard0.1 8 kA Frequency 45 Hz to 65 Hz 50 µA/A 2 95% Yes
Approved on 11
May 2017HV56n
Current transformer:
ratio error
Current transformer,
current transducer
Comparison with
standard or
comparison with
dedicated set-up
for high AC
current
0 5 % Frequency 45 Hz to 65 Hz 25E-06 2 95% NoApproved on 11
May 2017HV57n
Input current 1 A to 6 kA
Output current 10 mA to 20 A
Current transformer:
phase displacement
Current transformer,
current transducer
Comparison with
standard or
comparison with
dedicated set-up
for high AC
current
0 3.14 rad Frequency 45 Hz to 65 Hz 0.05 mrad 2 95% NoApproved on 11
May 2017HV58n
Input current 1 A to 6 kA
Output current 10 mA to 20 A
The BIPM key comparison database, June 2017 12/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
QuantityInstrument or
artifact
Instrument Type
or Method
Minimum
value
Maximum
valueunits Parameter Specifications Value Units
Coverage
Factor
Level of
Confidence
Is the expanded
uncertainty a
relative one?
Uncertainty
matrixComments
NMI Service
Identifier
Calibration or Measurement Services Measurand Level or RangeMeasurement
Conditions/Independent variablesExpanded Uncertainty
High DC current:
sources
High current DC
source
Zero flux
transformer100 500 A 20 µA/A 2 95% Yes
Approved on 07
April 2016HV59
High DC current:
meters
High current DC
meter
Zero flux
transformer100 500 A 20 µA/A 2 95% Yes
Approved on 07
April 2016HV60
RF power: absolute
power in coaxial linePower meter Direct 1 1 mW Frequency 10 MHz to 18 GHz 4 to 11 mW/W 2 95% Yes
Uncertainty
depends on
frequency and on
device match
Approved on 06
August 2013
201
Connector Type N
RF power:
calibration factor in
coaxial line
Power sensorPower splitter
system1 1 mW Frequency 10 MHz to 18 GHz
0.004 to
0.0112 95% No
Absolute
uncertainty of
calibration factor is
given. Uncertainty
depends on
frequency and on
device match
Approved on 06
August 2013
202
Connector Type N
Scattering
parameters:
reflection coefficient
(Sii) in coaxial line,
real and imaginary
Passive devicesVector network
analyser-1 1 Frequency 10 MHz to 18 GHz
0.013 to
0.0732.45 95% No REFLECTION
Uncertainty
depends on
frequency and
reflection
coefficient. At 10
MHz, S11 and S22
<= 0.2. VRC
magnitudes
greater than 0.2
are not measured
at 10 MHz.
Approved on 06
August 2013
203
Connector Type N
Scattering
parameters:
transmission
coefficient (Sij) in
coaxial line,
magnitude
Passive devicesVector network
analyser0 80 dB Frequency 10 MHz to 18 GHz
0.016 to
0.181dB 2 95% No ATTENUATION
At 10 MHz, S11
and S22 < 0.1
Approved on 06
August 2013
204
Connector Type N
S11 and S22 < 0.2
The BIPM key comparison database, June 2017 13/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
1_1_1
Relative expanded
uncertainty
Reference Standard used in
calibration
1 V 0.05 JAVS
1.018 V 0.05 JAVS
10 V 0.02 JAVS
1 V 0.2 Solid state voltage standard
1.018 V 0.2 Solid state voltage standard
10 V 0.2 Solid state voltage standard
The expanded uncertainty given in this table are expressed in µV/V
The BIPM key comparison database, June 2017 14/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
1_1_23
Expanded uncertaintyReference Standard used in
calibration
0 mV to 10 mV 0.1 Nanovoltmeter
10 mV to 10 V 0.3 + 2U DC voltmeter
100 mV 2.5USolid state voltage standard,
resistive voltage divider
1 V 0.7U Solid state voltage standard
10 V 0.5U Solid state voltage standard
100 V 2U Solid state voltage standard, divider
1000 V 3U Solid state voltage standard, divider
10 V to 100 V 10 + 6U DC voltmeter
100 V to 1000 V 10U DC voltmeter
U is expressed in volt
The expanded uncertainty given in this table are expressed in µV
The BIPM key comparison database, June 2017 15/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
1_2_2
Expanded uncertainty Reference Standard used in calibration
1 mV to 1 V 0.05 JAVS
1 V to 10 V 0.05 to 0.2 JAVS
0.1 V 3USolid state voltage standard, voltage
divider
1 V 1U Solid state voltage standard
10 V 1U Solid state voltage standard
100 V 2USolid state voltage standard, voltage
divider
1000 V 5USolid state voltage standard, voltage
divider
1 mV to 10 V 0.3 + 2U Multifunction calibrator
10 V to 100 V 10 + 6U Multifunction calibrator
100 V to 1000 V 10U Multifunction calibrator
U is expressed in volt
The expanded uncertainty given in this table are expressed in µV
The BIPM key comparison database, June 2017 16/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
2_1_1
Relative expanded uncertainty Method of measurement
0.1 mΩ to 1 mΩ 10 DCC bridge and range extender
1 mΩ to 10 mΩ 5 DCC bridge and range extender
10 mΩ to 100 mΩ 2 DCC bridge and range extender
0.1 Ω to 1 Ω 0.8 DCC bridge
10 mΩ 1 CCC bridge
The expanded uncertainty given in this table are expressed in μΩ/Ω
The BIPM key comparison database, June 2017 17/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
2_1_2
Relative expanded uncertainty Method of measurement
100 Ω 0.02Comparison with CCC bridge against
QHR
1 Ω to 10000 Ω 0.05Comparison with CCC bridge against
resistance standard
1 Ω to 100000 Ω 0.4Comparison with DCC bridge against
resistance standard
0.1 MΩ to 1 MΩ 2Comparison with DCC bridge against
resistance standard
The expanded uncertainty given in this table are expressed in μΩ/Ω
The BIPM key comparison database, June 2017 18/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
2_1_3
Relative expanded uncertainty Method of measurement
1 MΩ to 10 MΩ 3 DCC bridge or Modified Wheatstone bridge
10 MΩ to 100 MΩ 5 Modified Wheatstone bridge
0.1 GΩ to 1 GΩ 10 Modified Wheatstone bridge
1 GΩ to 10 GΩ 20 Modified Wheatstone bridge
10 GΩ to 100 GΩ 25 Modified Wheatstone bridge
0.1 TΩ to 1 TΩ 70 Modified Wheatstone bridge
1 TΩ to 10 TΩ 500 Modified Wheatstone bridge
10 TΩ to 100 TΩ 1500 Modified Wheatstone bridge
The expanded uncertainty given in this table are expressed in µΩ/Ω
The BIPM key comparison database, June 2017 19/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
2_1_4
Relative expanded uncertainty
10 µΩ to 100 µΩ 200
100 µΩ to 1 mΩ 70
1 mΩ to 10 mΩ 30
10 mΩ to 10 Ω 15
The expanded uncertainty given in this table are expressed in μΩ/Ω
The BIPM key comparison database, June 2017 20/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
2_1_5
Relative expanded uncertainty Method of measurement
1 Ω to 100 Ω 10 DCC bridge
100 Ω to 10000 Ω 5 DCC bridge
10 kΩ to 100 kΩ 10 Modified Wheatstone bridge
0.1 MΩ to 1 MΩ 25 Modified Wheatstone bridge
1 MΩ to 10 MΩ 50 Modified Wheatstone bridge
10 MΩ to 1000 MΩ 100 Modified Wheatstone bridge
The expanded uncertainty given in this table are expressed in μΩ/Ω
The BIPM key comparison database, June 2017 21/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
2_2_2
Relative expanded uncertainty
0.1 mΩ 25
1 mΩ 15
10 mΩ 6
100 mΩ 1.5
1 Ω to 100 kΩ 1.2
0.1 MΩ to 1 MΩ 4
10 MΩ 6
100 MΩ 12
1 GΩ 25
10 GΩ 35
100 GΩ 80
1 TΩ 150
10 TΩ 1600
100 TΩ 5000
The expanded uncertainty given in this table are expressed in μΩ/Ω
Best uncertainties can be applied when the calibrated instrument is a resistance bridge. Two standard resistors are
needed for calibration of a bridge and thus the uncertainties in this table are limited by combined uncertainties of the two
best reference resistors availlable at the NMI on each decade.
The BIPM key comparison database, June 2017 22/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
3_1_1
Expanded uncertainty
0 fA to 10 fA 0.1 fA
10 fA to 100 fA 0.1 fA to 1 fA
0.1 pA to 10 pA 0.1 fA to 10 fA
10 pA to 100 pA 6 fA to 60 fA
0.1 nA to 1 nA 20 fA to 200 fA
1 nA to 10 nA 0.1 pA to 1 pA
0.01 µA to 100 µA 0.2 pA to 2000 pA
The BIPM key comparison database, June 2017 23/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
3_1_2
Relative expanded uncertainty Method of measurement
0.1 mA to 100 mA 5 Voltage across resistance standard
0.1 A to 2 A 5 to 10 Voltage across resistance standard
2 A to 20 A 10 to 20 Voltage across resistance standard
0.1 mA to 1 A 10 Comparison with digital multimeter
The expanded uncertainty given in this table are expressed in μA/A
The BIPM key comparison database, June 2017 24/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
3_2_1
Expanded uncertainty Method of measurement
0 fA to 10 fA 0.1 fA Comparison with reference current source
10 fA to 100 fA 0.1 fA to 1 fA Comparison with reference current source
0.1 pA to 10 pA 0.1 fA to 10 fA Comparison with reference current source
10 pA to 100 pA 6 fA to 60 fA Comparison with reference current source
0.1 nA to 1 nA 20 fA to 200 fA Voltage across resistance standard
1 nA to 10 nA 0.1 pA to 1 pA Voltage across resistance standard
0.01 µA to 100 µA 0.2 pA to 2000 pA Voltage across resistance standard
The BIPM key comparison database, June 2017 25/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
3_2_2
Relative expanded uncertainty Method of measurement
0.1 mA to 100 mA 5 Voltage across resistance standard
0.1 A to 2 A 5 to 10 Voltage across resistance standard
2 A to 20 A 10 to 20 Voltage across resistance standard
0.1 mA to 1 A 10 Comparison with multifunction calibrator
The expanded uncertainty given in this table are expressed in μA/A
The BIPM key comparison database, June 2017 26/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
4_2_1
Relative expanded
uncertainty
10 pF 2
100 pF 2
1 nF 10
10 nF 30
100 nF 100
The expanded uncertainty given in this table are expressed in μF/F
The BIPM key comparison database, June 2017 27/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
4_2_2
Relative expanded
uncertainty
10 pF to 100 pF 10
100 pF to 1 nF 10
1 nF to 10 nF 30
10 nF to 100 nF 100
100 nF to 1 μF 100
The expanded uncertainty given in this table are expressed in μF/F
The BIPM key comparison database, June 2017 28/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
4_3_2
Method of
measurement
Reference standard used in
calibrationFrequency
Relative expanded
uncertainty
100 mH Series resonance Capacitance standard 1 kHz, 1.592 kHz 20
10 mH to 100 mH 2 DVM Resistance standard 50 Hz to 1000 Hz 50
100 µH to 10 mH 2 DVM Resistance standard 50 Hz to 2000 Hz 20
The expanded uncertainties given in this table are expressed in μH/H
The BIPM key comparison database, June 2017 29/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
ACDC_U_matrix
10 Hz 20 Hz30 Hz to
300 Hz
400 Hz to
10 kHz
20 kHz to
30 kHz50 kHz 70 kHz 100 kHz 200 kHz 300 kHz 500 kHz
700 kHz to
800 kHz1 MHz
2 mV to 20 mV 400 400 400 400 400 400 400 400 500 500 500 800 1200
20 mV to 200 mV 100 100 80 80 80 80 80 80 150 150 150 200 300
0.2 V to 0.5 V 10 5 5 5 5 6 7 7 15 15 20 40 50
0.5 V to 5 V 10 5 5 5 5 6 7 7 15 15 20 40 50
5 V to 10 V 10 5 5 5 5 6 7 7 15 15 20 40 50
10 V to 30 V 20 6 6 6 6 7 8 8 15 15 20 40 50
30 V to 100 V 30 6 6 6 6 8 10 15 20 - - - -
100 V to 300 V 30 7 7 7 7 10 15 20 - - - - -
300 V to 1000 V - - 10 10 15 25 50 100 - - - - -
The expanded uncertainties given in this table are expressed in µV/V
The BIPM key comparison database, June 2017 30/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
ACDC_U_matrix_2
10 Hz 20 Hz30 Hz to
300 Hz
400 Hz to
10 kHz
20 kHz to
30 kHz50 kHz 70 kHz 100 kHz 200 kHz 300 kHz 500 kHz
700 kHz to
800 kHz1 MHz
2 mV to 20 mV 400 400 400 400 400 400 400 400 500 500 500 800 1200
20 mV to 200 mV 100 100 80 80 80 80 80 80 150 150 150 200 300
0.2 V to 0.5 V 50 40 30 20 20 30 30 40 70 90 100 160 250
0.5 V to 5 V 50 40 30 20 20 30 30 40 70 90 100 160 250
5 V to 10 V 50 30 20 15 15 20 25 30 50 60 80 130 220
10 V to 30 V 50 30 20 20 20 20 25 30 50 60 80 130 220
30 V to 100 V 50 30 20 20 20 20 25 30 50 - - - -
100 V to 300 V 60 40 30 30 40 60 60 100 - - - - -
300 V to 1000 V - - 40 50 80 150 200 300 - - - - -
The expanded uncertainties given in this table are expressed in µV/V
The BIPM key comparison database, June 2017 31/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
AC-DC METER
20 Hz 40 Hz 53 Hz 400 Hz 1 kHz 20 kHz 50 kHz 100 kHz 500 kHz 1 MHz
100 mV 75 45 - 40 40 40 60 90 150 450
500 mV 70 40 - 35 35 35 55 80 140 400
1 V 50 20 - 17 17 17 35 60 110 400
5 V 50 20 - 19 19 19 35 60 110 400
10 V 50 20 - 19 19 19 35 60 110 400
100 V 55 30 - 25 25 25 60 100 - -
1000 V - - 40 40 40 - - - - -
The expanded uncertainties given in this table are expressed in µV/V
The BIPM key comparison database, June 2017 32/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
CALIBRATOR AC-DC
20 Hz 40 Hz 53 Hz 400 Hz 1 kHz 20 kHz 50 kHz
100 mV 75 45 - 40 40 40 60
500 mV 70 40 - 35 35 35 60
1 V 55 20 - 20 20 20 35
5 V 55 25 - 20 20 20 40
10 V 55 25 - 20 20 20 40
100 V 55 30 - 30 30 30 60
1000 V - - 45 45 45 - -
The expanded uncertainties given in this table are expressed in µV/V
The BIPM key comparison database, June 2017 33/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
ACV_matrix
1 Hz to 400 Hz 400 Hz to 1 kHz 1 kHz to 2 kHz
10 mV to 80 mV 50 50 100
80 mV to 8 V 20 50 100
8 V to 700 V 20 50 100
The expanded uncertainties given in this table are expressed in µV/V
The BIPM key comparison database, June 2017 34/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
CALIBRATOR ACV RMS
10 Hz 20 Hz 40 Hz 53 Hz 400 Hz 1 kHz 10 kHz 20 kHz 50 kHz 100 kHz 500 kHz 1 MHz
1 mV 1200 1200 1200 - 1200 1200 1200 1200 1200 1400 3300 5000
2 mV 590 590 590 - 590 590 590 590 590 680 1600 4600
10 mV 130 120 120 - 120 120 120 120 120 140 350 580
100 mV 50 50 30 - 30 30 30 30 35 60 220 450
1 V 40 40 20 - 15 15 15 15 30 50 110 440
10 V 40 40 20 - 20 20 20 20 30 45 140 400
100 V 40 40 30 - 20 20 20 20 35 40 - -
1000 V - - 50 50 50 50 50 55 65 95 - -
The expanded uncertainties given in this table are expressed in µV/V
The BIPM key comparison database, June 2017 35/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
DMM ACV RMS
20 Hz 40 Hz 53 Hz 400 Hz 1 kHz 20 kHz 50 kHz 100 kHz 500 kHz 1 MHz
2 mV 2300 2100 - 2100 2100 2100 2100 4200 12000 14000
20 mV 400 320 - 320 320 320 510 910 2300 3400
100 mV 250 150 - 150 150 150 300 760 1400 3000
200 mV 200 120 - 110 110 110 260 650 1300 2600
1 V 140 60 - 60 60 60 110 230 1100 1600
10 V 140 65 - 60 60 60 110 200 1200 2600
100 V 140 70 - 70 70 70 190 440 - -
1000 V - - 75 75 75 175 340 805 - -
The expanded uncertainties given in this table are expressed in µV/V
The BIPM key comparison database, June 2017 36/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
AC-DC CURRENT
40 Hz 400 Hz 1 kHz 10 kHz
10 mA 45 45 45 45
100 mA 45 45 45 45
1 A 55 55 55 55
10 A 100 100 100 200
20 A 100 100 100 250
The expanded uncertainties given in this table are expressed in µA/A
The BIPM key comparison database, June 2017 37/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
AC RMS CURRENT MFC
40 Hz 400 Hz 1 kHz 5 kHz 10 kHz
100 µA 80 80 80 90 110
1 mA 35 35 35 35 40
10 mA 35 35 35 35 50
100 mA 35 35 35 35 60
1 A 35 35 35 50 110
10 A 110 110 110 150 210
20 A 110 110 110 180 260
The expanded uncertainties given in this table are expressed in µA/A
The BIPM key comparison database, June 2017 38/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
AC RMS CURRENT DMM
40 Hz 400 Hz 1 kHz 5 kHz 10 kHz
100 µA 120 120 120 150 160
1 mA 70 70 70 80 90
10 mA 65 65 65 70 80
100 mA 60 60 60 70 85
1 A 50 45 45 80 120
10 A 120 120 120 200 250
20 A 150 150 150 230 300
The expanded uncertainties given in this table are expressed in µA/A
The BIPM key comparison database, June 2017 39/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
8_1_12
Output voltage from 0.1 V to 12 V Output voltage from 12 V to 1000 V
1 kV to 100 kV 5 10
100 kV to 200 kV 10 10
200 kV to 1000 kV 100 100
The expanded uncertainties given in this table are expressed in μV/V
Calibrations on voltages above 200 kV are performed in co-operation between MIKES, PTB, SP, UME and/or VSL.
The BIPM key comparison database, June 2017 40/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
8_1_3
Output voltage from 0.1 V to 12 V Output voltage from 12 V to 1000 V
1 kV to 100 kV 5E-06 10E-06
100 kV to 200 kV 10E-06 10E-06
200 kV to 1000 kV 100E-06 100E-06
The uncertainties are given relative to the measured value
Calibrations on voltages above 200 kV are performed in co-operation between MIKES, PTB, SP, UME and/or VSL.
The BIPM key comparison database, June 2017 41/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
8_2_1a
1 kV 1 kV to 10 kV 10 kV to 30 kV 30 kV to 200 kV
10 pF to 100 pF 50 50 50 50
100 pF 20 20 20 20
100 pF to 1 nF 30 30 30 30
1 nF 20 20 20 20
1 nF to 10 nF 30 30 30 30
10 nF 50 50 50 100
10 nF to 100 nF 50 50 50 100
100 nF to 100 μF 200 500 - -
The expanded uncertainty given in this table are expressed in μF/F
The BIPM key comparison database, June 2017 42/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
8_3_4
X/N-Ratio in V/V (nominal) 45 Hz to 60 Hz
0.02 to 111 20
111 to 1110 30
1110 to 11100 50
The expanded uncertainties given in this table are expressed in µV/V.
The BIPM key comparison database, June 2017 1/1
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
8_4_1
Relative expanded
uncertainty
50 mV to 300 V 1
300 V to 1 kV 1.5
1 kV to 400 kV 5
The expanded uncertainty given in this table are expressed in mV/V
The BIPM key comparison database, June 2017 44/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
8_4_2a
50 mV to 300 V 300 V to 1 kV 1 kV to 400 kV
0.84 µs to 1.56 µs 10 15 20
The expanded uncertainty given in this table are expressed in ms/s
The BIPM key comparison database, June 2017 45/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
8_4_2b
50 mV to 300 V 300 V to 1 kV 1 kV to 400 kV
40 µs to 60 µs 5 5 10
The expanded uncertainty given in this table are expressed in ms/s
The BIPM key comparison database, June 2017 46/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
8_4_3
Relativeexpanded
uncertainty
50 mV to 300 V 1
300 V to 200 kV 2
The expanded uncertainty given in this table are expressed in mV/V
The BIPM key comparison database, June 2017 47/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
8_4_4
50 mV to 300 V 300 V to 200 kV
2000 µs to 4000 µs 5 10
The expanded uncertainty given in this table are expressed in ms/s
The BIPM key comparison database, June 2017 48/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
Uncertainty table: REFLECTION
|VRC| 0.010 GHz 0.100 GHz 1.000 GHz 6.000 GHz 14.000 GHz 18.000 GHz
0 0.013 0.013 0.013 0.025 0.025 0.025
0.1 0.013 0.013 0.013 0.025 0.025 0.025
0.2 0.014 0.014 0.014 0.026 0.027 0.027
0.3 - 0.016 0.016 0.028 0.030 0.030
0.4 - 0.018 0.018 0.030 0.033 0.034
0.5 - 0.021 0.021 0.033 0.038 0.038
0.6 - 0.024 0.024 0.036 0.043 0.044
0.7 - 0.027 0.027 0.040 0.049 0.050
0.8 - 0.031 0.031 0.044 0.056 0.057
0.9 - 0.036 0.036 0.048 0.063 0.065
1 - 0.041 0.041 0.053 0.072 0.073
Uncertainties in real and imaginary part of reflection coefficient.
The column on the left (|VRC|) gives the magnitude of the reflection coefficient.
The expanded absolute uncertainties given in this table are dimensionless.
The BIPM key comparison database, June 2017 49/50
Calibration and Measurement Capabilities
Electricity and Magnetism, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
ATTENUATION
0.01 GHz 0.100 GHz 1 GHz 6 GHz 10.000 GHz 14.000 GHz 18.000 GHz
0 dB 0.016 0.016 0.025 0.055 0.057 0.057 0.057
10 dB 0.026 0.026 0.032 0.059 0.060 0.060 0.060
20 dB 0.043 0.043 0.047 0.068 0.070 0.070 0.070
30 dB 0.062 0.062 0.065 0.082 0.083 0.083 0.083
40 dB 0.082 0.082 0.084 0.097 0.099 0.099 0.099
50 dB 0.101 0.101 0.103 0.114 0.121 0.121 0.121
60 dB 0.121 0.121 0.123 0.132 0.181 0.181 0.181
70 dB 0.141 0.141 0.143 0.151 - - -
80 dB 0.165 0.165 0.166 0.173 - - -
The expanded absolute uncertainties given in this table are expressed in dB.
The BIPM key comparison database, June 2017 50/50