1
Slope at 20°C at 25°C Status 1pH unit 3pH units 1pH unit 3pH units >102 >59.3 >178.0 >60.3 >181.0 Poor 102% 59.3 178.0 60.3 181.0 OK 101% 58.8 176.3 59.8 179.3 Very Good 100% 58.2 174.5 59.2 177.5 Ideal 99% 57.6 172.8 58.6 175.7 Very Good 98% 57.0 171.0 58.0 173.9 Very Good 97% 56.4 169.3 57.4 172.2 Very Good 96% 55.8 167.5 56.8 170.4 Very Good 95% 55.3 165.8 56.2 168.6 Very Good 94% 57.0 171.0 58.0 173.9 OK 93% 54.1 162.3 55.0 165.1 OK 92% 53.5 160.5 54.4 163.3 OK <92 <53.5 <160.5 <54.4 <163.3 Poor How to check your pH electrode slope 1) Select the millivolt (mV) mode of your pH meter 2) Using a pH electrode, obtain mV readings of two fresh calibration buffers (i.e., pH 4.0 and pH 7.0 are best) 3) Determine the net mV change 4) Determine the net mV change per pH unit change and compare using the chart at right Cleaning Probes Cleaning a dirty, clogged or coated electrode can restore proper electrode performance and prolong the useful life of the electrode. Here are some cleaning suggestions: • Organic Oil/Grease Films—liquid dish soap and warm water are best. Isopropyl alcohol can be used on most electrodes if additional cleaning is needed. NOTE: Solvents such as acetone will damage plastic body electrodes! • Inorganic Salt Deposits—soak probe in 0.1M HCl as needed • Proteins—Use a protein-cleaning solution (13-642-584) If you have a refillable probe, drain the chamber and refill with correct reference filling solution. Rinse the probe tip with distilled or deionized water and condition the probe in storage solution before using again. pH Scale Battery Acid Lemon Juice Vinegar Orange Juice Milk Distilled Water Baking Soda Sea Water Milk of Magnesia Ammonia Lye 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Acid Rain Increasing Acidity Neutral Increasing Alkalinity F a x : 1 - 8 0 0 - 9 2 6 - 1 1 6 6 w w w . f s h e r s c i . c o m Did you know? Buffers should be at the same temperature as your sample. electrochemistry 101 with Dr. Seymour Blue For example, at 25°C: pH 4 = 170.5mV, pH 7 = -3.4mV: Net change = 170.5mV - (-3.4mV) = 173.9mV. Since pH 4 and pH 7 are 3 pH units apart, using the chart, 98% = Very Good Care and Maintenance of Electrodes Calibrate regularly with fresh buffers Rinse electrode tip well between readings On refillable electrodes be sure to check your fill level and open the fill hole before calibration or measurement Never rub the bulb to dry it; a static charge can develop! Store your electrode with fill-hole closed in electrode storage solution (Never store in deionized water!) Special electrode cleaning solutions (13-642-583) help with tough samples that coat your electrode bulb—proteins, bacteria, oils, and more Record temperature along with your pH value 888-358-4706 for your accumet specialist! (per NIST) Standardized Buffers pH Values of Buffers at Various Temperatures 0°C 1.67 4.00 6.98 7.11 9.46 10.32 12.79 5°C 1.67 4.00 6.95 7.08 9.40 10.25 12.73 10°C 1.67 4.00 6.92 7.06 9.33 10.18 12.67 20°C 1.67 4.00 6.87 7.01 9.23 10.06 12.52 25°C 1.68 4.01 6.86 7.00 9.18 10.01 12.46 30°C 1.68 4.02 6.85 6.98 9.14 9.97 12.36 40°C 1.69 4.04 6.84 6.97 9.07 9.89 12.17 50°C 1.71 4.06 6.83 6.97 9.01 9.83 11.96 60°C 1.72 4.09 6.84 6.97 8.96 9.79 11.73 70°C 1.74 4.13 6.85 6.99 8.92 9.78 11.47 80°C 1.77 4.16 6.86 7.03 8.89 9.78 11.19 90°C 1.79 4.21 6.88 7.08 8.85 9.80 10.89 Need Help? Call T e l : 1 - 8 0 0 - 7 6 6 - 7 0 0 0 ©2009 Thermo Fisher Scientific Inc. All rights reserved. Litho in U.S.A DC/MA 7/09 09_2129 BN0720094

Electrochemistry 101 with Dr. Seymour Blue Electrochemistry 101 with Dr. Seymour Blue Author: Fisher Scientific Subject: Tips for checking pH slope, cleaning probes, care and maintenance

  • Upload
    vuxuyen

  • View
    214

  • Download
    1

Embed Size (px)

Citation preview

Slopeat 20°C at 25°C

Status1pH unit 3pH units 1pH unit 3pH units

>102 >59.3 >178.0 >60.3 >181.0 Poor102% 59.3 178.0 60.3 181.0 OK101% 58.8 176.3 59.8 179.3 Very Good100% 58.2 174.5 59.2 177.5 Ideal99% 57.6 172.8 58.6 175.7 Very Good98% 57.0 171.0 58.0 173.9 Very Good97% 56.4 169.3 57.4 172.2 Very Good96% 55.8 167.5 56.8 170.4 Very Good95% 55.3 165.8 56.2 168.6 Very Good94% 57.0 171.0 58.0 173.9 OK93% 54.1 162.3 55.0 165.1 OK92% 53.5 160.5 54.4 163.3 OK<92 <53.5 <160.5 <54.4 <163.3 Poor

How to check your pH electrode slope1) Select the millivolt (mV) mode of your pH meter2) Using a pH electrode, obtain mV readings of two fresh

calibration buffers (i.e., pH 4.0 and pH 7.0 are best)3) Determine the net mV change4) Determine the net mV change per pH unit change and compare using the chart at right

Cleaning ProbesCleaning a dirty, clogged or coated electrode can restore proper electrode performance and prolong the useful life of the electrode. Here are some cleaning suggestions:

•OrganicOil/GreaseFilms—liquid dish soap and warm water are best. Isopropyl alcohol can be used on most electrodes if additional cleaning is needed. NOTE: Solvents such as acetone will damage plastic body electrodes!

•InorganicSaltDeposits—soak probe in 0.1M HCl as needed•Proteins—Use a protein-cleaning solution (13-642-584)

If you have a refillable probe, drain the chamber and refill with correct reference filling solution. Rinse the probe tip with distilled or deionized water and condition the probe in storage solution before using again.

pH Scale

BatteryAcid

LemonJuice

Vinegar

OrangeJuice

Milk

DistilledWater

BakingSodaSeaWater

MilkofMagnesia

Ammonia

Lye

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

IncreasingAcidity

IncreasingAlkalinity

Neutral

Battery Acid

Acid Rain

Lemon Juice

Vinegar

Milk

Baking SodaSea Water

Milk of Magnesia

Ammonia

Lye

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

AcidRain

IncreasingAcidity

Neutral

IncreasingAlkalinity

Fax: 1-800-926-1166

w

ww.f shersci.com

Did you know?Buffers should be at

the same temperature as your sample.

electrochemistry 101with Dr. Seymour Blue

For example, at 25°C: pH 4 = 170.5mV, pH 7 = -3.4mV: Net change = 170.5mV - (-3.4mV) = 173.9mV. Since pH 4 and pH 7 are 3 pH units apart, using the chart, 98% = Very Good

Care and Maintenance of Electrodes • Calibrate regularly with fresh buffers • Rinse electrode tip well between readings • On refillable electrodes be sure to check your fill level and open the fill

hole before calibration or measurement • Never rub the bulb to dry it; a static charge can develop! • Store your electrode with fill-hole closed in electrode storage solution

(Never store in deionized water!) • Special electrode cleaning solutions (13-642-583) help with

tough samples that coat your electrode bulb—proteins, bacteria, oils, and more

• Record temperature along with your pH value

888-358-4706 for your accumet specialist!

(per NIST) Standardized Buffers

pH Values of Buffers at Various Temperatures

0°C 1.67 4.00 6.98 7.11 9.46 10.32 12.795°C 1.67 4.00 6.95 7.08 9.40 10.25 12.73

10°C 1.67 4.00 6.92 7.06 9.33 10.18 12.6720°C 1.67 4.00 6.87 7.01 9.23 10.06 12.5225°C 1.68 4.01 6.86 7.00 9.18 10.01 12.4630°C 1.68 4.02 6.85 6.98 9.14 9.97 12.3640°C 1.69 4.04 6.84 6.97 9.07 9.89 12.1750°C 1.71 4.06 6.83 6.97 9.01 9.83 11.9660°C 1.72 4.09 6.84 6.97 8.96 9.79 11.7370°C 1.74 4.13 6.85 6.99 8.92 9.78 11.4780°C 1.77 4.16 6.86 7.03 8.89 9.78 11.1990°C 1.79 4.21 6.88 7.08 8.85 9.80 10.89

Need Help? Call

Tel: 1-800-766-700

0

©20

09 T

herm

o Fi

sher

Sci

entifi

c In

c. A

ll rig

hts

rese

rved

. L

itho

in U

.S.A

D

C/M

A

7/09

0

9_21

29

BN

0720

094