Electron Beam Therapy - Reloaded
Qiuwen Wu, Ph.D. Department of Radiation Oncology
Duke University
ET - Reloaded
Electron Photon Proton
• Depth dose and lateral profiles – Suitable for superficial tumors – Requires collimation close to skin
• Clinical e unchanged for decades and underutilized in cancer radiotherapy
• ~ 15% utilization
Introduction: e vs. γ vs. p
EAT – Electron Arc Therapy
• Electron beam delivered in partial arc
• Modifications to linac/cone and TPS – Shortened cone – Different PDDs – Cutout shape
• Patient-specific cast/shield and bolus
• Primarily for chest wall irradiation
• Univ. of Utah
Leavitt 1985 Electron ARC therapy: physical measurement and treatment planning techniques IJROBP 11
McNeely 1988 Electron arc therapy: chest wall irradiation of breast cancer patients IJROBP 14(6)
8/4/2016
2
ECT – Electron Conformal Therapy
• One or a few electron beams
• Bolus electron conformal therapy
– Tissue equivalent
– Modulate electron energy
– Keep PTV within 90% isodose volume
– Minimize dose to distal/underlying critical structures and normal tissue
Perkins 2001 A custom three-dimensional electron bolus technique
for optimization of postmastectomy irradiation IJROBP51
MERT – Modulated Electron Radiotherapy
• Multiple electron beams
• Energy modulation
– Bolus or energy selection
• Intensity modulation
– Cutout
– eMLC
– pMLC
– Scanning beam
Henzen et al 2014 Beamlet based direct aperture optimization for MERT using a photon MLC. Med Phys, 41
Henzen et al 2014 Monte Carlo based beam model using photon MLC for modulated electron radiotherapy. Med Phys, 41
: 6 MV e− : 6 / 9 / 12 MeV
12 segments
Standard MERT
DEAR – Dynamic Electron Arc Radiotherapy
• Electron radiation is delivered in ARC mode
• Electron applicator and cut-out are kept to provide lateral beam constriction
• Treatment couch is in simultaneous motion with gantry rotation to prevent collision. Beam always normal to skin and SSD = 100 cm
• Couch motion, gantry rotation, and dose rate are modulated to produce desirable dose distributions
Rodrigues, Yin, and Wu. "Dynamic Electron Arc Radiotherapy (DEAR): a feasibility study", Phys Med Biol, 2014. 59(2)
8/4/2016
3
DEAR design
DEAR design
Gantry
Couch
DEAR Photons
• DEAR: 6, 9, 12 MeV. • Photons: 6X, tangent, no wedge
Plan comparison: chest wall phantom
8/4/2016
4
DEAR delivery
Room View Patient View
DEAR delivery demo
-50
0
50
Gan
try
Ang
le (d
eg)
Expected
Actual
-4
-3
-2
-1
0
1
Cou
ch V
RT
pos
(cm
)
0 50 100 150 200-20
-15
-10
-5
0
Cou
ch L
AT
pos
(cm
)
Dose (MU)
Gantry Angle(deg)
Couch VRT (cm)
Couch LAT (cm)
0
1
2
3
4
Expected
Actual
-0.6
-0.4
-0.2
0
0.2
0.4
0 5 10 15 20 25 30-0.8
-0.6
-0.4
-0.2
0
Time (s)
Gantry Rot (deg/s)
Couch VRT (cm/s)
Couch LAT (cm/s)
Position Speed
Gantry and couch motion
8/4/2016
5
0
50
100
150
200
Dose
(MU)
Expected
Actual
120
125
130
135
140
Couc
h VRT
pos (
cm)
0 5 10 15 20 2595
100
105
110
Couc
h LAT
pos (
cm)
Time (s)
Beam Hold
Couch VRT (cm)
Couch LAT (cm)
Beam Hold
Beam hold
Dose (MU)
0
100
200
300
400
500
600
700
0 20 40 60 80 100 120
Do
se R
ate
(MU
/min
)
Time (s)
6e
Actual
Expected
0
100
200
300
400
500
600
700
0 20 40 60 80 100 120
Do
se R
ate
(MU
/min
)
Time (s)
6x
stdev = 1.6 MU/min stdev = 0.2 MU/min
Dose rate modulation
Static (16x10 cm2,G=0⁰)
depth 1.5 cm
DEAR (3x10 cm2,G=315-45⁰)
1.2 1.3
4.2 1.9
Dosimetry: 6e on cylindrical phantom
8/4/2016
6
• 6 MeV electron • D=1cm cutout • CR film • 148 CP (7MU/CP) • Beam hold (D->e)
Virtual scanning mode
Potential DEAR case
Photons DEAR
Need for fine energy resolution
8/4/2016
7
Small field electron beam dosimetry
Small field electron beam dosimetry
• Composite dose map
• 6 MeV
• 1x1 cm2 kernel
Small field dosimetry
• 6 MeV
• 10x10 cm2
• PDD
8/4/2016
8
Small field dosimetry
• 16 MeV
• 10x10 cm2
• PDD
DEAR Summary
Advantages • Uniform dose distributions over
large and curved targets while maintaining narrow penumbra
• Treated area > cone size
• DEAR delivery • Fixed cone/cutout
• DEAR delivery has high accuracy • Expected and actual plans agree
very well
• Trajectory log file can be used as a QC tool
• Dosimetry
Challenges • Small field dosimetry
• Not ready in clinical operation • Couch motion
• Coarse energy selection • Bolus
• Planning tools • Forward and inverse optimization
• Dynamic delivery collimators • Motorized electron collimator (MEC)
• Electron MLC (eMLC)
• QA Tools
Tech
no
log
y C
om
ple
xit
y
Photons Electrons
3DCRT ECT
ARC EAT
IMRT MERT
VMAT DEAR
FFF SFF
Summary: e vs. γ
VMATe
3D Printing
8/4/2016
9
Prototype eMLC
Eley, J. G., K. R. Hogstrom, et al. (2011). "Potential of discrete Gaussian edge feathering method for improving
abutment dosimetry in eMLC-delivered segmented-field electron conformal therapy." Medical Physics 38(12): 6610-6622. Looking to the Future
Antolak
Commercial eMLC
Gauer, T., D. Albers, et al. (2006). "Design of a computer-controlled multileaf collimator for advanced electron
radiotherapy." Physics in medicine and biology 51(23): 5987-6003. http://euromechanics.com/e_emlc.html https://www.youtube.com/watch?t=66&v=F0BBbHRrjBg
Antolak
8/4/2016
10
Slide 31 Hogstrom, 2003
Slide 32 Hogstrom, 2003