51
ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Embed Size (px)

Citation preview

Page 1: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

ELECTROPHYSIOLOGY LABEP: The Shocking Truth!

Nursing Grand Rounds

June 15, 2011

Kris Martin, RN

Julie MacDonald, RN

Page 2: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

WHERE ARE WE?

Wyman Ground

in between the OLD ICU and

Radiation Oncology

Page 3: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

What do we do?• Pacemaker and ICD implants

• Electrophysiology Studies

• TEE’s

• Cardioversions (Electrical and Chemical)

• Yearly Defibrillator testing

• Ablations- Atrial flutter and SVT, AVNRT

• Tilt-Table Testing

• Pacemaker/ICD Clinic

Page 4: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Why an EP study?Diagnoses abnormal heart rhythms and/or induces cardiac

arrhythmia's in order to:

• Determine the source of arrhythmia symptoms

• evaluate the effectiveness of certain medications in controlling the heart rhythm disorder

• predict the risk of a future cardiac event, such as Sudden Cardiac Death

• assess the need for an implantable device (ICD or pacemaker) or treatment procedure (ablation)

copyright Heart Rhythm Society 2004

Page 5: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

An Overview of the Procedure• Pt receives conscious sedation and local anesthesia.• Femoral VEIN is accessed-one or both sides.• EP catheters are guided under fluoroscopy to the

right atrium and ventricle via the IVC.• Electrodes at catheter tip collect data from various

points in the heart (electrical mapping).• Various arrhythmia's can then be induced or ablated,

or the decision to implant a pacemaker or ICD can be made.

Page 6: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

We Maintain a Sterile environment, same as the OR

Page 7: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Ablations• Radio frequency energy is delivered through EP

catheters to burn a small section of the myocardium.

• This prevents the arrhythmia from continuing or returning.

• At Mount Auburn, we do SVT and Atrial flutter ablations (Afib and VT ablations are not done here).

• EP studies and ablations are typically done as outpatient procedures.

Page 8: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

These are EP catheters seen on X-ray

RA

CS

RV

HIS

Page 9: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

TEE

• Transesophageal echocardiogram• Ultrasound probe passed into esophagus-posterior

view of heart.• Diagnoses thrombus, valve disease, septal defects

?source of embolus, bacteremia, or CVA/TIA.• Pre-cardioversion to rule out existing thrombus.• Patients get conscious sedation and generally need

to remain NPO for 4 hours after.

Page 10: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN
Page 11: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Cardioversion• Synchronized shock to terminate Afib or

flutter.

• Pts receive MAC (monitored anesthesia care-propofol) by anesthesiologist

• Chemical Cardioversions are when medications are given like Ibutilide or

Flecainide and EKG is monitored for conversion and arrhythmia.

Page 12: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Defibrillator Testing• DFTs: defibrillator threshold testing.• Usually done on ICD implant and yearly, or

if there is a potential device/lead issue.• Frequently outpatient basis.• Patient receives MAC (propofol).• The ICD programer is used to induce VF and

looks for the device to deliver a shock.

Page 13: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

ICD Pacemakers with leads

Page 14: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Pre-op Pacemaker/ICD implant• Pt must be NPO after midnight

• No applesauce with meds or

OJ for low blood sugar!

• AM meds are okay with a small sip of water

• Meds to hold:

heparin, multivitamins, oral hypoglycemics, insulin

• Check blood sugar if diabetic

• **LEFT sided IV**

Page 15: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Post-OP Care of Pacemaker/ ICD

• Watch site for S+S infection

pt will have prophylactic antibiotic • Watch site for hematoma or bleeding• Moderate amount of bleeding on DSD okay• Bedrest usually for 12 hours post op• Pain: Manual manipulation during implant

Percocet or Tylenol

Page 16: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Post Op Care• Keep arm in sling overnight, affected arm stays lower

than shoulder height, no extended reaching

this helps prevent dislodgment of leads!• Next day chest X-ray to confirm lead placement, and

device interrogation.• Pt will have a Green/purple folder in chart to take

home. Please make sure to give them to patient on discharge

• Call MD regarding procedure site, orders, function

of pacemaker.

Page 17: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Leads...• Thin wires with

flexible silicone coating

• Connect the pacemaker generator to the myocardium

• Sensing of heart activity and discharge of electrical impulse.

Page 18: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

• Leads are screwed into myocardium.

• 6-8 weeks for lead to mature: scar tissue forms around the lead and anchors it.

• Extended reaching or lifting during this time may dislodge the lead requiring re-do.

Page 19: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN
Page 20: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Pacemaker leads as seen on Xray-(Dual chamber system)

Atrial lead

Ventricular leads

Page 21: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Pacemakers and defibrillators are a complicatedbusiness and that is why we have specialists.

As a nurse you can not figure out what a pacemaker or ICD is doing by just looking at the monitor.

There are a few things that you should know that will help you decide what to do.

Page 22: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

What is a pacemaker or Defibrillator?

• Pacemaker is sensing and monitoring patients own heart rate and delivers an impulse in either the atrium or ventricle in the absence of a sensed beat.

• Defibrillator is sensing and monitoring the patients own heart rate for a fast rate usually set 188bpm to deliver a shock to convert to NSR. Can also pace if heart rate is slow usually 40 bpm.

Page 23: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

What do you need to know?• How many leads does the pacemaker have?

1 lead only sees what is happening in its chamber so a VVI pacemaker only knows what is happening in ventricle.

2 leads sees what is happening in both chambers the atrium and ventricle. Mode DDD/ DDI

• What is the lower rate set at?

If it is VVI @ 60 it will not pace if heart rate is

greater than 60bpm.

Page 24: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

A Dual chamber pacemaker or DDD will pacein Atrium at lower rate and then that sets up a timingcycle for the Ventricle to either pace or inhibit.In a DDD pacemaker you can see all different types of pacing on EGM strips:

A sensed V sensed = NSR/ AFIB, pacemaker is inbititedA paced V sensed = Sinus bradycardiaA sensed V paced = Heart blocks / CHB/2:1 blockA paced V paced = Heart block/Sinus arrest/asystole

Page 25: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Where will you find all this information?

You will find it in Meditech under CardiologyReports tab.If there is no report then they have not been seenat Mount Auburn Hospital Pacemaker Clinic.

Ask the patient/family if they have an ID card for theirpacemaker. Which will identify the manufacturer.Call pacemaker clinic @ 5622 to evaluate the device.

Page 26: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Pacemaker Report Pacemaker Report Clinic Tel: 617/499-5628

_____________________________________________________________________________

Date of Pacemaker check: 03/14/11

Pacemaker Type: Medtronic Sensia Implanted: 09/10/09

Pacemaker programmed rate: 70bpm Mode: DDDR

Response with magnet: Asynchronous pacing 3 at 100bpm then 85bpm

Presenting Rhythm: AV Paced

Underlying Rhythm: HB with PVC Pacemaker Dependent: N

Arrhythmia Recorded on Diagnostics: None

Battery Life Stable: Y

Impedance: A= 410 ohms RV= 531 ohms

Sensing: P-wave= 2.8 mv R-wave= 15.6 mv

LOC: A= 0.75 @ 0.4 ms RV= 0.75 @ 0.4 ms

Comments: In-house s/p fall. Appropriate sensing and capture testing. No

arrhythmia's recorded. Unable to promote intrinsic rhythm with AVD

at max, no changes made. Reprogramming for testing only. Follow

up in six months.

Page 27: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

When would you apply a magnet to a pacemaker? During electrocautery use in OR or GI Unit.

In a pacemaker a magnet causes the pacemaker to Asynchronously pace at a set rate usually 85bpm-100bpmWhat does this mean?It means that the pacemaker can no longer see what thepatients heart is doing and is going to pace at the magnet rate.If the patient is pacemaker dependent or has a HR below85-100 then you need to apply a magnet. If a magnet is not applied it could result in long pauses or asystole.

Page 28: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

What if the patient has a Higher HR?

If a patient has their own HR then you would not applya magnet since asynchronous pacing could cause apacing spike to come in on a T-wave resulting in VT!

Remember when you place the magnet on a pacemakerit can no longer see what the patient’s own heart is doing

A magnet should be available to place over device ifinhibition of pacing occurs resulting in pauses.

Page 29: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

When do you call pacemaker clinic?

This device is over sensing and thinks itsees ventricular activity and inhibitspacing.

Page 30: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Normal function

Non Sensing of R wave

Non Capture

What you might see on monitor

Page 31: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

What do you need to know about an ICD?

•You need to know if it is a single (VVI) or dual chamber(DDD) device and the lower rate.

• You need to know what the Zones are set at. VT zone or VF zone.

• The magnet response will always be: Inhibits therapies, no effect on pacemaker.

• If they are pacemaker dependent,they will need to be reprogrammed for surgery to an Asynchronous mode

Page 32: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

What do you need to know about a ICD?

• That it can overdrive pace a patient out of a fast HR if it falls into the VT zone.

• That it is not always able to distinguish between AFIB and VT especially if it is a single lead device. A VVI device doesn’t know what the atrium is doing.

• You can put a magnet over ICD to disable the therapies and prevent or stop inappropriate shocks.

Page 33: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

ICD Report ICD Report Clinic Tel: 617/499-5628________________________________________________________________________________Date of ICD check: 05/05/11ICD Type: Boston Scientific Cognis Implanted: 01/19/11VT Zone: 185 FVT: VFIB: 220ICD programmed rate: 60bpm Mode: DDDResponse with magnet: Inhibits therapies, no effect on pacemakerPresenting Rhythm: UnknownUnderlying Rhythm:Pacemaker Dependent:Arrhythmias Recorded on Diagnostics: noneBattery Life Stable: YImpedance: A= 439 ohms RV= 402 ohms LV= 949 ohmsSensing: P-wave= mv R-wave= 7.0 mv LV-wave= 19.3 mvLOC: A= @ ms RV= @ ms LV= @ msComments: Latitude transmission, BiV pacing 92% of the time. Follow up in

Page 34: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

This is what a magnet looks like some are blue

You can find them on all the code carts throughoutthe hospital, in the OR and GI units.

When would you place a magnet over an ICD?When would you place a magnet over an ICD?

If a patient is getting inappropriate shocks.

During cautery in OR or GI unit.

CMO status until device can be reprogrammed. A Doctors Order is need to apply a Magnet and disable a device.

Page 35: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

• If they are pacemaker dependent,they will need to be reprogrammed for surgery to an Asynchronous mode.

Remember that a Magnet only Inhibits therapies, has no effect on Pacemaker in an ICD Patient.

How do you know if they are pacemaker dependent?•Look at the report in Meditech.you will need to call Pacemaker clinic 5622 to arrange for someone to come and evaluate device and reprogram if necessary.

ICD Patients that are Pacemaker Dependent

Page 36: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

In OR, GI unit when cautery is being used the ICDshould be disabled by a magnet.Electrocautery can cause noise on the lead,an ICD would interpret it as fast heart rate and deliver an inappropriate shock.

The patient should always be on a monitor when the magnet is in place.The benefit of this is that if a VT/ VF episode occurs you can remove the magnet and the devicewill deliver a shock. If the device was programmed OFF then would need to externally shockthe patient. .

Page 37: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

This is an example of noiseon an ICD lead from Cautery.

The device thinks this is VF and would deliver a shock if a magnet was not applied.

Page 38: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

You can always call clinic, 5622 andwe will help you figure it out!

Page 39: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

If a patient has been made CMO you may place a magnet over device until it can be reprogrammed.

A note must be in the chart from attending physicianthat it has been discussed with patient and or family.

An order must be written to turn off ICD, before it can be done. If you are applying a magnet an order should be written.

This has no bearing on pacemaker portion of ICD and we do NOT turn off pacemakers in CMO patients.

Page 40: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

When would you apply a magnet to an ICD?•If a patient was receiving shocks that you/MD determine to be inappropriate you may apply a magnet to stop the therapies/shocks.•An example of this would be a patient who is in rapid AFIB with ventricular rate in 190bpm and has a single chamber device with a VT zone set for 188bpm. The ventricular lead only knows that the rate is above 188bpm and that it is programmed to give a shock.

Page 41: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Another example of when to place a magnet isthe patient is telling you they are getting Shockedand this is the rhythm on monitor/EKG.

This is NSR and if the patient is getting shocks from ICD, then you need to disable it with a magnetThe lead is most likely broken and over sensingthe device will continue to give shocks until it isdisabled.

Page 42: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

What if the device is not giving a shock and you see VT on monitor?The device will only do what it has been programmed to do

If a VT Zone is set at 188bpm and a patient is in a VT@a180bpm the device is never going to give a shock.You would need to externally shock the patient. The pads should be placed to avoid damaging the ICD.

Page 43: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

What if your patient tells you his device is BEEPINGor VIBRATING in their Chest?

They are NOT crazy! The devices are set up to alert patientsof a problem.You should notify the pacemaker clinic or cardiologist on call. They can call in a company representative to interrogate and find the problem.The beeping and or vibrating will continue every 6-10hours until device is reprogrammed.8am 2pm 8pm

Page 44: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

Why is Pacemaker/ ICD follow-up important?

We often find new onset AFIB. Device can date and timethe event, amount and duration of episodes and give atrial and ventricular rates.

Page 45: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

We evaluate Histograms to tell us if the heart rate has been fast or if it has been stuck at the lower rate. Medications areoften adjusted or Rate Response added.

0

5

10

15

20

25

30

35

40

45

50

% o

f B

eats

<40

50

60

70

80

90

100

110

120

130

140

150

160

170

180

180>

Rate (bpm)

Ventricular Long term Histogram

Sensed

Paced

Page 46: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

We find broken leads or batteries that need replacement.

Non capture orintermittentcapture

broken lead on x-ray

Thrombus/emboli

Page 47: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

We find non-sustained episodes of VT in pacemaker patients that otherwise would go undetected. This couldbe causing them dizziness or syncope.

New onset Afib is common to find also.

Page 48: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

In an ICD we can find episodes that received ATP or a shock. After reviewing with the cardiologist, their meds may be adjusted to prevent further therapies.

Page 49: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

In an ICD we can find episodes of AF getting inappropriatetherapies. Medications would be evaluated and adjusted.

Atrialventshock

Page 50: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

This is another exampleof inappropriate shock forSinus Tachycardiaand the patient should have medications adjustedto prevent their sinus ratefrom going this high.

Page 51: ELECTROPHYSIOLOGY LAB EP: The Shocking Truth! Nursing Grand Rounds June 15, 2011 Kris Martin, RN Julie MacDonald, RN

THANK YOU FOR COMING

The EP Staff

QUESTIONS?