Elemental, My Dear Elemental, My Dear WatsonWatson
Paul MiddlesteadUniversity of Ottawa, G.G. Hatch Laboratory
For 20th Continuous Flow Conference, Davis, 2014
This talk is for new usersThis talk is for new users We will touch the basics on Elemental We will touch the basics on Elemental
analysersanalysers A mix bag of tricks and advicesA mix bag of tricks and advices Refer to manufacturer’s instruction manualsRefer to manufacturer’s instruction manuals
Only endorsed products are: Only endorsed products are: Rickards Rickards RedRed and and Rickards White !Rickards White !
Who/what is this talk Who/what is this talk for?for?
Meet my familyMeet my family
Combustion
EA
NCHS
TC/EA
Pyrolysis
HO
Largecapacity
S only TOC Stableisotopes
15000C12000C
Leco is the most popular manufacturer of Leco is the most popular manufacturer of elemental analysers. elemental analysers. EveryEvery department department has one or two of those accumulating dust.has one or two of those accumulating dust.
Static combustion using oxygen: not Static combustion using oxygen: not suitable for IRMS.suitable for IRMS.
Meet Grand PaMeet Grand Pa
•Elemental Analyser as an…instrument•Schematics•Autosamplers•Flash combustion•Chemicals & Configurations•Gas Chromatography•Thermal Conductivity Detector•Typical run
•Common problems•EA-IRMS
Let’s touch onLet’s touch on
Simultaneous determination of Nitrogen, Simultaneous determination of Nitrogen, Carbon, Hydrogen, Sulfur, OxygenCarbon, Hydrogen, Sulfur, Oxygen
Measuring range: 100 ppm to 100%Measuring range: 100 ppm to 100% Sample size: 0.1 to 1000 mg / 0.1 to 25 ulSample size: 0.1 to 1000 mg / 0.1 to 25 ul Detection limit: 10 ppmDetection limit: 10 ppm Accuracy: 0.3% - 0.02% absoluteAccuracy: 0.3% - 0.02% absolute
EA as an …. instrumentEA as an …. instrument
CE instruments (Carlo Erba) models CE instruments (Carlo Erba) models 1108, 1110, NA 1500, NA 2100, Flash 1108, 1110, NA 1500, NA 2100, Flash 1112, Flash 20001112, Flash 2000
Costech Model ECS4010Costech Model ECS4010 Elementar cube familyElementar cube family Eurovector models EuroEa3028-HT, Eurovector models EuroEa3028-HT,
Ea3024-IRMS, pyrolysis modelEa3024-IRMS, pyrolysis model Sercon-Integra2 Sercon-Integra2 TC\EA ThermoTC\EA Thermo ……
Manufacturers, models Manufacturers, models blablablablablabla
EA : Bulk analysis of EA : Bulk analysis of NCHS/ONCHS/O
Organic compoundsOrganic compounds PharmaceuticalsPharmaceuticals OrganometallicsOrganometallics PetrochemicalsPetrochemicals Gasoline & fuelsGasoline & fuels GraphiteGraphite Carbides & nitridesCarbides & nitrides
Metals & alloysMetals & alloys PolymersPolymers ExplosivesExplosives HydrocarbonsHydrocarbons SoilsSoils Coal & cokeCoal & coke LiquidsLiquids
In short, you can combust your grandmother!
Ref gas 1
Elemental Analyser
(CNS)(OH)
Interface Isotope Ratio Mass Spectrometer
HeRef gas 2 2 m
Typical EA setup with Typical EA setup with IRMSIRMS
EA pictureEA picture
Elemental Analyser Elemental Analyser SchematicSchematic
TCD
CombustionChemicals1020C
ReductionChemicals650C
Gas chromatographyColumn in oven
Thermal Conduc-tivity Detector
Autosampler
N2
H2O
CO2
SO2
HeO2
WaterTrap EA line
out
Found the problem with Found the problem with the previous slide?the previous slide?
A beer if you do…A beer if you do…(5 seconds)(5 seconds)
Elemental Analyser Elemental Analyser SchematicSchematic
TCD
CombustionChemicals1020C
ReductionChemicals700C
Gas chromatographyColumn in oven
ThermocoupleDetector
Autosampler
N2
H2O
CO2
SO2
HeO2
WaterTrap EA line
out
A cold A cold Rickard’s RedRickard’s Red for me please.for me please.
1.1. Sample/Std is weighed and wrapped in tin Sample/Std is weighed and wrapped in tin foil, placed in culture tray/autosamplerfoil, placed in culture tray/autosampler
2.2. Sample drops is flash-combusted in OSample drops is flash-combusted in O22 in in He stream, causing rapid cracking and He stream, causing rapid cracking and oxidation of gases stoichiometrically oxidation of gases stoichiometrically equivalent to their elemental components, equivalent to their elemental components, some gas oxides need to be reduced before some gas oxides need to be reduced before obtaining final products (Nobtaining final products (N22, CO, CO22, H, H22O, SOO, SO22) ) i.e. oxidation and reducing reactions i.e. oxidation and reducing reactions involvedinvolved
Typical Analysis Procedure for Typical Analysis Procedure for EAEA
Typical Analysis Procedure for Typical Analysis Procedure for EAEA
3. Separation of gases by gas chromatography or chemical column traps
4. Detection by thermal conductivity detector (TCD) or IRMS.
5. Curve made from weighing certified stds (for isotope and/or quantitative measurements)
Some torture tools…Some torture tools…
Well well, got culture?Well well, got culture?
SSuuppeerr ssiizzee mmee
MAIS C’EST IMPOSSIBLE!MAIS C’EST IMPOSSIBLE!
C’EST FANTASTIQUE!C’EST FANTASTIQUE!
Filters? Careful!
Not flat pleaseNot flat please
Autosamplers (AS200)Autosamplers (AS200)
Zero BlankZero Blank
Zero Blank modifiedZero Blank modified
Special tools requiredSpecial tools required
ECS 4010
Take 20 minutes to study the gas schematics of your
instrument
Take 20 minutes to study the gas schematics of your
instrument
Be good to yourselfBe good to yourself
Sample droppingSample dropping
0 10 30 40
START
Sample drops
Combustion starts Injected O2Flash should occurwithin 2-3 secondsof being dropped
20
seconds
• PLATINUMPLATINUM: Absorbs much of the heat of the : Absorbs much of the heat of the reactor before passing it on to the sample, slowing reactor before passing it on to the sample, slowing down reactiondown reaction
• ALUMINIUMALUMINIUM: Supplies good oxidation flash and : Supplies good oxidation flash and prime sample ignition but does not melt or mix prime sample ignition but does not melt or mix with the sample thus does not promote inner with the sample thus does not promote inner oxidationoxidation
• SILVERSILVER: Melts at 960: Melts at 96000C, does not take part in the C, does not take part in the combustion, retains trace of carbon when in combustion, retains trace of carbon when in molten statemolten state
• TINTIN: Inexpensive & takes active part in process. : Inexpensive & takes active part in process. Melts at 235Melts at 23500C with very low enthalpy, intermixes C with very low enthalpy, intermixes with organic and inorganic substances, expedites with organic and inorganic substances, expedites the final oxidation reactionthe final oxidation reaction
Why you Tin man?Why you Tin man?
CC CO CO2 2 COCO22
Let’s get crackingLet’s get cracking
HH HH22OO HH22OO
SS SO SO2 2 + SO+ SO33 SOSO22
Combustion tube
Reduction tube
NN N N2 2 + NxOx+ NxOx N N22
Chemicals and Chemicals and ConfigurationsConfigurations
• System should be optimized for element(s) to be analyzed
• One-tube system: both oxidant catalyst and reducing chemicals in one tube; usually used if S is to be analyzed
• Two-tube system: one tube for oxidant, one tube for reducing chemicals; usually used for N, NC or NCH
• Chemical traps: Mg Perchlorate or Anhydrone will trap H2O, and Carbosorb will trap CO2 (and SO2)
ConfigurationConfigurationCNCN
ConfigurationConfigurationSS
Small is Small is beautifulbeautiful
Using grinding tools and diluted Nitric acid, one can re-use Using grinding tools and diluted Nitric acid, one can re-use combustion and reduction tubescombustion and reduction tubes
More torture toolsMore torture tools
Gas chromatographic columns will separate different Gas chromatographic columns will separate different components according to their polarity and molecular components according to their polarity and molecular size.size.
Factors influencing the quality of the chromatography: Factors influencing the quality of the chromatography: column length, size of packing, tube diameter, column length, size of packing, tube diameter, stationary stationary phase type, flow rate, temperature.phase type, flow rate, temperature.
Packed column:Packed column: packed polymer beads, different sizes packed polymer beads, different sizes available. High capacity , low resolution.available. High capacity , low resolution.
Capillary column:Capillary column: small capillary with polymeric film on small capillary with polymeric film on inner wall. High resolution, low capacity. inner wall. High resolution, low capacity.
Deactivated fused silica is free of adsorption problems Deactivated fused silica is free of adsorption problems encountered with most packings or capillaries columns.encountered with most packings or capillaries columns.
Gas ChromatographyGas Chromatography
Gas Separation/ GCGas Separation/ GC
He + N2 , CO2 , H2O , SO2 + He
He + SO2 + H2O + CO2 + N2 + He
INCREASING POLARITYOF GASES
What after we separate?What after we separate?
He + SO2 + H2O + CO2 + N2 + He
DETECTOR
TIME
DE
TE
CT
OR N2
CO2
SO2
H20
Gas Separation / Purge & Gas Separation / Purge & TrapTrap
TCD
SO2 H2O CO2
Thermal conductivity detector (TCD)Thermal conductivity detector (TCD) Heated filament from which heat is removed at Heated filament from which heat is removed at a constant rate by He gas stream. Change in a constant rate by He gas stream. Change in heat transfer is caused by presence of analyte heat transfer is caused by presence of analyte molecules with different thermal conductivities molecules with different thermal conductivities than He. Relatively low sensitivity, excellent than He. Relatively low sensitivity, excellent range and linearity. Non-destructive.range and linearity. Non-destructive.Electron capture detector (ECD)Electron capture detector (ECD)Electrons are captured by organic species in Electrons are captured by organic species in ionized carrier; used for trace sulfur ionized carrier; used for trace sulfur determination.determination.Isotope Ratio Mass Spectrometer (IRMS)Isotope Ratio Mass Spectrometer (IRMS)
DetectorsDetectors
Typical output of EA, text Typical output of EA, text bookbook
N2
CO2
H2OSO2
Balance: Get the best Balance: Get the best
A 0.01mg readability A 0.01mg readability translates in an error of translates in an error of 0.25% on a 2 mg sample 0.25% on a 2 mg sample (on 100% carbon)(on 100% carbon)
BalanceBalance
IAEA (and NIST) IAEA (and NIST) standardsstandards
Links available on Isogeochem
Best : %N, Best : %N, 1515N, %C, N, %C, 1313CC
Test different materials and pray Test different materials and pray
Make your own:Make your own:
CaffeineCaffeine
L-glutamic acidL-glutamic acid
Mixtures such as sucrose + potassium nitrateMixtures such as sucrose + potassium nitrate
Mix in solution of natural + Mix in solution of natural + enriched/depleted material then dried, enriched/depleted material then dried, powdered, sievedpowdered, sieved
STD: Making it on your STD: Making it on your own…own…
STD: Not making it on your STD: Not making it on your own…own…
Can also contact other university labs to Can also contact other university labs to buy or obtain their internal standards. buy or obtain their internal standards.
STD: Not making it on your own…STD: Not making it on your own…
ARNDT SCHIMMELMANNIndiana University http://mypage.iu.edu/~aschimme/hc.htm
Calculation MethodsCalculation Methods
1. K-Factor (Single point calibration with blank substraction)
Calculated % = K * (Area unk – Area blk)Weight unk
Where K = Weight std * Theor std %Area std - Area blk
Calculation MethodsCalculation Methods
2. Linear Fit (Least square linear regression)
Calculated % = m * Area unk + b Weight unk
Peak Area
Wt *
Th
std
%
Calculation MethodsCalculation Methods
3. Quadratic Fit (Least square quadratic regression)
Calculated % =
a * Area unk2 + b * Area unk + c Weight unk
Peak Area
Wt *
Th
std
%
An update…An update…
Most elemental analysers manufacturershave considerably refined their calculation methods, with more complex algorithms, low-high ranges and statistical tools.
Leak check (1)Leak check (1)
EA
Pressure check
•Cap exit port•Increase P of He to 1.3 bar, wait 3 minutes•Close regulator•Pressure gauge should not move for 5 minutes
EA
Flow check (requires electronic mass flow controller)
•Cap exit port•Monitor flow for 3 minutes•Flow should drop to 0 ml/min
Leak check (2)Leak check (2)
EA Inter-face
Leak check with IRMS
•EA has been leak checked•EA chemicals have outgassed (12 hours) •Tune to Ar (mass 40)•Should meet manufacturer’s requirements•Use Ar cylinder to hunt
MS
Ar
Elemental Analyser Schematic for Elemental Analyser Schematic for IRMSIRMS
TCD
Combustion tube Reduction tube
Water trap
Gas chromatographyColumn
ThermocoupleDetector
Autosampler
To MS
N2 CO2InterfaceDiluter & ref gas
Water is removed via Mg Perchlorate/sicapent Water is removed via Mg Perchlorate/sicapent traptrap---water and mass spectrometer ---water and mass spectrometer do not mixdo not mix--------
Configuration should be optimized for gas of Configuration should be optimized for gas of interestinterest
Leak free (mass 28 & 40, use Argon as leak probe)Leak free (mass 28 & 40, use Argon as leak probe) Low and stable background (mass Low and stable background (mass 2828, 18, 40, , 18, 40, 4444)) Dynamic range must be respected, do use target Dynamic range must be respected, do use target
beambeam Best sequence is carefully planned ie known Best sequence is carefully planned ie known
concentration of samplesconcentration of samples Garbage in……Garbage outGarbage in……Garbage out
Considerations for EA-IRMSConsiderations for EA-IRMS
Some useful infoSome useful infoLab No. Sample type Nitrogen Carbon Hydrogen Sulphur C/N Comments
97-256-010 Adult mysids 10,20 50,97 na na 5,00 Whall JD, Trent U.
97-220-003 Amphipods 6,81 33,20 5,66 0,00 4,88 Pastershank G., U of O/ Biology
97-256-004 Chironomids 8,98 37,23 na na 4,15 Whall JD, Trent U.
97-256-014 Clam 12,30 48,91 na na 3,98 Whall JD, Trent U.
97-220-004 Clam muscle 11,26 43,41 6,77 1,10 3,85 Pastershank G., U of O/ Biology
97-220-005Clam shell
0,05 11,91 0,19 0,00231,6
3Pastershank G., U of O/ Biology
97-220-018 Cragon (shrimp) 10,03 37,22 5,91 0,00 3,71 Pastershank G., U of O/ Biology
97-220-020 Diatom 0,00 0,74 0,09 0,00 N.A. Bad trace
97-220-002 Egg white 11,85 45,83 6,93 1,14 3,87 Pastershank G., U of O/ Biology
97-220-001 Egg yolk 5,41 62,69 9,73 0,00 11,59 Pastershank G., U of O/ Biology
97-220-009 Fish bone 6,43 19,53 3,51 0,00 3,04 Pastershank G., U of O/ Biology
97-220-008 Fish eye 12,90 48,70 7,05 2,13 3,77 Pastershank G., U of O/ Biology
97-220-013 Fish fin 10,01 34,36 5,35 0,00 3,43 Pastershank G., U of O/ Biology
97-220-014 Fish gill 13,28 47,10 7,06 0,00 3,55 Pastershank G., U of O/ Biology
97-220-015 Fish gut content 9,56 44,88 7,07 1,13 4,69 Pastershank G., U of O/ Biology
Best are quartz filters; they are stable. However Best are quartz filters; they are stable. However they are more expensive and offer less choice.they are more expensive and offer less choice.
Silver filters are great too.Silver filters are great too. Mostly glass filters, cheap, huge choice.Mostly glass filters, cheap, huge choice. How much to use? How much to use?
Filters, we can do that!Filters, we can do that!
Filters, we can do that!Filters, we can do that!
Area = *R2
In this case, a punch is about 16mm2 and our whole filter is about 1490mm2 .Our punch hole is roughly 1% of the filter.
Sediment contaminationSediment contamination
Inorganic Carbon
Average 13C: 0‰
= Total CarbonOrganic Carbon
Average 13C: -25‰
0% + 100% = -25 ‰1% + 99% = -
24.75 ‰5% + 95% = -23.75 ‰10% + 90% = -22.50 ‰50% + 50% = -12.5 ‰
Detection Limit? Hmmmm…Detection Limit? Hmmmm…
Every manufacturer has them, frankly they are all similar.In the end, you want an instrument that is linear with a decent sensitivity and a nice sample peak.Your source should be adjusted accordingly and you may have to sacrifice sensitivity in order to obtain stability and linearity. Hence, the detection limit may change according to the source parameters, the EA configuration and the condition of the chemicals.Be conservative.
Optimal amount for meOptimal amount for me
% Nitrogen Wt required
100 0.1mg
10 1.0mg
5 5mg
2 7mg
1 10mg
0.1 100mg
0.001 Yeah, right!
Optimal amount of Nitrogen: 0.1mg
Open split, magical stuffOpen split, magical stuff
He + CO2 + N2 + He
EA 100 ml/min
Fused silica capillary100 um, 2m, insertedabout 30cm in EA line,0.4 ml/min going intosource of IRMS
2mm or 1/16” SSline from EA
In real life…Open splitIn real life…Open split
EA
IRMSRef 1
Ref 2
He
Ref 3
The joy of sniffing EA onlyThe joy of sniffing EA only
Mass spectrometer outputMass spectrometer output
TCD output from EATCD output from EA
MagnetMagnetPeakjumPeakjumpp
N2
CO2
Mass 28, 29Mass 44,45,46
The joy of sniffing, totalThe joy of sniffing, total
Mass spectrometer outputMass spectrometer output
TCD output from EATCD output from EA
MagnetMagnetPeakjumPeakjumpp
N2
CO2
Mass 28, 29Mass 44,45,46
Masses 28 & 29 Masses 44, 45 & 46
To dilute or Not to dilute?To dilute or Not to dilute?
A useful suggestion…A useful suggestion…
I strongly recommend that you create a basic document for your users explaining
the limits and pitfalls of EA-IRMS analysis, this will save you (and your users) an
enormous amount of time.
Removing inorganic carbon is not trivial.Removing inorganic carbon is not trivial.
Alkalies (Sodium, Potassium, Calcium) are Alkalies (Sodium, Potassium, Calcium) are difficult to combust, catalist is definitively difficult to combust, catalist is definitively required to bind and help with oxydation.required to bind and help with oxydation.
Oxygen contribution to Oxygen contribution to 3434S is still a S is still a problem, especially with organic samples. problem, especially with organic samples.
Some unresolved issues Some unresolved issues with EAwith EA
Last suggestion and Last suggestion and only Official only Official
EndorsementEndorsement
The only product fully and officially endorsed by the author is:
Rickard’s White Rickard’s White
Thank youThank you