Analyt icalInstrumentsfor Sc ience
ELEMENTALSPECTROSCOPY
PRODUCT BROCHURE
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 2
ContentsPAGE NO.
Introduction 5
Detection Limits 7
AA500 Atomic Absorption Spectrometer 8
AAWin3 Software – NOW GLP Compliant 12
AA990F Atomic Absorption Spectrometer 18
AAWin2 Software 20
Atomic Absorption Spectrometer Qualification Kit 24
Atomic Fluorescence Spectroscopy 26
AF402 Simultaneous Atomic Fluorescence Spectrometer 28
PFWin Software 31
FP902 Flame Photometer 34
Consumables 38
3
Analyt icalInstrumentsfor Sc ience
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 4
4
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 5
AAS is an analytical procedure for the Qualitative and
Quantitative determination of chemical elements in a
sample. The technique is based on the Lambert-beer
principle which states that the absorbance of a solution
(A) is directly proportional to its path length (I) and its
concentration (C) when the wavelength of incidence
light remains fixed. This is summarized in the following
equation, where (ε) is the molar absorbtivity.
A= εIc
PG Instruments offer a range of instrumentation
designed to meet the requirements of all industry
sectors including Agricultural, Clinical, Environmental,
Food, Metal, Mining and Materials, Petrochemical
and Pharmaceutical. Further information about the
technique and products can be found in this brochure.
AAS Principle
Atomic Absorption Spectrophotometers (AAS) have
become an industry standard for many elemental
determinations within specific industries. There are
three AAS techniques including Flame, Graphite
Furnace, and Hydride Generation, all of which have
their own advantages and disadvantages due to
analytical limitations. The flame technique, being the
most common, will determine trace levels (usually ppm)
for most analytical elements and it is also the most cost
effective technique. If determinations of lower levels
are required (usually ppb) the graphite technique can
be used. There are a number of elements that form
volatile hydrides and these can be analysed using the
hydride technique.
Atomic Absorption Analysis
A hollow cathode lamp containing a coated cathode of
the element that is to be analysed is used as a light
source. The light source emits a beam of a specific
wavelength through the flame and into the
monochromator. The sample is aspirated into the flame
in the form of an aerosol which is converted to an
atomic vapour, atoms present in the vapour are then
raised from their ground state to their excited state. The
atoms absorb light of a specific wavelength when
moving to an excited energy state, this causes the
quantity of light absorbed to increase. The light
absorption then can be measured and used for a
Quantitative determination of the amount of analyte in
a sample.
Atomic Emission Analysis
For the determination of certain elements, usually those
excited at high wavelengths, a variation of the standard
flame technique can be performed which is called
atomic emission. As the technique relies on the spectral
emission from the flame there is no requirement for the
hollow cathode lamp. The sample is aspirated into the
flame which causes the ground state atoms to convert
to an excited state. As the excited state is very unstable
the atoms return to their ground state and emit an
emission spectrum consisting of a number of emission
wavelengths. The intensity of an emission line will
therefore increase as the number of excited atoms of
the specific element increases. The specific wavelength
of the emission can be used for a Quantitative
analysis to determine the concentration of a element in
a solution.
Instrumentation
The Atomic Absorption Spectrometer is made up of five
essential components. Each component is designed to
produce minimum disruption to the overall system and
many design features are installed to keep the signal to
noise ratio as low as possible.
l Light Source
l Burner Assembly
l Optics
l Detector
l Signal Processing
Since its development in the 1950’s the Atomic Absorption Spectrometer(AAS) has evolved into an accurate and reliable analytical tool usedextensively in today’s modern laboratory.
5
Atomic Absorption Spectroscopy
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 6
Techniques
There are three main techniques for AAS analysis each
having their strengths and weaknesses depending on
the type of analysis required. There are a number of
factors to consider when deciding which technique will
suit the application. For example:
l Detection limits l Working range
l Speed of analysis l Automation
l Sample volume l Sample matrix
l Equipment cost l Precision of results
l Element suitability l Available space required
l Clean area l Method development
l Running costs
Flame Analysis
Flame AAS is the standard technique used for general
purpose analysis. The readings are normally accepted to
be trace levels of ppm and nearly all elements can be
analysed to this level. There is a choice of Air/Acetylene
flame used for easily ionised elements and Nitrous
Oxide/Acetylene flame for the less easily excited
elements. The flame technique requires a number of
extra components to complete the installation.
l Oil free Air Compressor
l Ultra pure Acetylene Gas
l Nitrous Oxide Gas (optional)
l Natural Gas/LPG (optional)
l Autosampler (optional)
l Suitable Fume Extraction System
Graphite Analysis
The graphite analysis technique uses an assembly
containing a graphite tube to replace the burner. This
technique is used to analyse very low levels with low
sample volumes. Although the detection limit is very low,
the precision is almost always higher than that of the
flame technique. The graphite technique requires a
number of extra components to complete the installation.
l Water Circulating System for Cooling.
l Pure Argon Gas.
l High Current Power Supply
l Autosampler (optional)
l Suitable Fume Extraction System
Hydride Analysis
The hydride technique requires an additional unit to
add to the flame AAS system. The Hydride Generator is
used to produce a hydride vapour, for the hydride
forming elements, by means of a heated tube which fits
directly to the top of the flame burner. The unit can also
be used for cold vapour technique primarily for Mercury
analysis. The hydride technique requires the following
extra components to complete the installation.
l Pure Argon gas
l Suitable fume extraction system
6
1H
2He
3Li
11Na
19K
37Rb
55Cs
87Fr
4Be
12Mg
5B
13Al
20Ca
38Sr
56Ba
88Ra
21Sc
39Y
22Ti
40Zr
72Hf
104Rf
23V
41Nb
73Ta
105Db
24Cr
42Mo
74W
106Sg
25Mn
43Tc
75Re
107Bh
25Mn
43Tc
75Re
107Bh
25Mn
43Tc
75Re
107Bh
25Mn
43Tc
75Re
107Bh
25Mn
43Tc
75Re
107Bh
25Mn
43Tc
75Re
107Bh
25Mn
43Tc
75Re
107Bh
25Mn
43Tc
75Re
107Bh
25Mn
43Tc
75Re
107Bh
26Fe
44Ru
76Os
108Hs
26Fe
44Ru
76Os
108Hs
26Fe
44Ru
76Os
108Hs
26Fe
44Ru
76Os
108Hs
26Fe
44Ru
76Os
108Hs
26Fe
44Ru
76Os
108Hs
26Fe
44Ru
76Os
108Hs
26Fe
44Ru
76Os
108Hs
26Fe
44Ru
76Os
108Hs
26Fe
44Ru
76Os
108Hs
27Co
45Rh
77Ir
109Mt
28Ni
46Pd
78Pt
110Ds
29Cu
47Ag
79Au
111Rg
30Zn
48Cd
80Hg
112UUb
31Ga
49In
81Tl
113Uut
6C
14Si
32Ge
50Sn
82Pb
114Uuq
7N
15P
33As
51Sb
83Bi
115Uup
8O
16S
34Se
52Te
84Po
116Uuh
9F
17Cl
35Br
53I
85At
117Uus
10Ne
18Ar
36Kr
54Xe
86Rn
118Uuo
57La
89Ac
58Ce
90Th
59Pr
91Pa
75Re
107Bh
75Re
107Bh
75Re
107Bh
75Re
107Bh
75Re
107Bh
75Re
107Bh
75Re
107Bh
75Re
107Bh
76Os
108Hs
76Os
108Hs
76Os
108Hs
76Os
108Hs
76Os
108Hs
76Os
108Hs
76Os
108Hs
76Os
108Hs
76Os
108Hs
60Nd
92U
61Pm
93Np
62Sm
94Pu
63Eu
95Am
64Gd
96Cm
65Tb
97Bk
66Dy
98Cf
67Ho
99Es
68Er
100Fm
69Tm
101Md
70Yb
102No
71Lu
103Lr
Lanthanides
Actinides
GroupPeriod
1
2
3
4
5
6
7
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
Elements – Periodic Table
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 7
Detection limitsElement Wavelength Air/C2H2 N2O/C2H2/C2H2 Air/LPG Graphite Vapour
(nm) mg/l(ppm) mg/l(ppm) mg/l(ppm) µg/l(ppb) µg/l(ppb)
Aluminium (Al) 309.3 n/a 0.028 n/a 0.13 n/a
Antimony (Sb) 217.6 0.092 n/a 0.21 0.29 0.26
Arsenic (As) 193.7 0.12 n/a n/a 0.33 0.05
Barium (Ba) 553.6 n/a 0.031 n/a 0.23 n/a
Beryllium (Be) 234.9 n/a 0.0036 n/a 0.02 n/a
Bismuth (Bi) 223.1 0.0049 n/a 0.03 0.26 0.19
Boron (B) 249.7 n/a 3.5 n/a n/a n/a
Cadmium (Cd) 228.8 0.0028 n/a 0.008 0.01 n/a
Calcium (Ca) 422.7 n/a 0.0037 0.07 0.60 n/a
Chromium (Cr) 357.9 0.0054 n/a 0.6 0.025 n/a
Cobalt (Co) 240.7 0.01 n/a 0.018 2.00 n/a
Copper (Cu) 324.8 0.0045 n/a 0.0026 0.065 n/a
Dysprosium (Dy) 421.7 n/a 0.2 n/a n/a n/a
Europium (Eu) 459.4 n/a 0.04 n/a n/a n/a
Gold (Au) 242.8 0.013 n/a 0.026 3.00 n/a
Indium (In) 303.9 0.07 n/a 0.04 n/a n/a
Iron (Fe) 248.3 0.0043 n/a 0.015 0.06 n/a
Lead (Pb) 217.0 0.013 n/a 0.02 0.03 n/a
Lead (Pb) 283.3 0.016 n/a 0.04 0.08 n/a
Lithium (Li) 670.8 0.0021 n/a 0.004 0.70 n/a
Magnesium (Mn) 285.2 0.0022 n/a 0.002 0.09 n/a
Manganese (Mn) 279.5 0.0016 n/a 0.0019 0.03 n/a
Mercury (Hg) 253.7 n/a n/a n/a 40.0 0.15
Molybdenum(Mo) 313.3 n/a 0.021 n/a 0.14 n/a
Nickel (Ni) 232.0 0.008 n/a 0.012 0.065 n/a
Palladium (Pd) 247.6 0.03 n/a 0.018 0.04 n/a
Phosphorus (P) 213.6 n/a 10.0 n/a n/a n/a
Platinum (Pt) 265.9 0.12 n/a 0.22 0.04 n/a
Potassium (K) 766.5 0.0009 n/a 0.0041 0.40 n/a
Rhodium (Rh) 343.5 0.009 n/a 0.037 0.02 n/a
Rubidium (Rb) 780.0 0.003 n/a 0.005 n/a n/a
Scandium (Sc) 391.2 n/a 0.02 n/a n/a n/a
Selenium (Se) 196.0 0.23 n/a n/a 0.32 0.27
Silicon (Si) 251.6 n/a 0.30 n/a 17.0 n/a
Silver (Ag) 328.1 0.0032 n/a 0.0023 0.04 n/a
Sodium (Na) 589.0 0.0037 n/a 0.0086 n/a n/a
Strontium (Sr) 460.7 n/a 0.005 n/a 0.01 n/a
Tantalum (Ta) 271.5 n/a 2.2 n/a n/a n/a
Tellurium (Te) 214.3 0.054 n/a n/a 3.20 0.43
Thallium (Tl) 276.8 0.014 n/a n/a 0.015 n/a
Tin (Sn) 224.6 n/a 0.21 n/a 15.0 1.39
Titanium (Ti) 365.4 n/a 0.05 n/a 0.5 n/a
Tungsten (W) 255.1 n/a 1.2 n/a n/a n/a
Vanadium (V) 318.5 n/a 0.11 n/a 0.03 n/a
Zinc (Zn) 213.9 0.0033 n/a 0.0037 0.08 n/a
7
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 8
8
ATOM IC A B SOR P T I ON S P EC T ROMET E R
AA500
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 9
The AA500 Atomic Absorption Spectrometer is
a high performance automated instrument
designed to meet the requirements of the
modern laboratory. Due to its versatility and
performance it can be used for a wide range of
applications including:
l Agricultural l Clinical
l Environmental l Food
l Metal l Mining
l Geological l Petrochemical
l Pharmaceutical
The versatile instrument is available in threeconfigurations:
AA500F – The instrument is equipped with a flame
atomiser only. The positioning of which is fully controlled by
the embedded computer system and AA Win software.
Three flame options are available to the user with the
Air/Acetylene being the standard configuration. This
flame can be used for nearly all standard elements while
the N2O/Acetylene and the Air/LPG (Natural gas) are
available as an option for the more demanding of
elements. All three flame configurations offer coded
burner for full safety protection.
AA500G – The instrument is equipped with a Graphite
Furnace Atomiser only. The graphite head is fixed into the
optical path to maximise performance and eliminate drift.
The temperature of the transversely heated graphite tube
is accurately controlled by means of a precision feedback
system. Pyrolytically coated platform tubes are supplied
as standard to improve the performance and eliminate
many analytical problems associated with this technique.
AA500FG – The instrument is equipped with both Flame
Atomiser and Graphite Atomiser as described above. Both
configurations are installed into the instrument and can
be changed over by a simple selection in the versatile
AA Win software.
9
NEWUPGRADE
AVAILABLE
Full Automation for
multi-elem
ent analysis
NOW GLP COMPLIANT
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 10
F EATURES & FUNCT IONS
l Embedded PC system built into the instrument as
standard on all configurations. Pre-installed AAWin3
software, user manuals, cook book and Windows
operating system.
l AAWin3 software provides full control of the
instrument and autosampler with easy method
change for each technique. Now GLP compliant.
l Automatic 8 Hollow Cathode lamp turret controlled
and optimised by the AAWin3 software. Operating
lamp current and warm-up lamp current can be
individually controlled to eliminate drift commonly
associated with lamp warming.
l D2 lamp background correction system fitted as
standard to all configurations. High energy D2 lamp
and adjustable beam splitter mirror are optimised by
the AAWin3 software.
l Self Reversal background correction system fitted as
standard to all configurations. The high performance
background system uses the same hollow cathode
lamp as installed for the analysis. Minimum extra
components are required and optical alignment is
very simple. Self Reversal can be used for any element
at any wavelength making it extremely versatile.
l High precision minimal optics ensures maximum light
throughput to the computer controlled Czerny-
Turner monochromator.
l A universal autosampler is available as an optional
accessory which is conveniently mounted on the
front of the AA500 instrument.
l Optional XYZ autosampler provides appropriate
sample volume for multi element analysis, along with
scalable sample capacity for high sample throughput.
l Absorption and Emission modes are standard features
in the AAWin3 software as well as peak height, peak
area, sequential and manual integration modes.
F LAME ATOM I S ER F EATURES
The flame atomiser offers three flame options:
Air/acetylene is the standard configuration with the
N2O/acetylene and Air/LPG as options.
Air/Acetylenel The Air/Acetylene flame uses a 100mm single slot
burner for the standard configuration.
l The high sensitivity (Cu 2ppm >0.280abs) is due to
the inert adjustable nebuliser fitted as standard.
An acid resistant replacement is available with a
Pt/Ir capillary.
l The flame can be easily set from blue lean flame
through stoichiometric to fuel rich by means of
computer control.
N20/Acetylenel The N2O/Acetylene flame uses a 50mm single slot
burner and is available as an optional extra.
l This flame configuration is used to measure elements
less prone to ionization such as: Aluminium, Tin,
Titanium, Calcium, Vanadium and Molybdenum.
l Switching from Air/Acetylene to N2O/Acetylene to
Flame Off is fully controlled by the AAWinPro
software.
Air/Propane (LPG)l This flame uses a 3 slot burner and with the low
pressure requirement it is also much safer to operate.
l Due to the low temperature of the flame it is ideal for
analysing alkali metals such as: Potassium, Sodium and
Lithium, especially when used in the emission mode.
l Some remote areas of the world have difficulty
obtaining Acetylene gas of a high enough purity to
operate the flame correctly, LPG can give a real
alternative and offer comparable results for most
elements throughout the wavelength range.
10
AA500continued
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 11
11
Safety Featuresl Pressure monitoring for all gases
l Burner Identification
l Flame sensor
l Drain Trap level Sensor
l Gas Leak Detector
l Over Pressure in Premix
l Safety Cut off Switch
GRAPH I T E ATOM I S ER F EATURES
The integrated Graphite Furnace Atomiser is available
in two instrument configurations.
l In the AA500G instrument the graphite furnace head
is fixed into the light path so alignment with the
optical path is simple and accurate.
l In the AA500FG instrument the graphite furnace
head is fixed behind the flame atomiser assembly and
is motorised into position by a simple operation in
the AAWinPro software. The positions for the flame
and graphite are saved making it easy to swap
between modes for different analysis.
l The temperature of the transversely heated graphite
tube is accurately controlled by means of a precision
feedback system and has been designed to reduce
analytical problems normally associated with this
type of technique.
l Pyrolytically coated graphite tubes are used as standard
and are manufactured to improve performance as well
as increase the analytical life.
l Platform graphite tubes are supplied as standard and
will accept volumes up to 20µl. Non-platform
graphite tubes are also available as an optional extra.
l Up to 10 heat stages are available for the programming
of the graphite atomiser. These can be set and stored
within the AAWinPro software.
l The graphite tube is held in position by means of a gas
piston. Replacement of the graphite tube is performed
by a simple command in the AAWin Software.
l The graphite tube is efficiently cooled by an additional
water circulation system (supplied separately).
Safety Featuresl Argon Gas pressure Sensor
l Water flow sensor
l Over Temperature Sensor
l Broken graphite tube protection
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 12
12
AAWin3Software
The AAWin3 software allows the Analyst to control all
aspects of their analytical method whilst providing an
extensive range of tools for data collection, storage
and interpretation.
The software interface consists of three key work-
areas, whilst having toolbars to access many others.
These work areas allow the user to view real-time
signal acquisition, up-to-date display of calibration
curves and a flexible, sample table.
The AAWin3 software offers full automation allowing
the user to measure multiple elements sequentially by
means of a Sample Wizard and optional random access
autosampler.
AAWin3 Software is a powerful and intuitive software product designed toallow control and data acquisition from the PG Instruments AA500 seriesAtomic Absorption Spectrometer.
Lamp turret setup, operating and warm-up currents,
along with the desired analytical wavelength are
easily selected in the configuration.
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 13
Full Automation for Multi-Element Analysis
13For more details and ordering information please visit our website www.pginstruments.com
Ensure optimal peak position at the chosen analytical
line by scanning the emissions spectra.
Each stage of analysis setup is made quick and simple
by means of the Sample Wizard.
Obtain reliable and accurate results by using the
Energy control feature to manually optimise atomiser
position and setup.
Use the Auto-balance feature to ensure energy level,
and optical alignments are optimised when using
background correction.
Use the sample table to perform quick measurements
of both Standards and Samples.
Easily append the sample table to add new samples or
even revise calibration curves either by manual
introduction or using an Autosampler.
View up-to-date calibration curves in 1st, 2nd, or 3rd
order using a standard calibration or standard
addition.
Perform retrospective curve fits to ensure optimum
correlation.
View real-time signal acquisition for flame, graphite
furnace and hydride generation analysis.
NOWGLPCOM
PLIANT
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 14
14
SpecificationsSpectrometer System
Wavelength range 185nm - 910nm
Light Source Hollow cathode lamp (HCL), Deuterium Arc lamp (D2)
Modulation Square Wave Pulse
Modulation frequency 100hz Self reversal (SR) background, 400hz Deuterium Arc (D2) background
Monochromator
Grating 1800 grooves/mm diffraction grating
Blazed Wavelength 250nm
Focus 300nm
Bandwidth 0.1nm, 0.2nm, 0.4nm, 1.0nm, 2.0nm (Software selectable)
Scan mode Automatic
Photometric Type Single Beam
Wavelength Accuracy +/-0.15nm
Wavelength Resolution 0.2nm +/-0.02nm
Wavelength Reproducibility < 0.05nm
Baseline Stability 0.005Abs/30min
Background Correction Deuterium Arc (D2) 1.0Abs, Self Reversal (SR) 3.0Abs
Flame Analysis
Flame Types Air/Acetylene, Nitrous Oxide/Acetylene, Air/Propane (LPG)
Sensitivity (Cu) 2µg/ml Absorption >0.280Abs
Characteristic Concentration Cu < 0.02µg/ml, Ba < 0.15µg/ml ( N2O/Acetylene)
Detection Limit Cu < 0.004µg/ml
Repeatability Cu < 0.7% (Air/Acetylene Flame)Ba < 1.0% (Nitrous Oxide/Acetylene Flame)
Burner heads Titanium Alloy
Nebuliser Inert Adjustable Nebuliser, Flow Rate from 2ml - 6ml/min. Pt/Ir capillary available for high acid concentration
Pre-mix Chamber Corrosion Resistant
Atomiser Selection Automatic changeover (AA500FG), Manual (AA500F)
Safety Features Burner identification, Flame Sensor, Gas Leak Sensor, Low Gas Pressure Sensor,Drain Trap Sensor, Power Loss Protection
Graphite Furnace Analysis
Graphite Head Transversally Heated
Temperature Range Ambient – 2650C
Heating Program Up to 10 steps. Drying, Ashing, Atomisation, Cleaning
Feedback Voltage and Optical Temperature Control Feedback
Sensitivity (Cu) 50ng/ml Absorption > 0.40Abs
Detection Limit Cd < 0.004ng/ml
Repeatability Cu < 2.0%, Cd < 2.0%
Graphite Tubes Pyrolytically Coated with Platform
Sample Size Up to 20µl
Graphite Cooling Water Circulator Available
Safety Features Argon Pressure Sensor, Water Flow Sensor, Over Temperature Sensor, BrokenTube Protection
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 15
15
Specifications
Standard packaging
Spectrometer System
Data Processing
PC System Embedded PC, Windows Professional Operating System
Operating Program AA Win Pro software
Analytical Methods Flame AA, Flame AE, Graphite Furnace, Hydride Generation
Readout Mode Continuous, Manual, Peak Height, Peak Area
Calibrations Multi-Standard Calibration, Standard Addition, Interpolation
Data Storage Analytical results, instrument and measurement parameters, signal profile, calibration curve
Power Requirements
Main Unit 110/230Vac 50/60hz 200W
Graphite Power Supply 220Vac 50/60hz Instantaneous power 5KW Current rating??
Dimensions
Main unit 110cm x 54cm x 54cm 75Kg
Graphite Power Supply 50cm x 54cm x 54cm 70Kg
AA500F
1 Main unit
1 AA500F instruction Manual
1 Cook Book
1 Safety Manual
1 System validation document
1 Data coded Air/Acetylene burner
1 Inert adjustable Nebuliser
8 Hollow cathode Lamp Holders
1 Monitor
1 Keyboard and Mouse
1 Low Voltage power Supply
2 Power Cables
1 Gas tubing Air
1 Gas Tubing Acetylene
1 Drain Trap and Tubing
1 Tool Kit
1 Various gas Fittings
AA500G
1 Main unit
1 AA500G instruction Manual
1 Cook book
1 Safety Manual
1 System Validation document
8 Hollow cathode Lamp Holders
1 Monitor
1 Keyboard and Mouse
1 Low Voltage Power Supply
2 Power Cables
1 Tool Kit
1 Various Gas Fittings
10 Graphite Tubes
1 Gas Tubing Argon
1 Connection Cable (Graphite)
1 Pipette
1bag Pipette Tips
AA500FG
1 Main unit
1 AA500F instruction Manual
1 Cook Book
1 Safety Manual
1 System validation document
1 Data coded Air/Acetylene burner
1 Inert adjustable Nebuliser
8 Hollow cathode Lamp Holders
1 Monitor
1 Keyboard and Mouse
1 Low Voltage power Supply
2 Power Cables
1 Gas tubing Air
1 Gas Tubing Acetylene
1 Drain Trap and Tubing
1 Tool Kit
1 Various gas Fittings
10 Graphite tubes
1 Gas Tubing Argon
1 Connection Cable (graphite)
1 Pipette
1bag Pipette Tips
AA SPEC BRO – 060_Layout 1 09/10/2013 12:40 Page 16
Accessories
Flame Analysisl 38 sample positions (6 for standard samples, 32 for unknown
samples).
l Sample vial size 6ml
l Standard sample vial size 12ml
l Pressure protection for wash
l Position adjustment by AA Win software
l Reproducibility Cu < 0.6% (air/acetylene) Cu< 1.0% (air/LPG)
Ba <1.0% (nitrous oxide/acetylene)
Graphite Analysisl 76 sample positions (10 for standard samples, 6 for modifier
solutions, 60 for unknown samples)
l Sample vial size 1.5ml
l Modifier vial size 12ml
l Pressure protection for wash
l Up to 3 modifier additions
l Position adjustments by AA-Win Pro software
l Reproducibility Cu <2.0%, Cd <2.0%
Autosampler (500-004)The AS500 Autosampler is an accessory specifically designed for the AA500 Atomic Absorption Spectrometer. The AS500 can be used to automatically introduce standard and unknown samples to the atomiser for analysis. It can be used with both flame and graphite furnace instruments.
Nitrous Oxide/Acetylene Burner (500-023)
l Titanium alloy
construction
l 50mm single slot
l Coded for safety
l Direct replacement for
standard burner
l Designed for higher
temperature flame
l Adjustment of burner
rotation for sensitivity
optimisation
Air/Propane (LPG) Burner (500-030) XYZ Autosampler
l Titanium Alloy
construction
l 3 slot burner
l Coded for safety
l Direct replacement for
standard burner
l Designed for lower
temperature flame
l Adjustment of burner
rotation for sensitivity
optimisation
Inert Adjustable Nebuliser (500-011)
l Excellent RSD’s typically<1.0%
l Typical read back valuesbetter than 2%
l Enhanced sensitivity
l Can be used on all PGInstruments AA500 andAA990 series atomicabsorption systems
l Redesigned spray chamberto ensure excellentdrainage properties
For more details and ordering information please visit our website www.pginstruments.com16
l High Sample Volume for
multi-element analysis
l Scalable Sample capacity
for high sample
throughput
l Acid resistant nebuliserwith Pt/Ir capillary(500-012)
l Adjustable uptake from 2-6ml/min
l Fully adjustable impactbead
Patent pending
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 17
Hydride Unit (500-007)
l Universal design
l Full reaction cycle from
press of start button
l Supplied with
temperature controlled
heated absorption cell
l Cell is easily installed with
burner attachment
l Can be used for Mercury
cold vapour analysis
l Sensitivity for most
elements <1ng/ml
l Only 2.0ml – 2.5ml per
sample required
l RSD < 3%
l Quick analysis < 30
seconds per measurement
Air Compressor (GS-003)
Fume Extraction System
l Ultra low noise whenrunning – just 53-59dB (A)
l 100% Oil Free Design
l Efficient, Horizontal Twinpump unit
l Innovative CompactDesign, perfect for limitedspace
l 10 bar Maximum pressure
l 15 Litre tank for stable airsupply
l 104L/min supply at 3 bar
l Easy to ship only 14Kg
Water Circulator (GS-005)
For use with AA500G /
AA500FG
l Re-circulating water
cooler
l 2Kw cooling capacity
l 230v, 270 watts power
(110v available, please
call for details.)
l Rust free 5 Litre Water
Capacity
l 2 Litre per minute
flow rate
l Forced air cooling system
l 2x Male quick connection
points 2x Female quick
fittings supplied
l Dimensions (LWH)
250 x 245 x 700 mm
l Tubing and additional
connectors optional
Gas Regulators (GS-001, GS-002, GS-006, GS-007)
l Direct fitting to
gas cylinder
l Comply to British
Standards
l Supplied with hose
fitting and crimp
l Various gas cylinder
fittings available
l Regulators for Air,
Acetylene, Argon, LPG
l Specify fitment to
European standard
Acetylene Flame Arrestor (GS-004)
l Direct fitting to gas
regulator output
l Required for acetylene gas
l Non-resetting easy
to replace
For more details and ordering information please visit our website www.pginstruments.com 17
l Universal fitting
l Stainless steelconstruction
l All fittings supplied
l 150mm polymer fan
l 240Vac 50/60hz
l 300mm diameter hood,adjustable height
l Adjustable bends from 0° to 90°
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 18
18
ATOM IC A B SOR P T I ON S P EC T ROMET E R
AA990F
The PG990 Atomic Absorption Spectrophotometer is
a versatile entry level instrument with a computer
controlled Air/Acetylene flame for general laboratory
requirements. Due to its high sensitivity and excellent
performance the instrument can be used for a wide
range of applications including:
l Agricultural
l Clinical
l Environmental
l Food
l Metal
l Mining
l Geological
l Petrochemical
l Pharmaceutical
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 19
F EATURES & FUNCT IONS
l The instrument has a motorised 8 hollow cathode
lamp turret which allows the automatic positioning
and optimisation of each hollow cathode lamp by
the software.
l The control of the gas flows for the fuel gas (C2H2) of
the burner is also carried out directly from the software
thus allowing optimisation of the instrument for the
best analytical parameters for a selected analysis.
l Two methods of background correction are available.
The first utilises a Deuterium Arc lamp and the second
is the proven method of Self Reversal.
l High precision minimal optics ensures maximum light
throughput to the computer controlled Czerny-Turner
Monochromator.
l The location of the wavelength and peak selection is
automatically controlled from the software.
l The spectral bandwidth is automated and is available
with a choice of five slit sizes.
l The electronic parameters for the photomultiplier tube
detector, the hollow cathode lamp current and the
balancing of the absorbance and background energies
are controlled from the software.
l The ignition of the flame is computer controlled
and the various safety interlocks offer a very safe
operating system.
SAFETY
l The flame conditions are continuously monitored and
should the flow rates change, an audible alarm sounds.
l The pressure of the support gas (oxidant) is monitored
constantly. If the pressure changes then the flow of
the fuel gas will be stopped and the flame will be
safely extinguished.
l A sensor monitors the level of liquid in the drain
and will prevent ignition if too low. The flame will also
be extinguished if the level of liquid in the drain
changes significantly.
l A flame sensor monitors the flame and safely turns off the
gas flow to the burner if the flame suddenly extinguishes.
l The burner is identified by a switch making it
impossible to light without the burner being fitted.
l An emergency flame off button is installed in case a
problem is observed. The flame can be extinguished
safely.
19
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 20
20
AAWin2Software
The AAWin2 software allows the Analyst to control all
aspects of their analytical method whilst providing an
extensive range of tools for data collection, storage
and interpretation.
The software interface consists of three key work-areas,
whilst having toolbars to access many others. These
work areas allow the user to view real-time signal
acquisition, up-to-date display of calibration curves and
a flexible, sample table.
AAWin2 Software is a powerful and intuitive software product designed toallow control and data acquisition from the PG Instruments PG990 AtomicAbsorption Spectrometer.
Lamp turret setup, operating and warm-up currents,
along with the desired analytical wavelength are easily
selected in the configuration.
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 21
21For more details and ordering information please visit our website www.pginstruments.com
Ensure optimal peak position at the chosen analytical
line by scanning the emissions spectra.
Each stage of analysis setup is made quick and simple
by means of the Sample Wizard.
Obtain reliable and accurate results by using the
Energy control feature to manually optimise atomiser
position and setup.
Use the Auto-balance feature to ensure energy level.
Use the sample table to perform quick measurements
of both Standards and Samples.
Easily append the sample table to add new samples
or even revise calibration curves either by manual
introduction or using an Autosampler.
View up-to-date calibration curves in 1st, 2nd, or 3rd
order using a standard calibration or standard addition.
Perform retrospective curve fits to ensure optimum
correlation.
View real-time signal acquisition for flame and
hydride generation analysis.
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 22
22
SpecificationsSpectrometer system
Wavelength range 190nm - 900nm
Light Source Hollow cathode lamp (HCL), Deuterium Arc lamp (D2)
Modulation Square Wave Pulse
Modulation frequency 100hz Self reversal (SR) background, 400hz Deuterium Arc (D2) background
Monochromator
Grating 1800 grooves/mm diffraction grating
Blazed Wavelength 250nm
Focus 300nm
Bandwidth 0.1nm, 0.2nm, 0.4nm, 1.0nm, 2.0nm (Software selectable)
Scan mode Automatic
Photometric Type Single Beam
Wavelength Accuracy +/-0.15nm
Wavelength Resolution 0.2nm +/-0.02nm
Wavelength Reproducibility < 0.05nm
Baseline Stability 0.005Abs/30min
Background Correction Deuterium Arc (D2) 1.0Abs, Self Reversal (SR) 3.0Abs
Flame Analysis
Flame Type Air/Acetylene
Sensitivity (Cu) 2ug/ml Absorption >0.28Abs
Characteristic Concentration Cu < 0.02ug/ml
Detection Limit Cu < 0.004ug/ml
Repeatability Cu < 0.7%
Burner head Titanium Alloy
Nebuliser Inert Adjustable Nebuliser, Flow Rate from 2ml - 6ml/min. Pt/Ir capillary available for high acid concentration
Pre-mix Chamber Corrosion Resistant
Safety Functions Burner switch, Flame Sensor, Gas Leak Sensor, Low Gas Pressure Sensor, Drain Trap Sensor, Power Loss Protection
Data Processing
Operating Program AA Win software
Analytical Methods Flame AA, Flame AE, Hydride
Readout Mode Continuous, Manual, Peak Height, Peak Area
Calibrations Multi-Standard Calibration, Standard Addition, Interpolation
Data Storage Analytical results, Instrument and measurement parameters, Signal profile,Calibration curve
Power Requirements 220Vac 50/60hz 200W
Dimensions 110cm x 54cm x 54cm 75Kg
Standard packaging1 Main Unit1 PG990 Instruction Manual1 Cook Book1 Safety Manual1 System Validation Document
1 Air/Acetylene Burner1 Inert Adjustable Nebuliser8 Hollow Cathode Lamp Holders1 Power Cable1 Gas Tubing Air
1 Gas Tubing Acetylene1 Drain Trap and Tubing1 Tool Kit1 Various Gas Fittings
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 23
l Ultra low noise whenrunning – just 53-59dB (A)
l 100% Oil Free Design
l Efficient, Horizontal Twinpump unit
l Innovative CompactDesign, perfect forlimited space
l 10 bar Maximum pressure
l 15 Litre tank for stable airsupply
l 104L/min supply at 3 bar
l Easy to ship only 14Kgl Universal design
l Full reaction cycle frompress of start button
l Supplied withtemperature controlledheated absorption cell
l Cell is easily installed withburner attachment
l Can be used for Mercurycold vapour analysis
l Sensitivity for mostelements <1ng/ml
l Only 2.0ml – 2.5ml persample required
l RSD < 3%
l Quick analysis < 30seconds per measurement
l Sequential auto-samplerallows the automatedanalysis of 50 or more
samples and calibrationstandards.
l The system allows forautomatic update ofstandard values and curveparameters by using upto 8 standards, blanksand QC standards.
l A double wash station for probe wash avoidssample and standardcontamination.
l An inert Teflon probe issupplied as standard.
23
AccessoriesInert Adjustable Nebuliser (500-011) Flame Autosampler
For more details and ordering information please visit our website www.pginstruments.com
Hydride Unit (990-007)
Air Compressor (GS-003)
Acetylene Regulator (GS-001)
l Direct fitting to gas cylinder
l Comply to British Standards
l Supplied with hose fittingand crimp
l Various gas cylinderfittings available
l Specify fitment toEuropean standard
l Universal fitting
l Stainless steel construction
l All fittings supplied
l 150mm polymer fan
l 240Vac 50/60hz
l 300mm diameter hood,adjustable height
l Adjustable bends from 0° to 90°
Fume Extraction System
Acetylene Flame Arrestor (GS-004)
l Direct fitting to gasregulator output
l Required for acetylene gas
l Non-resetting easy to replace
l Excellent RSD’s typically<1.0%
l Typical read back valuesbetter than 2%
l Enhanced sensitivity
l Can be used on all PGInstruments AA500 andAA990 series atomicabsorption systems
l Redesigned spray chamberto ensure excellentdrainage properties
l Acid resistant nebuliserwith Pt/Ir capillary (500-012)
l Adjustable uptake from 2-6ml/min
l Fully adjustable impactbead
Patent pending
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 24
Atomic Absorption SpectrometerQualification Kit
24
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 25
25
The purpose of the Qualification Kit is to offer
both laboratory analysts and regional distributors
the flexibility to perform a full installation and
operational qualification on Atomic Absorption
Spectrometers manufactured by PG Instruments.
The work instructions within the qualification work book
have been developed with considerations of the
requirements of the Pharmacopoeia to ensure compliance
with good laboratory practice.
The documentation supplied with the kit, guides the user
through the qualification as it offers detailed instructions
on how to carry out the tests, document and record
results and perform any necessary corrective action. The
comprehensive kit offers all the materials required to
complete the qualification.
The contents of the kit are as follows:
l Qualified Hg Hollow Cathode Lamp – For determining
the wavelength accuracy.
l Qualified Cu Hollow Cathode Lamp – For determining
the flame and graphite analytical tests.
l Qualified Standard Solution – For determining the
flame analytical tests.
l Qualified Standard Solution – For determining the
graphite analytical tests.
l Filter Tool – For determining the photometric
accuracy.
l Certificates for Solutions.
l Qualification Workbook.
l USB Flash Drive containing qualification workbook
and qualification worksheets.
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 26
26
Atomic FluorescSpectroscopy
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 27
The AFS technique is widely used for the
determination of hydride forming elements
and Mercury analysis. The technique provides
an alternative to the conventional method of
connecting a hydride generator to an Atomic
Absorption Spectrophotometer (AAS). Due to
the high intensity Hollow Cathode lamp source
and low baseline noise offered by this
technique, the usable sensitivity is greatly
improved offering very low detection limits.
The AFS instrument fulfils the analysis by means of
simultaneous detection, accurate sample introduction
and stable furnace temperature control complementing
the more widely used AAS flame/graphite systems.
The Atomic Fluorescence Spectroscopy technique is used
to measure samples containing both hydride forming
elements and mercury in a variety of analytical sectors.
Due to its design and functionality, the instrument offers
enhanced sensitivity, reduced interference and excellent
limits of detection (ppt).
27
cence
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 28
28
S IMU LTANEOUS ATOM IC F L UOR E SC ENCE S P EC T ROMET E R
AF402
AA SPEC BRO – 060_Layout 1 09/10/2013 12:41 Page 29
The AF402 is an ideal instrument for elementalanalysis in various market and researchindustries including:
l Agricultural l Food and Beverage
l Public Health l Clinical
l Pharmaceutical l Geological
l Metallurgical l Petrochemical
STANDARD F EATURES
Light Source
l High Intensity Hollow Cathode lamps for greater
intensity and stability.
l Up to three lamps can be installed for simultaneous
analysis.
l All lamps are uniquely data coded offering important
information to the PF Win operating software.
Optical System
l Shielded optical design greatly reduces light interference.
l Enhanced signal to noise ratio for increased analytical
sensitivity.
l Unique optical configuration for increased fluorescence
intensity. Up to twice the fluorescence intensity found
in traditional AFS systems.
l High Quantum Solar Blind Photomultiplier Tube fitted
as standard to ensure optimum stability.
Atomiser System
l High precision quartz atomiser tube designed for
optimum performance, durability and long life.
l Self-igniting Argon/ Hydrogen flame with software
controlled heater coil.
l Adjustable height control for improved optimisation.
l Integrated fume exhaust system to decontaminate
pollution.
29
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 30
Hydride generator
l Built in Hydride generator with continuous flow.
l Multi-pump with individual pressure adjustment for
uptake control of reagents and carrier liquids.
l Introduction of sample using motorised syringe drive
and unique three way control valve. Software
controlled for greater accuracy.
l High efficiency glass liquid separator which optimises
the separation minimising pressure fluctuations and
improving measurement precision.
l Reagent compartment stores reagents safely and
cleanly. Storage for up to four containers.
l Low level reagent monitoring.
l Embedded PC system built into the instrument as
standard. Pre-installed PFWin software, user manual
and windows operational system.
Autosampler
l Integrated autosampler.
l XYZ configuration.
l 2 x 45 vial trays for automated sample introduction.
l Auto dilution using motorised syringe drive and
selection valve.
l Full calibration using only one stock standard solution.
l Fully controlled by the PF Win software.
Electronic Control System
l High technology electronics and PCB components.
l Rear mounted and separated electronic section to
eliminate carry over from hydride technique.
l Built in embedded PC operating system. No extra
PC control required.
l Windows operating software.
l PFWin instrument control software.
Detection Limits
Element Detection Limit(ug/l) RSD %
Arsenic (As) <0.01ug/l <1%
Bismuth (Bi) <0.01ug/l <1%
Cadmium (Cd) <0.001ug/l <1%
Germanium (Ge) <0.05ug/l <1%
Mercury Hg) <0.001ug/l <1%
Lead (Pb) <0.01ug/l <1%
Antimony (Sb) <0.01ug/l <1%
Selenium (Se) <0.01ug/l <1%
Tin (Sn) <0.01ug/l <1%
Tellurium (Te) <0.01ug/l <1%
Zinc (Zn) <1.0ug/l <1%
30
AF402continued
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 31
PFWinSoftwareThe intuitive software package provides ease of
operation, method development control and data
acquisition. Software wizards simplify the operation
and configuration.
AFWin offers real time display of fluorescence intensity.
Full adjustment of current for all hollow cathode lamps
is provided. Lamps are set for optimal performance.
Where parameters are input manually, PFWin software
will prompt the suggested parameters to complete the
configuration for the analysis.
Detailed information and results are shown for each
element analysed.
Calibration information as well as results for standards
are shown for each element.
Detailed results for the analysis can be seen and saved.
Reporting and QC control allows printouts of spectra,
standard calibration curves, analysis and signal data.
Report of operating parameters can be saved or printed
for future information.
31
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 32
32
SpecificationsSample Atomisation
Atomiser Quartz tube furnace with auto ignition of argon/hydrogen flame
Furnace heating Computer controlled heating
Hydride Generator Continuous flow high performance for cold-vapour mercury and hydride determinations of As, Se, Te, Bi, Sb, Sn, Zn, Pb, Cd, Ge
Gas Requirement High purity Argon gas(99.99%) 30psi pressure
Sample and delivery
Peristaltic Pump 3 channel pump with 3 adjustable pressure control clamps and programmable
speed control.
Exhaust System Built in with filter to decontaminate pollutants
Gas/Liquid Separator High efficiency two-stage gas-liquid separator.
Autosampler
Type Integrated XYZ
Sample Introduction High sample capacity, automated introduction of sample and standards bymotorised syringe drive.
Sample capacity 2 x 45 standard test tube sample racks
Optics
Optical Design Short focal length, non-dispersive integrated optical design.
Light source 3 channel simultaneous element analysis using computer controlled. Modulated and pulsed, hollow cathode light source.
Baseline stability <2%
Linear range >103
Computer system
Type Built in embedded PC system
Operating system Windows
Operating software PFWin software
Safety Alarm for low gas <60ml/min. Gas cut off in the event of power supply loss
Power requirements 220Vac ±10% 50/60hz, 300W
Dimensions 100.6cm x 58cm x 55cm
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 33
Argon Gas Regulator
l Direct fitting to gas cylinder
l Comply to British Standards
l Supplied with hose fitting and crimp
l Various gas cylinder fittings available
l Regulator for high purity Argon 99.99%
l Specify fitment to European standard
Accessories
33
Fume Extraction System
l Universal fitting
l Stainless steel construction
l All fittings supplied
l 150mm polymer fan
l 240Vac 50/60hz
l 300mm diameter hood, adjustable height
l Adjustable bends from 0° to 90°
For more details and ordering information please visit our website www.pginstruments.com
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 34
F LAME PHOTOMET E R
FP902
34
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 35
35
Flame photometry provides a simple method
of analysis and concentration determination of
elements which cannot easily be measured
using any other technique. PG Instruments
application notes for users of our versatile
new flame photometers cover a wide range
of industrial sectors, from food and drink to
cement, minerals and glass, as well as
agricultural, biological and clinical applications.
The PG Instruments FP902 Flame Photometer offers
the combination of clinical, industrial and education
applications in ONE instrument. It gives accurate, fast and
precise analysis of sodium, potassium, calcium and lithium
in one simple aspiration.
The FP902 uses a soft touch key pad for ease of data entry
and features electronic ignition, automatic flame failure
safety and will operate using butane, natural gas or LPG.
NOWWITH
BARIUM
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 36
36
TYP I CAL APP L I CAT IONS
l Soil and Agricultural Analysis
l Biological Fluids
l Petrochemical Industry
l Food and Drink Analysis
F EATURES & FUNCT IONS
l Five element measurement in a single aspiration
(Na, K, Li, Ca and Ba).
l Curve calibration using up to 5 standards.
l Calibration curve fit routines first order and
second order.
l Multiple calibration curve saving facility.
l Sample data storage.
l Printer attachment.
l Data displayed in concentration units ie. ppm,
meq etc.
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 37
Specifications
37
ModeGeneral Analytical
Serum UrineApplications
Na: 0.1-100 ppm Na: 0-200 Na: 0-250Up to 250 meq/l, 1:100 dil 1:100 dil 1:100 dil
K: 0.1-100 ppm K: 0-10 K: 0-100
RangeUp to 250 meq/l, 1:100 dil 1:100 dil 1:100 dil
Ca: 15-100 ppm Ca: - Li: 0-2 Ca: - Li: 0-10Up to 250 meq/l, 1:100 dil 1:5 dil 1:2 dil
Li: 0.5-100 ppmUp to 250 meq/l, 1:100 dil
Sensitivity General Mode: Na:0.5 ppm, K:0.5 ppm, Li:0.5 ppm, Ca:15 ppm, Ba:30 ppm
Filters Filters for Sodium, Potassium, Lithium, Calcium and Barium are included
Resolution 0.1 ppm/meq
Reproducibility + 2% FS, & + 2 Digits
Curve Fit Accuracy + 2% FS
Display 20 x 4 Alphanumeric backlit LCD display
Flame System LPG & dry oil free air
Detector Silicon Photodiode
Calibration Up to 5-Point Calibration with curve fitting software
Gas Control Fully Adjustable
Ignition System Auto Ignition
Atomiser Axial flow type
RS-232C Interface RS-232 PC Software available
Power 230 V + 10% AC, 50/60 Hz
Dimensions 365 x 245 x 220 mm (L x B x H) (Approx.)
Weight Main Unit: 7.5Kg. (Approx.)
Compressor UnitAir Supply Oil free compressor unit with pressure regulator
Combustion Gas LPG Controlled by precision regulator
Power 230 V + 10% AC, 50/60 Hz
Dimensions 290 x 255 x 210 mm (L x B x H) (Approx)
Weight 8Kg. (Approx.)
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 38
l 2 inch lamps available
l 70 single-element lamps
l More than 120 multielement lamps, up to 5 element per lamp
l Low current operation
l Long life >5000mA Hours
l Designed for use with SRbackground system
Consumables
Hollow Cathode Lamps Multi Element Standardsl 500ml volume bottles
l More than 40 individualelements available
l 1.000g/ml (1000ppm)concentration
l Individual elements madeup in a choice of acidsolutions
l Glass construction
l Fixed flow rate for easyinstallation.
l Acid resistant to 5%
l Easily replaced
Glass Fixed Flow Nebuliser Standard Nebuliser (Acid Resistant available)
38
AA500
l Pyrolytically coatedgraphite tube
l Supplied with or withoutplatform
l High quality long life tubes
l Can be used for largersamples >20ul
Graphite Tubes Sample Uptake Tubesl Replacement sample
uptake tubes
l Attached filter to eliminatesolid uptake
l Standard length
l Easily replaced
l High energy lamp
l For use with AA500instrument
l Easy fit
l No alignment required
Deuterium Lamp Auto Sampler Vials – Flame Samples
l Re-useable glass vials
l For use with AS500 AutoSampler
l Larger volume for samples6ml
l 11/2 inch standard lampfitting
l Excellent RSD’s typically<1.0%
l Typical read back valuesbetter than 2%
l Enhanced sensitivity
l Can be used on all PGInstruments AA500 andAA990 series atomicabsorption systems
l Redesigned spray chamberto ensure excellentdrainage properties
l Fully insert nebuliser
l Adjustable uptake from 2-6ml/min
l Fully adjustable impactbead
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 39
39
l Re-useable glass vials
l For use with AS500 AutoSampler
l Larger volume forstandards 12ml
Auto Sampler Vials – Flame Standards Auto Sampler Vials – Graphite Samples/Standards
l Re-usable vials
l For use with AS500autosampler
l 1.5ml volume for samplesand standards
l Re-useable glass vials
l For use with AS500 AutoSampler
l Larger volume for modifiersolution 12ml
Consumables
Auto Sample Vials – Graphite Modifier Solutions Micro Pipettel Used for manual injection
l Adjustable volume
l Accurate dispensingvolume
l For use with AA500Graphite Furnace
l Disposable tips
l For use with micro pipette
Micro Pipette Tips
AA500
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 40
Consumables
40
l Re-useable glass vials
l For use with AS990 AutoSampler
l Larger volume for samples
Auto Sampler Vials – Flame Samples/Standards Auto Sampler Vials – Wash
l For use with AS990 AutoSampler
l Large volume for wash
l Re-usable
Sample Uptake Tubesl Replacement sample
uptake tubes
l Attached filter to eliminatesolid uptake
l Standard length
l Easily replaced
AA990
l 2 inch lamps available
l 70 single-element lamps
l More than 120 multielement lamps, up to 5 element per lamp
l Low current operation
l Long life >5000mA Hours
l Designed for use with SRbackground system
Multi Element Standardsl 500ml volume bottles
l More than 40 individualelements available
l 1000g/ml (1000ppm)concentration
l Individual elements madeup in a choice of acidsolutionsl 11/2 inch standard lamp
fitting
Hollow Cathode Lamps
l Glass construction
l Fixed flow rate for easyinstallation.
l Acid resistant to 5%
l Easily replaced
Glass Fixed Flow Nebuliser Acid Resistant Nebuliser
l High energy lamp
l For use with AA500instrument.
l Easy fit
l No alignment required.
Deuterium Lamp
l Excellent RSD’s typically<1.0%
l Typical read back valuesbetter than 2%
l Enhanced sensitivity
l Can be used on all PGInstruments AA500 andAA990 series atomicabsorption systems
l Redesigned spray chamberto ensure excellentdrainage properties
l Fully insert nebuliser
l Adjustable uptake from 2-6ml/min
l Fully adjustable impactbead
AA SPEC BRO – 060_Layout 1 09/10/2013 12:42 Page 41
41
Consumables
l Re-usable glass cups
l Used in XYZ autosamplerfor standard solutions
Sample Cups Small Standard Solutions
l 500ml volume bottles
l Individual elementsavailable for Sb, As, Bi, Cd,Ge, Pb, Hg, Se, Te, Sn, Zn
l 1.000g/ml (1000ppm)concentration
Heater Coilsl Replacement wire coils
l Fits around the QuartzAtomiser
l Used for heating theatomiser when used forhydride analysis
l Regular replacementsuggested
AF402
l High performance – Highoutput lamps
l Lamps designed for usewith AF402 system
l Built in circuit for codingand usage recording
l Available for analysis of avariety of elementsincluding: Sb, As, Bi, Cd,Ge, Pb, Hg, Se, Te, Sn, Zn
Quartz Atomiserl Quartz Glass – to
withstand heat andchemicals
l Replacement required atregular intervals
l Inputs for gas and hydridevapout
Hollow Cathode Lamps
l Quartz Glass – towithstand high acidsolutions
l Designed to fit in the XYZautosampler
Injection Tubes Glass Mixing Chamberl Quartz glass
l Built in gas separator
l Easily replaced
l Accurate high quality
l Used for accurate uptakeof sample and carrier liquid
l Requires regular flushing
Syringe
AA SPEC BRO – 060_Layout 1 09/10/2013 12:43 Page 42
42
l Matching air compressordesigned only for use withFP902 instrument
l Oil free air
l Adjustable air supply withgauge
l 220v 50/60hz
Air Compressor Nebuliser
l Cross flow design
l Adjustable uptake rate
l High quality capillary tube
l Fits into premix chamber
l Designed for use with lowpressure propane flame
l Easily removed for cleaning
Consumables
Burner Capillary Tubesl Fit onto capillary tube of
nebuliser
l Optimised length
l Used for sample uptake
l Easily fitted and replaced
Sample Cupsl Re-usable glass cups
l Used for holding samplesand standard solutions
l Small volume size for usewith FP902 instrument
l Fine wire length
l Used to clean blockage incapillary tube of nebuliser
Cleaning Wires
FP902
AA SPEC BRO – 060_Layout 1 09/10/2013 12:43 Page 43
We reserve the right to modify, revise/upgrade, suspend or discontinue any Product in whole or in part, either temporarily or permanently, with or without notice.