ELITE SERIES SURGICAL TABLES

INCLUDING BATTERY MODELS

MODEL ELITE 6500

OPERATORS MANUAL

Page 2: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc
Page 3: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

19

MODEL ELITE 6500N

SKYTRON STANDARD LIMITED WARRANTYSKYTRON, a Division of the KMW Group, Inc. (SKYTRON) warrants all new products sold by it directly or through a dealer or other authorizedrepresentative, with exception to replacement parts, spares, bulbs (surgical lights), pads, and accessory items (surgical tables) to be free fromdefects in material or workmanship, under normal use and service, for a period of two (2) years. This warranty shall include the cost of repairor replacement of defective parts including the cost of service labor and travel time to the site of equipment use. Delays caused by the userin accessing the equipment for repair will be chargeable at the normal hourly rate for service by SKYTRON’s authorized service representative.The warranty period shall begin with the initial operation or one (1) year after receipt of the product, whichever shall occur first.

Replacement parts, spares, bulbs (surgical lights), pads and accessory items (surgical tables) are warranted to be free from defects in materialor workmanship, under normal use and service, for a period of ninety (90) days from receipt by the ultimate user, with exception to replacementparts supplied by SKYTRON, for products under warranty, which shall be covered for any remaining period of the original product warranty,or for 90 days, whichever is of greater benefit to the ultimate user.

SKYTRON’s responsibility and liability shall be limited to the repair or replacement of any part which we, SKYTRON, determine to be defectivewithin the applicable warranty period. Minor adjustments required as a result of normal wear during the use of the product within the warrantyperiod are not covered under warranty. The labor portion of this warranty is covered by SKYTRON’s Authorized Service Agent. Repairs madeby others are not authorized nor covered by SKYTRON with respect to labor costs.

SKYTRON shall not be liable for any other expense, loss or damage, whether direct, incidental, consequential or exemplary arising in connectionwith the sale or use of or the inability to use SKYTRON products.

NO EXPRESS WARRANTY IS GIVEN BY SKYTRON WITH RESPECT TO ITS PRODUCTS EXCEPT AS SPECIFICALLY SET FORTHHEREIN. ANY WARRANTY IMPLIED BY LAW, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULARPURPOSE, IS EXPRESSLY LIMITED TO THE TWO-YEAR AND 90-DAY TERMS SET FORTH ABOVE. THE FOREGOING STATEMENTSOF WARRANTY ARE EXCLUSIVE AND IN LIEU OF ALL OTHER REMEDIES.

No dealer, agent, employee or other representative of SKYTRON is authorized to extend or enlarge this warranty.

REV. 02/02

Although current at the time of publication, SKYTRON’s policy of continuous development makes thismanual subject to change without notice.

Page 4: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc
Page 5: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

21

TABLE OF CONTENTS

Title Page

SPECIFICATIONS ................................................................................................................................... 1

SPECIAL USER ATTENTION .................................................................................................................. 2

SECTION I INTRODUCTION ................................................................................................................... 51-1. General ......................................................................................................................................... 51-2. Power Requirements .................................................................................................................... 51-3. Pendant Control Unit .................................................................................................................... 61-4. Floor Lock/Brake System ............................................................................................................. 6

SECTION II BATTERY TABLE OPERATION .......................................................................................... 72-1. AC 120V Operation ...................................................................................................................... 72-2. Battery Operation ......................................................................................................................... 72-3. Automatic Shut-Off ....................................................................................................................... 82-4. Charging the Battery .................................................................................................................... 82-5. Emergency Back-Up Controls ...................................................................................................... 8

SECTION III OPERATION ....................................................................................................................... 93-1. Electrical Power ............................................................................................................................ 93-2. Positioning Functions ................................................................................................................... 9

a. Floor Lock/Brake system ........................................................................................................ 9b. Emergency Brake Release ..................................................................................................... 9c. Trendelenburg ...................................................................................................................... 10d. Lateral Tilt ............................................................................................................................. 10e. Back Section ........................................................................................................................ 10f. Elevation ............................................................................................................................... 11g. Leg Section .......................................................................................................................... 11h. Flex Positioning .................................................................................................................... 11i. Kidney Lift ............................................................................................................................. 11j. Return To Level .................................................................................................................... 12k. Pendant Control Storage ...................................................................................................... 12

3-3. Head Section .............................................................................................................................. 123-4. Leg Section Removal. ................................................................................................................ 123-5. 180 Degree Table Top Rotation. ................................................................................................ 133-6. Specialty Positioning .................................................................................................................. 13

SECTION IV MAINTENANCE ................................................................................................................ 164-1. Preventive Maintenance ............................................................................................................. 164-2. Cleaning Recommendations ...................................................................................................... 164-3. Service ....................................................................................................................................... 17

Page 6: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc
Page 7: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

1

®

Electrical SpecificationsPower requirementsCurrent LeakagePower Cord

120 VAC, 60Hz, 450 WattsLess than 100 micro amps

15 feet w/hospital grade connector(removeable on battery model)

10" 19-1/2" 21-1/2" 24"

15"

20"

3"

5"76"

TOP VIEW

8-1/4"

35-3/4"

5-3/4"

19"

21-3/4"

45" MAX26" MIN

SIDE VIEW END VIEW

90˚

60˚

14-1/2"

6500 Series General Purpose Surgical Table Specifications

Page 8: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

2

WARNINGDO NOT unlock brakes when a patientis on the table. An uneven patientweight load may cause instability.

NOTEIn case of a power failure or an electri-cal problem within the table, the emer-gency brake release system can beused to move the table. The controllever for this function is located on theside of the table base and is identifiedby an EMERGENCY BRAKE RELEASElabel. Turn the lever counterclockwiseto release the brakes.

WARNING•The Emergency Brake Release Valvemust be closed and tightened (clock-wise) before activating any hydraulicfunction.

•If the Emergency Brake Release Valvehas been operated, the UNLOCK but-ton on the pendant control will have tobe pressed before brakes will lock again.

NOTE•If brakes fail to set, make sure theEmergency Brake Release Lever is tight(clockwise).

•If the Emergency Brake Release Valvehas been operated, the UNLOCK but-ton on the pendant control will have tobe pressed before brakes will lock again.

NOTE•Turning the Main Power Switch ON willchange the table operation to 120 VACpower regardless of the position of theBATT button.

•If Battery operation is activated whileMain Power Switch is ON there will beapproximately a 4 second delay beforeBattery operation activates.

NOTEThe table will operate correctly on bat-tery power with the power cord con-nected to a wall outlet or disconnected.

The extreme positioning capabilities of the 6500Series Table requires special attention for possibleinterference points when using multiple functionpositioning. As with the operation of any surgicaltable, a certain amount of care should be exercisedto position the patient safely. Although the thickpads and sheets substantially protect the patient,pinch points, located at the joints of the top sectionshould always be considered. BE SURE THATTHE ARMS, HANDS AND FINGERS OF THEPATIENT AND THOSE OF THE OPERATINGROOM PERSONNEL ARE CLEAR OF ALL MOV-ING PARTS BEFORE MOVING THE TABLE.Proper restraints should always be used for patientsafety.

Certain accessories such as the Uro-Drain Tray,Armboards and X-Ray top can be damaged whenchanging the position of the table top sections.Always look first to see if a desired movement isgoing to interfere with any accessories in use.

The operator has the ultimate responsibility ofpreventing damage to the table and surroundingequipment or possible injury to the patient or staff.In general, common sense will dictate when thereis a potential hazard.

The following precautions should be reviewedby all personnel prior to operating the table.

WARNINGIndicates a possibility of personal injury.

CAUTIONIndicates a possibility of damage toequipment.

NOTEIndicates important facts or helpful hints.

WARNINGRisk of electrical shock. Make surepower cord is disconnected prior toaccessing fuses.

NOTEActivating any function button will acti-vate the brake system. Using the TABLEUP function to set the brakes providesa visual assurance that the brakes arelocked. As the brake cylinders areextending, the entire table will moveslightly. When the table top begins toelevate, the brakes are fully locked.

SPECIAL USER ATTENTION

Page 9: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

3

NOTEWith an evenly distributed patient weightload, all table positioning functions willoperate smoothly and quietly with apatient weight of up to 500 pounds.(400 pounds SN-1991-5&P).

WARNINGTo maximize patient safety, utilizeproper restraint methods during extremetrendelenburg positioning.

WARNINGTo maximize patient safety, utilizeproper restraint methods during extremelateral tilt positioning.

CAUTIONTo prevent damage to the kidney lift, asafety interlock prevents the back sec-tion from going above horizontal if thekidney lift is not all the way down.

WARNINGThe leg section may hit the table baseor the floor if both the leg and elevationsystems are placed in their full downposition.

NOTEWhen REFLEX button is activated, ifKidney Bridge is up, the back sectionwill not go above horizontal.

NOTETo prevent damage to the kidney lift, asafety interlock prevents the kidney liftfrom going up if the back section israised above horizontal.

NOTEElevation, kidney lift, and brake systemfunctions are not affected by the returnto level function.

WARNINGIf the patient position is reversed on thetable (patient's back on the leg section),the patient weight limit is 250 lbs.

WARNINGEnsure that the leg section is properlyengaged and secured to pins beforeuse to prevent injury.

NOTEBattery Operation must be turned OFFat the pendant control. It can not beturned Off using the main power switch.

NOTEWhen the amber light starts to blink(indicating low power in battery) thetable will operate for approximately 5continuous minutes, typically longenough to use the table for the rest ofthe day.

WARNINGThe charging system operates ONLYwhen the table is in AC120V operationmode.

NOTEThe table can be operated on 120VACpower while the battery is being re-charged. The green AC 120V indicatorlight (on the pendant control) will illumi-nate confirming 120VAC operation.

NOTEThe emergency back-up controlswitches will function when the table isoperating on 120VAC power, batterypower, or turned off.

NOTEThe main power switch can be placed inthe OFF position to completely deacti-vate all table functions if required dur-ing certain procedures or in case ofemergency.

WARNINGDO NOT unlock brakes when a patientis on the table. An uneven patientweight load may cause instability.

WARNINGThe Emergency Brake Release Valvemust be closed and tightened (clock-wise) before activating any hydraulicfunction.

•If the Emergency Brake Release Valvehas been operated, the UNLOCK but-ton on the pendant control will have tobe pressed before brakes will lock again.

SPECIAL USER ATTENTION

Page 10: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

4

CAUTIONCaution should be taken when cleaningthe table to prevent excessive fluid en-try into electrical connectors.

WARNINGAlways follow OSHA blood-borne patho-gens standards for protective clothing,including gloves, masks and eye pro-tection when cleaning the surgical table.

CAUTIONThoroughly read and follow themanufacturer's directions for all clean-ing fluids. DO NOT use cleaners con-taining phenolics.

CAUTIONWhen using spray cleaners DO NOTspray fluids directly into electrical re-ceptacles or micro switches.

CAUTIONBefore replacing pads on the table,make sure the pads and all matingsurfaces are completely dry. Moisturetrapped between the pads and matingsurfaces may cause distortion of tabletops.

NOTENormal table top position is with thehead (and back) section over the powercord end of the base.

WARNINGAlways lock the table top in positionafter rotation. DO NOT rotate the topwith an unevenly distributed patientweight load as instability may result.

WARNING•Make sure the TOP ROTATION LOCKHANDLE is tightened and the brakesare set before transferring the patient.

•Exercise caution with the table toprotated 90° to the base since an improp-erly distributed patient load may causethe table to be tipped over. A tablesupport rod is required for 90° position-ing. See Specialty Positioning.

NOTEAlways follow current AORN JournalGuidelines to ensure proper cleaningand disinfection procedure.

SPECIAL USER ATTENTION

Page 11: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

5

See figure 1-2. The main power ON/OFF switch islocated on the electrical enclosure.

WARNINGRisk of electrical shock. Make surepower cord is disconnected prior toaccessing fuses.

FUSE

MAIN POWERSWITCH

POWERCORD

ELECTRICALENCLOSURE

COVERPLATE

REMOVABLELEG SECTIONSEAT

SECTIONBACKSECTIONHEAD

SECTIONSIDERAIL

HEAD SECTIONLOCKING KNOB

PENDANT CONTROLSTORAGE BRACKET

TOP ROTATIONHANDLE

SERVICEACCESS COVER

POWERCORD

MAIN POWERSWITCH

EMERGENCYBRAKE RELEASE

FLOOR/LOCKBRAKE (4)

PENDANTCONTROL

LEG SECTIONLOCKING KNOB(EARLY MODELS)

LEG SECTIONRELEASE LEVER

Figure 1-1. Elite 6500

1-1. General

SKYTRON’s Elite 6500 Series Surgical Tables areelectro-hydraulically operated, general purpose sur-gical tables. See figure 1-1.

The electro-hydraulic positioning functions oper-ated by the hand-held, push button, pendant controlunit are: trendelenburg, lateral tilt, back section,elevation, leg section, flex/reflex, kidney lift, returnto level, and the floor lock/brake system.

Manual controls are provided for head sectionpositioning, table top rotation, emergency brakerelease and leg section removal.

1-2. Power Requirements

The Elite 6500 Surgical Table requires a 120VAC,60 Hz electrical power supply. The table is equippedwith a 15 foot long power cord with a standard threeprong, hospital grade plug. The electrical protec-tion fuses are located behind a cover plate in theelectrical enclosure on the front edge of the base. Figure 1-2. Power Switch and Fuse Location

SECTION I INTRODUCTION

Page 12: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

6

1-4. Floor Lock/Brake System

The floor lock/brake system consists of four self-leveling, hydraulic brake cylinders which raise andsupport the table base off from the casters. Pressthe TABLE UP button on the pendant control to setthe table’s brakes. An electronic timer will activatethe brake system until the brakes are completelyset, approximately 8-10 seconds.

NOTEActivating any function button will acti-vate the brake system. Using the TABLEUP function to set the brakes providesa visual assurance that the brakes arelocked. As the brake cylinders areextending, the entire table will moveslightly. When the table top begins toelevate, the brakes are fully locked.

To unlock the brakes, press the UNLOCK buttonand release. The brakes will retract automaticallyin approximately 7-8 seconds.

WARNINGDO NOT unlock brakes when a patientis on the table. An uneven patientweight load may cause instability.

NOTEIn case of a power failure or an electri-cal problem within the table, the emer-gency brake release system can beused to move the table. The controllever for this function is located on theside of the table base and is identifiedby an EMERGENCY BRAKE RELEASElabel. Turn the lever counterclockwiseto release the brakes.

WARNING•The Emergency Brake Release Valvemust be closed and tightened (clock-wise) before activating any hydraulicfunction.

•If the Emergency Brake Release Valvehas been operated, the UNLOCK but-ton on the pendant control will have tobe pressed before brakes will lock again.

REV TREND

REFLEXFLEX

TREND

TILT LEFT

BACK UP

TABLE UP

LEG UP

TILT RIGHT

BACK DOWN

TABLE DOWN

LEG DOWN

RETURN BRAKE UNLOCK

KIDNEY UP

KIDNEY DOWN

LATERALTILT LEFT

BACK UP

TABLE UP

LEG UP

FLEX

KIDNEYLIFT UP

RETURNTO LEVEL

LATERALTILT RIGHT

BACK DOWN

TABLE DOWN

LEG DOWN

REFLEX

KIDNEYLIFT DOWN

BRAKERELEASE

REVERSETRENDELENBURG

TRENDELENBURG

POWER INDICATOR

SIDE RAILCLIP

FUNCTIONBUTTONS

POWERINDICATOR

1-3. Pendant Control Unit

The hand-held pendant control unit (figure 1-3) hasa non-slip rubber cover which assures a positivegrip during use. A spring clip hanger located on theback of the control allows it to be stored on thetrendelenburg cylinder cap bracket or the tableside rails.

Figure 1-3. Pendant Control Unit

The function push buttons are identified with ab-breviated descriptions for all functions. See figure1-4. The trendelenburg and table up buttons arered, the remaining buttons are all black.

Figure 1-4. Function Push Buttons

Page 13: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

7

2-1. AC 120V Operation

a. Be sure the power cord is plugged into aproperly grounded, Hospital Grade, 120VAC out-let. Make sure the power cord is routed to the outletto prevent it from being in the way of operatingpersonnel.

WARNINGPrior to operating the table, observe alltable caution labels and review the SPE-CIAL USER ATTENTION section in thefront of this manual.

WARNINGPossible explosion hazard exists if tableis used in the presence of FLAMMABLEANESTHETICS.

b. Depress the main power ON/OFF switchlocated on the electrical enclosure. See figure 2-1. The green AC 120V, Power-On indicator lightlocated in the lower right corner of the pendantcontrol will illuminate. See figure 2-2.

SECTION II BATTERY TABLE OPERATIONc. The table is now ready for 120VAC operation.

Press the TABLE UP button on the pendant controlto set the table’s brakes. An electronic timer willactivate the brake system until the brakes arecompletely set, approximately 8-10 seconds. Pressthe desired function button to position the table.

NOTE•If brakes fail to set, make sure theEmergency Brake Release Lever is tight(clockwise).

•If the Emergency Brake Release Valvehas been operated, the UNLOCK but-ton on the pendant control will have tobe pressed before brakes will lock again.

NOTE•Turning the Main Power Switch ON willchange the table operation to 120 VACpower regardless of the position of theBATT button.

•If Battery operation is activated whileMain Power Switch is ON there will beapproximately a 4 second delay beforeBattery operation activates.

2-2. Battery Operation

a. Make sure the green, AC 120V, Power-Onindicator light, on the hand-held pendant control, isOFF. See figure 2-2. If the indicator light is ON,depress the main power ON/OFF switch, locatedon the electrical enclosure to turn AC120V opera-tion OFF.

NOTEThe table will operate correctly on bat-tery power with the power cord con-nected to a wall outlet or disconnected.

b. Press the BATT button on the hand-heldpendant control. The amber BATTERY indicatorlight, located in the lower left corner of the pendantcontrol, will illuminate. This confirms that the tableis now being operated with battery power.

c. The table is now ready to operate on batterypower. Press TABLE UP button on the pendantcontrol to set the table’s brakes. An electronic timerwill activate the brake system until the brakes arecompletely set, approximately 8-10 seconds. Pressdesired function button to position the table.

BATTERY CHARGINGINDICATOR LIGHT

MAIN POWER SWITCH ELECTRICAL ENCLOSURE

Figure 2-1. Main Power Switch

Figure 2-2. Pendant Control

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

BATT

LATERALTILT LEFT

BACK UPTABLE UP

LEG UPFLEX

KIDNEYLIFT UP

RETURNTO LEVEL

LATERALTILT RIGHT

BACK DOWN

TABLE DOWN

LEG DOWN

REFLEX

KIDNEYLIFT DOWN

BRAKERELEASE

REVERSETRENDELENBURG

TRENDELENBURG

AC120V OPERATIONINDICATOR LIGHT

BATTERYPOWER ON/OFF

BATTERY OPERATIONINDICATOR LIGHT

Page 14: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

8

d. To extend the battery charge life, turn theBATTERY power OFF with the pendant controlwhen the table is not going to be used.

NOTEBattery Operation must be turned OFFat the pendant control. It can not beturned Off using the main power switch.

2-3. Automatic Shut-Off

a. To prevent unnecessary discharge of thebattery, a timer is built into the battery circuit. Thistimer will automatically shut the battery power OFFafter 1½ hours of table inactivity.

b. To turn the table “ON” again, simply press theBATT button on the pendant control and the amberindicator light will illuminate. Select any controlbutton to operate the table.

2-4. Charging the Battery

a. If the battery needs to be charged whenoperating the table on battery power, the amberindicator light on the pendant control will begin toblink. At this time the battery needs to be re-charged.

NOTEWhen the amber light starts to blink(indicating low power in battery) thetable will operate for approximately 5continuous minutes, typically longenough to use the table for the rest ofthe day.

WARNINGThe charging system operates ONLYwhen the table is in AC120V operationmode.

b. To recharge the battery simply plug thepower cord into a 120VAC wall outlet, if notalready plugged in. Turn the main power ON/OFF switch ON by depressing it. The green AC120Voperation light, located on the lower left corner ofthe pendant control, will illuminate.

c. A full battery charge will last approximately 2weeks under normal operating conditions. How-ever, it is recommended to charge the batteries atthe end of each week to establish a normal routineprotocol.

FUNCTION CONTROLACCESS DOOR

Figure 2-3. Emergency Controls Location

b. In the event of either a power failure or aproblem with the hand-held pendant control, thetable can be operated using the emergency back-up switches. Simply push the desired emergencyswitch in the appropriate direction to operate thetable functions. See figure 2-4. These switches arespring-loaded so they return to the neutral or centerposition when released.

NOTEThe table can be operated on 120VACpower while the battery is being re-charged. The green AC 120V indicatorlight (on the pendant control) will illumi-nate confirming 120VAC operation.

2-5. Emergency Back-up Controls

a. The emergency back-up control switches arelocated under the access door on the serviceaccess cover in the table base. See figure 2-3.

Figure 2-4. Emergency Back-Up Controls

NOTEThe emergency back-up controlswitches will function when the table isoperating on 120VAC power, batterypower, or turned off.

LATERAL TILTLEFT/RIGHT

TABLEUP/DOWN

KIDNEYUP/DOWN

TRENDREV. TREND BACK

UP/DOWNLEGUP/DOWN

Page 15: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

9

Figure 3-3. Brake System Activation

Press the BRAKE UNLOCK button on the pendantcontrol to release the four self-leveling brake feet inorder to move the table. See figure 3-3. The brakedelay circuit automatically retracts the brake sys-tem with just one press of the BRAKE UNLOCKbutton. It takes approximately 7-8 seconds tototally release the system.

WARNINGDO NOT unlock brakes when a patientis on the table. An uneven patientweight load may cause instability.

b. Emergency Brake Release. In case of apower failure or an electrical problem within thetable, the emergency brake release system can beused to move the table. The control lever for thisfunction is located on the side of the table base andis identified by an EMERGENCY BRAKE RE-LEASE label. Turn the lever counterclockwise torelease the brakes. See figure 3-4.

3-1. Electrical Powera. Check to be sure the power cord is plugged

into a properly grounded, Hospital Grade, 120VACoutlet. Make sure the power cord is routed so as toprevent it from being in the way of the operatingpersonnel.

b. Depress “Main Power ON/OFF” switch onthe electrical enclosure. See figure 3-1. The greenPOWER ON indicator light on the pendant controlshould now be illuminated.

NOTEThe main power switch can be placed inthe OFF position to completely deacti-vate all table functions if required dur-ing certain procedures or in case ofemergency.

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

TABLE UP(BRAKE LOCK) BRAKE

UNLOCK

3-2. Positioning FunctionsThe hand-held pendant control (figure 3-2) acti-vates the following table functions:

MAINPOWER SWITCH

POWERCORD

ELECTRICALENCLOSURE

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

LATERALTILT LEFT

BACK UP

TABLE UP

LEG UP

FLEX

KIDNEYLIFT UP

RETURNTO LEVEL

LATERALTILT RIGHT

BACK DOWN

TABLE DOWN

LEG DOWN

REFLEX

KIDNEYLIFT DOWN

BRAKERELEASE

REVERSETRENDELENBURG TRENDELENBURG

POWER INDICATOR

Figure 3-2. Pendant Control Unit

Figure 3-1. Main Power Switch

a. Floor Lock/Brake System. To activate thebrakes without affecting table positioning, pressthe TABLE UP button. See figure 3-3. The eleva-tion cylinder will not function until the brakes arecompletely extended.

EMERGENCYBRAKE RELEASE

BRAKE (4)

SERVICEACCESS COVER

POWER CORD

Figure 3-4. Emergency Brake Release

SECTION III OPERATION

Page 16: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

10

Figure 3-6. Lateral Tilt Positioning

e. Back Section. To raise the back section,press the BACK UP button (figure 3-7). The backsection will raise up to 90° above horizontal. Tolower the back section, press the BACK DOWNbutton. The back section will go down to 40° belowhorizontal.

CAUTIONTo prevent damage to the kidney lift, asafety interlock prevents the back sec-tion from going above horizontal if thekidney lift is not all the way down.

WARNINGTo maximize patient safety, utilize properrestraint methods during extreme lat-eral tilt positioning.

WARNINGThe Emergency Brake Release Valvemust be closed and tightened (clock-wise) before activating any hydraulicfunction.

•If the Emergency Brake Release Valvehas been operated, the UNLOCK but-ton on the pendant control will have tobe pressed before brakes will lock again.

NOTEWith an evenly distributed patient weightload, all table positioning functions willoperate smoothly and quietly with apatient weight of up to 500 pounds.(400 pounds SN-1991-5&P).

c. Trendelenburg. To place the surgical table ina Trendelenburg (head down) position, press theTREND button (figure 3-5). Trendelenburg posi-tioning of up to 30° may be obtained. To place thetable in a reverse Trendelenburg (head up) posi-tion, press the REV TREND button. ReverseTrendelenburg positioning of up to 30° may beobtained.

WARNINGTo maximize patient safety, utilizeproper restraint methods during ex-treme Trendelenburg positioning.

Figure 3-5. Trendelenburg Positioning

d. Lateral Tilt. To achieve lateral tilt right (asviewed from the head end of the table), press theTILT RIGHT button (figure 3-6). Tilt of up to 30°may be obtained. To achieve lateral tilt left, pressthe TILT LEFT button. Tilt of up to 30° may beobtained.

Figure 3-7. Back Section Positioning

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

REVERSETRENDELENBURG

TRENDELENBURG

30˚

30˚

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

LATERALTILT LEFT

LATERALTILT RIGHT

30˚ 30˚

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

BACK UPBACKDOWN

40˚

90˚

Page 17: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

11

Figure 3-9. Leg Section Positioning

WARNINGThe Leg section may hit the table baseor the floor if both the leg and elevationsystems are placed in their full downposition.

h. Flex Positioning. To place the table top ina flex position from horizontal, press the FLEXbutton (figure 3-10). To return the table top to ahorizontal position or into a reflex position, pressthe RETURN or REFLEX button.

NOTEWhen REFLEX button is activated, ifKidney Bridge is up, the back sectionwill not go above horizontal.

Figure 3-10. Flex/Reflex Positioning

i. Kidney Lift. To raise the built-in kidney lift,press the KIDNEY UP button (figure 3-11). Up to5 inches of lift can be achieved. Press the KIDNEYDOWN button to lower the kidney lift.

NOTETo prevent damage to the kidney lift, asafety interlock prevents the kidney liftfrom going up if the back section israised above horizontal.

f. Elevation. To raise table top, press theTABLE UP button (figure 3-8). The table will lift apatient weight of 500 pounds up to a maximumheight of 45" (49" with X-Ray top and 2" pad). Tolower the table top, press the TABLE DOWN but-ton. The table top will go down to a minimum heightof 26".

Figure 3-11. Kidney Lift Positioning

Figure 3-8. Elevation Function

g. Leg Section. To lower the leg section, pressthe LEG DOWN button (figure 3-9). The legsection will go down to 105° below horizontal. Toraise the leg section, press the LEG UP button.The leg section will go up to 20° above horizontal.

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

TABLEUP

TABLEDOWN

45"

26"

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

LEG UPLEGDOWN

20˚

105˚

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

5"

KIDNEYUP

KIDNEYDOWN

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

FLEX REFLEX

Page 18: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

12

Figure 3-15. Repositioning Head Section(for use as a Foot Extension)

WARNINGIf the patient position is reversed on thetable (patient's back on the leg section),the patient weight limit is 250 lbs.

3-4. Leg Section Removal.

a. Early Table Models - For increased surgeonaccess in the perineal area, the leg section can beremoved from the table. To remove the leg section,loosen the two locking knobs located beneath theleg section and pull the section out. See figure 3-16. Press the LEG DOWN button on the pendantcontrol to position the leg section attachment pinsdown and out of the way.

To Install Leg Section, Press and hold the LEG UPbutton until the leg section attachment pins com-pletely stop before installing the leg section to thetable. Tighten locking knobs securely.

HEAD SECTIONRELEASE BAR

j. Return To Level. To return the table top to alevel position, press the RETURN button (figure 3-12).

NOTEElevation, kidney lift, and brake systemfunctions are not affected by the returnto level function.

Figure 3-14. Head Section Adjustment

b. By loosening two locking knobs beneath theback section, an additional 5" of longitudinal adjust-ment can be achieved. If desired, the head sectionmay be removed by loosening the locking knobsand pulling it straight out of the back section. 6500Series Tables have the capability of attaching thehead section to the leg section for use as a footextension ONLY. Do Not reverse the patient on thetable without first consulting with SKYTRON. Twolocking knobs are located on the inside of the legsection for securing the head section. See figure 3-15.

Figure 3-12. Return To Level

k. Pendant Control Storage. When the Pen-dant Control is not in use, it should be stored on aconvenient side or end rail. A bracket is locatedunder the table top for storage of the PendantControl when the table is not in use and duringcleaning. See figure 3-13.

Figure 3-13. Pendant Control StorageBracket

3-3. Head Section

a. A quick release positioning bar located underand to the front of the head section (figure 3-14) isused to raise or lower the head section. Pull therelease bar toward the head end to allow thesection to pivot up or down. Positioning from 60°above horizontal to 90° below horizontal in 15°increments is available. Release the bar to lock thehead section in position.

REVTREND

REFLEXFLEX

TREND

TILTLEFT

BACKUP

TABLEUP

LEGUP

TILTRIGHT

BACKDOWN

TABLEDOWN

LEGDOWN

RETURN BRAKEUNLOCK

KIDNEYUP

KIDNEYDOWN

RETURNTO LEVEL

SEATSECTIONBACK

SECTION

PENDANT CONTROLSTORAGE BRACKET TOP ROTATION

HANDLE

HEADSECTION

FOOT/LEGSECTION

LOCKINGKNOB

LOCKINGKNOB

Page 19: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

13

Figure 3-18. 180 Degree Top Rotation

b. The use of the optional support rod allows thetable top to be rotated 90° from the base. See figure3-19.

WARNING•Make sure the TOP ROTATION LOCKHANDLE is tightened and the brakesare set before transferring the patient.

•Exercise caution with the table toprotated 90° to the base since an improp-erly distributed patient load may causethe table to be tipped over. A tablesupport rod is required for 90° position-ing. See Specialty Positioning.

WARNINGEnsure that the leg section is properlyengaged and secured to pins beforeuse to prevent injury.

Figure 3-17. Leg Section Release Levers

3-5. 180 Degree Table Top Rotation.

NOTENormal table top position is with thehead (and back) section over the powercord end of the base.

a. The table top can be horizontally rotated 180°without having to rotate the entire table. To rotatethe top, turn the TOP ROTATION LOCK HANDLEcounterclockwise (figure 3-18), grasp the table bythe head end and rotate the top 180° counterclock-wise. Lock the top in position by tightening the TOPROTATION LOCK HANDLE clockwise.

CENTERCOLUMN

TOP ROTATIONHANDLE

WARNINGAlways lock the table top in positionafter rotation. DO NOT rotate the topwith an unevenly distributed patientweight load as instability may result.

Figure 3-19. 90 Degree Top Rotation

3-6. Specialty Positioning

The use of certain optional accessories availablefrom SKYTRON further extend the positioningcapabilities of the 6500 Series Tables. Refer to thefollowing "Positioning Guidelines" or contact yourSKYTRON representative for further details.

SUPPORT ROD

FOOT/LEG SECTION

LOCKING KNOBS

FOOT/LEG SECTION

LOCKING LEVERS

Figure 3-16. Leg Section Removal

b. Later Model Tables - To remove the legsection on the later model tables, simultaneouslydepress both release levers and pull the leg sectionout. See figure 3-17. Press the LEG DOWN buttonon the pendant control to position the leg sectionattachment pins down and out of the way.

To Install Leg Section Press, and hold the LEG UPbutton until the leg section attachment pins com-pletely stop. Install the leg section on the pins. Pullout on the leg section to make sure the releaselevers are completely locked.

Page 20: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

14

6500 Series Patient Positioning Guidelines

Cardiovascular. Accessories: 6-020-04 “L” type AnesthesiaScreen with Adjustable Wings; (2) 2-010-02 + 2-010-02-2PStandard Armboards with 2” pads; 6-010-41-V RestraintStrap; 4-030-02-B + 4-030-02-2P 8” Foot Extension with 2”Pad.

Cysto/ GYN Accessories: 2-010-02 + 2-010-02-4P StandardArmboard with 4” Pad; 2-020-01-1 Raised Armboard;6-040-05 Clark Socket; 4-080-01-1 Ankl-Lock III Stirrups; (2)6-040-03 Tri-Clamps; 5-010-09-B Radiographic Top;4-040-03 Uro Drain Tray.

Gall Bladder Accessories: 2-020-01-1 Raised Armboard;6-040-05 Clark Socket; 2-010-02 + 2-010-02-4P StandardArmboard with 4” Pad; 6-010-41-V Restraint Straps;4-030-02-B + 4-030-02-2P 8” Foot Extension with 2” Pad;5-010-09-B Radiographic Top.

Thoracic/Kidney Accessories: 2-010-2 + 2-010-02-2P Stan-dard Armboard with 2” pad; 2-020-01-1 Raised Armboard;6-040-05 Clark Socket; 3-010-01 Chest and Waist Supports;4-030-02-B + 4-030-02-2P 8” Foot Extension with 2” Pad.

Abdominal Accessories: 6-020-04 “L” Type AnesthesiaScreen with Adjustable Wings; (2) 2-010-02 + 2-010-02-4PStandard Armboards with 4” Pads; 6-010-41-V RestraintStrap; 4-030-02-B + 4-030-02-2P 8” Foot Extension with 2”Pad; 5-010-09-B Radiographic Top.

Ureteroscopy Accessories: 2-010-02 + 2-010-02-4P Stan-dard Armboard with 4” pad; 2-020-01-1 Raised Armboard;6-040-05 Clark Socket; 4-090-02-1 Levitator Stirrups; (2)6-040-05 Clark Sockets; 4-040-14 Urocatcher II; 4-040-14-1Sterile Pouch; 5-010-09-B Radiographic Top.

Page 21: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

15

6500 Series Patient Positioning Guidelines

Hip Revision Accessories: 2-020-01-1 Raised Armboard;6-040-05 Clark Socket; 2-010-02 + 2-010-02-2P StandardArmboard with 2” Pad.

Lumbar Accessories: (2) 2-010-02 + 2-010-02-2P StandardArmboards with 2” Pad; 6-010-41-V Restraint Strap.

Neurosurgery. Accessories: (2) 2-020-01-1 RaisedArmboard; (2) 6-040-03 Tri-Clamps; 6-010-41-V RestraintStrap; 4-030-02-B + 4-030-02-2P 8” Foot Extension with 2”Pad.

Ophthalmic Accessories: 2-010-02 + 2-010-02-2P StandardArmboards with 2” Pads; 4-030-02-B + 4-030-02-2P 8” FootExtension with 2” Pad; Optional: 7-010-07 Narrow HeadSection (not shown).

Shoulder Arthroscopy Accessories: 3-040-04 Power LiftShoulder Positioner; (2) 2-020-01-1 Raised Armboards; (2)6-040-05 Clark Sockets; 6-010-41-V Restraint Strap;4-030-02-B + 4-030-02-2P 8” Foot Extension with 2” Pad;3-010-01 Chest and Waist Supports.

Neurosurgery Accessories: 7-980-02 Multi-Purpose(“Sugita”) Head Frame; (2) 2-010-02 + 2-010-02-2P StandardArmboards with 2” Pads; 6-010-41-V Restraint Strap;4-030-02-B+ 4-030-02-2P 8” Foot Extension with 2” Pad.

Page 22: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

16

The following procedures should be followed whencleaning the surgical table between cases.

Place table top in level position prior to startingcleaning procedure.

WARNINGAlways follow OSHA blood-borne patho-gens standards for protective clothing,including gloves, masks and eye pro-tection when cleaning the surgical table.

Remove major contaminants from the table withdisposable materials following appropriate biohaz-ard waste disposal procedures.

Remove all table pads and place them on a flatsurface for cleaning.

CAUTIONThoroughly read and follow themanufacturer's directions for all clean-ing fluids. DO NOT use cleaners con-taining phenolics.

Apply cleaning fluid liberally to top and sides ofeach pad and wipe with a clean lint-free cloth.

Using a clean, damp, lint-free cloth, wipe the padsto remove the cleaning fluid.

Using a clean, dry, lint-free cloth, wipe the pads toremove all moisture.

Repeat the steps to clean the bottom of the eachpad.

CAUTIONWhen using spray cleaners DO NOTspray fluids directly into electrical re-ceptacles or micro switches.

Repeat cleaning procedure for all table surfacesincluding the top, sides, elevation column, baseand all accessories.

CAUTIONBefore replacing pads on the table, makesure the pads and all mating surfacesare completely dry. Moisture trappedbetween the pads and mating surfacesmay cause distortion of table tops.

4-1. Preventive Maintenance

The following preventive maintenance checks andservices are recommended to ensure the service-ability and proper operation of your SKYTRONSurgical Table, and should only be performed byqualified SKYTRON trained personel.

a. During normal cleaning, a general visualexamination should be made checking for leaks,loose bolts or parts, and cracked, chipped, ormissing paint. Any necessary repairs should bemade. Refer to the 6500 Maintenance Manual.

b. Semi-annually the following checks and ser-vices should be performed:

1. Check all hydraulic fittings, mini-valves andslave cylinders for proper operation andany signs of leaks.

2. Check the hydraulic speed controls andadjust if necessary.

3. Pressure check (with a gauge) the pres-sure relief valve.

4. Check all mechanical adjustments and ad-just as necessary.

5. Check hydraulic fluid level.

6. Lubricate the slider assembly.

7. Check function of foot leg release levers,(release knob early models). Lubricate asnecessary.

4-2. Cleaning Recommendations

NOTEAlways follow current AORN JournalGuidelines to ensure proper cleaningand disinfection procedure.

CAUTIONCaution should be taken when cleaningthe table to prevent excessive fluid en-try into electrical connectors.

SECTION IV MAINTENANCE

Page 23: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

17

When the cleaning procedure is complete, replaceall pads and accessories as applicable.

Remove pendant control from table side rail andapply cleaning solution to the pendant control andcord.

Use a clean cloth dampened with water to removecleaning solution.

Use another clean damp cloth to remove anyremaining residue.

Install pendant control on side rail for storage whencleaning procedure is complete.

4-3. Service

Table maintenance can be performed by trainedmaintenance personnel using SKYTRON autho-rized replacement parts and service techniques.Service instructions and parts are available fromSKYTRON.

Preventive Maintenance contracts are availablethrough your local SKYTRON representative.

To obtain service instructions, replacement parts,factory service or preventive maintenance con-tracts, contact the SKYTRON representative listedbelow.

Or contact:SKYTRON5000 36th Street S.E.Grand Rapids, MI 495121-800-SKYTRON (1-800-759-8766)Fax. 1-616-957-5053

Page 24: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc
Page 25: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc
Page 26: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc
Page 27: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc
Page 28: ELITE SERIES SURGICAL TABLES - Frank's Hospital Workshop€¦ · OPERATORS MANUAL. 19 MODEL ELITE 6500N SKYTRON STANDARD LIMITED WARRANTY SKYTRON, a Division of the KMW Group, Inc

225000 36th Street S.E., Grand Rapids, MI 49512

1-800-SKYTRON or 1-616-957-0500 • FAX 1-616-957-5053