Upload
kamlesh-gupta
View
68
Download
0
Embed Size (px)
DESCRIPTION
Environment lab manual for civil engineering students
Citation preview
Laboratory Manual
Waste-Water Analysis and Testing
Environment Engineering Lab
Department of Civil Engineering
Maulana Azad National Institute of Technology
Laboratory Manual
For
Water Analysis and Testing
CE-441
Environment Engineering Lab-
Department of Civil Engineering
Maulana Azad National Institute of Technology
Bhopal-462007
Water Analysis and Testing
-II
Department of Civil Engineering
Maulana Azad National Institute of Technology
INDEX
S.No. Name of Experiment Page
No.
1 Determination of total, dissolved, volatile, and fixed residue in a
sewage
2
2 Determination of turbidity 4
3 Estimation of pH 6
4 Estimation of chloride concentration 8
5 Determination of dissolved oxygen and percentage saturation 9
6 Determination of biochemical oxygen demand of waste water 12
7 Estimation of chemical oxygen demand 15
2
1. DETERMINATION OF TOTAL, SUSPENDED, DISSOLVED,
VOLATILE AND FIXED RESIDUE IN A SEWAGE/WATER
SAMPLE
The methods are gravimetric..Great care must be taken to obtain a representative sample
.the quantity of sample to be taken depends on the amount of suspended matter.
• Water 100ml to 500 ml
• Sewage 50 ml to
• Water effluent 50 ml to 100 ml
PROCEDURE
(A) TOTAL RESIDUE
• Place the required quantity of the sample in a dry constant weight dish or crucible.
• Evaporate to dryness in an oven at 103*C-105*C and dry to constant weight.
• Cool the dish in a desicator.
• Weigh and note the increase in weight.
• Total residue in mg/l =
(weight of crucible with residue - weight of empty crucible)×1000
sample (in ml)
(B) total solids : volatile and fixed
Ignite the residue obtained in (a) at 600 degree celcius in a muffle furnace (15 - 20 min),
cool ad weigh.
Total volatile residue in mg/l = (wt. of crucible with total residue – wt. of empty
crucible) – ( wt. of crucible with residue heated to
600*c) mg ×
1000))/ Sample(in ml)
Fixed residue mg/l = total residue - volatile residue
(c) Suspended and dissolved solids:
Filter the sample through whatman filter paper no 44 or 41. Take a suitamle quantity in
a weighed dry dish/crucible/corning or pyrex beaker. Evaporate to dryness at 103*C
-105*C .
• Cool container to a constant weight. Weigh and note the increase in weight.
3
• Dissolved residue mg/l = (weight of crucible with reside – weight of empty
crucible)mg × 1000/ sample(in ml)
• Suspended residue mg/l = total residue (a)-dissolved residue(c)
Sample
details
Type of
residue
weight
Source volume Empty
beaker
With residue Mg/l residue
Total residue
volatile
residue
fixed residue
dissolved
residue
Suspended
residue
Remarks
4
2. DETERMINATION OF TURBIDITY
Four turbidimeters are commonly employed in measurements.
Standard Jackson candle turbidimeter = 25 1000 Units (J.T.U.)
Baylis turbidimeter = 0 2 Units (J.T.U.)
Heliges / Aplab turbidimeter = 0 150 Units (J.T.U)
Nephlometer / Hach’s turbidimeter = 1 - 1000 N.T.U.
Standard Jackson Candle Turbidimeter
It consists of three parts: a calibrated glass tube, a holder and a candle (bee’s wax
candle designed to burn within the limits of 114 to 125 grains/hour). The glass tube and
the candle are supported in a vertical position so the centre line of the tube passes through
the centre line of the candle. The top of the support of the candle must be 7.6 cm below
the bottom of the glass tube.
Pour the sample in the glass tube until the image in the glass tube disappears. Note the
light path in cm (measure from inside bottom of the glass tube) and read in the table for
turbidity units. For turbidity exceeding 1000 units dilute the sample with turbidity free
water and proceed.
Cm. Turbidity Cm. Turbidity Cm. Turbidity Cm. Turbidity Cm. Turbidity
2.3 1000 5.8 380 8.1 270 13.5 160 28.1 75
2.6 900 5.9 370 8.4 260 14.4 150 29.8 70
2.9 800 6.1 360 8.7 250 15.4 140 31.8 65
3.2 700 6.3 350 9.1 240 16.6 130 34.1 60
3.5 650 6.4 340 9.5 230 18.0 120 36.7 55
3.8 600 6.6 330 9.9 220 19.6 110 39.8 50
4.1 550 6.8 320 10.3 210 21.5 100 43.5 45
4.5 500 7.0 310 10.8 200 22.6 95 48.1 40
4.9 450 7.3 300 11.4 190 23.8 90 54.0 35
5.5 400 7.5 290 12.5 180 25.1 85 61.8 30
5.6 390 7.8 280 12.7 170 26.7 80 72.9 25
5
Nephlometer
A turbidimeter consisting of a Nephlometer with a light source and one or more
photoelectric detectors and a read out mechanism at right angles to the path of light.
There are various commercial models available. Elico-Hach.
The equipment is calibrated with commercially available hydrazine sulphate
suspensions.
Measurements are made directly up to 40 NTU. If the turbidity is more, the sample is
diluted.
Calculation
NTU = Aх (B+C)/C
where
A = NTU of diluted sample
B = Volume of dilution water, ml
C = Sample of volume taken for dilution, ml.
Jackson’s Turbidimeter
Sample details Observations
Source Length path cm Turbidity JTU
Hach/Heliges/Aplab turbidimeter/
Nephlometer
No. of
bulb
Type of
filter
Reading on
scale
Turbidity
6
3. ESTIMATION OF pH
MATERIALS REQUIRED: Waste water and pH indicator
THEORY:
pH is the negative logarithm of hydrogen ion. For neutral water pH is 7 and acidic water
pH is less than 7 and more than 7for basic water.
Generally fresh sewage is alkaline in nature (pH>7) but as time passes its pH tends to fall
due to production of acids by bacterial action in anaerobic and nitrification processes.
The pH however rises upon the treatment of sewage.
There are two general methods for determination of pH
1. Colorimetric method or use of indicators
A number of pH indicators have been used. They are unreliable for measuring pH below
3 and above 10
2. Electrometric method
A number of electrodes have been suggested for the electrometric method of pH
measurement. The glass electrode is the standard electrode.
PROCEDURE:
Colorimetric method:
1. 10 ml of the sample is placed in each of the tubes of the pH comparator
2. The appropriate of PH indicator as indicated in the kit (0.3 ml of phenol red) is
added to one of them
3. It should be checked that the same indicator disc is fitted to the comparator
4. The tube in the comparator is placed in such a way that the colour standards are
opposite to the tube not containing the indicator
5. The tube is compared with the colour standards and colour nearest to the sample
6. The pH is noted.
Electrometeric method:
1. The pH meter is standardized by immersing the electrode in buffer solution of
known pH normally 4 and 9.2
2. The pH is read and adjusted correctly with the control knob, till the meter needle
indicates the correct value for the pH of the buffer solution
3. The electrodes is rinsed in distilled water and immersed in the sample and the
needle is settled at one point and pH value is read.
7
sample details Observations
Colorimetric Indicator Qty. of
indicator
Colour Ph
Elecrometric pH Buffer used
4. ESTIMATION OF THE CHLORIDE CONCENTRATION
Chlorides occur in all natural waters varying concentrations. Upland and mountain
streams are usually low in chloride concentration. Chlorides gain access in many ways
viz solvent power of water dissolve salts from top soil. Spray from ocean, seas invading
fresh waters during tides, underground formation and seepage, human excreta, industrial
wastes etc.
Reagents
(i)Potassium chromate indicator (8)
(ii)N/35.5 Silver nitrate solution (9)
Procedure
Take 100ml sample in two conical flasks. Add to both one drop of potassium
chromate indicator. Titrate with standard N/35.5 AgNO3 solution in one and compare
with other to distinguish change from yellow to brick red. Note the amount of titrant
used. The first permanent change of color should be noted.
Chlorides as Cl- = ml of AgnO3 used for sample X 1000
ml of sample
The pH should be between 6-8
Chemical reactions
XCl-
→ X
+ + Cl
-
AgNO3 →
Ag
+ + NO3
-
Ag
+ + Cl
-- → AgCl
2Ag+ + CrO4 →
Ag2CrO4
Data Sheet 4.7 Estimation of chloride concentration
Sample details Observation Chlorides
Source Volume Initial
burette
reading
Final burette
reading
ml AgNO3
Solution
used
Chlorides
mg/l
Remarks: The chloride values are normally low in surface inland waters. Ground waters
have comparatively higher values. Chlorides have sanitary significance in waters of low
chloride values.
8
5. DETERMINATION OF DISSOLVED OXYGEN AND PERCENTAGE
SATURATION
Dissolved oxygen vary in nature and waste waters to a large extent, from the state of no
oxygen to saturation level. The saturation values change with temperature, pressure,
altitude and chloride concentration. The test is very delicate for assessment of pollutional
load.
Reagents
(i) Manganous sulfate solution (10)
(ii) Alkaline potassium iodide (11)
(iii) N/40 Sodium thiosulphate solution (12)
(iv) Conc. Sulphuric acid
(v) Starch indicator
Procedure
Alsterberg modification of Winkler’s method
Find out the exact capacity of the bottle in ml. Collect the sample in a narrow mouth flat
stoppered reagent bottle of approximate 300 ml capacity.
It is always preferable to collect the sample through D.O sampler to avoid contact with
air. The bottle should be completely filled. Add 1.0 ml manganous sulphate solution by a
pipette, dipping the end blow the surface. Some water would overflow. Add 1.0ml
alkaline potassium iodide solution. Insert the stopper and mix thoroughly. Let the
precipitate settle. Add 2.0 ml conc. sulphuric acid. Dissolve the precipitate by vigorous
shaking. Take the calculated amount of aliquot as calculated below:
200 x exact capacity of bottle = ……….ml
exact capacity of bottle -4
Titrate with N/40 sodium thiosulphate solution using starch as indicator. Record the ml
titrant used.
Calculations
Since 1ml of N/40 Na2S2O3 = 0.2 mg oxygen, the ml of this solution used is equivalent to
mg/litre of dissolved oxygen.
Determine the temperature and chloride concentration and by reference to solubility table
on the next page find out the saturation value. Calculate percent saturation.
9
Chemical reactions
MnSO4 → Mn++
+ SO4- -
Alk K → K+ + I
- + OH
-
Mn++
+ 2OH- → Mn(OH)2
White precipitate means oxygen absent
Mn++
+ 2OH- +
1/2O2 → MnO2 + H2O
Brown precipitate means oxygen present
This is referred to as oxygen fixation.
MnO2 + 4H+ + 2I → IO2 + Mn
++ + 2 H2O
Released
I2 + 2 Na2S2O3 → Na2S4O6 + 2NaI
Solubility table oxygen at various temperatures (pressure 760 mm)
Chloride mg/litre Chloride mg/litre Temp
◦C 0 5000 Diff. per 100
Mg/litre
Temp
◦C 0 5000 Diff. per 100
Mg/litre
0 14.62 13.79 0.0165 16 9.95 9.46 0.0098
1 14.23 13.41 0.0160 17 9.74 9.26 0.0095
2 13.84 13.05 0.0154 18 9.54 9.07 0.0092
3 13.48 12.72 0.0149 19 9.35 8.89 0.0089
4 13.13 12.41 0.0144 20 9.17 8.73 0.0088
5 12.80 12.09 0.0140 21 8.99 8.57 0.0086
6 12.48 11.79 0.0135 22 8.83 8.42 0.0084
7 12.17 11.51 0.0130 23 8.68 8.22 0.0083
8 11.87 11.24 0.0125 24 8.53 8.12 0.0082
9 11.59 10.97 0.0121 25 8.38 7.96 0.0081
10 11.33 10.73 0.0118 26 8.22 7.81 0.0080
11 11.08 10.49 0.0114 27 8.07 7.67 0.0079
12 10.83 10.28 0.0110 28 7.92 7.53 0.0078
13 10.60 10.05 0.0107 29 7.77 7.39 0.0076
14 10.37 9.85 0.0104 30 7.63 7.25 0.0074
15 10.15 9.65 0.0100 -
10
Sample details Observations
Source Temp
Capacity
of
Bottle
Volume
Titrated Initial
burette
Reading
Final
burette
reading
ml n/40
Na2S2O3used
Dissolved
Oxygen
mg/l
Remarks: Dissolved oxygen is one parameter which indicates the health of a stream.
More oxygen better is the health (except when blooming with algae).
11
(BOD) OF WASTE WATER
The biochemichal oxygen demand (BOD) is the amount required by bacteria while
stabilizing decomposable organic matter under aerobic conditions. The quantity of
oxygen required may be taken as a measure of its content of decomposable organic
matter. The rate of BOD exertion is governed by the characteristics of its sewage, its
decomposable organic content, bacterial population and its temperature. The progressive
BOD exertion takes place in two stages
a. Carbonaceous
b. Nitrification
It has been observed that a large percentage of the total BOD is exerted is 5 days at 20
degrees Celsius. The value of 5 days at 20 deg celcius is to a reasonable extent
comparable to 4 days at 30 deg celcius and at 35 deg celcius.
The first stage BOD reaction is represented as:
Y=L(1-10^(-kt))
Where
Y=BOD at any time
L=ultimate BOD
K=reaction rate
T=time, days
Reagents
1)Phosphate buffer 13
2)Magnesium sulphate solution 14
3)Calcium chloride solution 15
4)Ferric chloride solution 16
6.DETERMINATION OF BIOCHEMICAL OXYGEN DEMAND
5)Manganous sulphate solution 10
6)Alkaline potassium iodide solution 11
7)N/40 sodium thiosulphate solution 12
8)Conc. Sulphuric acid
9)starch indicator
BOD measurable with different various dilutions
Range of BOD %Mixture
200-700 1.00
100-350 2.00
40-140 5.00
20-70 10.00
10-35 20.00
4-40 50.00
Procedure:
Preapare dilution water by adding 1.0 ml of phosphate buffer solution, magnesium
sulphate solution, calcium chloride solution and ferric chloride solution to 1.0 litre of
distilled water. Add 2.0 ml settled sewage and aerate. .Determine the exact capacity of
three BOD bottles. Find out the DO of undiluted sample as in 4.8 and designate Dos
Prepare the desired percent mixture by adding sample in dilution water. Fill up one
bottle with the mixture and the other one with dilution water blank. Incubate at a fixed
temperature for a definite time(20 deg celcius,5 days:30 deg celcius,4 days and 35 deg
celcius,3 days).Find out DO in both the bottles after incubation and designate:
Mixture as (DOi)
Blank(DOb)
Calculations
BOD mg/l=[(DOb- DOi) 100%-(DOi- DOs)]
13
Sample details %mixture DO3 DO2 DO1 BOD mg/l
Data sheet 4.9 Determination of BOD
Remarks: The sensitivity of test is dependent on the type of seed used. It is advisable to
use settled and acclimatized at 20 deg celcius seed. Checking also is possible by using a
synthetic sample mixture(mixture of glucose and glutamic acid, in equal amounts).1 mg
of glucose would be approximately equal to 1 mg TOD or 0.8 of BOD.
14
7.ESTIMATION OF CHEMICAL OXYGEN DEMAND
COD test is widely used for measuring the pollution strength of waste waters. All aoganic
compounds with few exceptions can be oxidized to carbon dioxide and water by the
action of strong oxidizing agents regardless of biological assimilability of subsantances.
Reagents
(1)standard potassium dichromate 0.25N (17)
(2)Sulphuric acid (with 1 gm of silver sulphate in every 0.75 ml acid
(3)Ferroin indicator solution (18)
(4)Standard ferrous ammonium sulphate solution (19)
Standardization procedure
Dilute 25 ml standard potassium dichromate solution 250ml with distilled water. Add
20ml conc Sulphuric acid. Titrate with ferrous ammonium sulphate using Ferroin
indicator to red end point.
Normality of FeSO4(NH4)2 SO4 = ml of K2Cr2O7Х0.25/ml of FeSO4(NH4)2 SO4
Procedure
The procedures are either through open reflux or closed treatment
Open reflux method
Take 50ml sample or a smaller amount diluted to 50.0ml in a refluxing flask, add boiling
chips and 1gm HgSO4. Cool the mixture. Add 0.25ml of 0.25N K2Cr2O7 solution and mix
again. Attach the condenser and start cooling water. Add remaining acid (70ml) through
the open end of the condenser, mix the reflux mixture. Apply heat and reflux the mixture
for two hour cool
Dilute the mixture to about 300ml and titrate excess of dichromate with Standard ferrous
ammonium sulphate solution using Ferroin indicator. The colour will change from yellow
to green blue and finally red. Record the ml of titrant used.
Reflux in the same manner a blank consisting of distilled water, equal to the volume of
the sample and the reagents.
Titrate the sample. Record the ml of titrant used.
Closed reflux method
Place 5ml or fraction diluted to 100ml of sample with distilled water in hard glass bottle
and add 25ml standard potassium dichromate solution. Carefully add 75ml conc sulphuric
acid mixing after each addition. Digest the mixture in pressure cooker or autoclave for
30min. repeat the procedure with 100ml distilled water and same reagent. Transfer the
contents to 500ml conical flask. Dilute the mixture to 350ml. titrate excess of dichromate
with Standard ferrous ammonium sulphate solution using Ferroin indicator. The end point
is red.
15
Table reagents quantities and normality for various sample sizes
Sample
size(ml)
0.25N
K2Cr2O7
(ml)
Conc. H2SO4.
with
AgSO4.ml
HgSO4.
ML
Normality of
FeSO4(NH4)2
SO4
Final vol before
titration ml
10 5 15 0.2 0.05 70
20 10 30 0.4 0.1 140
30 15 45 0.6 0.15 210
40 20 60 0.8 0.20 280
50 25 75 1.0 0.25 350
Calculation
COD (mg/l)= (A-B)Cх8000/ml of sample
Where
A = ml of FeSO4(NH4)2 SO4 for blank.
B = ml of FeSO4(NH4)2 SO4 for sample
C = Normality of
FeSO4(NH4)2 SO4 solutions determined above.
Data sheet for estimation of of COD
Sample
detail
Sample
source
Normality
of
K2Cr2O7
Amount of
K2Cr2O7
Added
Normality of
FeSO4(NH4)2
SO4
Ml of
FeSO4(NH4)2
SO4
COD
(MG/L)
16