42
E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 1 Zurzach 10.01.2010 Active scanning beams: 1. Modulating delivery Eros Pedroni Paul Scherrer Institute SWITZERLAND

Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 1

Zurzach10.01.2010

Active scanning beams: 1. Modulating delivery

Eros Pedroni

Paul Scherrer InstituteSWITZERLAND

Page 2: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 2

1. Beam delivery options 2. Scanning technology: experience with Gantry 1

3. Dose precision issues 4. Practical advantages of scanning

5. The organ motion problem – link to the next talk “new developments”

of

Wednesday 6. Questions

Summary

Page 3: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 3

1. BEAM DELIVERY OPTIONS FOR SCANNING

Page 4: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 4

Typical “physical”

scanning beam–

x= ±3 mm θ= ±10 mrad δp/p

= ±0.5%

1.1 The basic element –

the proton “pencil”

beam

Common

to all delivery methods–

Pencil beams with zero phase space and given initial energy

1 mm sigma

3 mm sigma

10 mm sigma

Monte Carlo

Scanning

Wobbling

After integration in the lateral direction the differences in depth disappear

Difference

between beam delivery methods?–

The “cumulated”

phase space

at

nozzle exit

(beam formation)–

The time structure

of the beam

and beam delivery sequence•

Which is relevant for

organ motion errors

Page 5: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 5

1.2 Options for beam spreading in the

lateral direction

Scattering (static) –

Constant particle fluence

(homogeneous dose field)

Single scattering (good penumbra –

but low efficiency)•

Double scattering (higher efficiency –

less good penumbra)

Contoured 2nd scatter (or double contoured to compensate energy mixing)–

Field size and depth dependent solutions

Uniform scanning (dynamic) •

Circular (wobbling) -

rectangular–

spiral –

BEV shapes

Conformal scanning (dynamic)–

Modulated particle fluence

(for 3d-conformation and IMPT)

Beam motion

magnetic•

Or through patient table motion

Performance issues:•

Precision of the dose shaping

Scanning speed –> repainting capability –> to reduce organ motion

sensitivity

Page 6: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 6

Lateral scanning options

Spot scanning–

Switch beam OFF in between spots

PSI Gantry 1–

Let beam ON in-between spots

GSI raster scanning•

Continuous scanning–

Magnet-driven scan

Dose shaping by changing the magnet speed (at constant dose rate)

Time-driven scan•

The beam moves with maximum speed –

the dose is painted by

modulating the beam intensity–

PSI Gantry 2

(feasibility ?)

Homogeneous energy layer

Non-homogeneous energy layer

Page 7: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 7

1.2 Options of beam formation in depth (Bragg peak)

Rotating wheel (very fast -

200 ms / cycle)–

Fast SOBP mixing (quasi-static)

Must be designed for each –

energy -

SOBP extent -

field size–

Not of interest in combination with scanning

Ridge filter –

(miniature structure blurred by angular confusion)–

SOBP energy mix (static)

Must be designed for each -

energy -

SOBP extent -

field size–

Concern –

angular confusion of the beam -

penumbra

Range shifter

(dynamic –

30 ms) -

> see PSI Gantry 1–

Plates (regular or digital) -

moving wedges -

water column

Range steps or continuous–

Concern –

beam spot broadening in the air gap

to the patient

Variable energy of the beam

(dynamic) -> see PSI Gantry 2–

Range steps

Concerns –

very steep Bragg peaks at the lowest energies•

Best in combination with a ridge filter or preabsorber?

Page 8: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 8

Range shifter

-> PSI Gantry 1–

Variable amount of material in the beam

Water-equivalent?•

Material, geometry, density of the plates

With zero air gapPencil beam size invariant with range

With non-zero air gap

-

problematic the •

Spot broadening due to MCS in the RS

Dynamic beam energy

-> PSI G2–

With degrader

and beam line

See “future developments talk”

Page 9: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 9

1.3 Relation between proton energy and fluence

for conformal dose shaping

Depth

Starting example : hat boxUniform proton fluenceFixed SOBP

Target

Gray shading = density of proton stops

Patient

Collimator

100%(0%)

100%(80-50%)

Fixed range modulation =unnecessarydose delivery

BEAM

Page 10: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 10

3d or 4d-shaped dose conformal scanning

Intensity modulation within energy-layerDelivery spot by spot

Collimator compensator scattering-wobbling

Collimator

OAR

Compensator

Stacking of energy-layersVariable range modulation

OAR

Compensator

MLC Shape shrinkingUniformScanning

Scanning:Freedom to deliverany pattern

UnnecessaryDose delivery

Page 11: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 11

2. PRACTICAL DEVELOPMENT OF SCANNING THE PSI EXPERIENCE WITH GANTRY 1

Page 12: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 12

2.1 PSI Scanning

Development started in 1989–

With limited resources

Compromises due to the limited space of the area

Parasitic use of the PSI beam•

With very long shut-downs

Page 13: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 13

The decision to build Gantry 1 was taken in 1992

Until May 2008 …the only gantry with scanning (1st patient in 1996)–

Magnetic scanning started before the last bending magnet

parallel scanning

(but only along one magnetic axis) •

gantry radius

reduced to only 2m

Eccentric mounting of the patient table on the gantry front wheel•

Patient moves away from the floor when treating with beam from below (a drawback)

α rotation

φ rotationβ rotationUpstream scanning –

eccentric design

Page 14: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 14

If I could do it again …

… Eccentric compact Gantry 2

Eccentric mounting as with Gantry 1 (R = 2m)–

But with rotation only on one side (0°

180°)

-

as with the new Gantry 2

With a counter-rotation–

False floor underneath the patient table moving with the gantry

360° 180°

Page 15: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 15

X Sweeper magnet most often used

Y Range shifter 2nd

loop

Gaussian pencil beam of 3 mm sigma–

Cartesian scanning (infinite SSD)

“Step and shoot”

spot on a 5 mm grid

Z Patient table slowest loop

Time

Spot-Dose Monitor + Fast Kicker

The sequence of the elements of scanning:

Discrete pencil beam scanning

2.1 Scanning on the PSI Gantry 1

Weak point: transverse scanning by moving patient table–

Slow motion ( no repainting possible)–

We can treat only non moving targets

Head, spinal chord and low pelvis

Page 16: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 16

Dimensional considerations for scanning

Reference size of 1 liter (very often much less, our max value 4

liters)–

Assumed beam size: 3 mm sigma (see next section)

Derived grid size: 5 mm

(21 spot lateral -

23 in depth)•

21 x 21 x 23 ~ 10’000 spots/liter

21 x 21 x 10 cm = 44 m path length–

Assumed treatment time: 1.5 minute

beam-ON time

10 ms/spot in average–

Due to the non-uniform spot weights (for a uniform SOBP)

Most distal spots: 60 ms•

Most proximal: 3 ms

(with beam spot weight optimization: we accept spots >= 0.5 ms)•

Treatment roughly proportional to -

VOLUME (the moving time)

and to BEV SURFACE (beam ON time of the high weighted distal Bragg peaks)•

Required intensity: 0.2 nA proton current–

beam ON time of 1 Min to deliver 1 Gy

in1 Liter

Page 17: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 17

Beam monitoring

Transmission ionization chambers (M1

and M2) –

proton flux–

5 (10) mm

in air –

2kV

voltage

response time < 100 μs

–> precision of controlling the dose •

1% of the mean spot time

(switching time of the kicker 50 μs)•

Delay of the current measurement and of the kicker subtracted in

advance from

preset -> 0.2% of mean spot time•

Strip-monitor chamber –

4 mm strips

Measure position and width of the beam after the delivery of each spot

Position resolution: < 0.5 mm–

Charge collection time ~ 0.8 ms

Wait 1 ms before reading scalers

at the end of the spot

Page 18: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 18

The reasons to switch off the beam in-between spots

Avoid dose uncertainties when stepping with beam ON to the next spot (transients)–

Dose errors (sweeper power supplies delay and non linearities)

Errors in checking the beam position with the strip monitor (time resolution of 0.8 ms at 1 cm/ms => 8 mm error)

Poor quality of the beam until 2006 (before COMET) –

Beam splitting

( 0.5% intensity) from the 2 mA

beam of the PSI ring

cyclotron and degrading it from 590 MeV down to 100-200 MeV•

After any repair -> bad vacuum -

> intensity spikes

Check monitor units precisely at the end of spot •

Conservative strategy –

Perform all calculations at the end of a static spot

Start next spot only when previous delivery shown to be correct -> overall dead time of 5 ms in-between spots

Best approach to control the dose with repainting

Beam IC

SweeperHall probe

Page 19: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 19

Sweeper magnets–

30 ms for a full sweep of 20 cm

For a 5 mm step•

Time to move the beam and to stabilize: 3 ms

< time to check previous spot: 5 ms

(Gantry 2 2ms)•

Range shifter–

40 plates (80 pneumatic valves)

4.5 mm thick each + one half plate •

Water equivalent arrangement

Dead time 50 ms

(30 ms for motion)•

Patient table–

Moves in steps of 5 mm

1-2 seconds

dead time per step•

Acceleration and decelaration

Smooth motion for patient comfort

Scanning motion devices

Page 20: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 20

The price to pay: the overall scanning dead time

Gantry 1–

Beam off after delivery of each spot

Sweeper dead time 10’000 x 5 ms = 50 s •

Range shifter 21x21x 50 ms= 22 s

Patient table 21 x 1s = 21 s–

1.5 minutes beam off vs. 1.5 minute beam on

Duty factor of our present discrete scanning low 50%•

Precision of dose delivery very good

Dose reproducibility of ~ 0.2%

Future: exploring more efficient solutions with Gantry 2–

Painting of lines instead of spots (check of dose delivery at the end of a line)

-> see “future developments”

presentation

Page 21: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 21

Beam delivery–

Small systematic errors in the sweeper calibration

Relative beam position precision of±

0.2 mm from one spot to next

Target motion during beam delivery (patient)

Statistical error–

Gets reduced by

sqrt( N*m*k)•

N= number of fraction

m= number of fields•

k= repainting number

Practical limit for Gantry 1–

Motions < ±

1-2 mm (including full fractionation N=30)

The required precision of placing the spots

Organ motion error is a function of ratio Motion size / pencil beam size

Page 22: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 22

The reproducibility of the beam tunes–

Automatic set-up of the beam energy without retuning the beam during treatments(cyclic ramping of the gantry magnets)

The beam appears at the correct position within 1.5 mm•

Position correction at the end of the first spot–

Correction allowed if within ±1.5 mm

All further spots•

Position deviation within ±

0.5 mm

of the expected value–

Interlock if deviation > 1.5 mm

On-line and off-line analysis of delivered spots–

Analysis of logged data

U sweeper position

Beam position error

1 mm

The precision of setting the beam (beam tunes)

Page 23: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 23

3. DOSE SHAPING PRECISION OF SCANNING

Page 24: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 24

3.1 Distal fall-off

Desirable Δp/pAt highest energy ~0.2% FWHM At lowest energy ~2%

FWHM

Momentum band dependence

Function of energy (linear with range)–

Range straggling

(σR/R ~1%) unavoidable

Momentum band

avoidable

Energy dependence

Very sharp Bragg peaksReason to use a pre-absorber

at low

energies (for variable beam energy)

302010 cm

Page 25: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 25

Distal fall-off

Scanning with variable energy–

In principle superior to scattering

No material in the beam (no compensator –

no range shifter)•

Minimalst

energy for the required range (physical limit)

Less range straggling•

Sharper distal fall-off

In practice the advantage compared to scattering is probably very marginal

Page 26: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 26

3.2 Lateral fall-off

Governed by the spot size within the patient (dose spot) –

… at the Bragg peak (the major effect)

Derived from –

The beam broadening due to the Multiple Coulomb Scattering in the patient

Given by the physics –

unavoidable•

In doesn't help to use a beam size which is much smaller than this effect

The beam size in air

(size of the beam at the exit of the nozzle)•

Given by the beam delivery system by design

Accelerator -

beam line -

and the material in the nozzle

The smaller the dose spot -> the sharper the lateral fall-off•

Sigma lateral fall-off ~=~ sigma dose spot

Page 27: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 27

Optimization of the beam size (spot size) •

Beam in vacuum–

Provide a beam source with small phase space

Beam in air–

Bring the vacuum very close to the patient

Beam monitors (material in the nozzle)–

Place material close to the nozzle exit

Reduce amount of material

(of low atomic number)–

Reduce air gap to the patient !!!!

Size = θ

x Distance θ

= F(E) x sqrt

(Material

/ S)

Beam pipe

Range shifter or pre-absorber –

Reduce air gap to the patient !!!!

Nozzle

Patient –

MCS–

You can do nothing about this

Page 28: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 28

The air gap problem when using a range shifter

119 MeV 177 MeV 214 MeV

Gap(cm) Gap(cm)

FWH

M (c

m)

FWH

M (c

m)

0 0 0

10

10 10

20

2020

3030

Beam blow-up due to MCS in the range shifter in front of the patient–

The reason with Gantry 2 to go for variable beam energy

Strategy of positioning beam modifiers in the beam•

Either very close

to the patient

Small air gap

Or very far

Loss of intensity•

But not in-between

The worse you can do

Similar problem with scattering –

air gap to compensator

lateral penumbra

Gap(cm)

FWH

M (c

m)

Page 29: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 29

Patient MCS

MCS in the patient PMCS•

Depends linearly with range–

Rule of thumb 1 cm FWHM at 20 cm

1.5 cm FWHM at 30 cm

Choice of the beam size (BS) in air –

Sum in quadrature

of the beam size in air (BS) and the MCS in the patient (Pmcs)

if we choose BS < 50%

Pmcs

we obtain sqrt( Pmcs^2 + BS^2) < 1.12

Pmcs–

A too small beam size -> needs smaller scan grid –> more spots -> more dead time

and is more sensitive to organ motion errors

Page 30: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 30

“Edge enhancement capability”

of scanning

Delivery of separated spots–

Variable choice of intensity

The dose lateral fall

can be made to be similar to the fall-off of the original beam Gaussian

Uniform fluence

of spots–

The case of collimation

Gaussian

folded with step function = error-function

Max difference–

Factor 1.7

(1.4)

Scanning with optimization can produce a sharper lateral fall-

off as compared to scattering

Page 31: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 31

Beam width at the Bragg peak as a function of the range

Assumed phase space: 6 mm FWHM

constant in vacuum at all energies•

Relevant for organ motion

errors

practical limit -

beam line

Nozzle material: as for

Gantry 1

practical limit -

monitoring–

MCS in the patient enhanced by factor a 1.4 in favor of scanning

Collimation improves precision only at ranges below 11 cm

Factor 1.4 higherdue to the error-

function

Scanning with collimationbetter

Scanning alone better

Idealized scattering(zero phase space)Realistic scanning

Page 32: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 32

3.2 Lateral fall-off

Scanning with variable energy (with or without collimator)–

In principle superior to scattering

No material in the beam (no compensator –

no range shifter)–

No MCS propagating in the air gap –

sharp beam

Less angular confusion –

sharp shadow when using collimators

Page 33: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 33

SCANNING ONLY APERTURE

CONTOURED APERTUREOAR-

SHIELDING BLOCK

Options for “collimated scanning”

Scanned field

Sensitive structure

LESS WEIGTH

BETTER VISIBILITYFOR BEV X-RAYS

Page 34: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 34

Possible scanning patterns

Criteria

Precision (edge of the field)

Duty factor (beam-off time)

“Topological”

scan (repainting capability)

Spot scanning (G1)

Spot scanning++ (G2) Line scanning (G2)

Raster scanning (GSI)

Contoured scanning (G2)IM

IM

Page 35: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 35

4. ADVANTAGES OF SCANNING

Page 36: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 36

All done by software –

with minimal equipment

No need to use individualized hardware–

Avoid fabrication and mounting of patient specific equipment in the nozzle

Apply dose fields sequences in one go without personnel entering in the treatment room

To reduce treatment time•

All fields of IMPT are delivered in the same fraction

Most efficient use of the beam–

“All”

used protons reach the target

Minimal neutron background

(for the patient) by default•

Less activation of equipment

(for the personnel) by default

Flexibility to treat from small to very large fields without changing equipment–

“All done by the beam”

Sharper penumbra than scattering (scanning alone or collimated) (Gantry 2)–

Less material before the patient (no compensator)

Page 37: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 37

Variable modulation of the range

For avoiding unnecessary 100% dose on the healthy tissues–

Especially relevant for large tumors

Page 38: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 38

4D –

(dose-modulated fields)

Dose tailored geometrically in 3d•

Dose shaping within the target–

Used for

Intensity modulated therapy

IMPT•

Biological targeting

Competition with IMRT

Page 39: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 39

Other possible scanning related “advantages”

Gantry design with “upstream scanning”–

Reduce the radius of the gantry

No additional radial distance for the spreading the beam–

Parallelism of scanning

–source at infinite distance

Simplify dosimetry –

treatment planning –

field patching -

collimation and compensation

Capability to simulate scattering (repainted BEV box scans)–

Scanning can simulate and improve scattering (variable range modulation)

The opposite is probably not true–

Provide backward compatibility with the more established techniques using one and the same nozzle (see future developments -> Gantry 2)

The only drawback compared to scattering–

The SENSITIVITY TO ORGAN MOTION ERRORS

A problem common to all dynamic beam delivery methods including IMRT

Page 40: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 40

5. THE ORGAN MOTION PROBLEM

Page 41: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 41

5.1 An unsolved problem: organ motion errors of scanning

Disturbance of the lateral dose fall-off (same problem for scattering and scanning)

Add safety margins or•

Reduce with Gating

or

Tracking

Disturbance of the dose homogeneity–

Scattering –

highly repainted -

insensitive

Single painted scanning -

very sensitive–

Repainted scanning

-

acceptable

Alone

for medium motion•

With Gating

or Tracking

for large motion

The experience of treating moving targets with scanning is still inexistent–

WE HAVE TO LEARN HOW TO DO THAT

Prospective solutions …

see next talk on future developments with Gantry 2

Page 42: Eros Pedroni Paul Scherrer Institute SWITZERLAND Zurzach 10.01 · E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010

E. Pedroni Center for Proton Radiation Therapy - Paul Scherrer Institute - Proton Therapy Winter School 10 -01-2010 42

THANK YOU

A very exciting field …

Based on a beautiful idea

The next step –

scanning speed