Fadahunsi Ct Seminar

Embed Size (px)

Citation preview

  • 8/6/2019 Fadahunsi Ct Seminar

    1/16

    .

    (a) briefly describe the difference

    between a 4th

    generation and a Helical CTscanner. What are the advantages of

    Helical scanner over conventional

    scanner.(b) Concerning CT, explain the following

    including factors affecting them.

    (i) CT contrast(ii) spatial contrast

    Why is molybdenum usually used as

    DF1

  • 8/6/2019 Fadahunsi Ct Seminar

    2/16

  • 8/6/2019 Fadahunsi Ct Seminar

    3/16

    Filter. Illustrate your answer with an appropriate x-ray spectra diagram.

    Describe the properties of screen/film combinationused in mammography.

    ANSWER

    In the 4th generation CT scan the x-ray tube has to

    be provided with a high voltage supply and thedetectors have to pass their signals to thecomputer, the gantry is also wired to an externalequipment, because of these, at the completion ofevery slice after accelerating the x-ray tube to a

    rotational scanning speed, the system mustdecelerate to a gradual stop and then repeat theprocess by rotating in the opposite direction inorder to unwind the high voltage, control, powerand signal cables take up assembly.A

  • 8/6/2019 Fadahunsi Ct Seminar

    4/16

    In the helical scanner, a slip ring technology has

    been incorporated which allows gantry control

    and data signal to pass from and into acontinuously rotating gantry. The high voltage

    supply used by the x-ray generator uses a

    compact high frequency transformer that ismounted within the rotating gantry. All power

    required by the gantry is passed through the slip

    rings thereby allowing it to rotate continuously

    without the need to stop after every slices.

  • 8/6/2019 Fadahunsi Ct Seminar

    5/16

    ADVANTAGES

    The basic advantage of continuous rotation is to permitfaster rotation times.

    The continuously rotating gantry allows continuousacquisition of data, reducing the inter scan delays. It is

    possible to do dynamic scans at up to 12 scans per minute.A patient volume is scanned continuously without gaps and

    with fewer problems with patient motion particularlypatient breathing and this is helpful in reformatted sagittaland coronal images.

    Any slice can be selected within the volume to display insteadof the fixed slice position of the conventional 4th generationscan.

    There is reduced use of contrast medium.

  • 8/6/2019 Fadahunsi Ct Seminar

    6/16

    CT contrast or contrast resolution or densityresolution is the ability of an imaging system to

    display an image of a relatively large object thatis only slightly different in density from itssurroundings. For the image to be visible, theobject must produce enough change in the no of

    transmitted photons to overcome statisticalfluctuation in transmitted photons caused bynoise i.e. in the computation of any one pixelvalue there is error in the form of statisticalvariation and it is this variation which limits theultimate contrast resolution and this variation isreferred to as image noise which manifest itselfas a grainy background or mottle.

  • 8/6/2019 Fadahunsi Ct Seminar

    7/16

    Since noise is the ultimate limitation in the accuracy ofcontrast resolution, factors that affect or induce noise affectcontrast and these factors are

    (1) photon flux--- in any ray measurement the statisticalvariation associated with that measurement is directlyproportional to the number of photons detected, the relativeimage noise is affected inversely as the square root of thedetected photon flux. large photon fluxes therefore reduces

    relative noise although increases patient dose. The photon intensity is affected by the efficiency of the

    detector, photon flux, size of patient and the presence of highattenuation material and in turn the uncertainty in themeasurement. A large patient thus will have a more mottled

    image than a small one for the same tube output. Presence ofhigh contrast objects such as bone will cause an increase innoise by reducing the transmitted photon flux.

  • 8/6/2019 Fadahunsi Ct Seminar

    8/16

    X-ray scatter---- Because any particular detector cannot discriminatebetween primary photon directly from the source and a scattered photon

    arising from an area not in line of the ray projection, this alters the recordedvalues in adjacent detectors and produces a decrease in difference betweenadjacent measurements with a concomitant decrease in contrast resolution.

    Voxel length---- if the slice thickness is reduced by 50% the detectorcollimator size must be reduced accordingly which in turn reduces theintensity by the same amount. To retain the same accuracy requires

    restoration of the detected photon intensity which means doubling thedose.

    Similarly decreasing pixel size increases the relative inaccuracy unless thephoton flux is raised. If the pixel side length is reduced by half the voxelvolume will be reduced by a factor of four, the SD of the CT value woulddouble and the image will become more mottled in appearance and the

    contrast will deteriorate. Filters ------- These are used to remove image blurring created by back

    projection process by accentuating high frequency component found in thedata and image noise increases with this thereby decreasing contrast.

  • 8/6/2019 Fadahunsi Ct Seminar

    9/16

  • 8/6/2019 Fadahunsi Ct Seminar

    10/16

    Filters---- the various filters control the amount

    of image blurring created by accentuating high

    frequency component found in the data. For a

    crisp image the high spatial frequencies are

    accentuated, and this has the effect of

    sharpening the edges and improving spatialresolution.

    Opening size of detector aperture----- This is the

    effective size of each detector in the imageplane. A small detector aperture produces a

    narrow ray, less blur and better image detail.

    When a portion of the detector is covered,

    however, the geometric efficiency is reduced.

  • 8/6/2019 Fadahunsi Ct Seminar

    11/16

    An increase in radiation exposure to the patient

    is then required to produce the same image

    quality with respect to noise.

    Ray sample interval----- If the spacing between

    rays significantly exceeds the dimension of small

    objects or anatomical detail, the detail will notappear on the image. The rays must be

    sufficiently close during the scanning procedure

    to measure any anatomical detail that is toappear in the image.

    Voxel and pixel size-------since a specific pixel can

    have only one CT number value, there can be no

    detail within a ixel.

  • 8/6/2019 Fadahunsi Ct Seminar

    12/16

    In other words all detail within the tissue voxel

    represented by a pixel is blurred together andassigned a single value. Anatomical detail within

    the voxel cannot be imaged therefore small

    voxels are needed when image detail is required.

    Three factors determine voxel size: field of view,

    matrix size and slice thickness.

    Since mammography is a technique for the

    radiographic examination of the breast whichconsist of connective tissue, glandular tissue,

    skin and fat, and all must be visualized but all

    have very similar attenuation coefficient.

  • 8/6/2019 Fadahunsi Ct Seminar

    13/16

    In order to maximize difference in attenuation

    coefficient of these various structures a low

    energy x-rays spectrum needs to be used. For

    this imaging situation, the use of a broad

    spectrum bremsstrahlung spectrum from

    tungsten is far from ideal and it is now standardpractice to make use of the characteristic

    radiation fro a lower atomic number material.

    Molybdenum has an intermediate atomicnumber of 42, has a k edge at 20keV and

    produces two rather intense characteristic x-ray

    energies: k-alpha radiation, at 17.9keV, and k-

    beta, at 19.5keV.

  • 8/6/2019 Fadahunsi Ct Seminar

    14/16

    The x-ray tube potential is about 30kVp.

    When the tube is operated at 30 to 40 kVp, a

    molybdenum tube will also produce

    considerable bremsstrahlung with high energies

    greater than 20keV. This higher energy radiation

    will reduce contrast in breast soft tissuestructures. To reduce the amount of higher

    energy radiation in the molybdenum tube

    spectrum, a molybdenum filter of 0.030mmthickness is commonly used.

    This filter will attenuate x-rays just above the

    20keV K edge of Mo very strongly, but will

    transmit 57% of the 17.5K-alpha and 67%of the

  • 8/6/2019 Fadahunsi Ct Seminar

    15/16

    19.6keV K- beta radiation of molybdenum.

    The screens used in mammography are rare

    earth screens in which the principal interaction is

    the photoelectric effect with the L shell electrons

    (the K edge for rare earth materials is in the

    region of 50keV). Because of the low x-rayenergy, the screen are thinner than for routine

    radiography, thus reducing unsharpness.

    A single screen is used with a single sidedemulsion. The screen (and emulsion) is on the

    distal side of the film, furthest from the x-ray

    tube and patient. This arrangement is used

    because the highest proportion of the photon

  • 8/6/2019 Fadahunsi Ct Seminar

    16/16

    interactions and thus light production, will be

    on the side of the screen closest to the film and

    this serves to minimize further the screen

    unsharpness.

    They are able to produce a limiting resolution

    of 151p/mm or better.