6
Introduction Flexichamber ® is a compact, anti-static valved collapsible holding chamber used to deliver medication from most pressurized Metered Dose Inhalers (pMDIs). This device is designed to be used for delivering aerosol medication into the lungs. When you release a spray of aerosol from the MDI into the device, the medication is held in the reservoir bag for a moment. The reservoir bag removes most larger aerosol drug particles while allowing smaller particles to enter into the lungs during inhalation. The Flexichamber ® reservoir bag pulls inward during inhalation enabling the patient to appreciate the depth of inhalation of the medication. The compact design of the device makes it convenient to carry. The removable Backpiece allows the device to be easily cleaned. Please discuss the use of Flexichamber ® with your physician, pharmacist, respiratory therapist, or other healthcare professional. Always read and follow instructions as provided in this product leaflet and as directed by your healthcare professional. Indications For Use Flexichamber ® is an anti-static valved collapsible holding chamber intended to be used by patients under the care or treatment of a physician or licensed healthcare professional. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers (pMDIs). The intended environments for use include home, hospitals, and clinics. This product is intended for patients who can follow verbal instructions. Instructions For Use 1. Remove device from carton and outer plastic wrap covering the device (See Figure 1). 2. Before using the product, carefully examine the device for any damage (such as loose, broken, or missing parts; or holes, tears, or rips in the Reservoir Bag or Backpiece). Check for and remove any foreign objects in the chamber (Reservoir Bag) and mouthpiece to prevent possible choking hazards. This product contains a duckbill-shaped inhalation valve which is a small part and could pose a choking hazard if removed. Do not remove the inhalation valve. Proper disassembly and cleaning instructions are provided with this leaflet. If the product is damaged or is missing any parts then it should be replaced immediately. If necessary, use the MDI alone until a replacement Flexichamber ® is obtained. If the patient’s symptoms worsen, please seek immediate medical care. 3. To prepare the device for use, remove the Cover from the device (See Figure 2). With the Mouthpiece facing you, place thumbs on the Tabs, and push the Backpiece up toward you (See Figure 3). 4. Once the Mouthpiece has popped out of the base, gently pull it away from the Backpiece to expand the Reservoir Bag (See Figure 4). 5. Remove the cap from the MDI (See Figure 5). 6. Shake the MDI immediately before each use as described in the instructions provided with the MDI medication (See Figure 6). 7. Insert the MDI into the Backpiece of the device (See Figure 7). Ensure the MDI is securely inserted into the Backpiece. 8. Grasp the Backpiece firmly, place Mouthpiece into mouth and close lips around Mouthpiece to form an effective seal. Press down on the MDI to dispense one dose of medication into the Reservoir Bag. Administer only one (1) puff at a time (See Figure 8). 9. Inhale slowly and deeply through the mouth until a full breath has been taken. The Reservoir Bag may pull inward slightly as you inhale (See Figure 9). If possible, hold breath for 5-10 seconds, while keeping lips tightly around the Mouthpiece to maintain seal for 2-3 breaths. 10. Repeat steps 8-9, following the instructions provided with the MDI and as directed by your physician, including amount of time to wait between doses. 11. Important Note: Before repeating steps 8-9 ensure the Reservoir Bag has been restored to its full size (no longer in a deflated or collapsed state) by pulling the Mouthpiece gently away from the Backpiece (See Figure 4). (Continued on reverse side) Flexichamber ® Single patient, multi-use, disposable Anti-static Valved Collapsible Holding Chamber Instructions for use: Before using Flexichamber ® , it is important to read and understand these instructions and the instructions provided with your Metered Dose Inhaler (MDI). Figure 1 Figure 4 Figure 7 Figure 2 Figure 5 Figure 8 Figure 3 Figure 6 Figure 9 MOUTHPIECE BACKPIECE RESERVOIR BAG COVER TAB INHALATION VALVE

Flexichamber - Avadel Pharmaceuticals · clinics. This product is intended for patients who can follow verbal instructions. Instructions For Use 1.ve device from carton and outer

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Flexichamber - Avadel Pharmaceuticals · clinics. This product is intended for patients who can follow verbal instructions. Instructions For Use 1.ve device from carton and outer

IntroductionFlexichamber® is a compact, anti-static valved collapsible holding chamber used to deliver medication from most pressurized Metered Dose Inhalers (pMDIs). This device is designed to be used for delivering aerosol medication into the lungs.When you release a spray of aerosol from the MDI into the device, the medication is held in the reservoir bag for a moment. The reservoir bag removes most larger aerosol drug particles while allowing smaller particles to enter into the lungs during inhalation.The Flexichamber® reservoir bag pulls inward during inhalation enabling the patient to appreciate the depth of inhalation of the medication. The compact design of the device makes it convenient to carry. The removable Backpiece allows the device to be easily cleaned.Please discuss the use of Flexichamber® with your physician, pharmacist, respiratory therapist, or other healthcare professional. Always read and follow instructions as provided in this product leaflet and as directed by your healthcare professional.

Indications For UseFlexichamber® is an anti-static valved collapsible holding chamber intended to be used by patients under the care or treatment of a physician or licensed healthcare professional. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers (pMDIs). The intended environments for use include home, hospitals, and clinics. This product is intended for patients who can follow verbal instructions.

Instructions For Use1. Remove device from carton and outer plastic wrap covering

the device (See Figure 1).

2. Before using the product, carefullyexamine the device for any damage(such as loose, broken, or missingparts; or holes, tears, or rips in theReservoir Bag or Backpiece). Checkfor and remove any foreign objectsin the chamber (Reservoir Bag) andmouthpiece to prevent possiblechoking hazards. This productcontains a duckbill-shaped inhalationvalve which is a small part andcould pose a choking hazard ifremoved. Do not remove theinhalation valve. Proper disassemblyand cleaning instructions areprovided with this leaflet. If theproduct is damaged or is missingany parts then it should be replacedimmediately. If necessary, use theMDI alone until a replacementFlexichamber® is obtained. If thepatient’s symptoms worsen, pleaseseek immediate medical care.

3. To prepare the device for use, removethe Cover from the device (SeeFigure 2). With the Mouthpiece

facing you, place thumbs on the Tabs, and push the Backpiece up toward you (See Figure 3).

4. Once the Mouthpiece haspopped out of the base, gentlypull it away from the Backpieceto expand the Reservoir Bag(See Figure 4).

5. Remove the cap from the MDI(See Figure 5).

6. Shake the MDI immediatelybefore each use as describedin the instructions providedwith the MDI medication(See Figure 6).

7. Insert the MDI into the Backpieceof the device (See Figure 7).Ensure the MDI is securelyinserted into the Backpiece.

8. Grasp the Backpiece firmly, placeMouthpiece into mouth and closelips around Mouthpiece to forman effective seal. Press down onthe MDI to dispense one dose ofmedication into the Reservoir Bag.Administer only one (1) puff at atime (See Figure 8).

9. Inhale slowly and deeply throughthe mouth until a full breath hasbeen taken. The Reservoir Bagmay pull inward slightly as youinhale (See Figure 9). If possible,hold breath for 5-10 seconds,while keeping lips tightly aroundthe Mouthpiece to maintain sealfor 2-3 breaths.

10. Repeat steps 8-9, followingthe instructions provided withthe MDI and as directed by yourphysician, including amount oftime to wait between doses.

11. Important Note: Before repeatingsteps 8-9 ensure the ReservoirBag has been restored to its fullsize (no longer in a deflated orcollapsed state) by pulling theMouthpiece gently away fromthe Backpiece (See Figure 4).

(Continued on reverse side)

Flexichamber®

Single patient, multi-use, disposable

Anti-static Valved Collapsible Holding Chamber Instructions for use: Before using Flexichamber®, it is important to read and understand these instructions and the instructions provided with your Metered Dose Inhaler (MDI).

Figure 1

Figure 4

Figure 7

Figure 2

Figure 5

Figure 8

Figure 3

Figure 6

Figure 9

MOUTHPIECE BACKPIECE

RESERVOIR BAG

COVER TAB

INHALATIONVALVE

Page 2: Flexichamber - Avadel Pharmaceuticals · clinics. This product is intended for patients who can follow verbal instructions. Instructions For Use 1.ve device from carton and outer

When finished taking your medication as prescribed by your physician, remove the MDI from the Backpiece and close the device by carefully twisting the Mouthpiece and pushing toward the Backpiece, not allowing the Reservoir Bag to become lodged between the Backpiece and the Mouthpiece (See Figure 10). Replace the Cover.Helpful Hints1. Take slow deep breaths during inhalations.2. For small children and elderly who may have difficulty using

Flexichamber®, allow another adult to assist with administration.3. Keep Flexichamber® closed and with Cover in place when not in use.

Important Cleaning InstructionsClean Flexichamber® by hand weekly, or more frequently, if required depending on your use of the product. Flexichamber® is NOT dishwasher or microwave safe.1. Remove the Cover.2. With the Mouthpiece facing you,

place thumbs on the Tabs, and pushthe Backpiece up toward you to openthe device.

3. Remove the Backpiece from thedevice. Do not remove any otherparts during disassembly and cleaningas it may result in damage to thedevice and present a possible chokinghazard (See Figure 11).

4. Wash the parts in lukewarm waterusing a mild, liquid dish detergentand a soft cloth. Do not use abrasivedetergents or cloths, or any cleaningagents other than mild dish detergent,as this may result in tears, holes, orother damage to the Reservoir Bag(See Figure 12).

5. Use fresh, lukewarm water to rinsethe device and its parts.

6. To dry, gently shake off excess water.Let all parts air dry completely,keeping the device in its open,extended position (See Figure 13).Do not place near artificial heat, suchas a dishwasher or oven.

7. When parts are completely dry,reassemble the Backpiece firmly intoplace. Visually inspect to ensure theBackpiece is on tightly and no otherparts are loose, missing, or damaged.Check for and remove any foreignobjects or materials.

8. Replace the Cover and close thedevice by carefully twisting and pushing down toward theBackpiece (See Figure 14).

9. Store in a cool, dry place.10. This is a multi-use device, but should be examined prior to

each use for signs of damage. Check for loose, broken, ormissing parts; or holes, tears or rips in the Reservoir Bag orBackpiece. Damaged devices should be replaced immediately.

Important Information Concerning Your Medication1. Package insert dosing instructions should be consulted for all

metered-dose inhalers (MDIs) when used with Flexichamber®.Dosage and administration recommendations vary for differentMDIs. The limitations and conditions of use for each productshould be considered before utilizing this device.

2. Product may need to be replaced after repeated use. Properstorage, cleaning, and environmental conditions will affect theproduct life span.

3. Store in a cool, dry place.4. Do not share this device with anyone. It is intended for use with

a single patient.5. Do not place in dishwasher. Wash Flexichamber® by hand using

lukewarm water and a mild, non-abrasive dish detergent.6. If you notice medication build-up within the Reservoir Bag, wash

as described in the Important Cleaning Instructions section of thisproduct insert.

7. This product contains small parts (duckbill-shaped inhalationvalve) which could pose a possible choking hazard if removed andingested. Never remove the inhalation valve. Do not removeany other parts, except as described in the cleaning instructions.

8. Prior to each use, check the device for and remove any foreignobjects or materials. Check for any damage to the product. Donot use if damaged.

Do not leave device unattended with children.

Note: You will need a physician’s prescription for a new Flexichamber® device.

Caution: Federal law restricts this device to sale by, or on the order of, a physician.

HOW SUPPLIED: Flexichamber® Anti-static NDC 13551-901-01 Valved Collapsible Holding Chamber

Figure 11

Figure 10

Figure 12

Figure 13

Figure 14

Distributed By: Avadel Pharmaceuticals (USA), Inc., Chesterfield, MO 63005www.avadel.com

Rev 11/16Patent flexichamber.com1116 FLX DEV IFU WEB ENG

Page 3: Flexichamber - Avadel Pharmaceuticals · clinics. This product is intended for patients who can follow verbal instructions. Instructions For Use 1.ve device from carton and outer

BackgroundBefore using the Flexichamber® Face Mask with the Flexichamber® Anti-static Valved Collapsible Holding Chamber, it is important to read and understand these instructions, the instructions provided with the Flexichamber® Anti-static Valved Collapsible Holding Chamber, and the instructions provided with your Metered Dose Inhaler (MDI). Use as directed by your healthcare professional.

Indications For UseThe Flexichamber® Face Mask is intended to be used with the Flexichamber® Anti-static Valved Collapsible Holding Chamber and by patients under the care or treatment of a physician or licensed healthcare professional. The Flexichamber® Face Mask and the Flexichamber® Anti-static Valved Collapsible Holding Chamber are intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers. The intended environments for use include home, hospitals, and clinics. This product is intended for patients who can follow verbal instructions.

IntroductionFor most children, the MDI delivers medicine most effectively when given through a valved holding chamber. A valved holding chamber contains a one-way valve that does not allow the patient to breathe back into the chamber. This feature is very important because it allows young children to breathe normally and still get a reliable dose of medication. The mask that is attached to the chamber fits over the child’s nose and mouth and forms a seal. This seal keeps the medication from leaking out into the air.

Instructions For Use1. Remove the Flexichamber® Face

Mask from the carton and outerplastic wrap covering the mask.(See Figure 1).

2. Attach the Flexichamber® Face Maskto the end of the Flexichamber®

Anti-static Valved Collapsible Holding Chamber (See Figure 2).

3. Apply the mask over the mouth andnose and ensure there is a good seal(See Figure 3).

4. Press down on the MDI to dispenseone dose of medication into theReservoir Bag. Administer only one(1) puff at a time (See Figure 4).

5. Inhale slowly and deeply throughthe nose and mouth until a full breathhas been taken. The Reservoir Bagmay pull inward slightly as you inhale(See Figure 5). If possible, hold breathfor 5-10 seconds, while maintaininga seal with the mask.

6. Repeat steps 4-5, following theinstructions provided with the MDI,the Flexichamber® Anti-static ValvedHolding Chamber, and as directedby your physician, including amountof time to wait between doses.

7. Important Note: Before repeatingsteps 4-5 ensure the Reservoir Baghas been restored to its full size(no longer in a deflated or collapsedstate) by pulling the Mouthpiecegently away from the Backpiece(See Figure 6).

When finished taking your medication as prescribed by your physician, remove the MDI from the Backpiece; remove the face mask, and then close the device by carefully twisting the Mouthpiece and pushing toward the Backpiece, not allowing the Reservoir Bag to become lodged between the Backpiece and the Mouthpiece (See Figure 7). Replace the Cover.

Important Cleaning InstructionsClean the Flexichamber® Face Mask by hand weekly, or as required depending on your use of the product. The mask is NOT dishwasher or microwave safe.1. Remove the mask from the

Flexichamber® device.2. Wash in lukewarm water using a

mild, liquid dish detergent and a softcloth. Do not use abrasive detergentsor cloths as this may result in damageto the face mask (See Figure 8).

3. Use fresh lukewarm water to rinsethe mask.

4. To dry, gently shake off excess water. Let the mask air dry. Donot place near artificial heat, such as a dishwasher or oven.

5. When completely dry, store the Flexichamber® Face Mask withthe Flexichamber® Anti-static Valved Collapsible Holding Chamber.

6. Store in a cool, dry place.

Caution: Federal law restricts this device to sale by, or on the order of, a physician.

HOW SUPPLIED:The Flexichamber® Face Mask comes in three (3) sizes:Small Child Mask NDC # 13551-903-01 Large Child Mask NDC # 13551-903-02 Small Adult Mask NDC # 13551-903-03

Flexichamber® Face Mask For Use With The Flexichamber® Anti-static Valved Collapsible Holding Chamber Single patient, multi-use, disposable

Figure 1

Figure 2

Figure 3

Figure 4

Figure 5

Figure 6

Figure 7

Figure 8

Distributed By: Avadel Pharmaceuticals (USA), Inc., Chesterfield, MO 63005www.avadel.com

Rev 11/161116 FLX MSK IFU WEB ENG

Page 4: Flexichamber - Avadel Pharmaceuticals · clinics. This product is intended for patients who can follow verbal instructions. Instructions For Use 1.ve device from carton and outer

IntroducciónFlexichamber® es una cámara espaciadora plegable, antiestática, con válvula y compacta, utilizada para administrar medicamento desde la mayoría de los inhaladores de dosis medidas presurizadas (IDMp). Este dispositivo está diseñado para enviar medicamentos en aerosol a los pulmones.

Cuando usted libera un chorro pulverizado de aerosol desde el IDM hacia el dispositivo, el medicamento se mantiene en la bolsa de depósito durante un momento. La bolsa de depósito elimina las partículas más grandes del fármaco en aerosol, mientras permite que las partículas más pequeñas ingresen en los pulmones durante la inhalación.

La bolsa de depósito Flexichamber® se retrae durante la inhalación, lo que permite al paciente mejorar la profundidad de la inhalación del medicamento. El diseño compacto del dispositivo hace que sea cómodo de transportar. La pieza posterior extraíble permite la fácil limpieza del dispositivo.

Analice el uso de Flexichamber® con su médico, farmacéutico, terapeuta respiratorio u otro profesional sanitario. Siempre lea y siga las instrucciones según se establece en el folleto de este producto y según las indicaciones de su profesional sanitario.

Indicaciones de usoFlexichamber® es una cámara espaciadora plegable, antiestática y con válvula destinada a ser utilizada por pacientes que se encuentran en tratamiento o a cargo de un médico o un profesional sanitario matriculado. Estos pacientes utilizan el dispositivo para administrar medicamento en aerosol desde la mayoría de los inhaladores de dosis medidas presurizadas (IDMp). El dispositivo está diseñado para ser utilizado en entornos tales como hogares, hospitales y clínicas. Este producto es para pacientes que pueden seguir indicaciones verbales.

Instrucciones de uso1. Retire el dispositivo del envase

de cartón y del plástico externo que lo recubre (ver Figura 1).

2. Antes de usar el producto, examinedetenidamente el dispositivo paradetectar cualquier posible daño (comopiezas sueltas, rotas o faltantes; oagujeros, roturas o rasgaduras en la bolsade depósito o en la pieza posterior).Inspeccione y retire cualquier objetoextraño en la cámara (bolsa de depósito)y la boquilla para evitar cualquier posibleriesgo de atragantamiento. Este productocontiene una válvula de inhalación conforma de “pico de pato”, que es unapieza pequeña y podría provocar unriesgo de atragantamiento si se retira. Noretire la válvula de inhalación. Este folletobrinda instrucciones para desarmar ylimpiar el producto de manera adecuada.Si el producto se daña o carece de algunapieza debe reemplazarse de inmediato.En caso de ser necesario, use solamenteel IDM hasta que se obtenga unaFlexichamber® de reemplazo. Si lossíntomas del paciente empeoran, busqueatención médica de inmediato.

3. Para preparar el dispositivo parasu uso, retire la tapa del dispositivo(ver Figura 2). Con la boquilla frentea usted, coloque los dedos pulgaressobre las lengüetas y empuje la piezaposterior hacia usted (ver Figura 3).

4. Una vez que la boquilla ha salido de labase, tire de ella alejándola de la piezaposterior para expandir la bolsa dedepósito (ver Figura 4).

5. Quite la tapa del IDM (ver Figura 5).

6. Agite el IDM inmediatamente antesde cada uso, según se describe en lasinstrucciones que se proporcionan conel medicamento del IDM (ver Figura 6).

7. Inserte el IDM en la pieza posterior deldispositivo (ver Figura 7). Asegúrese deque el IDMen la pieza posterior.

8. colóquese la boquilla en la boca y cierre los labios alrededor de la boquilla para

abajo sobre el IDM para enviar una dosis de medicamento a la bolsa de depósito. Administre solo una (1) descarga por vez (ver Figura 8).

9. Inhale lenta y profundamente por la bocahasta que haya realizado una inspiracióncompleta. La bolsa de depósito podríaretraerse hacia adentro levemente amedida que usted inhala (ver Figura 9).Si es posible, aguante la respiración entre5 y 10 segundos mientras mantiene los

la boquilla para mantener el sello entre2 y 3 respiraciones.

10. Repita los pasos 8 y 9, siguiendo lasinstrucciones provistas con el IDM ysegún lo indique su médico, lo queincluye la cantidad de tiempo que debeesperar entre las dosis.

11. Nota importante: Antes de repetir lospasos 8 y 9 asegúrese de que la bolsade depósito haya recuperado su tamañocompleto (es decir, que ya no tenga una

suavemente la boquilla de la piezaposterior (ver Figura 4).

(Continúa al dorso)

Flexichamber®Para un solo paciente, multiuso, desechable

Cámara espaciadora plegable, antiestática y con válvula Instrucciones de uso: Antes de usar Flexichamber®, es importante leer y comprender estas instrucciones y las instrucciones proporcionadas en su inhalador de dosis medidas (IDM).

Figura 1

Figura 2

Figura 3

Figura 4

Figura 7

Figura 5

Figura 8

Figura 6

Figura 9

BOQUILLA PIEZA POSTERIOR

BOLSA DE DEPÓSITO

TAPA LENGÜETA

VÁLVULA DEINHALACIÓN

Page 5: Flexichamber - Avadel Pharmaceuticals · clinics. This product is intended for patients who can follow verbal instructions. Instructions For Use 1.ve device from carton and outer

Consejos útiles1. Respire lenta y profundamente durante

las inhalaciones.

2. para usar la Flexichamber®, por lo que es posible que necesiten la ayuda de un adulto para administrarlo.

3. Mantenga la Flexichamber® cerrada y tapada cuando no esté en uso.

Instrucciones importantes de limpiezaLimpie la Flexichamber® a mano semanalmente, o con más frecuencia, en dependencia del uso del producto. La Flexichamber® NO es apta para lavavajillas u horno microondas.1. Retire la tapa.

2. Con la boquilla frente a usted, coloquelos dedos pulgares sobre las lengüetasy empuje la pieza posterior hacia arribay hacia usted para abrir el dispositivo.

3. Retire la pieza posterior del dispositivo.No retire ninguna otra pieza mientrasdesarma y limpia el dispositivo, yaque esto puede dañarlo y presentarun posible riesgo de atragantamiento(ver Figura 11).

4. Lave las piezas en agua tibia condetergente para vajilla líquido suave ycon un paño suave. No use detergentesni paños abrasivos, ni ningún agentelimpiador que no sea detergente paravajilla de acción leve, ya que esto puedeprovocar roturas, agujeros u otros dañosa la bolsa de depósito (ver Figura 12).

5. Use agua fresca y tibia para enjuagarel dispositivo y sus piezas.

6. Para secar, agite suavemente paraeliminar el exceso de agua. Dejeque todas las piezas se sequencompletamente al aire, manteniendoel dispositivo en su posición abiertay extendida (ver Figura 13). No locoloque cerca de fuentes de calor

7. Cuando las piezas estén completamentesecas, vuelva a colocar la pieza posterior

visualmente para asegurar que la pieza

y que no haya otras piezas sueltas,o piezas que falten o estén dañadas.Inspeccione y retire cualquier objetoo material extraño.

8. .

9. Almacene en un lugar frío y seco.

10.

Información importante sobre su medicamento1. o

del paquete para todos los inhaladores de dosis medidas (IDM), cuando estos se usan con la Flexichamber® . Las recomendaciones

Se deben considerar las limitaciones y condiciones de uso de cada producto antes de usar este dispositivo.

2.

3. Almacene en un lugar frío y seco.4. No comparta este dispositivo con ninguna otra persona. Este debe

ser utilizado por un solo paciente.5. No lave el dispositivo en el lavavajillas. Lave la Flexichamber® a mano,

con agua tibia y detergente para vajilla de acción leve y no abrasivo.6. Si nota acumulación de medicamento dentro de la bolsa de

depósito, lávela según se describe en la sección de instruccionesimportantes de limpieza en el folleto de este producto.

7. Este producto contiene piezas pequeñas (válvula de inhalación conforma de pico de pato) que podrían plantear un posible riesgo deatragantamiento si se retiran e ingieren. Nunca retire la válvula deinhalación. No quite ninguna otra pieza, excepto según se describeen las instrucciones de limpieza.

8. Antes de cada uso, inspeccione el dispositivo y quite cualquier

No lo use si está dañado.

No deje el dispositivo sin supervisión al alcance de los niños.

Nota: Necesitará una receta de un médico para obtener un nuevo dispositivo Flexichamber®.

Precaución: Según la ley federal, este dispositivo solo puede ser vendido por un médico o por indicación de este.

PRESENTACIÓN: Cámara espaciadora plegable, NDC 13551-901-01 antiestática y con válvula Flexichamber®

Distribuido por: Avadel Pharmaceuticals (USA), Inc., Chesterfield, MO 63005www.avadel.comRev 11/16Patente flexichamber.com1116 FLX DEV IFU WEB SP

Figura 11

Figur 10

Figura 12

Figura 13

Figura 14

Cuando termine de tomar el medicamento según lo recetado por su médico, saque el inhalador de dosis medidas (IDM) de la pieza posterior y cierre el dispositivo con cuidado girando la boquilla y ejerciendo presión hacia la pieza posterior, sin permitir que la bolsa del depósito se aloje entre la pieza posterior y la boquilla (Ver Figura 10). Vuelva a colocar la tapa.

Vuelva a colocar la tapa y cierre el dispositivo girando con cuidado y ejerciendo presión hacia abajo hacia la pieza posterior (Ver Figura 14).

Este es un dispositivo multiuso, pero debe examinarse antes de cada uso para detectar señales de daños. Verifique si hay piezas sueltas, rotas o faltantes; o agujeros, roturas o rasgaduras en la bolsa del depósito o en la pieza posterior. Los dispositivos dañados deben reemplazarse inmediatamente.

Es posible que sea necesario reemplazar el producto después de un uso repetido. El almacenamiento, la limpieza y las condiciones ambientales adecuadas tendrán su impacto en la vida útil del producto.

Figura 10

Page 6: Flexichamber - Avadel Pharmaceuticals · clinics. This product is intended for patients who can follow verbal instructions. Instructions For Use 1.ve device from carton and outer

Instrucciones importantes de limpiezaLimpie la máscara facial Flexichamber®

a mano semanalmente, o según se requiera en dependencia del uso del producto. La máscara NO es apta para lavavajillas u horno microondas.1. Retire la máscara del dispositivo

Flexichamber®.2. Lave en agua tibia con detergente para

vajilla líquido de acción leve y conun paño suave. No use detergentesni paños abrasivos, ya que esto podríadañar la máscara facial (ver Figura 8).

3. Use agua tibia para enjuagarla máscara.

4. Para secar, agite suavemente paraeliminar el exceso de agua. Dejeque la máscara se seque al aire. Nolo coloque cerca de fuentes de calorartificiales, como lavavajillas u hornos.

5. Cuando esté completamente seca, almacene la máscara facialFlexichamber® con la cámara espaciadora plegable, antiestáticay con válvula Flexichamber®.

6. Almacene en un lugar frío y seco.

Precaución: Según la ley federal, este dispositivo solo puede ser vendido por un médico o por indicación de este.

PRESENTACIÓN:La máscara facial Flexichamber® se presenta en tres (3) tamaños:Máscara pequeña para niños NDC # 13551-903-01 Máscara grande para niños NDC # 13551-903-02 Máscara pequeña para adultos NDC # 13551-903-03

Máscara facial Flexichamber®

Para ser utilizada con la cámara espaciadora plegable, antiestática y con válvula Flexichamber®

Para un solo paciente, multiuso, desechable

Figura 1

Figura 2

Figura 3

Figura 4

Figura 5

Figura 6

Figura 7

Figura 8

AntecedentesAntes de usar la máscara facial Flexichamber® con la cámara espaciadora plegable, antiestática y con válvula Flexichamber®, es importante leer y comprender estas instrucciones, las instrucciones proporcionadas con la cámara espaciadora plegable, antiestática y con válvula Flexichamber® y las instrucciones provistas con su inhalador de dosis medidas (IDM). Se debe usar según las indicaciones de su profesional sanitario.

Indicaciones de usoLa máscara facial Flexichamber® está diseñada para usarse con la cámara espaciadora plegable, antiestática y con válvula Flexichamber® utilizada por pacientes que se encuentran en tratamiento o a cargo de un médico o un profesional sanitario matriculado. La máscara facial Flexichamber® y la cámara espaciadora plegable, antiestática y con válvula Flexichamber®

están diseñadas para ser usadas por estos pacientes con el propósito de administrar medicamento en aerosol desde la mayoría de los inhaladores de dosis medidas presurizadas. El dispositivo está diseñado para ser utilizado en entornos tales como hogares, hospitales y clínicas. Este producto es para pacientes que pueden seguir indicaciones verbales.

IntroducciónPara la mayoría de los niños, el IDM proporciona el medicamento de la manera más eficaz cuando se administra a través de una cámara espaciadora con válvula. Una cámara espaciadora con válvula contiene una válvula de una sola vía, la cual no permite que el paciente vuelva a respirar dentro de la cámara. Esta característica es muy importante porque permite que los niños pequeños respiren normalmente y que aun así obtengan una dosis confiable de medicamento. La máscara que se adjunta a la cámara se adapta a la nariz y a la boca del niño y forma un sello. El sello no permite que el medicamento se escape al aire.

Instrucciones de uso1. Retire la máscara facial Flexichamber®

del envase de cartón y del plásticoexterno que la recubre (ver Figura 1).

2. Fije la máscara facial Flexichamber®

al extremo de la cámara espaciadoraplegable, antiestática y con válvulaFlexichamber® (ver Figura 2).

3. Coloque la máscara sobre la bocay la nariz para asegurar un buen sello(ver Figura 3).

4. Presione hacia abajo sobre el IDMpara enviar una dosis de medicamentoa la bolsa de depósito. Administre solouna (1) descarga por vez (ver Figura 4).

5. Inhale lenta y profundamente a travésde la nariz y la boca hasta que hayarealizado una inspiración completa.La bolsa de depósito podría retraersehacia adentro levemente a medidaque usted inhala (ver Figura 5).Si es posible, aguante la respiraciónentre 5 y 10 segundos mientrasmantiene un sello con la máscara.

6. Repita los pasos 4 y 5, siguiendolas instrucciones provistas conel IDM, la cámara espaciadoraplegable, antiestática y con válvulaFlexichamber® y según lo indiquesu médico, lo que incluye la cantidadde tiempo que debe esperar entrelas dosis.

7. Nota importante: Antes de repetirlos pasos 4 y 5 asegúrese de que la

Cuando termine de tomar el medicamento según lo recetado por su médico, saque el inhalador de dosis medidas (IDM) de la pieza posterior y cierre el dispositivo con cuidado girando la boquilla y ejerciendo presión hacia la pieza posterior, sin permitir que la bolsa del depósito se aloje entre la pieza posterior y la boquilla (Ver Figura 7). Vuelva a colocar la tapa.

bolsa de depósito haya recuperado su tamaño completo (es decir, que ya no tenga una forma desinflada o colapsada) retirando suavemente la boquilla de la pieza posterior (ver Figura 6).

Distribuido por: Avadel Pharmaceuticals (USA), Inc., Chesterfield, MO 63005www.avadel.comRev 11/161116 FLX MSK IFU WEB SP