Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
CTU IN PRAGUE ATLAS SCT
Flow-meter calibrations for the ID
of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: R. Marek, M. Doubek and M. Vítek
31.10.2009
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 2
Abstract This report presents results obtained during measurement of detailed characteristics of the
thermal mass flow-meter Bronkhorst. The initial impulse for this measurement came from the ID
community of ATLAS experiment as a preparatory work for the thermal characteristic
measurement of the detector components where the Bronkhorst thermal mass flow-meter is
employed for the C3F8 vapor mass flow determination. The calibration took place at Prévessin
(H8) where the C3F8 experimental cooling circuit has been built by the department of Physics
from the Czech Technical University in Prague. Within the scope of this measurement several
other flow-meters were either calibrated or used as the reference and some of them will be used
in other projects at CERN.
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 3
Content Abstract ........................................................................................................................................... 2
Introduction ..................................................................................................................................... 4 Referential flow-meter .................................................................................................................... 4 Investigated and calibrated flow-meters ......................................................................................... 5 Calibration setup ............................................................................................................................. 6 Calibration of the Bronkhorst thermal mass flow-meter ................................................................ 7
Summary ....................................................................................................................................... 11 List of stable conditions for Bronkhorst thermal mass flow-meter .............................................. 12 Calibration sheet for Bronkhorst flow-meter ................................................................................ 14
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 4
Introduction
Calibration of vapor flow-meters inside the two-phase cooling circuit has many advantages.
The most important one is the possibility to use a referential flow-meter for liquid phase which is
generally more precise than vapor flow-meters. The cooling circuit has been prepared by the
Department of Physics from the Czech Technical University in Prague and modified for the
purpose of flow-meter calibration with mass flow capacity currently standing at 5 g/s of C3F8
vapor.
The flow-meter calibration is a rather time consuming procedure due to the necessity to
achieve steady and balanced flow pattern. This fact enforced us to combine several other flow-
meters to use the set-up and time efficiently.
The cooling circuit is equipped with ELMB-based DAQ system using PVSS software. The
calibration of the DAQ system has been checked with Keithley precision multimeter.
Figure 1: Cooling circuit located in Bldg. 887 H8 in Prévessin
Referential flow-meter
All calibrations have been conducted and compared with CORI-flow
type mass flow-meter by Bronkhorst. Its main advantages are high
accuracy, fast response time and mainly the direct mass flow
measurement. The mass flow is calculated from the phase shifting of
two conversely vibrating tube loops. The phase shift is increasing with
higher mass flow. The calibration fit is provided by the manufacturer.
Specification:
Mass flow range: 0…92 ln/min; (0…13.9 g/s)
Temperature range: 0…70 °C
Analog output: 0…5 V
Supply voltage: +15…24 V
A vapor phase volume flow-meter by Schlumberger served as the
second verification of the mass flow. The flow-meter is located at compressor suction line and is
coupled with pressure and temperature sensors for the adequate mass flow calculation.
Figure 2: CORI-Flow
mass flow-meter
IST thermal mass flow sensors
on vapor line
Bronkhorst flow-meter
Compressor station IST reading electronics Distribution manifold
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 5
Investigated and calibrated flow-meters The EL-FLOW type sensor by Bronkhorst is a
thermal mass flow-meter specially designed for
mass flow measurement of C3F8 vapor. High
precision, pressure and temperature independency
plus a short reaction time belong to its main
advantages.
Specification:
Mass flow range: 0…100 Nl/min; (0…15 g/s)
Analog output: 0…5 V
Supply voltage: +15…24 V
Figure 3: Bronkhorst thermal mass flow meter
Sensor FS5L (liquid type) by Innovative Sensor
Technology (IST) is fixed in a brass thick-walled
tube (Figure 4.) coupled with special readout
electronics supplied by the IST. It is a very
promising and cheap thermal flow-meter. It
determines mass flow by measuring the resistance of
a RTD (temperature variable resistor) element
heated by an electrical current. An additional
temperature sensor is present on the chip for
compensating the temperature of the liquid.
Figure 4: IST FS5L
Swissflow SF800 turbine volume flow-meters use
an IR sensor to measure revolutions of a
lightweight turbine subjected to liquid flow. Its
output signal is of a rectangular shape. We have
coupled all three of them with
frequency/voltage transducers and calibrated them
in conjunction.
Figure 5: Swissflow SF800 flow meters with two types
of frequency transducers
Sensors IST FS5 for vapor are basically of the
same principle as the FS5L type. The only
difference is the adjustment of the sensor itself for
the low density vapor flow. The sensor requires also
differently tuned readout electronics.
Figure 6: IST FS5 in a Swagelok Tee
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 6
Calibration setup By manipulating the temperature of water circuit supplied by chiller that first cools down the
refrigerant entering the CORI-Flow and warms up vapor in heat exchanger later and thanks to
the power change of the heater located on vapor line, the temperature of the vapor may be easily
changed between 15 to 40 °C.
Simplified schematic of the circuit modification during the calibration runs can be seen on
Figure 7. Except from SF800 sensors all flow-meters were calibrated at the same time.
Swissflow sensors were located on three parallel liquid lines. These have had to be opened one at
a time when calibration run took place.
AC
Condenser
AC
3 separated lines Swissflow
flowmeters
CapillariesWater HEX
Heater
IST 8 mm
IST 8 mm
IST 10 mmTwo-stage
compressor unit
CORI
referential
flowmeter
Vapor Liquid
IST liquid
Bronkhorst
thermal flowmeter
Schlumberger
Figure 7: Simplified cooling circuit scheme with all flow-meters
Readings of the mass flow from the flow-meters CORI-Flow and Schlumberger differ in
almost constant ratio 1:0.874. Schlumberger gives a lower flow rate. Therefore Schlumberger
served as a second reference sensor in cases with high deviation of the measured points from the
final fit. CORI / Schlumberger
y = 0.8743x
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5
CORI [g/s]
Sch
lum
berg
er
[g/s
]
Figure 8: Comparison of the mass flow indicated by Schlumberger and CORI
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 7
Calibration of the Bronkhorst thermal mass flow-meter
The measurement took place between the 14th
to 18th
of October 2009 with some additional
measurements on the 21st, 22
nd, 23
rd and on the 28
th of October 2009.
Out of thirty thousand collected data points 82 fully stable sections of working points have
been evaluated in order to cover all thinkable dependencies as well as their tendencies. Vibration
from compressor station has been eliminated by the solid anti-vibration insulation of the
referential CORI flow-meter as well as the Bronkhorst thermal flow-meter.
Data have been recorded in maximal possible frequency. Their behavior has been deeply
observed and all intervals of representative stable values have been carefully chosen. Time
required for the stabilization of the flow pattern within the limit spans from several minutes to
half an hour depending on the nature of the changes in mass flow setting. Temperature
instabilities usually consumed the most of time to achieve a steady flow condition.
Figure 9 shows all the measured and evaluated working points of the Bronkhorst flow-meter
with mass flow of up to 5.0 g/s, which is a current limit of the cooling circuit.
Simple characteristics of the Bronkhorst thermal flowmeter
All measured data
dm /dt = 2.9612U [g/s; V]
0.0
1.0
2.0
3.0
4.0
5.0
6.0
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0
Flowmeter signal voltage U [V]
C3F
8 m
assf
low
dm
/dt
[g/s
]
Figure 9: All measured data in a linear characteristic
Other objective of the calibration was to study the temperature dependency of the Bronkhorst
sensor, if any. By exploring the mass flow difference between a temperature independent fit (see
Fig. 9) and evaluated data points, the temperature dependency is evident (see Fig. 10). We
assume that this dependency has a linear tendency in relation to temperature. For a range
between 10 and 35 °C its convexity is barely visible. It is well within the data dispersion. With
increased temperature the real mass flow is lower than what the Bronkhorst sensor indicates.
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 8
Massflow difference to previous fit according to temperature
All measured data
-0.3
-0.2
-0.1
0.0
0.1
0.2
0.3
0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0
Vapor temperature T [°C]
C3F
8 m
as
sfl
ow
err
or
dm
e/d
t [g
/s]
Figure 10: Mass flow difference to simple linear fit.
Figure 11 shows mass flow difference between evaluated data points and the final suggested
fit. The data are classified according to the single property (vapor pressure, supply voltage, etc.)
by which they differ from our nominal working parameters – low vapor pressure and 16.2 V of
supply voltage.
Working points with higher than nominal vapor pressure are presented also on Fig. 12. It is
safe to say that there is no visible dependency on vapor pressure in the range of (1…2.5) bara for
the measured mass flow interval of (0…5) g/s.
Figure 13 presents mass flow error reflecting temperature of the liquid at the referential CORI
flow-meter. All data points are within the data dispersion except several data points at low
temperature. We have no satisfactory explanation for the sudden decrease of the signal.
Exceeding the dew point temperature might have something to do with it. However no further
investigation of this phenomenon has been conducted. Those points have been excluded from
calculation of the final fit.
Supply voltage dependency has not been proven. The results may indicate a very slight rising
tendency but within the data dispersion (see Fig. 14). Only four working points with elevated
supply voltage have been measured. For the best results the use of 16 V supply voltage is
recommended.
These values have been omitted from
the final calculation (vide infra).
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 9
Massflow difference to the final fit for classified data
-0.20
-0.15
-0.10
-0.05
0.00
0.05
0.10
0.15
0.20
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0
Flowmeter signal voltage U [V]
C3F
8 m
as
sfl
ow
dm
/dt [
g/s
]
Nominal High vapor pressure High supply voltage Cold liquid at CORI
Fig.ure 11: Difference of mass flow to the final fit
Massflow difference to the final fit according to vapor pressure
-0.20
-0.15
-0.10
-0.05
0.00
0.05
0.10
0.15
0.20
0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6
Vapor pressure p [bara]
C3F
8 m
as
sfl
ow
dm
/dt [
g/s
]
Fig.ure 12: Mass flow difference to the final fit according to vapor pressure
These working points have been
excluded from the calculation.
These working points have been
excluded from the calculation.
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 10
Massflow difference to the final fit according to liquid temperature
-0.20
-0.15
-0.10
-0.05
0.00
0.05
0.10
0.15
0.20
10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 20.0
Liquid temperatore T [°C]
C3F
8 m
as
sfl
ow
dm
/dt [
g/s
]
Figure 13: Mass flow difference to the final fit according to liquid temperature at CORI referential flow-
meter
Massflow difference to the final fit according to supply voltage
-0.20
-0.15
-0.10
-0.05
0.00
0.05
0.10
0.15
0.20
15.0 16.0 17.0 18.0 19.0 20.0 21.0 22.0 23.0 24.0 25.0
Flowmeter supply voltage U [V]
C3F
8 m
as
sfl
ow
dm
/dt [
g/s
]
Fig.ure 14: Mass flow difference to the final fit according to supply voltage
These working points have been
excluded from the calculation.
These working points have been
excluded from the calculation.
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 11
Calibration of Swissflow SF800 liquid flow-meters coupled with frequency
transducers
Swissflow calibration
dm /dt = 1.6113U + 0.5512 [g/s; V]
dm /dt = 1.6811U + 0.737 [g/s; V]
dm /dt = 2.0471U + 0.1645 [g/s; V]
1.2
1.25
1.3
1.35
1.4
1.45
1.5
1.55
1.6
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8
Transducer's signal voltage [V]
Ma
ss
flo
w [
g/s
]
Swissflow 1 Swissflow 2 Swissflow 3
Linear (Swissflow 1) Linear (Swissflow 2) Linear (Swissflow 3)
Figure 15: Swissflow SF800 calibration fits
Summary
The Bronkhorst thermal flow-meter behaved as expected. The temperature dependency causes
rising of the output signal by 0.34% per 1 °C in increased temperature. For mass flow of 10 g/s
and maximally expected difference in temperature of 10 °C the temperature dependency might
induce the error up to 0.34 g/s. The calibration fit includes a correction for temperatures different
from 17 °C that has been the most frequent value during experimental runs in the PIT.
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 12
List of stable conditions for Bronkhorst thermal mass flow-meter
Time
identification
Referential mass flow-meter Vapor
temperature
Bronkhorst
signal
Difference to
the final fit
Comment
[hh:mm:ss] CORI [g/s] Schlumberger [g/s] [°C] [V] [g/s]
- 0 0 - 0 0 No offset
12:06:00 4.01 3.49 19.1 1.332 0.043
12:43:10 4.73 4.13 18.8 1.577 0.034
13:09:18 4.49 3.93 19.0 1.499 0.033
13:39:20 4.38 3.84 19.1 1.462 0.035
14:03:03 3.81 3.3 19.4 1.269 0.037
14:19:23 3.70 3.24 19.5 1.232 0.039
14:42:20 2.40 2.13 20.0 0.798 0.035
15:03:52 2.57 2.27 20.1 0.854 0.038
15:25:05 2.96 2.6 20.0 0.985 0.040
16:00:07 1.68 1.5 20.5 0.559 0.030
16:18:57 1.52 1.36 20.5 0.503 0.030
16:37:52 1.43 1.27 20.6 0.473 0.029
16:59:10 1.28 1.13 20.6 0.424 0.029 2.11 bara
17:23:28 3.73 3.25 19.6 1.240 0.048 2.19 bara
17:38:45 3.99 3.4 19.5 1.328 0.043 2.20 bara
17:59:47 4.39 3.79 19.4 1.462 0.043 2.21 bara
18:37:45 4.46 3.88 18.2 1.483 0.038
18:56:07 4.49 3.92 28.5 1.559 0.005
12:15:35 4.78 4.11 9.9 1.559 -0.006
12:39:20 4.77 4.09 13.9 1.592 -0.051
13:05:00 4.76 4.13 20.4 1.621 -0.046
13:28:08 4.75 4.15 24.3 1.629 -0.013
13:56:00 4.73 4.16 28.9 1.640 0.018
14:06:43 4.73 - 29.4 1.633 0.043 23V supply
14:28:57 4.73 4.11 10.9 1.517 0.090 23V supply
14:52:57 4.85 4.18 12.5 1.598 -0.009
15:12:50 4.80 4.13 14.5 1.598 -0.033
15:28:52 4.78 4.11 16.8 1.608 -0.040
15:45:28 4.83 4.18 22.8 1.656 -0.041
15:59:57 4.86 4.24 22.6 1.656 -0.004
16:25:27 4.96 4.3 12.5 1.618 0.032
16:48:35 4.99 4.34 18.3 1.684 -0.036
17:11:52 4.61 4.01 19.9 1.559 -0.018
17:30:25 4.61 4.01 13.4 1.513 0.020
18:17:25 4.59 - 12.6 1.521 -0.037 2.27 bara
18:47:58 4.60 - 12.3 1.524 -0.041 2.27 bara
19:55:00 4.59 - 19.2 1.552 -0.023 2.27 bara
10:19:05 3.97 3.45 16.1 1.329 -0.021
10:48:37 4.39 3.83 15.8 1.469 -0.033
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 13
11:14:55 4.37 3.84 23.2 1.511 -0.060
11:36:55 4.00 3.46 16.2 1.330 0.003
11:53:25 3.98 3.44 20.8 1.360 -0.038
12:10:15 3.97 3.46 28.3 1.392 -0.040
12:26:52 3.67 3.22 24.8 1.255 0.013
12:43:20 3.69 3.23 16.8 1.216 0.041
13:04:28 2.37 2.13 18.1 0.798 -0.007
13:27:47 2.35 2.11 24.5 0.814 -0.023
13:45:10 2.34 2.11 34.0 0.841 -0.030
13:56:17 2.80 2.5 32.9 0.990 -0.003
14:14:53 2.81 2.5 21.4 0.941 0.033
14:33:02 2.81 2.5 18.0 0.935 0.016
14:51:57 2.35 2.1 18.5 0.790 0.000 1.62 bara
15:12:45 2.32 2.07 18.6 0.780 -0.003 2.52 bara
15:36:30 2.33 2.08 18.6 0.786 -0.016 1.85 bara
15:52:22 1.02 0.87 19.8 0.348 -0.015
16:19:05 1.16 1.04 25.1 0.406 -0.023
16:42:10 1.14 1.04 20.4 0.392 -0.020
14:43:20 4.47 3.93 15.2 1.498 -0.043
15:53:10 4.42 3.88 18.6 1.463 0.059
16:06:02 4.42 3.88 18.3 1.467 0.045 23V supply
16:17:57 4.41 3.88 18.2 1.471 0.024
16:44:07 4.49 3.97 23.2 1.529 0.000
17:03:22 4.46 - 21.8 1.504 0.024
17:23:35 2.86 2.6 23.7 0.991 -0.046 12.3 °C Liq.
17:47:57 2.71 2.62 19.2 0.978 -0.193 11.2 °C Liq.
18:09:02 2.32 2.22 20.2 0.835 -0.153 10.9 °C Liq.
18:20:05 2.35 - 20.2 0.849 -0.169 10.9 °C Liq.
18:36:25 2.44 - 20.0 0.893 -0.211 10.9 °C Liq.
10:06:27 4.13 3.63 15.8 1.380 -0.022 12.8 °C Liq.
10:36:30 5.03 - 32.9 1.782 -0.016
11:54:02 4.90 - 11.3 1.595 0.028
12:33:12 4.78 - 12.1 1.572 -0.010
13:26:58 3.84 - 21.0 1.290 0.023
13:44:47 3.90 - 16.6 1.290 0.028
15:10:00 2.64 2.4 20.3 0.906 -0.041 12.4 °C Liq.
16:00:50 2.78 - 20.1 0.958 -0.056 12.3 °C Liq.
16:28:27 2.85 - 18.9 0.976 -0.058 12.4 °C Liq.
16:48:53 2.85 - 23.3 0.994 -0.068 12.4 °C Liq.
17:21:05 4.56 - 20.4 1.540 0.002
17:40:22 4.15 - 27.3 1.440 -0.015
10:58:23 4.98 4.37 32.0 1.720 0.085
17:19:42 2.78 - 29.4 0.949 0.056 23V supply Table 1: List of stable conditions
Flow-meter calibrations for the ID of the ATLAS Experiment
Supervisors: Stephen McMahon and Vaclav Vacek Collaborators: M. Doubek, R. Marek and M. Vitek Page 14
Calibration sheet for Bronkhorst flow-meter M. Doubek, R. Marek, M. Vítek and V. Vacek
CTU Prague, 31.10.2009
F-113AC-AAD-99-V pmax = 100 bar
/min 0...100 g/s 15 ... 0 nlm
RS232 + analog (n/c control)
Analog output 0…5 V
Supply voltage +15…24 Vdc
Viton sealed
FC g 05.9C) 0 ;bar (1 l 1for C V; /min;l ; 170676.08693.19
C V; g/s; ; 170102.099695,2
)( ;
83ann
00
TUUm
TUUm
TfaTTkUaUm
Bronkhorst flow meter characteristics
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0
Signal voltage [V]
Ma
ss
flo
w [
g/s
]
0
10
20
30
40
50
60
70
80
90
100
Ma
ss
flo
w [
l n/m
in]
T = 10 °C T = 15 °C T = 17 °C T = 19 °C T = 25 °C
RS232 wiring used in calibration
1 Not connected
2 Analog output 0…5 V
3 Not connected
4 Ground
5 Not connected
6 Not connected
7 Supply voltage +15…24 Vdc
8 Ground
9 Shield
Bronkhorst thermal flow meter
Extrapolated data Calibration have been
conducted from 0 to 5.0 g/s
Calibration sheet for Bronkhorst flow-meter M. Doubek, R. Marek, M. Vítek and V. Vacek
CTU Prague, 31.10.2009
CTU IN PRAGUE ATLAS SCT
F-113AC-AAD-99-V pmax = 100 bar
/min 0...100 g/s 15 ... 0 nlm
RS232 + analog (n/c control)
Analog output 0…5 V
Supply voltage +15…24 Vdc
Viton sealed
FC g 05.9C) 0 ;bar (1 l 1for C V; /min;l ; 170676.08693.19
C V; g/s; ; 170102.099695,2
)( ;
83ann
00
TUUm
TUUm
TfaTTkUaUm
Bronkhorst flowmeter characteristics
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0
Signal voltage [V]
Ma
ss
flo
w [
g/s
]
0
10
20
30
40
50
60
70
80
90
100
Ma
ss
flo
w [
l n/m
in]
T = 10 °C T = 15 °C T = 17 °C T = 19 °C T = 25 °C
RS232 wiring used in calibration
1 Not connected
2 Analog output 0…5 V
3 Not connected
4 Ground
5 Not connected
6 Not connected
7 Supply voltage +15…24 Vdc
8 Ground
9 Shield
Bronkhorst thermal flow meter
Calibration have been
conducted from 0 to 5.0 g/s Extrapolated data