Focal plane of lens Focal plane of lens full field microscope, magnification defined by image distance/object distance scanning microscope, magnification defined by display size / scan size object objec t In epi-illuminated light microscopy, , all arrows are double headed. image object image detector

Focal plane of lens

  • Upload
    cleave

  • View
    54

  • Download
    4

Embed Size (px)

DESCRIPTION

object. image. detector. full field microscope, magnification defined by image distance/object distance. scanning microscope, magnification defined by display size / scan size. image. Focal plane of lens. In epi-illuminated light microscopy, , all arrows are double headed. - PowerPoint PPT Presentation

Citation preview

Page 1: Focal plane of lens

Focal plane of lens

Focal plane of lens

full field microscope, magnification defined by image distance/object distance

scanning microscope, magnification defined by display size / scan size

object

object

In epi-illuminated light microscopy, , all arrows are double headed.

image

object

image

detector

Page 2: Focal plane of lens

Probe formation and signal collection in light microscopy (case study using Marshall U. MRC 1024 CSLM)

In conventional microscopy the image point intensities of tiny objects are defined primarily by the collection pathway, in CSLM the illumination and detection pathways are equally important to defining this spot that is imaged. Both pathways travel through the objective lens and the confocal scanhead. X Y

Z not shown (height is intensity)

laser in, signal from sample out

sample

Page 3: Focal plane of lens

Where do the excitation and emission filters go? Where is the confocal iris?

Page 4: Focal plane of lens

Illumination and detection path in MRC1024 confocal

gas, KrAr laser

optical filters (Neutral Density and excitation λ)

sample

scan mirrors (galvo driven x & y)

objective lens

optical fiber to carry laser to scanhead

dichroic mirror (beam splitter) split excitation from emission ) dichroic mirror

objective lens

scan mirrors

EX EM

emission λ filter

confocal iris

PMT detector

sample

Page 5: Focal plane of lens

488nm 568nm 647nm

Low pressure gasses emit discreet lines of color as in this low pressure Hg spectrum (grey) overlaid with 3 of the major lines from a KrAr mixed gas laser, also low pressure (color).

Different light sources have different spectra

This is just one of the Hg lines at much higher spectral resolution.

Page 6: Focal plane of lens

High pressure gasses in arc lamps emit relatively broad spectrum light (many more vibrational energy states than low pressure). Black body radiation as emitted from the tungsten filament in halogen/tungsten or standard tungsten filament sources varies with filament temperature and are UV and blue poor as well as being inefficient (heat).

Metal halide are very similar to gas arc but include metal iodides and bromides in the gas mix. Longer life and tunable spectrum (based on the metal used).

gas

gas

black body

black body

gas

All adapted from Murphy 2001

Page 7: Focal plane of lens

Neutral Density Filtering (all λ are affected equally)

Beer / Lambert relationship

Transmittance: T = P / P0% Transmittance: %T = 100 T

A=-log TA = log10 P0 / PA = log10 1 / %T

Absorbance (OD):

Lets say this filter blocks 70% of the incident light so its T=.30

So, this filter has an A=.125

What happens if we double the width of the filter? T= ? A=?

What is the A if the filter blocks out 99.7% of the incident light?

A = ε (pathlength) concentration ε depends on the filter material, pathlength is thickness of filter, concentration is the conc. of material with ε in the filter

P0

P

Page 8: Focal plane of lens

Short and long pass filtering with colored (colloidal) glass filters; these do not affect all wavelengths equally

515 LP or

OG 515

A<515=-log T<515

Absorbance <515nm

Page 9: Focal plane of lens

Filtering with interference filters (most but not all band pass filters are interference filters)

All of our confocal band pass filters and dichroic mirrors are interference filters, see website below or Slayter & Slayter 4.3. Also see the optical thin film example on course website.

http://hyperphysics.phy-astr.gsu.edu/hbase/hph.html

These filters are given absorbance and OD values but they DO NOT follow the Beer Lambert relationship with thickness and concentration.

Page 10: Focal plane of lens

Purdue Univ.

Detector

Emission light (pink color) follows the same path as the excitation light (red) until we reach the dichroic mirror (DM). This DM prevents reflected laser signal from causing decreased contrast of our fluorescence signal.

DM

Laser

Signal in

Signal out

sample on microscope

Page 11: Focal plane of lens

Murphy 2001These objective lenses are corrected for chromatic and spherical aberrations and curvature of field. Apochromats are corrected at more wavelengths (colors) than achromats.

Many lens elements with complex shapes are needed to bend and fold the light so that rays of different colors and those closer or further from the optical axis are focused to the same place. When the objective lens is perfectly designed and built, only then do we achieve ‘diffraction limited resolution’ as defined by the next slide.

Page 12: Focal plane of lens

Gaussian image points vs. Abbe’s theory of image formation; Gaussian ray tracing brings rays to an infinitely small and unacheivable point in space (what is dmin).

Abbe’s theory of image formation states that a small interference pattern is formed by the perfect lens, this is the diffraction limited resolution. If you can develop

technology that elegantly overcomes this barrier, you could win the Nobel prize.

Gaussian theory works well for image formation of features larger than the dmin.

Abbe’s theory works for describing features that are approximately the size of dmin or ~ the wavelength of light (or electrons) used.

dmin = 1.22(λ) / NAcond + NAobj

dmin = 1.22(λ) / 2 sin α η

dmin = .61(λ) / NAobj (for epi- microscopy)

NA (numericalaperture) = sin α η

Page 13: Focal plane of lens

What does the confocal iris or pinhole do?

An interference pattern like this is projected at the PMT for each object/ image point as the beam scans the sample. The confocal iris or pinhole selects only part of this Airy pattern to reach the PMT, this can improve resolution. Its effect on the z-axis diffraction pattern (not shown) is even more dramatic. This is why with confocal we can take optical z-sections.

Murphy 2001

This is a 2D view of a 3D phenomenon.

Iris and diaphragm

Airy pattern with central disk.

Page 14: Focal plane of lens

From Pawley, 1995

Z

X

Y

Page 15: Focal plane of lens

Total voltage drop in a PMT may equal thousands of volts.

Photocathode, usually coated, must be sensitive to Einstein’s photoelectric effect.

This PMT detector gives no spatial information on its own, only gives counts. The eye does give spatial information directly. It has an array of detectors (neurons). A ccd or film camera also directly record spatial information.

http://www.olympusmicro.com/primer/flash/photomultiplier/index.html

Eye is most sensitive to green light, PMTs can have photocathodes made of materials sensitive to specific colors (Murphy spectra p.25 )

Adapted from university of British Columbia Physics

(- - -)

(-)

(- -)

(+)

(+++)

(++++)

(+++++)

(++)

emission filter

Yes, these are secondary e- like in SEM.

Page 16: Focal plane of lens

Adapted from university of British Columbia Physics

(- - -)

(-)

(- -)

(+)

(+++)

(++++)

(+++++)

(++)

emission filter

Now lets include system noise into the detector. Lets assume that we get the equivalent of 1 photon/second of noise due to a light leak (this could also be electronic noise). Lets also assume that we get 16 photons/sec in real signal. Our PMT converts each photon into 10 electronic counts (gain of 10).

Given the above information, what is the signal to noise (S/N) ratio (in electronic counts) per pixel in two cases; 1-scan rate= 1 pixel/second 2- scan rate = 1 pixel/10seconds? This can also be called signal/background ratio.

One step further, all detector systems have some inherent ability to detect differences in intensity. It is not the signal /noise ratio that is important here, it is the absolute difference in signal intensity between 2 sample points.

Try this with this 4x4 pixel, 8 bit image. Which pairs of pixels can be differentiated from each other?

Each pixel pair has the same signal / signal ratio but the absolute differences vary.

When sampling with more pixels, adjust scan rate to increase photons/pixel; this is a good idea unless you are worried about beam damage! The top row of 3 scans (not in box) was done at constant scan rate (what is the scan rate for the top row?) The photon flux from the sample in all cases is 16 photons/second.

fast (? sec/scan)

slow (4 sec/scan) Slowest (? sec/scan)

40 80

2010

4 8

21

Page 17: Focal plane of lens