1
35 cmH 2 O with a PEEP of 5 cmH 2 O are considered is 210 ml, while 300 ml is set on the ventilator [2]. 2. The second issue, which may possibly explain the results reported by Zanetta et al. is related to the technical limita- tion of the Venturi inspiratory flow sys- tems. With these systems, inspiratory flow may be generated either at 100% or at 60% FIO 2 . Nevertheless, at 60% FIO 2 , the pressurisation gas source may be insufficient to overcome a high re- sistive load, subsequently resulting in an increased FIO 2 and a reduction in the flow delivered [3]. This phenome- non is particularly marked when the set inspiratory flow is low. Moreover, this is interestingly resolved by switching FIO 2 from 60 to 100% FIO 2 . Figure 1 illustrates the switch from a FIO 2 of 100% to 60%, with a Vt set at 300 ml and a given resistance of 50 cmH 2 O.s/l. In this experiment, the fall in peak pres- sure clearly suggests the weakness of the Venturi flow delivery system. Implication of these findings in clinical practice Although the ability of transport ventilators to deliver set Vt may be a relevant issue in patients exhibiting high resistive load, in fact PEEP and FIO 2 are at least as impor- tant to guarantee oxygenation in the major- ity of critically ill patients. For the purpose of limiting volume loss, increasing the set Vt according to the expected compressed Vt, as well as the use of 100% FIO 2 may be safely recommended in patients com- bining low inspiratory Vt and high resis- tive load. Whether a sophisticated ICU ventilator, that was not specifically designed for transport, should be used compared to a simpler transport ventilator remains con- troversial and warrants further investiga- tion. References 1. Zanetta G, Robert D, Guérin C (2002) Evaluation of ventilators used during transport of ICU patients – a bench study. Intensive Care Med 28:443–451 2. McGough EK, Banner MJ, Melker RJ (1992) Variations in tidal volume with portable transport ventilators. Respir Care 37:233–239 3. Kacmarek RM, Hess DR (1993) Airway pressure, flow and volume waveforms, and lungs mechanics during mechanical ventilation. In: RM Kacmarek, DR Hess, JK Stoller (eds) Monitoring in res- piratory care. Mosby, Saint-Louis, pp 497–543 L. Breton · G. Minaret · J. Aboab J.-C. Richard ( ) Medical Intensive Care Unit, Rouen University Hospital, 1, Rue de Germont, 76000 Rouen, France e-mail: [email protected] Tel.: +33-232-888261 Fax: +33-232-888314 Intensive Care Med (2002) 28:1181 DOI 10.1007/s00134-002-1390-7 CORRESPONDENCE Lucie Breton Guillaume Minaret Jérôme Aboab Jean-Christophe Richard Fractional inspired oxygen on transport ventilators: an important determinant of volume delivery during assist control ventilation with high resistive load Received: 15 May 2002 Accepted: 30 May 2002 Published online: 5 July 2002 © Springer-Verlag 2002 Sir: Zanetta et al. reported an interesting bench test study dealing with the specific problem of ventilators designed for trans- port [1]. The authors showed that the per- formance of transport and intensive care (ICU) ventilators, set in a volume-con- trolled mode and submitted to various combinations of resistive and elastic loads, were very inhomogeneous. These findings are of high clinical relevance, since the loss of tidal volume (Vt) they observed reached 30% of the set Vt when high resis- tance was considered. As discussed by the authors, two main mechanisms may explain, at least in part, these findings: (1) gas compression in the ventilatory circuit; (2) technical limits of the Venturi flow delivery systems. A better understanding of the respective role played by these two features may permit the user to limit significantly the clinical impact of problems related to Vt delivery. 1. Except for the Horus (Taema, Antony, France), no ventilator investigated in this study compensated for gas com- pression. This might explain the excel- lent performance exhibited by the Horus, whatever the condition tested, com- pared with both portable and other ICU ventilators. Nevertheless, increasing the set Vt according to a theoretical amount of gas compressed in the circuit may ef- ficiently prevent this problem. In fact, when the compliance of the circuit is known, the following equation allows for the calculation of the volume loss [C circuit *(P peak –PEEP), where C circuit is circuit compliance, P peak peak pressure and PEEP positive end-expiratory pres- sure]. For instance, the volume effec- tively inflated into the lungs when a C circuit of 3 ml/cmH 2 O and a P peak at Fig. 1 Inspiratory flow and airway pressure traces for a tidal volume (Vt) set at 300 ml on a ventilator (Osiris 1, Taema, Antony, France) and an imposed resistance (R) at 50 cmH 2 O.s/l during the FIO 2 switch from position 100% (O 2 ) to 60% (Air-O 2 ). The change in FIO 2 resulted in the peak pressure fall, due to the reduction in flow delivery. In this example, the volume reduction reached approximately 30%

Fractional inspired oxygen on transport ventilators: an important determinant of volume delivery during assist control ventilation with high resistive load

Embed Size (px)

Citation preview

35 cmH2O with a PEEP of 5 cmH2Oare considered is 210 ml, while 300 mlis set on the ventilator [2].

2. The second issue, which may possiblyexplain the results reported by Zanettaet al. is related to the technical limita-tion of the Venturi inspiratory flow sys-tems. With these systems, inspiratoryflow may be generated either at 100%or at 60% FIO2. Nevertheless, at 60%FIO2, the pressurisation gas source maybe insufficient to overcome a high re-sistive load, subsequently resulting inan increased FIO2 and a reduction inthe flow delivered [3]. This phenome-non is particularly marked when the setinspiratory flow is low. Moreover, thisis interestingly resolved by switchingFIO2 from 60 to 100% FIO2. Figure 1illustrates the switch from a FIO2 of100% to 60%, with a Vt set at 300 mland a given resistance of 50 cmH2O.s/l.In this experiment, the fall in peak pres-sure clearly suggests the weakness ofthe Venturi flow delivery system.

Implication of these findingsin clinical practice

Although the ability of transport ventilatorsto deliver set Vt may be a relevant issue inpatients exhibiting high resistive load, infact PEEP and FIO2 are at least as impor-tant to guarantee oxygenation in the major-ity of critically ill patients. For the purposeof limiting volume loss, increasing the setVt according to the expected compressedVt, as well as the use of 100% FIO2 may

be safely recommended in patients com-bining low inspiratory Vt and high resis-tive load.

Whether a sophisticated ICU ventilator,that was not specifically designed fortransport, should be used compared to asimpler transport ventilator remains con-troversial and warrants further investiga-tion.

References

1. Zanetta G, Robert D, Guérin C (2002)Evaluation of ventilators used duringtransport of ICU patients – a benchstudy. Intensive Care Med 28:443–451

2. McGough EK, Banner MJ, Melker RJ(1992) Variations in tidal volume withportable transport ventilators. RespirCare 37:233–239

3. Kacmarek RM, Hess DR (1993) Airwaypressure, flow and volume waveforms,and lungs mechanics during mechanicalventilation. In: RM Kacmarek, DRHess, JK Stoller (eds) Monitoring in res-piratory care. Mosby, Saint-Louis,pp 497–543

L. Breton · G. Minaret · J. AboabJ.-C. Richard (✉ )Medical Intensive Care Unit, Rouen University Hospital, 1, Rue de Germont, 76000 Rouen, Francee-mail: [email protected].: +33-232-888261Fax: +33-232-888314

Intensive Care Med (2002) 28:1181DOI 10.1007/s00134-002-1390-7 C O R R E S P O N D E N C E

Lucie BretonGuillaume MinaretJérôme AboabJean-Christophe Richard

Fractional inspired oxygenon transport ventilators: an important determinantof volume delivery during assistcontrol ventilation with high resistive load

Received: 15 May 2002Accepted: 30 May 2002Published online: 5 July 2002© Springer-Verlag 2002

Sir: Zanetta et al. reported an interestingbench test study dealing with the specificproblem of ventilators designed for trans-port [1]. The authors showed that the per-formance of transport and intensive care(ICU) ventilators, set in a volume-con-trolled mode and submitted to variouscombinations of resistive and elastic loads,were very inhomogeneous. These findingsare of high clinical relevance, since theloss of tidal volume (Vt) they observedreached 30% of the set Vt when high resis-tance was considered.

As discussed by the authors, two mainmechanisms may explain, at least in part,these findings: (1) gas compression in theventilatory circuit; (2) technical limits ofthe Venturi flow delivery systems. A betterunderstanding of the respective role playedby these two features may permit the userto limit significantly the clinical impact ofproblems related to Vt delivery.

1. Except for the Horus (Taema, Antony,France), no ventilator investigated inthis study compensated for gas com-pression. This might explain the excel-lent performance exhibited by the Horus,whatever the condition tested, com-pared with both portable and other ICUventilators. Nevertheless, increasing theset Vt according to a theoretical amountof gas compressed in the circuit may ef-ficiently prevent this problem. In fact,when the compliance of the circuit isknown, the following equation allowsfor the calculation of the volume loss[Ccircuit*(Ppeak–PEEP), where Ccircuit iscircuit compliance, Ppeak peak pressureand PEEP positive end-expiratory pres-sure]. For instance, the volume effec-tively inflated into the lungs when aCcircuit of 3 ml/cmH2O and a Ppeak at

Fig. 1 Inspiratory flow and airway pressure traces for a tidal volume (Vt) set at 300 ml ona ventilator (Osiris 1, Taema, Antony, France) and an imposed resistance (R) at50 cmH2O.s/l during the FIO2 switch from position 100% (O2) to 60% (Air-O2). Thechange in FIO2 resulted in the peak pressure fall, due to the reduction in flow delivery. Inthis example, the volume reduction reached approximately 30%