ftir july2010

Embed Size (px)

Citation preview

  • 7/28/2019 ftir july2010

    1/76

    Molecular absorption

    spectroscopy

    FTIR

  • 7/28/2019 ftir july2010

    2/76

    IR spectral regions

    REGION WAVELENGTHS

    (), m WAVENUMBERS( ), cm-1 FREQUENCY (), HzNear 0.78 to 2.5 12800 to 4000 3.8 x 1014 to 1.2 x 1014

    Middle 2.5 to 50 400 to 200 1.2 x 1014

    to 6.0 x 1012

    Far 50 to 1000 200 to 10 6.0 x 1014 to 2.0 x 1011

    Most

    used

    2.5 to 15 4000 to 670 1.3 x 1014 to 2.0 x 1013

  • 7/28/2019 ftir july2010

    3/76

    IR absorption spectrum of a

    polystyrene film

  • 7/28/2019 ftir july2010

    4/76

    IR spectra

    Frequency is seldom used as the abscissa

    Inconvenient size of unit

    1.3 x 1014

    to 2.0 x 1013

    Hz or s-1

    This axis is often referred to frequency

    The terminology is not correct

  • 7/28/2019 ftir july2010

    5/76

    IR spectra

    Plot of %T vs. cm-1

    Linear wavenumber scale is preferred

    Direct proportionality betweenwavenumber and both energy and

    frequency

    hcchhE

  • 7/28/2019 ftir july2010

    6/76

    Molecules

    Are composed of atoms held toghether bychemical bonds

    The atoms in a molecule are always

    moving or vibrating The intensities of vibrations increase when

    IR radiation is absorbed

  • 7/28/2019 ftir july2010

    7/76

    Each chemical bond requires a precise amount

    of energy to make it vibrate Each frequency of IR radiation provides a preciseamount

    Radiation is absorbed by a molecule only if the

    frequency of the radiation provides energy in theprecise amount required by one of the bonds in

    the molecule

  • 7/28/2019 ftir july2010

    8/76

    Molecules can be large or small

    The atoms they contain may be the same

    or different

    The bonds between them may vary from

    weak to small

  • 7/28/2019 ftir july2010

    9/76

    IR active

    To absorb IR radiation

    The molecule absorbs IR radiation that

    corresponds to the energy difference for

    the vibrational transition

    Energy is quantized

    A molecule must undergo a net change

    in dipole moment as it vibrates or

    rotates

  • 7/28/2019 ftir july2010

    10/76

    Dipole moment

    The charge distribution around a molecule

    is not symmetric because one of the atom

    has a higher electron density

    Example HCl

    Is determined by the magnitude of the

    charge difference and the distance

    between the two centers of charge

  • 7/28/2019 ftir july2010

    11/76

    DIPOLE MOMENTS

    Only bonds which have significant dipole moments willabsorb infrared radiation.

    Bonds which do not absorb infrared include:

    Symmetrically substituted alkenes and alkynes

    C C RR

    R

    R R

    R

    Many types of C-C Bonds

    Symmetric diatomic molecules

    H-H Cl-Cl

  • 7/28/2019 ftir july2010

    12/76

    C

    OThe carbonyl group is oneof the strongest absorbers

    O H C OAlso O-H and C-O bonds

    d+d-

    STRONG ABSORBERS

    + +

    - -

    C

    O

    C

    O

    d+d-

    oscillating dipoles couple and

    energy is transferred

    infrared beam

  • 7/28/2019 ftir july2010

    13/76

    Rotational transitions

    Small energy is required to cause a change in

    rotational level

    100 cm-1 or > 100m

    Rotational levels are quantized

    Absorption of gases in the FIR region is

    characterized by discrete, well defined lines

    In liquids and solids; intramolecular collisionsand interactions cause broadening of the lines

    into a continuum

  • 7/28/2019 ftir july2010

    14/76

    Vibrational-rotational

    transitions Vibrational energy levels are quantized

    E between quantum states correspond to MIR

    region for most molecules

    Spectrum of a gas

    A series of closely spaced lines

    There are several rotational energy levels for each

    vibrational level

    Spectra ofsolids and liquids

    Broad vibrational bands

    Rotation is restricted in solids and liquids

  • 7/28/2019 ftir july2010

    15/76

    Depending on the number of ways the

    bond can move (bend, stretch etc.), each

    type of bond may absorb IR radiation at

    one or more specific frequencies.

  • 7/28/2019 ftir july2010

    16/76

    Types of molecular vibrations

    Stretching

    A continuous change in the interatomic distance

    along the axis of the bands between the two atoms

    Symmetrical

    Asymmetrical

    Bending

    A change in the angle between two bonds

    4 types Scissoring

    Rocking

    Wagging

    Twisting

  • 7/28/2019 ftir july2010

    17/76

    Stretching vibrations

    Linear molecule

  • 7/28/2019 ftir july2010

    18/76

    Stretching vibrations

  • 7/28/2019 ftir july2010

    19/76

    Bending vibrations

  • 7/28/2019 ftir july2010

    20/76

    Hookes Law

  • 7/28/2019 ftir july2010

    21/76

    =1

    2pcK

    larger K,

    higher frequency

    larger atom masses,

    lower frequency

    constants

    2150 1650 1200

    C=C > C=C > C-C=

    C-H > C-C > C-O > C-Cl > C-Br

    3000 1200 1100 750 650

    increasing K

    increasing

    1m

    21

    21

    mm

    mm

  • 7/28/2019 ftir july2010

    22/76

    Stretching vibrations of CO2

  • 7/28/2019 ftir july2010

    23/76

    Stretching vibrations of CO2

  • 7/28/2019 ftir july2010

    24/76

  • 7/28/2019 ftir july2010

    25/76

    Bond order affects bond strength, so bond order

    affects the position of absorption bands

    The approximate wavenumber of an absorption can be

    calculated from Hookes law

    is the wavenumber of the stretching vibration

    f is the force constant

    m1 and m2 are the masses of the atoms

    2

    1

    21

    21 )(2

    1

    mmmmf

    cp

  • 7/28/2019 ftir july2010

    26/76

    The exact position of the absorption band

    depends on

    electron delocalization

    the electronic effect of neighboring substituents

    hydrogen bonding

    CH3CCH2CH2CH3

    O O O-

    C O C O

    at 1720 cm1 at 1680 cm1

  • 7/28/2019 ftir july2010

    27/76

    Putting an atom other than carbon next to the carbonyl group

    causes the position of the carbonyl absorption band to shift

    The predominant effect of the nitrogen of an amide is electron

    donation by resonance

    The predominant effect of the oxygen of an ester is inductive

    electron withdrawal

  • 7/28/2019 ftir july2010

    28/76

    The position of a CO absorption varies because

    CH3CH2 OH

    CH3CH2 O CH2CH3

    C

    O

    H3C OHC

    O-

    H3C OH

    C

    O

    H3C O CH3C

    O-

    H3C O CH3

    ~1050 cm1

    ~1050 cm1

    ~1250 cm1

    ~1250 cm1 and 1050 cm1

  • 7/28/2019 ftir july2010

    29/76

    FTIR spectrometer

    The main optical components are

    The IR source

    The interferometer The beamsplitter

    The laser

    The IR detector

  • 7/28/2019 ftir july2010

    30/76

    IR source

    The IR source produces IR radiation beam

    that travels through the spectrometer

    passing through the sample and to the

    detector

  • 7/28/2019 ftir july2010

    31/76

    Interferometer

    Beamsplitter

    Two mirrors

    fixed

    movable

  • 7/28/2019 ftir july2010

    32/76

    Beamsplitter

    Separates the IR beam into two beams of

    nearly equal energy

    One beam is reflected from the beam splitter

    to a fixed mirror

    The other beam is transmitted to a moving

    mirror and back to the beam splitter where

    the beams recombine

  • 7/28/2019 ftir july2010

    33/76

    Laser

    Used as an internal calibrator

    control the moving mirrors position signals the capture of data

  • 7/28/2019 ftir july2010

    34/76

    IR transducer

    After being absorbed at specific

    frequencies by the sample,

    the remainder of the IR beam is focused onto

    the detector

    The detector produces signal to the

    amount of IR radiation striking it and sends

    the signal to the computer for processing

  • 7/28/2019 ftir july2010

    35/76

    FTIR spectrometer

  • 7/28/2019 ftir july2010

    36/76

    FTIR spectrometer

  • 7/28/2019 ftir july2010

    37/76

    Emits IR radiation across the region of

    interest

    IR beam is deflected off a mirror

    Directs the IR beam into the interferogram

    Where the spectral encoding takes place

  • 7/28/2019 ftir july2010

    38/76

    Interferometer

    IR source

    Beamsplitter

    Moving

    mirror

    Fixed

    mirror

    Detector

    Sample

    He-Ne Laser

  • 7/28/2019 ftir july2010

    39/76

  • 7/28/2019 ftir july2010

    40/76

    The computer reads the interferogram and

    uses Fourier transformation to decode the

    intensity information for each frequency

    and presents a spectrum

  • 7/28/2019 ftir july2010

    41/76

    The spectrometer measures the intensity

    of a speciallyencoded IR beam after it has

    passed through the sample

    The resulting signal, interferogram

    Contains infromation about all frequencies

    present in the beam

  • 7/28/2019 ftir july2010

    42/76

  • 7/28/2019 ftir july2010

    43/76

    Laser

    Produces a single frequency of red light

    that follows the same path as the IR beam

    Calibrates the instrment internally

  • 7/28/2019 ftir july2010

    44/76

  • 7/28/2019 ftir july2010

    45/76

    FOURIER TRANSFORMATION

  • 7/28/2019 ftir july2010

    46/76

  • 7/28/2019 ftir july2010

    47/76

    Applications of IR spectrometry

    Major application (MIR region)

    Structural investigations of molecular

    compounds particularly organic

    compounds

    M j li ti f IR

  • 7/28/2019 ftir july2010

    48/76

    Major applications of IR

    spectrometry

  • 7/28/2019 ftir july2010

    49/76

    NIR and FIR regions

    NIR region (4000 to 14,000 cm-1)

    Quantitative determination of low MW

    hydrocarbons, H2O, CO2, S

    FIR region (15 to 100 m)

    Structural determination of inorganic and

    metal-organic species

  • 7/28/2019 ftir july2010

    50/76

    What produces the IR spectrum

    Materials that contain C (organic

    compounds)

    Many compounds; natural and synthetic

    inorganic materials

  • 7/28/2019 ftir july2010

    51/76

    Sample handling

    No solvent is transparent throughout the

    region of interest

    Sample handling

    Most difficult

    Time-consuming

    part of IR spectrometric analysis

  • 7/28/2019 ftir july2010

    52/76

    Gases

    Sample expands into an evacuated

    cylindrical cells with suitable windows (Fig

    16.3)

    Variable lengths

    A few cm to 10 m or longer

  • 7/28/2019 ftir july2010

    53/76

    Measuring gases

    Placed in a sealed cell

    Specified T and P can be maintained

    Gases are often present at low

    concentrations

    Mirrors are used to deflect the IR beam back

    and forth before the beam is allowed to exit

    Increased the interaction between gas sample andIR beam

    To acquire more spectral information

    This gas cell provides a fixed

    pathlength that is 2 m long

  • 7/28/2019 ftir july2010

    54/76

    Solutions

    Water and alcohols are not suitable

    solvents

    SOLVENTS FOR IR

  • 7/28/2019 ftir july2010

    55/76

    SOLVENTS FOR IR

    SPECTROMETRYNo solvent is transparent throughout the region of interest

  • 7/28/2019 ftir july2010

    56/76

    Liquids

    Pure (neat) liquid

    Very thin film, a short pathlength

    A drop of neat liquid is squeezed between

    two salt plates

    Held together by capillary action

    Placed in the beam of IR radiationNon reproducible transmittance data

    Satisfactory for qualitative analysis

  • 7/28/2019 ftir july2010

    57/76

    Measuring liquids

    Liquid cells Holds the liquid between two crystal made

    from materials that completely transmit IR

    radiation

  • 7/28/2019 ftir july2010

    58/76

    Cells

    Liquid cells

    (compared to UV/Vis spectroscopy)

    Narrow pathlengths

    0.01 to 1 mm

    Relatively high sample concentrations

    0.1% to 10%

    Short pathlengths

    Low molar absorbtivity

  • 7/28/2019 ftir july2010

    59/76

    Liquid cells

    Designed for easy assembly/disassembly

    Use Teflon spacers

    Allow variations in path lengths

    Fixed path-length cells

    Use a syringe to fill or empty

    Demountable IR cell for liquid

  • 7/28/2019 ftir july2010

    60/76

    Demountable IR cell for liquid

    sample

    C IR i d t i l

  • 7/28/2019 ftir july2010

    61/76

    Common IR window materials

  • 7/28/2019 ftir july2010

    62/76

    Selection of window materials

    Cost

    Range of transparency

    Solubility in solvent

    Reactivity with sample or solvent

    Most common NaCl and KBr

  • 7/28/2019 ftir july2010

    63/76

    Solids

    Mulls

    Pellets

  • 7/28/2019 ftir july2010

    64/76

    Solids

    Solid in a liquid or solid matrix

    Samples are ground to fine powder

    To avoid effects of scattered radiation

  • 7/28/2019 ftir july2010

    65/76

    Pelleting

    Most common technique; KBr pelleting

    Halide salts have a property ofcold flow

    Translucent property when pressure is

    applied to finely powdered salts

    ~1 mg (or less) sample is mixed with ~

    100 mg dried KBr

    Mix and grind in a mortar and pestle

    Apply 10,000 to 15,000 psi in a die

    Produce a transparent disk or pellet

  • 7/28/2019 ftir july2010

    66/76

    Pelleting

    Best done in vacuum or store the pellet in

    a dessicator before measurement

    Bands at 3450 and 1640 cm-1 due to

    absorbed moisture

  • 7/28/2019 ftir july2010

    67/76

    Mulls

    Solids

    Not soluble in an IR transparent solvent

    KBr pelleting is not convenient

    Grind ~ 2 5 mg fine powder in a drop of

    mulling agents

    Heavy hydrocarbon (Nujol)

    Fluorolube (halogenated polymer)

    Examine as thin film between two salt

    discs

  • 7/28/2019 ftir july2010

    68/76

    Qualitative analysis

    FTIR, NMR and MS

    Indentify

    Organic

    Inorganic species

    Biological

    Limitations to the use of

  • 7/28/2019 ftir july2010

    69/76

    Limitations to the use of

    correlation chartsTo determine the identity or structure Not possible using the correlation charts

    alone (only as a guide).

    Overlapping of group frequencies Instrumental limitations

    Spectral variations as a function of physical

    states Neat liquid

    Solution

    Pellet

    Mull

    Table of group frequencies for

  • 7/28/2019 ftir july2010

    70/76

    Table of group frequencies for

    organic functional groups

  • 7/28/2019 ftir july2010

    71/76

  • 7/28/2019 ftir july2010

    72/76

  • 7/28/2019 ftir july2010

    73/76

    Group frequency region of MIR

  • 7/28/2019 ftir july2010

    74/76

    Group frequency region of MIR

    Active or inactive vibrations in

  • 7/28/2019 ftir july2010

    75/76

    Active or inactive vibrations in

    IR spectrum?

  • 7/28/2019 ftir july2010

    76/76