Upload
rzi-danil-ishutin
View
233
Download
0
Embed Size (px)
Citation preview
7/28/2019 ftir july2010
1/76
Molecular absorption
spectroscopy
FTIR
7/28/2019 ftir july2010
2/76
IR spectral regions
REGION WAVELENGTHS
(), m WAVENUMBERS( ), cm-1 FREQUENCY (), HzNear 0.78 to 2.5 12800 to 4000 3.8 x 1014 to 1.2 x 1014
Middle 2.5 to 50 400 to 200 1.2 x 1014
to 6.0 x 1012
Far 50 to 1000 200 to 10 6.0 x 1014 to 2.0 x 1011
Most
used
2.5 to 15 4000 to 670 1.3 x 1014 to 2.0 x 1013
7/28/2019 ftir july2010
3/76
IR absorption spectrum of a
polystyrene film
7/28/2019 ftir july2010
4/76
IR spectra
Frequency is seldom used as the abscissa
Inconvenient size of unit
1.3 x 1014
to 2.0 x 1013
Hz or s-1
This axis is often referred to frequency
The terminology is not correct
7/28/2019 ftir july2010
5/76
IR spectra
Plot of %T vs. cm-1
Linear wavenumber scale is preferred
Direct proportionality betweenwavenumber and both energy and
frequency
hcchhE
7/28/2019 ftir july2010
6/76
Molecules
Are composed of atoms held toghether bychemical bonds
The atoms in a molecule are always
moving or vibrating The intensities of vibrations increase when
IR radiation is absorbed
7/28/2019 ftir july2010
7/76
Each chemical bond requires a precise amount
of energy to make it vibrate Each frequency of IR radiation provides a preciseamount
Radiation is absorbed by a molecule only if the
frequency of the radiation provides energy in theprecise amount required by one of the bonds in
the molecule
7/28/2019 ftir july2010
8/76
Molecules can be large or small
The atoms they contain may be the same
or different
The bonds between them may vary from
weak to small
7/28/2019 ftir july2010
9/76
IR active
To absorb IR radiation
The molecule absorbs IR radiation that
corresponds to the energy difference for
the vibrational transition
Energy is quantized
A molecule must undergo a net change
in dipole moment as it vibrates or
rotates
7/28/2019 ftir july2010
10/76
Dipole moment
The charge distribution around a molecule
is not symmetric because one of the atom
has a higher electron density
Example HCl
Is determined by the magnitude of the
charge difference and the distance
between the two centers of charge
7/28/2019 ftir july2010
11/76
DIPOLE MOMENTS
Only bonds which have significant dipole moments willabsorb infrared radiation.
Bonds which do not absorb infrared include:
Symmetrically substituted alkenes and alkynes
C C RR
R
R R
R
Many types of C-C Bonds
Symmetric diatomic molecules
H-H Cl-Cl
7/28/2019 ftir july2010
12/76
C
OThe carbonyl group is oneof the strongest absorbers
O H C OAlso O-H and C-O bonds
d+d-
STRONG ABSORBERS
+ +
- -
C
O
C
O
d+d-
oscillating dipoles couple and
energy is transferred
infrared beam
7/28/2019 ftir july2010
13/76
Rotational transitions
Small energy is required to cause a change in
rotational level
100 cm-1 or > 100m
Rotational levels are quantized
Absorption of gases in the FIR region is
characterized by discrete, well defined lines
In liquids and solids; intramolecular collisionsand interactions cause broadening of the lines
into a continuum
7/28/2019 ftir july2010
14/76
Vibrational-rotational
transitions Vibrational energy levels are quantized
E between quantum states correspond to MIR
region for most molecules
Spectrum of a gas
A series of closely spaced lines
There are several rotational energy levels for each
vibrational level
Spectra ofsolids and liquids
Broad vibrational bands
Rotation is restricted in solids and liquids
7/28/2019 ftir july2010
15/76
Depending on the number of ways the
bond can move (bend, stretch etc.), each
type of bond may absorb IR radiation at
one or more specific frequencies.
7/28/2019 ftir july2010
16/76
Types of molecular vibrations
Stretching
A continuous change in the interatomic distance
along the axis of the bands between the two atoms
Symmetrical
Asymmetrical
Bending
A change in the angle between two bonds
4 types Scissoring
Rocking
Wagging
Twisting
7/28/2019 ftir july2010
17/76
Stretching vibrations
Linear molecule
7/28/2019 ftir july2010
18/76
Stretching vibrations
7/28/2019 ftir july2010
19/76
Bending vibrations
7/28/2019 ftir july2010
20/76
Hookes Law
7/28/2019 ftir july2010
21/76
=1
2pcK
larger K,
higher frequency
larger atom masses,
lower frequency
constants
2150 1650 1200
C=C > C=C > C-C=
C-H > C-C > C-O > C-Cl > C-Br
3000 1200 1100 750 650
increasing K
increasing
1m
21
21
mm
mm
7/28/2019 ftir july2010
22/76
Stretching vibrations of CO2
7/28/2019 ftir july2010
23/76
Stretching vibrations of CO2
7/28/2019 ftir july2010
24/76
7/28/2019 ftir july2010
25/76
Bond order affects bond strength, so bond order
affects the position of absorption bands
The approximate wavenumber of an absorption can be
calculated from Hookes law
is the wavenumber of the stretching vibration
f is the force constant
m1 and m2 are the masses of the atoms
2
1
21
21 )(2
1
mmmmf
cp
7/28/2019 ftir july2010
26/76
The exact position of the absorption band
depends on
electron delocalization
the electronic effect of neighboring substituents
hydrogen bonding
CH3CCH2CH2CH3
O O O-
C O C O
at 1720 cm1 at 1680 cm1
7/28/2019 ftir july2010
27/76
Putting an atom other than carbon next to the carbonyl group
causes the position of the carbonyl absorption band to shift
The predominant effect of the nitrogen of an amide is electron
donation by resonance
The predominant effect of the oxygen of an ester is inductive
electron withdrawal
7/28/2019 ftir july2010
28/76
The position of a CO absorption varies because
CH3CH2 OH
CH3CH2 O CH2CH3
C
O
H3C OHC
O-
H3C OH
C
O
H3C O CH3C
O-
H3C O CH3
~1050 cm1
~1050 cm1
~1250 cm1
~1250 cm1 and 1050 cm1
7/28/2019 ftir july2010
29/76
FTIR spectrometer
The main optical components are
The IR source
The interferometer The beamsplitter
The laser
The IR detector
7/28/2019 ftir july2010
30/76
IR source
The IR source produces IR radiation beam
that travels through the spectrometer
passing through the sample and to the
detector
7/28/2019 ftir july2010
31/76
Interferometer
Beamsplitter
Two mirrors
fixed
movable
7/28/2019 ftir july2010
32/76
Beamsplitter
Separates the IR beam into two beams of
nearly equal energy
One beam is reflected from the beam splitter
to a fixed mirror
The other beam is transmitted to a moving
mirror and back to the beam splitter where
the beams recombine
7/28/2019 ftir july2010
33/76
Laser
Used as an internal calibrator
control the moving mirrors position signals the capture of data
7/28/2019 ftir july2010
34/76
IR transducer
After being absorbed at specific
frequencies by the sample,
the remainder of the IR beam is focused onto
the detector
The detector produces signal to the
amount of IR radiation striking it and sends
the signal to the computer for processing
7/28/2019 ftir july2010
35/76
FTIR spectrometer
7/28/2019 ftir july2010
36/76
FTIR spectrometer
7/28/2019 ftir july2010
37/76
Emits IR radiation across the region of
interest
IR beam is deflected off a mirror
Directs the IR beam into the interferogram
Where the spectral encoding takes place
7/28/2019 ftir july2010
38/76
Interferometer
IR source
Beamsplitter
Moving
mirror
Fixed
mirror
Detector
Sample
He-Ne Laser
7/28/2019 ftir july2010
39/76
7/28/2019 ftir july2010
40/76
The computer reads the interferogram and
uses Fourier transformation to decode the
intensity information for each frequency
and presents a spectrum
7/28/2019 ftir july2010
41/76
The spectrometer measures the intensity
of a speciallyencoded IR beam after it has
passed through the sample
The resulting signal, interferogram
Contains infromation about all frequencies
present in the beam
7/28/2019 ftir july2010
42/76
7/28/2019 ftir july2010
43/76
Laser
Produces a single frequency of red light
that follows the same path as the IR beam
Calibrates the instrment internally
7/28/2019 ftir july2010
44/76
7/28/2019 ftir july2010
45/76
FOURIER TRANSFORMATION
7/28/2019 ftir july2010
46/76
7/28/2019 ftir july2010
47/76
Applications of IR spectrometry
Major application (MIR region)
Structural investigations of molecular
compounds particularly organic
compounds
M j li ti f IR
7/28/2019 ftir july2010
48/76
Major applications of IR
spectrometry
7/28/2019 ftir july2010
49/76
NIR and FIR regions
NIR region (4000 to 14,000 cm-1)
Quantitative determination of low MW
hydrocarbons, H2O, CO2, S
FIR region (15 to 100 m)
Structural determination of inorganic and
metal-organic species
7/28/2019 ftir july2010
50/76
What produces the IR spectrum
Materials that contain C (organic
compounds)
Many compounds; natural and synthetic
inorganic materials
7/28/2019 ftir july2010
51/76
Sample handling
No solvent is transparent throughout the
region of interest
Sample handling
Most difficult
Time-consuming
part of IR spectrometric analysis
7/28/2019 ftir july2010
52/76
Gases
Sample expands into an evacuated
cylindrical cells with suitable windows (Fig
16.3)
Variable lengths
A few cm to 10 m or longer
7/28/2019 ftir july2010
53/76
Measuring gases
Placed in a sealed cell
Specified T and P can be maintained
Gases are often present at low
concentrations
Mirrors are used to deflect the IR beam back
and forth before the beam is allowed to exit
Increased the interaction between gas sample andIR beam
To acquire more spectral information
This gas cell provides a fixed
pathlength that is 2 m long
7/28/2019 ftir july2010
54/76
Solutions
Water and alcohols are not suitable
solvents
SOLVENTS FOR IR
7/28/2019 ftir july2010
55/76
SOLVENTS FOR IR
SPECTROMETRYNo solvent is transparent throughout the region of interest
7/28/2019 ftir july2010
56/76
Liquids
Pure (neat) liquid
Very thin film, a short pathlength
A drop of neat liquid is squeezed between
two salt plates
Held together by capillary action
Placed in the beam of IR radiationNon reproducible transmittance data
Satisfactory for qualitative analysis
7/28/2019 ftir july2010
57/76
Measuring liquids
Liquid cells Holds the liquid between two crystal made
from materials that completely transmit IR
radiation
7/28/2019 ftir july2010
58/76
Cells
Liquid cells
(compared to UV/Vis spectroscopy)
Narrow pathlengths
0.01 to 1 mm
Relatively high sample concentrations
0.1% to 10%
Short pathlengths
Low molar absorbtivity
7/28/2019 ftir july2010
59/76
Liquid cells
Designed for easy assembly/disassembly
Use Teflon spacers
Allow variations in path lengths
Fixed path-length cells
Use a syringe to fill or empty
Demountable IR cell for liquid
7/28/2019 ftir july2010
60/76
Demountable IR cell for liquid
sample
C IR i d t i l
7/28/2019 ftir july2010
61/76
Common IR window materials
7/28/2019 ftir july2010
62/76
Selection of window materials
Cost
Range of transparency
Solubility in solvent
Reactivity with sample or solvent
Most common NaCl and KBr
7/28/2019 ftir july2010
63/76
Solids
Mulls
Pellets
7/28/2019 ftir july2010
64/76
Solids
Solid in a liquid or solid matrix
Samples are ground to fine powder
To avoid effects of scattered radiation
7/28/2019 ftir july2010
65/76
Pelleting
Most common technique; KBr pelleting
Halide salts have a property ofcold flow
Translucent property when pressure is
applied to finely powdered salts
~1 mg (or less) sample is mixed with ~
100 mg dried KBr
Mix and grind in a mortar and pestle
Apply 10,000 to 15,000 psi in a die
Produce a transparent disk or pellet
7/28/2019 ftir july2010
66/76
Pelleting
Best done in vacuum or store the pellet in
a dessicator before measurement
Bands at 3450 and 1640 cm-1 due to
absorbed moisture
7/28/2019 ftir july2010
67/76
Mulls
Solids
Not soluble in an IR transparent solvent
KBr pelleting is not convenient
Grind ~ 2 5 mg fine powder in a drop of
mulling agents
Heavy hydrocarbon (Nujol)
Fluorolube (halogenated polymer)
Examine as thin film between two salt
discs
7/28/2019 ftir july2010
68/76
Qualitative analysis
FTIR, NMR and MS
Indentify
Organic
Inorganic species
Biological
Limitations to the use of
7/28/2019 ftir july2010
69/76
Limitations to the use of
correlation chartsTo determine the identity or structure Not possible using the correlation charts
alone (only as a guide).
Overlapping of group frequencies Instrumental limitations
Spectral variations as a function of physical
states Neat liquid
Solution
Pellet
Mull
Table of group frequencies for
7/28/2019 ftir july2010
70/76
Table of group frequencies for
organic functional groups
7/28/2019 ftir july2010
71/76
7/28/2019 ftir july2010
72/76
7/28/2019 ftir july2010
73/76
Group frequency region of MIR
7/28/2019 ftir july2010
74/76
Group frequency region of MIR
Active or inactive vibrations in
7/28/2019 ftir july2010
75/76
Active or inactive vibrations in
IR spectrum?
7/28/2019 ftir july2010
76/76