Installation & Operating Manual Temperature Electronics Ltd. Unit 2, Wren Nest Road Glossop, SK13 8HB United Kingdom Authorised Australian Distributor: T: 08 9409 1789 | F: 08 9309 1206 A: 45 Attwell Street, Landsdale WA 6065 E: [email protected] www.johndecplastics.com.au FUME HOOD AUTO-SASH CONTROLLER

FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Installation & Operating Manual

Temperature Electronics Ltd.Unit 2, Wren Nest RoadGlossop, SK13 8HBUnited Kingdom

Authorised Australian Distributor:

T: 08 9409 1789 | F: 08 9309 1206 A: 45 Attwell Street, Landsdale WA 6065

E: [email protected] www.johndecplastics.com.au

FUME HOODAUTO-SASH CONTROLLER

Page 2: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Table of Contents

1 Safety 6

1.1 Safety Practices 61.2 Warning Notices 61.3 Precautions 71.4 General Operating Conditions 71.5 Environmental Conditions 71.6 Electrical Safety 81.7 Electrical Protection 91.8 EMC Compliance 91.9 Warning Labels 9

2 Overview of the fume hood Auto-Sash Controller 11 52.1 Features 112.2 Auto-Sash Control Options 122.2.1 Auto-Sash Controller with Tiptronic Open/Close Control 122.2.2 Auto-Sash Controller with Sash Low Micro-Switch 132.2.3 Auto-Sash Controller with Sash Low Proximity Switch 132.3 Control System Components 142.3.1 Auto-Sash Control Unit 152.3.2 Personnel Sensor 162.3.3 Keypad/Alarm Panel 172.3.4 Sash Light Curtain 182.3.5 Under Sash Sensor 192.3.6 Sash Low Switch 192.3.7 Sash Position Sensor 202.3.8 Tilt Switch 202.3.9 Foot Switch 212.3.10 Sash Drive Unit 21

3 Installing the Auto-Sash Control System Components 23

3.1 Auto-Sash Control Unit 233.1.1 Mounting the Unit 233.1.2 Wiring in the Auto-Sash Control Unit 243.2 Auto-Sash Keypad/Alarm Panel 263.2.1 Fitting the Auto-Sash Keypad/Alarm Panel 263.3 Auto-Sash Motor Drive Unit 273.3.1 Rack and Pinion Motor Drive 283.3.2 Chain and Sprocket Drive 313.3.3 Wire and Pulley Drive 333.4 Sash Light Curtain 343.4.1 Installing the Sash Light Curtain Control Box 34

Page 3: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Table of Contents

3.4.2 Installing the Sash Light Curtain Transmitter and Receiver 343.4.3 Aligning the Light Curtain Transmitter and Receiver 353.5 Installing and Aligning the Under Sash Sensor 393.6 Tilt Switch 413.7 Sash Low Switch 423.7.1 Installing the Sash Low Micro-Switch 423.7.2 Installing the Sash Low Proximity Switch 433.8 Sash Position Sensor 443.9 Personnel Sensor 463.9.1 Installing the Personnel Sensor 463.9.2 Adjusting the Personnel Sensor Beam Width 483.9.3 Personnel Sensor Dip Switch Settings 503.10 Auto-Sash System Control Box - General Wiring Diagram 51

4 Setting Up and Calibration of Auto-Sash Control Unit 53

4.1 Starting Up the Auto-Sash Control Unit 534.2 Calibrating the Auto-Sash Control Unit for Tiptronic Operation using the Auto-Sash Keypad 544.3 Calibrating the Sash Lock Function 554.4 Calibrating the Safety Light Curtain 574.5 Calibrating the Tilt Switch 604.6 Calibrating the Sash Position Sensor 604.7 Calibrating the Personnel Sensor 604.8 Calibrating the Sash Low Switch 60

5 System Menus 62

5.1 Accessing System Menus 645.2 Navigating the System Menus 655.2.1 Keypad Tones 655.2.2 Sash 1 Settings Menu 655.2.3 BMS Input Settings Menu 665.2.4 BMS Output Settings Menu 67

Page 4: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Table of Contents

6 Operation 70

6.1 Control Philosophy 706.2 Modes of Operation 706.2.1 User Present 706.2.2 User Not Detected 716.2.3 Auto-Open(UserConfigurable) 716.2.4 Sash Lock 726.2.5 BuildingManagementSystem(BMS)Inputs 726.3 Operation and Alarm Indication 726.3.1 RUN Mode 726.3.2 Alarm Conditions 73

7 Troubleshooting 77

7.1 Faults and Alarms 77

8 Limitation of Warranty and Liability 79

9 Auto-SashControllerSeriesSpecification 81

Page 5: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

1. SAFETY

Page 6: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.6

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

1 SAFETY

1.1 Safety Practices

ThisdocumentdescribesthegeneralsafetypracticesandprecautionsthatmustbeobservedwhenoperatingtheAuto-SashController.

Thisadviceisintendedtosupplement,notsupersede,thenormalsafetycodesintheuser’scountry.Theinformationprovideddoesnotcovereverysafetyprocedurethatshouldbefollowed.Ultimately,maintenanceofasafelaboratoryenvironmentistheresponsibilityoftheuserandtheuser’sorganisation.

PleaseconsultalldocumentationsuppliedwiththeAuto-SashControllerbeforestartingtowork.Carefullyreadthesafetyinformationinthisdocumentandintheotherdocumentationsupplied.Whensettinguptheequipmentorperforminganalysisormaintenanceprocedures,strictlyfollowtheinstructionsprovided.

1.2 Warning Notices

WithinthisUserGuideWARNINGSareusedtohighlightinformationorinstructionsthatmustbefollowedinordertoavoidpersonalinjurytoyourselforotherpeopleinthevicinity,eg.switchoffthemainsvoltagebeforeanymaintenance.

WARNINGS appear as below:

Page 7: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.7

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

1.3 Precautions

The following precautions must be observed when using the Auto-Sash Controller and associated systems:

•BesurethatthevoltageoftheAuto-SashControllerequipmentcorrespondstothevoltageavailablewhereitistobeinstalled.

•NeverremovethesideorbackpanelsoftheAuto-SashControllerwithoutfirstshutting down the equipment and disconnecting the mains cord.

1.4 General Operating Conditions

TheAuto-SashControllerandequipmenthavebeendesignedandtestedinaccordancewiththesafetyrequirementsoftheInternationalElectrotechnicalCommission(IEC).TheAuto-SashControllerconformstoIEC61010-1(SafetyRequirementsforelectricalequipmentformeasurement,controlandlaboratoryuse)asitappliestoIECClass1(earthed)appliances,andthereforemeetstherequirementsofECdirective73/23/EEC.

Ifpossible,avoidanyadjustment,maintenanceorrepairtotheequipmentwhilstcoversareopenoritisoperative.However,ifanyadjustment,maintenanceorrepairisnecessarywhilethecoversareopen,thismustbedonebyaskilledpersonwhoisawareofthehazardsinvolved.

Whenever circumstances arise that mean an Auto-Sash Controller may be unsafe, make it inoperative. In particular, an Auto-Sash Controller may be unsafe if it:

•Showsvisibledamage.

•Failstoperformcorrectly.

•Hasbeensubjectedtoseveretransportstresses.

•Hasbeensubjectedtoprolongedstorageinunfavorableconditions.

1.5 Environmental Conditions

The Auto-Sash Controller should only be used under the following conditions:

• Indoors.

•Inambienttemperaturesbetween5°Cand40°C.

•Withrelativehumiditybelow80%fortemperaturesupto31°C,decreasinglinearlyto50%relativehumidityat40°C.

•Electricalsupplyfluctuationsnotexceeding+10%ofthenominalvoltage.

Page 8: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.8

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

1.6 Electrical Safety

The Auto-Sash Controller and associated equipment are designed to protect the user frompotentialelectricalhazards.Thissectiondescribessomerecommendedelectricalsafetypractices.

TheAuto-SashControllerandassociatedequipmentmustbecorrectlyconnectedtoasuitableelectricalsupply.Thesupplymusthaveacorrectlyinstalledprotectiveconductor(earthorground)andmustbeinstalledandcheckedbyaqualifiedelectricianbeforeinitial power up.

When working with the Auto-Sash Controller System:

• Connect the equipment to a correctly installed mains power outlet that has a protective conductor connection.

• Do not operate the equipment with any covers or internal parts removed.

•Disconnecttheequipmentfromalllivevoltagesourcesbeforeopeningittomakeanyadjustments,replacements,maintenanceorrepair.Iftheopenedequipmentmustbeoperatedforfurtheradjustment,maintenanceorrepair,thismustonlybedonebyasupplier’s Service Engineer.

Ifitispossiblethattheequipmentisnolongerelectricallysafeforuse,maketheequipment inoperative and secure it against any unauthorised or unintentional operation.

The electrical safety of the equipment is likely to be impaired if:

•Ithasanysignsofvisibledamage.

•Ifithasbeensubjectedtoprolongedstorageinunfavourableconditions.

•Ifithasbeensubjectedtoseverestressduringtransportation.

Page 9: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.9

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

1.7 Electrical Protection

Observe the following electrical protection precautions:

•Insulation:ClassIratingforexternalcircuits.OnlyconnectequipmentthatmeetstherequirementsofIEC61010-1,IEC60950orequivalentstandards.

•InstallationCategory:Theequipmentisabletowithstandtransientover-voltagestypicallypresentonthemainssupply.Thenormalleveloftransientover-voltagesisimpulsewithstand(overvoltage)CategoryIIofIEC60364-4-443.

• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however,temporaryconductivitycausedbycondensationmustbeexpected.

1.8 EMC Compliance

EC DirectiveThe Auto-Sash Controller System is designed and tested to meet the requirements oftheECdirective89/336/EECand93/68/EECandcomplieswiththeEMCstandardEN61326(EMCstandardforelectricalequipmentformeasurement,controlandlaboratoryuse)andEN55011(ISM)ClassA(RFemissions).

FCC Rules and RegulationsTheAuto-SashControllerSystemisclassifiedasadigitaldeviceusedexclusivelyasindustrial,commercialormedicaltestequipment.ItisexemptfromthetechnicalstandardsspecifiedinPart15oftheFCCRulesandRegulationsbasedonSection15.103(c).

1.9 Warning Labels

Warninglabelsattachedtotheequipmentdrawattentiontospecifichazards-refertothisguideandotherdocumentationprovidedwiththeequipmentformoredetailsconcerningpotentialhazardsandanyprecautionsorotheractionsthatmustbetaken.

Page 10: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

2. OVERVIEW OF THE FUME HOODAUTO-SASH CONTROLLER

Page 11: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.11

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

TheAuto-SashControllerautomatestheopeningandclosingofthefumehoodsash.Automationofthesashmovementprovidesanincreasedlevelofsafetytothecontrolledenvironmentassashescannotbe‘accidentally’leftopenandcanrespondtosafetycriticalinputsfromexternalsourcessuchasfirealarmsandventilationsystemfailures.

Suchautomationalsofacilitatescostsavings,iflinkedtoventilationcontrols,asautomaticallyclosinganunattendedfumehoodallowsextractorfanspeedsandelectricityconsumptiontobereduced.

2.1 Features

The Auto-Sash Controller has the following features/functionality:

1. Auto-Close-Thesashcanbeautomaticallyclosedwhenthefumehoodisunattended.

2. Auto-Open-Thesashcanautomaticallyopen(orre-openifpreviouslyautomaticallyclosed)whentheoperatorreturns.

3. Sash Lock-Thesashcanbeprogrammedtolock,ie.Auto-SashDriveinhibited,whenmovedtoauserdefinedposition,eg.fullyopen,toallowforloadingorsettingupofequipmentinsidetheFumeCupboard.

4. Manual Sash Open/Close-Allowsthesashtobeopenedorclosedmanually.

5. Keypad Pushbutton Open/Close-ThesashcanbeopenedorclosedusingtheKeypadpushbuttons.

6. Footswitch-Thesashcanbeopenedorclosedusinganoptional,thirdpartysupplied,footswitch.

7. Tilt Switch-Anoptional,thirdpartysupplied,tiltswitchcanbeusedtodisabletheAuto-SashControlUnitwhenthesashisopenedformaintenance.

8. Building Management System (BMS) Inputs - The sash can automatically close whenasignalisreceivedfromtheBMS,eg.intheeventofafirealarm.

9. Building Management System (BMS) Outputs - The Auto-Sash Controller can sendanumberofsignalstotheBMSregardingthecurrentstatusofthefumehood, eg. sash position, alarm status.

2 OVERVIEW OF THE FUME HOODAUTO-SASH CONTROLLER

Page 12: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.12

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.2 Auto-Sash Control Options

Three main control options can be utilised by the Auto-Sash Controller to control the opening and closing of the fume hood sash:•Auto-SashControllerfittedwithTiptronicOpen/CloseControl.•Auto-SashControllerfittedwithSashLowMicro-Switch.•Auto-SashControllerfittedwithSashLowProximitySwitch.

2.2.1 Auto-Sash Controller with Tiptronic Open/Close Control

TiptronicOpen/Closecontrolisthedefaultcontrolsetup,usingaSashPositionSensor(sprungpotentiometer)withadrawwire(1.2mlong)attachedtoeitherthesash or the counterweight. The Auto-Sash Controller can then be calibrated to give one of the following methods of operation:

Method A - Close only(bottompositiononlycalibrated)Thesashautomaticallydrivestothebottompositionwhentheoperatorisnotpresent. Pressing the Downpushbuttondrivesthesashtothebottompositionwhenthe operator is present. The Uppushbuttonisdisabled.

Method B - 2-Position(bottomandmid-positionscalibrated)Thesashautomaticallydrivestothebottompositionwhentheoperatorisnotpresent.PressingtheDownpushbuttondrivesthesashtothebottompositionwhentheoperatorispresent.PressingtheUppushbuttondrivesthesashtothemid-position when the operator is present.

IftheTiptronicfeatureisenabled,manuallytip(move)thesashupwards-thesashdrives to the mid-position. Manually tip the sash downwards - the sash drives to the bottomposition.

IftheAutoOpenfeatureisenabled,thesashautomaticallydrivestothemid-positionwhentheoperatorispresent(slighttimedelayfollowinguserdetection).

IftheAutoLockfeatureisenabledandaSashLockpositionhasbeencalibrated,oncethesashisraisedtoapositionequaltoorabovetheSashLockpositionthesashlocksinitscurrentpositionandtheAutoClosefeatureisdisabled.ThisfacilitatesloadingoftheFumeCupboard.

Page 13: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.13

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

Method C - 3-Position(bottom,midandtoppositionscalibrated)Thesashautomaticallydrivestothebottompositionwhentheoperatorisnotpresent.PressingtheDownpushbuttondrivesthesashtothebottompositionwhentheoperatorispresent.PressingtheUppushbuttondrivesthesashtothenextcalibratedpositionwhentheoperatorispresent,ie.ifthesashislowerthanthemid-positionthesashisdriventothemid-position,ifthesashishigherthanmid-positionthesashisdriven to the top position.

IftheTiptronicfeatureisenabled,manuallytip(move)thesashupwards-thesashdrivestothenextcalibratedposition.Manuallytipthesashdownwards-thesashdrivestothebottomposition.

IftheAutoOpenfeatureisenabled,thesashautomaticallydrivestothemid-positionwhentheoperatorispresent(slighttimedelayfollowinguserdetection),ifthesashislower than mid position.

IftheAutoLockfeatureisenabledandaSashLockpositionhasbeencalibrated,oncethesashisraisedtoapositionequaltoorabovetheSashLockpositionthesashlocksinitscurrentpositionandtheAutoClosefeatureisdisabled.ThisfacilitatesloadingoftheFumeCupboard.

2.2.2 Auto-Sash Controller with Sash Low Micro-SwitchIftheTiptronicOpen/Closecontroloptionisnotchosen,aSashLowMicro-SwitchcanbespecifiedforfittingbyTELorbythefumehoodmanufacturer.Theswitchcanbearrangedtobeclosedoropenwhenthesashisclosed,dependingonhowandwheretheswitchisfitted.Thewiringarrangementmustbemodifiedtosuittheparticularinstallationandtheoperationalrequirementsofthefumehood.

The following functionality is available:

•Thesashautomaticallydrivestothebottompositionwhentheoperatorisnotpresent.

•PressingtheDownpushbuttondrivesthesashtothebottompositionwhentheoperator is present.

•TheUppushbuttonisdisabled.

2.2.3 Auto-Sash Controller with Sash Low Proximity SwitchIftheTiptronicOpen/Closecontroloptionisnotchosen,aSashLowProximitySwitchcanbespecifiedforfittingbyTELorthefumehoodmanufacturer.Theswitchcanbearrangedtobeclosedoropenwhenthesashisclosed,dependingonhowandwheretheswitchisfitted.Thewiringarrangementmustbemodifiedtosuittheparticularinstallationandtheoperationalrequirementsofthefumehood.

The following functionality is available:

•Thesashautomaticallydrivestothebottompositionwhentheoperatorisnotpresent.

•PressingtheDownpushbuttondrivesthesashtothebottompositionwhentheoperator is present.

•TheUppushbuttonisdisabled.

Page 14: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.14

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.3 Control System Components

TheAuto-SashControlSystemcomprisesacontrolunitaccompaniedbyanumberofadditional components and options depending on the installation.

Note: ThisexampleinstallationisfittedwithaPulleyandWireSashDrive.

Page 15: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.15

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.3.1 Auto-Sash Control UnitTheAuto-SashControlUnitprocessestheinputsfromthesensorsandswitchesassociatedwiththesashcontrolfunctionandoutputsrelevantsignalstothemotordrive(s)andalarmpanel.

Page 16: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.16

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.3.2 Personnel SensorThePersonnelSensorisaPassiveInfra-Red(PIR)occupancydetectorwhichdetectsthepresenceorabsenceofthefumehoodoperatorandpreventsthesashclosingautomaticallywhentheoperatorispresentatthefumehood.

Note:Iftheoperatorisdetected,thesashwillnotautomaticallyclosebutcanstillbeopenedandclosedeithermanuallyorbyusingtheUp/Downpushbuttons.

Onpowerup,thePersonnelSensor‘learns’thereflectioncharacteristicsoftheenvironmentwithinitsfieldofviewandstoresthisinformationasreferencebackgrounddata.Thesensorthen‘re-learns’thebackgroundeverythreeminutes,sothatifastationaryobject,suchasastool,isleftinthefielditwillbecomepartofthebackgroundand‘ignored’bythesensor.Then,whentheoperatorisnolongerpresentatthefumehood,thesashwillcloseautomatically,followingapre-setdelaytime,aslongastheSashLightCurtainisnotbrokenandthesashisnotalreadyclosed.

Page 17: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.17

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.3.3 Keypad/Alarm PanelTheKeypad/AlarmPanelprovidesavisualindicationofthestatusoftheAuto-Sash,includinganyfaultconditions,viaaLiquidCrystalDisplay(LCD).

Page 18: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.18

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.3.4 Sash Light CurtainTheSashLightCurtainisasafetydevicedesignedtodetectwhethertheopeningbelowthesashisclearofobstructionsbeforethesashisautomaticallyclosed.Whenanobstructionisdetected,thesashdriveisdisabledandthealarmactivated.TheSashLightCurtaincomprisesaninfraredtransmittermountedononesideofthefumehoodandaninfraredreceivermountedontheoppositesideofthefumehood’sinnerchamber.

Theinfraredtransmitterhasmultiple,closelyspaced,LightEmittingDiodes(LEDs)tofacilitatethedetectionofsmallobjects.

Objectsaredetectedwheninfraredlightfromthetransmitterisnotdetectedbythereceiverduetoanobstructioninitspath.

A separate Sash Light Curtain Control Box includes a sensitivity potentiometer thatallowsthebeamsensitivitytobesetto‘see’glassorotherobjectspartiallytransparenttoinfraredlight.

Page 19: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.19

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.3.5 Under Sash SensorThe Under Sash Sensor is used as an alternative to the Sash Light Curtain and is designedtodetectwhethertheopeningbelowthesashisclearfromobstructionsbeforethesashisdrivenclosed.

Asinglesensorononesideofthefumehooddetectsthereflectionofabeamoflightfromaself-adhesivestripofretro-reflectivetapemountedontheoppositeinsidewallofthefumehood.

Thesensorincorporatesapotentiometerwhichallowsthebeamsensitivitytobesetto‘see’glassobjects.

2.3.6 Sash Low SwitchThe Sash Low switch is used when the Tiptronic control option (see Section 2.2 - Auto-SashControlOptions)hasnotbeenchosenandisusedtosignaltotheAuto-SashControlUnitthatthefumehoodsashisclosed.

Theswitchcanbeeitheramechanicalrollerlevertypemicro-switchoraninductiveproximity switch.

Page 20: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.20

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.3.7 Sash Position SensorTheSashPositionSensor,comprisingaflexiblecable,aspring-loadedspoolandapotentiometer, is used in conjunction with the Tiptronic control option to detect and measureverticalsashmovementandthepositionofthefumehoodsash.

Duringtheinitialon-screencalibrationsetup(seeSection4-SettingUpandCalibrationofAuto-SashControlUnit),thesashcontrollerwilldetectthedirectionoftravelofthesensorandthenprovideinformationonthepositionofthesashtotheAuto-SashControlUnittoenablecalibrationofthesashopenandsashclosedpositions.

2.3.8 Tilt SwitchTheTiltSwitchisanoptional,thirdpartysupplied,componentfittedtothefumehoodtop panel which operates to isolate the Auto-Sash controls when the top panel is openedtofacilitateservicingofthefumehood.

Page 21: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.21

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.3.9 Foot SwitchThe Foot Switch is an optional, third party supplied, componentwhichcanbepositionedonthefloorinfrontofthefumehoodandisusedtoopenthesash.

2.3.10 Sash Drive UnitTheSashDriveUnitisagearedelectricmotorandclutchassemblyusedtodrivethesash open and closed.

Three types of drive are available, none of which preclude manual opening and closing of the fume hood:

• Rack and Pinion drive.

• Chain and Sprocket drive.

• Pulley and Wire drive.

Page 22: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

3. INSTALLING THE AUTO-SASH CONTROL SYSTEM COMPONENTS

Page 23: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.23

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

The following section outlines the installation of the various components of the Auto-Sash control system.

Asthesizeandformatofindividualfumehoodsmayvaryconsiderably,specificinstallationinstructionsarenotpossible,thoughtheprinciplesoutlinedbelowshouldremain valid in all cases.

Note:Beforestartinginstallation,ensurethatthesashisfreetomoveinitsrunners.Usesiliconsprayonthesashrunnersoradjustthesashrunner’salignmentifthesashdoesnotmovefreely.

3.1 Auto-Sash Control Unit

3.1.1 Mounting the UnitThe Auto-Sash Control Unit can be mounted as follows:

•Screwedflattotheoutsideofthefumehoodroof(horizontalmounting).

Note: Forhorizontalmounting,thecontrolboxshouldalwaysbemountedwiththelid/frontpanelfacingupwards.

•Screwedtotheoutsidewalloffumehoodwall(verticalmounting).

OncetheAuto-SashControlUnithasbeenpositioned,markitspositionusingthefixingholesintherearofthehousingasatemplate.Then,attachtheAuto-SashControl Unit with screws.

3 INSTALLING THE AUTO-SASH CONTROL SYSTEM COMPONENTS

Page 24: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.24

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.1.2 Wiring in the Auto-Sash Control UnitOnafumehoodwithaninnerandoutersidewall,cablestoandfromsensors,themotordriveandtheAuto-SashControlUnitcanberoutedthroughthesidewallandinto the service void.

Onasinglewalltypefumehood,cablestoandfromsensors,themotordriveandtheAuto-SashControlUnitshouldberouteduptheoutsideofthesidewallusingsuitabletrunkingorfixingpadsandties.

When wiring in the Auto-Sash Control Unit, observe the following:

•Inallcases,cablesshouldbekeptclearofmovingpartssuchascomponentsofthemotor drive and sash.

•Inallcases,grommetsshouldbeusedwherecablespassthroughsidewallstopreventabrasionofthecableinsulationand/orsidewall.WhereacableissuppliedwithaDINconnector,thegrommetshouldhaveaminimuminternaldiameterof17mm(0.67”).

•Ensurethatanybendingofthecables,suchaswhenpassingat90°throughasidewall,doesnotleadtothembecomingoverstressed.

Page 25: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.25

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

Page 26: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.26

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.2 Auto-Sash Keypad/Alarm Panel

AlternativeAuto-SashKeypad/AlarmPanelscanbesuppliedforUSandEUmarkets.Bothshouldbemountedonthefrontofthefumehood,ensuringthereisadequateclearancebehindforthecablesconnectingthekeypadtotheAuto-SashControlUnit.

3.2.1 Fitting the Auto-Sash Keypad/Alarm Panel

1. FittheAuto-SashKeypadtothefumehoodusingeitherasinglegangsocketforUKapplicationsor,forUSapplications,usingacut-outtothe(Version2)dimensions shown in Figure 20 and Figure 21:

Page 27: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.27

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2. Connect the Keypad/Alarm Panel to the Auto-Sash Control Unit. See the wiring schematic in Section 3.10 - Auto-Sash System Control Box - General Wiring Diagram and Figure 22.

3.3 Auto-Sash Motor Drive Unit

AMotorDriveUnitisusedtoautomaticallyraiseandlowerthefumehood sash.

Three types of drive are available:

• Rack and Pinion Motor Drive.

• Chain and Sprocket Motor Drive.

• Wire and Pulley Motor Drive.

Theoptimalchoiceofsashdrivetypedependsonwhethertheinstallationispartofanewbuildoraretro-fit,thesizeandconfigurationofthefumehoodanditslocation.

Noneoftheseoptionsprecludemanualoperationofthesash.

Page 28: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.28

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.3.1 Rack and Pinion Motor DriveTheRackandPinionMotorDriveutilisesarackfitteddirectlyontothesashframeandamotordriveassemblyfittedtoaframeworkabovethefumehood.

Thisdrivesystemisidealforretro-fitinstallationswherethereisnoaccesstothesashwiresorcounterweightsand/orwherethesashisfittedtoametalframework,eg.combinationtypesashfumehoods.

3.3.1.1 Installing the Rack and Pinion Drive

1.Raisethesashtothefullyopenposition.

2.Offerupthemotorassemblyandracktoensurethatthereisaclearanceofatleast1.2m(47.24”)abovethetopofthesash.

•Therackissuppliedin1m(39.37”)lengths.

3.Themotordriveassemblyshouldbefittedtoarigidframeworkandpositionedcentrallyabovethesash.

ToensurethattheT-barfixingclearsthebottomofthemotorassembly,thereshouldbeatleast50mmbetweenthetopofthesashandthebottomofthemotordriveassembly,whenthesashisfullyopened

Note:Ensurethattherackisofsufficientlengthtostillprotrudethroughthetopofthemotorassemblywhenthesashisfullyclosed.

4.Oncethemotordriveassemblyisfixedtothefumehood,theT-barshouldbefixedtotherackandofferedupthesashframe.

5.EnsurethattherackisfreetomoveoverthefullsashopeningbeforemarkingupandfixingtheT-bartothesashframe.

Note:Therackmusttorunverticallyat90degreestothehorizontalaxisofthesashframe.

Page 29: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.29

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

6.Ifasashlowswitchisfittedinsteadofasashpositionsensor,therackwillneedtobecuttothecorrectlengthtoenabletheswitchtooperate.

Closethesashfullyandmarktherackjustbelowthepositionoftheproximitylowswitchatthetopofthemotorassemblytoensurethattherackisclearoftheswitch when the sash is closed.

7.ConnecttheSashDrivetotheAuto-SashControlUnitusingthecablesprovided(redandblackconnectors).SeethewiringschematicinSection3.10-Auto-SashSystem Control Box - General Wiring Diagram.

Note:Itisimportanttoobservethecorrectpolaritywhenmakingelectricalconnections.

8.ConnecttheclutchassemblytotheAuto-SashControlUnitusingthecablesprovided(blueandyellowconnectors).

Page 30: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.30

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

Page 31: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.31

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.3.2 Chain and Sprocket Drive

Two types of chain and sprocket drive are available:

•ShaftDrive.

• Counterweight Drive.

3.3.2.1 Chain and Sprocket Shaft DriveWiththeShaftDrivearrangement,themotordriveutilisesataperboresprocketandarollerchaintodriveashaftrunningbetweenthecounterweightpulleys.

Themotorassemblyissuppliedfittedtoanadjustablebracketsothatthecorrectchaintensioncanbeset.

TheShaftDrivearrangementcanaccommodateoneortwosashwires.

3.3.2.2 Installing the Chain and Sprocket Shaft Drive

1.Fitthetaperboresprockettotheshaftandensurethatthesprocketissecurelyfitted.Themotordriveandsprocketshouldbepositionedcentrallyontheshaftwherepossibleandthemotorassemblyshouldbefittedtoarigidframeworktosupport the motor.

2.Looselyfitthechainbeforefixingthemotorplatetoensurethechainiscorrectlytensioned and aligned.

Page 32: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.32

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.3.2.3 Chain and Sprocket Counterweight DriveWith the Counterweight Drive arrangement, the chain and sprocket drive system isusedtodirectlyliftthesashcounterweightbyfittingarollerchaintothecounterweightandutilisingaseparatecounterweightandfreewheelingsprockettomaintain the correct chain tension.

3.3.2.4 Installing the Chain and Sprocket Counterweight Drive

1.Oncethemotor,chainandfreewheelingsprocketarefitted,manuallyopenandclosethesashtoensurethatthemotordriveisfreerunningandthechainisnottoo slack or too tight.

2. Fit the motor so that the drive sprocket is in line with the counterweight pulley.

3.Fitoneendofthechaintothecounterweightorsashwireincloseproximitytothecounterweight.

4.Fitthefreewheelingsprocketinlinewiththedrivesprocket.

5.Looselyfitthechainoverthefreewheelingpulleyanddeterminethebestpositionforthechaincounterweight.

6.Fitthechaincounterweightusingacounterweightguideifnecessary.

7.Oncethemotor,chainandcounterweightarefittedmanuallyopenandclosethesashtoensurethemotordrive,chainandcounterweightarefreerunning.

Page 33: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.33

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.3.3 Wire and Pulley DriveWith the Wire and Pulley Drive, the pulley motor drives the sash wire and can accept bothsingleanddualsashwiresystems.Themotorassemblyisfittedinlinewiththesash wires and uses two guide pulleys and a drive pulley.

Thepulleydriveisidealfornewfumehoodsbutcanberetro-fittedifalongerreplacementsashwirecanbefitted.

3.3.3.1 Installing the Wire and Pulley Drive

Themotorassemblycanbefittedhorizontallyorverticallyoneithersideofthefume hood, as follows:

1.Fitthemotorassemblysothattheguidepulleysareinlinewiththesashwire.

2.Runthesashwireoverthefirstguidepulley,aroundthedrivepulleyandthenoverthefinalguidepulley.

3. Once the motor drive is installed, open and close the sash to ensure that the sash cantravelfreely.

Page 34: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.34

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.4 Sash Light Curtain

The Sash Light Curtain consists of three main components and associated wiring:

• Sash Light Curtain Control Box.

• LED Transmitter.

• LED Receiver.

The Sash Light Curtain Control Box is connected to the main Auto-Sash Control Unit.

3.4.1 Installing the Sash Light Curtain Control Box

1.FixtheSashLightCurtainControlBoxontothetopofthefumehood,usingthetabsattachedtothecontrolboxandthescrewsprovided,ensuringthatthereisaccess to the terminals to connect the wiring.

2.ConnecttotheAuto-SashControlUnit,usingthepre-assembledcablesandconnectors supplied, as per the wiring schematic in Section 3.10 - Auto-Sash System Control Box - General Wiring Diagram.

3.4.2 Installing the Sash Light Curtain Transmitter and Receiver

1.Findasuitablepositionascloseaspossibletothebackofthesashtrack,ensuringthatthesashhandlewillnotprotrudeintothelightbeamandthattherearenootherobstructionsoneachsidewall.

Page 35: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.35

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.Carefullyremovethefrontplasticcoversfromtherighthandsideofthelightcurtainreceiverandthelefthandsideofthelightcurtaintransmittercases.

3. Position the light curtain transmitter and receiver on oppositesidewallsensuringthatthebackofthesashhandledoesnotpenetratethelightbeampaths. Thesashlightcurtaintransmitterandreceivershouldbefittedsothattheircableentry/exitpointsareatthetop. Ensurethatthelightbeampathsarefreefromotherobstructions.

3.4.3 Aligning the Light Curtain Transmitter and ReceiverMounting the light curtain receiver and transmitter at the same height on opposing sidesofthefumehoodopeningwillusuallybesufficienttoalignthemcorrectlywithnofurtheradjustmentrequired.

Asmallerrorinalignmentisacceptable,theallowableerrorbeingdependentonthespacingbetweenthetransmitterandreceiver.

Figurebelow,illustratesthemaximumallowabletransmittertoreceivermisalignment/displacementfora0to3m(0to118.11”)rangeoffumehoodwidths.Ascanbeseen,foratypicalfumehoodwidthof1to2m(39.37to78.740”),themaximumallowablemisalignment/displacementrangesfromapproximately20to40mm(0.79to1.58”).

Page 36: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.36

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

4.Theareaprotectedbythelightbeamsshouldcovertheopenareaunderthesash,typically500mm(20”)fromitsbase. Thelightcurtaintransmitterandreceivercanbepositionedabovethebottomofthesashinthefumehoodaslongastheyarefreefromobstructions. Inatypicalinstallation,thelightpathwouldbealignedflushtothebottomofthesash’sbase.

Note:Thereisa123mm(4.84”)gapfromthetopofthelightcurtaintransmittertothefirstLEDdownanda22mm(0.87”)gapfromthebottomofthetransmittertothefirstLEDup.

Page 37: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.37

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

5. Oncethelightcurtaintransmitterandreceiverhavebeenpositioned,markthesidewallusingthefixingholesinthetransmitter and receiver as templates.

6.Drilltheholesforthefixingscrews,usingthe2mm(0.08”)drillbitprovidedandattachthetransmitterandreceiver.

7. Onaconventionalfumehoodwithaninnerandoutersidewall,thelightcurtaintransmitterandreceivercablescanrunthrough the sidewall into the service void and up to the Sash LightCurtainControlBoxontopofthefumehood. Ifthefumehoodisasinglewalltype,drilltheholeinthetopsoffitpanelandrunthecablesupthesidewallsusingsuitabletrunkingorfixingpadsandties. Ifagrommetisused,drillasuitableholeinthesidewalltoaccept the grommet. Ifnogrommetisused,drilla17mm(0.67”)holetoclearthecableDINconnector.

Note: Thehole’scentreshouldbeaminimumof35mm(1.38”)fromthetopofthelightbeampath,sothatthecableisnotstressedwhenitbends90°throughtheinnersidewall.

8.PassthecablesthroughthesidewallandconnecttotheSashLightCurtainControlBoxonthetopofthefumehood.

9. Connect the light curtain transmitter and receiver to the SashLightCurtainControlBoxusingtheplug-inRJ45cablesprovided.

Page 38: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.38

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

10.ConnecttheoutputfromtheSafetyCurtainControlBoxtothe Auto-Sash Control Unit. See Section 3.10 - Auto-Sash System Control Box - General Wiring Diagram using the pre-assembledcablesprovided.

11. Fix the light curtain transmitter and receiver to the sidewall usingtheirfixingholesandre-fittheplasticcovers.

Page 39: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.39

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.5 Installing and Aligning the Under Sash Sensor

The Under Sash Sensor consists of two components:

• Sash Sensor

• Retro-reflectiveTapeStrip

1.Fitthesensorsothatitslensisclearofobstructionsandbelowthebottomofthesashhandle.Asuitablebracketshouldbeusedsothatthesensorpositioncanbeadjustedtothebestposition.

2.Fixthe5mcableattachedtothesensortothesashwirealongitslength(wherepossible)usingcabletiesandattachtothesashcounterweighttokeepthecabletaught as the sash opens and closes. Thecableslackfromthesashcounterweightothesashcontrollerboxcanbeencasedinplasticsheathingtoguidethecableandstopitfromsnagging. Thecablecanbeattachedtothesashusingsmalldiametertrunkingifitisnotpossibletofixthecabletothesashwire. Arollerbearingorpulleyshouldbeusedwherethecablerunsoveranymetalsurface.

Page 40: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.40

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.Fixtheretro-reflectivetapestriptotheinsideofthefumehoodontheoppositeside to the sensor. Thecentreoftheretro-reflectivetapestripshouldbepositionedinlinewiththecentreofthesensorwherepossiblehowever,asmalloffsetbetweenthesensorandretro-reflectivetapestripisacceptable.Themaximumoffsetvalues,forarangeoffumehoodwidths,areillustratedinFigure40overleaf.

Page 41: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.41

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

For example, on a 1000mm wide Fume Hood the sensor can detect the retro-reflectivestripupto+/-20mmeithersideoftheUnderSashSensor’scentreline.

4. Connect to the Auto-Sash Control Unit as per the wiring diagram in Section 3.10 - Auto-Sash System Control Box - General Wiring Diagram.

3.6 Tilt Switch

AsTiltSwitchesareprovidedinavarietyofforms,specificinstallationinstructions are not possible, though the following applies in all cases:

•EnsurethattheTiltSwitchisinstalledinthecorrectorientationieis‘open’whenthefrontpanelisclosedand‘closed’whenthefrontpanelisopen.

• Connect the Tilt Switch to the Auto-Sash Control Unit as per the wiring schematic in Section 3.10 - Auto-Sash System Control Box - General WiringDiagramandtheschematicinError!Referencesourcenotfound.

•IfaTiltSwitchisnotused,linkout/connecttogethertherelevantterminalson the Auto-Sash Control Unit – see wiring schematic in Section 3.10 - Auto-Sash System Control Box - General Wiring Diagram.

Page 42: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.42

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.7 Sash Low Switch

Two types of Sash Low switch can be employed:

• Micro-Switch.

• Proximity Switch.

Inbothcases,theswitchcanbeemployedsothatthesashissignalledasbeingopenwhentheswitchis‘open’or‘closed’dependingontherequirementsoftheparticularinstallation.

3.7.1 Installing the Sash Low Micro-SwitchTheSashLowMicro-Switchisachange-overtypeswitchsocanbesetuptoeithermakeorbreakwhenthesashisclosedaslongasthecorrectterminalsareusedontheswitchforeachtypeofactivation.Typically,theswitchwouldbefittedsothattheleverengages(pushesin)whenthesashisopen.

Connect the Sash Low Micro-Switch to the Auto-Sash Control Unit as per the wiring schematic in Section 3.10 - Auto-Sash System Control Box - General Wiring Diagram and the schematic in Figure 42.

Iftheswitchisinstalledsothattheleverpushesinwhenthesashisclosed,connectthe wiring to the Auto-Sash Control Unit to terminals C and NO on the switch.

Iftheswitchisinstalledsothattheleverpushesinwhenthesashisopen,connectthe wiring to the Auto-Sash Control Unit to terminals C and NC on the switch.

Page 43: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.43

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.7.2 Installing the Sash Low Proximity SwitchThe Sash Low Proximity Switch is an inductive switch that operates when it comes withinrangeofametalplateattachedtothesash,ie.whenthesashisclosed.Dependingonthetypeofswitchselected,thecircuitcanbeeithermadeorbrokenwhenthesashisclosed.Typically,theswitchwouldbefittedsothattheswitchisclosed when the sash is open.

Connect the Sash Low Proximity Switch to the Auto-Sash Control Unit as per the wiring schematic in Section 3.10 - Auto-Sash System Control Box - General Wiring Diagram and the schematic in Figure 43.

Note 1: Inbothcases,iftheAuto-Sashcontrolisbeingretro-fittedtoafumehood,ensuresufficientfixingbracketadjustmentisprovidedtoallowforanyover-travelofthe sash.

Page 44: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.44

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.8 Sash Position Sensor

Thesashpositionsensorisfittedwithamountingbracketwhichcanbefittedinfourdifferentorientationstoenablethesensordrawwiretobecorrectlyaligned.

1.Offerupthesensortodeterminethebestfixingpositiononthefumehoodframework. Forbestresults,thesensordrawwireshouldrunstraightoutofthesensorhousinginlinewiththecablehousingextension.Incaseswherethewirehasrunatanangletothesensor,ensuretheangleisassmallaspossibletoensurethatthewiredoesnotrubonanypartofthesensorbodyincludingthecablehousingextension.

2.Markthefixingholesintheframeworkandattachthesensor. Thefixingbrackethastwo0.37mm(0.015”)diameterfixingholes.

Note:Ensurethatthesensorisinasuitablepositiontoensurethatthedrawwiredoesnotoverextend.Thesensordrawwirefullstrokeis1270mm(50”)maximum.

Page 45: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.45

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

4.Oncethesensorisfitted,manuallyopenandclosethesashtoensurethatthesensor is installed correctly and does not snag.

3. Attach the draw wire to either:

•Thesashusingasuitablysizedscrew-thedrawwirehasa9.53mm(0.38”)diameterfixingringwitha4.85mm(0.18”)diameterfixingholeinitscentre.

•Thesashcableusingacabletie,ensuringthatthefixingringdoesnotrunoverpulleys etc.

•ThecounterweightusingacabletieasshowninFigure46.

Page 46: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.46

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.9 Personnel Sensor

The Personnel Sensor is supplied as a single transmitter/receiver unit.

3.9.1 Installing the Personnel Sensor

1.UsethemountingtemplateillustratedinFigure47andfixingscrewsprovidedtofixthePersonnelSensorcentrallytothefrontofthefumehood,abovethefumehoodsash,usingthefixingholesinthebackofthesensor. RefertothetableinFigure47forthecorrectmountingholedimensionsandlocations.

ThePersonnelSensorshouldbepositionedsothatitpointsstraightdown,withthehorizontal(long)axisofthesensorparalleltothehorizontal(long)axisofthefumehood. See Figure 48.

Note: Donotmountthesensorhigherthan2500mm(98”)fromthelaboratoryfloor.

Page 47: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.47

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

2.Fittheconnectioncabletothesensorcableinsidethesensorhousing.

3.Plugthe4-wayterminalblockintotheAuto-SashControlUnit.Seewiringschematics in Section 3.10 - Auto-Sash System Control Box - General Wiring Diagram and Figure 49.

Page 48: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.48

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.9.2 Adjusting the Personnel Sensor Beam WidthItispossibletoadjustthewidthofthePersonnelSensor’sbeamtosuitdifferentwidthsoffumehoods.

To adjust the beam width:

1.Accessthetwopotentiometersunderthesensor’sfrontcoverasillustratedinFigure50below.

2. ReferringtoError!Referencesourcenotfound.below,adjustthewidthofthePersonnelSensorbeamusingthetwopotentiometerstoachievethebestfieldcoverageinfrontofthefumehoodforthehood’sparticularwidth.

Page 49: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.49

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.9.2.1 Adjusting the Beam DepthReferringtoFigure52belowadjustthesensorbytiltingthebottomofthesensortowardsorawayfromthefrontwallofthefumehood,sothatitsbeamisatoptimumanglefromthefumehoodtoachievethebestfieldcoveragefortheparticularfumehood’s height.

ThebodyofthePersonnelSensorcanbetiltedfrom0oto5ointhreeincrementstoachieve this.

Page 50: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.50

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.9.3 Personnel Sensor Dip Switch SettingsThefollowingPersonnelSensorfunctionsandparametersarecontrolledbyDipswitches, located within the Sash Light Curtain Control Box. See Section 4 - Setting Up andCalibrationofAuto-SashController.

Switches 1 and 2: Presence Timer.SelectthetimedelaybeforethePersonnelSensor‘re-learns’thebackgroundifastationaryobjectisdetected.

2,15,60or180seconddelaysareselectedbysettingassociatedswitches1and2tocombinationsofON/OFF.SeeFigure53.

Note: Thetimerwillresetifanymovementisdetectedduringthetimeperiod.

Switches 3 and 4: Pattern Depth.Setsthedistanceofthedetectionfieldawayfromthesensor.

Switches 5 and 6: Frequency.Allowstheselectionofuptofourdifferentsensorfrequencies(H;MH;MandL)iftwoormoresensorbeampathsoverlap.

Switch 7: Monitor Mode.SelectstheMonitorMode.The‘Snow’settingisonlyapplicabletooutdoorapplicationsandthereforeisnotused.Theswitchshouldbesetto‘Normal’.

Switch 8: Self Diagnostics.Ifthesensorissuspectedofnotworking,selectingswitch8to‘ON’willstartaselfdiagnostictestwhenthesensorisswitchedon.IfthetestdetectsafaulttheLEDwillcontinuallyflickerredandgreen.

Page 51: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.51

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3.10 Auto-Sash System Control Box - General Wiring Diagram

Page 52: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

4. SETTING UP AND CALIBRATION OF AUTO-SASH CONTROL UNIT

Page 53: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.53

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

BeforeStartinguptheAuto-SashController,ensurethatallAuto-SashControlSystemcomponentshavebeencorrectlyinstalledandthatallconnectionshavebeenmade.

Note 1:Checkthatthefumehoodsashandassociatedmotordrives,pulleys,operatingcables,racksandchainsmovefreelyovertheirentireoperatingrangeswithoutbindingorcatchingonadjacentcomponents.

Note 2:Checkthatallcablesaresecuredawayfrommovingcomponentsandthatnone are overstressed or damaged.

Note 3:Checkthesecurityofallelectricalconnections.

4.1 Starting Up the Auto-Sash Control Unit

TheAutoSashControllerisstartedbypressingthePowerUp/MenuEntryButtonillustrated in Figure 5 / Figure 6.

1. Switch on the Auto-Sash Control Unit. Onpowerup,theAuto-SashControlUnitperformsaself-testofitsfunctions,LEDsandtheaudiblealarmwhichtakesapproximately3seconds. Duringtheself-testperiod,allinputsandoutputsarede-activated.

2. At the end of the self-test, the Auto-Sash Control Unit will do one of two things: If the Auto-Sash Control Unit has previously been calibrated: • The Auto-Sash Controller will enter Normal Operating Mode and display its

current status or alarm condition on the display. If the Auto-Sash Control Unit has not been calibrated:

•WhentheTiltSwitchinputisopen,thedisplaywillshow‘SashDriveInhibited’.

•WhentheTiltSwitchinputisclosed,thedisplaywillshow‘Auto-SashNotConfigured’.

4 SETTING UP AND CALIBRATION OFAUTO-SASH CONTROL UNIT

Page 54: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.54

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

4.2 Calibrating the Auto-Sash Control Unit for Tiptronic Operation using the Auto-Sash Keypad

The Auto-Sash Control Unit needs to have each of its required stationary positionsdefinedforTiptronicfunctionalityasfollows:

1. Press and hold the Enterbuttonforapproximately5secondsoruntilthemainmenu appears.

TheLCDdisplaywillshowthecurrenthardwareandsoftwareversionseg.‘1.HW2.0FW1.1’.

2. Use the Downarrowtoscrolldownandselect‘2.Setup’.

3. Press Enter.

4.EnterthePassword(factorydefaultis1001)usingtheUp arrow and Enter buttons.IntheSet-Upmenu,thecontrollerwilldisplay‘KeypadTones’.

5. Use the Downarrowtoscrollthroughthemenuandselect‘Sash1Settings’.

6. Press Enter.

IntheS1(Sash1)Settingsmenu,thedisplaywillshow‘1.S1Enabled’.

7. Use the Downarrowbuttontoscrollthroughthemenuandselect‘8.S1Calibration’.

8. Press Enter.Thedisplaywillshowonscreeninstructions.Asanexample,forthisstageoftheprocess,thefollowingmessagewillbedisplayed:

“Manually move sash to center and then press the Up/Down keys to find sash down direction. Press Cancel to abort Calibration”

9.Movethesashtothemiddleofthesashopening(eg.250mmfromfumehoodbase).

10. Press the Up and Downarrowbuttonstofindwhicharrowactivatessashclosing.

11.Pressthebuttonwhichclosesthesash,thenpressEnter. This tells the controller which direction the motor needs to turn to close the sash.

12. Move the sash to the lowest position required.

13. Press Enter.

Page 55: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.55

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

14. Either: PressQuit(Mutebutton)ifthisistheonlypresetpositionrequired, or:Setthesashtothenormalworkingheight(eg.500mm)

15. Press Enteragainfortwopositionoperation.

16. Either: Press Quit (Mutebutton)ifonlytwopresetpositions(‘lowest’and‘normal’)are required, or: Setthesashtothefullopeningheight(eg.750mm).

17. Press Enteragainforthreepositionoperation. Lowest,normalandfullyopenpositionsarenowcalibrated.

18. Press EntertoquitCalibrationmode.

19. Using the Downarrowbutton,select‘BacktoSetupmenu’.

20. Press Enter.

21. Use the Downarrowbuttonandselect‘ExitandSaveChanges’.

22. Press Enter.

The Controller will now enter Run mode.

Note:EventhoughtheAuto-SashControllerdriveallowsforaccelerationanddeceleration,thesashmayslightlyoverrunthecalibrationpointsiftheAuto-Sashmotordriveissettoaveryfastspeed.Topreventthis,whencalibratingthesashpositions,allowamarginforthesashtoover-runthelowest,normalandfullyopenpositionssetduringcalibration.SeeSection4.2-CalibratingtheAuto-SashControlUnitforTiptronicOperationusingtheAuto-SashKeypadabove.

For example:

Tocalibratethelowerstopheightwiththisallowance,closethesashfullyandthenopen5mm(0.197”)beforecalibrating.

Tocalibratethenormalstopheightwiththisallowance,openthesashtothenormalheightandthenlowerby5mm(0.197”)beforecalibrating.

4.3 Calibrating the Sash Lock Function

TherequiredSashLockpositionisdefinedasfollows:

1. Press and Hold the Enterbuttonforapproximately5secondsoruntiltheMain menu appears.

2.ThedisplaywillshowthecurrentHardwareandSoftwareversionseg.“1.HW2.0FW1.1”

Page 56: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.56

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3. Use the Downarrowtoscrolldown,select“2.Setup”andpressEnter.

4.EnterthePassword(Factorydefaultis1001)usingtheUp arrow and Enter buttons.

5.IntheSetupmenu,thecontrollerwilldisplay“KeypadTones”

6. Use the Downarrowtoscrollthroughthemenuandselect“Sash1Settings”andpress Enter.

7. In the S1 Settingsmenuthedisplaywillshow“1.S1Enabled”

8. Use the Downarrowtoscrollthroughthemenuandselect“11.S1AutoLock”andpress Enter.

9.SettoEnabledandpressEnter.

10. Use the Downarrowtoscrollthroughthemenuandselect“11b.S1LockAlarm”and press Enter.

11. Use the Up and Down arrows to set the required alarm delay time and press Enter.

12. Use the Downarrowtoscrollthroughthemenuandselect“11c.S1LockPosCalibration”.

13. Press Enter.

The display will show on screen instructions. As an example, for this stage of the process, the following message will be displayed:

“Lock the sash then move the sash to the lowest lock position. Enter to accept. Cancel to abort”

14. Open the sash to the Lock position and press Enter.

15.Ifthepositionchosenistoolow(lowerthanthecalibratedBottom,MiddleorToppositions)“Badposition-tryagain”willbedisplayed.

16.Ifthechosenpositionisacceptable,thelockpositionwillbeset.

17. Use the Downbuttonandselect”BacktoSetupmenu”andpressEnter.

18. Use the Downbuttonandselect”ExitandSaveChanges”andpressEnter.

19. The Auto-Sash Controller will now go to RUN mode.

Page 57: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.57

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

4.4 Calibrating the Safety Light Curtain

Adjust the range and sensitivity of the Safety Light Curtain as follows:

1.EnsurethattheareabetweentheSafetyLightCurtaintransmitterandreceiverisfreefromobstructions.

2. EnsurethatthelightbeampathsarenotbrokenandthatthePersonnelSensorisdisabledsothatthesashdoesnotAuto-ClosewhilsttheSafetyLightCurtainisbeingcalibrated.

Note:IfitisnotpossibletodisablethePersonnelSensor,removethemotor/clutchconnectorfromtheAuto-SashControlUnit.ThiswillactivatetheSashControlAlarmaftertheselectedtimedelayandtheresetbuttonwillneedtobepressedoncetheSafetyLightCurtainhasbeencalibrated.

3.RemovethecoverfromtheSafetyLightCurtainControlBox.

Page 58: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.58

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

4. Power up the Auto-Sash Control Unit.

5.CheckthattheSafetyLightCurtainTransmitterandReceiverareconnected correctly and are operational. TheredPowerOnLEDshouldnowbevisibleonthereceiver. See Figure 55.

6.EnsurethatSwitch1(sw1)(seeFigure54)issettothecorrectrange. See table overleaf:

Page 59: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.59

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

6. EnsurethatSwitch2(sw2)(seeFigure54)issettoOff(OperationMode).

7.EnsurethatSwitch3(sw3)issettoOn(NormalOperation) Abuzzercanbesettosoundwhenthebeamisbrokentoaidsettingupthebeam.Toenablethebuzzer,setSwitch4(sw4)toOn. ThesignalLEDalsoindicatesbeamstatus;greenifthedetectionareaisclearandredifanobjectortheoperatorisdetected.

8.Adjustthegainpotentiometerfullycounter-clockwise(tominimum),sothatthetransmitterandreceivercannot‘see’eachotherie.thereceivercannolongerdetectthebeamsentfromthetransmitter.

9.Slowlyadjustthegainpotentiometer(seeFigure54above)clockwise(tomaximum)untilthetransmitterandreceivercan‘see’eachotherie.thereceivercandetectthebeamsentfromthetransmitter.

10.SetSwitch3(sw3)toOfftostorethesettings.

11. Turn the gain potentiometer approximately 5 degrees clockwise to ensure sufficientgain.

12.Useaglassobject(eg.abeaker)tocheckthatsuchobjectscanbreakthebeamacrosstheentireopenfaceofthefumehood. Iftheglassobjectdoesnotbreakthebeamoverthefullarea,slowlyadjustthegainpotentiometercounter-clockwiseuntilinsertingtheglassobjectbreaksthebeam.

13.DisablethebuzzerifrequiredbysettingSwitch4(sw4)toOff. Ifthesashlightbeamsonlyintermittently‘see’glassobjects,ensuresw1issettoOff(shortrange)andrepeatsteps5to14.

Page 60: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.60

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

If adjusting the Gain Potentiometer does not allow the transmitter and receiver to ‘see’ each other and always indicates a broken beam, inspite of a clear path between the light beams:

Either:Switchsw2On(Diagnosticmode).Afterapproximately15secondstheRXERRandTXERRLEDs(seeFigure54above)willindicateifthereisafaultybeamie:

•RXERRRedLEDOn–Faultyreceiver.•TXERRRedLEDOn–Faultytransmitter.

Or:Cleanorreplacethefaultycomponentandsetsw2backtoOff(Operationmode).

4.5 Calibrating the Tilt Switch

NousercalibrationoftheTiltSwitchispossiblehowever,ifaTiltSwitchisfitted,ensurethatthetoppanelofthefumehoodisclosedbeforeswitchingontheAuto-Sash Control Unit.

4.6 Calibrating the Sash Position Sensor

NousercalibrationoftheSashPositionSensorispossible,theunitsimplyprovidessash position data to the Auto-Sash Controller relative to its starting position.

4.7 Calibrating the Personnel Sensor

Thesensorisself-calibratingandneedsnoadjustment.Toensureoptimumperformance,ensurethedetectionfieldinfrontofthefumehoodisclearfromobjectsandpersonnelwhenpoweringuptheAuto-SashControlUnit.

Thesensorwillbefullyoperationalapproximately10secondsafterpowerup.TheLEDonthesensor’sfaciawillbegreenifthedetectionareaisclearandredifanobjectortheoperatorisdetected.

To check that the sensor is operational, listen to hear whether the internal relay energisesandde-energises(clicks)whenstandinginfrontofandthenwalkingawayfromthefumehood.

4.8 Calibrating the Sash Low Switch

NousercalibrationoftheSashLowSwitchispossible.

Page 61: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

5. SYSTEM MENUS

Page 62: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.62

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

5 SYSTEM MENUSTheSystemMenusallowtheOperatortodefineparametersaffectingtheoperationoftheAuto-SashControllerandassociatedsystems.

TheoverallstructureoftheSystemMenusandroutestoaccessindividualmenuitemsareillustratedintheflowchartoverleaf:

Page 63: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.63

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

Page 64: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.64

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

5.1 Accessing System Menus

To access the System Menus, press and hold the Keypad Enterbuttonforapproximately 5 seconds or until the Main/Setup menu is displayed.

1. Navigate the Main/Setup menu using the Up and Down arrow buttonsonKeypad.

2. Select the required menu

3. Press Enterbutton

4. Navigate the chosen menu using the Up and Downarrowbuttons. Select from the following System Menus: • Keypad Tones. • Sash 1 Settings. • BMS Input Settings. • BMS Output Settings.

5.SelectamenuentrybypressingtheEnterbutton.

6.Selectamenuentryvalue/parameterbyusingtheUp and Downarrowbuttonstodisplay the required value/parameter.

7. Press Enterbutton.

8. Press Cancelbuttontoexitmenuentryandreturntomenu.

9. To return to Main/Setupmenu,scrolldowntothebottomofthecurrentmenuand select <Back to Setup Menu.

10. Press Enterbutton.

11. Scroll down through Main/Setupmenuandselectoneof: •ExitandAbandonChanges • Exit and Save Changes

Page 65: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.65

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

5.2 Navigating the System Menus

5.2.1 Keypad TonesEnablesordisablestonesoundingwhenakeyispressedontheKeypad.Select from: 1. Enabled. 2. Disabled.

5.2.2 Sash 1 Settings Menu

1. S1 Enabled PermanentlyenablesordisablestheAuto-SashControllermotordrive.Select from: 1. No. 2. Yes.

2a. S1 Close DelaySetsthetimedelaybeforethesashcloses(0to3600seconds).

2b. S1 Close AlarmSetsthetimedelaybetweentheCloseAlarmactivatingandthesashclosing(0to10seconds).

3. S1 Auto-OpenEnablesordisablesAuto-OpenfeatureorsetstoAuto-CloseOnly.

3a. S1 Open DelaySetsthetimedelaybeforethesashopens(0to10seconds).

3b. S1 Open AlarmSetstimedelaybetweenOpenAlarmactivatingandthesashopening(0to10seconds).

4. S1 Motor SpeedSetsmotordrivespeed(0to100%).

5. S1 Tilt SwitchSets input activation.Select from: 1. Normally Open. 2. Normally Closed. 3. Disabled.

6. S1 Light CurtainSets input activation.Select from: 1. Disabled. 2. N/CContact(NormallyClosed). 3. N/OContact(NormallyOpen).

Page 66: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.66

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

7. S1 Personnel (Sensor)Sets input activation.Select from: 1. Disabled 2. N/CContact(NormallyClosed) 3. N/OContact(NormallyOpen)

8. S1 CalibrationCalibratessashpositionsensorandmotordirection.SeeSection4.2.

9. S1 TiptronicEnablesordisablestheTiptronicfeature.Select from: 1. Enabled 2. Disabled

10. S1 MaxTiptimeSetsTiptronictouchsensitivity(0to1500ms–0to1.5seconds)

Note: Thisistheminimumtimethesashmusttraveltobeoperatedmanually.ManualmovementforlessthanthespecifiedtimewillcausethesashtoAuto-Drivetothecalibratedposition.

11. S1 Auto LockEnablesordisablestheAuto-Lockfeature(seeSection4.2).Select from: 1. Enabled 2. Disabled

11b. Lock AlarmSetsthealarmdelaytime-timebeforealarmactivatesifsashremainslocked(1to30minutes).

11c. Lock Pos CalibrationCalibratestheLockposition(seeSection4.3).

5.2.3 BMS Input Settings Menu

1. Fire Alarm Select from: 1. Disabled 2. Input 1 Open 3. Input 1 Closed

2. Open SashSelect from: 1. Disabled 2. Input 1 Open 3. Input 1 Closed

Page 67: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.67

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

3. Close SashSelect from: 1. Disabled 2. Input 1 Open 3. Input 1 Closed

4. EV Open SashSelect from: 1. Disabled 2. Input 1 Open 3. Input 1 Closed

5. EV Close SashSelect from: 1.Disabled 2. Input 1 Open 3. Input 1 Closed

6. Foot SW OpenSelect from: 1. Disabled 2. Input 1 Open 3. Input 1 Closed

7. Foot SW ClosedSelect from: 1. Disabled 2. Input 1 Open 3. Input 1 Closed 5.2.4 BMS Output Settings Menu

1. Alarm SounderEnablesordisablesRemoteAlarmSounder.

2. Analogue VoltsSelectsanalogueoutputvoltageforsashpositionindication.Select from: 1.0-10V 2.1-10V 3.0-5V 4.1-5V

3. Analogue CurrentSelectsanalogueoutputcurrentforsashpositionindication.Select from: 1. 0 - 20mA 2. 4 - 20mA

Page 68: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.68

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

4. Relays 1 to 4 (RLY 1; RLY 2; RLY3; RLY4)For each relay, select from:1. Always Open2. Always Closed3. Open on Sash Alarm4. Close on Sash Alarm5.OpenonObstructionDetected6.CloseonObstructionDetected7.OpenonUser(Detected)8.CloseonUser(Detected)9. Open on Sash Open10. Close on Sash Open11. Open on Sash Locked12. Close on Sash Locked

Page 69: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

6. OPERATIONS

Page 70: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.70

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

6.1 Control Philosophy

Whenallcomponentsareinstalled,setupandcalibratedcorrectly,theAuto-SashControllerwillclosethesashafterapresetdelay,providedtheareainfrontofthefumehoodisclear.

The following additional protocols also apply:

•iftheuserreturnstothefumehoodduringthedelayperiod,thecontroltimerwillstopandresetoncetheuserhasleftthefumehoodagain.•iftheuserreturnstothefumehoodoncethesashisclosing,thedrivewillstopassoonasit‘sees’theuser.•ifanobstructionisdetectedunderthesash,theAuto-SashControllerwillnotopenorclosethesashandwillgiveavisualalarmtoindicateafaultcondition.•ifanobstructionisdetectedwhilstthesashisclosing,thesashdrivewillstopandgiveavisualindicationtoindicateafaultcondition.Thefaultalarmcanberesetbypressingtheresetbutton.•oncethesashisclosingatimerisinitiated,ifthesashdoesnotclosewithinthespecifiedtime,thesashdrivewillstopandwillgiveavisualalarmtoindicateafaultcondition.•wheneverauserispresentinfrontofthefumehood,thesashcanbeoperatedbyhand.

6.2 Modes of Operation

The Tiptronic Auto-Sash Controller has the following modes of operation:• User Present.• User Not Detected.•Auto-Open(UserConfigurable).•BuildingManagementSystem(BMS)Inputs.• Sash Lock.

6 OPERATION

Page 71: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.71

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

6.2.1 User PresentWhen the user is detected the sash can be moved in the following ways:• Manually

IftheTiptronicfeatureisenabled,thesashcanbeopenedorclosedmanuallyaslongasitiskeptmovingformorethanauserdefinedperiodoftimerangingfrom0to1500ms(0to1.5seconds).

Note:Thisistheminimumtimethesashmusttraveltobeoperatedmanually.MovementforlessthanthespecifiedtimewillleadtothesashAuto-Drivingtotheappropriatecalibratedposition.

• Tiptronic

Ifthesashismovedmanuallyforlessthanitsuserdefinedperiodoftime(0to1.5seconds),thesashwilldrivetoitsuserdefinedposition.

•Up/DownPushbuttons

Ifthesebuttonsarepressed,thesashwilldrivetoitsprogrammedposition.

•OptionalFootswitch(OpenandCloseoptions)

Ifoperatedthesashwilldrivetoitsprogrammedfullyopenorfullyclosedposition

6.2.2 User Not DetectedWhen the user is not detected, the sash will Auto-Close to its lowest programmed position following a pre-set delay time unless:

•TheSafetyLightCurtainbeamisbroken(ie.anobjectisdetectedinthesashopening).• The Auto-Sash Controller detects no sash movement whilst the motor is operating.

6.2.3 Auto-Open(UserConfigurable)When enabled, the sash will automatically open when the Operator returns following a pre-set time delay:•ifthesashwasAuto-Closed,thesashwillautomaticallyopenbacktothesamepositionthesashwasinbeforeitautomaticallyclosed.

•ifthesashwasclosedbyanyothermethod(manually,Tiptronically,pushbutton,BMSetc.),thesashwillopenautomaticallyandreturntoitscalibratedposition.

The sash will not open automatically if:

•anobjectisdetectedinthesashopening.• the controller detects no sash movement whilst the motor is operating.

Page 72: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.72

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

6.2.4 Sash LockThesashcanbesettoLock(Auto-Sashdriveinhibited)whenthesashissettoapresetposition(egfullyopen).

Anaudibleandvisualalarmwillactivatefollowingapresetalarmtimedelaytoremindthe user to lower the sash.

Thisfunctionisusedwhenthesashneedstoremainintheopenpositionforaprolongedperiodoftimetoallowloadingorsettingupofequipmentinsidethefumehood. 6.2.5 BuildingManagementSystem(BMS)InputsThe sash can be set to operate from BMS inputs as follows:• Open/CloseThesashwillonlyAuto-Driveopen/closedatthestart/endoftheday,followinganinputtotheAuto-SashControllerfromaBMS,iftheOperatorisnotpresentandthenoobjectisdetectedinthesashopening.•EV(Emergency)Open/CloseEVaudiblealarmtonewillsoundwhilsttheEVinputisactive.ThesashwillonlyAuto-Driveopen/closeiftheOperatorisnotpresentandnoobjectis detected in the sash opening.InEVmode,thesashcanbemanuallymovediftheOperatorisdetected.ThesashwillthenAuto-Driveopen/closedagain,iftheEVinputisstillactive,whentheOperator is no longer detected.• Fire Alarm CloseThefirealarmaudiblealarmtonewillsoundwhilstthefirealarminputisactive.ThesashwillonlyAuto-Driveopen/closediftheuserisnotpresentandnoobjectisdetected in the sash opening.InFireAlarmClosemode,thesashcanbemanuallymovediftheOperatorisdetected.ThesashwillthenAuto-Driveopen/closedagain,iftheEVinputisstillactive,when the Operator is no longer detected.

6.3 Operation and Alarm Indication

6.3.1 RUN ModeIn RUN mode, the display will show the current status:

“User present” Theuserisdetected,thesashdriveisinhibited,Tiptronicandmanualdriveareenabled.

“XX Seconds to close”Theuserisnotdetected;thesashwillclosefollowingthecountdowndelaytime.

“Sash Closing”Displayedwhenthesashisdrivingclosed(Tiptronic,Auto-Closeorpushbutton).

Page 73: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.73

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

“Sash Closed”Whentheuserisnotpresentandthesashhasclosed.(keypadbacklightdimmed).

“XX Seconds to open”Whentheuserisdetected;thesashwillopenfollowingthecountdowndelaytime.(Auto-Openenabled)

“Sash Opening”Displayedwhenthesashisdrivingopen(Auto-Open,Tiptronicorpushbutton).

“Operator Manual Move”Displayedifthesashismanuallyopenedorclosed.

“Sash Inhibited”DisplayediftheTiltswitchinputisopen,driveisinhibiteduntiltheTiltswitchisclosed.

“Sash Disabled”Displayedifthesashdrivehasbeendisabledinthemenusettings.

“Sash Locked”DisplayedifthesashisatorabovethecalibratedLockposition.

IftheAuto-Openfeatureisenabledinthemenuandisswitchedonfromthekeypad,an“*”(asterix)willbedisplayedinthelefthandcornerofthescreen,eg.“*Userpresent”.

ToenableAuto-OpenpressandholdtheMutebuttonfor5secondsoruntilthe“*”symbolisdisplayed.

TodisableAuto-OpenpressandholdtheMutebuttonfor5secondsoruntilthe“*”symboldisappears.

Auto-Opencanbesetto‘Disabled’,‘Enabled’,‘EnabledifAuto-Closed’.

Note:Ifsetto‘EnabledifAuto-Closed’,thesashwillonlyAuto-OpenifthesashhasAuto-Closed.

ClosingthesashmanuallyorbyusingthepushbuttonswillinhibittheAuto-Openwhentheuserreturnstothefumehood.

6.3.2 Alarm ConditionsInanAlarmConditionoriginatingfromwithinthefumehood,thedisplaywillshowthecurrent alarm status:

“Obstruction Detected”Displayedifthesashbeamisbrokenwhenthesashstartstoclose–removeobstructionandpresstheMutebuttontoresetthealarm.

Page 74: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.74

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

1. “Sash Fault”Displayedifthesashdoesnotcloseoropenwhenexpected-pressMutetoresetalarm.Checksashdriveforfaults.

2. “Remove Sash Lock/Cancel to Reset”Displayedifthesashisatorabovethecalibratedlockpositionforlongerthanthepreset alarm time - press Mute or lower the sash to reset the alarm.

Following an input from the BMS, the display will show one of the following messages:

1. “Fire Alarm”Displayedalongwiththenormaltext(eg.“FireAlarm,UserPresent”)whenthefirealarm input is activated.

Inthissituation,thesashwilldriveclosediftheuserisnotdetectedandthesashisclear.

Thesashcanbeoperatedmanuallyiftheuserisdetected(Auto-Open,Tiptronicandpushbuttonsaredisabled).

2. “BMS open”DisplayedalongwiththenormaltextiftheuserisnotdetectedwhentheBMSOpeninput is activated.

Inthissituation,thesashwilldriveopeniftheuserisnotpresentandthesashisclear.Thesashcanbeoperatednormallyiftheuserisdetected(Auto-Open,Tiptronicandpushbuttonsareactive).

3. “BMS close”WhentheBMScloseinputisactivated,thesashwilldriveclosediftheuserisnotpresent and the sash is clear.

Thesashcanbeoperatednormallyiftheuserisdetected(Auto-Open,Tiptronicandkeypadpushbuttonsareactive).

4. “Emergency (up)”DisplayedalongwiththenormaltextiftheuserisnotdetectedwhentheEVOpeninput is activated.

Inthissituation,thesashwilldriveopeniftheuserisnotpresentandthesashisclear.Thesashcanbeoperatedmanuallyiftheuserisdetected(Auto-Open,Tiptronicandpushbuttonsaredisabled).

5. “Emergency (dn)”DisplayedalongwiththenormaltextiftheuserisnotdetectedwhentheEVCloseinput is activated.

Page 75: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.75

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

Inthissituation,thesashwilldriveclosediftheuserisnotpresentandthesashisclear.

Thesashcanbeoperatedmanuallyiftheuserisdetected(Auto-Open,Tiptronicandpushbuttonsaredisabled).

6. “Sash Opening”DisplayediftheSashOpenFootswitchhasbeenoperatedwhilsttheOperatorispresent.

7. “Sash Closing”DisplayediftheSashCloseFootswitchhasbeenoperatedwhilsttheOperatorispresent.

Page 76: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

7. TROUBLESHOOTING

Page 77: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.77

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

7.1 Faults and Alarms

Sash does not Auto-Drive

a)Checkthatthereismainspowertothecontrolunitandalloftheconnectionsarecorrect.

b)CheckthattheSash1SettingsMenuentryS1EnabledissettoEnabled(“SashDisabled”willbedisplayedinrunmodeifsettoDisabled).

c)ChecktheTiltSwitchisconnectedorlinkedoutatthecontrolboxterminals.

Motor drives but clutch does not engageCheck that the clutch is connected.

Fault LED is illuminatedCheckthattheareabetweenthesashsensortransmitterandreceiverisclearandthat the sensors are aligned – power up again to reset.

Sash drive does not stop when sash is closedCheck that the Sash Low switch is connected and operational.

Sash drive is slow or fastAdjusttheSash1SettingsMenuentryS1MotorSpeed(0–100%).

7 TROUBLESHOOTING

Page 78: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

8. LIMITATION OF WARRANTYAND LIABILITY

Page 79: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.79

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

Sellerwarrantsthatthisproduct,undernormaluseandserviceasdescribedintheoperator’smanualshallbefreefromdefectsinworkmanshipandmaterialforaperiodoftwelve(12)months,orthelengthoftimespecifiedintheoperator’smanual,fromthedateofshipmenttothecustomer.

This limited warranty is subject to the following exclusion:

1.Batteriesandcertainothercomponentswhenindicatedinspecificationsarewarrantedforaperiodof90daysfromthedateofshipmenttothecustomer.

2. With respect to any repair services rendered, Seller warrants that the parts repairedorreplacedwillbefreefromdefectsinworkmanshipandmaterial,undernormaluse,foraperiodof90daysfromthedateofshipmenttothecustomer

3.Sellerdoesnotprovideanywarrantyonfinishedgoodsmanufacturedbyothers.Onlytheoriginalmanufacturer’swarrantyapplies.

4.UnlessspecificallyauthorisedinaseparatewritingbySeller,Sellermakesnowarrantywithrespectto,andshallhavenoliabilityinconnectionwith,anygoodswhichareincorporatedintootherproductsorequipmentbytheBuyer.AllgoodsreturnedunderwarrantyshallbeattheBuyer’sriskofloss,Seller’sfactoryprepaid,andwillbereturnedatSeller’sriskofloss,Buyer’sfactoryprepaid.

TheforegoingisINLIEUOFallotherwarrantiesandissubjecttotheconditionsandLIMITATIONSstatedherein.NOOTHEREXPRESSORIMPLIEDWARRANTYOFFITNESSFOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.

THEEXCLUSIVEREMEDYOFTHEUSERORPURCHASER,ANDTHELIMITOFLIABILITYOF SELLER FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES IN CONNECTION WITH THIS PRODUCT (INCLUDING CLAIMS BASED ON CONTRACT NEGLIGENCE, STRICT LIABILITY,OTHERTORT,OROTHERWISE)SHALLBETHERETURNOFTHEPRODUCTTO THE FACTORY OR DESIGNATED LOCATION AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF THE SELLER, THE REPAIR OR REPLACEMENT OFTHEPRODUCT.INNOEVENTSHALLSELLERBELIABLEFORANYSPECIAL,INCIDENTAL OR CONSEQUENTIAL DAMAGES.SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING, REASSEMBLY OR REINSTALLATION COSTS OR CHARGES. NO ACTION, REGARDLESS OF FORM, MAY BE BROUGHT AGAINST THE SELLER MORE THAN ONE YEAR AFTER THE CAUSE OF ACTION HAS ACCRUED.

ThepurchaserandallusersaredeemedtohaveacceptedthetermsofthisLIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limitedwarrantyofSeller.ThisLIMITATIONOFWARRANTYANDLIABILITYmaynotbeamendedormodifiednormayanyofitstermsbewaivedexceptbyawritingsignedbyanauthorisedrepresentativeoftheSeller.

8 LIMITATION OF WARRANTY AND LIABILITY

Page 80: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

9. AUTO-SASH CONTROLLER SERIES SPECIFICATION

Page 81: FUME HOOD AUTO-SASH CONTROLLER - TEL...• Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, temporary conductivity caused by condensation must

Version 01 / p.81

FUME HOOD AUTO-SASH CONTROLLER Installation & Operating Manual

9 AUTO-SASH CONTROLLER SERIES SPECIFICATION