G1888-90009 Headspace Sampler

Embed Size (px)

DESCRIPTION

G1888-90009 Headspace Sampler

Citation preview

  • Agilent Technologies

    Agilent G1888 Network Headspace Sampler

    Site Prep and Installation Guide

    siteprep.book Page 1 Tuesday, February 17, 2004 9:18 AM

  • 2 Site Preparation and Installation Guide

    Notices Agilent Technologies, Inc. 2004

    No part of this manual may be reproduced in any form or by any means (including elec-tronic storage and retrieval or translation into a foreign language) without prior agree-ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

    Manual Part NumberG1888-90009

    EditionFirst edition, March 2004

    Printed in US

    Agilent Technologies, Inc.2850 Centerville Road Wilmington, DE 19808-1610 USA

    Warranty

    The material contained in this docu-ment is provided as is, and is sub-ject to being changed, without notice, in future editions. Further, to the max-imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par-ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec-tion with the furnishing, use, or per-formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.

    Safety Notices

    CAUTION

    A CAUTION notice denotes a haz-ard. It calls attention to an operat-ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

    WARNING

    A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per-formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi-tions are fully understood and met.

    AcknowledgementsSwagelok is a registered trademark of the Swagelok Company.

    Microsoft and Windows are registered trademarks of the Microsoft Corporation.

    siteprep.book Page 2 Tuesday, February 17, 2004 9:18 AM

  • Site Preparation and Installation Guide 3

    Manual Overview

    This manual contains four chapters and one appendix.

    1 Introduction

    This chapter gives an overview of the manual and contains safety information.

    2 Site Preparation

    This chapter provides the requirements for preparing the site before installing the Headspace Sampler. It covers the requirements for temperature and humidity, space, power, gas, and venting.

    3 Installation

    Installation describes how to install the Headspace Sampler. It includes instructions for unpacking the unit, checking the contents, connecting the power and gas supplies, making connections to an integrator or Agilent data system, and connecting the unit to a GC.

    4 Running Checkout Samples

    This chapter provides information on preparing and running the checkout sample to confirm that the sampler is operating correctly.

    A Connecting a Swagelok Fitting

    This appendix describes proper Swagelok connections.

    siteprep.book Page 3 Tuesday, February 17, 2004 9:18 AM

  • Site Preparation and Installation Guide 4

    siteprep.book Page 4 Tuesday, February 17, 2004 9:18 AM

  • Site Pr

    Contents

    1 Introduction

    About

    Ma

    Ele

    MaGasGen

    Safety

    InfoSymEleSou

    Fuses

    Cleani

    Recycl

    2 Site Preparation

    Introdu

    siteprep.book Page 5 Tuesday, February 17, 2004 9:18 AMTemperature and Humidity Requirements 21

    Performance range 21Optimum performance 21ep and Installation Guideny internal parts of the instrument carry dangerous voltages 11ctrostatic discharge is a threat to instrument electronics 11ny parts are dangerously hot 12es 12eral warnings 13

    and Regulatory Certifications 14

    rmation 14bols 15

    ctromagnetic compatibility 15nd Emission Certification for Federal Republic of Germany 16

    17

    ng 17

    ing the Product 17

    ction to Site Preparation 20Video supplements 10

    Important Safety Warnings 11this Manual 105

  • 6Space Requirements 22

    Power Requirements 23

    LinPow

    SupReg

    Ventin

    Sampl

    3 Installation

    Install

    Unpac

    Place t

    Conne

    Conne

    FloTyp

    MPEPC

    Contro

    Con

    Set

    4 Running Checkout Sa

    Runnin

    siteprep.book Page 6 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    C connections 42 connections 46

    l Setup 50

    nect the control cables 50

    communication options 52

    mples

    g Checkout Samples 56Set the flow mode on the Headspace Sampler 40Connect on/off valves to the gas input fittings 41ply tubing 25ulators 25

    g Requirements 28

    er Considerations 29

    ation 32

    k the Unit and Check the Contents 33

    he Unit in Position 34

    ct the Power Cord 35

    ct the Plumbing and Transfer Line 37

    w modes 37ical installation procedures 39Gas Requirements 24

    External Plumbing 25e voltage 23er line cord 23

  • Site Prep and Installation Guide

    Prepare the checkout sample 56Run the sample 576850/6890 GC checkout method 58

    A Connecting a Swagel

    Prepar

    ObjMa

    Proced

    siteprep.book Page 7 Tuesday, February 17, 2004 9:18 AM7ok Fitting

    ation 70

    ective 70terials needed 70

    ure 71

  • 8siteprep.book Page 8 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

  • Agilent G1888 Network Headspace SamplerSite Prep and Installation Guide

    1Introduction

    About t

    Vide

    Importa

    Man

    Elec

    Man

    Gas

    Gen

    Safety

    Info

    Sym

    Elec

    Sou

    Fuses

    Cleanin

    Recycli

    This csafety

    siteprep.book Page 9 Tuesday, February 17, 2004 9:18 AM9Agilent Technologies

    his Manual 10

    o supplements 10

    nt Safety Warnings 11

    y internal parts of the instrument carry dangerous voltages 11

    trostatic discharge is a threat to instrument electronics 11

    y parts are dangerously hot 12

    es 12

    eral warnings 13

    and Regulatory Certifications 14

    rmation 14

    bols 15

    tromagnetic compatibility 15

    nd Emission Certification for Federal Republic of Germany 16

    17

    g 17

    ng the Product 17

    hapter gives an overview of the manual and provides information.

  • 10

    Introduction

    About this Manual

    This manual explains the requirements and procedures for installThe mperforpreparincludvariou

    For opUser I

    Video supplements

    Click tvideo

    siteprep.book Page 10 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guideing the Agilent Technologies G1888A Headspace Sampler. anual is intended for Agilent Service personnel ming the installation. It includes basic information for ing a site and for installing the Headspace Sampler, ing attaching it to the GC and setting flow rates for s inlet configurations.

    erating directions, see the Agilent Headspace Sampler nformation CD ROM.

    his icon when it appears throughout the manual to view supplements to certain procedures.

  • Introduction

    Site Pr

    Important Safety Warnings

    Before moving on, there are several important safety notices that yoSampl

    Many internal parts

    If the

    All

    Electrostatic dischar

    The pr

    voltageunless

    WARNING If the preplac

    siteprep.book Page 11 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 11damaged by electrostatic discharge. Do not touch any of the boards unless it is absolutely necessary. If you must handle ge is a threat to instrument electronics

    inted circuit (PC) boards in the instrument can be

    be difficult to accidentally make contact with dangerous s. Unless specifically instructed to, never remove a cover

    the heated zones are turned off.

    ower cord insulation is frayed or worn, the cord must be ed. Contact your Agilent service representative. The internal wires and cables connected to these boards

    The wires for any heater (such as the oven)

    WARNING All these parts are shielded by covers. With the covers in place, it shouldn:

    electronics boards in the instrumentpower switch is off, potentially dangerous voltages exist on:

    The wiring between the instrument power cord and the AC power supply, the AC power supply itself, and the wiring from the AC power supply to the power switch

    With the power switch on, potentially dangerous voltages also exist ou should always keep in mind when using the Headspace er.

    of the instrument carry dangerous voltages

    instrument is connected to a power source, even if the

  • 12

    Introduction

    them, wear a grounded wrist strap and take other antistatic precautions. Wear a grounded wrist strap any time you must remove the electronics cover.

    Many parts are dang

    Many enouglimited

    The

    The

    You shtempefirst setempeIf you and w

    Gases

    Do

    Do Samto t

    Weainju

    Fasimm

    Storele

    siteprep.book Page 12 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide Do not put gas cylinders in the path of a hot air vent (including a GC oven exhaust).not use flammable carrier gases.

    not use hydrogen as a carrier gas in the Headspace pler. Hydrogen creates a potential explosion hazard due

    he venting of gases during operation and standby.

    r eye protection when using compressed gas to avoid eye ry.

    ten all compressed gas cylinders securely to an ovable structure or permanent wall.

    re and handle compressed gases in accordance with vant safety codes.instrument that you will be maintaining before you begin working on it.

    WARNING Be careful when working behind the instrument. During cool-down cycles, the instrument emits hot exhaust which can cause burns.erously hot

    parts of the instrument operate at temperatures high h to cause serious burns. These parts include but are not to:

    carousel and its contents

    sample probe/loop assembly

    ould always cool these areas of the instrument to room rature before working on them. They will cool faster if you t the temperature of the heated zone to room

    rature. Turn the zone off after it has reached the setpoint. must perform maintenance on hot parts, use a wrench ear gloves. Whenever possible, cool the part of the

  • Introduction

    Site Prep and Installation Guide

    General warnings

    Perform periodic leak checks on supply lines, fittings, and pneumatic plumbing to prevent a potentially hazardous con

    To ato ldiscsolu

    siteprep.book Page 13 Tuesday, February 17, 2004 9:18 AM13

    dition.

    void a potential shock hazard when using liquid solution ocate leaks, turn the main power switch off and onnect the main power cord. Be careful not to spill leak tion on electrical leads.

  • 14

    Introduction

    Safety and Regulatory Certifications

    The Headspace Sampler conforms to the following safety standa

    Can

    CSAUL

    Inte

    Eur

    The inElectrInterfe

    CIS

    This ISISM es

    The insystem

    Information

    The AfollowclassifCatego

    This urecognthe insmanufbe impHeadsfrom aoperat

    siteprep.book Page 14 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation GuideM device complies with Canadian ICES-001. Cet appareil t conforme a la norme NMB001 du Canada.

    strument is designed and manufactured under a quality registered to ISO 9001.

    gilent Technologies Headspace Sampler meets the ing IEC (International Electro-technical Commission) ications: Safety Class I, Transient Overvoltage ry II, Pollution Degree 2.

    nit has been designed and tested in accordance with ized safety standards and is designed for use indoors. If trument is used in a manner not specified by the acturer, the protection provided by the instrument may aired. Whenever the safety protection of the Agilent pace Sampler has been compromised, disconnect the unit ll power sources and secure the unit against unintended ion. IEC/EN 61326

    AUS/NZ rds:

    adian Standards Association (CSA): C22.2 No. 1010.1

    /Nationally Recognized Test Laboratory (NRTL): 61010A1

    rnational Electrotechnical Commission (IEC): 610101

    oNorm (EN): 610101

    strument conforms to the following regulations on omagnetic Compatibility (EMC) and Radio Frequency rence (RFI):

    PR 11/EN 55011: Group 1, Class A

  • Introduction

    Site Prep and Installation Guide

    Refer servicing to qualified service personnel. Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard.

    Symbols

    Warniduringinstrusafety

    Electromagnetic com

    This dOpera

    1 Thi

    2 Thi

    See acinform

    Indica

    Indica

    Indica

    Indica

    Indica

    siteprep.book Page 15 Tuesday, February 17, 2004 9:18 AM15interference that may cause undesired operation.patibility

    evice complies with the requirements of CISPR 11. tion is subject to the following two conditions:

    s device may not cause harmful interference.

    s device must accept any interference received, including

    companying instructions for more ation.

    tes a hot surface.

    tes hazardous voltages.

    tes earth (ground) terminal.

    tes explosion hazard.

    tes electrostatic discharge hazard.instrument. Agilent Technologies assumes no liability for the customers failure to comply with these requirements.ngs in the manual or on the instrument must be observed all phases of operation, service, and repair of this ment. Failure to comply with these precautions violates standards of design and the intended use of the

  • 16

    Introduction

    If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try one or more of the

    1 Rel

    2 Mov

    3 Pludevcirc

    4 Mak

    5 Makdev

    6 Conexp

    7 ChaTecequ

    Sound Emission Cert

    (Type

    Schall

    Schall(Typp

    siteprep.book Page 16 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    test).

    druckpegel

    druckpegel LP < 70 dB(A) nach DIN-EN 27779 rufung).Sound pressure

    Sound pressure Lp < 70 dB(A) according to DIN-EN 27779following measures:

    ocate the radio or antenna.

    e the device away from the radio or television.

    g the device into a different electrical outlet, so that the ice and the radio or television are on separate electrical uits.

    e sure that all peripheral devices are also certified.

    e sure that appropriate cables are used to connect the ice to peripheral equipment.

    sult your equipment dealer, Agilent Technologies, or an erienced technician for assistance.

    nges or modifications not expressly approved by Agilent hnologies could void the users authority to operate the ipment.

    ification for Federal Republic of Germany

  • Introduction

    Site Pr

    Fuses

    Table 1 lists the fuses required for proper operation. These fuses s

    Cleaning

    Recycling the Product

    Table 1

    Fuse de

    F2, F3

    F3

    F3, F4

    F1

    siteprep.book Page 17 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 17For recycling, contact your local Agilent sales office.To clean the unit, disconnect the power and wipe down with a damp, lint-free cloth.hould only be accessed by Agilent service personnel.

    Fuses

    signation Location Fuse rating and type

    Power line module 10A 250V, glass tube

    Terminal near transformer

    10A 250V, glass tube

    Power board 6A 125V, glass tube

    Power board 1A 250V, glass tube

  • 18

    Introduction

    siteprep.book Page 18 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

  • Agilent G1888A Headspace SamplerSite Preparation and Installation

    2Site Preparation

    Introdu

    Temper

    Perf

    Opti

    Space R

    Power

    Line

    Pow

    Gas Re

    Externa

    Sup

    Reg

    Venting

    Sample

    This cbefore

    siteprep.book Page 19 Tuesday, February 17, 2004 9:18 AM19Agilent Technologies

    ction to Site Preparation 20

    ature and Humidity Requirements 21

    ormance range 21

    mum performance 21

    equirements 22

    Requirements 23

    voltage 23

    er line cord 23

    quirements 24

    l Plumbing 25

    ply tubing 25

    ulators 25

    Requirements 28

    r Considerations 29

    hapter describes requirements and tasks to complete installing the Headspace Sampler in the laboratory.

  • 20

    Site Preparation

    Introduction to Site Preparation

    This chapter provides site preparation information for the Aglienfor:

    Tem

    Spa

    Pow

    Gas

    Ext

    Ven

    ContaPrepapropercomplat wwHeads

    siteprep.book Page 20 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guidet G1888A Headspace Sampler, covering the requirements

    perature and humidity

    ce

    er

    ernal plumbing

    ting

    ct an Agilent sales representative for the current Site ration Checklist. Use the checklist to ensure that the equipment and parts are present, and that installation is

    eted properly. The checklist can also be found on the web w.agilent.com/chem. Click Library and search for the pace Sampler model number.

  • Site Preparation

    Site Pr

    Temperature and Humidity Requirements

    Performance range

    The Hfrom 5humid

    The mturneddepentempeturnedtempe

    Optimum performanc

    The op(58 Fconde

    siteprep.book Page 21 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 21eadspace Sampler operates in environment temperatures C to 45 C (41 F to 113 F) and 5% to 90% relative

    ity .

    inimum operating temperature for the oven while it is on is 40 C. However, the actual minimum temperature

    ds on the environment. Generally, the minimum oven rature will equal 20 C above ambient. The oven can be off to use the Headspace Sampler closer to ambient

    rature.

    e

    timum temperature and humidity are 20 C to 27 C to 80 F) and 50% to 60% relative humidity (without nsation).

  • 22

    Site Preparation

    Space Requirements

    The Headspace Sampler requires a laboratory bench with an unobs46 cm30 kg

    The unabove the Heobstruinterfe

    Other chromrequirinstrushows

    Figure

    64 cm

    siteprep.book Page 22 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guidetructed surface approximately 64 cm (25 inches) deep by (18.1 inches) wide and capable of supporting at least (102 lb.).

    it is approximately 56 cm (22 inches) high. The area it must be clear. Allow enough room to open the top lid of adspace Sampler. Shelves or any overhanging ctions limit access to the top of the instrument and re with cooling.

    devices used with the Headspace Sampler (gas atographs, detectors, integrators, or workstations) will e additional bench space. See the manual for each ment for the applicable space requirements. Figure 1 instrument footprints for a sample installation.

    1 Example of space needed on lab bench

    Headspace Sampler

    6890 GC

    PC Monitor

    PC Keyboard46 cm

    190 cm

  • Site Preparation

    Site Pr

    Power Requirements

    Line voltage

    The Hsuppli

    115750

    230750

    The H230 VAacceptmaximdedica

    Power line cord

    To progroundaccord(IEC)

    When

    recept

    siteprep.book Page 23 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 23acle is properly grounded.three-conductor power line cord provides power and grounds the instrument. A properly grounded receptacle has its ground contact connected to a suitable earth ground. Verify that the eadspace Sampler operates from one of these line voltage es:

    VAC single phase (90 VAC to 130 VAC), 50 to 60 Hz, VA max

    VAC single phase (198 VAC to 264 VAC), 50 to 60 Hz, VA max

    eadspace Sampler is shipped with the voltage set to C. The maximum power the Headspace Sampler can is 750 VA. The recommended neutral to ground level is a um of 3 VAC. The Headspace Sampler should have a ted ground.

    tect users, the instrument panels and cabinet are ed through the three-conductor power line cord in ance with International Electrotechnical Commission

    requirements.

    plugged into a properly grounded receptacle, the

  • 24

    Site Preparation

    Gas Requirements

    The gas entering the Headspace Sampler becomes part of the

    WARNING Do notpotentstandb

    siteprep.book Page 24 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide use hydrogen in the Headspace Sampler. Hydrogen creates a ial explosion hazard due to venting during operation and y.carrier gas for the GC column. It may come from a pressure-regulated purified source (see Figure 3 on page 27) or from the GC itself. A gas supply connection for vial pressurization is also required. Gas at 550 kPa (80 psi) is recommended.

    We recommend instrument or chromatographic purity grades specifically intended for chromatographic use. The gas purities in Table 2 are recommended for optimal chromatographic performance.

    Table 2 Gas purity requirements

    Gas Purity

    Helium 99.9995% pure

    Nitrogen 99.9995% pure

  • Site Preparation

    Site Pr

    External Plumbing

    Supply tubing

    The Helectrorespecsourceneede

    If extrtubinginstrucontamcontam

    rate fowhen instru

    Gas coSwageSwage

    Do nothread

    Regulators

    Two-ssupplipossibSuitab

    If cylintempeat the

    siteprep.book Page 25 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 25r the particular gas. Tubing of 1/8-inch od is adequate the supply is less than 3 meters (10 feet) away from the ment. Use 1/4-inch od tubing for longer distances.

    nnection to the Headspace Sampler is made with 1/8-inch lok bulkhead fittings. Never use Teflon tape on lok-type fittings.

    t use pipe dope or liquid thread sealant on any pipe s. Use only PTFE plumbing tape for these fittings.

    tage regulators are recommended. Consult a local gas er for the type and size of cylinder valve connections. If le, obtain Compressed Gas Association (CGA) numbers. le two-stage regulators are also available from Agilent.

    ders and regulators are stored outdoors where rature fluctuates, use an additional single-stage regulator back of the instrument.The required tube diameter depends on the distance from the supply or source to the Headspace Sampler and the total flow eadspace Sampler ships with plumbing kits for both nic and manual pressure control (EPC and MPC, tively). The kits contain enough tubing to reach a gas 3 meters from the Headspace Sampler. Extra plumbing is

    d if the gas supply is not within 3 meters.

    a plumbing is needed, use only preconditioned copper (part number 5180-4196) to supply gases to the ment. Ordinary copper tubing contains oils and

    inants. Plastic tubing is permeable to oxygen and other inants that can damage columns and detectors.

  • 26

    Site Preparation

    If a single gas source supplies several instruments (via a manifold or T-connections), the pressure in the lines will fluctuate. Add single-stage regulators at the back of each instru

    Figure

    Gas supply

    Single-stage p

    regulator

    source

    siteprep.book Page 26 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation GuideSet regulators to deliver 410 to 690 kPa (60 to 100 psi) to the instrument. Flow and pressure regulators require at least 140 kPa (20 psi) pressure drop across them to maintain regulation. See Table 3 for the correct regulator to use with each outlet fitting.

    Table 3 CGA Numbers and Pressure Regulator Kit Part Numbers

    Cylinder outlet fittingCGA number

    Part number for pressure regulator kit with 1/8-inch Swagelok outlet

    346 5183-4641

    540 5183-4643

    580 5183-4644

    590 5183-4645ment to stabilize the pressure. See Figure 2 below.

    2 Single stage regulator used to stabilize pressure

    Headspaceunit

    ressure

    80-120 psig 60-100 psig550-830 kPa 410-690 kPa

  • Site Preparation

    Site Pr

    The gas to the Headspace Sampler becomes the carrier gas. Traps are recommended to protect the columns. See Figure 3.

    Figure

    Two-stage regulator

    On/Off

    Oxygen

    Moistu

    Hydroc

    Gas sup

    Gas sup

    To HeaSample

    siteprep.book Page 27 Tuesday, February 17, 2004 9:18 AM27ep and Installation Guide

    3 Plumbing a pressure-regulated gas purified source

    valve

    trap

    re trap

    arbon trap

    ply

    ply on/off valve

    dspacer

  • 28

    Site Preparation

    Venting Requirements

    During normal operation, some of the vial headspace gas is ventedcompoplace tatmos1/8-inthe loc

    siteprep.book Page 28 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide4 Location of vent line on back of instrumentFigure

    Vent line to the outside of the Headspace Sampler. If the nents of this gas are expected to be toxic or noxious, he unit within a fume hood or attach a vent system, at

    pheric pressure, to the port on the sampler unit. Use ach Swagelok fitting for the connection. See Figure 4 for ation of the vent line.

  • Site Preparation

    Site Pr

    Sampler Considerations

    One or more optional parts and kits may be required for the install

    Theand232

    To eflowmumodmod

    To sPre

    Theon aG23

    siteprep.book Page 29 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 29ation, depending on the intended use.

    Headspace Sampler ships with a deactivated flowpath a 1-mL sample loop. A 3-mL sample loop (part number 1700004) can be ordered separately.

    nable software method control of the Headspace Sampler s and pressures, an Agilent 6890 or Agilent 6850 GC

    st have an auxiliary electronic pressure control (EPC) ule. Order kit G1888-60705. The kit contains the EPC ule, the bleeder weldment, and the interface kit.

    plice the transfer line into the inlet, order the Manual ssure Control Interface Kit G1290-60515.

    re is an optional wire support for the transfer line for use n Agilent 6890 GC. Order Transfer Line Support Kit 19-60600.

  • 30

    Site Preparation

    siteprep.book Page 30 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

  • Agilent G1888 Network Headspace SamplerSite Prep and Installation Guide

    3Installation

    Installa

    Unpack

    Place t

    Connec

    Connec

    Flow

    Typi

    Set

    Con

    MPC

    EPC

    Con

    Set

    Control

    This cin the installHeads

    siteprep.book Page 31 Tuesday, February 17, 2004 9:18 AM31Agilent Technologies

    tion 32

    the Unit and Check the Contents 33

    he Unit in Position 34

    t the Power Cord 35

    t the Plumbing and Transfer Line 37

    modes 37

    cal installation procedures 39

    the flow mode on the Headspace Sampler 40

    nect on/off valves to the gas input fittings 41

    connections 42

    connections 46

    nect the control cables 50

    communication options 52

    Setup 50

    hapter describes steps to install the Headspace Sampler laboratory. It is assumed that the GC has already been ed, is working properly, and is available for the pace Sampler Installation.

  • 32

    Installation

    Installation

    To ins

    1 Unp

    2 Pla

    3 Con

    4 Con(GC

    5 Conopt

    6 Per

    ContaInstallpropercomplat wwHeads

    CAUTION The Hecompaconnec

    siteprep.book Page 32 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guidetall the Headspace Sampler:

    ack the unit and check the contents. See page 33.

    ce the unit in a suitable location. See page 34.

    nect the power cord. See page 35.

    nect the headspace sampler to the gas chromatograph ). See page 37.

    nect the control cables to the GC and set communication ions. See page 50.

    form the checkout procedure. See page 56.

    ct an Agilent sales representative for the current ation Checklist. Use the checklist to ensure that the equipment and parts are present, and that installation is

    eted properly. The checklist can also be found on the web w.agilent.com/chem. Click Library and search for the pace Sampler model number.

    adspace Sampler contains metric fittings which are not tible with Swagelok or other fittings except for the external gas tions on the rear of the unit.

  • Installation

    Site Pr

    Unpack the Unit and Check the Contents

    Inspect the shipping containers for damage. If a container is damagyour loinspec

    Checkany diRetaincompl

    siteprep.book Page 33 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 33ed or shows signs of stress, notify both the carrier and cal Agilent office. Keep all shipping materials for tion by the carrier.

    the items received against the packing lists. If there are screpancies, notify your local Agilent office immediately. the shipping containers until all contents are checked for eteness and instrument performance is verified.

  • 34

    Installation

    Place the Unit in Position

    1 Place the Headspace Sampler on a bench top to the left of the GC.

    2 FreSamUsenec

    3 Poseasily reach the GC inlet without kinks or sharp bends.

    CAUTION The spobstructop of t

    Figure

    Release

    siteprep.book Page 34 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guidee the transfer line from the side of the Headspace pler. Open the plastic collar holding the line in place. a small flat blade screwdriver to press the release tab if essary. See Figure 5.

    ition the Headspace Sampler so that the transfer line can

    ace above the Headspace Sampler must be free of overhanging tions that could interfere with cooling and limit access to the he instrument or prevent the top from opening.

    5 The transfer line release tab

    tab

  • Installation

    Site Pr

    Connect the Power Cord

    1 Verify the proper operating voltage setting. The Headspace Samneeloca

    2 Plugro

    Figure

    Voltage

    Fuses

    siteprep.book Page 35 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 35pler is shipped with the voltage set to 230 VAC. If ded, set the voltage selector to 115 VAC. Also note the tion of the fuses. See Figure 6.

    g the Headspace Sampler power cord into a properly unded outlet. See page 23 for power requirements.

    6 Power connection

    selector

  • 36

    Installation

    3 Turn on the Headspace Sampler to verify that it is working properly. The Headspace Sampler performs a self-test when it starts up. Verify that each test is labeled OK. The self-test infodia

    a P

    b SP

    c S

    Usescre

    See thdescri

    Figure

    Shu

    Tray

    Caro

    Oven

    Loop

    siteprep.book Page 36 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    rmation can also be viewed at any time from the gnostic screen.

    ress the Menu key.

    croll down to Advanced Functions with the cursor keys. ress Enter.

    croll down to Diagnostic and press Enter.

    the cursor arrows to scroll through the multiple test ens. A typical diagnostic test screen is shown in Figure 7.

    e G1888A Headspace Sampler operating manual for a ption of the system diagnostic messages.

    7 Self-test screen

    DIAGNOSTIC

    NOT READY

    tter:

    usel OK

    Sensor

    OK

    OK

    OK

    Sensor OK

    SELF TEST

  • Installation

    Site Pr

    Connect the Plumbing and Transfer Line

    Flow modes

    The Aflow m

    Ele

    EPCconvialsetp

    Man

    MPregucar

    See Filocatio

    Figure

    Pressur

    Flow co

    siteprep.book Page 37 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 37gilent G1888A Headspace Sampler can operate in two odes:

    ctronic pressure control (EPC)

    mode requires an EPC inlet and an Auxiliary EPC trol module in the GC. The modules provide carrier and pressurization gas control via GC or Agilent data system oints.

    ual pressure control (MPC)

    C mode uses the manual flow controller and pressure lator on the G1888A Headspace sampler to control the

    rier and vial pressurization gases.

    gure 8 for the MPC flow controller and pressure regulator ns.

    8 Flow controller and pressure regulator location

    e regulator

    ntroller

  • 38

    Installation

    There are several ways to connect the Agilent Headspace Sampler to the GC. The type of connection depends on the GC model, the selected flow mode, and how the instruments are to be use

    EPC co

    Direct insertreal mconsisduplicnorma

    Direct the sa6890 G

    MPC c

    Needlefrom tinsertinjecti

    Needlefrom t23225

    siteprep.book Page 38 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    d. The possibilities include:

    nnections

    connection to a split/splitless inlet The sampler is ed between the inlet flow module and the inlet, allowing ethod control, via the GC, of flows and pressures for tent results between setups. The system is easy to ate and monitor but requires time to convert back to l inlet injection for nonheadspace samples.

    connection to a volatiles interface The transfer line from mpler is connected to the volatiles interface. See the Cs user information for instructions.

    onnections

    connection to a split/splitless inlet The transfer line he sampler ends in a needle (part no. 2322590005) that is ed through the inlet septum. Conversion to normal on is rapid and simple.

    connection to a cool on-column inlet The transfer line he sampler ends in a small od needle (part no. 90004) that is inserted through the inlet septum.

  • Installation

    Site Prep and Installation Guide

    Typical installation procedures

    For MPC installations with a split/splitless inlet:

    1 VerSam

    2 ConHea

    3 Consou

    4 Conspli

    For EP

    1 Set to E

    2 ConHea

    3 Conmod

    4 Conspli(pa

    Altern

    To insGC, co

    siteprep.book Page 39 Tuesday, February 17, 2004 9:18 AM39

    ify that the flow mode toggle switches on the Headspace pler are set to MPC (page 40).

    nect the on/off valves to the gas input fittings on the dspace Sampler (page 41).

    nect the gas input fittings to a pressure-regulated gas rce (page 42).

    nect the transfer line using a needle connection to a t/splitless inlet (page 43).

    C installations with a split/splitless inlet:

    the flow mode toggle switches on the Headspace Sampler PC (page 40).

    nect the on/off valves to the gas input fittings on the dspace Sampler (page 41).

    nect the vial pressurization gas to the GC auxiliary EPC ule (page 46).

    nect the carrier gas and transfer line directly to the t/splitless inlet by splicing into the carrier gas line ge 48).

    ative installations

    tall the Headspace Sampler on a 4890D or 5890 Series II ntact Agilent Service.

  • 40

    Installation

    Set the flow mode on the Headspace Sampler

    Choose the flow mode for the carrier gas and vial pressurization gas and set the flow mode toggle switches on the headspace

    Figure

    siteprep.book Page 40 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    hessampler.

    1 Open the lid of the Headspace Sampler. Locate the two tabs near the top of the Headspace Sampler on the left and right sides. Push the tabs towards the back of the instrument and lift the lid. Use caution when opening the lid due to its weight. The lid locks in place when fully opened.

    2 Set the toggle switches for carrier gas and vial pressurization gas on the underside of the Headspace Sampler lid to the necessary positions. See Figure 9. The factory default setting is MPC.

    9 The pressure control toggle switc

    Vial pressurization gas

    Carrier gas

    MPC

    EPC

  • Installation

    Site Prep and Installation Guide

    Connect on/off valves to the gas input fittings

    Install two on/off valves and tubing on the gas input fittings as shown in Figure 10. The figure shows a connection to the MPC

    Figure

    Carri

    Vial

    On/Ofor le

    For M

    For E

    siteprep.book Page 41 Tuesday, February 17, 2004 9:18 AM41input fittings.

    For EPC, use the fittings on the right

    For MPC, use the fittings on the left

    10 On/Off valves connected

    er gas in

    pressurization gas in

    ff valvesak testing

    PC

    PC

  • 42

    Installation

    MPC connections

    Carrier gas and vial pressurization gas connections

    If bothInstallpressuFigure

    Figure

    CarrierVial pr

    siteprep.book Page 42 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation GuideIf only one gas input fitting uses MPC, run a line from a pressure-regulated gas source directly to the on/off valve connected to the MPC gas input fitting.

    11 Carrier and vial pressure connections

    gas inessure

    T-fitting

    gas source

    On/Off valves

    regulated the carrier gas and vial pressurization gas are MPC, a T-fitting to the on/off valves. Run a line from a re-regulated gas source to the T-fitting as shown in 11.

    To pressure

  • Installation

    Site Prep and Installation Guide

    Needle connection from transfer line to a split/splitless inlet

    If the Headspace Sampler uses MPC for the carrier gas, the transfer line connects to the GC inlet by a needle passing throug

    This sydoes nmoduldilutioinlet c

    The Houter-This nod neeinlet a

    A smafor onOrder

    Followprocedfor a 6

    1 Instspli

    2 Rem

    3 Attwre

    4 Inse

    siteprep.book Page 43 Tuesday, February 17, 2004 9:18 AM43

    h the inlet septum.

    stem easily converts to normal injection. The installation ot disturb the normal plumbing between the inlet flow e and the inlet. Normal inlet flow reduces to minimize n of the headspace sample. Most of the flow through the omes from the sampler.

    eadspace Sampler shipping kit includes a 0.7 mm diameter (od) transfer line needle (part no. 2322590005). eedle is also called a transfer line terminal. The 0.7 mm dle is most appropriate for use with the purged/packed nd the split/splitless inlet.

    ller diameter needle (part no. 2322590004) must be used -column injection and may be used with other inlets. this needle separately.

    these steps to connect the transfer line to the GC. This ure is described for the 6850 GC. Installation is similar 890 GC.

    all a deactivated direct liner (part no. 5181-8818) in the t/splitless inlet.

    ove the standard septum nut from the inlet.

    ach the transfer line to the needle. Use two 7-mm nches to tighten it.

    rt the needle through the strain relief coupling.

  • 44

    Installation

    5 Push an 11-mm through-hole septum onto the needle. Carefully remove any pieces of septum cored by the needle. Be sure the needle bore is clear. Figure 12 shows the asse

    Figure

    End of t

    Top of

    siteprep.book Page 44 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    mbly.

    12 The needle assembly

    ransfer line

    inlet

  • Installation

    Site Prep and Installation Guide

    6 Insert the tip of the needle into the inlet. Force the septum (it is oversized) into the recess in the top of the inlet. See Figure 13 for the assembled view.

    7 Scr

    When using contrometerrate.

    Figure

    Septum

    siteprep.book Page 45 Tuesday, February 17, 2004 9:18 AM45

    ew on and tighten the strain relief coupling.

    operating, set a small flow of carrier gas through the inlet the GC inlet control. Set most of the flow using the flow ller in the Headspace Sampler. Use an electronic flow

    connected to the inlet split vent to verify the actual flow

    13 The transfer line attached to the inlet

    is forced into recess

  • 46

    Installation

    EPC connections

    Connect the vial pressurization gas for an EPC setup

    An Augas. ThEPC cfor a 6

    1 If thG19pro

    2 InstG18inst

    3 Locthepne

    4 Remmodthe

    Figure

    Frits

    Output

    siteprep.book Page 46 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    xiliary gas channel in the GC supplies vial pressurization e following steps describe installation for a 6890 GC with

    ontrol of headspace. The installation procedure is similar 850 GC.

    e GC does not have an Auxiliary EPC module, install the 40A Headspace Interface Kit. Follow the instructions vided with that kit.

    all the external sampler interface kit (part no. 88-60705) into the GC. Follow the procedure for alling onto a back-pressure regulated system.

    ate the block on the Auxiliary EPC module that connects three gas outlet tubes for the auxiliary channels to the umatics module.

    ove the screw that holds the block to the pneumatics ule. Pull the block free of the module and rotate it so that

    frits are on top. See Figure 14.

    14 The Auxiliary EPC module gas outlet block

    Aux 5

    Aux 4

    Aux 3

    Aux 3

    Aux 4

    Aux 5

    s

  • Installation

    Site Prep and Installation Guide

    5 Choose a channel for the vial pressurization control. Remove that channels frit from the block. Remove the O-ring that seals the channel.

    Figure

    Heads

    GC

    Vial pr

    To Aux

    Bleed w

    Brass

    siteprep.book Page 47 Tuesday, February 17, 2004 9:18 AM47

    ler to the auxiliary module6 Place an O-ring on a zero-resistance brass tube frit (part no. G1570-20540). Place the O-ring/frit combination in the block.

    7 Reconnect the block to the pneumatics module. Tighten the screw.

    8 Use tubing and Swagelok fittings to connect the appropriate Auxiliary module output, the bleed weldment, and the VIAL PRESSURE fitting to the on/off valve on the back of the Headspace Sampler as shown in Figure 15. Do not disturb the brass end on the bleed weldment.

    15 Connecting the Headspace Samp

    pace Sampler

    essure fitting

    iliary module output

    eldment

    end

  • 48

    Installation

    Direct connection from the transfer line to a split/splitless inlet

    The GC inlet control channel supplies carrier gas. The following steps describe installation for a 6890 GC with EPC control of headsp

    1 Inst

    2 LocFigu

    3 Cut2 in

    a C

    b Hf

    c F

    4 Slid

    Figure

    Carrier

    siteprep.book Page 48 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    ace. The installation procedure is similar on a 6850 GC.

    all a split inlet liner (part no. 5183-4647) in the inlet.

    ate the carrier line to the split/splitless inlet shown in re 16.

    the carrier line to the split/splitless inlet about 1 to ches from the inlet body.

    arefully score the circumference of the tubing.

    old the tubing with pliers on each side of the score and lex it until it breaks free.

    ile the edges until they are smooth.

    e the nut onto the inlet side of the carrier line.

    16 The split/splitless inlet carrier gas line

    gas line

  • Installation

    Site Prep and Installation Guide

    5 Attach a zero-dead-volume (ZDV) union to the short piece of carrier line attached to the inlet as shown in Figure 17.

    6 AttSee

    7 Usecarshiprea

    8 Con

    Figure

    Figure

    ZDV un

    Carrier

    Ferrule

    Nut

    ZDV un

    Transfe

    siteprep.book Page 49 Tuesday, February 17, 2004 9:18 AM49

    ach the headspace sampler transfer line to the ZDV union. Figure 18.

    the other ZDV union to connect the free end of the rier gas line to the length of tubing provided in the ping kit. Route the tubing out the back of the GC to the

    r panel of the sampler.

    nect the tubing to the CARRIER fitting on the sampler.

    17 Attaching the ZDV union to the carrier line

    18 Connecting the transfer line to the carrier line

    ion

    line

    ion

    r line

  • 50

    Installation

    Control Setup

    Connect the control cables

    ConneBe surthe fre

    Depenneedeused w

    Figurecablinsoftwa

    .

    Table 4

    Sample

    LAN*

    Remote

    APG Y-c

    Serial (R

    Remote

    *Throug

    siteprep.book Page 50 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guidect the remote start/stop cable to the back of the sampler. e to connect the spade lug to the ground screw. Connect e end of the cable to the GC.

    ding on the setup, other cable connections may be d. Refer to Table 4 for an overview of cable connections ith the Headspace Samplers.

    19 shows some common cabling configurations. These g configurations are valid for the G2070AA and G2922AA re set.

    Cables for Headspace Samplers

    r port Connect to Part number

    Hub* 8121-0940

    6820, 6850 and 6890 GCs G1530-60930

    able 6890N and 5973N G1530-61200

    S-232) PC G1530-60600

    4890 and 5890 Series II GCs 35900-60700

    h a network hub or similar connection.

  • Installation

    Site Pr

    Figure

    21

    1 1Cables

    1

    2

    3

    4

    5

    siteprep.book Page 51 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 51

    19 Common cabling configurations

    Workstation

    Printer

    Hub or switch

    G1888A 6890N

    Workstation

    Printer

    Hub or switch

    G1888A 6850

    21

    1 1

    Local Talk

    Local Talk

    BNC

    BNC

    Workstation

    Printer

    G1888A 6820

    23

    5 3

    Workstation

    Printer

    Hub or switch

    G1888A 5973N 6890N

    421

    1 11

    8121-0940, Cable, 7.5-m 100-BaseT LAN

    G1530-60930, 2-m APG remote cable, 9-pin male to 9-pin male

    .

    .

    G1530-60600, 2-m RS-232 cable, 9-pin female to 9-pin female

    G1530-61200, 2-m APG Y-cable, two 9-pin male, one 9-pin female

    8121-1013, USB-DB9 RS-232 adapter cable

    .

    .

    .

  • 52

    Installation

    Set communication options

    If using RS-232 or LAN communications, configure the communications as described below.

    LAN s

    If the AddreFunct

    1 Pre

    2 Scr

    3 Scrpre

    4 Theand

    After cusing

    1 Pre

    2 Scr

    3 Scrpre

    Figure

    siteprep.book Page 52 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    ettings

    Headspace Sampler is connected to a LAN, set the IP ss, subnet mask, and gateway using the Advanced ions menu.ss the Menu key.

    oll down to Advanced Functions. Press Enter.

    oll down to LAN Config. and press Enter or ss . followed by 4 on the numeric keypad.

    LAN Config. screen is shown in Figure 20. Use the cursor keypad to enter new values. Press Enter to set the value.

    onfiguring the LAN settings, disable the RS-232 port the Advanced Functions menu. ss the Menu key.

    oll down to Advanced Functions. Press Enter.

    oll down to Enable RS232 and press Enter or ss . followed by 6 on the numeric keypad.

    20 The LAN Config. screen

    SET

    IP Address

    Sub MaskRunning

    LAN Config.

    Gateway

    10 .1 .1 .105

    255 .255 .255 .0

    10 .1 .1 .1

  • Installation

    Site Prep and Installation Guide

    4 The Enable RS232 screen is shown in Figure 21. Use the cursor to choose No from the ADV. FUNCT list. Press Enter.

    Headsthe RS

    Additio

    If usinenable

    1 Pre

    2 Scr

    3 Scrpre

    Figure

    siteprep.book Page 53 Tuesday, February 17, 2004 9:18 AM53

    pace Sampler automatically resets after after disabling -232 port.

    nal RS-232 settings

    g RS-232 communications with the Headspace Sampler, RS-232 capabilities using the Advanced Functions. ss the Menu key.

    oll down to Advanced Functions. Press Enter.

    oll down to Enable RS232 and press Enter or ss . followed by 6 on the numeric keypad.

    21 The Enable RS232 screen

    13. Zones Calibration

    14. LAN Configuration

    15. MAC AddressYes

    ADV. FUNCTION

    No16. Enable RS232

  • 54

    Installation

    4 The Enable RS232 screen is shown in Figure 22. Use the cursor to choose Yes from the ADV. FUNCT list. Press Enter.

    Figure

    siteprep.book Page 54 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    22 The Enable RS232 screen

    13. Zones Calibration

    14. LAN Configuration

    15. MAC AddressYes

    ADV. FUNCTION

    No16. Enable RS232

  • Agilent G1888 Network Headspace SamplerSite Prep and Installation Guide

    4Running Checkout Samples

    Runnin

    Prep

    Run

    6850

    This cSampl

    siteprep.book Page 55 Tuesday, February 17, 2004 9:18 AM55Agilent Technologies

    g Checkout Samples 56

    are the checkout sample 56

    the sample 57

    /6890 GC checkout method 58

    hapter describes the checkout steps for the Headspace er.

  • 56

    Running Checkout Samples

    Running Checkout Samples

    Prepare the checkout sample

    The chethanocapilla

    The chheadsp20 mLinstru

    1 Snacon

    2 Dis

    3 HolsampraFigu

    4 Thecombotrem

    5 Drocap

    Figure

    1. Gr

    siteprep.book Page 56 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guideeckout sample is three detector-specific compounds in l. The box contains 1 mL of sample and a bottle of 5-L ry pipettes.

    eckout method calls for a 5-L sample in a 20-mL ace vial. The default Headspace Sampler vial size is

    . See the Headspace Sampler Operating Manual for ctions to set the vial size if needed.

    p the top off the checkout sample ampoule. Transfer the tents to a screw-cap bottle.

    pense one capillary pipette

    d the pipette with clean tweezers. With the checkout ple bottle and the pipette as close to horizontal as

    ctical, dip the end of the pipette into the sample. See re 23

    pipette will fill by capillary action. When it has filled pletely, pull the pipette out of the sample. Wipe the

    tom edge of the pipette against the edge of the vial to ove any droplets on the outside.

    p the filled pipette into a 20-mL headspace vial. Crimp a on the vial.

    23 Filling a pipette

    asp 2. Fill 3. Wipe

  • Running Checkout Samples

    Site Prep and Installation Guide

    Run the sample

    1 Set the method parameters on the GC and Headspace sampler. See Table 5 on page 58 for checkout settings or load

    siteprep.book Page 57 Tuesday, February 17, 2004 9:18 AM57the stored checkout method as described in 6850/6890 GC checkout method on page 58. If needed, install the checkout column and liner.

    2 If using MPC, set the carrier gas flow as follows:

    a Set the GC inlet split ratio and column head pressure to the values given in Table 7 on page 61.

    b Set the split flow to 4 mL/min. The total flow will read 11 mL/min.

    c With an electronic flow meter connected to the GC split vent, turn up the headspace carrier gas flow until the flow rate out the split vent is approximately 80 mL/min. Remove the flow meter. See Figure 24.

    NOTE The GC and Headspace pressure readings should be similar. Record the headspace pressure readingit can be used to set up future checkout runs.

  • 58

    Running Checkout Samples

    The repages

    6850/6890 GC check

    Load t

    The Hthe chchecko

    1 Pre

    2 HigEnt

    3 UseFun

    Figure

    6850 Ser i esGC System

    Split ve

    Electronic flowmeter

    siteprep.book Page 58 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    sults should be similar to the chromatograms shown on 65 to 68.

    out method

    he existing checkout method

    eadspace Sampler ships with a stored method containing eckout settings. Do the following steps to load the ut method.

    ss the Menu key.

    hlight Advanced Functions with the cursor keys. Press er.

    the cursor keys to select Stored Meth. (Advanced ction number 7) and press Enter. 3 Run the checkout sample.24 The electronic flow meter connected

    nt

    6850 GC shown

  • Running Checkout Samples

    Site Prep and Installation Guide

    4 The screen shown in Figure 25 appears. Use the cursor keys to select Checkout. Press Enter.

    Create

    If the acceptto creasectionthat ap

    Figure

    Table 5

    Headsp

    Oven tem

    Valve te

    Transfer

    Loop vol

    Vial pres

    GC cycle

    Vial equ

    Pressur

    siteprep.book Page 59 Tuesday, February 17, 2004 9:18 AM59

    a new checkout method

    checkout settings in the stored methods screen are not able, use Tables 5 through 12 and Figures 26 through 29 te a new Headspace Sampler checkout method. This covers several inlets and detector types. Use the settings ply.

    25 The Stored Method screen

    Headspace Sampler settings

    ace method

    perature 100 C

    mperature 110 C

    line temperature 115 C

    ume 1 mL

    sure 15 psi (See Table 11 if using EPC control)

    time 5.5 min

    ib time 7.0 min

    ization time 0.08 min

    DIAGNOSTIC

    READY

    Checkout

    OQ/PV

    MeOH in H2O

    STORED METH.Default

  • 60

    Running Checkout Samples

    Loop fill

    Loop eq

    Injection

    Table 5 Headspace Sampler settings (continued)

    Headspace method

    siteprep.book Page 60 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    time 0.5 min

    uib time 0.05 min

    time 1.0 min

  • Running Checkout Samples

    Site Prep and Installation Guide

    Table 6 GC oven settings

    Oven

    Initial te

    Initial tim

    Rate

    Maximu

    Equilibra

    Run tim

    Table 7

    Inlet (vo

    Mode

    Initial te

    Pressur

    Split rat

    Split flo

    Total flo

    Gas sav

    Gas type

    Liner

    siteprep.book Page 61 Tuesday, February 17, 2004 9:18 AM61

    mperature 100 C

    e 4.00 min

    0 (Off)

    m temperature 300 C

    tion time 1.00 min

    e 4.00 min

    Inlet settings

    latiles or split/splitless)

    Split

    mperature 250 C (On)

    e 19.5 psi (On)

    io 20:1 (G1289)

    w 80 mL/min (G1289)

    w 87 mL/min (G1289)

    er Off

    Helium

    5181-8818

  • 62

    Running Checkout Samples

    Table 8 Column settings

    Column

    Model n

    Maximu

    Length

    Diamete

    Film thic

    Mode

    Initial flo

    Initial pr

    Average

    Outlet p

    siteprep.book Page 62 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    umber 689019091J-413, HP-5 5% Phenyl methyl silicone685019091Z-413E, HP-1 Methyl silicone

    m temperature 325 C

    30 m

    r 320 m

    kness 0.25 m

    Constant flow

    w 4.0 mL/min

    essure 19.5 psi

    velocity 60 cm/s

    ressure Ambient

  • Running Checkout Samples

    Site Prep and Installation Guide

    Table 9 Detector settings (FID, NPD, ECD)

    Detector parame

    FID NPD ECD

    Tempera

    Hydroge

    Air flow

    Mode

    Makeup

    Makeuptype

    Flame

    Electrom

    Lit offse

    Adjust o

    Bead

    Equilibratime

    siteprep.book Page 63 Tuesday, February 17, 2004 9:18 AM63

    ters

    ture 300 C 300 C

    n flow 30 mL/min 2.0 mL/min

    400 mL/min 60 mL/min Constant makeup

    Constant makeup Constant makeup

    30 mL/min

    flow 25 mL/min 8 mL/min Nitrogen or argon/methane

    gas Nitrogen or helium Nitrogen or helium

    On On

    eter On On

    t 2.0

    ffset 30

    On

    tion 0

  • 64

    Running Checkout Samples

    Use thcontro

    Table 10 5973 MSD settings

    5973 MSD unique parameters

    Column

    Solvent

    Run tim

    HeadspaGC cycle

    Scan ran

    Method

    Inlet pre

    Column

    Split rat

    Injector

    Table 1

    Aux Pre

    Gas type

    Initial pr

    Initial tim

    Rate

    siteprep.book Page 64 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    e settings given in Table 11 when using a GC with EPC l.

    6890/5973A MSD 19091S-433, HP-5MS 5% Phenyl methyl siloxane

    delay 3 min

    e 6.5 min

    ce event time for time

    7 min

    ge 45 to 300

    used Atune file

    ssure 12 psi

    flow 1.1 mL/min with foreline reading of 60

    io 80:1

    External device

    1 GC Aux pressure settings for EPC control

    ssure 3 (for vial pressurization on G1888A)

    Helium

    essure 15 psi (On)

    e 0 min

    0 (Off)

  • Running Checkout Samples

    Site Prep and Installation Guide

    Table 12 Checkout sample

    Checkout sample

    Figure

    0.0

    pA

    20

    40

    60

    80

    100

    120

    140

    160

    FID

    siteprep.book Page 65 Tuesday, February 17, 2004 9:18 AM65Type Headspace OQ/PV Standard

    Part number 5182-9733

    Size 5 L (by micro-pipette) in 20-mL headspace vial

    26 FID checkout chromatogram

    min0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0

    1 A, (HS_ECD\HSVI0004.D)

    0.8

    58

    0.8

    80

    2.0

    37

    2.4

    76

    2.8

    25

    Ethanol(40,000 counts typical)

    Dichlorobenzene

    t-Butyl disulfide

    Nitrobenzene

  • 66

    Running Checkout Samples

    Figure

    0.0

    pA

    50

    100

    150

    200

    250

    NPD1 B, (HS_NPD\HSNPD002.D)

    2.3

    84

    siteprep.book Page 66 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    27 NPD checkout chromatogram

    min0.5 1.0 1.5 2.0 2.5 3.0 3.5

    Nitrobenzene

    Ethanol

  • Running Checkout Samples

    Site Pr

    Figure

    0

    Hz

    0

    20000

    40000

    60000

    80000

    100000

    120000

    ECD1 A, (HS_ECD\HSEC0000.D)

    2.0

    21

    siteprep.book Page 67 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 67

    28 ECD checkout chromatorgram

    min1 2 3 4

    0.8

    50

    2.4

    61

    2.8

    10

    Dichlorobenzene

    Ethanol

  • 68

    Running Checkout Samples

    Figure 29 MSD checkout chromatorgram

    Dichlorobenzene

    NitrobenzeneAbu

    ndan

    ce

    1000000

    800000

    600000

    400000

    200000

    03.20 3.60 4.00 4.40

    siteprep.book Page 68 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation GuideTIC: 0201002.D

    t-Butyl disulfide

    Time

    4.80 5.20 5.60 6.00 6.40

  • Agilent G1888 Network Headspace SamplerSite Preparation and Installation

    AConnecting a Swagelok Fitting

    Prepara

    Proced

    This c

    siteprep.book Page 69 Tuesday, February 17, 2004 9:18 AM69Agilent Technologies

    tion 70

    ure 71

    hapter describes proper Swagelok connections.

  • 70

    Connecting a Swagelok Fitting

    Preparation

    Objective

    To mawithou

    Materials needed

    1/8

    1/8

    Fro

    Twonut

    siteprep.book Page 70 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guideke a leak-free tubing connection that can be taken apart t damaging the fitting

    -inch (or 1/4-inch, if used) preconditioned copper tubing

    -inch (or 1/4-inch, if used) Swagelok nuts

    nt and back ferrules

    7/16-inch (for 1/8-inch nuts) or 9/16-inch (for 1/4-inch s) wrenches

  • Connecting a Swagelok Fitting

    Site Pr

    Procedure

    1 Place a Swagelok nut, back ferrule, and front ferrule to the tub

    2 Cla

    3 Pus

    4 Makthe

    Figure

    Front

    Back

    Nut

    Tubin

    Figure

    Plug

    Nut

    held

    siteprep.book Page 71 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 71h the tubing into the stainless steel plug (Figure 31).

    e sure that the front ferrule is touching the plug. Slide Swagelok nut over the ferrule and thread it onto the plug.

    s very costly to repair.

    31 Assembling the fitting

    or fitting

    and ferrules

    Tubing in a viseCAUTIONDo not

    Use a separate stainless steel fitting in a vise for initial tightening of the nut. Do not use an inlet or detector fitting. Strong forces are required to properly set the ferrules, and damage to an inlet or detector fitting iing as shown in Figure 30.

    mp a stainless steel plug or similar fitting in a bench vise.

    30 Swagelok nuts and ferrules

    ferrule

    ferrule

    g

    IMPORTANT!The narrow end of the backferrule fits into the rear ofthe front ferrule.

  • 72

    Connecting a Swagelok Fitting

    5 Push the tube fully into the plug, then withdraw it approximately 1 to 2 mm. Finger-tighten the nut (Figure 32).

    6 Mar

    7 ForwreFortigh

    Figure

    Figure

    Fronferr

    Pencil l

    siteprep.book Page 72 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide

    k the nut with a pencil line (Figure 33).

    1/8-inch Swagelok fittings, use a pair of 7/16-inch nches to tighten the fitting 3/4 of a turn (Figure 34). 1/4-inch fittings, use a pair of 9/16-inch wrenches to ten them 1-1/4 turn (Figure 34).

    32 Insert the tubing

    33 Marking the fitting

    tule

    Backferrule

    Nut

    Insert tubing fully

    Withdraw tubing 1-2 mm

    Tighten nut

    ine

  • Connecting a Swagelok Fitting

    Site Prep and Installation Guide

    8 Remnutthefitti

    9 A cthacruferr

    Figure

    Figure

    siteprep.book Page 73 Tuesday, February 17, 2004 9:18 AM73

    ove the plug from the fitting. To connect the tubing, with and ferrules, to another fitting, finger-tighten the nut, n use a wrench to tighten it 3/4 of a turn (1/8-inch ngs) or 1-1/4 of a turn (1/4-inch fittings).

    orrectly-swaged connection is shown in Figure 35. Note t the end of the tubing in a correctly-swaged fitting is not shed and does not interfere with the action of the ules.

    34 Final tightening

    35 Completed fitting

  • Agilent Technologies

    siteprep.book Page 74 Tuesday, February 17, 2004 9:18 AM Agilent Technologies, Inc.

    Printed in USA, March 2004

    ContentsIntroductionAbout this ManualVideo supplements

    Important Safety WarningsMany internal parts of the instrument carry dangerous voltagesElectrostatic discharge is a threat to instrument electronicsMany parts are dangerously hotGasesGeneral warnings

    Safety and Regulatory CertificationsInformationSymbolsElectromagnetic compatibilitySound Emission Certification for Federal Republic of Germany

    FusesCleaningRecycling the Product

    Site PreparationIntroduction to Site PreparationTemperature and Humidity RequirementsPerformance rangeOptimum performance

    Space RequirementsFigure 1 Example of space needed on lab bench

    Power RequirementsLine voltagePower line cord

    Gas RequirementsExternal PlumbingSupply tubingRegulatorsFigure 2 Single stage regulator used to stabilize pressureFigure 3 Plumbing a pressure-regulated gas purified source

    Venting RequirementsFigure 4 Location of vent line on back of instrument

    Sampler Considerations

    InstallationInstallationUnpack the Unit and Check the ContentsPlace the Unit in PositionFigure 5 The transfer line release tab

    Connect the Power CordFigure 6 Power connectionFigure 7 Self-test screen

    Connect the Plumbing and Transfer LineFlow modesFigure 8 Flow controller and pressure regulator location

    Typical installation proceduresSet the flow mode on the Headspace SamplerFigure 9 The pressure control toggle switches

    Connect on/off valves to the gas input fittingsFigure 10 On/Off valves connected

    MPC connectionsFigure 11 Carrier and vial pressure connectionsFigure 12 The needle assemblyFigure 13 The transfer line attached to the inlet

    EPC connectionsFigure 14 The Auxiliary EPC module gas outlet blockFigure 15 Connecting the Headspace Sampler to the auxiliary moduleFigure 16 The split/splitless inlet carrier gas lineFigure 17 Attaching the ZDV union to the carrier lineFigure 18 Connecting the transfer line to the carrier line

    Control SetupConnect the control cablesFigure 19 Common cabling configurations

    Set communication optionsFigure 20 The LAN Config. screenFigure 21 The Enable RS232 screenFigure 22 The Enable RS232 screen

    Running Checkout SamplesRunning Checkout SamplesPrepare the checkout sampleFigure 23 Filling a pipette

    Run the sampleFigure 24 The electronic flow meter connected

    6850/6890 GC checkout methodFigure 25 The Stored Method screenFigure 26 FID checkout chromatogramFigure 27 NPD checkout chromatogramFigure 28 ECD checkout chromatorgramFigure 29 MSD checkout chromatorgram

    Connecting a Swagelok FittingPreparationObjectiveMaterials needed

    ProcedureFigure 30 Swagelok nuts and ferrulesFigure 31 Assembling the fittingFigure 32 Insert the tubingFigure 33 Marking the fittingFigure 34 Final tighteningFigure 35 Completed fitting

    MPC transfer line: EPC connection: