Upload
poetaoscuro
View
357
Download
8
Embed Size (px)
DESCRIPTION
G1888-90009 Headspace Sampler
Citation preview
Agilent Technologies
Agilent G1888 Network Headspace Sampler
Site Prep and Installation Guide
siteprep.book Page 1 Tuesday, February 17, 2004 9:18 AM
2 Site Preparation and Installation Guide
Notices Agilent Technologies, Inc. 2004
No part of this manual may be reproduced in any form or by any means (including elec-tronic storage and retrieval or translation into a foreign language) without prior agree-ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part NumberG1888-90009
EditionFirst edition, March 2004
Printed in US
Agilent Technologies, Inc.2850 Centerville Road Wilmington, DE 19808-1610 USA
Warranty
The material contained in this docu-ment is provided as is, and is sub-ject to being changed, without notice, in future editions. Further, to the max-imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par-ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec-tion with the furnishing, use, or per-formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Safety Notices
CAUTION
A CAUTION notice denotes a haz-ard. It calls attention to an operat-ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per-formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi-tions are fully understood and met.
AcknowledgementsSwagelok is a registered trademark of the Swagelok Company.
Microsoft and Windows are registered trademarks of the Microsoft Corporation.
siteprep.book Page 2 Tuesday, February 17, 2004 9:18 AM
Site Preparation and Installation Guide 3
Manual Overview
This manual contains four chapters and one appendix.
1 Introduction
This chapter gives an overview of the manual and contains safety information.
2 Site Preparation
This chapter provides the requirements for preparing the site before installing the Headspace Sampler. It covers the requirements for temperature and humidity, space, power, gas, and venting.
3 Installation
Installation describes how to install the Headspace Sampler. It includes instructions for unpacking the unit, checking the contents, connecting the power and gas supplies, making connections to an integrator or Agilent data system, and connecting the unit to a GC.
4 Running Checkout Samples
This chapter provides information on preparing and running the checkout sample to confirm that the sampler is operating correctly.
A Connecting a Swagelok Fitting
This appendix describes proper Swagelok connections.
siteprep.book Page 3 Tuesday, February 17, 2004 9:18 AM
Site Preparation and Installation Guide 4
siteprep.book Page 4 Tuesday, February 17, 2004 9:18 AM
Site Pr
Contents
1 Introduction
About
Ma
Ele
MaGasGen
Safety
InfoSymEleSou
Fuses
Cleani
Recycl
2 Site Preparation
Introdu
siteprep.book Page 5 Tuesday, February 17, 2004 9:18 AMTemperature and Humidity Requirements 21
Performance range 21Optimum performance 21ep and Installation Guideny internal parts of the instrument carry dangerous voltages 11ctrostatic discharge is a threat to instrument electronics 11ny parts are dangerously hot 12es 12eral warnings 13
and Regulatory Certifications 14
rmation 14bols 15
ctromagnetic compatibility 15nd Emission Certification for Federal Republic of Germany 16
17
ng 17
ing the Product 17
ction to Site Preparation 20Video supplements 10
Important Safety Warnings 11this Manual 105
6Space Requirements 22
Power Requirements 23
LinPow
SupReg
Ventin
Sampl
3 Installation
Install
Unpac
Place t
Conne
Conne
FloTyp
MPEPC
Contro
Con
Set
4 Running Checkout Sa
Runnin
siteprep.book Page 6 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
C connections 42 connections 46
l Setup 50
nect the control cables 50
communication options 52
mples
g Checkout Samples 56Set the flow mode on the Headspace Sampler 40Connect on/off valves to the gas input fittings 41ply tubing 25ulators 25
g Requirements 28
er Considerations 29
ation 32
k the Unit and Check the Contents 33
he Unit in Position 34
ct the Power Cord 35
ct the Plumbing and Transfer Line 37
w modes 37ical installation procedures 39Gas Requirements 24
External Plumbing 25e voltage 23er line cord 23
Site Prep and Installation Guide
Prepare the checkout sample 56Run the sample 576850/6890 GC checkout method 58
A Connecting a Swagel
Prepar
ObjMa
Proced
siteprep.book Page 7 Tuesday, February 17, 2004 9:18 AM7ok Fitting
ation 70
ective 70terials needed 70
ure 71
8siteprep.book Page 8 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
Agilent G1888 Network Headspace SamplerSite Prep and Installation Guide
1Introduction
About t
Vide
Importa
Man
Elec
Man
Gas
Gen
Safety
Info
Sym
Elec
Sou
Fuses
Cleanin
Recycli
This csafety
siteprep.book Page 9 Tuesday, February 17, 2004 9:18 AM9Agilent Technologies
his Manual 10
o supplements 10
nt Safety Warnings 11
y internal parts of the instrument carry dangerous voltages 11
trostatic discharge is a threat to instrument electronics 11
y parts are dangerously hot 12
es 12
eral warnings 13
and Regulatory Certifications 14
rmation 14
bols 15
tromagnetic compatibility 15
nd Emission Certification for Federal Republic of Germany 16
17
g 17
ng the Product 17
hapter gives an overview of the manual and provides information.
10
Introduction
About this Manual
This manual explains the requirements and procedures for installThe mperforpreparincludvariou
For opUser I
Video supplements
Click tvideo
siteprep.book Page 10 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guideing the Agilent Technologies G1888A Headspace Sampler. anual is intended for Agilent Service personnel ming the installation. It includes basic information for ing a site and for installing the Headspace Sampler, ing attaching it to the GC and setting flow rates for s inlet configurations.
erating directions, see the Agilent Headspace Sampler nformation CD ROM.
his icon when it appears throughout the manual to view supplements to certain procedures.
Introduction
Site Pr
Important Safety Warnings
Before moving on, there are several important safety notices that yoSampl
Many internal parts
If the
All
Electrostatic dischar
The pr
voltageunless
WARNING If the preplac
siteprep.book Page 11 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 11damaged by electrostatic discharge. Do not touch any of the boards unless it is absolutely necessary. If you must handle ge is a threat to instrument electronics
inted circuit (PC) boards in the instrument can be
be difficult to accidentally make contact with dangerous s. Unless specifically instructed to, never remove a cover
the heated zones are turned off.
ower cord insulation is frayed or worn, the cord must be ed. Contact your Agilent service representative. The internal wires and cables connected to these boards
The wires for any heater (such as the oven)
WARNING All these parts are shielded by covers. With the covers in place, it shouldn:
electronics boards in the instrumentpower switch is off, potentially dangerous voltages exist on:
The wiring between the instrument power cord and the AC power supply, the AC power supply itself, and the wiring from the AC power supply to the power switch
With the power switch on, potentially dangerous voltages also exist ou should always keep in mind when using the Headspace er.
of the instrument carry dangerous voltages
instrument is connected to a power source, even if the
12
Introduction
them, wear a grounded wrist strap and take other antistatic precautions. Wear a grounded wrist strap any time you must remove the electronics cover.
Many parts are dang
Many enouglimited
The
The
You shtempefirst setempeIf you and w
Gases
Do
Do Samto t
Weainju
Fasimm
Storele
siteprep.book Page 12 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide Do not put gas cylinders in the path of a hot air vent (including a GC oven exhaust).not use flammable carrier gases.
not use hydrogen as a carrier gas in the Headspace pler. Hydrogen creates a potential explosion hazard due
he venting of gases during operation and standby.
r eye protection when using compressed gas to avoid eye ry.
ten all compressed gas cylinders securely to an ovable structure or permanent wall.
re and handle compressed gases in accordance with vant safety codes.instrument that you will be maintaining before you begin working on it.
WARNING Be careful when working behind the instrument. During cool-down cycles, the instrument emits hot exhaust which can cause burns.erously hot
parts of the instrument operate at temperatures high h to cause serious burns. These parts include but are not to:
carousel and its contents
sample probe/loop assembly
ould always cool these areas of the instrument to room rature before working on them. They will cool faster if you t the temperature of the heated zone to room
rature. Turn the zone off after it has reached the setpoint. must perform maintenance on hot parts, use a wrench ear gloves. Whenever possible, cool the part of the
Introduction
Site Prep and Installation Guide
General warnings
Perform periodic leak checks on supply lines, fittings, and pneumatic plumbing to prevent a potentially hazardous con
To ato ldiscsolu
siteprep.book Page 13 Tuesday, February 17, 2004 9:18 AM13
dition.
void a potential shock hazard when using liquid solution ocate leaks, turn the main power switch off and onnect the main power cord. Be careful not to spill leak tion on electrical leads.
14
Introduction
Safety and Regulatory Certifications
The Headspace Sampler conforms to the following safety standa
Can
CSAUL
Inte
Eur
The inElectrInterfe
CIS
This ISISM es
The insystem
Information
The AfollowclassifCatego
This urecognthe insmanufbe impHeadsfrom aoperat
siteprep.book Page 14 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation GuideM device complies with Canadian ICES-001. Cet appareil t conforme a la norme NMB001 du Canada.
strument is designed and manufactured under a quality registered to ISO 9001.
gilent Technologies Headspace Sampler meets the ing IEC (International Electro-technical Commission) ications: Safety Class I, Transient Overvoltage ry II, Pollution Degree 2.
nit has been designed and tested in accordance with ized safety standards and is designed for use indoors. If trument is used in a manner not specified by the acturer, the protection provided by the instrument may aired. Whenever the safety protection of the Agilent pace Sampler has been compromised, disconnect the unit ll power sources and secure the unit against unintended ion. IEC/EN 61326
AUS/NZ rds:
adian Standards Association (CSA): C22.2 No. 1010.1
/Nationally Recognized Test Laboratory (NRTL): 61010A1
rnational Electrotechnical Commission (IEC): 610101
oNorm (EN): 610101
strument conforms to the following regulations on omagnetic Compatibility (EMC) and Radio Frequency rence (RFI):
PR 11/EN 55011: Group 1, Class A
Introduction
Site Prep and Installation Guide
Refer servicing to qualified service personnel. Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard.
Symbols
Warniduringinstrusafety
Electromagnetic com
This dOpera
1 Thi
2 Thi
See acinform
Indica
Indica
Indica
Indica
Indica
siteprep.book Page 15 Tuesday, February 17, 2004 9:18 AM15interference that may cause undesired operation.patibility
evice complies with the requirements of CISPR 11. tion is subject to the following two conditions:
s device may not cause harmful interference.
s device must accept any interference received, including
companying instructions for more ation.
tes a hot surface.
tes hazardous voltages.
tes earth (ground) terminal.
tes explosion hazard.
tes electrostatic discharge hazard.instrument. Agilent Technologies assumes no liability for the customers failure to comply with these requirements.ngs in the manual or on the instrument must be observed all phases of operation, service, and repair of this ment. Failure to comply with these precautions violates standards of design and the intended use of the
16
Introduction
If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try one or more of the
1 Rel
2 Mov
3 Pludevcirc
4 Mak
5 Makdev
6 Conexp
7 ChaTecequ
Sound Emission Cert
(Type
Schall
Schall(Typp
siteprep.book Page 16 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
test).
druckpegel
druckpegel LP < 70 dB(A) nach DIN-EN 27779 rufung).Sound pressure
Sound pressure Lp < 70 dB(A) according to DIN-EN 27779following measures:
ocate the radio or antenna.
e the device away from the radio or television.
g the device into a different electrical outlet, so that the ice and the radio or television are on separate electrical uits.
e sure that all peripheral devices are also certified.
e sure that appropriate cables are used to connect the ice to peripheral equipment.
sult your equipment dealer, Agilent Technologies, or an erienced technician for assistance.
nges or modifications not expressly approved by Agilent hnologies could void the users authority to operate the ipment.
ification for Federal Republic of Germany
Introduction
Site Pr
Fuses
Table 1 lists the fuses required for proper operation. These fuses s
Cleaning
Recycling the Product
Table 1
Fuse de
F2, F3
F3
F3, F4
F1
siteprep.book Page 17 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 17For recycling, contact your local Agilent sales office.To clean the unit, disconnect the power and wipe down with a damp, lint-free cloth.hould only be accessed by Agilent service personnel.
Fuses
signation Location Fuse rating and type
Power line module 10A 250V, glass tube
Terminal near transformer
10A 250V, glass tube
Power board 6A 125V, glass tube
Power board 1A 250V, glass tube
18
Introduction
siteprep.book Page 18 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
Agilent G1888A Headspace SamplerSite Preparation and Installation
2Site Preparation
Introdu
Temper
Perf
Opti
Space R
Power
Line
Pow
Gas Re
Externa
Sup
Reg
Venting
Sample
This cbefore
siteprep.book Page 19 Tuesday, February 17, 2004 9:18 AM19Agilent Technologies
ction to Site Preparation 20
ature and Humidity Requirements 21
ormance range 21
mum performance 21
equirements 22
Requirements 23
voltage 23
er line cord 23
quirements 24
l Plumbing 25
ply tubing 25
ulators 25
Requirements 28
r Considerations 29
hapter describes requirements and tasks to complete installing the Headspace Sampler in the laboratory.
20
Site Preparation
Introduction to Site Preparation
This chapter provides site preparation information for the Aglienfor:
Tem
Spa
Pow
Gas
Ext
Ven
ContaPrepapropercomplat wwHeads
siteprep.book Page 20 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guidet G1888A Headspace Sampler, covering the requirements
perature and humidity
ce
er
ernal plumbing
ting
ct an Agilent sales representative for the current Site ration Checklist. Use the checklist to ensure that the equipment and parts are present, and that installation is
eted properly. The checklist can also be found on the web w.agilent.com/chem. Click Library and search for the pace Sampler model number.
Site Preparation
Site Pr
Temperature and Humidity Requirements
Performance range
The Hfrom 5humid
The mturneddepentempeturnedtempe
Optimum performanc
The op(58 Fconde
siteprep.book Page 21 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 21eadspace Sampler operates in environment temperatures C to 45 C (41 F to 113 F) and 5% to 90% relative
ity .
inimum operating temperature for the oven while it is on is 40 C. However, the actual minimum temperature
ds on the environment. Generally, the minimum oven rature will equal 20 C above ambient. The oven can be off to use the Headspace Sampler closer to ambient
rature.
e
timum temperature and humidity are 20 C to 27 C to 80 F) and 50% to 60% relative humidity (without nsation).
22
Site Preparation
Space Requirements
The Headspace Sampler requires a laboratory bench with an unobs46 cm30 kg
The unabove the Heobstruinterfe
Other chromrequirinstrushows
Figure
64 cm
siteprep.book Page 22 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guidetructed surface approximately 64 cm (25 inches) deep by (18.1 inches) wide and capable of supporting at least (102 lb.).
it is approximately 56 cm (22 inches) high. The area it must be clear. Allow enough room to open the top lid of adspace Sampler. Shelves or any overhanging ctions limit access to the top of the instrument and re with cooling.
devices used with the Headspace Sampler (gas atographs, detectors, integrators, or workstations) will e additional bench space. See the manual for each ment for the applicable space requirements. Figure 1 instrument footprints for a sample installation.
1 Example of space needed on lab bench
Headspace Sampler
6890 GC
PC Monitor
PC Keyboard46 cm
190 cm
Site Preparation
Site Pr
Power Requirements
Line voltage
The Hsuppli
115750
230750
The H230 VAacceptmaximdedica
Power line cord
To progroundaccord(IEC)
When
recept
siteprep.book Page 23 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 23acle is properly grounded.three-conductor power line cord provides power and grounds the instrument. A properly grounded receptacle has its ground contact connected to a suitable earth ground. Verify that the eadspace Sampler operates from one of these line voltage es:
VAC single phase (90 VAC to 130 VAC), 50 to 60 Hz, VA max
VAC single phase (198 VAC to 264 VAC), 50 to 60 Hz, VA max
eadspace Sampler is shipped with the voltage set to C. The maximum power the Headspace Sampler can is 750 VA. The recommended neutral to ground level is a um of 3 VAC. The Headspace Sampler should have a ted ground.
tect users, the instrument panels and cabinet are ed through the three-conductor power line cord in ance with International Electrotechnical Commission
requirements.
plugged into a properly grounded receptacle, the
24
Site Preparation
Gas Requirements
The gas entering the Headspace Sampler becomes part of the
WARNING Do notpotentstandb
siteprep.book Page 24 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide use hydrogen in the Headspace Sampler. Hydrogen creates a ial explosion hazard due to venting during operation and y.carrier gas for the GC column. It may come from a pressure-regulated purified source (see Figure 3 on page 27) or from the GC itself. A gas supply connection for vial pressurization is also required. Gas at 550 kPa (80 psi) is recommended.
We recommend instrument or chromatographic purity grades specifically intended for chromatographic use. The gas purities in Table 2 are recommended for optimal chromatographic performance.
Table 2 Gas purity requirements
Gas Purity
Helium 99.9995% pure
Nitrogen 99.9995% pure
Site Preparation
Site Pr
External Plumbing
Supply tubing
The Helectrorespecsourceneede
If extrtubinginstrucontamcontam
rate fowhen instru
Gas coSwageSwage
Do nothread
Regulators
Two-ssupplipossibSuitab
If cylintempeat the
siteprep.book Page 25 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 25r the particular gas. Tubing of 1/8-inch od is adequate the supply is less than 3 meters (10 feet) away from the ment. Use 1/4-inch od tubing for longer distances.
nnection to the Headspace Sampler is made with 1/8-inch lok bulkhead fittings. Never use Teflon tape on lok-type fittings.
t use pipe dope or liquid thread sealant on any pipe s. Use only PTFE plumbing tape for these fittings.
tage regulators are recommended. Consult a local gas er for the type and size of cylinder valve connections. If le, obtain Compressed Gas Association (CGA) numbers. le two-stage regulators are also available from Agilent.
ders and regulators are stored outdoors where rature fluctuates, use an additional single-stage regulator back of the instrument.The required tube diameter depends on the distance from the supply or source to the Headspace Sampler and the total flow eadspace Sampler ships with plumbing kits for both nic and manual pressure control (EPC and MPC, tively). The kits contain enough tubing to reach a gas 3 meters from the Headspace Sampler. Extra plumbing is
d if the gas supply is not within 3 meters.
a plumbing is needed, use only preconditioned copper (part number 5180-4196) to supply gases to the ment. Ordinary copper tubing contains oils and
inants. Plastic tubing is permeable to oxygen and other inants that can damage columns and detectors.
26
Site Preparation
If a single gas source supplies several instruments (via a manifold or T-connections), the pressure in the lines will fluctuate. Add single-stage regulators at the back of each instru
Figure
Gas supply
Single-stage p
regulator
source
siteprep.book Page 26 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation GuideSet regulators to deliver 410 to 690 kPa (60 to 100 psi) to the instrument. Flow and pressure regulators require at least 140 kPa (20 psi) pressure drop across them to maintain regulation. See Table 3 for the correct regulator to use with each outlet fitting.
Table 3 CGA Numbers and Pressure Regulator Kit Part Numbers
Cylinder outlet fittingCGA number
Part number for pressure regulator kit with 1/8-inch Swagelok outlet
346 5183-4641
540 5183-4643
580 5183-4644
590 5183-4645ment to stabilize the pressure. See Figure 2 below.
2 Single stage regulator used to stabilize pressure
Headspaceunit
ressure
80-120 psig 60-100 psig550-830 kPa 410-690 kPa
Site Preparation
Site Pr
The gas to the Headspace Sampler becomes the carrier gas. Traps are recommended to protect the columns. See Figure 3.
Figure
Two-stage regulator
On/Off
Oxygen
Moistu
Hydroc
Gas sup
Gas sup
To HeaSample
siteprep.book Page 27 Tuesday, February 17, 2004 9:18 AM27ep and Installation Guide
3 Plumbing a pressure-regulated gas purified source
valve
trap
re trap
arbon trap
ply
ply on/off valve
dspacer
28
Site Preparation
Venting Requirements
During normal operation, some of the vial headspace gas is ventedcompoplace tatmos1/8-inthe loc
siteprep.book Page 28 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide4 Location of vent line on back of instrumentFigure
Vent line to the outside of the Headspace Sampler. If the nents of this gas are expected to be toxic or noxious, he unit within a fume hood or attach a vent system, at
pheric pressure, to the port on the sampler unit. Use ach Swagelok fitting for the connection. See Figure 4 for ation of the vent line.
Site Preparation
Site Pr
Sampler Considerations
One or more optional parts and kits may be required for the install
Theand232
To eflowmumodmod
To sPre
Theon aG23
siteprep.book Page 29 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 29ation, depending on the intended use.
Headspace Sampler ships with a deactivated flowpath a 1-mL sample loop. A 3-mL sample loop (part number 1700004) can be ordered separately.
nable software method control of the Headspace Sampler s and pressures, an Agilent 6890 or Agilent 6850 GC
st have an auxiliary electronic pressure control (EPC) ule. Order kit G1888-60705. The kit contains the EPC ule, the bleeder weldment, and the interface kit.
plice the transfer line into the inlet, order the Manual ssure Control Interface Kit G1290-60515.
re is an optional wire support for the transfer line for use n Agilent 6890 GC. Order Transfer Line Support Kit 19-60600.
30
Site Preparation
siteprep.book Page 30 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
Agilent G1888 Network Headspace SamplerSite Prep and Installation Guide
3Installation
Installa
Unpack
Place t
Connec
Connec
Flow
Typi
Set
Con
MPC
EPC
Con
Set
Control
This cin the installHeads
siteprep.book Page 31 Tuesday, February 17, 2004 9:18 AM31Agilent Technologies
tion 32
the Unit and Check the Contents 33
he Unit in Position 34
t the Power Cord 35
t the Plumbing and Transfer Line 37
modes 37
cal installation procedures 39
the flow mode on the Headspace Sampler 40
nect on/off valves to the gas input fittings 41
connections 42
connections 46
nect the control cables 50
communication options 52
Setup 50
hapter describes steps to install the Headspace Sampler laboratory. It is assumed that the GC has already been ed, is working properly, and is available for the pace Sampler Installation.
32
Installation
Installation
To ins
1 Unp
2 Pla
3 Con
4 Con(GC
5 Conopt
6 Per
ContaInstallpropercomplat wwHeads
CAUTION The Hecompaconnec
siteprep.book Page 32 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guidetall the Headspace Sampler:
ack the unit and check the contents. See page 33.
ce the unit in a suitable location. See page 34.
nect the power cord. See page 35.
nect the headspace sampler to the gas chromatograph ). See page 37.
nect the control cables to the GC and set communication ions. See page 50.
form the checkout procedure. See page 56.
ct an Agilent sales representative for the current ation Checklist. Use the checklist to ensure that the equipment and parts are present, and that installation is
eted properly. The checklist can also be found on the web w.agilent.com/chem. Click Library and search for the pace Sampler model number.
adspace Sampler contains metric fittings which are not tible with Swagelok or other fittings except for the external gas tions on the rear of the unit.
Installation
Site Pr
Unpack the Unit and Check the Contents
Inspect the shipping containers for damage. If a container is damagyour loinspec
Checkany diRetaincompl
siteprep.book Page 33 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 33ed or shows signs of stress, notify both the carrier and cal Agilent office. Keep all shipping materials for tion by the carrier.
the items received against the packing lists. If there are screpancies, notify your local Agilent office immediately. the shipping containers until all contents are checked for eteness and instrument performance is verified.
34
Installation
Place the Unit in Position
1 Place the Headspace Sampler on a bench top to the left of the GC.
2 FreSamUsenec
3 Poseasily reach the GC inlet without kinks or sharp bends.
CAUTION The spobstructop of t
Figure
Release
siteprep.book Page 34 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guidee the transfer line from the side of the Headspace pler. Open the plastic collar holding the line in place. a small flat blade screwdriver to press the release tab if essary. See Figure 5.
ition the Headspace Sampler so that the transfer line can
ace above the Headspace Sampler must be free of overhanging tions that could interfere with cooling and limit access to the he instrument or prevent the top from opening.
5 The transfer line release tab
tab
Installation
Site Pr
Connect the Power Cord
1 Verify the proper operating voltage setting. The Headspace Samneeloca
2 Plugro
Figure
Voltage
Fuses
siteprep.book Page 35 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 35pler is shipped with the voltage set to 230 VAC. If ded, set the voltage selector to 115 VAC. Also note the tion of the fuses. See Figure 6.
g the Headspace Sampler power cord into a properly unded outlet. See page 23 for power requirements.
6 Power connection
selector
36
Installation
3 Turn on the Headspace Sampler to verify that it is working properly. The Headspace Sampler performs a self-test when it starts up. Verify that each test is labeled OK. The self-test infodia
a P
b SP
c S
Usescre
See thdescri
Figure
Shu
Tray
Caro
Oven
Loop
siteprep.book Page 36 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
rmation can also be viewed at any time from the gnostic screen.
ress the Menu key.
croll down to Advanced Functions with the cursor keys. ress Enter.
croll down to Diagnostic and press Enter.
the cursor arrows to scroll through the multiple test ens. A typical diagnostic test screen is shown in Figure 7.
e G1888A Headspace Sampler operating manual for a ption of the system diagnostic messages.
7 Self-test screen
DIAGNOSTIC
NOT READY
tter:
usel OK
Sensor
OK
OK
OK
Sensor OK
SELF TEST
Installation
Site Pr
Connect the Plumbing and Transfer Line
Flow modes
The Aflow m
Ele
EPCconvialsetp
Man
MPregucar
See Filocatio
Figure
Pressur
Flow co
siteprep.book Page 37 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 37gilent G1888A Headspace Sampler can operate in two odes:
ctronic pressure control (EPC)
mode requires an EPC inlet and an Auxiliary EPC trol module in the GC. The modules provide carrier and pressurization gas control via GC or Agilent data system oints.
ual pressure control (MPC)
C mode uses the manual flow controller and pressure lator on the G1888A Headspace sampler to control the
rier and vial pressurization gases.
gure 8 for the MPC flow controller and pressure regulator ns.
8 Flow controller and pressure regulator location
e regulator
ntroller
38
Installation
There are several ways to connect the Agilent Headspace Sampler to the GC. The type of connection depends on the GC model, the selected flow mode, and how the instruments are to be use
EPC co
Direct insertreal mconsisduplicnorma
Direct the sa6890 G
MPC c
Needlefrom tinsertinjecti
Needlefrom t23225
siteprep.book Page 38 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
d. The possibilities include:
nnections
connection to a split/splitless inlet The sampler is ed between the inlet flow module and the inlet, allowing ethod control, via the GC, of flows and pressures for tent results between setups. The system is easy to ate and monitor but requires time to convert back to l inlet injection for nonheadspace samples.
connection to a volatiles interface The transfer line from mpler is connected to the volatiles interface. See the Cs user information for instructions.
onnections
connection to a split/splitless inlet The transfer line he sampler ends in a needle (part no. 2322590005) that is ed through the inlet septum. Conversion to normal on is rapid and simple.
connection to a cool on-column inlet The transfer line he sampler ends in a small od needle (part no. 90004) that is inserted through the inlet septum.
Installation
Site Prep and Installation Guide
Typical installation procedures
For MPC installations with a split/splitless inlet:
1 VerSam
2 ConHea
3 Consou
4 Conspli
For EP
1 Set to E
2 ConHea
3 Conmod
4 Conspli(pa
Altern
To insGC, co
siteprep.book Page 39 Tuesday, February 17, 2004 9:18 AM39
ify that the flow mode toggle switches on the Headspace pler are set to MPC (page 40).
nect the on/off valves to the gas input fittings on the dspace Sampler (page 41).
nect the gas input fittings to a pressure-regulated gas rce (page 42).
nect the transfer line using a needle connection to a t/splitless inlet (page 43).
C installations with a split/splitless inlet:
the flow mode toggle switches on the Headspace Sampler PC (page 40).
nect the on/off valves to the gas input fittings on the dspace Sampler (page 41).
nect the vial pressurization gas to the GC auxiliary EPC ule (page 46).
nect the carrier gas and transfer line directly to the t/splitless inlet by splicing into the carrier gas line ge 48).
ative installations
tall the Headspace Sampler on a 4890D or 5890 Series II ntact Agilent Service.
40
Installation
Set the flow mode on the Headspace Sampler
Choose the flow mode for the carrier gas and vial pressurization gas and set the flow mode toggle switches on the headspace
Figure
siteprep.book Page 40 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
hessampler.
1 Open the lid of the Headspace Sampler. Locate the two tabs near the top of the Headspace Sampler on the left and right sides. Push the tabs towards the back of the instrument and lift the lid. Use caution when opening the lid due to its weight. The lid locks in place when fully opened.
2 Set the toggle switches for carrier gas and vial pressurization gas on the underside of the Headspace Sampler lid to the necessary positions. See Figure 9. The factory default setting is MPC.
9 The pressure control toggle switc
Vial pressurization gas
Carrier gas
MPC
EPC
Installation
Site Prep and Installation Guide
Connect on/off valves to the gas input fittings
Install two on/off valves and tubing on the gas input fittings as shown in Figure 10. The figure shows a connection to the MPC
Figure
Carri
Vial
On/Ofor le
For M
For E
siteprep.book Page 41 Tuesday, February 17, 2004 9:18 AM41input fittings.
For EPC, use the fittings on the right
For MPC, use the fittings on the left
10 On/Off valves connected
er gas in
pressurization gas in
ff valvesak testing
PC
PC
42
Installation
MPC connections
Carrier gas and vial pressurization gas connections
If bothInstallpressuFigure
Figure
CarrierVial pr
siteprep.book Page 42 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation GuideIf only one gas input fitting uses MPC, run a line from a pressure-regulated gas source directly to the on/off valve connected to the MPC gas input fitting.
11 Carrier and vial pressure connections
gas inessure
T-fitting
gas source
On/Off valves
regulated the carrier gas and vial pressurization gas are MPC, a T-fitting to the on/off valves. Run a line from a re-regulated gas source to the T-fitting as shown in 11.
To pressure
Installation
Site Prep and Installation Guide
Needle connection from transfer line to a split/splitless inlet
If the Headspace Sampler uses MPC for the carrier gas, the transfer line connects to the GC inlet by a needle passing throug
This sydoes nmoduldilutioinlet c
The Houter-This nod neeinlet a
A smafor onOrder
Followprocedfor a 6
1 Instspli
2 Rem
3 Attwre
4 Inse
siteprep.book Page 43 Tuesday, February 17, 2004 9:18 AM43
h the inlet septum.
stem easily converts to normal injection. The installation ot disturb the normal plumbing between the inlet flow e and the inlet. Normal inlet flow reduces to minimize n of the headspace sample. Most of the flow through the omes from the sampler.
eadspace Sampler shipping kit includes a 0.7 mm diameter (od) transfer line needle (part no. 2322590005). eedle is also called a transfer line terminal. The 0.7 mm dle is most appropriate for use with the purged/packed nd the split/splitless inlet.
ller diameter needle (part no. 2322590004) must be used -column injection and may be used with other inlets. this needle separately.
these steps to connect the transfer line to the GC. This ure is described for the 6850 GC. Installation is similar 890 GC.
all a deactivated direct liner (part no. 5181-8818) in the t/splitless inlet.
ove the standard septum nut from the inlet.
ach the transfer line to the needle. Use two 7-mm nches to tighten it.
rt the needle through the strain relief coupling.
44
Installation
5 Push an 11-mm through-hole septum onto the needle. Carefully remove any pieces of septum cored by the needle. Be sure the needle bore is clear. Figure 12 shows the asse
Figure
End of t
Top of
siteprep.book Page 44 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
mbly.
12 The needle assembly
ransfer line
inlet
Installation
Site Prep and Installation Guide
6 Insert the tip of the needle into the inlet. Force the septum (it is oversized) into the recess in the top of the inlet. See Figure 13 for the assembled view.
7 Scr
When using contrometerrate.
Figure
Septum
siteprep.book Page 45 Tuesday, February 17, 2004 9:18 AM45
ew on and tighten the strain relief coupling.
operating, set a small flow of carrier gas through the inlet the GC inlet control. Set most of the flow using the flow ller in the Headspace Sampler. Use an electronic flow
connected to the inlet split vent to verify the actual flow
13 The transfer line attached to the inlet
is forced into recess
46
Installation
EPC connections
Connect the vial pressurization gas for an EPC setup
An Augas. ThEPC cfor a 6
1 If thG19pro
2 InstG18inst
3 Locthepne
4 Remmodthe
Figure
Frits
Output
siteprep.book Page 46 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
xiliary gas channel in the GC supplies vial pressurization e following steps describe installation for a 6890 GC with
ontrol of headspace. The installation procedure is similar 850 GC.
e GC does not have an Auxiliary EPC module, install the 40A Headspace Interface Kit. Follow the instructions vided with that kit.
all the external sampler interface kit (part no. 88-60705) into the GC. Follow the procedure for alling onto a back-pressure regulated system.
ate the block on the Auxiliary EPC module that connects three gas outlet tubes for the auxiliary channels to the umatics module.
ove the screw that holds the block to the pneumatics ule. Pull the block free of the module and rotate it so that
frits are on top. See Figure 14.
14 The Auxiliary EPC module gas outlet block
Aux 5
Aux 4
Aux 3
Aux 3
Aux 4
Aux 5
s
Installation
Site Prep and Installation Guide
5 Choose a channel for the vial pressurization control. Remove that channels frit from the block. Remove the O-ring that seals the channel.
Figure
Heads
GC
Vial pr
To Aux
Bleed w
Brass
siteprep.book Page 47 Tuesday, February 17, 2004 9:18 AM47
ler to the auxiliary module6 Place an O-ring on a zero-resistance brass tube frit (part no. G1570-20540). Place the O-ring/frit combination in the block.
7 Reconnect the block to the pneumatics module. Tighten the screw.
8 Use tubing and Swagelok fittings to connect the appropriate Auxiliary module output, the bleed weldment, and the VIAL PRESSURE fitting to the on/off valve on the back of the Headspace Sampler as shown in Figure 15. Do not disturb the brass end on the bleed weldment.
15 Connecting the Headspace Samp
pace Sampler
essure fitting
iliary module output
eldment
end
48
Installation
Direct connection from the transfer line to a split/splitless inlet
The GC inlet control channel supplies carrier gas. The following steps describe installation for a 6890 GC with EPC control of headsp
1 Inst
2 LocFigu
3 Cut2 in
a C
b Hf
c F
4 Slid
Figure
Carrier
siteprep.book Page 48 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
ace. The installation procedure is similar on a 6850 GC.
all a split inlet liner (part no. 5183-4647) in the inlet.
ate the carrier line to the split/splitless inlet shown in re 16.
the carrier line to the split/splitless inlet about 1 to ches from the inlet body.
arefully score the circumference of the tubing.
old the tubing with pliers on each side of the score and lex it until it breaks free.
ile the edges until they are smooth.
e the nut onto the inlet side of the carrier line.
16 The split/splitless inlet carrier gas line
gas line
Installation
Site Prep and Installation Guide
5 Attach a zero-dead-volume (ZDV) union to the short piece of carrier line attached to the inlet as shown in Figure 17.
6 AttSee
7 Usecarshiprea
8 Con
Figure
Figure
ZDV un
Carrier
Ferrule
Nut
ZDV un
Transfe
siteprep.book Page 49 Tuesday, February 17, 2004 9:18 AM49
ach the headspace sampler transfer line to the ZDV union. Figure 18.
the other ZDV union to connect the free end of the rier gas line to the length of tubing provided in the ping kit. Route the tubing out the back of the GC to the
r panel of the sampler.
nect the tubing to the CARRIER fitting on the sampler.
17 Attaching the ZDV union to the carrier line
18 Connecting the transfer line to the carrier line
ion
line
ion
r line
50
Installation
Control Setup
Connect the control cables
ConneBe surthe fre
Depenneedeused w
Figurecablinsoftwa
.
Table 4
Sample
LAN*
Remote
APG Y-c
Serial (R
Remote
*Throug
siteprep.book Page 50 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guidect the remote start/stop cable to the back of the sampler. e to connect the spade lug to the ground screw. Connect e end of the cable to the GC.
ding on the setup, other cable connections may be d. Refer to Table 4 for an overview of cable connections ith the Headspace Samplers.
19 shows some common cabling configurations. These g configurations are valid for the G2070AA and G2922AA re set.
Cables for Headspace Samplers
r port Connect to Part number
Hub* 8121-0940
6820, 6850 and 6890 GCs G1530-60930
able 6890N and 5973N G1530-61200
S-232) PC G1530-60600
4890 and 5890 Series II GCs 35900-60700
h a network hub or similar connection.
Installation
Site Pr
Figure
21
1 1Cables
1
2
3
4
5
siteprep.book Page 51 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 51
19 Common cabling configurations
Workstation
Printer
Hub or switch
G1888A 6890N
Workstation
Printer
Hub or switch
G1888A 6850
21
1 1
Local Talk
Local Talk
BNC
BNC
Workstation
Printer
G1888A 6820
23
5 3
Workstation
Printer
Hub or switch
G1888A 5973N 6890N
421
1 11
8121-0940, Cable, 7.5-m 100-BaseT LAN
G1530-60930, 2-m APG remote cable, 9-pin male to 9-pin male
.
.
G1530-60600, 2-m RS-232 cable, 9-pin female to 9-pin female
G1530-61200, 2-m APG Y-cable, two 9-pin male, one 9-pin female
8121-1013, USB-DB9 RS-232 adapter cable
.
.
.
52
Installation
Set communication options
If using RS-232 or LAN communications, configure the communications as described below.
LAN s
If the AddreFunct
1 Pre
2 Scr
3 Scrpre
4 Theand
After cusing
1 Pre
2 Scr
3 Scrpre
Figure
siteprep.book Page 52 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
ettings
Headspace Sampler is connected to a LAN, set the IP ss, subnet mask, and gateway using the Advanced ions menu.ss the Menu key.
oll down to Advanced Functions. Press Enter.
oll down to LAN Config. and press Enter or ss . followed by 4 on the numeric keypad.
LAN Config. screen is shown in Figure 20. Use the cursor keypad to enter new values. Press Enter to set the value.
onfiguring the LAN settings, disable the RS-232 port the Advanced Functions menu. ss the Menu key.
oll down to Advanced Functions. Press Enter.
oll down to Enable RS232 and press Enter or ss . followed by 6 on the numeric keypad.
20 The LAN Config. screen
SET
IP Address
Sub MaskRunning
LAN Config.
Gateway
10 .1 .1 .105
255 .255 .255 .0
10 .1 .1 .1
Installation
Site Prep and Installation Guide
4 The Enable RS232 screen is shown in Figure 21. Use the cursor to choose No from the ADV. FUNCT list. Press Enter.
Headsthe RS
Additio
If usinenable
1 Pre
2 Scr
3 Scrpre
Figure
siteprep.book Page 53 Tuesday, February 17, 2004 9:18 AM53
pace Sampler automatically resets after after disabling -232 port.
nal RS-232 settings
g RS-232 communications with the Headspace Sampler, RS-232 capabilities using the Advanced Functions. ss the Menu key.
oll down to Advanced Functions. Press Enter.
oll down to Enable RS232 and press Enter or ss . followed by 6 on the numeric keypad.
21 The Enable RS232 screen
13. Zones Calibration
14. LAN Configuration
15. MAC AddressYes
ADV. FUNCTION
No16. Enable RS232
54
Installation
4 The Enable RS232 screen is shown in Figure 22. Use the cursor to choose Yes from the ADV. FUNCT list. Press Enter.
Figure
siteprep.book Page 54 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
22 The Enable RS232 screen
13. Zones Calibration
14. LAN Configuration
15. MAC AddressYes
ADV. FUNCTION
No16. Enable RS232
Agilent G1888 Network Headspace SamplerSite Prep and Installation Guide
4Running Checkout Samples
Runnin
Prep
Run
6850
This cSampl
siteprep.book Page 55 Tuesday, February 17, 2004 9:18 AM55Agilent Technologies
g Checkout Samples 56
are the checkout sample 56
the sample 57
/6890 GC checkout method 58
hapter describes the checkout steps for the Headspace er.
56
Running Checkout Samples
Running Checkout Samples
Prepare the checkout sample
The chethanocapilla
The chheadsp20 mLinstru
1 Snacon
2 Dis
3 HolsampraFigu
4 Thecombotrem
5 Drocap
Figure
1. Gr
siteprep.book Page 56 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guideeckout sample is three detector-specific compounds in l. The box contains 1 mL of sample and a bottle of 5-L ry pipettes.
eckout method calls for a 5-L sample in a 20-mL ace vial. The default Headspace Sampler vial size is
. See the Headspace Sampler Operating Manual for ctions to set the vial size if needed.
p the top off the checkout sample ampoule. Transfer the tents to a screw-cap bottle.
pense one capillary pipette
d the pipette with clean tweezers. With the checkout ple bottle and the pipette as close to horizontal as
ctical, dip the end of the pipette into the sample. See re 23
pipette will fill by capillary action. When it has filled pletely, pull the pipette out of the sample. Wipe the
tom edge of the pipette against the edge of the vial to ove any droplets on the outside.
p the filled pipette into a 20-mL headspace vial. Crimp a on the vial.
23 Filling a pipette
asp 2. Fill 3. Wipe
Running Checkout Samples
Site Prep and Installation Guide
Run the sample
1 Set the method parameters on the GC and Headspace sampler. See Table 5 on page 58 for checkout settings or load
siteprep.book Page 57 Tuesday, February 17, 2004 9:18 AM57the stored checkout method as described in 6850/6890 GC checkout method on page 58. If needed, install the checkout column and liner.
2 If using MPC, set the carrier gas flow as follows:
a Set the GC inlet split ratio and column head pressure to the values given in Table 7 on page 61.
b Set the split flow to 4 mL/min. The total flow will read 11 mL/min.
c With an electronic flow meter connected to the GC split vent, turn up the headspace carrier gas flow until the flow rate out the split vent is approximately 80 mL/min. Remove the flow meter. See Figure 24.
NOTE The GC and Headspace pressure readings should be similar. Record the headspace pressure readingit can be used to set up future checkout runs.
58
Running Checkout Samples
The repages
6850/6890 GC check
Load t
The Hthe chchecko
1 Pre
2 HigEnt
3 UseFun
Figure
6850 Ser i esGC System
Split ve
Electronic flowmeter
siteprep.book Page 58 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
sults should be similar to the chromatograms shown on 65 to 68.
out method
he existing checkout method
eadspace Sampler ships with a stored method containing eckout settings. Do the following steps to load the ut method.
ss the Menu key.
hlight Advanced Functions with the cursor keys. Press er.
the cursor keys to select Stored Meth. (Advanced ction number 7) and press Enter. 3 Run the checkout sample.24 The electronic flow meter connected
nt
6850 GC shown
Running Checkout Samples
Site Prep and Installation Guide
4 The screen shown in Figure 25 appears. Use the cursor keys to select Checkout. Press Enter.
Create
If the acceptto creasectionthat ap
Figure
Table 5
Headsp
Oven tem
Valve te
Transfer
Loop vol
Vial pres
GC cycle
Vial equ
Pressur
siteprep.book Page 59 Tuesday, February 17, 2004 9:18 AM59
a new checkout method
checkout settings in the stored methods screen are not able, use Tables 5 through 12 and Figures 26 through 29 te a new Headspace Sampler checkout method. This covers several inlets and detector types. Use the settings ply.
25 The Stored Method screen
Headspace Sampler settings
ace method
perature 100 C
mperature 110 C
line temperature 115 C
ume 1 mL
sure 15 psi (See Table 11 if using EPC control)
time 5.5 min
ib time 7.0 min
ization time 0.08 min
DIAGNOSTIC
READY
Checkout
OQ/PV
MeOH in H2O
STORED METH.Default
60
Running Checkout Samples
Loop fill
Loop eq
Injection
Table 5 Headspace Sampler settings (continued)
Headspace method
siteprep.book Page 60 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
time 0.5 min
uib time 0.05 min
time 1.0 min
Running Checkout Samples
Site Prep and Installation Guide
Table 6 GC oven settings
Oven
Initial te
Initial tim
Rate
Maximu
Equilibra
Run tim
Table 7
Inlet (vo
Mode
Initial te
Pressur
Split rat
Split flo
Total flo
Gas sav
Gas type
Liner
siteprep.book Page 61 Tuesday, February 17, 2004 9:18 AM61
mperature 100 C
e 4.00 min
0 (Off)
m temperature 300 C
tion time 1.00 min
e 4.00 min
Inlet settings
latiles or split/splitless)
Split
mperature 250 C (On)
e 19.5 psi (On)
io 20:1 (G1289)
w 80 mL/min (G1289)
w 87 mL/min (G1289)
er Off
Helium
5181-8818
62
Running Checkout Samples
Table 8 Column settings
Column
Model n
Maximu
Length
Diamete
Film thic
Mode
Initial flo
Initial pr
Average
Outlet p
siteprep.book Page 62 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
umber 689019091J-413, HP-5 5% Phenyl methyl silicone685019091Z-413E, HP-1 Methyl silicone
m temperature 325 C
30 m
r 320 m
kness 0.25 m
Constant flow
w 4.0 mL/min
essure 19.5 psi
velocity 60 cm/s
ressure Ambient
Running Checkout Samples
Site Prep and Installation Guide
Table 9 Detector settings (FID, NPD, ECD)
Detector parame
FID NPD ECD
Tempera
Hydroge
Air flow
Mode
Makeup
Makeuptype
Flame
Electrom
Lit offse
Adjust o
Bead
Equilibratime
siteprep.book Page 63 Tuesday, February 17, 2004 9:18 AM63
ters
ture 300 C 300 C
n flow 30 mL/min 2.0 mL/min
400 mL/min 60 mL/min Constant makeup
Constant makeup Constant makeup
30 mL/min
flow 25 mL/min 8 mL/min Nitrogen or argon/methane
gas Nitrogen or helium Nitrogen or helium
On On
eter On On
t 2.0
ffset 30
On
tion 0
64
Running Checkout Samples
Use thcontro
Table 10 5973 MSD settings
5973 MSD unique parameters
Column
Solvent
Run tim
HeadspaGC cycle
Scan ran
Method
Inlet pre
Column
Split rat
Injector
Table 1
Aux Pre
Gas type
Initial pr
Initial tim
Rate
siteprep.book Page 64 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
e settings given in Table 11 when using a GC with EPC l.
6890/5973A MSD 19091S-433, HP-5MS 5% Phenyl methyl siloxane
delay 3 min
e 6.5 min
ce event time for time
7 min
ge 45 to 300
used Atune file
ssure 12 psi
flow 1.1 mL/min with foreline reading of 60
io 80:1
External device
1 GC Aux pressure settings for EPC control
ssure 3 (for vial pressurization on G1888A)
Helium
essure 15 psi (On)
e 0 min
0 (Off)
Running Checkout Samples
Site Prep and Installation Guide
Table 12 Checkout sample
Checkout sample
Figure
0.0
pA
20
40
60
80
100
120
140
160
FID
siteprep.book Page 65 Tuesday, February 17, 2004 9:18 AM65Type Headspace OQ/PV Standard
Part number 5182-9733
Size 5 L (by micro-pipette) in 20-mL headspace vial
26 FID checkout chromatogram
min0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0
1 A, (HS_ECD\HSVI0004.D)
0.8
58
0.8
80
2.0
37
2.4
76
2.8
25
Ethanol(40,000 counts typical)
Dichlorobenzene
t-Butyl disulfide
Nitrobenzene
66
Running Checkout Samples
Figure
0.0
pA
50
100
150
200
250
NPD1 B, (HS_NPD\HSNPD002.D)
2.3
84
siteprep.book Page 66 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
27 NPD checkout chromatogram
min0.5 1.0 1.5 2.0 2.5 3.0 3.5
Nitrobenzene
Ethanol
Running Checkout Samples
Site Pr
Figure
0
Hz
0
20000
40000
60000
80000
100000
120000
ECD1 A, (HS_ECD\HSEC0000.D)
2.0
21
siteprep.book Page 67 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 67
28 ECD checkout chromatorgram
min1 2 3 4
0.8
50
2.4
61
2.8
10
Dichlorobenzene
Ethanol
68
Running Checkout Samples
Figure 29 MSD checkout chromatorgram
Dichlorobenzene
NitrobenzeneAbu
ndan
ce
1000000
800000
600000
400000
200000
03.20 3.60 4.00 4.40
siteprep.book Page 68 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation GuideTIC: 0201002.D
t-Butyl disulfide
Time
4.80 5.20 5.60 6.00 6.40
Agilent G1888 Network Headspace SamplerSite Preparation and Installation
AConnecting a Swagelok Fitting
Prepara
Proced
This c
siteprep.book Page 69 Tuesday, February 17, 2004 9:18 AM69Agilent Technologies
tion 70
ure 71
hapter describes proper Swagelok connections.
70
Connecting a Swagelok Fitting
Preparation
Objective
To mawithou
Materials needed
1/8
1/8
Fro
Twonut
siteprep.book Page 70 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guideke a leak-free tubing connection that can be taken apart t damaging the fitting
-inch (or 1/4-inch, if used) preconditioned copper tubing
-inch (or 1/4-inch, if used) Swagelok nuts
nt and back ferrules
7/16-inch (for 1/8-inch nuts) or 9/16-inch (for 1/4-inch s) wrenches
Connecting a Swagelok Fitting
Site Pr
Procedure
1 Place a Swagelok nut, back ferrule, and front ferrule to the tub
2 Cla
3 Pus
4 Makthe
Figure
Front
Back
Nut
Tubin
Figure
Plug
Nut
held
siteprep.book Page 71 Tuesday, February 17, 2004 9:18 AMep and Installation Guide 71h the tubing into the stainless steel plug (Figure 31).
e sure that the front ferrule is touching the plug. Slide Swagelok nut over the ferrule and thread it onto the plug.
s very costly to repair.
31 Assembling the fitting
or fitting
and ferrules
Tubing in a viseCAUTIONDo not
Use a separate stainless steel fitting in a vise for initial tightening of the nut. Do not use an inlet or detector fitting. Strong forces are required to properly set the ferrules, and damage to an inlet or detector fitting iing as shown in Figure 30.
mp a stainless steel plug or similar fitting in a bench vise.
30 Swagelok nuts and ferrules
ferrule
ferrule
g
IMPORTANT!The narrow end of the backferrule fits into the rear ofthe front ferrule.
72
Connecting a Swagelok Fitting
5 Push the tube fully into the plug, then withdraw it approximately 1 to 2 mm. Finger-tighten the nut (Figure 32).
6 Mar
7 ForwreFortigh
Figure
Figure
Fronferr
Pencil l
siteprep.book Page 72 Tuesday, February 17, 2004 9:18 AMSite Prep and Installation Guide
k the nut with a pencil line (Figure 33).
1/8-inch Swagelok fittings, use a pair of 7/16-inch nches to tighten the fitting 3/4 of a turn (Figure 34). 1/4-inch fittings, use a pair of 9/16-inch wrenches to ten them 1-1/4 turn (Figure 34).
32 Insert the tubing
33 Marking the fitting
tule
Backferrule
Nut
Insert tubing fully
Withdraw tubing 1-2 mm
Tighten nut
ine
Connecting a Swagelok Fitting
Site Prep and Installation Guide
8 Remnutthefitti
9 A cthacruferr
Figure
Figure
siteprep.book Page 73 Tuesday, February 17, 2004 9:18 AM73
ove the plug from the fitting. To connect the tubing, with and ferrules, to another fitting, finger-tighten the nut, n use a wrench to tighten it 3/4 of a turn (1/8-inch ngs) or 1-1/4 of a turn (1/4-inch fittings).
orrectly-swaged connection is shown in Figure 35. Note t the end of the tubing in a correctly-swaged fitting is not shed and does not interfere with the action of the ules.
34 Final tightening
35 Completed fitting
Agilent Technologies
siteprep.book Page 74 Tuesday, February 17, 2004 9:18 AM Agilent Technologies, Inc.
Printed in USA, March 2004
ContentsIntroductionAbout this ManualVideo supplements
Important Safety WarningsMany internal parts of the instrument carry dangerous voltagesElectrostatic discharge is a threat to instrument electronicsMany parts are dangerously hotGasesGeneral warnings
Safety and Regulatory CertificationsInformationSymbolsElectromagnetic compatibilitySound Emission Certification for Federal Republic of Germany
FusesCleaningRecycling the Product
Site PreparationIntroduction to Site PreparationTemperature and Humidity RequirementsPerformance rangeOptimum performance
Space RequirementsFigure 1 Example of space needed on lab bench
Power RequirementsLine voltagePower line cord
Gas RequirementsExternal PlumbingSupply tubingRegulatorsFigure 2 Single stage regulator used to stabilize pressureFigure 3 Plumbing a pressure-regulated gas purified source
Venting RequirementsFigure 4 Location of vent line on back of instrument
Sampler Considerations
InstallationInstallationUnpack the Unit and Check the ContentsPlace the Unit in PositionFigure 5 The transfer line release tab
Connect the Power CordFigure 6 Power connectionFigure 7 Self-test screen
Connect the Plumbing and Transfer LineFlow modesFigure 8 Flow controller and pressure regulator location
Typical installation proceduresSet the flow mode on the Headspace SamplerFigure 9 The pressure control toggle switches
Connect on/off valves to the gas input fittingsFigure 10 On/Off valves connected
MPC connectionsFigure 11 Carrier and vial pressure connectionsFigure 12 The needle assemblyFigure 13 The transfer line attached to the inlet
EPC connectionsFigure 14 The Auxiliary EPC module gas outlet blockFigure 15 Connecting the Headspace Sampler to the auxiliary moduleFigure 16 The split/splitless inlet carrier gas lineFigure 17 Attaching the ZDV union to the carrier lineFigure 18 Connecting the transfer line to the carrier line
Control SetupConnect the control cablesFigure 19 Common cabling configurations
Set communication optionsFigure 20 The LAN Config. screenFigure 21 The Enable RS232 screenFigure 22 The Enable RS232 screen
Running Checkout SamplesRunning Checkout SamplesPrepare the checkout sampleFigure 23 Filling a pipette
Run the sampleFigure 24 The electronic flow meter connected
6850/6890 GC checkout methodFigure 25 The Stored Method screenFigure 26 FID checkout chromatogramFigure 27 NPD checkout chromatogramFigure 28 ECD checkout chromatorgramFigure 29 MSD checkout chromatorgram
Connecting a Swagelok FittingPreparationObjectiveMaterials needed
ProcedureFigure 30 Swagelok nuts and ferrulesFigure 31 Assembling the fittingFigure 32 Insert the tubingFigure 33 Marking the fittingFigure 34 Final tighteningFigure 35 Completed fitting
MPC transfer line: EPC connection: