G.8262 SyncE Conformance TestingApplications User Guide
Citation preview
Calnex Solutions Ltd Page 1 of 40
Reg. SC299625
G.8262 SyncE Conformance Testing
Applications User Guide
1. Hardware and Software required .............................................................................................................. 2
2. Connecting an EEC to the Paragon-X ......................................................................................................... 3
3. How to Configure the Paragon-X for ALL G.8262 Tests ............................................................................. 5
4. Measuring Frequency Accuracy G.8262 Section 6 .................................................................................. 8
5. Pull-in, Hold-in, and Pull-out ranges G.8262 Section 7 ......................................................................... 10
6. Wander (Noise) Generation G.8262 Section 8 ..................................................................................... 13
7. Jitter Generation G.8262 Section 8.3 .................................................................................................... 16
8. Wander (Noise) Tolerance G.8262 Section 9 ........................................................................................ 17
9. Jitter Tolerance G.8262 Section 9.2 ...................................................................................................... 22
10. Wander (Noise) Transfer G.8262 Section 10 ........................................................................................ 23
11. Phase Transient Response G.8262 Section 11 ...................................................................................... 31
12. Appendix 1 - G.8262 Testing; Practical interpretation guidance ............................................................. 36
Calnex Solutions Ltd Page 2 of 40
Reg. SC299625
1. Hardware and Software required
Paragon-X
o Option 110 GbE electrical and Optical
o Option 111 10GbE interface (optional)
o Option 120 Delay and Header Capture
o Option 213 SyncE measurements
o Option 223 MTIE/TDEV internal wander generation
o Paragon- X Software X.10.14 or higher
Synchronisation (Reference) Frequency Source
Calnex Solutions Ltd Page 3 of 40
Reg. SC299625
2. Connecting an EEC to the Paragon-X
The front panel of the Paragon-X provides the following interfaces for testing:
1. 100MbE Electrical or Optical (SGMII SFP)
2. 1GbE Electrical or Optical (SFP) with option 110 fitted
3. 10GbE Optical (XFP or SFP+) with option 111 fitted
Paragon-X Front Panel Connections
10GbE Port 1
100MbE/1GbE Port 1
10GbE Port 2
100MbE/1GbE Port 2
Status Display
PC Controller Port
Calnex Solutions Ltd Page 4 of 40
Reg. SC299625
The Paragon-X accepts the following reference clocks which should be applied to one of the
Reference Inputs on the back panel:
1. 2.048MHz
2. 10MHz
3. E1 (2.048Mb/s)
4. DS1 (T1) (1.544Mb/s)
Fig. 1 Paragon-X Reference Inputs (Back Panel)
There are two reference inputs
BNC with 75 impedance (the connector on the left in the figure).
Bantam connector with 100 impedance.
o 3 pole bantam jack o 4.39mm Diameter o Tip; Positive, Ring; Negative, Sleeve; Ground
Calnex Solutions Ltd Page 5 of 40
Reg. SC299625
3. How to Configure the Paragon-X for ALL G.8262 Tests
a) Verify your physical connections have been completed per section 2.
b) Start the Paragon-X GUI
c) Select the Operating Mode button.
d) Select Sync-E then select Close
e) Click Setup Interface and then Physical Settings.
Calnex Solutions Ltd Page 6 of 40
Reg. SC299625
f) Verify that the SyncE Clock Rx > Tx is not checked
g) Select the reference clock source.
h) Select the line rate and Electrical/Optical (100M/1GbE/10GbE).
Note: If using SFPs, XFPs or SFP+s, BOTH Port 1 and Port 2 optical transceivers must be
inserted into the Paragon-X.
i) Select the Close button to close the Physical Settings window, and then select Close again to
close the setup interface window.
Note: once you have selected Sync-E mode, the Paragon-X GUI will display SYNCE LOCK as grey
(This verifies that the SyncE Rx > Tx box is unchecked).
ESMC Generation
Ethernet Synchronisation Message Channel (ESMC) is a Point to Point Protocol in which each EEC
generates and terminates an ESMC PDU. This ESMC PDU is normally transmitted once per second
on the Slow Protocol Channel and contains the Quality Level (QL) TLV of the associated clock . The
receiving EEC uses this QL to select the best quality clock. The EEC then regenerates the ESMC
Calnex Solutions Ltd Page 7 of 40
Reg. SC299625
PDU with the appropriate QL TLV value to indicate the quality of the clock being transmitted to the
next EEC.
If the EEC supports ESMC, it is possible for the Paragon-X to use this functionality to indicate if an
EEC switches clock reference or goes into holdover.
The Paragon-X can generate ESMC messages on both Ports 1 and 2. For G.8262 conformance
testing, ESMC messages are generated on Port 2 with a used defined Quality Level (QL) such as
PRC.
a) Generate ESMC messages by selecting button. The following window will open.
b) Check the box Enable TX+Rx Mode
c) Select Port 2 tab (the output from the Paragon and input to EEC)
d) Select SSM code QL-PRC, which tells the EEC that the clock from the Paragon-X is of PRC quality.
e) Set rate to 1000mS.
f) Click and then
g) The ESMC Generation button will go red to indicate the Paragon-X is generating ESMC packets.
After any setup changes to the Paragon-X, please ensure the EEC has had time to settle
before making any measurements
Calnex Solutions Ltd Page 8 of 40
Reg. SC299625
4. Measuring Frequency Accuracy G.8262 Section 6
Input Stimulus Pass/Fail Criteria Notes
EEC Option 1
Free run +/- 4.6 ppm
Recommend to test for
an hour, longer if close
to limits
EEC Option 2
Hold over +/- 4.6 ppm
Recommend to test for
an hour, longer if close
to limits
Measurement Setup a) Connect the EEC to Paragon-X to EEC as shown above
b) Set up the Paragon-X GUI as per Section 3.
Measurement Process
a) Select the Start Capture button to start measurement.
b) To stop the measurement after a pre-defined period for running the test, select Stop Capture
button.
Calnex Solutions Ltd Page 9 of 40
Reg. SC299625
c) Paragon-X will provide the ppm frequency accuracy at the bottom of the graph in red text (as shown
in the screen shot at the top of the next page).
Calnex Solutions Ltd Page 10 of 40
Reg. SC299625
5. Pull-in, Hold-in, and Pull-out ranges G.8262 Section 7
Pullin Range (G.8262 Section 7.1) The Pull-in range is defined as the largest offset between a slave clock's reference frequency and a
specified nominal frequency, within which the slave clock will achieve locked mode.
Input Stimulus Pass/Fail Criteria Notes
EEC
Option 1
and
EEC
Option 2
Apply a large Frequency
offset ensuring EEC is in
holdover. Reduce offset
until EEC locks.
EEC starts unlocked with large offset applied
EEC locks before offset reaches +/- 4.6ppm
Lock can also be
monitored by using
ESMC (if supported)
Holdin Range (G.8262 Section 7.2) Hold-in range is defined as the largest offset between a slave clock's reference frequency and a
specified nominal frequency, within which the slave clock maintains lock as the frequency varies
arbitrarily slowly over the frequency range.
Input Stimulus Pass/Fail Criteria Notes
EEC
Option 1 Not Applicable
EEC
Option 2
EEC is locked to the clock
from the Paragon-X. The
Frequency is then offset
to +/-4.6ppm
EEC should remained locked at an offset at +/-4.6ppm
Lock can also be
monitored by using
ESMC (if supported)
Calnex Solutions Ltd Page 11 of 40
Reg. SC299625
Pullout Range (G.8262 Section 7.3) Pull-out range is defined as the offset between a slave clock's reference frequency and a specified
nominal frequency, within which the slave clock stays in the locked mode and outside of which the
slave clock cannot maintain locked mode, irrespective of the rate of the frequency change.
Input Stimulus Pass/Fail Criteria Notes
EEC
Option 1
EEC is locked to the clock
from the Paragon-X. The
Frequency is then offset
until the EEC unlocks
EEC should remain locked at an offset at +/-4.6ppm but lock should extend beyond this.
G.8262 states this is for
further study
EEC
Option 2 Not Applicable
Measurement Setup a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
b) Set up the Paragon-X GUI as per Section 3, including setting up ESMC with QL=PRC if using ESMC
to monitor which clock the EEC is locked to.
Measurement Process
a) select the Start Capture button to start measurement.
b) Select the Wander and Jitter Generation button. The Frequency Offset window will appear
Calnex Solutions Ltd Page 12 of 40
Reg. SC299625
c) In the Frequency Offset window, add the Frequency Offset required and click the Apply Offset
button
d) To remove the Frequency Offset and return it back to 0ppm click the Remove Offset
button
e) To stop the measurement after a pre-defined period for running the test, select Stop Capture
button.
f) The Paragon-X TIE graph indicates if the EEC is in or out of lock as shown in the screenshot below
Calnex Solutions Ltd Page 13 of 40
Reg. SC299625
6. Wander (Noise) Generation G.8262 Section 8
Input Stimulus Pass/Fail Criteria
Notes
(G.8262 masks)
EEC Option 1
(Constant Temp)
- Locked Mode
- Wander Free reference
- Constant temperature
MTIE & TDEV Pass/Fail masks shown in G.8262 Section 8.1.1
MTIE Table 1/ Figure 1
TDEV Table 3/ Figure 2
EEC Option 1
(Temp effects)
- Locked Mode
- Wander Free reference
- Temperature effects
MTIE Pass/Fail masks shown in G.8262 Section 8.1.2.
MTIE Table 1/ Figure 1
TDEV G.8262 states
for further study
EEC Option 2
(Constant Temp)
- Locked Mode
- Wander Free reference
- Constant temperature
MTIE & TDEV Pass/Fail masks shown in G.8262 Section 8.1.2
MTIE Table 4/ Figure 3
TDEV Table 5/ Figure 4
G.8261 Section 8.2 also mentions measurements in Non-locked mode and refers to section 11.2
Long-term phase transient response (Holdover) and will not be covered in this section.
Measurement Setup a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
g) Set up the Paragon-X as described in section 3, including setting up ESMC with QL=PRC if using
ESMC to monitor which clock the EEC is locked to.
Measurement Process
a) Select the Start Capture button to start measurement.
Calnex Solutions Ltd Page 14 of 40
Reg. SC299625
b) To stop the measurement, select Stop Capture. . It is suggested that the test
should run for approx 3,000 seconds
c) The graph will show the captured TIE
d) To display the MTIE and TDEV graphs click on the bottom right hand side of the
graph and then MTIE/TDEV Analysis from the displayed menu.
e) A separate window will open showing the TIE in graphical form at the top, with an MTIE/TDEV
graph at the bottom.
Calnex Solutions Ltd Page 15 of 40
Reg. SC299625
f) The MTIE/ TDEV analysis can be carried out against the G.8262 masks for Wander Generation
which can be selected from the measurement mask pull down selection in the left hand window.
g) Pass/Fail indication against the masks is shown also in this left hand window and will have a
green background for Mask Pass and a red background for Mask Fail
Calnex Solutions Ltd Page 16 of 40
Reg. SC299625
7. Jitter Generation G.8262 Section 8.3
Testing Jitter generation is planned for a future Paragon-X release
Calnex Solutions Ltd Page 17 of 40
Reg. SC299625
8. Wander (Noise) Tolerance G.8262 Section 9
Input Stimulus Pass/Fail Criteria Notes
EEC Option 1
MTIE Wander
Table 7/Figure 5
TDEV Wander
Table 8/Figure 6
Sinusoidal Wander
Table 9/Figure 7
The EEC is;
i. Maintaining the clock within performance limits.
ii. Not causing any alarms.
iii. Not causing the clock to switch reference.
iv. Not causing the clock to go into holdover.
To Check whether the
EEC is switching
references or going
into holdover, the
Paragon can measure
the wander and/or
ESMC QL of the EEC
output.
EEC Option 2
TDEV Wander
Table 10/Figure 8
The EEC is;.
i. Maintaining the clock within performance limits
ii. Not causing any alarms.
iii. Not causing the clock to switch reference.
iv. Not causing the clock to go into holdover.
To Check whether the
EEC is switching
references or going
into holdover, the
Paragon can measure
the wander and/or
ESMC QL of the EEC
output.
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
b) Set up the Paragon-X as described in section 3, including setting up ESMC with QL=PRC if using
ESMC to monitor if the EEC is switching clock references or going into holdover.
Calnex Solutions Ltd Page 18 of 40
Reg. SC299625
Measurement Process
a) Select the Wander and Jitter Generation button on the Paragon-X GUI.
b) The Following window will now be open.
c) Select the Wander Tolerance Tab and the following window will open
d) There are three methods of generating wander into the EEC:
i. MTIE/TDEV modes Fastest and most effective way to evaluate EEC
ii. Table (Sine) Wander Can be used for finding Maximum Tolerable Wander
iii. Single (Sine) Wander Can be used for troubleshooting
Calnex Solutions Ltd Page 19 of 40
Reg. SC299625
e) MTIE/TDEV Wander
The Paragon-X generates MTIE and TDEV wander as defined in G.8262.
i. Select the wander mask required from the drop down list. The mask and the maximum
running time will be shown on the right hand side of the window.
G.8262 Option 1 MTIE - Running Time is 1000s
G.8262 Option 1 TDEV - Running Time is 12000s
G.8262 Option 2 TDEV - Running Time is 12000s
ii. Click Generate Wander to start the test. The elapsed time will be displayed on the bottom
right hand side of the window.
iii. The test will stop after the max Running time has reached, the test can also be stopped
manually by clicking Stop Wander.
f) Table (Sinusoidal)
Can be used to test Maximum Tolerable Wander
Calnex Solutions Ltd Page 20 of 40
Reg. SC299625
The user can enter up to 10 different wander parameter sets in the table. The Paragon-X will then
automatically add each of the specified wander sets in turn giving full indication of progress in the
Status column. Switching between the different sets is always done at a zero crossing to prevent
phase steps.
i. Enter the frequency, amplitude, and dwell time (number of cycles the frequency/amplitude
pair will be run) for each wander test point.
The same frequency with different amplitudes can be entered to find maximum
tolerable wander
The Restore Defaults button when checked will restore the values to that defined in
table 9 in G.8262.
Only rows that have the enable check box ticked will be executed in the test. To
skip over a selection, un-tick the enable box for that selection.
ii. Click Generate Wander to start the test
iii. Click Stop Wander to stop the test. The test will terminate at the next zero crossing
iv. A pop up box stating how long to the next zero crossing is displayed, click the Stop
v. Immediate button if it is desired to stop the test instantly.
g) Single Sinusoidal Wander
Can be used for troubleshooting issues at a specific frequency
i. Enter the frequency and amplitude of the desired wander
ii. Click Generate Wander to start the test
iii. Click Stop Wander to stop the test. The test will terminate at the next zero crossing of the
wander frequency
A pop up box stating how long to the next zero crossing is displayed.
Calnex Solutions Ltd Page 21 of 40
Reg. SC299625
h) G.8262 states that with the wander applied that the EEC should
i. Maintain the clock within the prescribed performance limits (the exact performance limits are
for further study)
ii. Not cause any alarms
iii. Not cause the clock to switch references
iv. Not cause the clock to go into holdover
For further insight, use the Paragon-X ESMC generation/capture capability and TIE graph to
check whether the EEC is switching clock reference or going into holdover.
i) Ensure the Paragon-X has been set up to Generate ESMC messages with QL=PRC as described
in section 3 of this document.
j) Select the Start Capture button to start the measurement.
k) To show both the TIE and ESMC graph at the same time click and then Show 2nd
Graph and then select ON
l) To change what a graph is displaying,
i. click one of the graphs (the selected graph will be non-Grey
ii. Click and then Graph Display Mode, finally select the graph to display
Time Interval Error (TIE) vs Nominal
ESMC Rx Quality vs Time: (Port 1 Rx)
The screenshot above shows an EEC that has switched references or gone into holdover (top graph
is ESMC, bottom graph shows the difference in Frequency between the reference and the recovered
clock from the EEC)
m) To stop the measurement, select Stop Capture.
Calnex Solutions Ltd Page 22 of 40
Reg. SC299625
9. Jitter Tolerance G.8262 Section 9.2
Testing Jitter tolerance is planned for a future Paragon-X release
Calnex Solutions Ltd Page 23 of 40
Reg. SC299625
10. Wander (Noise) Transfer G.8262 Section 10
Input Stimulus Pass/Fail Criteria Notes
EEC Option 1 Not defined The phase gain of the EEC should be smaller than 0.2 dB (2.3%).
There is no definition of the
input stimulus to be used in
G.8262. Without further
guidance from the Standards,
it is suggested that the
amplitude and frequency
values associated with mask
points labelled f1, f2 & f3 on
G.8262 Section 9.1.1, Table 9
& Figure 7 are used.
EEC Option 2
TDEV Wander
Table 10/Figure 8
Measure EEC output against TDEV Pass/Fail masks shown in G.8262 Section 10..2 Table 13/Figure 11
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
b) Set up the Paragon-X as described in section 3, including setting up ESMC with QL=PRC if using
ESMC to monitor which clock the EEC is locked to.
Calnex Solutions Ltd Page 24 of 40
Reg. SC299625
Measurement Process
a) Select the Wander and Jitter Generation button on the Paragon-X GUI.
b) The Following window will now be open.
c) Select the Wander Transfer Tab and the following window will open
Calnex Solutions Ltd Page 25 of 40
Reg. SC299625
G.8262 Option 1 Wander Transfer Test: Paragon-X automated testing
a) Select the G.8262 Option 1 tab and then the Table tab.
b) Enter the frequency, amplitude, and dwell time (number of cycles the frequency/amplitude pair will be
run) for each wander test point.
i. The Restore Defaults button when checked will restore the values to that defined in table 9 in
G.8262.
ii. Only rows that have the enable check box ticked will be executed in the test. To skip over a
selection, un-tick the enable box for that selection.
c) To calibrate the Paragon-X, connect a short Ethernet cable between Port 1 and Port 2 and click the
button.
Calnex Solutions Ltd Page 26 of 40
Reg. SC299625
d) When the calibration is finished the Status window will show
e) Remove the short Ethernet cable between Port 1 and Port 2 and connect the EEC under test as
shown in the setup diagram at the start of this section.
f) Click the Generate Wander button to start the test.
g) The Paragon-X will show the status of the test, the measured Gain (dB) value and also a Pass/Fail
indication on the right hand side of the window.
Calnex Solutions Ltd Page 27 of 40
Reg. SC299625
G.8262 Option 1 Wander Transfer Test: Paragon-X single frequency test
This capability can be used for fault finding issues at specific frequencies
a) Select the G.8262 Option 1 tab and then the Single tab.
b) Enter the frequency and amplitude of the wander to be generated
c) To calibrate the Paragon-X, connect a short Ethernet cable between Port 1 and Port 2 and click the
button.
d) When the calibration is finished the Status window will show
e) Remove the short Ethernet cable between Port 1 and Port 2 and replace with the EEC under test.
Calnex Solutions Ltd Page 28 of 40
Reg. SC299625
f) Click the Generate Wander button to start the test.
g) The GUI will show the estimated completion time at the bottom of the screen. The measured Gain
(dB) value and also a Pass/Fail indication (Green/Red background) will be displayed.
Calnex Solutions Ltd Page 29 of 40
Reg. SC299625
G.8262 Option 2 Wander Transfer Test:
a) Select the G.8262 Option 2 tab.
b) Press Generate Wander button to start the test.
c) The amount of the time until completion of the test is shown at the bottom right hand side of the
screen.
d) At any time during the test, it is possible to view an updated output TDEV graph by clicking the
TDEV Results button. The TDEV graph will then be displayed.
Calnex Solutions Ltd Page 30 of 40
Reg. SC299625
e) At the end of the test the Paragon-X will automatically show the final MTIE graph and show Pass/Fail
against the MTIE mask.
Calnex Solutions Ltd Page 31 of 40
Reg. SC299625
11. Phase Transient Response G.8262 Section 11
Input Stimulus Pass/Fail Criteria Notes
EEC Option 1
EEC input reference is lost for 15 seconds and a 2nd reference input signal, traceable to the same reference clock, is available simultaneously
Maximum phase transient at the output due to reference switching to meet mask in G.8262 Figure 12
To emulate the loss of the
link either
Change ESMC
QL=DNU
Remove the cable
between port 2 and
EEC
EEC Option 2
EEC input reference
is lost for 15 seconds
and a 2nd reference
input signal, traceable
to the same reference
clock, is available
simultaneously
EEC output should meet MTIE mask defined by table 15/ Figure 14 in section 11.4.2 of G.8262
To emulate the loss of the
link either:
Change ESMC
QL=DNU
Remove the cable
between port 2 and
EEC
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
b) Set up the Paragon-X as described in section 3 (including ESMC generation if using the ESMC
method of switching clock references, ensuring SSM code is set to PRC)
c) Select Configure Capture , select 10mSecs sample period and then close.
Calnex Solutions Ltd Page 32 of 40
Reg. SC299625
Measurement Process
a) There are two methods for determining Phase Transient Response with the Paragon-X;
i. Use ESMC Generation (if supported by the DUT).
ii. Remove the link between Port 2 on the paragon-X and the EEC input port.
b) Select the Start Capture button to start measurement.
c) If using the Ethernet cable removal method.
d) If using ESMC method
i. Select button. The following window will open
ii. Ensure SSM code is set to QL PRC
iii. In the SSM Code drop down menu select QL DNU/DUS and then click .
e) To stop the measurement click Stop Capture. .
Calnex Solutions Ltd Page 33 of 40
Reg. SC299625
Measuring Results
The switching transient will be seen on the Paragon-X TIE graph.
The Paragon-X graphs can show the TIE and also the ESMC QL for each port
In the screenshot bellow the ESMC graph shows the time and QL change of the ESMC transition on
Port 2 Tx of the Paragon-X. This can then be compared in time to the TIE graph showing the
Transient.
Method of evaluating the capture for phase transient response is dependent upon whether EEC Option
1 or Option 2 is being evaluated.
EEC Option 1
a) To view the output TIE for EEC Option 1, closely monitor the output TIE graph on Paragon-X for the
duration (15s) of the test. While the slave clock is acquiring the new reference, the output phase
transient should be within the limits of figure 12 of G.8262 (provided below)
Calnex Solutions Ltd Page 34 of 40
Reg. SC299625
EEC Option 2
a) To view the output MTIE mask for EEC Option 2 ONLY, perform the following:
b) To display the MTIE and TDEV graphs click on the bottom right hand side of the TIE
graph and then MTIE/TDEV Analysis from the displayed menu.
c) A separate window will open showing the TIE in graphical form at the top, with an MTIE/TDEV graph
at the bottom.
d) The MTIE analysis can be carried out against the G.8262 masks for Wander Transient which can
be selected from the measurement mask pull down selection in the left hand window.
Calnex Solutions Ltd Page 35 of 40
Reg. SC299625
e) Pass/Fail indication against the masks is shown also in this left hand window and will have a green
background for Mask Pass and a red background for Mask Fail
Calnex Solutions Ltd Page 36 of 40
Reg. SC299625
12. Appendix 1 - G.8262; Practical interpretation guidance
Frequency Accuracy G.8262 Section 6
Input Stimulus Pass/Fail Criteria Notes
EEC Option 1
Free run +/- 4.6 ppm
Recommend to test
for an hour, longer if
close to limits
EEC Option 2
Hold over +/- 4.6 ppm
Recommend to test
for an hour, longer if
close to limits
Pullin Range (G.8262 Section 7.1)
Input Stimulus Pass/Fail Criteria Notes
EEC
Option 1
and
EEC
Option 2
Apply a large Frequency
offset ensuring EEC is in
holdover. Reduce offset
until EEC locks.
EEC starts unlocked with large offset applied
EEC locks before offset reaches +/- 4.6ppm
Lock can also be
monitored by using
ESMC (if supported)
Holdin Range (G.8262 Section 7.2)
Input Stimulus Pass/Fail Criteria Notes
EEC
Option 1 Not Applicable
EEC
Option 2
EEC is locked to the clock
from the Paragon-X. The
Frequency is then offset
to +/-4.6ppm
EEC should remained locked at an offset at +/-4.6ppm
Lock can also be
monitored by using
ESMC (if supported)
Calnex Solutions Ltd Page 37 of 40
Reg. SC299625
Pull out Range (G.8262 Section 7.3)
Input Stimulus Pass/Fail Criteria Notes
EEC
Option 1
EEC is locked to the clock
from the Paragon-X. The
Frequency is then offset
until the EEC unlocks.
EEC should remain locked at an offset at +/-4.6ppm but lock should extend beyond this.
G.8262 states this is for
further study
EEC
Option 2 Not Applicable
Wander Generation (G.8262 Section 8)
Input Stimulus Pass/Fail Criteria Notes
EEC Option 1
(Constant Temp)
- Locked Mode
- Wander Free reference
- Constant temperature
MTIE & TDEV Pass/Fail masks shown in G.8262 Section 8.1.1
MTIE Table 1/ Figure 1
TDEV Table 3/ Figure 2
EEC Option 1
(Temp effects)
- Locked Mode
- Wander Free reference
- Temperature effects
MTIE Pass/Fail masks shown in G.8262 Section 8.1.2.
MTIE Table 1/ Figure 1
TDEV G.8262 states
for further study
EEC Option 2
(Constant Temp)
- Locked Mode
- Wander Free reference
- Constant temperature
MTIE & TDEV Pass/Fail masks shown in G.8262 Section 8.1.2
MTIE Table 4/ Figure 3
TDEV Table 5/ Figure 4
Calnex Solutions Ltd Page 38 of 40
Reg. SC299625
Wander Tolerance (G.8262 Section 9)
Input Stimulus Pass/Fail Criteria Notes
EEC Option 1
MTIE Wander
Table 7/Figure 5
TDEV Wander
Table 8/Figure 6
Sinusoidal Wander
Table 9/Figure 7
The EEC is;
v. Maintaining the clock within performance limits.
vi. Not causing any alarms.
vii. Not causing the clock to switch reference.
viii. Not causing the clock to go into holdover.
To Check whether the
EEC is switching
references or going
into holdover, the
Paragon can measure
the wander and/or
ESMC QL of the EEC
output.
EEC Option 2
TDEV Wander
Table 10/Figure 8
The EEC is;.
v. Maintaining the clock within performance limits
vi. Not causing any alarms.
vii. Not causing the clock to switch reference.
viii. Not causing the clock to go into holdover.
To Check whether the
EEC is switching
references or going
into holdover, the
Paragon can measure
the wander and/or
ESMC QL of the EEC
output.
Wander Transfer (G.8262 Section 10)
Input Stimulus Pass/Fail Criteria Notes
EEC Option 1 Not defined The phase gain of the EEC should be smaller than 0.2 dB (2.3%).
There is no definition of the
input stimulus to be used in
G.8262. Without further
guidance from the Standards,
it is suggested that the
amplitude and frequency
values associated with mask
points labelled f1, f2 & f3 on
G.8262 Section 9.1.1, Table 9
& Figure 7 are used.
EEC Option 2
TDEV Wander
Table 10/Figure 8
Measure EEC output against TDEV Pass/Fail masks shown in G.8262 Section 10..2 Table 13/Figure 11
Calnex Solutions Ltd Page 39 of 40
Reg. SC299625
Transient Response (G.8262 Section 11)
Input Stimulus Pass/Fail Criteria Notes
EEC Option 1
EEC input reference is lost for 15 seconds and a 2nd reference input signal, traceable to the same reference clock, is available simultaneously
Maximum phase transient at the output due to reference switching to meet mask in G.8262 Figure 12
To emulate the loss of the
link either
Change ESMC
QL=DNU
Remove the cable
between port 2 and
EEC
EEC Option 2
EEC input reference
is lost for 15 seconds
and a 2nd reference
input signal, traceable
to the same reference
clock, is available
simultaneously
EEC output should meet MTIE mask defined by table 15/ Figure 14 in section 11.4.2 of G.8262
To emulate the loss of the
link either:
Change ESMC
QL=DNU
Remove the cable
between port 2 and
EEC
For more information on the Calnex Paragon-X and to take advantage of Calnexs extensive experience in