Gas and Liquid Sampling Device

Embed Size (px)

Citation preview

  • 8/8/2019 Gas and Liquid Sampling Device

    1/13

    AC Liquefied Gas & Gas SamplingDevice

  • 8/8/2019 Gas and Liquid Sampling Device

    2/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 2 of 13Version 2004/4.0

    Introduction

    The AC Liquefied and Gas Sampling Device is a sampling module that allows the user to

    reproducibly introduce gaseous and liquefied (LPG) samples into a gas analyzer. A built-

    in heated vaporizer vaporizes the liquefied samples, so they are introduced into the gas

    chromatograph in the gaseous state. The unit is placed adjacent to the gas

    chromatograph and connected to the gas sample loop(s) of the gas analyzer. It is

    configured with a heated vaporizer, two rotameters, a selection valve and a clamp to

    hold the sample bombs.

    Partnumber Description

    68100.000 Gas and Liquid Sampling Device excluding vaporizer

    68100.001 Gas and Liquid Sampling Device including 110 V vaporizer

    Options :

    - low temperature (24-104C) *

    - high temperature (104-194C) *

    68100.002 Gas and Liquid Sampling Device including 220 V vaporizer

    Options :

    - low temperature (24-104C) *- high temperature (104-194C) *

    * NOTE: Temperature adjustable inside the vaporizer, depending on the desired

    temperature.

  • 8/8/2019 Gas and Liquid Sampling Device

    3/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 3 of 13

    Configuration

    The sampling device is a separate unit configured with a heated vaporizer, a rotameter,

    a clamp to hold the sample bombs, two selection valves and the necessary plumbing

    material. The schematic overview can be found in figure 1.

    Heated Vaporizer

    to Gas SamplingValve GC

    L

    G

    out

    in

    Gaseous Sample in

    Liquefied Sample in

    from Gas SamplingValve GC

    Selection Valve

    rotameter

    Vent

    Needle valve

    Outlet pressureadjustment

    2

    10

    4

    3

    7

    1

    5 6

    8

    9

    Selection Valve13

    13

    12

    A

    B

    Figure 1 Schematics

    Principle

    Both Liquefied and gaseous samples can be mounted on the device. With the Selection

    Valve A, the proper position is chosen.

    So for gaseous samples, position 3 (of selection valve A) is selected. The sample from

    the sample bomb directly flows via Selection Valve A through the gas sample valve of

    the gas chromatograph. It then goes through the rotameter to Vent. The effluent flow canbe adjusted with the needle valve (8) underneath the rotameter.

    For Liquefied samples the Selection Valve A must be put in the Liquid position (4) and

    selection valve B in position (12). So the liquefied sample is introduced in the Heated

    Vaporizer. Here it is vaporized. The effluent gas goes to the gas sampling valve of the

    gas chromatograph and finally again through the needle valve to vent. The effluent flow

    should be adjusted with the outlet pressure adjustment (10), a special tool to adjust the

    outlet pressure is delivered with the system.

    Version 2004/4.0

  • 8/8/2019 Gas and Liquid Sampling Device

    4/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 4 of 13

    Configuration

    Figure 2 Over view

    1. Gas Sample In (male Swagelok 1/8)

    2. Rotameter

    3. Selection Valve A Position: Gas (3)

    4. Selection Valve A Position: Liquid (4)

    5. To Gas Sampling Valve Gas Chromatograph (male Swagelok 1/8)

    6. From Gas Sampling Valve Gas Chromatograph (male Swagelok 1/8)7. Liquefied Sample in (male Swagelok 1/8)

    8. Needle Valve Rotameter

    9. Vent (male Swagelok 1/8)

    10. Outlet pressure adjustment Heated vaporizer (5/32 Allen wrench)

    11. Power supply

    12. Selection Valve B Position: to GC (12)

    13. Selection Valve B Position: to vent (13)

    Version 2004/4.0

  • 8/8/2019 Gas and Liquid Sampling Device

    5/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 5 of 13

    Power supply Configuration

    Figure 3 Front and Rear view

    Version 2004/4.0

  • 8/8/2019 Gas and Liquid Sampling Device

    6/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 6 of 13

    Installation

    The Sampling device must be placed adjacent to the gas chromatograph on the

    lefthandside. The Gas sampling valve of the chromatograph must then be connected to

    the Sampling device.

    Using 1/16 Stainless steel capillary tubing, position 5 is connected to the Gas Sample

    in from the gas chromatograph. Position 6 is connected to the Gas sample out from the

    gas chromatograph.

    It is recommended to connect a length of 1/8 tubing from the Vent position (9) on the

    rear side of the device to a fume hood.

    The power plug can be plugged in, this is the power supply of the heated vaporizer.

    Please refer to figure 4 for the location of the temperature adjusting screw. This screw is

    used to set the temperature of the heated vaporizer depending on the boiling point range

    of the sample.

    Figure 4 Wiring diagram

    Version 2004/4.0

  • 8/8/2019 Gas and Liquid Sampling Device

    7/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 7 of 13

    Gaseous Samples

    To inject gaseous samples, the gas sample bomb must be vertically mounted in the

    clamp. The top outlet valve must be connected to the gas sample in (1) position on the

    device and the selection valve A must be switched to the Gas position. (3)

    Now with the needle valve of the rotameter (8) in the closed position (turned entirely

    clockwise) the outlet valve of the sample bomb is opened. Then slowly the rotameter

    needle valve is opened until a constant flow is obtained. (2 on scale corresponds with

    about 30 mLs/minute)

    The sample loop in the gas chromatograph is now being flushed. After one minute the

    analysis can be started.

    To obtain proper results, it is important to introduce and analyse samples and calibration

    blends in the same way. Therefore if analysis is done whilst the sample loop is being

    flushed, the flow through the gas sample loop must always be identical for sample and

    calibration analyses.

    If atmospheric injection is preferred, care must be taken to allow the same wait time after

    the flow through the sample has been stopped for both sample and calibration runs.

    (The stopwatch function of the GC display can be helpful in this)

    Heated Vaporizer

    L

    out

    inLiquefied Sample in

    from Gas SamplingValve GC

    Selection Valve

    rotameter

    Vent

    Needle valve

    Outlet pressureadjustment

    2

    10

    4

    7

    6

    8

    9

    Selection Valve13

    13

    12

    A

    B

    to Gas SamplingValve GC

    G

    Gaseous Sample in

    3

    1

    5

    Figure 5 Flow path gaseous samples

    Version 2004/4.0

  • 8/8/2019 Gas and Liquid Sampling Device

    8/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 8 of 13Version 2004/4.0

    Liquefied Samples

    To inject liquefied samples, the sample bomb must be mounted in the clamp, and the

    bottom valve must be connected to the Liquefied Sample in position of the device*.

    The selection valve A must be switched to the Liquid position. (4), selection valve B

    must be in position (12).

    The outlet pressure adjustment (10) can be adjusted with the special tool supplied with

    the system.

    Now with the outlet pressure adjustment (10) in the closed position and the needle valveof the rotameter (8) opened, the outlet valve of the sample bomb is opened. Then slowly

    the outlet pressure adjustment (10) is opened until a steady flow (about 30 ml/minute) is

    obtained. (2 on scale corresponds with about 30 ml /minute). The sample loop in the gas

    chromatograph is now being flushed. Several minutes of time must be allowed to pass in

    order to properly flush the vaporizer and to obtain a homogeneous sample.

    Prior to analysis, selction valve B must be switched to position (13) in order to vent

    excess sample. The pressure in the sample loop now becomes atmospheric.

    To obtain proper results, it is important to introduce and analyse samples and calibrationblends in the same way. Therefore if analysis is done whilst the sample loop is being

    flushed, the flow through the gas sample loop must always be identical for sample and

    calibration analyses.

    Please refer to figure 6.

  • 8/8/2019 Gas and Liquid Sampling Device

    9/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 9 of 13

    Heated Vaporizer

    to Gas SamplingValve GC

    L

    G

    out

    in

    Gaseous Sample in

    Liquefied Sample in

    from Gas SamplingValve GC

    Selection Valve

    rotameter

    Vent

    Needle valve

    Outlet pressureadjustment

    2

    10

    4

    3

    7

    1

    5 6

    8

    9

    Selection Valve1312

    A

    B

    Figure 6 Flow path liquefied samples

    NOTE: A too high sample flow can be the cause of a not totally evaporated sample in

    the sample loop.

    NOTE: If the sample bomb is equipped with a diptube then the liquid out of the diptube

    must be connected to the Liquefied Sample in position of the device!

    NOTE: To vent excess sample, selection valve B must be switched to position (13).

    Version 2004/4.0

  • 8/8/2019 Gas and Liquid Sampling Device

    10/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 10 of 13

    Heated Vaporizer

    The heated vaporizer is built in to vaporize the liquefied samples. It consists of a heat

    exchanger and a pressure control section. The liquefied sample enters the heated body,

    where it is vaporized. The outlet pressure is controlled by a regulator (10) which can be

    accessed at the back of the sampling device.

    Figure 7 Heated vaporizer

    Version 2004/4.0

  • 8/8/2019 Gas and Liquid Sampling Device

    11/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 11 of 13Version 2004/4.0

  • 8/8/2019 Gas and Liquid Sampling Device

    12/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 12 of 13Version 2004/4.0

    Part numbers:

    28024.210 Heated Vaporizer (220V) 24 104 C

    28024.220 Heated Vaporizer (220V) 104 194 C

    28024.211 Heated Vaporizer (110V) 24 104 C

    28124.221 Heated Vaporizer (110V) 104 194C

    Standard Specifications

    Body: Stainless steel

    Model:/Port configuration: Side entry

    Seat: Peek

    Temp range: 75 F to 220 F (24 C to 104 C)

    215 F to 380 F (104 C to 194 C)

    Outlet range: 0-10 Psi

    Power: 100 watt

    Max pressure inlet: 2000 Psi

    Typical applications are

    Vaporize liquid hydrocarbons Ethylene, LPG, Butane. Natural gas sampling

    Heated gas samples to prevent condensable drop out

    Installation and Usage

    Installation Vaporizer involves plumbing lines and electrical connection, use only

    with AC applications.

    Use instructions regarding connections as described below:

    Allways check the connection for leaks.

  • 8/8/2019 Gas and Liquid Sampling Device

    13/13

    AC Liquefied Gas & Gas Sampling Device

    Hardware section

    Page 13 of 13Version 2004/4.0

    General mounting instructions

    Disconnect power before servicing or replacement.

    Plug in the power cord to the specified voltage.

    Adjusting the proper temperature

    Disconnect power.

    Remove left cover from unit.

    Open the vaporizer cylindrical cover plate to get access to the power cable.

    Adjust the proper temperature setting. (position 1-6)