103
GC and GCMS Fundamentals of Instrument Troubleshooting and Maintenance August 1, 2016 Kirk Lokits Ph.D. Applications Chemist

GC and GCMS Fundamentals of Instrument Troubleshooting and Maintenance · 2018. 4. 4. · GC and GCMS Fundamentals of Instrument Troubleshooting and Maintenance August 1, 2016 Kirk

  • Upload
    others

  • View
    21

  • Download
    2

Embed Size (px)

Citation preview

  • GC and GCMS Fundamentals of Instrument

    Troubleshooting and Maintenance

    August 1, 2016

    Kirk Lokits Ph.D.

    Applications Chemist

  • 1971 5930A with Oscilloscope and Strip Chart

    Timeline:

    • Intel introduces first microprocessor, 4004

    • IBM introduces 8” floppy disk

    Page 2

  • 2016 5977B/7890B with Sampler Tower and Tray

    Page 3

  • Page 4

    “Everything was just fine

    and then this happened!”

    How do I go about

    TROUBLESHOOTING?”

  • Page 5

    “Everything was just fine and then

    this happened!”

    Logic = Something changed (slowly or sudden)

    = Something is different

    Track Events – log book

    -Changed column, liner, septum, syringe, etc.

    -Injected samples, other method, etc.

    -Did maintenance, cut column, inlet flush, etc.

  • Page 6

    Troubleshooting Starts with Isolating the problem

    There are 5 basic areas from where the problem arises

    FLOW

    INJECTOR

    COLUMN

    DETECTOR

    ELECTRONICS

    But of course it can always be some COMBINATION

    Knowing what can & can’t cause the symptom is the key

    Logical Troubleshooting

  • Page 7

    Typical Problems of Optimized Methods becoming

    Unoptimized…and the Reason Why.

    • Column Installation – Correct Process

    • Peak Tailing – Flow Path or Activity

    • Bonus Peaks – In Sample or Back Flash (Carry Over)

    • Split Peaks – Injector Problems, Mixed Solvent

    • No Peaks – Wasn’t Introduced, Wasn’t Detected

    • Response Changes – Activity, Injector Discrimination, Detector Problem

    • Peak Fronting – Overload or Solubility Mismatch, Injector Problems

    • Shifting Retention – Leaks, Column Aging, Contamination or Damage

    • Loss of Resolution – Separation Decreasing, Peak Broadening

    • Baseline Disturbances – Column Bleed, Contamination, Electronics

    • Noisy or Spiking Baseline – Electronics or Contaminated Detector

    • Quantitation Problems – Activity, Injector or Detector Problems

  • Page 8

    Column Installation Procedure

    • Install the column

    • Leak and installation check

    • Column conditioning

    • Bleed profile

    • Test mix

  • Page 9

    Column Installation

    What type of ferrule should I use?

    •Graphite

    •Graphite/Vespel

  • Self Tightening Column Nuts

    Page 10

    For mass spec transfer line

    p/n 5190-5233

    For inlet or detector

    p/n 5190-6194

  • What to Use When?

    Application Requirement

    Self Tightening

    Column nut with

    short

    Graphite/Polyimide

    blend ferrule

    UltiMetal Plus

    Flexible Metal

    ferrule

    Ease of Use - finger tighten / no wrench x

    Inertness for active analytes x

    Installed or used hardware x

    Same style ferrule at inlet and MS interface x

    High temp above 350 x

    Mass Spec Interface fitting x

    Capillary Flow Technology device x

    Pre-swage for precise height into fitting x x

    Re-use of ferrule x

    Page 11

  • Solving Shrinking Ferrule Syndrome

    Page 12

  • Page 13

    Column InstallationMeasuring the right distance

    White out Septa

  • Page 14

    Cutting The Column

    Gently scribe through the polyimide coating.Do not attempt to cut the glass.

    Recommended tools:Diamond or carbide tipped pencil; or sapphire cleaving tool, ceramic wafer Ocular

    Do not use:Scissors, file, etc.

  • Page 15

    Example of a Bad Cut

  • Page 16

    Examples of Column Cuts

    Good

    Bad

  • Page 17

    Column Installation

    How tight is

    tight?

  • Page 18

    Overtightened Ferrule

  • Page 19

    Column InstallationLeak Check

    DO NOT USE SNOOP

    Electronic leak detector

    IPA/Water (50:50)

    Inject a non-retained peak

  • Page 20

    Leak and Installation CheckInject a non-retained compound vs DB-1

    The peak should be sharp and symmetrical

    Detector Compound

    FID Methane or Butane

    ECD MeCl2 (headspace or diluted)

    NPD CH3CN-acetonitrile (headspace or diluted)

    TCD Air

    MS Air or Butane

  • Page 21

    Non-Retained Peak Shapes

    Good Installation Improper Installation orInjector Leak

    Check for: -Too low of a split ratio

    -Injector or septum leak

    -Liner problem:

    (broken, leaking, misplaced)

    -Column position in injector and detector

  • Page 22

    Column Conditioning

    System must be leak free before conditioning column

    Heat the column to the lower of:

    Isothermal maximum temperature OR

    20° to 30°C above highest operation temperature

    Temperature programming is not necessary

    Stop conditioning when the stable baseline is obtained:

    1 to 2 hours in most cases

  • Page 23

    Generating a Bleed Profile

    Temperature program the column without an

    injection*

    *DB-1 30m x .32mm I.D., .25µm

    Temperature program // 40°C, hold 1 min // 20°/min to 320°C, hold 10 min.

    Time (min.)0 5 10 15 20 25

    6000

    7000

    8000

    9000

    1.0e4

    1.1e4

    1.2e4

    1.3e4

  • Page 24

    Peak Tailing

    INJECTOR or COLUMN is Active

    -Reversible adsorption of active compounds

    (-OH, -NH, -SH)

    FLOW problem - dead volume, obstruction, poor installation, or severe column contamination

    Miscellaneous - co-elution, polarity mismatch between phase, solute or solvent, and some compounds always tail

    *Tip = Inject a light hydrocarbon, should not tail unless flow path problem.

  • Page 25

    Bonus Peaks or Ghost Peaks

    Contamination in INJECTOR, COLUMN or FLOW (carrier gas)

    -Carry-over from a back flash or previous sample

    -Bad tank of gas or traps have expired

    -Septum bleed

    *TIP = Run a blank run…it should be blank!

  • Ghost Peaks

    Back flash

    Back flash occurs

    when vapor

    condenses on a

    cooler part of the

    injector. This

    condensed sample

    will still have a

    “headspace” which

    can be swept forward

    to the column and

    cause ghost peaks.

    Page 26

  • Ghost Peaks

    Steps to Prevent Backflash• Reduce injection volume

    • Lower inlet temperature

    • Consider a large volume liner

    • Pressure pulse the injection (if possible)

    • Change solvents

    Page 27

  • Page 28

    Split Peaks

    INJECTOR (poor sample introduction)

    -Injecting the sample twice (some how?)

    -Mixed sample solvent (polarity difference)

    -Sample in syringe needle (manual inject)

    INJECTOR (activity)

    -Breakdown (not really a split peak, 2 peaks)

    -Sample degradation in injector

    VOLATILITY

    High boilers dropping out on Cold Spots

    -Transfer line temps

    -Unions or fittings not tracking column temp

  • Page 29

    Logical Troubleshooting Tips

    Look and Ask…Which peaks? Early or late? Trends?

    = Injection technique - erratic

  • Page 30

    No Peaks

    DETECTOR (not on or not operational)

    INJECTOR (not working)

    -Plugged syringe/plunger not moving

    -Wrong injector (or detector)

    -Huge leak (older systems)

    -No carrier gas flow

    NOT the COLUMN Unless…

    -Broken column or No column

  • Page 31

    A Real Troubleshooting Example

    No Peaks

    10 20 30 40 50

    14

    16

    18

    20

    22

    24

    26

    28

    30

  • Page 32

    Is the flame Lit?

    put glass piece over FID outlet----Answer in this case, Water condensation

    look at output in instrument gauge-- is the digital value greater than 0.0?

    Answer in this case is approximately 16.2 pico amps

    Is there flow through the column?

    disconnect column from detector and measure flow with a meter

    Answer in this case was YES THERE IS FLOW

    Assess the observations

    Flame is lit and we have flow from end of column

    Hypothesis: Sample not getting on column-syringe plugged?

    Take syringe out and make injection manually on a dry paper towel

    Answer – towel stays dry (Syringe was clogged with septum)

    Pull plunger out top, add solvent and replace plunger will usually dislodge septum particle

    (should hear a little pop) If you can’t dislodge plug, Replace syringe

    Reassemble the Injector & Re-inject

    Logical Steps Taken to Find Peaks(most of our problems are leaks and plugs)

  • Page 33

    Peak ResponseAll Change in Size

    DETECTOR (response problem)

    -Settings or flows changed

    -Electronics failing

    -Split ratio set incorrectly

    -Wrong purge activation time

    -Septum purge flow too high

    -Injector temperature too low*

    INJECTOR

    -Leaky syringe

    *Tip = Is it all of them or some of them, if all then injector or detector

  • Page 34

    Peak ResponseSome Change in Size

    INJECTOR or COLUMN is active/contaminated

    -Irreversible adsorption of active compounds (-OH, -NH, -SH)

    -Decomposition of sample

    -Temperature Change – Discrimination

    -Evaporation from sample

    *Tip = If only some change, then which ones? If active compounds then

    activity. If tracks volatility then cold spots or inlet discrimination.

  • Page 35

    Peak FrontingShark Fin Shaped or Just Slight

    COLUMN (contaminated)

    -Overload (More pronounced with large solute and phase polarity differences)

    INJECTOR

    -Column installation

    -Compound very soluble in injection solvent (need retention gap)

    -Mixed sample solvent

    OTHER

    -Co-elution

    -Breakdown

  • Page 36

    Retention Time Shift

    2.00

    3.25

    4.75

    2.75

    4.00

    5.50

    INJECTOR

    -Leak in the septum

    -Change in injection solvent

    -Large change in sample concentration

    FLOW

    -Change in gas velocity

    COLUMN

    -Contamination

    -Damaged stationary phase

    -Loss of stationary phase

    -Change in temperature

    1.75

    3.00

    4.50

  • Page 37

    min8 9 10 11

    1400 ng

    7 ng

    Effect of Sample Overload on

    Retention Time and Peak Shape

  • Page 38

    Loss of Resolution

    Resolution is a function of separation and peak width

    Separation

    Peak Width

  • Page 39

    Loss of Resolution - Separation Decrease

    COLUMN

    -Different column temperature

    -Contamination (more phase?)

    -Matrix components co-eluting

    -Different column phase?

    Separation

    Peak Width

  • Page 40

    Loss of Resolution - Peak Broadening

    FLOW

    -Change in carrier gas velocity

    -Make-up gas

    COLUMN

    -Contamination

    -Phase degradation

    INJECTOR (efficiency)

    -Settings, Liner, Installation, etc.

    Peak Width

    Separation

  • Page 41

    Baseline Disturbances

    Sudden Changes, Wandering, or Drifting

    COLUMN or DETECTOR

    -Not fully conditioned or stabilized (electronics)

    -Contamination

    FLOW

    -Changes in carrier and/or detector gas flows

    -Valves switching, leaks

    WANDER

    DRIFT

  • Page 42

    Noisy Baseline

    FLOW

    -Contaminated gas

    -Incorrect detector settings

    COLUMN

    -Bleed if at high temperature

    -In detector flame (poor installation)

    MILD

    SEVERE

    DETECTOR

    -Air leak - ECD, TCD

    -Electronics malfunction

  • Page 43

    Spiking Baseline

    DETECTOR

    -Particles entering the detector

    -Random: poor connection

    -Regular: nearby "cycling" equipment (electronics)

  • Page 44

    Quantitation Problems

    DETECTOR

    -Poor stability (electronics) or Baseline disturbances (contamination)

    -Outside detector's linear range or wrong settings

    Activity (adsorption) in INJECTOR or COLUMN

    INJECTOR

    -Technique, settings, conditions

    -Syringe worn

    OTHER

    -Co-elution

    -Matrix effects

    -Sample evaporation – leaky vials

    -Sample decomposition

  • Page 45

    TROUBLESHOOTING “TOOLS”

    Bleed Profile: baseline problems

    Inject a non-retained peak: peak shape problems

    Test mix: all problems

    Isolate the components: all problems

    Jumper Tube Test: baseline problems

    Static Pressure Test: check for leaks

  • GENERATING A BLEED PROFILE

    Time (min.)0 5 10 15 20 256000

    7000

    8000

    9000

    1.0e4

    1.1e4

    1.2e4

    1.3e4

    *DB-1 30m x .32mm I.D., .25µm

    Temperature program // 40°C, hold 1 min //

    20°/min to 320°C, hold 10 min.

    Use when the column is new (for future

    reference) or there is a baseline problem

    Page 46

  • Good Installation Improper Installation orInjector Leak

    Potential problems:

    Injector or septum leak

    Too low of a split ratio

    Liner problem(broken, leaking, misplaced)

    Column position in injector and detector

    NON-RETAINED PEAK SHAPES

    Used to Check

    Flowpath

    Page 47

  • Page 48

    Test Mixes

    Used to determine how "good" the column is

    #1

  • Page 49

    Column Performance Summary

    THEORETICAL PLATES/METER: MIN SPEC ACTUAL

    COATING EFFICIENCY:

    RETENTION INDEX:

    PEAK HEIGHT RATIO:

    MIN SPEC MAX SPEC ACTUAL

    PENTADECANE

    PENTADECANE

    1-UNDECANOL

    ACENAPHTHYLENE

    4-CHLOROPHENOL/METHYL NONANOATE

    4-PROPYLANILINE/METHYL NONANOATE

    0.83

    1.14

    1371.04 1372.04 1371.43

    1459.34 1460.34 1459.53

    3900 4389

    90.0 95.5

    COMPOUNDIDENTIFICATION

    RETENTION TIME(T )R

    PARTITION RATIO

    (k)

    PEAK WIDTH(W 1/2)

    1,6-HEXANEDIOL

    4-CHLOROPHENOL

    METHYL NONANOATE

    4-PROPYLANILINE

    TRIDECANE

    1-UNDECANOL

    ACENAPHTHYLENE

    PENTADECANE

    Approximately 5-10 ng on column

    (t )o

    2.51

    2.95

    3.21

    3.81

    4.20

    5.52

    8.00

    9.58

    0.9

    1.3

    1.5

    1.9

    2.2

    3.3

    5.2

    6.4

    0.019

    0.022

    0.022

    0.026

    0.027

    0.036

    0.053

    0.062

    1.29

    PART NO:

    COLUMN I.D. NO.:

    LIQUID PHASE:

    FILM THICKNESS:

    COLUMN DIMENSIONS:

    m X mm

    TEMPERATURE LIMITS:

    C TO C

    1225032

    3303121

    DB-5

    0.25 µm

    30 0.252

    -60° 325° ( C PROGRAM)350°

  • TEST MIXUsed to determine how “good” the column is or if the problem is related to

    the chemical properties of the analytes.

    TEST TEMPERATURE: C135°

    CARRIER GAS: ( )1.2(H )2 38.7

    INJECTION: SPLIT ANALYST: ERIK

    cmsec min

    mL

    RETENTION TIME (MIN)

    Max.Bleed10.0 pA

    325

    9.0 pA

    135

    0 5

    Page 50

  • Page 51

    Test Mixture Components

    Compounds

    Hydrocarbons

    FAME’s, PAH’s

    Alcohols

    Acids

    Bases

    Purpose

    Efficiency

    Retention

    Retention

    Activity

    Acidic Character

    Basic Character

  • Page 52

    Own Test Mixture

    • More specific to your application

    • Selective detectors

    • Concentrations specific to your application

    • Use same instrument conditions

    • Easiest to simply inject a calibration standard

    • Store for future measure of column performance

  • ISOLATE THE COMPONENTS

    Put in a known good column

    Move column to a different GC, inlet or detector

    Simplify the system:

    • example - Direct injection instead of P&T sample introduction

    Page 53

  • JUMPER TUBE TESTPurpose

    Helps to locate the source of contamination or noise

    Isolates GC components

    Page 54

  • JUMPER TUBE TEST Isolate the Detector

    Remove column from the detector

    Cap detector and turn on

    Blank run

    Page 55

  • JUMPER TUBE TEST Isolation of Detector - Results

    Detector is the problem

    Detector OK

    Page 56

  • JUMPER TUBE TEST Isolate the Injector

    Connect the injector and detector

    1-2 meters deactivated fused silica tubing

    Turn on carrier gas

    Blank run

    Page 57

  • JUMPER TUBE TEST Isolate the Injector - Results

    Injector OK

    Injector, lines or carriergas contaminated

    Page 58

  • JUMPER TUBE TEST Isolate the Column

    Reinstall the column

    Setup as before

    Blank Run

    Page 59

  • JUMPER TUBE TEST Isolate the Column - Results

    Problem returns: It’s the column

    Problem gone: Previous leak, solid debris, or installation

    problem

    Page 60

  • STATIC PRESSURE CHECK

    Checks for Leaks in Injector or Gas Lines

    Seal all injector outlets

    Supply pressure from gas cylinder

    up to 20psi higher than analysis

    Turn off cylinder

  • Page 62

    Common Causes of Column Performance

    Degradation and or Short Column Life

    • Physical damage to the polyimide coating

    • Thermal damage

    • Oxidation (O2 damage)

    • Chemical damage by samples

    • Contamination

  • Page 63

    Physical Damage to The Polyimide Coating

    • The smaller the tubing diameter, the more flexible it is.

    • Avoid scratches and abrasions

    • Immediate breakage does not always occur upon physical damage

  • Page 64

    NOT what you want your column to look like!

  • Page 65

    Thermal Damage

    Degradation of the stationary phase is increased at higher temperatures. Breakage along the polymer backbone.

    Dimethylpolysiloxane

    CH3 CH3CH3

    CH3 CH3CH3

    O

    Si SiSi

    OOO

  • Page 66

    Thermal DamageWhat To Do If It Happens

    • Disconnect column from detector

    • “Bake out” overnight at isothermal limit

    • Remove 10-15 cm from column end

  • Page 67

    Thermal Damage

    • Rapid degradation of the stationary phase

    caused by excessively high temperatures

    Isothermal limit = Indefinite time

    Programmed limit = 5-10 minutes

    • Temporary "column failure" below lower

    temperature limit

  • Page 68

    Oxidation (O2 Damage)

    Oxygen in the carrier gas rapidly degrades the stationary phase. The damage is accelerated at higher temperatures. Damage along the polymer backbone is irreversible.

    Dimethylpolysiloxan

    e

    CH3 CH3CH3

    CH3 CH3CH3

    O

    Si SiSi

    OOO

    O2

  • Group/Presentation Title

    Agilent Restricted

    Page 69

    Oxygen DamageWhat To Do If It Happens

    •Rapid damage to the column

    •Usually results in irreversible column damage

  • Page 70

    How to Prevent Column Damage by Oxygen

    • High quality carrier gas (4 nine's or greater)

    • Leak free injector and carrier lines

    Change septa

    Maintain gas regulator fittings

    • Appropriate impurity traps

  • Page 71

    OVEN TEMP 100

    Configurations for Carrier Gas Purifiers

    IndicatingMoisture Trap

    High CapacityOxygen Trap

    Indicating Oxygen Trap

    GC

  • Page 72

    Common Causes of Leaks

    Re-use and mis-installation.

    • Leak from O-ring, Gold Seal, ferrules, column nuts

    • O-rings are elastomer compression fittings designed for one use, not

    perfectly elastic.

    • Gold seals are designed for one use, knife edge cuts into gold layer

    giving leak tight seal w/o shrinkage or potential organic contaminants

    from polyimide out-gassing/degradation.

    • Re-using could result in overlap in seal rings, resulting in a leak.

    • Over-tightening of fittings

  • Page 73

    Leaks Due to Septum Nut

    • With repeated use, conical needle guide gets

    worn, out of round, and needs replacement as

    septum can begin to “bulge” out, especially with

    excessive tightening,

    • Septa fail faster because needle is not guided with as

    much precision.

    • Under or Over tightening—tighten nut until c-clamp on

    top stops turning, then ½ to ¾ turn more.

    • Non-Agilent septa may be too thin, too thick, or out of

    round like die-cut septa and may not seal as well.

    • “Use Environments” that decrease lifetime, like using

    non-Agilent Autosamplers (ours are precisely aligned),

    manual injection, larger gauge syringes

    • Replace septum nut annually for peace of mind.

  • Page 74

    Avoid High Temperatures

    Hotter temperatures degrade the phase faster

    Maintain column at or below isothermal limit

    Minimize time at programmable temperature limit

    Lower temperature limit is performance related and

    not related to phase damage

  • Page 75

    Avoid Chemical Damage

    There are more things that don’t go through the

    column than do.

    There are nasty things in samples.

    Inorganic acids and bases will react with and

    damage most stationary phases.

    • i.e. HCl, H2SO4, KOH, NaOH, etc.

    CONCLUSION: Don’t inject anything on the column!

    - It will last much longer!

  • Page 76

    Chemical DamageWhat To Do If It Happens

    • Remove 1/2 - 1 meter from the front of the

    columns

    • Severe cases may require removal of up to 5

    meters

  • Page 77

    Column: DB-5ms, 30m x 0.25mm, 0.25um

    Carrier: H2, 60 cm/sec, constant flow

    Injector: split 1:20, 250C

    Detector: FID, 320C, N2 makeup gas

    Oven: 40C for 0.75 min, 40-325C at 20C/min, 325C for 30 min

    Red: WD-40

    Blue: system blank

    Contamination of system by residue on

    fingers during column installation

    Procedure:

    (1) One very small drop of liquid placed on one fingertip.

    (2) Fingertip was wiped with paper towel to remove as much of the offending material as possible.

    (3) Lightly touched the part of the column sticking up above the ferrule.

    (4) Installed column into injector.

    (5) Set oven temperature to 40C.

    (6) Started oven temperature program as soon as oven reached 40C.

  • Page 78

    Column: DB-5ms, 30m x 0.25mm, 0.25um

    Carrier: H2, 60 cm/sec, constant flow

    Injector: split 1:20, 250C

    Detector: FID, 320C, N2 makeup gas

    Oven: 40C for 0.75 min, 40-325C at 20C/min, 325C for 30 min

    Red: liquid soap

    Blue: system blank

    Contamination from Liquid Soap

    Procedure:

    (1) One very small drop of liquid placed on one fingertip.

    (2) Fingertip was wiped with paper towel to remove as much of the offending material as possible.

    (3) Lightly touched the part of the column sticking up above the ferrule.

    (4) Installed column into injector.

    (5) Set oven temperature to 40C.

    (6) Started oven temperature program as soon as oven reached 40C.

  • Page 79

    Contamination from Hand Lotion

    Column: DB-5ms, 30m x 0.25mm, 0.25um

    Carrier: H2, 60 cm/sec, constant flow

    Injector: split 1:20, 250C

    Detector: FID, 320C, N2 makeup gas

    Oven: 40C for 0.75 min, 40-325C at 20C/min, 325C for 30 min

    Procedure:

    (1) One very small drop of liquid placed on one fingertip.

    (2) Fingertip was wiped with paper towel to remove as much of the offending material as possible.

    (3) Lightly touched the part of the column sticking up above the ferrule.

    (4) Installed column into injector.

    (5) Set oven temperature to 40C.

    (6) Started oven temperature program as soon as oven reached 40C.

    Red: hand lotion

    Blue: system blank

  • Page 80

    Contamination of system by residue on fingers

    during column installation

    Column: DB-5ms, 30m x 0.25mm, 0.25um

    Carrier: H2, 60 cm/sec, constant flow

    Injector: split 1:20, 250oC

    Detector: FID, 320oC, N2 makeup gas

    Oven: 40oC for 0.75 min, 40-325oC at 20oC/min,

    325oC for 30 min

    Red: French Fry

    Blue: system blank

    Procedure:

    (1) Held French fry for 5 seconds.

    (2) Fingertip was wiped with paper towel to remove as much of the offending material as possible.

    (3) Lightly touched the part of the column sticking up above the ferrule.

    (4) Installed column into injector.

    (5) Set oven temperature to 40oC.

    (6) Started oven temperature program as soon as oven reached 40oC.

  • Page 81

    Tips to Maximize Septum Life,

    Minimize Septum Leaks

    – Use Agilent Gold Standard, HP Point, 23-26 gauge taper

    syringes. The point style cores septa significantly less when

    used with CenterGuide Septa. Taper minimizes septum

    coring/wear.

    – Use Agilent CenterGuide Septa. The molded hole

    minimizes septa coring, counter-intuitive, but true.

    HP-Point Style

    Solid Septum CenterGuide Septum

  • Page 82

    Easier question: What does poor inertness look like?

    Symptoms of poor GC system inertness:

    * Tailing peaks

    * Reduced peak response

    * No peak response

    * Extra peaks!

    * Poor linearity of a peak – usually at low

    concentrations

    What Does GC System Inertness Look Like?

  • Page 83

    Possible Inertness Problem Areas

    Inlet -liner, liner packing, gold seal, stainless steel

    Consumables - septa, syringe, vial, caps, inserts, solvents

    Column

    GC Detector - source geometry, material, column interface,

    acquisition rates?

    Temperatures - inlet, xfer line, source, quads, oven

    Other method factors, i.e. standards & sample preparation

  • Page 84

    How important is each component’s inertness to

    overall Flowpath Inertness?

    GC Flowpath Surface Areas

    l (cm) d (cm) pi

    Surface Area

    (cm2)

    Liner 7 0.2 3.142 4.4

    Seal 0.4 0.8 3.142 1.0

    Column 3000 0.025 3.142 235.6

  • Split Liners – What’s What?

    Straighttube

    Straighttube withglass wool

    Invertedcup

    BaffleFixed glass

    wool

    Page 85

  • Page 86

    Splitless Injection Liners

    Liner Part No. Comments

    5181-3316

    Single taper, deactivated, 900L volume. Taper isolates sample

    from metal seal, reducing breakdown of compounds that are

    active with metals. For trace samples, general application.

    5062-3587

    Single taper, deactivated, with glass wool, 900L volume.

    Glass wool aides volatilization and protects column. For trace

    (dirty) samples.

    5181-3315

    Double taper, deactivated, 800L volume. Taper on inlet

    reduces chance for backflash into carrier gas lines. High

    efficiency liner for trace, active samples.

    G1544-80730

    G1544-80700

    Direct connect liners, single and dual taper, deactivated.

    Capillary column press fits into liner end, eliminating sample

    exposure to inlet. Ultimate protection for trace, active samples.

    Side hole permits use with EPC.Side hole

  • Page 87

    •Liners become contaminated with use,

    collecting non-volatiles, salts, excess reagents,

    etc., or become damaged/cracked.

    •Should inspect and replace liners often.

    •Handle with gloves and forceps.

    •Insert into or remove liners only from cool

    injection ports.

    •Replacing with a new liner is recommended, to

    ensure reproducibility

    Liner Maintenance

  • Page 88

    What about “Baking Out” Columns?

    SURE! But…

    Only removes semi-volatile residues

    Limit to 1-2 hours

    May polymerize some contaminants

    Reduces column life

  • Page 89

    Common Care and Maintenance Scheme

    1. Bake out the column for no more than 2 hours.

    2. Change the Liner and cut off 6”-1ft of the inlet end of the column. (Change gold seal)

    3. Cut off more column. (repeat as necessary)

    4. Periodically clean the injector.

  • Page 90

    MS Troubleshooting Process (same as GC)

    1. Isolate the problem.

    (Blank Run, Inject Un-retained Compound, Jumper Tube Test)

    2. Change only one variable at a time.

    3. Compare before/after chromatograms.

    (Peak shape, response, retention, baseline rise, background, look for trends, etc.)

    4. Utilize Technical Support.

  • Page 91

  • Page 92

  • Page 93

  • Page 94

  • Page 95

  • AMU gain, offsetEntrance Lens,

    Repeller

    Ion SourceVolume

    Inlet (column)Filament

    FilamentDrawout

    Ion Focus

    Mass axis gain, offsetoffset

    Tuning Parameters - EI Inert Source

    Element Voltages Effect

    Filament Emission 0-315ua

    Electron Energy 10-241.5 volts

    Energy of electron beam

    Number of electrons generated

    Repeller 0 – 42.7 volts Pushes ions out of the source

    Drawout Ground potential Entrance aperture to lens stack

    Ion Focus 0 – 242.0 volts Relative abundance

    Entrance Lens 0 – 128 mV / amu Relative abundance

    Entrance Lens Offset 0 – 127.5 volts Relative abundance

    AMU Gain 0 – 4095 Peak Width

    AMU Offset 0 – 255 Peak Width

    Guide Rod or DIF +- 650 volts Focus / Guide Ions

    HED +-10,000 volts Converts ions to electrons

    Electron Multiplier 0 – 3000 volts Sensitivity

    Mass Axis Gain 2047 Mass assignment

    Mass Axis Offset 499 Mass assignment

    Electron Multiplier

    HED

    Guide Rod or DIF

    Page 96

  • AMU gain, offsetEntrance Lens

    Repeller

    Ion SourceVolume

    Inlet (column)Filament

    FilamentExtractor

    Ion Focus

    Mass axis gain, offset

    Tuning Parameters - EI Extraction Source 5977

    Element Voltages Effect

    Filament Emission current 0-315ua / Electron Energy 10-241.5 volts Number of electrons generated / Energy of electron beam

    Ion Body -4.24 – 20 volts Reference for Electron Energy

    Repeller 0 – 42.7 volts Pushes ions out of the source

    Extractor -10 – 10 volts Entrance aperture to lens stack

    Ion Focus 0 – 242.0 volts Relative abundance

    Entrance Lens 0 – 128 mV / amu Relative abundance

    Entrance Lens Offset 0 – 127.5 volts Relative abundance

    AMU Gain 0 – 4095 Peak Width

    AMU Offset 0 – 255 Peak Width

    Guide Rod or DIF +- 650 volts Focus / Guide Ions

    HED +-10,000 volts Converts ions to electrons

    Electron Multiplier 0 – 3000 volts Sensitivity

    Mass Axis Gain 2047 Mass assignment

    Mass Axis Offset 499 Mass assignment

    Electron Multiplier

    HED

    Guide RodIon Body

    Quadrupole

    Page 97

  • Autotune report evaluation

    Typical relative abundanceBase peak should be 69 or 219

    502 ≥2.4%

    Proper isotope ratios1.1, 4.4, 10.1% respectively

    ±20%

    Consistent mass peak widths ±0.05 of each other (0.48-0.65) Proper absolute abundance

    Minimum usually set to

    400,000-650,000 for 69

    Symmetrical smooth peak shapesNo split peaks ≥10%

    Total peak height

    Low water and air18 to 69 ≤20%

    28 to 69 ≤10%

    Correct mass assignments±0.1 a.m.u.

    Correct mass assignments

    ±0.1 a.m.u.

    Appropriate EM voltage ≤3000 maximum

    Page 98

  • Good Source Cleaning Practices

    Use aluminum oxide sand powder, green sandpaper,

    Bar Keepers Friend®, flexible sand sticks (LCMS), Soft Scrub®

    Lightly clean flat surfaces and focus on inside surface of lens

    openings

    Examine repeller surface and ensure flat rounded surface

    Solvent wash cleaned source parts in 1water, 2acetone or

    methanol, 3dichloromethane or hexane

    (don’t solvent wash ceramics or polymer insulators)

    Change your filaments

    Page 99

  • Troubleshooting Tools Needed for MS

    Know good pressure readings for your system (high/rough vacuum)

    Known good tune voltages of a clean well functioning system

    Know your background abundances (H2O, N2, CO2, column bleed ions i.e. m/z 207)

    Good column background of blank baseline of column bleed

    Good chromatogram of your mid to low standard or calibrator (prefer in matrix)

    Page 100

  • Potential MS Leak Sites

    O-ring around analyzer

    MS column interface connection

    GC inlet

    Pumping system problems

    Page 101

  • Vacuum Leak Troubleshooting Process

    Isolate the system compartments (Gas flow, GC inlet, column, MS interface, MS)

    Check age of gas traps, isolate from flow path (can become saturated)

    Check septa, liner O-ring, Au seal, ferrule/column installation

    Use dust off and manual tune window to monitor at m/z 51 and 65

    Cap column off from inlet with a septa and allow MS to pump for ~ 30 mins.

    Cap off MS with solid blank ferrule

    Measure the flow coming out of rough pump

    Page 102

  • Questions…

    Thank you for your attention

    103