Select Start

Reaction starts: the green LED of the section lights up

Test time

Reaction ends: the green LED flashes

Remove the strip(s) and SPR(s) of the sectionResults print out automatically

Interpretation and measurement range

HOW TO PERFORM MAINTENANCE?

Weekly: - cleaning of the SPR blocks

Monthly: - cleaning of the reagent strip area- performing VIDAS QCV test- cleaning of the two optical lenses

02-1

0 /

008G

B990

71C

/ Th

is d

ocum

ent i

s no

t leg

ally

bin

ding

. bio

Mér

ieux

rese

rves

the

right

to m

odify

spe

cific

atio

ns w

ithou

t not

ice

/ BI

OM

ERIE

UX

and

the

blue

logo

, are

regi

ster

ed a

nd p

rote

cted

trad

emar

ksbe

long

ing

to b

ioM

érie

ux S

.A. o

r one

of i

ts s

ubsi

diar

ies

/ bi

oMér

ieux

SA

RCS

Lyon

673

620

399

/ P

rinte

d in

Fra

nce

/ TH

ERA

Con

seil

/ RC

S Ly

on B

398

160

242

If required, select User ID, if not press

q

SEEPACKAGE

INSERT

SEEPACKAGE

INSERT

bioMérieux S.A.69280 Marcy l’EtoileFranceTel.: 33 (0)4 78 87 20 00Fax: 33 (0)4 78 87 20 90www.biomerieux.com

SEE

MINI VIDAS

USER

MANUAL

SEE LENS

CLEANER

PACKAGE

INSERT

GCS / IMMUNOASSAYSSIMPLIFIED USER GUIDE

mini VIDAS®

2010/02/01

HOW TO KNOW IF A CALIBRATION IS VALID?

To do this, proceed as follows:

Press until the MAIN MENU is displayed

Select the test code in the list by pressing on the key to scroll down/up if necessary

Test code

● If the lot does not feature in the list, the lot has not been calibrated:

Press three times to return to the MAIN MENU

and follow the protocol on page 4

● If the lot features in the list, check whether the calibrationis still valid:

Press on the key corresponding to the lot required, and the following screen will be displayed:

Control the lot expiration date displayed on the screen

● If the calibration has expired:

Press three times to return to the MAIN MENU

and follow the protocol on page 5

● If the calibration is still valid:

Press three times to return to the MAIN MENU and follow the protocol on page 6

HOW TO UPDATE A PROTOCOL?Always read the protocole barcode before reading the MLE card

MAIN MENU

Select the Utility Menu

Select the Master lot menu

Select List Stored Stds

● If the test does not feature in the list, the test has not been calibrated:

Press twice to return to the MAIN MENU

and follow the protocol on page 4

● If the test features in the list, check whether the calibration isstill valid:

Select the lot if necessarySelect Assay Update

q q

q

q

q

q

SEEPACKAGE

INSERT

2 3

!IMPORTANT: The key enables the user to return to the MAIN MENU at any time.

Fulfil the warning instructions and click on Yes to continue

Read the protocole barcode printed on last page of the package insert with the barcode wand

Select Confirm to Quit

HOW TO READ A MLE CARD?

MAIN MENU

Selectthe Master Lot Menu

Place the MLE card on the white plastic card support

Slide the card support into an empty section

Select Read Master Lot

Select the chosen section

The message “Please wait”is displayed on the screen

As soon as the message disappears,remove the card and support

Press to return to the MAIN MENU

Then calibrate using the protocol below

HOW TO RUN STANDARDS (S) AND CONTROLS (C) FOR CALIBRATION?

If an error is made when selecting a standard, control or sample:

• select the position n° to be modified (1 to 6)• select “Assay”• select “Clear”

• select “Clear this position’s setup only”

MAIN MENU

Select Status Screen

Type 1

Repeat this operation for each standardused for the test

Select section A or B

Select S for the standard, type 1, then press

q

q

q

q

qq

q

SEEPACKAGE

INSERT

4 5

!

Select C for control, type 1 for C1then press

Repeat this operation for each control used for the test

Then follow the protocol below from

HOW TO IDENTIFY SAMPLES IN STATUS SCREEN?

MAIN MENU

Select Status Screen

Type 1 *

Enter the Sample ID (see below) and, if necessary,the dilution factor

Press

After identifying the last patient, press until the work list for the section is displayed, then start running the samples using the protocol below

Sample identification(maximum of 12 alphanumeric characters)

Numeric characters are entered using the keyboardAlphabetic characters are entered by selecting each character on the screen:

Use the arrow pad on the keyboard to move the cursor and select the characters, then validate them one by one by pressing

Sample ID can also be entered by using the barcode wand

HOW TO RUN SAMPLES?

If required, leave reagents 30 min at room temperature before use

Check the type of sample and pre-process samples if required

Place the strip(s) and SPR(s) into the section indicated on the screen

Dispense the sample volume into the sample well of the strip

Repeat the same actions for each sample

*

Select a free section

Select Sample ID

q

q q

q

q

SEEPACKAGE

INSERT

IMPORTANT: The key enables the user to return to the Main Menu at any time.

SEEPACKAGE

INSERT

6 7

Select Start

Reaction starts: the green LED of the section lights up

Test time

Reaction ends: the green LED flashes

Remove the strip(s) and SPR(s) of the sectionResults print out automatically

Interpretation and measurement range

HOW TO PERFORM MAINTENANCE?

Weekly: - cleaning of the SPR blocks

Monthly: - cleaning of the reagent strip area- performing VIDAS QCV test- cleaning of the two optical lenses

02-1

0 /

008G

B990

71C

/ Th

is d

ocum

ent i

s no

t leg

ally

bin

ding

. bio

Mér

ieux

rese

rves

the

right

to m

odify

spe

cific

atio

ns w

ithou

t not

ice

/ BI

OM

ERIE

UX

and

the

blue

logo

, are

regi

ster

ed a

nd p

rote

cted

trad

emar

ksbe

long

ing

to b

ioM

érie

ux S

.A. o

r one

of i

ts s

ubsi

diar

ies

/ bi

oMér

ieux

SA

RCS

Lyon

673

620

399

/ P

rinte

d in

Fra

nce

/ TH

ERA

Con

seil

/ RC

S Ly

on B

398

160

242

If required, select User ID, if not press

q

SEEPACKAGE

INSERT

SEEPACKAGE

INSERT

bioMérieux S.A.69280 Marcy l’EtoileFranceTel.: 33 (0)4 78 87 20 00Fax: 33 (0)4 78 87 20 90www.biomerieux.com

SEE

MINI VIDAS

USER

MANUAL

SEE LENS

CLEANER

PACKAGE

INSERT

GCS / IMMUNOASSAYSSIMPLIFIED USER GUIDE

mini VIDAS®

2010/02/01