AUTHORIZED DISTRIBUTOR - GasDetectorsUSA.com - Houston, Texas USA [email protected] GDS-68XP QuickStart Guide Mount Ver?cally 1 3 Keep Away From Vibra?on & Shock 2 Don’t Face Into Direct Sunlight 4 +24VDC INPUT 4-20mA OUTPUT 110VAC INPUT FOR HEATER PURGE AIR INLET SAMPLE INLET FILTER DRAIN SAMPLE OUTLET ETHERNET, OPTIONAL RELAYS and MODBUS SEE GDS-68XP MANUAL FOR SAMPLE INLET GUIDELINES SEE GDS-68XP MANUAL FOR SAMPLE OUTLET GUIDELINES - 832-615-3588

GDS-68XP QuickStart Guide - Gasdetectorsusa.com · GDS-68XP QuickStart Guide 1 Mount Ver?cally 3 Keep Away From Vibra?on & Shock 2 Don’t Face Into Direct Sunlight 4 +24VDC INPUT

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: GDS-68XP QuickStart Guide - Gasdetectorsusa.com · GDS-68XP QuickStart Guide 1 Mount Ver?cally 3 Keep Away From Vibra?on & Shock 2 Don’t Face Into Direct Sunlight 4 +24VDC INPUT

AUTHORIZED DISTRIBUTOR - GasDetectorsUSA.com - Houston, Texas USA

[email protected]

GDS-68XP QuickStart Guide

Mount Ver?cally 1 3 Keep Away From Vibra?on & Shock

2 Don’t Face Into Direct Sunlight

4 +24VDC INPUT 4-20mA OUTPUT

110VAC INPUT FOR HEATER

PURGE AIR INLET

SAMPLE INLET

FILTER DRAIN SAMPLE OUTLET

ETHERNET, OPTIONAL RELAYS and MODBUS

SEE GDS-68XP MANUAL FOR SAMPLE INLET

GUIDELINES SEE GDS-68XP MANUAL FOR

SAMPLE OUTLET

GUIDELINES

- 832-615-3588

Page 2: GDS-68XP QuickStart Guide - Gasdetectorsusa.com · GDS-68XP QuickStart Guide 1 Mount Ver?cally 3 Keep Away From Vibra?on & Shock 2 Don’t Face Into Direct Sunlight 4 +24VDC INPUT

MEAS-1% N

E

RR

T

F0.05

2.21

ppm H2S

Raw Sensor

MEAS-1% N

E

RR

T

F0.05

2.21

ppm H2S

Raw Sensor

MEAS-1% N

E

RR

T

F0.05

2.21

ppm H2S

Raw Sensor

Step 1: Disconnect DC Power

Step 2: Disconnect flow cell inlet

Step 3: Disconnect flow cell outlet

Step 4: Unscrew sensor head cover and flow cell together

Step 5: Pull sensor straight down to remove (do not twist).

Step 6: Verify new sensor part number against exis?ng sensor.

Step 7: Align arrow on new sensor with engraved arrow on sensor head and push straight up.

Step 8: Carefully reassemble sensor head cover / flow cell and rea?ach inlet and outlet tubing. Make sure outlet tubing does not touch heater.

Step 9: Apply DC power and allow sensor to stabilize (see manual).

Step 10: Recalibrate.

GDS-68XP Sensor Replacement