GDS-68XP QuickStart Guide
Mount Ver?cally 1 3 Keep Away From Vibra?on & Shock
2 Don’t Face Into Direct Sunlight
4 +24VDC INPUT 4-20mA OUTPUT
110VAC INPUT FOR HEATER
PURGE AIR INLET
SAMPLE INLET
FILTER DRAIN SAMPLE OUTLET
ETHERNET, OPTIONAL RELAYS and MODBUS
SEE GDS-68XP MANUAL FOR SAMPLE INLET
GUIDELINES SEE GDS-68XP MANUAL FOR
SAMPLE OUTLET
GUIDELINES
- 832-615-3588
MEAS-1% N
E
RR
T
F0.05
2.21
ppm H2S
Raw Sensor
MEAS-1% N
E
RR
T
F0.05
2.21
ppm H2S
Raw Sensor
MEAS-1% N
E
RR
T
F0.05
2.21
ppm H2S
Raw Sensor
Step 1: Disconnect DC Power
Step 2: Disconnect flow cell inlet
Step 3: Disconnect flow cell outlet
Step 4: Unscrew sensor head cover and flow cell together
Step 5: Pull sensor straight down to remove (do not twist).
Step 6: Verify new sensor part number against exis?ng sensor.
Step 7: Align arrow on new sensor with engraved arrow on sensor head and push straight up.
Step 8: Carefully reassemble sensor head cover / flow cell and rea?ach inlet and outlet tubing. Make sure outlet tubing does not touch heater.
Step 9: Apply DC power and allow sensor to stabilize (see manual).
Step 10: Recalibrate.
GDS-68XP Sensor Replacement
1
2
3
4
9
5
7
8
6