GE Healthcare
Discovery* RF180 PRODUCT DATA SHEET Remote X-Ray Imaging R&F System
2 | GE Healthcare
System Overview
One system designed for all.Flexible, precise, and low dose, the Discovery RF180 is a powerful remote imaging system that fosters great diagnostic certainty, all from the first image. Simple and intuitive, it provides a streamlined clinical workflow optimized for quality-infused value-based care. It’s also one of the only systems that can conduct a wide variety of exams, including tomosynthesis and pasting, helping ensure unprecedented performance, uptime, and fast ROI.
Main features
• Table-top height adjustable• Source Image Distance (SID) up to 180 cm (71”)• Patented autofocusing grid• Total patient accessibility with open design rear access • Maximum patient's weight with full movements 266 kg (586 lbs)• Innovative table joysticks controls• Smart digital system with multi touch screen display• Radiography and fluoroscopy 43 x 43 cm digital detector (17 x 17”)• Full DICOM functionality• Advanced Applications available• Upgradeable with advanced applications
Main applications
• Gastroenterology• Skeleton• Thorax and lungs• Pediatrics• Urology and gynecology• Emergency/traumatology
System Features
All about easy. Your day streamlined.Serve a high patient volume, optimize the patient experience, and help deliver the best care possible with fast and streamlined workflow. Intuitive touchscreen user interface and AutoRF simplify exam set-up with just a few clicks and less need to physically move equipment. Large, elevating easy-access table with touch controls accommodates a variety of patient sizes and ages and enables easy positioning. Features true rear accessibility to facilitate patient transfer and radiologist approach.
Exam table
90/90 universal remote system table with elevating movement independent from the table tilting and designed for 43 x 43 cm (17 x 17”) dynamic flat panel. Minimum table height of only 47cm.Single suspended patient tabletop with FULL rear access to the patient that simplifies the transfer from/to the stretcher, ensures immediate intervention and less operator’s physical effort.X-ray tube/detector support with maximum longitudinal travel, coupled with longitudinal motorized table movements, yield patient coverage up to 305 cm.
X-ray tube/detector speed adjustable in continuous mode by the console control joystick. Movement also controlled by fixed speed from the frontal detector support console.
Exams on stretchered patient
Radiological examinations can be achieved outside the table-top and on stretchered patient.
Operator interface
Main contact points between operator and system.
Remote console
• Innovative user-friendly table control console• Control console with joysticks and stop push button to assist the
operator in the most suitable table positioning
Product Data Sheet | 3
Smart and highly innovative digital system
• Smart integrated 23-inch monitor touch screen display• Generator, table, and digital system controls are integrated in the
monitor display.• Multi touch gesture to optimize the workflow - easy to customize
The system stores acquisition, processing, and visualization parameters related to anatomic part and patient size. This means single-screen control of the total examination, avoiding additional system interactions.
Post-processing functions are available to modify the images and simplify its management.
Table integrated keyboard
• Table movements control• X-ray tube assembly movement control• X-ray collimator beam control• Emergency push stop
Lead free collimator with touch screen display
• Touchscreen interface display with collimator controls integrated.• Table and X-ray tube assembly movement controls• Touch screen collimator display• LED control, blades opening/closing control; Source Image
Distance (SID) control (step less adjustable), automatic/manual control. Additional filtration
• Key button screen change• Table movement controls: elevation, transversal, longitudinal
patient table top movements• X-ray tube assembly• Tube inclination angle • Orthogonal X-ray tube positioning
4 | GE Healthcare
Auto RF Features
Suite of features to further simplify your workflow includes:
Auto-protocol assist
System will automatically transition directly to the Acquire screen when the protocol code downloaded from the HIS/ RIS (automatically performed with worklist refresh) matches the exam code contained in the protocol database. This tool eliminates the user steps required to select patient exam types and initiate an exam.
Auto positioning
The table automatically reaches a pre-set anatomical position if the auto-positioning mode is enabled by the service team during in-stallation.
The available positions are: automatic table tilts, X-ray tube drive position, longitudinal detector position, patient table-top lateral position, source image distance SID, collimation and grid parameters.
Auto field of view
Auto Field of View enables the user to pre-define the collimation size on an individual view basis.
Auto focusing grid
Grid with an exclusive autofocusing device that automatically set the correct grid focalization according to the selected SID.Easy grid removing to allow dose reduction.
All about confidence – Diagnostic Benefits
This intuitive, powerful imaging tool also helps to deliver great diagnostic certainty. It has been reimagined with the latest premium FPD low-dose technology, specific pediatric features and radiation-free collimation and positioning to minimize dose. With the ability to perform a wide range of exams, and advanced apps like DSA, tomosynthesis and auto image paste the Discovery RF180 is designed to deliver fast, high-quality images.
Tomographic technique
The remote-controlled table performs high level tomographic exams. The digital control assures precision with four different angles, increasing and decreasing automatic cutting plane.
Auto image paste (optional function)
The auto image paste function is an optional module of the table system that involves the automatic acquisition and recomposition of a set of radiography images. The pasted image has all the source pixels; it can be viewed on the monitor, processed, printed or sent via network by usual digital image system functions. The system includes the spine and leg automatic reconstruction with patient lying or standing. The operator can program the images reconstruction with selectable length of 60 – 90 –120 cm (24 – 35 – 47”).
Tomosynthesis (optional function)
The system offers the possibility of a high innovative diagnostics examination in the medical field: "tomosynthesis". The application range vary from chest, abdomen and orthopedics examination. The tomosynthesis function acquires images that allow the analysis of an entire volume using a series of low dose rapid exposures.
Tomosynthesis involves a series of x-ray exposures during a single tomographic sweep with a fixed image receptor; the system then reconstructs the data to visualize multiple level planes (slices) from the surface of the image receptor up through the imaged anato-my. It removes overlapping/overlying structures and enhances the conspicuity of anatomy in the different slices. Thanks to the innovative approach in image acquisition, it is possible to achieve very high-quality images.
Digital subtraction angiography (DSA) (optional function)
The DSA option gives the following possibilities:• Real-time acquisition images in subtract mode.• Possibilities to program up to six different phases, each phase can
be set timing and acquisition rate.• Road-mapping function with maximum opacity function and
subtraction.• 8 reference images selectable on dedicated monitor
(reference monitor).• Review runs sequence in cine loop mode.• Easy change of mask image in post-processing mode with direct
result viewing.• Calibration measurement directly on catheter.• Percentage stenosis calculation.
Auto image processing
High-definition images display optimized within a short time.In acquisition sequence, real-time display preview format. After acquisition, automatic final format images availability.
Compressor device (optional function)
Compressor device with motorized parking. Mechanical and electrical safety (double safety).
FFD1
FFD2
Product Data Sheet | 5
Performance amplified – Maximize room utilization
Boost uptime and patient volume to accelerate ROI with an all-in-one system that supports a wide range of examinations. Designed to keep your fluoroscopy department busy, the Discovery RF180 can perform well beyond simple radiography and fluoroscopy. Your system utilization is optimized by the following features:
• Variable SID with maximum 180 cm permits chest exams with no additional equipment
• One of the few systems to accommodate bariatric exams• Accessible equipment for mobility-impaired patients, form
pediatric to geriatric• Extended movements enable standing lower limb exams without
need for patient to climb• Supports a wide variety of advanced exams (tomosynthesis, auto
image paste, DSA)• Remote service capabilities to quickly identify problems and
resolve issues
EXAM TABLE
Tilting Motorized from + 90° to – 90°
Tilting speed Variable, continuous from 0 to 5°/s
PATIENT TABLETOP
Dimensions 246 x 80 cm (97 x 31”)
Radio-transparent area 237 x 57 cm (93 x 22”)
Material Carbon fiber
Patient tabletop type Flat or Concave (upon request)
Patient tabletop movements type With longitudinal travel
Minimum floor distance From 47.3 cm to 100 cm (18.6 to 39,4”)
Inherent filtration 0.5 mm (0.019”) Al/eq a 100 kVp
Maximum patient's weight 266 kg – 586 lbs (323 kg – 712 lbs with limitations)
Vertical travel speed Step less from 0 to 45 mm (2”)/s
Transversal motorized travel ±17,5 cm (7”)
Lateral travel speed Step less from 0 to 25 mm (1.18”)/s
6 | GE Healthcare
Longitudinal travel ±50 cm (20”)
Longitudinal travel speed 45 mm (2.4”)/s
Maximum tabletop rear access distance 52.5 cm (20.5”) max
X-RAY TUBE ASSEMBLY
Longitudinal motorized travel (X-ray tube only) 195 cm (77”)
Longitudinal motorized travel X-ray tube/detector
162 cm (64”)
Longitudinal speed X-ray tube assembly Motorized movement. Step less from 0 to 150 mm (6”)/s
X-ray beam assembly angulations ± 40°
Focus / detector distance From 115 to 180 cm (45 to 71”) step less, motorized
Focus/floor distance (Tilt +90°) From 40.5 to 226.5 cm (15.9 to 88.9”)
X-ray tube rotation movement 90°/180° manual movement
COMPRESSION CONE (OPTION)
Compression deviceMotorized compressor cone with double safety system (mechanical and electrical)
TOMOGRAPHY
Stratigraphy type Linear tomography in any tilting table angle
Focus/film distance 115 cm (45”)
Layer cut adjustmentFrom 0 to 300 mm (0 to 12”) to with millimetric adjustment
Pre-programmed modes Automatic increment/decrement of the layer cut to be set by the operator before starting the tomographic technique
Product Data Sheet | 7
AUTOMATIC COLLIMATOR WITH TOUCH SCREEN DISPLAY
Collimator type Motorized with square field limitation device
Lamp type simulation Led lamp 7 W
Lighting average (IEC 601.1.3) > 200 lux a FFD = 100 cm (39”)
Fixed filter 0.5 mm Al (0.019”)
Optional additional filters1 mm Al + 0.1 mm Cu / 1 mm Al + 0.2 mm Cu / 2 mm Al + 0.3 mm Cu
DIGITAL IMAGE PROCESSOR
Composition
Processor unit for digital image processing Main Controller for X ray synchronization and measure of the X ray dose when the AEC is selected PU device, for detector connection and image acquisition
Operating system Microsoft Windows 7
Connection netSpeed 10/100/1000 Mbit/s (auto)
Protocol TCP/IP
IP address customizable in 4 bytes
IP mask customizable in 4 bytes
DETECTOR
Pixium RF4343 Pixium RF4343FL
Type Amorphous silicon
Technology Caesium Iodide
Matrix size 2880 x 2880 pixel
Dynamic range – linear response 16 bits
Pixel pitch 148 μm
Acquisition area
Nominal: 43 x 43 cm (17 X 17”)Zoom 1: 30 x 30 cm (12 x 12“)
Zoom 2: 20 x 20 cm (8 x 8”)Zoom 3: 15 x 15 cm (6 x 6”)
8 | GE Healthcare
Maximum acquisition rate
Continuous fluoroscopy 30 img/s 20 img/s
Pulsed fluoroscopy 15 img/s 12 img/s
Spatial resolution 3.4 lp/mm
DQE 65% (@ 0 lp/mm)
MTF @ 1 lp/mm 63% 55%
MTF @ 2 lp/mm 32% 25%
Cover filtration < 0.3 mm (0.12”) Al/eq a 100 kVp)
Grid features 80 l/cm, 12:1 carbon fiber
Grid focalization From 115 to 180 cm (45 to 71”) with "
AUTOFOCUSING GRID" device
Grid removal Manual removal
Continuous digital fluoroscopy
Size MatrixFrequency HIRIS
RF4343Frequency HIRIS
RF4343FL
43 x 43 cm (17 x 17”) 960 x 960 x 16 bit 18 img/s 16 img/s
30 x 30 cm (12 x 12”) 1024 x 1024 x 16 bit 15 img/s 12 img/s
20 x 20 cm (8 x 8”) 672 x 672 x 16 bit 30 img/s 20 img/s
15 x 15 cm (6 x6”) 1024 x 1024 x 16 bit 15 img/s 6 img/s
Last Image Hold
Possibility to record a fluoro loop or LIH
Noise reduction with “motion sensitivity”
Multi step edge enhancement
Digital image inversion (Horizontal/Vertical)
Product Data Sheet | 9
Pulsed digital fluoroscopy
Size MatrixFrequency HIRIS
RF4343Frequency HIRIS
RF4343FL
43 x 43 cm (17 x 17”) 960 x 960 x 16 bit 15 img/s max (1) 12 img/s max (1)
30 x 30 cm (12 x 12”) 1024 x 1024 x 16 bit 15 img/s max (1) 12 img/s max (1)
20 x 20 cm (8 x 8”) 672 x 672 x 16 bit 15 img/s max (1) 12 img/s max (1)
15 x 15 cm (6 x6”) 1024 x 1024 x 16 bit 15 img/s max (1) 6 img/s max (1)
(1) Adjustable from 1 img/s
Last Image Hold
Possibility to record a fluoro loop or LIH
Noise reduction with “motion sensitivity”
Multi step edge enhancement
Digital image inversion (Horizontal/Vertical)
Digital radiography
Size MatrixFrequency HIRIS
RF4343Frequency HIRIS
RF4343FL
43 x 43 cm (17 x 17”) 2880 x 2880 x 16 bit 3 img/s max (1) 2 img/s max (1)
43 x 43 cm (17 x 17”) 1440 x 1440 x 16 bit 8 img/s max (2) 6 img/s max (2)
(1) HR mode – high resolution(2) HS mode – high speed
Direct hard disk image saving, Multi-step edge enhancement, H/V Digital image inversion
Anatomical program
Library up to 99 parts of the body, each one with 99 projections performed with seven different patient types and for a total of 68.607 ways of using the system. The expositions parameters (kV, mA automatic exposure meter dominant) and the irradiated field can be changed by the operator during the examination.
10 | GE Healthcare
Automatic exposure
Completed with measure chamber at three dominant exposure meter which are selectable independently by the operator or during anatomical programming.Ability to link a predefined dose value between 0,5 uGy and 5,0 uGy (at the detector level) at each single anatomical program in function of the patient size (seven sizes). Generator automatic kV and mA set values. Exposures time fixed by the automatic exposure meter for constantly maintaining the operator pre-set dose and based on the exam type selection.Generator pulsed fluoroscopy dynamic kV value variation to optimize the image for each acquisition rate.
Exams documentations management
Each acquired image has indicated:- Patient name- Label number- Birth date- Date and time of the exam execution- Institute name- Radiographic parameters (kV, mAs),- Dose (mGy cm2)Display images parameters and reference scale.Archives querying and sorting by exam date or patient name.Possibility to define the default output (printers, writer/burner CD/DVD, workstation or PACS). Multi-Store.Indication on monitor of the exam sent to PACS or printer.Management of network printers and film optimization with several reproduction modalities.
Real-time functions processing
Algorithms studied for each exam type to optimize image processing and display. High definition images display optimized within a short time.In acquisition sequence, real-time display preview format. After acquisition, automatic final format images availability. Application of post-processing algorithms without altering the source data.
Anatomic Tissue Harmonization (ATH) & Fluoro Tissue Harmonization (FTH)
Radiographic (ATH) and fluoroscopic (FTH) software real time images processing to improve the acquired images with contrast increase, brightness and noise reduction. Increasing of the dynamic range for the acquired images.Adjustable anatomical program parameters and data customization during installation and in function of operator needs.
Product Data Sheet | 11
Post-processing functions
- Single image display or multi-image simultaneously displayed
- Automatic contrast and brightness adjustment- Window Level and gamma adjustment- Zoom with variable enlargement 1:3- Grey Scale Inversion- H&V image reverse- 90° image rotation and free rotation- SHARP (edges) and SMOOTH (faded edges) spatial
filters, kernel size, and weight of the applied filter- Electronic shutters (square, and quadrilateral)- Virtual collimation (open/close the collimator
controls on LIH to get the required result without X-ray emission)
- Virtual scan (optional): this function allows patient centering without X-ray emission (this function uses the LIH in fluoroscopy or pulsed fluoroscopy that can be shifted on the monitor without X-rays emission).
- Cine-loop of dynamic images sequence with different speed
- Text and marker overlay- Graphic calculation of angles and distances (mm/pixel)- Grid overlay- COBB angle and orthopedic measurement (option)- Image export in RAW, JPEG and MP4- Reject analysis (flag the image currently shown as
rejected and then specify the reason for its rejection)
Operator interface
Table joysticks control console.Digital system with multi touchscreen controls integrated:• generator and table management• images display• post-processing images management• images management• patient data management• Work list communication, RIS-PACS system
printers, writer/burner CD/DVD• fluoro loop, possibility to record a loop after
fluoroscopy• User Login Management (Max 50 Users)• 7 selectable Patient sizesThe anatomical mode program can be checked by the operator through the image processor monitor and through the work list.
12 | GE Healthcare
Connectivity
Ethernet TCP/IP network interface via DICOM protocolSTANDARD DICOMDICOM StoreDICOM Send serviceDICOM Print classDICOM work listDICOM MPPSOPTIONAL DICOMDICOM storage commitmentDICOM query/retrieveDICOM DOSE SRDICOM Media interchange (CD/DVD)
Supported MonitorsControl room monitorOptional monitors connection
Optional special procedures
IMAGE RECONSTUCTION (AUTO IMAGE PASTE)Automatic acquisition and re-composition of a set of radiography images (each time a different section of the patient is irradiated until a complete large format image is obtained). Typically for exams concerning the spine or legs.TOMOSYNTHESISDigital image processing function producing multiple slices starting from a single image set acquired through tomography with limited rotation angles.ANGIOGRAPHY (DSA)• Mask shift• Image subtraction• Pixel shift• Vascular tracing• Land marking• QA analysis
MONITOR
Monitor type and dimension 23” color monitor
Native resolution 1920x1080
Viewing angle 178° H/V
LCD Technology 260 cd/m² (typical)
Contrast 1000:1 typ.
Monitor weight 6.6 Kg (14.5 lbs)
Monitor dimension (mm) 556.7 x 143.9-360.7 x 89-401.3 (22x5.7-14 x 3.5-15.8”)
Product Data Sheet | 13
GENERATOR
Type 65 R/F 80 R/F
Frequency High frequency Output – (maximum 400 kHz)
Power 65 kW 80 kW
Output parameters G650 G800
KV mA mA
80 800 1000
100 630 800
150 400 500
Starter speed Anode rotation 3000/9000 rpm
Line voltage rate ±10%
Radiography
kVp Range/Steps 40 - 150 kV in 1 kV increments
High voltage ripple <1kV at 110 kV
mA range/steps (*Rénard)(1 mA/0,1 mA optional steps)
10 – 800 mA* R’10
10 – 1000 mA * R’10
Time range 1.0 to 6300 milliseconds
mAs Range (no AEC) 0.1 – 1000 mAs/ R’10 *
Continuous Fluoroscopy
kVp Range/Steps 40 - 125 kV in 1 kV steps
High voltage ripple <1kV at 110 kV 5mA
mA range/steps 0.5 – 10 mA in 0,1 mA steps
Pulsed Fluoroscopy
kVp Range/Steps 40-125 kV in 1 kV steps
mA Range/steps 5-99 mA in 1 mA steps
14 | GE Healthcare
RADIATION SOURCE ASSEMBLY - G292
Anode type Rotating anode 3.000/10.000 rpm
Anode material RTMC
Anode diameter (mm) 102 mm
Anode angle 12°
Maximum tension 150 kV
Focal spots dimension 0,6/1,2 mm
Maximum power to 10.000 rpm 40/100 kW
Maximum power to 3.000 rpm 30/60 kW
Anode heath capacity 445 kJ (600 kHU)
Maximum Anode cooling rate 125 kHU/min (1540 W)
Housing heath capacity 1480 kJ (2000 kHU)
Maximum housing dissipation capacity 445 Watts (600 HU/sec)
OTS - OVERHEAD TUBE SUSPENSION (OPTION)
Type Manual
Rails Longitudinal x Transversal
Standard length 441 x 301 cm
Optional length441 x 361 cm361 x 301 cm
Transversal Bridge
Transversal carriage rails travel 354 - 274 cm
Ceiling Stand
Transversal carriage rails travel 217 - 277 cm
Elements number 5 extruded aluminum
Motorized vertical travel 170 cm
Type of movement Manual
Balancing method Motorized at “0 strength”
Adjustment of balancing strength Automatic
Product Data Sheet | 15
Radiation Source Assembly Support
X-ray tube rotation+/- 160° around vertical axis+/- 155° around horizontal axis
Focus/X-ray source axis minimum distance 42,9 cm
Focus/ceiling minimum distance 95,5
Focus/floor minimum distance 33
Rotation movements brakes Electro mechanics
Ceiling Suspension Control Panel
Operator interface Color display touch screen 10,4”
Brakes unblocks “Capacitive” handle functions
Typical exam room height 295 cm
Manual Collimator
Collimator type manual collimator front panel with knobs
Rectangular field coverage 48x48 cm @1m
SID operative 90 ÷ 200 cm
Minimum collimation field 0 x 0 cm
Inherent filtration 2 mmAl/eq
Scattered radiation <40 mR/h
Manual hardening filters2 mm Al1 mm Al + 0.1 mm Cu1 mm Al + 0.2 mm Cu
Simulation light Power LED
Lightening average > 160 lux at FFD = 100 cm
Light field edge contrast ratio > 4 @SID = 100 cm
X-ray field/light field accuracy <1% SID
Scale /light field size accuracy < 2% SID
SID indication accuracy < 2% SID
16 | GE Healthcare
Radiation Source Assembly RTM 101
Type Type RTM 101 / HS
X-ray tube assembly Rotating anode 3.000/10.000 rpm
Anode material Rhenium, Tungsten, Molybdenum
Anode diameter (mm) 102 mm
Anode angle 12,5°
X-Ray tube assembly maximum voltage 150 kV
Focal spots dimension 0,6/1,2 mm
Maximum power to 10.000 rpm 40/100 kW
Maximum power to 3.000 rpm 26/63 kW
Anode heath capacity 400 kHU - 300 kJ
Housing heath capacity 1.280 kJ
Anode dissipation capacity 125 kHU/min - 1000 W
Housing dissipation capacity 370 W (with fan)
Total minimum filtration (housing/X-ray tube/collimator)
> 2,5 mm Al/eq
INSTALLATION TECHNICAL DATA (OTS)SYSTEM POWER SUPPLY
Ceiling stand power supply (set during the installation)
115V-230V (± 10%) monophasic
Voltage power supply tolerance ± 10%
Nominal line frequency 50/60 Hz
Line tolerance ± 1 Hz
Standby power 300 VA
Peak power 2.000 VA
Weights And Dimensions
Ceiling stand extendable elements weights with transversal bridge (including collimator and x-ray tube)
300 Kg*
Ceiling rails weights (longitudinal) 70 Kg*
Product Data Sheet | 17
Thermal Dissipation
Kalos ceiling suspension Approx. 1000 kcal/h
* heaviest configuration
WALL BUCKY STAND (OPTION)
Model WS Vertically adjustable
Model WST Vertically adjustable with tilting device
Column structure with floor base and wall fixing.Bucky for radiographic cassettes up to 35x43 cm (14 x 17”) with counter-balanced vertical movement Complete with grid (R12/L90/F180) and pre-set to receive an automatic exposure meter.Mechanical locking on all movements.
WS WST
Bucky Vertical Travel 151.5 cm (59.65”) 140 cm (55.12”)
Min distance floor to bucky center vertical position / horizontal position
40 cm (15.75”)37.5 cm (14.76”)/ 68.5 cm
(26.97”)
Max distance floor to bucky center vertical position 191.5 cm (75.39”) 177.5 cm (69.88”)
Tilting positions NoYes, 0°, +90° (with bucky
rotation), –20°
WIRELESS DETECTORS (OPTION)
35 X 43 Wi-Fi Pixium 3543Ez Detector
GAD Csl
Technology Amorphous Silicon
Scintillator Gadolinium oxysulphide Caesium Iodide
Pixel area 35 x 43 cm (14 x 17“)
Matrix size 2400 x 2880 pixel
Pixel pitch 148 μm
Data conversion 16 bit
Spatial resolution 3,4 pl/mm
DQE 37% typ. 66% typ.
18 | GE Healthcare
Batteries chargerYes, up to three batteries simultaneously
(two batteries included)
Weight battery included 2.8 kg (6.2 lbs)
24 X 30 Wi-Fi Pixium 2430Ez Detector
Technology Amorphous Silicon
Scintillator Caesium Iodide
Pixel area 24 x 30 cm (9 x 12“)
Matrix size 1560 x 1920 pixel
Pixel pitch 148 μm
Data conversion 16 bit
Spatial resolution 3.4 lp/mm
DQE 66% typ.
Batteries charger Yes, up to two batteries included
Weight battery included 1.58 kg (2.2 lbs)
INSTALLATION DATA
System Power Supply
System power supply 400 V AC, Triphase
Voltage power supply tolerance ± 10%
Nominal line frequency 50/60 Hz
Line frequency tolerance ± 2%
Maximum absorbed power (Gen. 65kW) Apparent: 95 KVA; Active 65 KW
Maximum absorbed power (Gen. 80kW) Apparent: 115 kVA; Active 80 kW
Line protection device Magneto thermal differential 63A
Maximum power line resistance 0.1 ohm
Standby power 2 kVA
Weight And Dimension
Maximum table layout (w x h x d) 2,460 x 1,770 x 2,010 mm (97 x 70 x 79”)
Table weight 1,380 kg (3,042 lbs)
Product Data Sheet | 19
Control console weight 8 kg (18 lbs)
Collimator weight 14 kg (31 lbs)
Radiation source assembly weight 24 kg (53 lbs)
Generator cabinet 91 kg (200 lbs)
Monitor weight (with support) 6.6 kg (14.5 lbs)
Monitor dimensions (with support) 556.7 x 143.9-360.7 x 89-401.3 (22x5.7-14 x 3.5-15.8”)
Minimum exam room size for full functionality 5,200 x 4,850 mm controls area included (205 x 191”)
Enviromental Conditions In Use
Operating temperature From +15 to +35 °C
Humidity From 30 to 75 % not condensing
Product may not be available in all countries and regions.
Contact a GE Healthcare Representative for more information.