    gGE Healthcare

    TechnicalPublications

    Direction DOC1462293

    Revision 2

    LOGIQ F SERIES*

    DICOM CONFORMANCE STATEMENT

    Note: In the file, LOGIQ F SERIES means LOGIQ F3 & LOGIQ F5 & LOGIQ F6 & LOGIQ F8

    Do not duplicate

    Copyright 2013 by General Electric Co.

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    2/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    2

    THIS PAGE LEFT INTENTIONALLY BLANK

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    3/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    3

    CONFORMANCE STATEMENT OVERVIEW

    The LOGIQ F Series is a self-contained networked computer system used for acquiring ultrasound diagnostic medicalimages. The system implements the necessary DICOM services to download worklists from an information system, save

    acquired ultrasound images to a network storage device or media, print to a networked hardcopy device, query and moveultrasound images from a networked storage and inform the information system about the work actually done. The system

    conforms to the DICOM standard to allow the sharing of medical information with other digital imaging systems. LOGIQ F

    SeriesProduct is compliance to DICOM 3.0.

    Table 0.1 provides an overview of the network services supported by LOGIQ F Series.

    Table 0.1 NETWORK SERVICES

    SOP Classes User of Service

    (SCU)

    Provider of

    Service (SCP)

    Transfer

    Ultrasound Image Storage Yes Yes

    Ultrasound Image Storage (Retired) Yes Yes

    Ultrasound Multi-frame Image Storage Yes Yes

    Ultrasound Multi-frame Image Storage (Retired) Yes Yes

    Enhanced US Volume Storage Yes Yes

    Secondary Capture Image Storage Yes Yes

    Comprehensive SR Yes No

    Key Object Selection Document Yes No

    Verification SOP Class Yes Yes

    Query/Retrieve

    Study Root Query/Retrieve Information Model FIND Yes No

    Study Root Query/Retrieve Information Model MOVE Yes No

    Print Management

    Basic Grayscale Print Management Meta SOP Class Yes No

    Basic Color Print Management Meta SOP Class Yes No

    Basic Annotation Box Yes No

    Printer SOP Class Yes No

    Basic Film Session SOP Class Yes No

    Basic Film Box SOP Class Yes No

    Basic Grayscale Image Box SOP Class Yes No

    Basic Color Image Box SOP Class Yes No

    Workflow Management

    Storage Commitment Push Model SOP Class Yes No

    Modality Performed Procedure Step SOP Class Yes No

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    4/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    4

    Modality Worklist Information Model FIND SOP Class Yes No

    Table 0.2 provides an overview of the Media Storage Application Profilessupported by LOGIQ F Series.

    Table 0.2 - MEDIA SERVICES

    Media Storage Application Profile Write Files

    (FSC or FSU)

    Read Files

    (FSR)

    Compact Disk - Recordable

    General Purpose CD-R Yes Yes

    DVD

    General Purpose DVD Interchange with JPEG Yes Yes

    General Purpose DVD Interchange with JPEG 2000 Yes Yes

    USB

    General Purpose USB Media Interchange with JPEG Yes Yes

    General Purpose USB Media Interchange with JPEG 2000 Yes Yes

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    5/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    5

    TABLE OF CONTENTS

    1. INTRODUCTION ............................................................................................................... 9

    1.1 Overview ...................................................................................................................................................................... 9

    1.2 Overall DICOM Conformance Statement Document Structure .......................................................................... 11

    1.3 Intended Audience .................................................................................................................................................... 13

    1.4 Scope and Field of Application ................................................................................................................................ 13

    1.5 Important Remarks .................................................................................................................................................. 13

    1.6 References .................................................................................................................................................................. 14

    1.7 Definitions .................................................................................................................................................................. 14

    1.8 Symbols and Abbreviations ...................................................................................................................................... 15

    2. NETWORK CONFORMANCE STATEMENT.................................................................. 18

    2.1 Introduction ............................................................................................................................................................... 18

    2.2 Implementation Model ............................................................................................................................................. 18

    2.3 AE Specifications ....................................................................................................................................................... 21

    2.4 COMMUNICATION PROFILES ........................................................................................................................... 44

    3. MEDIA STORAGE CONFORMANCE STATEMENT ...................................................... 47

    3.4 AUGMENTED AND PRIVATE APPLICATION PROFILES ............................................................................ 51

    4. ULTRASOUND IMAGE INFORMATION OBJECT IMPLEMENTATION........................ 53

    4.1 Introduction ............................................................................................................................................................... 53

    4.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 53

    4.3 IOD Module Table .................................................................................................................................................... 53

    4.4 Information Module Definitions .............................................................................................................................. 54

    5. ULTRASOUND MULTI-FRAME IMAGE INFORMATION OBJECT IMPLEMENTATION66

    5.1 Introduction ............................................................................................................................................................... 66

    5.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 66

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    6/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    6

    5.3 IOD Module Table .................................................................................................................................................... 66

    5.4 Information Module Definitions .............................................................................................................................. 67

    6. ENHANCED US VOLUME INFORMATION OBJECT IMPLEMENTATION ................... 69

    6.1 Introduction ............................................................................................................................................................... 69

    6.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 69

    6.3 IOD Module Table .................................................................................................................................................... 69

    6.4 Information Module Definitions .............................................................................................................................. 70

    7. SECONDARY CAPTURE INFORMATION OBJECT IMPLEMENTATION .................... 80

    7.1 Introduction ............................................................................................................................................................... 80

    7.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 80

    7.3 IOD Module Table .................................................................................................................................................... 80

    7.4 Information Module Definitions .............................................................................................................................. 81

    8. COMPREHENSIVE STRUCTURED REPORT INFORMATION OBJECT

    IMPLEMENTATION .................................................................................................................. 83

    8.1 Introduction ............................................................................................................................................................... 83

    8.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 83

    8.3 IOD Module Table .................................................................................................................................................... 83

    8.4 Information Module Definitions .............................................................................................................................. 84

    8.5 Standard Extended and Private Data Attributes ................................................................................................... 89

    8.6 Standard Extended and Private Context Groups .................................................................................................. 89

    8.7 Standard, Standard Extended and Private Templates .......................................................................................... 89

    9. KEY OBJECT SELECTION DOCUMENT INFORMATION OBJECT IMPLEMENTATION91

    9.1 Introduction ............................................................................................................................................................... 91

    9.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 91

    9.3 IOD Module Table .................................................................................................................................................... 91

    9.4 Information Module Definitions .............................................................................................................................. 92

    9.5 Standard, Standard Extended and Private Templates .......................................................................................... 93

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    7/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    7

    10. MODALITY WORKLIST INFORMATION MODEL DEFINITION .................................... 94

    10.1 Introduction ............................................................................................................................................................... 94

    10.2 Modality Worklist Information Model Description .............................................................................................. 94

    10.3 Modality Worklist Information Model Entity-Relationship Model ..................................................................... 94

    10.4 Information Model Module Table ........................................................................................................................... 96

    10.5 Information Model Keys .......................................................................................................................................... 97

    11. MODALITY PERFORMED PROCEDURE STEP IMPLEMENTATION ........................ 103

    11.1 Introduction ............................................................................................................................................................. 103

    11.2 RELATIONSHIP BETWEEN SCHEDULED AND PERFORMED PROCEDURE STEPS ......................... 103

    11.3 MODALITY PERFORMED PROCEDURE STEP MODULE TABLE ........................................................... 103

    11.4 MODALITY PERFORMED PROCEDURE STEP MODULE DEFINITIONS .............................................. 104

    11.5 SOP Common Module ............................................................................................................................................ 104

    11.6 Performed Procedure Step Relationship Module ................................................................................................ 104

    11.7 Performed Procedure Step Information Module ................................................................................................. 106

    11.8 Image Acquisition Results Module ........................................................................................................................ 106

    12. STORAGE COMMITMENT PUSH MODEL IMPLEMENTATION ................................. 108

    12.1 Introduction ............................................................................................................................................................. 108

    12.2 Storage Commitment Push Model Sop Class Definition ..................................................................................... 108

    13. BASIC DIRECTORY INFORMATION OBJECT IMPLEMENTATION .......................... 112

    13.1 Introduction ............................................................................................................................................................. 112

    13.2 Basic Directory Iod Implementation ..................................................................................................................... 112

    13.3 Basic Directory Entity-Relationship Model .......................................................................................................... 112

    13.4 IOD Module Table .................................................................................................................................................. 113

    13.5 Information Module Definitions ............................................................................................................................ 114

    13.6 PRIVATE DATA DICTIONARY ......................................................................................................................... 121

    14. PRINT MANAGEMENT IMPLEMENTATION................................................................ 123

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    8/214

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    9/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    9

    1. INTRODUCTION

    1.1 OVERVIEW

    This DICOM Conformance Statement is divided into Sections as described below:

    Section 1 (Introduction),which describes the overall structure, intent, and references for this Conformance Statement

    Section 2 (Network Conformance Statement), which specifies the GEHC equipment compliance to the DICOM

    requirements for the implementation of Networking features.

    Section 3 (Media Storage Conformance Statement),which specifies the GEHC equipment compliance to the DICOM

    requirements for the implementation of Media Storage features.

    Section 4 (Ultrasound Image Information Object Implementation),which specifies the GEHC equipment compliance to

    DICOM requirements for the implementation of an UltrasoundImage Information Object.

    Section 5 (Ultrasound Multi-Frame Image Information Object Implementation),which specifies the GEHC equipmentcompliance to DICOM requirements for the implementation of an Ultrasound Multi-Frame Image Information.

    Section 6 (Enhanced US Volume Information Object Implementation),which specifies the GEHC equipment complianceto DICOM requirements for the implementation of an Enhanced US Volume Information.

    Section 7 (Secondary Capture Object Implementation), which specifies the GEHC equipment compliance to DICOMrequirements for the implementation of a Secondary Capture Information Object.

    Section 8 (Comprehensive Structured Report Object Implementation), which specifies the GEHC equipment compliance

    to DICOM requirements for the implementation of a Comprehensive Structured Reporting Information Object.

    Section 9 (Key Object Selection Document Implementation), which specifies the GEHC equipment compliance to

    DICOM requirements for the implementation of a Key Object Selection Document Object.

    Section 10 (Modality Worklist Information Model), which specifies the GEHC equipment compliance to DICOMrequirements for the implementation of the Modality Worklist service.

    Section 11 (Modality Performed Procedure Step Implementation),which specifies the GEHC equipment compliance toDICOM requirements for the implementation of Modality Performed Procedure Step Service.

    Section 12 (Storage Commitment PushModel SOP Implementation),which specifies the GEHC equipment complianceto DICOM requirements for the implementation of the Storage CommitmentPush Model Service.

    Section 13 (Basic Directory Information Object Implementation), which specifies the GEHC equipment compliance toDICOM requirements for the implementation of a Basic Directory Information Object.

    Section 14 (Print Management), which specifies the GEHC equipment compliance to DICOM requirements for the

    implementation of Basic Print Meta SOP Classes (Gray and Color).

    Section 15 (Study Root Query/Retrieve Information Model), which specifies the GEHC equipment compliance to

    DICOM requirements for the Study Root Query/Retrieve Information Model.

    Section 16 (DICOM SR Templates),which specifies all LOGIQ F Seriessupported DICOM SR Templates.

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    10/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    10

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    11/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    11

    1.2 OVERALL DICOM CONFORMANCE STATEMENT DOCUMENT STRUCTURE

    The Documentation Structure of the GEHC DICOM Conformance Statements is shown in the Illustration below.

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    12/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    12

    This document specifies the DICOM implementation. It is entitled:

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    13/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    13

    LOGIQ F Series

    Conformance Statement for DICOM

    Direction DOC1462293

    This DICOM Conformance Statement documents the DICOM Conformance Statement and Technical Specification requiredto interoperate with the GEHC network interface.

    The GEHC Conformance Statement, contained in this document, also specifies the Lower Layer communications which itsupports (e.g., TCP/IP). However, the Technical Specifications are defined in the DICOM Part 8 standard.

    For more information regarding DICOM, copies of the Standard may be obtained on the Internet at http://medical.nema.org.Comments on the Standard may be addressed to:

    DICOM SecretariatNEMA

    1300 N. 17thStreet, Suite 1752

    Rosslyn, VA 22209

    USA

    Phone: +1.703.841.3200

    1.3 INTENDED AUDIENCE

    The reader of this document is concerned with software design and/or system integration issues. It is assumed that the

    reader of this document is familiar with the DICOM Standard and with the terminology and concepts which are used in thatStandard.

    1.4 SCOPE AND FIELD OF APPLICATION

    It is the intent of this document to provide an unambiguous specification for GEHC implementations. This specification,

    called a Conformance Statement, includes a DICOM Conformance Statement and is necessary to ensure proper processing

    and interpretation of GEHC medical data exchanged using DICOM. The GEHC Conformance Statements are available to

    the public.

    The reader of this DICOM Conformance Statement should be aware that different GEHC devices are capable of usingdifferent Information Object Definitions. For example, a GEHC CT Scanner may send images using the CT InformationObject, MR Information Object, Secondary Capture Object, etc.

    Included in this DICOM Conformance Statement are the Module Definitions which define all data elements used by thisGEHC implementation. If the user encounters unspecified private data elements while parsing a GEHC Data Set, the user is

    well advised to ignore those data elements (per the DICOM standard). Unspecified private data element information is

    subject to change without notice. If, however, the device is acting as a "full fidelity storage device", it should retain and re-transmit all of the private data elements which are sent by GEHC devices.

    1.5 IMPORTANT REMARKS

    The use of these DICOM Conformance Statements, in conjunction with the DICOM Standards, is intended to facilitatecommunication with GE imaging equipment. However, by itself, it is not sufficient to ensure that inter-operation will be

    successful. The user (or user's agent)needs to proceed with caution and address at least four issues:

    Integration - The integration of any device into an overall system of interconnected devices goes beyond the scope of

    standards (DICOM v3.0), and of this introduction and associated DICOM Conformance Statements when interoperability

    with non-GE equipment is desired. The responsibility to analyze the applications requirements and to design a solution thatintegrates GE imaging equipment with nonGE systems is the user'sresponsibility and should not be underestimated. The

    useris strongly advised to ensure that such an integration analysis is correctly performed.

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    14/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    14

    Validation - Testing the complete range of possible interactions between any GE device and nonGE devices, before theconnection is declared operational, should not be overlooked. Therefore, the usershould ensure that any nonGE provider

    accepts full responsibility for all validation required for their connection with GE devices. This includes the accuracy of the

    image data once it has crossed the interface between the GE imaging equipment and the nonGE device and the stability ofthe image data for the intended applications.

    Such a validation is required before any clinical use (diagnosis and/or treatment) is performed. It applies when imagesacquired on GE imaging equipment are processed/displayed on a non-GE device, as well as when images acquired on non-

    GE equipment is processed/displayed on a GE console or workstation.

    Future Evolution- GE understands that the DICOM Standard will evolve to meet the user's growing requirements. GE is

    actively involved in the development of the DICOM Standard. DICOM will incorporate new features and technologies andGE may follow the evolution of the Standard. The GEHC protocol is based on DICOM as specified in each DICOM

    Conformance Statement. Evolution of the Standard may require changes to devices which have implemented DICOM. In

    addition, GE reserves the right to discontinue or make changes to the support of communications features (on its

    products) described by these DICOM Conformance Statements. The user should ensure that any nonGE provider,

    which connects with GE devices, also plans for the future evolution of the DICOM Standard. Failure to do so will likely

    result in the loss of function and/or connectivity as the DICOM Standard changes and GE Products are enhanced to support

    these changes.

    Interaction - It is the sole responsibility of the nonGE provider to ensure that communication with the interfacedequipment does not cause degradation of GE imaging equipment performance and/or function.

    1.6 REFERENCES

    NEMA PS3 Digital Imaging and Communications in Medicine (DICOM) Standard, available free at

    http://medical.nema.org/

    1.7 DEFINITIONS

    Informal definitions are provided for the following terms used in this Conformance Statement. The DICOM Standard is the

    authoritative source for formal definitions of these terms.

    Abstract Syntax the information agreed to be exchanged between applications, generally equivalent to a Service/Object

    Pair (SOP) Class. Examples : Verification SOP Class, Modality Worklist Information Model Find SOP Class, ComputedRadiography Image Storage SOP Class.

    Application Entity (AE) an end point of a DICOM information exchange, including the DICOM network or mediainterface software; i.e., the software that sends or receives DICOM information objects or messages. A single device may

    have multiple Application Entities.

    Application Entity Title the externally known name of an Application Entity, used to identify a DICOM application to

    other DICOM applications on the network.

    Application Context the specification of the type of communication used betweenApplication Entities. Example: DICOMnetwork protocol.

    Association a network communication channel set up betweenApplication Entities.

    Attribute a unit of information in an object definition; a data element identified by a tag. The information may be acomplex data structure (Sequence), itself composed of lower level data elements. Examples: Patient ID (0010,0020),

    Accession Number (0008,0050), Photometric Interpretation (0028,0004), Procedure Code Sequence (0008,1032).

    Information Object Definition (IOD) the specified set ofAttributesthat comprise a type of data object; does not representa specific instance of the data object, but rather a class of similar data objects that have the same properties. The Attributes

    may be specified as Mandatory (Type 1), Required but possibly unknown (Type 2), or Optional (Type 3), and there may be

    conditions associated with the use of an Attribute (Types 1C and 2C). Examples: MR Image IOD, CT Image IOD, Print Job

    IOD.

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    15/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    15

    Joint Photographic Experts Group (JPEG) a set of standardized image compression techniques, available for use byDICOM applications.

    Media Application Profile the specification of DICOM information objects and encoding exchanged on removable media

    (e.g., CDs)

    Module a set ofAttributeswithin anInformation Object Definitionthat are logically related to each other. Example: PatientModule includes Patient Name, Patient ID, Patient Birth Date, and Patient Sex.

    Negotiation first phase of Associationestablishment that allows Application Entities to agree on the types of data to beexchanged and how that data will be encoded.

    Presentation Context the set of DICOM network services used over an Association, as negotiated between ApplicationEntities; includesAbstract Syntaxes and Transfer Syntaxes.

    Protocol Data Unit (PDU) a packet (piece) of a DICOM message sent across the network. Devices must specify the

    maximum size packet they can receive for DICOM messages.

    Security Profile a set of mechanisms, such as encryption, user authentication, or digital signatures, used by an Application

    Entityto ensure confidentiality, integrity, and/or availability of exchanged DICOM data

    Service Class Provider (SCP) role of an Application Entitythat provides a DICOM network service; typically, a server

    that performs operations requested by another Application Entity (Service Class User). Examples: Picture Archiving andCommunication System (image storage SCP, and image query/retrieve SCP), Radiology Information System (modality

    worklist SCP).

    Service Class User (SCU) role of anApplication Entitythat uses a DICOM network service; typically, a client. Examples:imaging modality (image storage SCU, and modality worklist SCU), imaging workstation (image query/retrieve SCU)

    Service/Object Pair (SOP) Class the specification of the network or media transfer (service) of a particular type of data(object); the fundamental unit of DICOM interoperability specification. Examples: Ultrasound Image Storage Service, Basic

    Grayscale Print Management.

    Service/Object Pair (SOP) Instance an information object; a specific occurrence of information exchanged in a SOPClass. Examples: a specific x-ray image.

    Tag a 32-bit identifier for a data element, represented as a pair of four digit hexadecimal numbers, the group and the

    element. If the group number is odd, the tag is for a private (manufacturer-specific) data element. Examples: (0010,0020)[Patient ID], (07FE,0010) [Pixel Data], (0019,0210) [private data element]

    Transfer Syntax the encoding used for exchange of DICOM information objects and messages. Examples: JPEGcompressed (images), little endian explicit value representation.

    Unique Identifier (UID) a globally unique dotted decimal string that identifies a specific object or a class of objects; anISO-8824 Object Identifier. Examples: Study Instance UID, SOP Class UID, SOP Instance UID.

    Value Representation (VR) the format type of an individual DICOM data element, such as text, an integer, a personsname, or a code. DICOM information objects can be transmitted with either explicit identification of the type of each data

    element (Explicit VR), or without explicit identification (Implicit VR); with Implicit VR, the receiving application must use a

    DICOM data dictionary to look up the format of each data element.

    1.8 SYMBOLS AND ABBREVIATIONS

    AE Application Entity

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    16/214

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    17/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    17

    PDU Protocol Data Unit

    R Required (Key Attribute)

    RIS Radiology Information System

    SC Secondary Capture

    SCP Service Class Provider

    SCU Service Class User

    SOP Service-Object Pair

    SPS Scheduled Procedure Step

    SR Structured Reporting

    TCP/IP Transmission Control Protocol/Internet Protocol

    U Unique (Key Attribute)

    UL Upper Layer

    US Ultrasound

    VR Value Representation

    KOS Key Object Selection

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    18/214

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    19/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    19

    ILLUSTRATION 21

    LOGIQFSERIESNETWORK APPLICATION MODEL AND DATA FLOW DIAGRAM

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    20/214

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    21/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    21

    There are six local real-world activities that occur in LOGIQ F Series- Exam Save, Echo, Worklist Query, Image

    Store/End Exam,Image Printand Exam Query/Retrieve

    Exam save initiates a connection with the DICOM SCP and transmits images and results to the DICOM SCP. If Storage

    Commitment is configured a commitment request will be sent for the images and results.

    Echoinitiates a connection with the DICOM SCP, posts a Verification request and closes the connection. It also responds toincoming Verification requests (for service use).

    Worklist Query initiates a connection with the DICOM SCP, performs a query and retrieves the matching entries to theproduct.

    Image Store/End exam: If Modality Performed Procedure Step is configured N-CREATE and N-SET messages will be sentfor the exam.

    Image Printwill send images to a DICOM Print SCP.

    Exam Query/Retrieveinitiates a connection with the DICOM SCP, performs a query and retrieves selected examination.

    2.2.2 Functional Definition of AE's

    Application Entity LOGIQ F Seriessupports the following functions:

    Initiates a DICOM association to send images and results.

    Initiates a DICOM verification to assist in network diagnostics.

    Initiates a DICOM worklist query to receive worklist information.

    Initiates a DICOM association to notify start of examination.

    Initiates a DICOM association to notify end of examination.

    Initiates a DICOM association to request storage commitment of images.

    Responds to replies for storage commitment requests of images.

    Initiates a DICOM association to print images.

    Initiates a DICOM association to query for and retrieve examinations.

    Responds to storage requests for images triggered by examination retrieve requests.

    2.2.3 Sequencing of Real-World Activities

    Non Applicable.

    2.3 AE SPECIFICATIONS

    2.3.1 LOGIQ F SeriesAE Specification

    This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an SCU and/or as anSCP:

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    22/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    22

    SOP Class Name SOP Class UID SCU SCP

    Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 Yes Yes

    Ultrasound Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.6 Yes Yes

    Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 Yes Yes

    Ultrasound Multi-frame Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.3 Yes Yes

    Enhanced US Volume Storage 1.2.840.10008.5.1.4.1.1.6.2 Yes Yes

    Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 Yes Yes

    Comprehensive SR 1.2.840.10008.5.1.4.1.1.88.33 Yes No

    Key Object Selection Document Storage 1.2.840.10008.5.1.4.1.1.88.59 Yes No

    Verification SOP Class 1.2.840.10008.1.1 Yes Yes

    Study Root Query/Retrieve Information Model - FIND 1.2.840.10008.5.1.4.1.2.2.1 Yes No

    Study Root Query/Retrieve Information Model - MOVE 1.2.840.10008.5.1.4.1.2.2.2 Yes No

    Modality Worklist Information Model - FIND 1.2.840.10008.5.1.4.31 Yes No

    Modality Performed Procedure Step 1.2.840.10008.3.1.2.3.3 Yes No

    Storage Commitment Push Model 1.2.840.10008.1.20.1 Yes No

    Basic Grayscale Print Management Meta SOP Class 1.2.840.10008.5.1.1.9 Yes No

    Basic Color Print Management Meta SOP Class 1.2.840.10008.5.1.1.18 Yes No

    Basic Annotation Box 1.2.840.10008.5.1.1.15 Yes No

    Printer SOP Class 1.2.840.10008.5.1.1.16 Yes No

    Basic Film Session SOP Class 1.2.840.10008.5.1.1.1 Yes No

    Basic Film Box SOP Class 1.2.840.10008.5.1.1.2 Yes No

    Basic Grayscale Image Box SOP Class 1.2.840.10008.5.1.1.4 Yes No

    Basic Color Image Box SOP Class 1.2.840.10008.5.1.1.4.1 Yes No

    2.3.1.1 Association Establishment Policies

    2.3.1.1.1 General

    The DICOM Application Context Name (ACN), which is always proposed, is:

    Application Context Name 1.2.840.10008.3.1.1.1

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    23/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    23

    The Maximum Length PDU negotiation is included in all association establishment requests. The maximum length PDUfor an association initiated by LOGIQ F Seriesis:

    Maximum Length PDU 32768

    The maximum length PDU is a fixed number - not configurable.

    The SOP Class Extended Negotiation is not supported.

    The user information Items sent by this product are:

    Maximum PDU Length

    Implementation UID

    Implementation Version Name

    2.3.1.1.2 Number of Associations

    The LOGIQ F SeriesAE will initiate multiple DICOM associations. The maximum numbers of associations are based on theconnectivity service configuration.

    2.3.1.1.3 Asynchronous Nature

    Asynchronous mode is not supported. All operations will be performed synchronously.

    2.3.1.1.4 Implementation Identifying Information

    The Implementation UID for this DICOM Implementation is:

    LOGIQ F SeriesImplementation UID 1.2.840.113619.6.384

    LOGIQ F SeriesImplementation Version Name LOGIQ F Series

    2.3.1.2 Association Initiation Policy

    The LOGIQ F SeriesAE attempts to establish a new association with a remote device due to six Real-World Activities:

    Exam save initiated by the operator for images, key object selection document and structured reports and sending request forStorage Commitment.

    Verification, which verifies application level communication between peer DICOM AEs for service purposes.

    Worklist initiated by the operator for receiving worklist information.Image Store/End Exam sending messages to Modality Performed Procedure Step.

    Print initiated by the operator for a specific image or group of images.

    Exam Query/Retrieve initiated by the operator for receiving examination information and selecting examination to retrieve.

    2.3.1.2.1 Real-World Activity - Exam save Operation

    2.3.1.2.1.1 Associated Real-World Activity

    Upon a request by the operator (manual or automatic), images will be sent to a DICOM Storage SCP.

    2.3.1.2.1.2 Proposed Presentation Context Tables

    The Proposed Presentation Context Table depends on compression according to the following table:

    Presentation Context Table Proposed

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    24/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    24

    Abstract Syntax Transfer Syntax Role ExtendedNegotiation

    Name UID Name List UID List

    Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to None

    Ultrasound Image

    Storage

    1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little Endian

    Explicit VR Big EndianImplicit VR Little Endian

    1.2.840.10008.1.2.1

    1.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound Multi-

    frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 Explicit VR Little Endian

    Explicit VR Big EndianImplicit VR Little Endian

    1.2.840.10008.1.2.1

    1.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Enhanced US Volume

    Storage

    1.2.840.10008.5.1.4.1.1.6.2 Explicit VR Little Endian

    Explicit VR Big EndianImplicit VR Little Endian

    1.2.840.10008.1.2.1

    1.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Secondary CaptureImage Storage

    1.2.840.10008.5.1.4.1.1.7 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Presentation Context Table: Still Compression set to RLE, Clip/Volume Compression set to RLE Secondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound Image

    Storage

    1.2.840.10008.5.1.4.1.1.6.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Presentation Context Table: Still Compression set to JPEG Lossy, Clip/Volume Compression set to

    JPEG LossySecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Presentation Context Table: Still Compression set to JPEG Lossless, Clip/Volume Compression set to

    JPEG LosslessSecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Lossless Non

    Hierarchical (Proc 14)1.2.840.10008.1.2.4.70 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound Image 1.2.840.10008.5.1.4.1.1.6 JPEG Lossless Non 1.2.840.10008.1.2.4.70 SCU None

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    25/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    25

    Storage (retired) Hierarchical (Proc 14)

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound Multi-

    frame Image Storage(retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG Lossless Non

    Hierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Presentation Context Table: Still Compression set to JPEG 2000 Lossy, Clip/Volume Compression

    set to JPEG 2000 LossySecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression

    set to JPEG 2000 LosslessSecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to RLE Secondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian

    Explicit VR Big EndianImplicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.2

    1.2.840.10008.1.2

    SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.2

    1.2.840.10008.1.2

    SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to JPEG

    LossySecondary Capture 1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    26/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    26

    Image Storage Explicit VR Big EndianImplicit VR Little Endian

    1.2.840.10008.1.2.21.2.840.10008.1.2

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.2

    1.2.840.10008.1.2

    SCU None

    Ultrasound Image

    Storage (retired)

    1.2.840.10008.5.1.4.1.1.6 Explicit VR Little Endian

    Explicit VR Big EndianImplicit VR Little Endian

    1.2.840.10008.1.2.1

    1.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to JPEG

    LosslessSecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian

    Explicit VR Big EndianImplicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.2

    1.2.840.10008.1.2

    SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to JPEG

    2000 LossySecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian

    Explicit VR Big EndianImplicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to JPEG

    2000 LosslessSecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian

    Explicit VR Big EndianImplicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    27/214

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    28/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    28

    Secondary CaptureImage Storage

    1.2.840.10008.5.1.4.1.1.7 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Enhanced US Volume

    Storage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Presentation Context Table: Still Compression set to RLE, Clip/Volume Compression set to JPEG

    2000 LosslessSecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Presentation Context Table: Still Compression set to JPEG Lossy, Clip/Volume Compression set to

    NoneSecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound Image

    Storage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Presentation Context Table: Still Compression set to JPEG Lossy, Clip/Volume Compression set to

    RLESecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    29/214

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    30/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    30

    Storage Explicit VR Big EndianImplicit VR Little Endian

    1.2.840.10008.1.2.21.2.840.10008.1.2

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.2

    1.2.840.10008.1.2

    SCU None

    Presentation Context Table: Still Compression set to JPEG Lossless, Clip/Volume Compression set to

    RLESecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Lossless Non

    Hierarchical (Proc 14)1.2.840.10008.1.2.4.70 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Presentation Context Table: Still Compression set to JPEG Lossless, Clip/Volume Compression set to

    JPEG LossySecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Lossless Non

    Hierarchical (Proc 14)1.2.840.10008.1.2.4.70 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound Multi-

    frame Image Storage(retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Presentation Context Table: Still Compression set to JPEG Lossless, Clip/Volume Compression set to

    JPEG 2000 LossySecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Lossless Non

    Hierarchical (Proc 14)1.2.840.10008.1.2.4.70 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Presentation Context Table: Still Compression set to JPEG Lossless, Clip/Volume Compression set to

    JPEG 2000 LosslessSecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Lossless Non

    Hierarchical (Proc 14)1.2.840.10008.1.2.4.70 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound Image

    Storage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG Lossless Non

    Hierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    31/214

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    32/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    32

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound Multi-

    frame Image Storage(retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG Lossless Non

    Hierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Presentation Context Table: Still Compression set to JPEG 2000 Lossy, Clip/Volume Compression

    set to JPEG 2000 LosslessSecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression

    set to NoneSecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Image

    Storage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 Explicit VR Little EndianExplicit VR Big Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCU None

    Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression

    set to RLESecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 Run Length Encoding 1.2.840.10008.1.2.5 SCU None

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    33/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    33

    Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression

    set to JPEG LossySecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Image

    Storage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None

    Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression

    set to JPEG LosslessSecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Multi-frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG Lossless NonHierarchical (Proc 14)

    1.2.840.10008.1.2.4.70 SCU None

    Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression

    set to JPEG 2000 LossySecondary Capture

    Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None

    Ultrasound Multi-

    frame Image Storage

    1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Ultrasound Multi-frame Image Storage

    (retired)

    1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None

    Presentation Context Table for Structured Reports

    ComprehensiveStructured Report 1.2.840.10008.5.1.4.1.1.88.33 Explicit VR Little EndianImplicit VR Little Endian 1.2.840.10008.1.2.11.2.840.10008.1.2 SCU None

    Presentation Context Table for Key Object Selection Document Key Object Selection

    Document1.2.840.10008.5.1.4.1.1.88.59 Explicit VR Little Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.1

    1.2.840.10008.1.2

    SCU None

    This operation also sends a Storage Commitment Request, with the following proposed presentation context. The result fromthe SCP is expected on another association for the Storage Commitment result.

    Presentation Context Table Proposed

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    34/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    34

    Presentation Context Table Proposed

    Abstract Syntax Transfer Syntax Role Extended

    NegotiationName UID Name List UID List

    Storage

    Commitment PushModel SOP Class

    1.2.840.10008.1.20.1 Explicit VR Little Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.1

    1.2.840.10008.1.2 SCU None

    2.3.1.2.1.2.1 SOP Specific Conformance Statement for Image Storage SOP Classes

    For these SOP class all status codes with status Refused or Error are treated as failures and terminate the association and

    operation. On a failure, the request will be put in a holding queue for the user to manually retry the request. All status codes

    with status Warning or Success are treated as successes.

    Service

    Status

    Further Meaning Status

    Codes

    Application

    Behavior When

    receiving Status

    Codes

    Failure Refused: Out of Resources A7xx Terminateassociation. Mark the

    job failed.Error: Data Set does not match SOPClass

    A9xx

    Error: Cannot understand Cxxx

    Warning Coercion of Data Elements B000 Treat the job success.

    Data Set does not match SOP Class B007

    Elements Discarded B006

    Success 0000

    * Any other status code * Terminate

    association. Mark thejob failed.

    C-STORE SCU operation supports an Association Timer and Operation Inactivity Timer with default time out value of

    30 seconds each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1, 2, 3, 4, 5, 10,20, 30, 60, 120 and 360 seconds.

    2.3.1.2.2 Real-World Activity - Echo Operation

    2.3.1.2.2.1 Associated Real-World Activity

    The user may initiate a DICOM Verification Request in the config screen.Associations will be released upon the receipt of each C-ECHO confirmation.

    2.3.1.2.2.2 Proposed Presentation Context Table

    Presentation Context Table - Proposed

    Abstract Syntax Transfer Syntax Role Extended

    NegotiationName UID Name List UID List

    Verification SOP Class 1.2.840.10008.1.1Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    35/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    35

    Implicit VR Little Endian 1.2.840.10008.1.2

    2.3.1.2.2.2.1 SOP Specific DICOM Conformance Statement for Verify SOP Class

    The AE provides standard conformance to the Verification SOP Class as an SCU. In the event that the SCP does not respond

    for some reason, the operation will time out and LOGIQ F Serieswill close the association.

    C-ECHO SCU operation supports an Association Timer and Operation Inactivity Timer with default time out value of 30

    seconds and 10secs each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1 to 10

    seconds.

    2.3.1.2.3 Real-World Activity - Worklist Query Operation

    2.3.1.2.3.1 Associated Real-World Activity

    The user may initiate a DICOM Worklist Query in Search screen, which will send a C-FIND-RQ to the Worklist SCP.

    Associations will be released upon the receipt of C-FIND-RSP confirmation.

    2.3.1.2.3.2 Proposed Presentation Context Tables

    Presentation Context Table Proposed

    Abstract Syntax Transfer Syntax Role Extended

    NegotiationName UID Name List UID List

    Modality Worklist InformationModel FIND

    1.2.840.10008.5.1.4.31Explicit VR Little EndianImplicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2

    SCU None

    2.3.1.2.3.2.1 SOP Specific DICOM Conformance Statement for Worklist SOP Classes

    The LOGIQ F Seriesincludes matching keys in the Modality Worklist queries as described in Section 10.All status codes with status Refused or Error are treated as failures and terminate the association and operation. On a failure,

    the user will be informed and the last successful query will be used as Worklist. All status codes with status Warning orSuccess are treated as successes. Application doesnt support C-FIND-CANCEL request.

    Service Status Status Code Further Meaning Application Behavior

    When receiving

    Status Codes

    Refused A700 Out of resources Terminate theassociation andoperation

    0122 SOP Class not

    Supported

    Failed A900 Identifier does notmatch SOP Class

    Cxxx Unable to process

    Success 0000 Matching is complete

    No final identifier is

    supplied

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    36/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    36

    Pending FF00 Matches are

    continuing Current

    Match is supplied and

    any Optional Keys

    were supported in thesame manner

    Receiving process of

    the matches continues.

    FF01 Matches are

    continuing Warning

    that one or more

    Optional Keys were

    not supported for

    existence for this

    Identifier

    * * Any other status code. Terminate the

    association and

    operation

    Worklist operation supports an Association Timer and Operation Inactivity Timer with default time out value of 30

    seconds each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1, 2, 3, 4, 5, 10,

    20, 30, 60, 120 and 360 seconds.

    2.3.1.2.4 Real-World Activity - Image Store/End exam Operation

    2.3.1.2.4.1 Associated Real-World Activity

    The Modality Performed Procedure Step messages are sent when the first image is acquired for the start of an exam and when

    the exam is ended (for the case where there are no images, the N-CREATE is sent when the exam is ended). For an exam

    with saved images or results, the N-SET will be sent with status COMPLETED. For an exam without saved images or

    results, the N-SET will be sent with status DISCONTINUED.

    2.3.1.2.4.2 Proposed Presentation Context Table

    Presentation Context Table Proposed

    Abstract Syntax Transfer Syntax Role Extended

    Name UID Name List UID List Negotiation

    Modality Performed Procedure

    Step SOP Class1.2.840.10008.3.1.2.3.3

    Explicit VR Little Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.1

    1.2.840.10008.1.2

    SCU None

    2.3.1.2.4.2.1 SOP Specific DICOM Conformance Statement for Modality Performed Procedure Step Class

    LOGIQ F Seriesincludes attributes in the Modality Performed Procedure Step N-CREATE and N-SET as described in

    Section 10.2.

    LOGIQ F Seriesincludes attributes in the Modality Performed Procedure Step N-SET as described in Section 11. The

    mapping from Worklist attributes is described in Section 10.5. LOGIQ F Seriessends N-SET after the exam is ended. The N-

    SET will include all acquired images UIDs and the status of COMPLETED or DISCONTINUED.

    For this SOP class, all status codes with status Refused or Error are treated as failures and terminate the association and

    operation. All status codes with status Warning or Success are treated as successes.

    Modality Performed Procedure Step N-CREATE response status Handling Behavior

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    37/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    37

    Service Status Status Code Further Meaning Application Behavior

    When receiving

    Status Codes

    Success 0000 Success

    * * Any other status Treat as failure.

    Aborts the association.

    Modality Performed Procedure Step N-SET response status Handling Behavior

    Service Status Status Code Further Meaning Application Behavior

    When receiving

    Status Codes

    Failure 0110H Processing Failure.

    Performed ProcedureStep Object may no

    longer be updated

    Aborts the operation.

    Success 0000 Success Completes the

    operation.

    * * Any other status Aborts the operation.

    MPPS operation supports an Association Timer and Operation Inactivity Timer with default time out value of 30 seconds

    each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1, 2, 3, 4, 5, 10, 20, 30, 60,120 and 360 seconds.

    2.3.1.2.5 Real-World Activity - Image Print Operation

    2.3.1.2.5.1 Associated Real-World Activity

    Upon a request by the operator, print jobs will be sent to a DICOM Print SCP. If an error occurs during the transmission, thecurrent association is released and a new association initiated. The maximum number of retries is configurable.

    2.3.1.2.5.2 Proposed Presentation Context Tables

    The following table is used:

    Presentation Context Table - Proposed

    Abstract Syntax Transfer Syntax Role Extended

    Name UID Name List UID List Negotiation

    Basic Grayscale PrintManagement Meta SOP Class

    1.2.840.10008.5.1.1.9Explicit VR Little EndianImplicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2

    SCU None

    Basic Color Print ManagementMeta SOP Class

    1.2.840.10008.5.1.1.18 Explicit VR Little EndianImplicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2

    SCU None

    Basic Annotation Box SOPClass

    1.2.840.10008.5.1.1.15Explicit VR Little EndianImplicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2

    SCU None

    2.3.1.2.5.2.1 SOP Specific DICOM Conformance Statement for all Print Management SOP Classes

    All status codes with status Refused or Error are treated as failures and terminate the association and operation. All status

    codes with status Warning or Success are treated as successes.

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    38/214

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    39/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    39

    2.3.1.2.6.2 Proposed Presentation Context Tables

    Presentation Context Table Proposed

    Abstract Syntax Transfer Syntax Role Extended

    NegotiationName UID Name List UID List

    Study Root Query/RetrieveInformation Model - FIND

    1.2.840.10008.5.1.4.1.2.2.1 Explicit VR Little EndianImplicit VR Little Endian

    1.2.840.10008.1.2.11.2.840.10008.1.2

    SCU None

    Study Root Query/Retrieve

    Information Model - MOVE

    1.2.840.10008.5.1.4.1.2.2.2Explicit VR Little Endian

    Implicit VR Little Endian

    1.2.840.10008.1.2.1

    1.2.840.10008.1.2

    SCU None

    2.3.1.2.6.2.1 SOP Specific DICOM Conformance Statement for Study Root Query/Retrieve Information Model -

    FIND SOP Classes

    Only a single information model, Study Root, is supported.

    All queries are initiated at the highest level of the information model (the STUDY level), and then for each response

    received, recursively repeated at the next lower levels (the SERIES and then IMAGE levels).

    C-FIND-CANCEL request is not supported.

    All status codes with status Refused or Error are treated as failures and terminate the association and operation. All status

    codes with status Warning or Success are treated as successes. LOGIQ F Serieswill only support hierarchical query.

    Service

    Status

    Status

    Code

    Further Meaning Application Behavior When Receiving Status

    Code

    Failure A700 Refused: Out of resources Operation aborted.

    A900 Error: Identifier does not matchSOP Class

    C000-

    CFFF

    Error: Unable to process

    0122 SOP Class Not Supported

    Cancel FE00 Matching terminated due to

    cancel

    Display the results in UI.

    Success 0000 Matching is complete - No finalidentifier is supplied

    Display the results in UI.

    Pending FF00 Matches are continuing -

    Current Match is supplied andany Optional Keys were

    supported in the same manner

    as Required Keys.

    Stores the results internally and waits for the more

    results.

    FF01 Matches are continuing -Warning that one or more

    Optional Keys were not

    supported for existence and/or

    matching for this Identifier

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    40/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    40

    * * Any other status code. Operation aborted.

    C-FIND SCU operation supports an Association Timer and Operation Inactivity Timer with default time out value of 30

    seconds each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1, 2, 3, 4, 5, 10,20, 30, 60, 120 and 360 seconds.

    LOGIQ F Seriesquery responses for Key Object Selection documents and Structured Report documents.

    Study Level attributes for the Study Root Query/Retrieve Information Model is described in Section 15.2.

    Types of Matching:

    " Single Value matching (S)

    " Universal Matching (U)

    " Wild Card Matching (*)

    " Range of date, Range of Time (R)

    The types of Matching supported by the C-FIND SCU are: S indicates the identifier attribute uses Single Value Matching,

    an R indicates Range Matching, a * indicates wildcard matching, a U indicates Universal Matching, and UNIQUE

    indicates that this is the Unique Key for that query level, in which case Universal Matching or Single Value Matching is used

    depending on the query level.

    Filtering is supported means that matching strings can be controlled from the Search screen at Data Transfer. All other

    matching fields can be configured in Config screen to be either enabled, enabled with a matching string value or disabled.

    The constant value will be used as entered by user.

    2.3.1.2.6.2.2 SOP Specific DICOM Conformance Statement for Study Root Query/Retrieve Information Model -

    MOVE SOP Classes

    All status codes with status Refused or Error are treated as failures and terminate the association and operation. All status

    codes with status Warning or Success are treated as successes.

    C-MOVE-CANCEL request is not supported.

    Service

    Status

    Status

    Code

    Further Meaning Application Behavior When Receiving Status

    Code

    Failure A701 Refused: Out of resources -Unable to calculate number of

    matches

    Retrieve operation will be aborted.

    A702 Refused: Out of resources -Unable to perform sub-

    operations

    A801 Refused: Move Destination

    Unknown

    A900 Error: Identifier does not match

    SOP Class

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    41/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    41

    C000-CFFF

    Error: Unable to process

    0122 SOP Class Not Supported

    Cancel FE00 Sub-operations terminated dueto a Cancel indication

    Warning B000 Sub-operations Complete - Oneor more Failures.

    Imports selected patient(s).

    Success 0000 Sub-operations Complete - NoFailure.

    Pending FF00 Sub-operations are continuing - Stores the patients internally until final response.

    * * Any other status code. Retrieve operation will be aborted.

    C-MOVE SCU operation supports an Association Timer and Operation Inactivity Timer with default time out value of

    30 seconds each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1, 2, 3, 4, 5, 10,

    20, 30, 60, 120 and 360 seconds.

    2.3.1.3 Association Acceptance Policy

    The AE accepts an association when it receives a Verification Request from another network device or a store request from a

    Q/R SCP or a Storage Commitment result from a Storage Commitment SCP.

    2.3.1.3.1 Real-World Activity - Echo operation

    2.3.1.3.1.1 Associated Real-World Activity

    An incoming Verification Request will cause the AE to accept the association and respond with a Verification Response.

    2.3.1.3.1.2 Accepted Presentation Context Table

    Presentation Context Table - Accepted

    Abstract Syntax Transfer Syntax Role Extended

    Name UID Name List UID List Negotiation

    Verification SOP Class 1.2.840.10008.1.1Explicit VR Little

    Endian Implicit VR

    Little Endian

    1.2.840.10008.1.2.1

    1.2.840.10008.1.2

    SCP None

    2.3.1.3.1.2.1 SOP Specific DICOM Conformance Statement for Verify SOP Class

    The AE provides standard conformance to the Verification SOP Class as an SCP. The port number used is configured on

    config screen, default is 104.

    2.3.1.3.1.3 Presentation Context Acceptance Criterion

    No criterion.

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    42/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    42

    2.3.1.3.1.4 Transfer Syntax Selection Policies

    The selected transfer syntax is based on the proposed transfer syntax list. The priority order is Explicit VR Little Endian andImplicit VR Little Endian.

    2.3.1.3.2 Real-World Activity - Exam Query/Retrieve Operation

    2.3.1.3.2.1 Associated Real-World Activity

    If the user has initiated a retrieve by a C-MOVE-RQ, the AE will accept associations for C-STORE-RQs. The images will bestored locally.

    2.3.1.3.2.2 Accepted Presentation Context Table

    Abstract Syntax Transfer Syntax Role Extended

    Negotiation

    Name UID Name List UID List

    Ultrasound ImageStorage

    1.2.840.10008.5.1.4.1.1.6.1JPEG 2000 LosslessJPEG 2000 LossyJPEG Baseline

    JPEG Lossless NonHierarchical (Proc 14)Run Length Encoding

    Exp VR Little EndianExplicit VR Big EndianImp VR Little Endian

    1.2.840.10008.1.2.4.901.2.840.10008.1.2.4.911.2.840.10008.1.2.4.50

    1.2.840.10008.1.2.4.70

    1.2.840.10008.1.2.5

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCP None

    Ultrasound Multi-frame

    Image Storage

    1.2.840.10008.5.1.4.1.1.3.1JPEG 2000 LosslessJPEG 2000 LossyJPEG BaselineJPEG Lossless Non

    Hierarchical (Proc 14)Run Length Encoding

    Exp VR Little EndianExplicit VR Big Endian

    Imp VR Little Endian

    1.2.840.10008.1.2.4.901.2.840.10008.1.2.4.911.2.840.10008.1.2.4.501.2.840.10008.1.2.4.70

    1.2.840.10008.1.2.5

    1.2.840.10008.1.2.11.2.840.10008.1.2.2

    1.2.840.10008.1.2

    SCP None

    Enhanced US VolumeStorage

    1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 LosslessJPEG 2000 LossyJPEG BaselineJPEG Lossless NonHierarchical (Proc 14)Run Length Encoding

    Exp VR Little EndianExplicit VR Big EndianImp VR Little Endian

    1.2.840.10008.1.2.4.901.2.840.10008.1.2.4.911.2.840.10008.1.2.4.501.2.840.10008.1.2.4.70

    1.2.840.10008.1.2.5

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCP None

    Ultrasound ImageStorage (retired)

    1.2.840.10008.5.1.4.1.1.6JPEG 2000 LosslessJPEG 2000 LossyJPEG Baseline

    JPEG Lossless NonHierarchical (Proc 14)Run Length Encoding

    Exp VR Little EndianExplicit VR Big EndianImp VR Little Endian

    1.2.840.10008.1.2.4.901.2.840.10008.1.2.4.911.2.840.10008.1.2.4.50

    1.2.840.10008.1.2.4.70

    1.2.840.10008.1.2.5

    1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2

    SCP None

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    43/214

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    44/214

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    45/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    45

    Element Name Tag VR VM Description

    Private Creator 7FE1,00xxLO 1 GEMS_Ultrasound_MovieGroup_001

    This means that all private tags starting with 7FE1,xx will belong to the GEMS_Ultrasound_MovieGroup_001.

    If so configured, the product will send preview image in private data elements designated by the Private Creator element:Element Name Tag VR VM Description

    Private Creator 6003,00xxLO 1 GEMS_Ultrasound_ImageGroup_001

    This means that all private tags starting with 6003,xx will belong to the GEMS_Ultrasound_ImageGroup_001.

    The product will send measurements raw data information in private data elements designated by the Private Creator

    element:

    Element Name Tag VR VM Description

    Private Creator 6005,00xxLO 1 GE_GROUP

    This means that all private tags starting with 6005,xx will belong to the GE_GROUP.

    2.6 CONFIGURATION

    2.6.1 AE Title/Presentation Address Mapping

    The Local AE title is configurable through the Config screen, see below.

    2.6.2 Configurable Parameters

    Network:

    Local IP address

    Local IP netmask

    Default Gateway Local network speed information

    Local:

    Local AE Title

    Local port number

    Verification:

    The AE Title, IP address and port number of the SCP

    Max retries, Retry interval, Timeout

    Modality Worklist:

    The AE Title, IP address and port number of the SCP

    Max retries, Retry interval, Timeout

    Refresh interval - the interval between downloads from the worklist. Searching for a patient within the time of a

    refresh interval does not perform a new worklist query.

    Disabling/enabling and setting constant values for query fields This is a way to disable search criteria in the

    worklist query or setting them to a fixed value.

    Maximum number of downloaded entries

    Storage:

    The AE Title, IP address and port number of the SCP

    Max retries, Retry interval, Timeout

  • 8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2

    46/214

    LOGIQ F SERIES

    GE Healthcare DICOM CONFORMANCE STATEMENT

    DOC1462293REV 2

    46

    Enable/disable raw data

    Frame rate reduction

    Enable/disable multi-frame

    Compression selections

    Color support

    Association strategies: one association per image or one association per exam Include structured report (SR).

    Included Key Object Selection Document for Image Rejection (KO