8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
1/214
gGE Healthcare
TechnicalPublications
Direction DOC1462293
Revision 2
LOGIQ F SERIES*
DICOM CONFORMANCE STATEMENT
Note: In the file, LOGIQ F SERIES means LOGIQ F3 & LOGIQ F5 & LOGIQ F6 & LOGIQ F8
Do not duplicate
Copyright 2013 by General Electric Co.
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
2/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
2
THIS PAGE LEFT INTENTIONALLY BLANK
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
3/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
3
CONFORMANCE STATEMENT OVERVIEW
The LOGIQ F Series is a self-contained networked computer system used for acquiring ultrasound diagnostic medicalimages. The system implements the necessary DICOM services to download worklists from an information system, save
acquired ultrasound images to a network storage device or media, print to a networked hardcopy device, query and moveultrasound images from a networked storage and inform the information system about the work actually done. The system
conforms to the DICOM standard to allow the sharing of medical information with other digital imaging systems. LOGIQ F
SeriesProduct is compliance to DICOM 3.0.
Table 0.1 provides an overview of the network services supported by LOGIQ F Series.
Table 0.1 NETWORK SERVICES
SOP Classes User of Service
(SCU)
Provider of
Service (SCP)
Transfer
Ultrasound Image Storage Yes Yes
Ultrasound Image Storage (Retired) Yes Yes
Ultrasound Multi-frame Image Storage Yes Yes
Ultrasound Multi-frame Image Storage (Retired) Yes Yes
Enhanced US Volume Storage Yes Yes
Secondary Capture Image Storage Yes Yes
Comprehensive SR Yes No
Key Object Selection Document Yes No
Verification SOP Class Yes Yes
Query/Retrieve
Study Root Query/Retrieve Information Model FIND Yes No
Study Root Query/Retrieve Information Model MOVE Yes No
Print Management
Basic Grayscale Print Management Meta SOP Class Yes No
Basic Color Print Management Meta SOP Class Yes No
Basic Annotation Box Yes No
Printer SOP Class Yes No
Basic Film Session SOP Class Yes No
Basic Film Box SOP Class Yes No
Basic Grayscale Image Box SOP Class Yes No
Basic Color Image Box SOP Class Yes No
Workflow Management
Storage Commitment Push Model SOP Class Yes No
Modality Performed Procedure Step SOP Class Yes No
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
4/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
4
Modality Worklist Information Model FIND SOP Class Yes No
Table 0.2 provides an overview of the Media Storage Application Profilessupported by LOGIQ F Series.
Table 0.2 - MEDIA SERVICES
Media Storage Application Profile Write Files
(FSC or FSU)
Read Files
(FSR)
Compact Disk - Recordable
General Purpose CD-R Yes Yes
DVD
General Purpose DVD Interchange with JPEG Yes Yes
General Purpose DVD Interchange with JPEG 2000 Yes Yes
USB
General Purpose USB Media Interchange with JPEG Yes Yes
General Purpose USB Media Interchange with JPEG 2000 Yes Yes
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
5/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
5
TABLE OF CONTENTS
1. INTRODUCTION ............................................................................................................... 9
1.1 Overview ...................................................................................................................................................................... 9
1.2 Overall DICOM Conformance Statement Document Structure .......................................................................... 11
1.3 Intended Audience .................................................................................................................................................... 13
1.4 Scope and Field of Application ................................................................................................................................ 13
1.5 Important Remarks .................................................................................................................................................. 13
1.6 References .................................................................................................................................................................. 14
1.7 Definitions .................................................................................................................................................................. 14
1.8 Symbols and Abbreviations ...................................................................................................................................... 15
2. NETWORK CONFORMANCE STATEMENT.................................................................. 18
2.1 Introduction ............................................................................................................................................................... 18
2.2 Implementation Model ............................................................................................................................................. 18
2.3 AE Specifications ....................................................................................................................................................... 21
2.4 COMMUNICATION PROFILES ........................................................................................................................... 44
3. MEDIA STORAGE CONFORMANCE STATEMENT ...................................................... 47
3.4 AUGMENTED AND PRIVATE APPLICATION PROFILES ............................................................................ 51
4. ULTRASOUND IMAGE INFORMATION OBJECT IMPLEMENTATION........................ 53
4.1 Introduction ............................................................................................................................................................... 53
4.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 53
4.3 IOD Module Table .................................................................................................................................................... 53
4.4 Information Module Definitions .............................................................................................................................. 54
5. ULTRASOUND MULTI-FRAME IMAGE INFORMATION OBJECT IMPLEMENTATION66
5.1 Introduction ............................................................................................................................................................... 66
5.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 66
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
6/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
6
5.3 IOD Module Table .................................................................................................................................................... 66
5.4 Information Module Definitions .............................................................................................................................. 67
6. ENHANCED US VOLUME INFORMATION OBJECT IMPLEMENTATION ................... 69
6.1 Introduction ............................................................................................................................................................... 69
6.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 69
6.3 IOD Module Table .................................................................................................................................................... 69
6.4 Information Module Definitions .............................................................................................................................. 70
7. SECONDARY CAPTURE INFORMATION OBJECT IMPLEMENTATION .................... 80
7.1 Introduction ............................................................................................................................................................... 80
7.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 80
7.3 IOD Module Table .................................................................................................................................................... 80
7.4 Information Module Definitions .............................................................................................................................. 81
8. COMPREHENSIVE STRUCTURED REPORT INFORMATION OBJECT
IMPLEMENTATION .................................................................................................................. 83
8.1 Introduction ............................................................................................................................................................... 83
8.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 83
8.3 IOD Module Table .................................................................................................................................................... 83
8.4 Information Module Definitions .............................................................................................................................. 84
8.5 Standard Extended and Private Data Attributes ................................................................................................... 89
8.6 Standard Extended and Private Context Groups .................................................................................................. 89
8.7 Standard, Standard Extended and Private Templates .......................................................................................... 89
9. KEY OBJECT SELECTION DOCUMENT INFORMATION OBJECT IMPLEMENTATION91
9.1 Introduction ............................................................................................................................................................... 91
9.2 LOGIQ F SeriesMapping of DICOM Entities ...................................................................................................... 91
9.3 IOD Module Table .................................................................................................................................................... 91
9.4 Information Module Definitions .............................................................................................................................. 92
9.5 Standard, Standard Extended and Private Templates .......................................................................................... 93
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
7/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
7
10. MODALITY WORKLIST INFORMATION MODEL DEFINITION .................................... 94
10.1 Introduction ............................................................................................................................................................... 94
10.2 Modality Worklist Information Model Description .............................................................................................. 94
10.3 Modality Worklist Information Model Entity-Relationship Model ..................................................................... 94
10.4 Information Model Module Table ........................................................................................................................... 96
10.5 Information Model Keys .......................................................................................................................................... 97
11. MODALITY PERFORMED PROCEDURE STEP IMPLEMENTATION ........................ 103
11.1 Introduction ............................................................................................................................................................. 103
11.2 RELATIONSHIP BETWEEN SCHEDULED AND PERFORMED PROCEDURE STEPS ......................... 103
11.3 MODALITY PERFORMED PROCEDURE STEP MODULE TABLE ........................................................... 103
11.4 MODALITY PERFORMED PROCEDURE STEP MODULE DEFINITIONS .............................................. 104
11.5 SOP Common Module ............................................................................................................................................ 104
11.6 Performed Procedure Step Relationship Module ................................................................................................ 104
11.7 Performed Procedure Step Information Module ................................................................................................. 106
11.8 Image Acquisition Results Module ........................................................................................................................ 106
12. STORAGE COMMITMENT PUSH MODEL IMPLEMENTATION ................................. 108
12.1 Introduction ............................................................................................................................................................. 108
12.2 Storage Commitment Push Model Sop Class Definition ..................................................................................... 108
13. BASIC DIRECTORY INFORMATION OBJECT IMPLEMENTATION .......................... 112
13.1 Introduction ............................................................................................................................................................. 112
13.2 Basic Directory Iod Implementation ..................................................................................................................... 112
13.3 Basic Directory Entity-Relationship Model .......................................................................................................... 112
13.4 IOD Module Table .................................................................................................................................................. 113
13.5 Information Module Definitions ............................................................................................................................ 114
13.6 PRIVATE DATA DICTIONARY ......................................................................................................................... 121
14. PRINT MANAGEMENT IMPLEMENTATION................................................................ 123
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
8/214
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
9/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
9
1. INTRODUCTION
1.1 OVERVIEW
This DICOM Conformance Statement is divided into Sections as described below:
Section 1 (Introduction),which describes the overall structure, intent, and references for this Conformance Statement
Section 2 (Network Conformance Statement), which specifies the GEHC equipment compliance to the DICOM
requirements for the implementation of Networking features.
Section 3 (Media Storage Conformance Statement),which specifies the GEHC equipment compliance to the DICOM
requirements for the implementation of Media Storage features.
Section 4 (Ultrasound Image Information Object Implementation),which specifies the GEHC equipment compliance to
DICOM requirements for the implementation of an UltrasoundImage Information Object.
Section 5 (Ultrasound Multi-Frame Image Information Object Implementation),which specifies the GEHC equipmentcompliance to DICOM requirements for the implementation of an Ultrasound Multi-Frame Image Information.
Section 6 (Enhanced US Volume Information Object Implementation),which specifies the GEHC equipment complianceto DICOM requirements for the implementation of an Enhanced US Volume Information.
Section 7 (Secondary Capture Object Implementation), which specifies the GEHC equipment compliance to DICOMrequirements for the implementation of a Secondary Capture Information Object.
Section 8 (Comprehensive Structured Report Object Implementation), which specifies the GEHC equipment compliance
to DICOM requirements for the implementation of a Comprehensive Structured Reporting Information Object.
Section 9 (Key Object Selection Document Implementation), which specifies the GEHC equipment compliance to
DICOM requirements for the implementation of a Key Object Selection Document Object.
Section 10 (Modality Worklist Information Model), which specifies the GEHC equipment compliance to DICOMrequirements for the implementation of the Modality Worklist service.
Section 11 (Modality Performed Procedure Step Implementation),which specifies the GEHC equipment compliance toDICOM requirements for the implementation of Modality Performed Procedure Step Service.
Section 12 (Storage Commitment PushModel SOP Implementation),which specifies the GEHC equipment complianceto DICOM requirements for the implementation of the Storage CommitmentPush Model Service.
Section 13 (Basic Directory Information Object Implementation), which specifies the GEHC equipment compliance toDICOM requirements for the implementation of a Basic Directory Information Object.
Section 14 (Print Management), which specifies the GEHC equipment compliance to DICOM requirements for the
implementation of Basic Print Meta SOP Classes (Gray and Color).
Section 15 (Study Root Query/Retrieve Information Model), which specifies the GEHC equipment compliance to
DICOM requirements for the Study Root Query/Retrieve Information Model.
Section 16 (DICOM SR Templates),which specifies all LOGIQ F Seriessupported DICOM SR Templates.
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
10/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
10
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
11/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
11
1.2 OVERALL DICOM CONFORMANCE STATEMENT DOCUMENT STRUCTURE
The Documentation Structure of the GEHC DICOM Conformance Statements is shown in the Illustration below.
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
12/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
12
This document specifies the DICOM implementation. It is entitled:
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
13/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
13
LOGIQ F Series
Conformance Statement for DICOM
Direction DOC1462293
This DICOM Conformance Statement documents the DICOM Conformance Statement and Technical Specification requiredto interoperate with the GEHC network interface.
The GEHC Conformance Statement, contained in this document, also specifies the Lower Layer communications which itsupports (e.g., TCP/IP). However, the Technical Specifications are defined in the DICOM Part 8 standard.
For more information regarding DICOM, copies of the Standard may be obtained on the Internet at http://medical.nema.org.Comments on the Standard may be addressed to:
DICOM SecretariatNEMA
1300 N. 17thStreet, Suite 1752
Rosslyn, VA 22209
USA
Phone: +1.703.841.3200
1.3 INTENDED AUDIENCE
The reader of this document is concerned with software design and/or system integration issues. It is assumed that the
reader of this document is familiar with the DICOM Standard and with the terminology and concepts which are used in thatStandard.
1.4 SCOPE AND FIELD OF APPLICATION
It is the intent of this document to provide an unambiguous specification for GEHC implementations. This specification,
called a Conformance Statement, includes a DICOM Conformance Statement and is necessary to ensure proper processing
and interpretation of GEHC medical data exchanged using DICOM. The GEHC Conformance Statements are available to
the public.
The reader of this DICOM Conformance Statement should be aware that different GEHC devices are capable of usingdifferent Information Object Definitions. For example, a GEHC CT Scanner may send images using the CT InformationObject, MR Information Object, Secondary Capture Object, etc.
Included in this DICOM Conformance Statement are the Module Definitions which define all data elements used by thisGEHC implementation. If the user encounters unspecified private data elements while parsing a GEHC Data Set, the user is
well advised to ignore those data elements (per the DICOM standard). Unspecified private data element information is
subject to change without notice. If, however, the device is acting as a "full fidelity storage device", it should retain and re-transmit all of the private data elements which are sent by GEHC devices.
1.5 IMPORTANT REMARKS
The use of these DICOM Conformance Statements, in conjunction with the DICOM Standards, is intended to facilitatecommunication with GE imaging equipment. However, by itself, it is not sufficient to ensure that inter-operation will be
successful. The user (or user's agent)needs to proceed with caution and address at least four issues:
Integration - The integration of any device into an overall system of interconnected devices goes beyond the scope of
standards (DICOM v3.0), and of this introduction and associated DICOM Conformance Statements when interoperability
with non-GE equipment is desired. The responsibility to analyze the applications requirements and to design a solution thatintegrates GE imaging equipment with nonGE systems is the user'sresponsibility and should not be underestimated. The
useris strongly advised to ensure that such an integration analysis is correctly performed.
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
14/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
14
Validation - Testing the complete range of possible interactions between any GE device and nonGE devices, before theconnection is declared operational, should not be overlooked. Therefore, the usershould ensure that any nonGE provider
accepts full responsibility for all validation required for their connection with GE devices. This includes the accuracy of the
image data once it has crossed the interface between the GE imaging equipment and the nonGE device and the stability ofthe image data for the intended applications.
Such a validation is required before any clinical use (diagnosis and/or treatment) is performed. It applies when imagesacquired on GE imaging equipment are processed/displayed on a non-GE device, as well as when images acquired on non-
GE equipment is processed/displayed on a GE console or workstation.
Future Evolution- GE understands that the DICOM Standard will evolve to meet the user's growing requirements. GE is
actively involved in the development of the DICOM Standard. DICOM will incorporate new features and technologies andGE may follow the evolution of the Standard. The GEHC protocol is based on DICOM as specified in each DICOM
Conformance Statement. Evolution of the Standard may require changes to devices which have implemented DICOM. In
addition, GE reserves the right to discontinue or make changes to the support of communications features (on its
products) described by these DICOM Conformance Statements. The user should ensure that any nonGE provider,
which connects with GE devices, also plans for the future evolution of the DICOM Standard. Failure to do so will likely
result in the loss of function and/or connectivity as the DICOM Standard changes and GE Products are enhanced to support
these changes.
Interaction - It is the sole responsibility of the nonGE provider to ensure that communication with the interfacedequipment does not cause degradation of GE imaging equipment performance and/or function.
1.6 REFERENCES
NEMA PS3 Digital Imaging and Communications in Medicine (DICOM) Standard, available free at
http://medical.nema.org/
1.7 DEFINITIONS
Informal definitions are provided for the following terms used in this Conformance Statement. The DICOM Standard is the
authoritative source for formal definitions of these terms.
Abstract Syntax the information agreed to be exchanged between applications, generally equivalent to a Service/Object
Pair (SOP) Class. Examples : Verification SOP Class, Modality Worklist Information Model Find SOP Class, ComputedRadiography Image Storage SOP Class.
Application Entity (AE) an end point of a DICOM information exchange, including the DICOM network or mediainterface software; i.e., the software that sends or receives DICOM information objects or messages. A single device may
have multiple Application Entities.
Application Entity Title the externally known name of an Application Entity, used to identify a DICOM application to
other DICOM applications on the network.
Application Context the specification of the type of communication used betweenApplication Entities. Example: DICOMnetwork protocol.
Association a network communication channel set up betweenApplication Entities.
Attribute a unit of information in an object definition; a data element identified by a tag. The information may be acomplex data structure (Sequence), itself composed of lower level data elements. Examples: Patient ID (0010,0020),
Accession Number (0008,0050), Photometric Interpretation (0028,0004), Procedure Code Sequence (0008,1032).
Information Object Definition (IOD) the specified set ofAttributesthat comprise a type of data object; does not representa specific instance of the data object, but rather a class of similar data objects that have the same properties. The Attributes
may be specified as Mandatory (Type 1), Required but possibly unknown (Type 2), or Optional (Type 3), and there may be
conditions associated with the use of an Attribute (Types 1C and 2C). Examples: MR Image IOD, CT Image IOD, Print Job
IOD.
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
15/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
15
Joint Photographic Experts Group (JPEG) a set of standardized image compression techniques, available for use byDICOM applications.
Media Application Profile the specification of DICOM information objects and encoding exchanged on removable media
(e.g., CDs)
Module a set ofAttributeswithin anInformation Object Definitionthat are logically related to each other. Example: PatientModule includes Patient Name, Patient ID, Patient Birth Date, and Patient Sex.
Negotiation first phase of Associationestablishment that allows Application Entities to agree on the types of data to beexchanged and how that data will be encoded.
Presentation Context the set of DICOM network services used over an Association, as negotiated between ApplicationEntities; includesAbstract Syntaxes and Transfer Syntaxes.
Protocol Data Unit (PDU) a packet (piece) of a DICOM message sent across the network. Devices must specify the
maximum size packet they can receive for DICOM messages.
Security Profile a set of mechanisms, such as encryption, user authentication, or digital signatures, used by an Application
Entityto ensure confidentiality, integrity, and/or availability of exchanged DICOM data
Service Class Provider (SCP) role of an Application Entitythat provides a DICOM network service; typically, a server
that performs operations requested by another Application Entity (Service Class User). Examples: Picture Archiving andCommunication System (image storage SCP, and image query/retrieve SCP), Radiology Information System (modality
worklist SCP).
Service Class User (SCU) role of anApplication Entitythat uses a DICOM network service; typically, a client. Examples:imaging modality (image storage SCU, and modality worklist SCU), imaging workstation (image query/retrieve SCU)
Service/Object Pair (SOP) Class the specification of the network or media transfer (service) of a particular type of data(object); the fundamental unit of DICOM interoperability specification. Examples: Ultrasound Image Storage Service, Basic
Grayscale Print Management.
Service/Object Pair (SOP) Instance an information object; a specific occurrence of information exchanged in a SOPClass. Examples: a specific x-ray image.
Tag a 32-bit identifier for a data element, represented as a pair of four digit hexadecimal numbers, the group and the
element. If the group number is odd, the tag is for a private (manufacturer-specific) data element. Examples: (0010,0020)[Patient ID], (07FE,0010) [Pixel Data], (0019,0210) [private data element]
Transfer Syntax the encoding used for exchange of DICOM information objects and messages. Examples: JPEGcompressed (images), little endian explicit value representation.
Unique Identifier (UID) a globally unique dotted decimal string that identifies a specific object or a class of objects; anISO-8824 Object Identifier. Examples: Study Instance UID, SOP Class UID, SOP Instance UID.
Value Representation (VR) the format type of an individual DICOM data element, such as text, an integer, a personsname, or a code. DICOM information objects can be transmitted with either explicit identification of the type of each data
element (Explicit VR), or without explicit identification (Implicit VR); with Implicit VR, the receiving application must use a
DICOM data dictionary to look up the format of each data element.
1.8 SYMBOLS AND ABBREVIATIONS
AE Application Entity
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
16/214
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
17/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
17
PDU Protocol Data Unit
R Required (Key Attribute)
RIS Radiology Information System
SC Secondary Capture
SCP Service Class Provider
SCU Service Class User
SOP Service-Object Pair
SPS Scheduled Procedure Step
SR Structured Reporting
TCP/IP Transmission Control Protocol/Internet Protocol
U Unique (Key Attribute)
UL Upper Layer
US Ultrasound
VR Value Representation
KOS Key Object Selection
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
18/214
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
19/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
19
ILLUSTRATION 21
LOGIQFSERIESNETWORK APPLICATION MODEL AND DATA FLOW DIAGRAM
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
20/214
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
21/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
21
There are six local real-world activities that occur in LOGIQ F Series- Exam Save, Echo, Worklist Query, Image
Store/End Exam,Image Printand Exam Query/Retrieve
Exam save initiates a connection with the DICOM SCP and transmits images and results to the DICOM SCP. If Storage
Commitment is configured a commitment request will be sent for the images and results.
Echoinitiates a connection with the DICOM SCP, posts a Verification request and closes the connection. It also responds toincoming Verification requests (for service use).
Worklist Query initiates a connection with the DICOM SCP, performs a query and retrieves the matching entries to theproduct.
Image Store/End exam: If Modality Performed Procedure Step is configured N-CREATE and N-SET messages will be sentfor the exam.
Image Printwill send images to a DICOM Print SCP.
Exam Query/Retrieveinitiates a connection with the DICOM SCP, performs a query and retrieves selected examination.
2.2.2 Functional Definition of AE's
Application Entity LOGIQ F Seriessupports the following functions:
Initiates a DICOM association to send images and results.
Initiates a DICOM verification to assist in network diagnostics.
Initiates a DICOM worklist query to receive worklist information.
Initiates a DICOM association to notify start of examination.
Initiates a DICOM association to notify end of examination.
Initiates a DICOM association to request storage commitment of images.
Responds to replies for storage commitment requests of images.
Initiates a DICOM association to print images.
Initiates a DICOM association to query for and retrieve examinations.
Responds to storage requests for images triggered by examination retrieve requests.
2.2.3 Sequencing of Real-World Activities
Non Applicable.
2.3 AE SPECIFICATIONS
2.3.1 LOGIQ F SeriesAE Specification
This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an SCU and/or as anSCP:
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
22/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
22
SOP Class Name SOP Class UID SCU SCP
Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 Yes Yes
Ultrasound Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.6 Yes Yes
Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 Yes Yes
Ultrasound Multi-frame Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.3 Yes Yes
Enhanced US Volume Storage 1.2.840.10008.5.1.4.1.1.6.2 Yes Yes
Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 Yes Yes
Comprehensive SR 1.2.840.10008.5.1.4.1.1.88.33 Yes No
Key Object Selection Document Storage 1.2.840.10008.5.1.4.1.1.88.59 Yes No
Verification SOP Class 1.2.840.10008.1.1 Yes Yes
Study Root Query/Retrieve Information Model - FIND 1.2.840.10008.5.1.4.1.2.2.1 Yes No
Study Root Query/Retrieve Information Model - MOVE 1.2.840.10008.5.1.4.1.2.2.2 Yes No
Modality Worklist Information Model - FIND 1.2.840.10008.5.1.4.31 Yes No
Modality Performed Procedure Step 1.2.840.10008.3.1.2.3.3 Yes No
Storage Commitment Push Model 1.2.840.10008.1.20.1 Yes No
Basic Grayscale Print Management Meta SOP Class 1.2.840.10008.5.1.1.9 Yes No
Basic Color Print Management Meta SOP Class 1.2.840.10008.5.1.1.18 Yes No
Basic Annotation Box 1.2.840.10008.5.1.1.15 Yes No
Printer SOP Class 1.2.840.10008.5.1.1.16 Yes No
Basic Film Session SOP Class 1.2.840.10008.5.1.1.1 Yes No
Basic Film Box SOP Class 1.2.840.10008.5.1.1.2 Yes No
Basic Grayscale Image Box SOP Class 1.2.840.10008.5.1.1.4 Yes No
Basic Color Image Box SOP Class 1.2.840.10008.5.1.1.4.1 Yes No
2.3.1.1 Association Establishment Policies
2.3.1.1.1 General
The DICOM Application Context Name (ACN), which is always proposed, is:
Application Context Name 1.2.840.10008.3.1.1.1
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
23/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
23
The Maximum Length PDU negotiation is included in all association establishment requests. The maximum length PDUfor an association initiated by LOGIQ F Seriesis:
Maximum Length PDU 32768
The maximum length PDU is a fixed number - not configurable.
The SOP Class Extended Negotiation is not supported.
The user information Items sent by this product are:
Maximum PDU Length
Implementation UID
Implementation Version Name
2.3.1.1.2 Number of Associations
The LOGIQ F SeriesAE will initiate multiple DICOM associations. The maximum numbers of associations are based on theconnectivity service configuration.
2.3.1.1.3 Asynchronous Nature
Asynchronous mode is not supported. All operations will be performed synchronously.
2.3.1.1.4 Implementation Identifying Information
The Implementation UID for this DICOM Implementation is:
LOGIQ F SeriesImplementation UID 1.2.840.113619.6.384
LOGIQ F SeriesImplementation Version Name LOGIQ F Series
2.3.1.2 Association Initiation Policy
The LOGIQ F SeriesAE attempts to establish a new association with a remote device due to six Real-World Activities:
Exam save initiated by the operator for images, key object selection document and structured reports and sending request forStorage Commitment.
Verification, which verifies application level communication between peer DICOM AEs for service purposes.
Worklist initiated by the operator for receiving worklist information.Image Store/End Exam sending messages to Modality Performed Procedure Step.
Print initiated by the operator for a specific image or group of images.
Exam Query/Retrieve initiated by the operator for receiving examination information and selecting examination to retrieve.
2.3.1.2.1 Real-World Activity - Exam save Operation
2.3.1.2.1.1 Associated Real-World Activity
Upon a request by the operator (manual or automatic), images will be sent to a DICOM Storage SCP.
2.3.1.2.1.2 Proposed Presentation Context Tables
The Proposed Presentation Context Table depends on compression according to the following table:
Presentation Context Table Proposed
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
24/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
24
Abstract Syntax Transfer Syntax Role ExtendedNegotiation
Name UID Name List UID List
Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to None
Ultrasound Image
Storage
1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little Endian
Explicit VR Big EndianImplicit VR Little Endian
1.2.840.10008.1.2.1
1.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound Multi-
frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 Explicit VR Little Endian
Explicit VR Big EndianImplicit VR Little Endian
1.2.840.10008.1.2.1
1.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Enhanced US Volume
Storage
1.2.840.10008.5.1.4.1.1.6.2 Explicit VR Little Endian
Explicit VR Big EndianImplicit VR Little Endian
1.2.840.10008.1.2.1
1.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Secondary CaptureImage Storage
1.2.840.10008.5.1.4.1.1.7 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Presentation Context Table: Still Compression set to RLE, Clip/Volume Compression set to RLE Secondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound Image
Storage
1.2.840.10008.5.1.4.1.1.6.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Presentation Context Table: Still Compression set to JPEG Lossy, Clip/Volume Compression set to
JPEG LossySecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Presentation Context Table: Still Compression set to JPEG Lossless, Clip/Volume Compression set to
JPEG LosslessSecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Lossless Non
Hierarchical (Proc 14)1.2.840.10008.1.2.4.70 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound Image 1.2.840.10008.5.1.4.1.1.6 JPEG Lossless Non 1.2.840.10008.1.2.4.70 SCU None
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
25/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
25
Storage (retired) Hierarchical (Proc 14)
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound Multi-
frame Image Storage(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG Lossless Non
Hierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Presentation Context Table: Still Compression set to JPEG 2000 Lossy, Clip/Volume Compression
set to JPEG 2000 LossySecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression
set to JPEG 2000 LosslessSecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to RLE Secondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian
Explicit VR Big EndianImplicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.2
1.2.840.10008.1.2
SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.2
1.2.840.10008.1.2
SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to JPEG
LossySecondary Capture 1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
26/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
26
Image Storage Explicit VR Big EndianImplicit VR Little Endian
1.2.840.10008.1.2.21.2.840.10008.1.2
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.2
1.2.840.10008.1.2
SCU None
Ultrasound Image
Storage (retired)
1.2.840.10008.5.1.4.1.1.6 Explicit VR Little Endian
Explicit VR Big EndianImplicit VR Little Endian
1.2.840.10008.1.2.1
1.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to JPEG
LosslessSecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian
Explicit VR Big EndianImplicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.2
1.2.840.10008.1.2
SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to JPEG
2000 LossySecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian
Explicit VR Big EndianImplicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Presentation Context Table: Still Compression set to None, Clip/Volume Compression set to JPEG
2000 LosslessSecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian
Explicit VR Big EndianImplicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
27/214
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
28/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
28
Secondary CaptureImage Storage
1.2.840.10008.5.1.4.1.1.7 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Enhanced US Volume
Storage
1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Presentation Context Table: Still Compression set to RLE, Clip/Volume Compression set to JPEG
2000 LosslessSecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Presentation Context Table: Still Compression set to JPEG Lossy, Clip/Volume Compression set to
NoneSecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound Image
Storage
1.2.840.10008.5.1.4.1.1.6.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Presentation Context Table: Still Compression set to JPEG Lossy, Clip/Volume Compression set to
RLESecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
29/214
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
30/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
30
Storage Explicit VR Big EndianImplicit VR Little Endian
1.2.840.10008.1.2.21.2.840.10008.1.2
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.2
1.2.840.10008.1.2
SCU None
Presentation Context Table: Still Compression set to JPEG Lossless, Clip/Volume Compression set to
RLESecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Lossless Non
Hierarchical (Proc 14)1.2.840.10008.1.2.4.70 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Presentation Context Table: Still Compression set to JPEG Lossless, Clip/Volume Compression set to
JPEG LossySecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Lossless Non
Hierarchical (Proc 14)1.2.840.10008.1.2.4.70 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound Multi-
frame Image Storage(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Presentation Context Table: Still Compression set to JPEG Lossless, Clip/Volume Compression set to
JPEG 2000 LossySecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Lossless Non
Hierarchical (Proc 14)1.2.840.10008.1.2.4.70 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Presentation Context Table: Still Compression set to JPEG Lossless, Clip/Volume Compression set to
JPEG 2000 LosslessSecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG Lossless Non
Hierarchical (Proc 14)1.2.840.10008.1.2.4.70 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound Image
Storage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG Lossless Non
Hierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
31/214
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
32/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
32
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound Multi-
frame Image Storage(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG Lossless Non
Hierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Presentation Context Table: Still Compression set to JPEG 2000 Lossy, Clip/Volume Compression
set to JPEG 2000 LosslessSecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression
set to NoneSecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Image
Storage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 Explicit VR Little EndianExplicit VR Big Endian
Implicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCU None
Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression
set to RLESecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 Run Length Encoding 1.2.840.10008.1.2.5 SCU None
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
33/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
33
Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression
set to JPEG LossySecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Image
Storage
1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG Baseline 1.2.840.10008.1.2.4.50 SCU None
Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression
set to JPEG LosslessSecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Multi-frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG Lossless NonHierarchical (Proc 14)
1.2.840.10008.1.2.4.70 SCU None
Presentation Context Table: Still Compression set to JPEG 2000 Lossless, Clip/Volume Compression
set to JPEG 2000 LossySecondary Capture
Image Storage1.2.840.10008.5.1.4.1.1.7 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6 JPEG 2000 Lossless 1.2.840.10008.1.2.4.90 SCU None
Ultrasound Multi-
frame Image Storage
1.2.840.10008.5.1.4.1.1.3.1 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Ultrasound Multi-frame Image Storage
(retired)
1.2.840.10008.5.1.4.1.1.3 JPEG 2000 Lossy 1.2.840.10008.1.2.4.91 SCU None
Presentation Context Table for Structured Reports
ComprehensiveStructured Report 1.2.840.10008.5.1.4.1.1.88.33 Explicit VR Little EndianImplicit VR Little Endian 1.2.840.10008.1.2.11.2.840.10008.1.2 SCU None
Presentation Context Table for Key Object Selection Document Key Object Selection
Document1.2.840.10008.5.1.4.1.1.88.59 Explicit VR Little Endian
Implicit VR Little Endian
1.2.840.10008.1.2.1
1.2.840.10008.1.2
SCU None
This operation also sends a Storage Commitment Request, with the following proposed presentation context. The result fromthe SCP is expected on another association for the Storage Commitment result.
Presentation Context Table Proposed
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
34/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
34
Presentation Context Table Proposed
Abstract Syntax Transfer Syntax Role Extended
NegotiationName UID Name List UID List
Storage
Commitment PushModel SOP Class
1.2.840.10008.1.20.1 Explicit VR Little Endian
Implicit VR Little Endian
1.2.840.10008.1.2.1
1.2.840.10008.1.2 SCU None
2.3.1.2.1.2.1 SOP Specific Conformance Statement for Image Storage SOP Classes
For these SOP class all status codes with status Refused or Error are treated as failures and terminate the association and
operation. On a failure, the request will be put in a holding queue for the user to manually retry the request. All status codes
with status Warning or Success are treated as successes.
Service
Status
Further Meaning Status
Codes
Application
Behavior When
receiving Status
Codes
Failure Refused: Out of Resources A7xx Terminateassociation. Mark the
job failed.Error: Data Set does not match SOPClass
A9xx
Error: Cannot understand Cxxx
Warning Coercion of Data Elements B000 Treat the job success.
Data Set does not match SOP Class B007
Elements Discarded B006
Success 0000
* Any other status code * Terminate
association. Mark thejob failed.
C-STORE SCU operation supports an Association Timer and Operation Inactivity Timer with default time out value of
30 seconds each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1, 2, 3, 4, 5, 10,20, 30, 60, 120 and 360 seconds.
2.3.1.2.2 Real-World Activity - Echo Operation
2.3.1.2.2.1 Associated Real-World Activity
The user may initiate a DICOM Verification Request in the config screen.Associations will be released upon the receipt of each C-ECHO confirmation.
2.3.1.2.2.2 Proposed Presentation Context Table
Presentation Context Table - Proposed
Abstract Syntax Transfer Syntax Role Extended
NegotiationName UID Name List UID List
Verification SOP Class 1.2.840.10008.1.1Explicit VR Little Endian 1.2.840.10008.1.2.1 SCU None
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
35/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
35
Implicit VR Little Endian 1.2.840.10008.1.2
2.3.1.2.2.2.1 SOP Specific DICOM Conformance Statement for Verify SOP Class
The AE provides standard conformance to the Verification SOP Class as an SCU. In the event that the SCP does not respond
for some reason, the operation will time out and LOGIQ F Serieswill close the association.
C-ECHO SCU operation supports an Association Timer and Operation Inactivity Timer with default time out value of 30
seconds and 10secs each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1 to 10
seconds.
2.3.1.2.3 Real-World Activity - Worklist Query Operation
2.3.1.2.3.1 Associated Real-World Activity
The user may initiate a DICOM Worklist Query in Search screen, which will send a C-FIND-RQ to the Worklist SCP.
Associations will be released upon the receipt of C-FIND-RSP confirmation.
2.3.1.2.3.2 Proposed Presentation Context Tables
Presentation Context Table Proposed
Abstract Syntax Transfer Syntax Role Extended
NegotiationName UID Name List UID List
Modality Worklist InformationModel FIND
1.2.840.10008.5.1.4.31Explicit VR Little EndianImplicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2
SCU None
2.3.1.2.3.2.1 SOP Specific DICOM Conformance Statement for Worklist SOP Classes
The LOGIQ F Seriesincludes matching keys in the Modality Worklist queries as described in Section 10.All status codes with status Refused or Error are treated as failures and terminate the association and operation. On a failure,
the user will be informed and the last successful query will be used as Worklist. All status codes with status Warning orSuccess are treated as successes. Application doesnt support C-FIND-CANCEL request.
Service Status Status Code Further Meaning Application Behavior
When receiving
Status Codes
Refused A700 Out of resources Terminate theassociation andoperation
0122 SOP Class not
Supported
Failed A900 Identifier does notmatch SOP Class
Cxxx Unable to process
Success 0000 Matching is complete
No final identifier is
supplied
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
36/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
36
Pending FF00 Matches are
continuing Current
Match is supplied and
any Optional Keys
were supported in thesame manner
Receiving process of
the matches continues.
FF01 Matches are
continuing Warning
that one or more
Optional Keys were
not supported for
existence for this
Identifier
* * Any other status code. Terminate the
association and
operation
Worklist operation supports an Association Timer and Operation Inactivity Timer with default time out value of 30
seconds each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1, 2, 3, 4, 5, 10,
20, 30, 60, 120 and 360 seconds.
2.3.1.2.4 Real-World Activity - Image Store/End exam Operation
2.3.1.2.4.1 Associated Real-World Activity
The Modality Performed Procedure Step messages are sent when the first image is acquired for the start of an exam and when
the exam is ended (for the case where there are no images, the N-CREATE is sent when the exam is ended). For an exam
with saved images or results, the N-SET will be sent with status COMPLETED. For an exam without saved images or
results, the N-SET will be sent with status DISCONTINUED.
2.3.1.2.4.2 Proposed Presentation Context Table
Presentation Context Table Proposed
Abstract Syntax Transfer Syntax Role Extended
Name UID Name List UID List Negotiation
Modality Performed Procedure
Step SOP Class1.2.840.10008.3.1.2.3.3
Explicit VR Little Endian
Implicit VR Little Endian
1.2.840.10008.1.2.1
1.2.840.10008.1.2
SCU None
2.3.1.2.4.2.1 SOP Specific DICOM Conformance Statement for Modality Performed Procedure Step Class
LOGIQ F Seriesincludes attributes in the Modality Performed Procedure Step N-CREATE and N-SET as described in
Section 10.2.
LOGIQ F Seriesincludes attributes in the Modality Performed Procedure Step N-SET as described in Section 11. The
mapping from Worklist attributes is described in Section 10.5. LOGIQ F Seriessends N-SET after the exam is ended. The N-
SET will include all acquired images UIDs and the status of COMPLETED or DISCONTINUED.
For this SOP class, all status codes with status Refused or Error are treated as failures and terminate the association and
operation. All status codes with status Warning or Success are treated as successes.
Modality Performed Procedure Step N-CREATE response status Handling Behavior
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
37/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
37
Service Status Status Code Further Meaning Application Behavior
When receiving
Status Codes
Success 0000 Success
* * Any other status Treat as failure.
Aborts the association.
Modality Performed Procedure Step N-SET response status Handling Behavior
Service Status Status Code Further Meaning Application Behavior
When receiving
Status Codes
Failure 0110H Processing Failure.
Performed ProcedureStep Object may no
longer be updated
Aborts the operation.
Success 0000 Success Completes the
operation.
* * Any other status Aborts the operation.
MPPS operation supports an Association Timer and Operation Inactivity Timer with default time out value of 30 seconds
each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1, 2, 3, 4, 5, 10, 20, 30, 60,120 and 360 seconds.
2.3.1.2.5 Real-World Activity - Image Print Operation
2.3.1.2.5.1 Associated Real-World Activity
Upon a request by the operator, print jobs will be sent to a DICOM Print SCP. If an error occurs during the transmission, thecurrent association is released and a new association initiated. The maximum number of retries is configurable.
2.3.1.2.5.2 Proposed Presentation Context Tables
The following table is used:
Presentation Context Table - Proposed
Abstract Syntax Transfer Syntax Role Extended
Name UID Name List UID List Negotiation
Basic Grayscale PrintManagement Meta SOP Class
1.2.840.10008.5.1.1.9Explicit VR Little EndianImplicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2
SCU None
Basic Color Print ManagementMeta SOP Class
1.2.840.10008.5.1.1.18 Explicit VR Little EndianImplicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2
SCU None
Basic Annotation Box SOPClass
1.2.840.10008.5.1.1.15Explicit VR Little EndianImplicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2
SCU None
2.3.1.2.5.2.1 SOP Specific DICOM Conformance Statement for all Print Management SOP Classes
All status codes with status Refused or Error are treated as failures and terminate the association and operation. All status
codes with status Warning or Success are treated as successes.
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
38/214
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
39/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
39
2.3.1.2.6.2 Proposed Presentation Context Tables
Presentation Context Table Proposed
Abstract Syntax Transfer Syntax Role Extended
NegotiationName UID Name List UID List
Study Root Query/RetrieveInformation Model - FIND
1.2.840.10008.5.1.4.1.2.2.1 Explicit VR Little EndianImplicit VR Little Endian
1.2.840.10008.1.2.11.2.840.10008.1.2
SCU None
Study Root Query/Retrieve
Information Model - MOVE
1.2.840.10008.5.1.4.1.2.2.2Explicit VR Little Endian
Implicit VR Little Endian
1.2.840.10008.1.2.1
1.2.840.10008.1.2
SCU None
2.3.1.2.6.2.1 SOP Specific DICOM Conformance Statement for Study Root Query/Retrieve Information Model -
FIND SOP Classes
Only a single information model, Study Root, is supported.
All queries are initiated at the highest level of the information model (the STUDY level), and then for each response
received, recursively repeated at the next lower levels (the SERIES and then IMAGE levels).
C-FIND-CANCEL request is not supported.
All status codes with status Refused or Error are treated as failures and terminate the association and operation. All status
codes with status Warning or Success are treated as successes. LOGIQ F Serieswill only support hierarchical query.
Service
Status
Status
Code
Further Meaning Application Behavior When Receiving Status
Code
Failure A700 Refused: Out of resources Operation aborted.
A900 Error: Identifier does not matchSOP Class
C000-
CFFF
Error: Unable to process
0122 SOP Class Not Supported
Cancel FE00 Matching terminated due to
cancel
Display the results in UI.
Success 0000 Matching is complete - No finalidentifier is supplied
Display the results in UI.
Pending FF00 Matches are continuing -
Current Match is supplied andany Optional Keys were
supported in the same manner
as Required Keys.
Stores the results internally and waits for the more
results.
FF01 Matches are continuing -Warning that one or more
Optional Keys were not
supported for existence and/or
matching for this Identifier
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
40/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
40
* * Any other status code. Operation aborted.
C-FIND SCU operation supports an Association Timer and Operation Inactivity Timer with default time out value of 30
seconds each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1, 2, 3, 4, 5, 10,20, 30, 60, 120 and 360 seconds.
LOGIQ F Seriesquery responses for Key Object Selection documents and Structured Report documents.
Study Level attributes for the Study Root Query/Retrieve Information Model is described in Section 15.2.
Types of Matching:
" Single Value matching (S)
" Universal Matching (U)
" Wild Card Matching (*)
" Range of date, Range of Time (R)
The types of Matching supported by the C-FIND SCU are: S indicates the identifier attribute uses Single Value Matching,
an R indicates Range Matching, a * indicates wildcard matching, a U indicates Universal Matching, and UNIQUE
indicates that this is the Unique Key for that query level, in which case Universal Matching or Single Value Matching is used
depending on the query level.
Filtering is supported means that matching strings can be controlled from the Search screen at Data Transfer. All other
matching fields can be configured in Config screen to be either enabled, enabled with a matching string value or disabled.
The constant value will be used as entered by user.
2.3.1.2.6.2.2 SOP Specific DICOM Conformance Statement for Study Root Query/Retrieve Information Model -
MOVE SOP Classes
All status codes with status Refused or Error are treated as failures and terminate the association and operation. All status
codes with status Warning or Success are treated as successes.
C-MOVE-CANCEL request is not supported.
Service
Status
Status
Code
Further Meaning Application Behavior When Receiving Status
Code
Failure A701 Refused: Out of resources -Unable to calculate number of
matches
Retrieve operation will be aborted.
A702 Refused: Out of resources -Unable to perform sub-
operations
A801 Refused: Move Destination
Unknown
A900 Error: Identifier does not match
SOP Class
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
41/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
41
C000-CFFF
Error: Unable to process
0122 SOP Class Not Supported
Cancel FE00 Sub-operations terminated dueto a Cancel indication
Warning B000 Sub-operations Complete - Oneor more Failures.
Imports selected patient(s).
Success 0000 Sub-operations Complete - NoFailure.
Pending FF00 Sub-operations are continuing - Stores the patients internally until final response.
* * Any other status code. Retrieve operation will be aborted.
C-MOVE SCU operation supports an Association Timer and Operation Inactivity Timer with default time out value of
30 seconds each. Operation Inactivity Timer is configurable and different values that can be assigned to it are 1, 2, 3, 4, 5, 10,
20, 30, 60, 120 and 360 seconds.
2.3.1.3 Association Acceptance Policy
The AE accepts an association when it receives a Verification Request from another network device or a store request from a
Q/R SCP or a Storage Commitment result from a Storage Commitment SCP.
2.3.1.3.1 Real-World Activity - Echo operation
2.3.1.3.1.1 Associated Real-World Activity
An incoming Verification Request will cause the AE to accept the association and respond with a Verification Response.
2.3.1.3.1.2 Accepted Presentation Context Table
Presentation Context Table - Accepted
Abstract Syntax Transfer Syntax Role Extended
Name UID Name List UID List Negotiation
Verification SOP Class 1.2.840.10008.1.1Explicit VR Little
Endian Implicit VR
Little Endian
1.2.840.10008.1.2.1
1.2.840.10008.1.2
SCP None
2.3.1.3.1.2.1 SOP Specific DICOM Conformance Statement for Verify SOP Class
The AE provides standard conformance to the Verification SOP Class as an SCP. The port number used is configured on
config screen, default is 104.
2.3.1.3.1.3 Presentation Context Acceptance Criterion
No criterion.
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
42/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
42
2.3.1.3.1.4 Transfer Syntax Selection Policies
The selected transfer syntax is based on the proposed transfer syntax list. The priority order is Explicit VR Little Endian andImplicit VR Little Endian.
2.3.1.3.2 Real-World Activity - Exam Query/Retrieve Operation
2.3.1.3.2.1 Associated Real-World Activity
If the user has initiated a retrieve by a C-MOVE-RQ, the AE will accept associations for C-STORE-RQs. The images will bestored locally.
2.3.1.3.2.2 Accepted Presentation Context Table
Abstract Syntax Transfer Syntax Role Extended
Negotiation
Name UID Name List UID List
Ultrasound ImageStorage
1.2.840.10008.5.1.4.1.1.6.1JPEG 2000 LosslessJPEG 2000 LossyJPEG Baseline
JPEG Lossless NonHierarchical (Proc 14)Run Length Encoding
Exp VR Little EndianExplicit VR Big EndianImp VR Little Endian
1.2.840.10008.1.2.4.901.2.840.10008.1.2.4.911.2.840.10008.1.2.4.50
1.2.840.10008.1.2.4.70
1.2.840.10008.1.2.5
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCP None
Ultrasound Multi-frame
Image Storage
1.2.840.10008.5.1.4.1.1.3.1JPEG 2000 LosslessJPEG 2000 LossyJPEG BaselineJPEG Lossless Non
Hierarchical (Proc 14)Run Length Encoding
Exp VR Little EndianExplicit VR Big Endian
Imp VR Little Endian
1.2.840.10008.1.2.4.901.2.840.10008.1.2.4.911.2.840.10008.1.2.4.501.2.840.10008.1.2.4.70
1.2.840.10008.1.2.5
1.2.840.10008.1.2.11.2.840.10008.1.2.2
1.2.840.10008.1.2
SCP None
Enhanced US VolumeStorage
1.2.840.10008.5.1.4.1.1.6.2 JPEG 2000 LosslessJPEG 2000 LossyJPEG BaselineJPEG Lossless NonHierarchical (Proc 14)Run Length Encoding
Exp VR Little EndianExplicit VR Big EndianImp VR Little Endian
1.2.840.10008.1.2.4.901.2.840.10008.1.2.4.911.2.840.10008.1.2.4.501.2.840.10008.1.2.4.70
1.2.840.10008.1.2.5
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCP None
Ultrasound ImageStorage (retired)
1.2.840.10008.5.1.4.1.1.6JPEG 2000 LosslessJPEG 2000 LossyJPEG Baseline
JPEG Lossless NonHierarchical (Proc 14)Run Length Encoding
Exp VR Little EndianExplicit VR Big EndianImp VR Little Endian
1.2.840.10008.1.2.4.901.2.840.10008.1.2.4.911.2.840.10008.1.2.4.50
1.2.840.10008.1.2.4.70
1.2.840.10008.1.2.5
1.2.840.10008.1.2.11.2.840.10008.1.2.21.2.840.10008.1.2
SCP None
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
43/214
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
44/214
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
45/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
45
Element Name Tag VR VM Description
Private Creator 7FE1,00xxLO 1 GEMS_Ultrasound_MovieGroup_001
This means that all private tags starting with 7FE1,xx will belong to the GEMS_Ultrasound_MovieGroup_001.
If so configured, the product will send preview image in private data elements designated by the Private Creator element:Element Name Tag VR VM Description
Private Creator 6003,00xxLO 1 GEMS_Ultrasound_ImageGroup_001
This means that all private tags starting with 6003,xx will belong to the GEMS_Ultrasound_ImageGroup_001.
The product will send measurements raw data information in private data elements designated by the Private Creator
element:
Element Name Tag VR VM Description
Private Creator 6005,00xxLO 1 GE_GROUP
This means that all private tags starting with 6005,xx will belong to the GE_GROUP.
2.6 CONFIGURATION
2.6.1 AE Title/Presentation Address Mapping
The Local AE title is configurable through the Config screen, see below.
2.6.2 Configurable Parameters
Network:
Local IP address
Local IP netmask
Default Gateway Local network speed information
Local:
Local AE Title
Local port number
Verification:
The AE Title, IP address and port number of the SCP
Max retries, Retry interval, Timeout
Modality Worklist:
The AE Title, IP address and port number of the SCP
Max retries, Retry interval, Timeout
Refresh interval - the interval between downloads from the worklist. Searching for a patient within the time of a
refresh interval does not perform a new worklist query.
Disabling/enabling and setting constant values for query fields This is a way to disable search criteria in the
worklist query or setting them to a fixed value.
Maximum number of downloaded entries
Storage:
The AE Title, IP address and port number of the SCP
Max retries, Retry interval, Timeout
8/10/2019 Gehc Dicom Conformance Logiq f Series Doc1559107 Rev2
46/214
LOGIQ F SERIES
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1462293REV 2
46
Enable/disable raw data
Frame rate reduction
Enable/disable multi-frame
Compression selections
Color support
Association strategies: one association per image or one association per exam Include structured report (SR).
Included Key Object Selection Document for Image Rejection (KO