GEN seriesGN611B
Isolated 1 kV 200 kS/sInput Card
Special features
- 6 analog channels- Isolated, balanced differential
inputs- ± 10 mV to ± 1000 V input range- Basic accuracy 0.02%- Basic power accuracy 0.02%- 600 V RMS CAT II reinforced
isolation, tested up to 6.4 kV- Analog/digital anti-alias filters- 200 kS/s sample rate- 18 bit resolution- Real-time cyclic calculators- Real-time formula database
calculators (option)- Real-time data output (option)- Triggering on real-time results
Isolated 1 kV 200 kS Input CardThe isolated balanced differential input offersvoltage ranges from ± 10 mV to ± 1000 V.Tested up to 6.4 kV, the reinforced isolationallows for safe measurements up to600 V RMS CAT II (IEC61010-1:2010 safetystandard).Optimum anti-alias protection is achieved bythe 7-pole analog anti-alias filter combined witha fixed 2 MS/s sampling Analog-to-Digitalconverter. At lower sample rates the digital antialias filters allow for a large range of high orderfilter characteristics with precise phase matchand ultra low noise output.The two Timer/Counters and the G070Atorque/RPM adapter allow for direct interfacingto HBM torque transducers or other torque andspeed sensors.
Besides the instantaneous values, the real-time cycle detection allows for cycle basedreal-time calculations like: TrueRMS on allanalog channels; torque, angle, speed on allTimer/Counter channels.The real-time formula database calculatorsoption offers math routines to solve almost anyreal-time mathematical challenge likeobtaining mechanical power and/or multi-phase (not limited to three) electric power(P, Q, S) or even efficiency calculations.Real-time calculated results can be used totrigger the recording. Every cycle based resultfrom the real-time formula database can betransferred in real-time to the EtherCAT®
output card option of the mainframe.
Dat
a sh
eet
B4375-1.2 en HBM: public
Capabilities OverviewModel GN611BMaximum sample rate per channel 200 kS/sMemory per card 200 MBAnalog channels 6Anti-alias filters Fixed bandwidth analog AA-filter combined with sample rate tracking digital AA-filterADC resolution 18 bitIsolation Channel to channel and channel to chassisInput type Analog, isolated balanced differentialPassive voltage/current probes Special designed matching probes only (e.g. Elas HDP)Sensors Not supportedTEDS Not supportedReal-time cycle based calculators 32; Cycle and Timer based calculations with triggering on calculated resultsReal-time formula database calculators (option) Extensive set of user programmable math routines
EtherCat® output Maximum 1000 updates per second, with 1 ms output latencyDigital Event/Timer/Counter 16 digital events and 2 Timer/Counter channelsStandard data streaming (up to 200 MB/s) Not supportedFast data streaming (up to 1 GB/s) SupportedSlot width 1
Block Diagram
1010
AC/DC/GND
+
-
F
A
F
n-1
n-1
z-1
z-1
Σ
a1
a2 b2
b1
x(n) y(n)
A
F(x)
T
Sample Rate
Channel &
Card Trigger
Communication& Memory &
Recording controlSynchronization
& Sample Rate
Real-Time
Calculator
Event TimerCounter
Input & Output
2.0 or 1.024 MS/s
2.0 or 1.024 MS/s
Backplane
Channel Amplifier Analog Anti-Alias Filter ADC Isolation
Digital Filter &
Sample Rate selection
System Trigger Bus
Communication and
Fast Data Streaming
Acquisition Control
Master Time Base
Event/Timer/Counter
6
/ /
Figure 1.1: Block Diagram
Note The typical and maximum specification given in this data sheet are based on respectively 1 σ (68.27%) and5 σ (99.9999%) statistical evaluations of calibration results. Cards exceeding the maximum specifications arenot released for sale. All specifications are defined at 23 °C ± 2 °C, unless specified differently.The specifications listed are valid for cards that have been calibrated and are used in the same mainframe andslots as they were at the time of calibration. When the card is removed from its original location and placed inanother slot and/or mainframe, the Offset error, Gain error and MSE specifications are expected to increase(up to double the original specification) due to thermal differences within the configurations.
HBM: public 2 B4375-1.2 en
Analog Input SectionChannels 6Connectors Fully isolated 4 mm banana plugs (plastic), 2 per channel (red and black)Input type Analog, isolated balanced differentialInput impedance 2 * 1 MΩ ± 1% // 33 pF ± 10% ranges larger than ± 5 V. All other ranges 57 pF ± 10%Input coupling
Coupling modes AC, DC, GNDAC coupling frequency 48 Hz ± 5 Hz (-3 dB)
0.1 1 10 100 1000 10000Frequency [Hz]
-50
-40
-30
-20
-10
0
Ma
gn
itu
de
[d
B]
AC coupling response [dB]
0
10
20
30
40
50
60
70
80
90
100
0.1 1 10 100 1000 10000
Ma
gn
itu
de
[%
]
Frequency [Hz]
AC coupling response [%]
Figure 1.2: Representative AC coupling responseRanges ± 10 mV, ± 20 mV, ± 50 mV, ± 0.1 V, ± 0.2 V, ± 0.5 V, ± 1 V, ± 2 V, ± 5 V, ± 10 V, ± 20 V,
± 50 V, ± 100 V, ± 200 V, ± 500 V, ± 1000 VOffset ± 50% in 1000 steps (0.1%);
± 1000 V range has fixed 0% offsetCommon mode (referred to system ground)
Ranges Less than ± 10 V Larger than or equal to ± 10 VRejection (CMR) > 80 dB @ 80 Hz (100 dB typical) > 60 dB @ 80 Hz (80 dB typical)
Maximum common mode voltage 7 V RMS 1000 V RMS
-120
-100
-80
-60
-40
-20
0
0.01 0.1 1 10 100 1000
Frequency [kHz]
Common mode response
< ± 10 V ranges
≥ ± 10 V ranges
Mag
nit
ud
e [
dB
]
Figure 1.3: Representative common mode response
B4375-1.2 en 3 HBM: public
Analog Input SectionInput overload protection
Overvoltage impedance change The activation of the overvoltage protection system results in a reduced input impedance.The overvoltage protection is not active for as long as the input voltage remains less than200% of the selected input range or 1250 V, whichever value is the smallest.
Maximum nondestructive voltage ± 2000 V DCMaximum overload without auto range 200% of selected range
Automatic auto range When overload causes the amplifier to overheat, the amplifier increases its range in stepsof a factor of 10 until the overload ceases. When the overload exceeds 1000 V, the inputsignal is disconnected and the amplifier input is grounded. When the temperature returnsto normal, the range that was originally selected is restored. The automatic auto rangecannot be turned off.
Overload recovery time Restored to 0.1% accuracy in less than 5 μs after 200% overload
Maximum Static Error (MSE)Maximum static error (MSE) Typical Maximum
All filters 0.02% of Full Scale 0.05% of Full Scale ± 10 μV
Ranges [V]Ranges [V]
Detail MSE (all filters)MSE (all filters)
Typical
Maximum
All filters Typical [%] Maximum [%]
± 0.01 V 0.0200.020
0.020
0.020
0.020
0.0200.020
0.020
0.0200.020
0.020
0.020
0.020
0.020
0.020
0.020
0.1000.075
0.060
0.055
0.053
0.0510.051
0.051
0.0510.051
0.051
0.051
0.051
0.051
0.051
0.051
± 0.02 V
± 0.05 V
± 0.1 V
± 0.2 V
± 0.5 V± 1 V
± 2 V
± 5 V± 10 V
± 20 V
± 50 V
± 100 V
± 200 V
± 500 V
± 1000 V
Err
or
[%]
Err
or
[%]
0
0.02
0.04
0.06
0.08
0.1
0.12
Typical
Maximum
0
0.01
0.02
0.03
0.04
0.05
0.06
± 0
.01
± 0
.02
± 0
.05
± 0
.1
± 0
.2
± 0
.5
± 1
± 2
± 5
± 1
0
± 2
0
± 5
0± 1
00
± 2
00
± 5
00
± 1
000
Figure 1.4: Maximum static error (MSE)
HBM: public 4 B4375-1.2 en
DC GainDC Gain error Typical Maximum
All filters 0.1% of Full Scale 0.1% of Full Scale
Maximum
Typical
0.00
0.05
0.10
Err
or
[%]
Ranges [V]
DC gain error (all filters)
± 0
.01
± 0
.02
± 0
.05
± 0
.1
± 0
.2
± 0
.5
± 1
± 2
± 5
± 1
0
± 2
0
± 5
0
± 1
00
± 2
00
± 5
00
± 1
000
All filters Typical [%] Maximum [%]
± 0.01 V 0.100.100.100.100.100.100.100.100.100.100.100.100.100.100.100.10
0.100.100.100.100.100.100.100.100.100.100.100.100.100.100.100.10
± 0.02 V± 0.05 V± 0.1 V± 0.2 V± 0.5 V± 1 V± 2 V± 5 V± 10 V± 20 V± 50 V± 100 V± 200 V± 500 V± 1000 V
Figure 1.5: DC gain errorDC gain error drift Typical Maximum
All filters ±20 ppm/°C (±12 ppm/°F) ±35 ppm/°C (±20 ppm/°F)
Ranges [V]
Dri
ft [
pp
m]
Typical
Maximum
Typical
Maximum
Dri
ft [
pp
m/°
C]
DC gain drift ± 10 V range (all filters)
0
5
10
15
20
25
30
35
40
45DC gain drift (all filters)
0/(32) 10/(50) 20/(68) 30/(86) 40/(104)
Temperature [°C/(°F)]
-1000
-800
-600
-400
-200
0
200
400
600
800
1000
± 0
.01
± 0
.02
± 0
.05
± 0
.1
± 0
.2
± 0
.5
± 1
± 2
± 5
± 1
0
± 2
0
± 5
0± 1
00
± 2
00
± 5
00
± 1
000
All filters Typical
[ppm/°C]
Maximum
[ppm/°C]
± 0.01 V± 0.02 V± 0.05 V± 0.1 V± 0.2 V± 0.5 V± 1 V± 2 V± 5 V± 10 V± 20 V± 50 V± 100 V± 200 V± 500 V± 1000 V
20202020202020202020202020202020
35353535353535353535353535353535
Figure 1.6: DC gain error drift
B4375-1.2 en 5 HBM: public
DC OffsetDC Offset error Typical Maximum
All filters 0.01% of Full Scale 0.01% of Full Scale ± 10 μV
Ranges [V]
Offset error (all filters)
0.000
0.005
0.010
0.015
Err
or
[%]
Ranges [V]
Detail offset error (all filters)
Typical
MaximumE
rro
r [%
]
Typical
Maximum
0
0.01
0.02
0.03
0.04
0.05
0.06
0.07
± 0
.01
± 0
.02
± 0
.05
± 0
.1
± 0
.2
± 0
.5
± 1
± 2
± 5
± 1
0
± 2
0
± 5
0± 1
00
± 2
00
± 5
00
± 1
000
± 0
.01
± 0
.02
± 0
.05
± 0
.1
± 0
.2
± 0
.5
± 1
± 2
± 5
± 1
0
± 2
0
± 5
0± 1
00
± 2
00
± 5
00
± 1
000
All Filters Typical [%] Maximum [%]
± 0.01 V 0.010 0.060± 0.02 V 0.010 0.035
± 0.05 V 0.010 0.020± 0.1 V 0.010 0.015
± 0.2 V 0.010 0.013± 0.5 V 0.010 0.011
± 1 V 0.010 0.011
± 2 V 0.010 0.011± 5 V 0.010 0.011± 10 V 0.010 0.011
± 20 V 0.010 0.011
± 50 V 0.010 0.011
± 100 V 0.010 0.011
± 200 V 0.010 0.011± 500 V 0.010 0.011± 1000 V 0.010 0.011
Figure 1.7: DC offset errorDC Offset drift Typical Maximum
All filters ±(30 ppm + 10 μV)/°C (±(17 ppm + 6 μV)/°F) ±(80 ppm + 10 μV)/°C (±(45 ppm + 6 μV)/°F)
-2500
-2000
-1500
-1000
-500
0
500
1000
1500
2000
2500
Ranges [V]
DC offset drift (all filters)
Dri
ft [
pp
m]
All filters ±10 V range
0/(32) 10/(50) 20/(68) 30/(86) 40/(104)
Temperature [°C/(°F)]
Typical
Maximum
Typical
Maximum
Dri
ft [
pp
m/°
C]
0
100
200
300
400
500
600
700
± 0
.01
± 0
.02
± 0
.05
± 0
.1
± 0
.2
± 0
.5
± 1
± 2
± 5
± 1
0
± 2
0
± 5
0
± 1
00
± 2
00
± 5
00
± 1
000
All filters Typical
[ppm/°C]
Maximum
[ppm/°C]
± 0.01 V 530 580± 0.02 V 280 330± 0.05 V 130 180± 0.1 V 80 130± 0.2 V 55 105± 0.5 V 40 90± 1 V 55 85± 2 V 33 83± 5 V 31 81± 10 V 31 81± 20 V 31 81± 50 V 31 81± 100 V 31 81± 200 V 31 81± 500 V 31 81± 1000 V 31 81
Figure 1.8: DC Offset drift
RMS NoiseRMS Noise (50 Ω terminated) Typical Maximum
All filters 0.01% of Full Scale ± 20 μV 0.02% of Full Scale ± 20 μV
Ranges [V]
Typical
Maximum
Typical
Maximum
No
ise
[%
]
No
ise
[%
]
Ranges [V]
RMS noise (all filters)
0
0.02
0.04
0.06
0.08
0.1
0.12
0.14Detail RMS noise (all filters)
0
0.01
0.02
0.03
0.04
0.05
± 0
.01
± 0
.02
± 0
.05
± 0
.1
± 0
.2
± 0
.5
± 1
± 2
± 5
± 1
0
± 2
0
± 5
0
± 1
00
± 2
00
± 5
00
± 1
000
± 0
.01
± 0
.02
± 0
.05
± 0
.1
± 0
.2
± 0
.5
± 1
± 2
± 5
± 1
0
± 2
0
± 5
0
± 1
00
± 2
00
± 5
00
± 1
000
Range [V] Typical [%] Maximum [%]
± 0.01 V 0.110
0.060
0.0300.020
0.015
0.0120.011
0.011
0.0110.011
0.011
0.011
0.0110.011
0.0110.011
0.120
0.070
0.0400.030
0.025
0.0220.021
0.021
0.0210.021
0.021
0.021
0.0210.021
0.0210.021
± 0.02 V
± 0.05 V± 0.1 V
± 0.2 V
± 0.5 V± 1 V
± 2 V
± 5 V± 10 V
± 20 V
± 50 V
± 100 V± 200 V
± 500 V± 1000 V
Figure 1.9: RMS noise
HBM: public 6 B4375-1.2 en
Basic Power AccuracyThe GN610B is calibrated and checked at 53 Hz voltage and current inputs using burden resistors. During calibration burden resistors areattached to three voltage channels to enable current measurements.Specifications are given for the 2.5 Ω burden. Using the 1.0 Ω or 10.0 Ω burden will give different current ranges but identical results.
2.5 Ω Burdenspans
1.264 A DC 800 mA DC 400 mA DC 160 mA DC 80 mA DC 40 mA DC
0 - 100 HzSine waveCF: 1.41
Cos Phi : 1
Burdenranges
440 mA RMS 280 mA RMS 140 mA RMS 56 mA RMS 28 mA RMS 14 mA RMS
IT200 ranges N/A 200 A RMS(1) 140 A RMS 56 A RMS 28 A RMS 14 A RMS
IT400 ranges N/A 400 A RMS(1) 280 A RMS 112 A RMS 56 A RMS 28 A RMS
IT700 ranges 700 A RMS(1) 490 A RMS 245 A RMS 98 A RMS 49 A RMS 24.5 A RMS
Voltagespans
Voltageranges
Typical Typical Typical Typical Typical Typical
40 V DC 14.1 V RMS 0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.1% range
0.02% reading+ 0.1% range
0.02% reading+ 0.15% range
100 V DC 35.3 V RMS 0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.1% range
0.02% reading+ 0.1% range
0.02% reading+ 0.15% range
200 V DC 70.7 V RMS 0.02% reading+0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.1% range
0.02% reading+ 0.1% range
0.02% reading+ 0.15% range
400 V DC 141 V RMS 0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.1% range
0.02% reading+ 0.1% range
0.02% reading+ 0.15% range
1 kV DC 353 V RMS 0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.1% range
0.02% reading+ 0.1% range
0.02% reading+ 0.15% range
2 kV DC 707 V RMS 0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.05% range
0.02% reading+ 0.1% range
0.02% reading+ 0.1% range
0.02% reading+ 0.15% range
(1) With the right setup the LEM sensor can be used for this range, however this input range is not specified in the LEM current sensordata sheet.
B4375-1.2 en 7 HBM: public
Isolation
±1000 V
RM
S
600 V
CA
T II
Rein
forc
ed
-
+
ADC
Isolated channel
±1000 V
RM
S
600 V
CA
T II
Rein
forc
ed
±2000 V
RM
S
±2000 V
RM
S
-
+
ADC
Isolated channel
Chassis
±1000 V
RM
S
600 V
CA
T II
±1000 V RMS
600 V CAT II Reinforced
±1000 V RMS
600 V CAT II Reinforced
Figure 1.10: Isolation 1kV card overview CAT II CAT IIIChannel to chassis (earth) 1000 V RMS 600 V RMS (1) 300 V RMS(1)
Channel to channel 2000 V RMS (2) (2)
(1) IEC61010-1 category voltage ratings are RMS voltages.(2) Channel to channel CAT II and CAT III ratings are not a valid method to specify.
Isolation and Input Type TestingIEC61010-1:2010 and EC61010-2-30:2010 isolation tests
Channel to channel 3510 V RMS and 4935 V DC for 5 s3260 V RMS and 4596 V DC for 1 minute
Channel to chassis 3510 V RMS and 4935 V DC for 5 s3260 V RMS and 4596 V DC for 1 minute
Channel to channel impulse 6400 V peak using a 2 Ω series resistorRise time 1.2 μs, 50% amplitude reduction in 50 μs
Channel to chassis impulse 6400 V peak using a 2 Ω series resistorRise time 1.2 μs, 50% amplitude reduction in 50 μs
50%
Upk
1 µs µs50
Time
Am
plit
ude
Figure 1.11: Example of 1.2/50 μs impulseInput impulse test
Channel positive to negative input 4000 V peak using a 12 Ω series resistor, rise time 1.2 μs, 50% amplitude reduction in50 μs
HBM: public 8 B4375-1.2 en
Analog to Digital ConversionSample rate per channel 0.1 S/s to 200 kS/sADC resolution; one ADC per channel 18 bitADC type Successive Approximation Register (SAR); Analog Devices AD7986BCPZTime base accuracy Defined by mainframe: ± 3.5 ppm; aging after 10 years ± 10 ppmBinary sample rate Supported; produces rounded BIN values when calculating FFTsMaximum binary sample rate 204.8 kS/sExternal time base frequency 0 S/s to 200 kS/sExternal time base frequency divider Divide external clock by 1 to 220
External time base level TTLExternal time base minimum pulse width 200 ns
Anti-Alias FiltersNote on phase matching channels. Every filter characteristic and/or filter bandwidth selection comes with it's own specific phase response.Using different filter selections (Bessel IIR/Butterworth IIR/etc.) or different filter bandwidths can result in phase mismatches between channels.
SAR ADC Sample Rate Selection
Analog Input
Digital Filter (Anti-Alias)
1 out of N
A
F
Analog Anti-Alias Filter
1010A
F
y(n)
n-1
n-1
Σx(n)
z-1
z-1
Figure 1.12: Combined analog and digital anti-alias filter block diagramAnti-aliasing is prevented by a steep, fixed frequency analog anti-alias filter in front of the Analog to Digital Converter (ADC). The ADC alwayssamples at a fixed sample rate. The fixed sample rate of the ADC avoids the need for different analog anti-alias filter frequencies.Directly behind the ADC, the high precision digital filter is used as anti-alias protection before the digital downsampling to the desired usersample rate is performed. The digital filter is programmed to a fraction of the user sample rate and automatically tracks any user sample rateselection. Compared to analog anti-alias filters, the programmable digital filter offers additional benefits like higher order filter with steep roll-off, a larger selection of filter characteristics, noise-free digital output and no additional phase shifts between channels that use the same filtersettings.Bessel IIR When Bessel IIR filter is selected, this is always a combination of an analog Bessel anti-
alias filter and a digital Bessel IIR filter to prevent aliasing at lower sample rates.Bessel filters are typically used when looking at signals in the time domain. They are bestused for measuring transient signals or sharp edge signals like square waves or stepresponses.
Butterworth IIR When Butterworth IIR filter is selected, this is always a combination of an analog Butterworthanti-alias filter and a digital Butterworth IIR filter to prevent aliasing at lower sample rates.This filter is best used when working in the frequency domain. When working in the timedomain, this filter is best used for signals that are (close to) sine waves.
Elliptic IIR When Elliptic IIR filter is selected, this is always a combination of an analog Butterworthanti-alias filter and a digital Elliptic IIR filter to prevent aliasing at lower sample rates.This filter is best used when working in the frequency domain. When working in the timedomain, this filter is best used for signals that are (close to) sine waves.
B4375-1.2 en 9 HBM: public
Bessel IIR Filter (Digital Anti-Alias)
δs
ωp ωs
Passband Stopband
δp : Passband ripple
δs: Stopband attenuation
ωp
ωc
ωs
: Passband frequency
: Corner frequency
: Stopband frequency
Frequency [kHz]
Magnitu
de [dB
]
ωc
1 + δp1 - δp
-3 dB
Figure 1.13: Digital Bessel IIR FilterWhen Bessel IIR filter is selected, this is always a combination of an analog Bessel anti-alias filter and a digital Bessel IIR filter.
Analog anti-alias filter bandwidth 400 kHz ± 25 kHz (-3 dB)Analog anti-alias filter characteristic 7-pole Bessel, optimal step response
Bessel IIR filter characteristic 8-pole Bessel style IIRBessel IIR filter user selection Auto tracking for sample rate divided by: 10, 20, 40, 100
The user selects a division factor from the current sample rate; software then adjusts thefilter when the sample rate is changed.
Bessel IIR filter bandwidth (ωc) User selectable from 0.4 Hz to 20 kHz
Bessel IIR 0.1 dB passband (ωp)(1) DC to 0.14 * ωc
Bessel IIR filter stopband attenuation (δs) 60 dBBessel IIR filter roll-off 48 dB/octave
100
Ma
gn
itu
de
[%
]
102.33
101.74
101.16
100.58
99.43
98.86
98.29
97.72
100
102.33
101.74
101.16
100.58
99.43
98.86
98.29
97.72
Ma
gn
itu
de
[%
]
Frequency [kHz]
Mag
nit
ud
e [
dB
]
1001010.1
± 10 V: Bessel 20 kHz Passband flatness0.2
Specification line
1001010.1
Specification line
0.1
0
0
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Mag
nit
ud
e [
dB
]
± 10 V: Bessel 20 kHz Overview
Frequency [kHz]
Mag
nit
ud
e [
dB
]
± 2 V: Bessel 20 kHz Passband flatness0.2
0.1
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Mag
nit
ud
e [
dB
]
± 2 V: Bessel 20 kHz Overview
-130
-110
-90
-70
-50
-30
-10
10
-130
-110
-90
-70
-50
-30
-10
10
0.1 1 10 100 1000
0.1 1 10 100 1000
Figure 1.14: Representative Bessel IIR examples
(1) Measured using a Fluke 5700A calibrator, DC normalized
HBM: public 10 B4375-1.2 en
Butterworth IIR Filter (Digital Anti-Alias)
Passband Stopband
Ma
gnitude
[d
B]
δp: Passband ripple
δs: Stopband attenuation
ωp:
ωc:
Passband frequency
Corner frequency
Stopband frequencyωs:ωp ωs
Frequency [kHz] ωc
1 + δp
1 - δp
-3 dB
δs
Figure 1.15: Digital Butterworth IIR FilterWhen Butterworth IIR filter is selected, this is always a combination of an analog Butterworth anti-alias filter and a digital Butterworth IIR filter.
Analog anti-alias filter bandwidth 465 kHz ± 25 kHz (-3 dB)Analog anti-alias filter characteristic 7-pole Butterworth, extended passband response
Butterworth IIR filter characteristic 8-pole Butterworth style IIRButterworth IIR filter user selection Auto tracking for sample rate divided by: 4, 10, 20, 40
The user selects a division factor from the current sample rate; software then adjusts thefilter when the sample rate is changed.
Butterworth IIR filter bandwidth (ωc) User selectable from 1 Hz to 50 kHz
Butterworth IIR 0.1 dB passband (ωp)(1) DC to 0.7 * ωc
Butterworth IIR filter stopband attenuation (δs) 75 dBButterworth IIR filter roll-off 48 dB/octave
100
Ma
gn
itu
de
[%
]
102.33
101.74
101.16
100.58
99.43
98.86
98.29
97.72
100
102.33
101.74
101.16
100.58
99.43
98.86
98.29
97.72
Ma
gn
itu
de
[%
]
Frequency [kHz]
Mag
nit
ud
e [
dB
]
1001010.1
± 10 V: Butterworth 50 kHz Passband flatness0.2
Specification line
1001010.1
Specification line
0.1
0
0
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Mag
nit
ud
e [
dB
]
± 10 V: Butterworth 50 kHz Overview
Frequency [kHz]
Mag
nit
ud
e [
dB
]
± 2 V: Butterworth 50 kHz Passband flatness0.2
0.1
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Mag
nit
ud
e [
dB
]
± 2 V: Butterworth 50 kHz Overview
-130
-110
-90
-70
-50
-30
-10
10
-130
-110
-90
-70
-50
-30
-10
10
0.1 1 10 100 1000
0.1 1 10 100 1000
Figure 1.16: Representative Butterworth IIR examples
(1) Measured using a Fluke 5700A calibrator, DC normalized
B4375-1.2 en 11 HBM: public
Elliptic IIR Filter (Digital Anti-Alias)
Passband Stopband
Frequency [kHz]
δp: Passband ripple
δs: Stopband attenuation
ωp:
ωc:
Passband frequency
Corner frequency
Stopband frequencyωs:ωp=ωc ωs
Magnitude [dB
]
1 + δp
1 - δp
δs
Figure 1.17: Digital Elliptic IIR FilterWhen Elliptic IIR filter is selected, this is always a combination of an analog Butterworth anti-alias filter and a digital Elliptic IIR filter.
Analog anti-alias filter bandwidth 465 kHz ± 25 kHz (-3 dB)Analog anti-alias filter characteristic 7-pole Butterworth, extended passband response
Elliptic IIR filter characteristic 7-pole Elliptic style IIRElliptic IIR filter user selection Auto tracking for sample rate divided by: 4, 10, 20, 40
The user selects a division factor from the current sample rate; software then adjusts thefilter when the sample rate is changed.
Elliptic IIR filter bandwidth (ωc) User selectable from 1 Hz to 50 kHz
Elliptic IIR 0.1 dB passband (ωp)(1) DC to ωc
Elliptic IIR filter stopband attenuation (δs) 75 dBElliptic IIR filter roll-off 72 dB/octave
100
Ma
gn
itu
de
[%
]
102.33
101.74
101.16
100.58
99.43
98.86
98.29
97.72
100
102.33
101.74
101.16
100.58
99.43
98.86
98.29
97.72
Ma
gn
itu
de
[%
]
Frequency [kHz]
Mag
nit
ud
e [
dB
]
1001010.1
± 10 V: Elliptic 50 kHz Passband flatness0.2
Specification line
1001010.1
Specification line
0.1
0
0
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Mag
nit
ud
e [
dB
]
± 10 V: Elliptic 50 kHz Overview
Frequency [kHz]
Mag
nit
ud
e [
dB
]
± 2 V: Elliptic 50 kHz Passband flatness0.2
0.1
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Mag
nit
ud
e [
dB
]
± 2 V: Elliptic 50 kHz Overview
-130
-110
-90
-70
-50
-30
-10
10
-130
-110
-90
-70
-50
-30
-10
10
0.1 1 10 100 1000
0.1 1 10 100 1000
Figure 1.18: Representative Elliptic IIR examples
(1) Measured using a Fluke 5700A calibrator, DC normalized
HBM: public 12 B4375-1.2 en
Channel to Channel Phase MatchUsing different filter selections (Bessel IIR/Butterworth IIR/etc.) or different filter bandwidths results in phase mismatches between channels.All specifications are typical static values and measured using a 10 kHz sine wave and 200 kS/s sample rate. < ±10V spans ≥ ±10V spans Combined spansBessel IIR, Filter frequency 20 kHz
Channels on card 0.01° (3 ns) 0.04° (13 ns) 0.27° (76 ns)GN611B Channels within mainframe 0.01° (3 ns) 0.06° (17 ns) 0.27° (76 ns)
Butterworth IIR, Filter frequency 50 kHz Channels on card 0.02° (6 ns) 0.04° (13 ns) 0.27° (76 ns)
GN611B Channels within mainframe 0.02° (6 ns) 0.06° (17 ns) 0.27° (76 ns)Elliptic IIR, Filter frequency 50 kHz
Channels on card 0.02° (6 ns) 0.04° (13 ns) 0.27° (76 ns)GN611B Channels within mainframe 0.02° (6 ns) 0.06° (17 ns) 0.27° (76 ns)
GN611B channels across mainframes Defined by synchronization method used (None, IRIG, GPS, Master/Slave, PTP)
Channel to Channel CrosstalkChannel to channel crosstalk is measured with a 50 Ω termination resistor on the input and uses sine wave signals on the channel above andbelow the channel being tested. To test Channel 2, Channel 2 is terminated with 50 Ω and Channels 1 and 3 are connected to the sine wavegenerator.
Frequency [kHz]
Ma
gn
itu
de
[d
B]
± 10 mV: Crosstalk overview
-140
-160
-180
-120
-100
-80
-60
-40
-20
0
0.01 0.1 1 10 100 1000
Figure 1.19: Representative Channel to Channel crosstalk
On-board MemoryPer card 200 MB (100 MS @ 16 bits, 50 MS @ 18 bits storage)Organization Automatically distributed amongst channels enabled for storage or real-time calculationsMemory diagnostics Automatic memory test when system is powered on but not recordingStorage sample size User selectable 16 or 18 bits
16 bits, 2 bytes/sample18 bits, 4 bytes/sample
B4375-1.2 en 13 HBM: public
Digital Event/Timer/CounterThe Digital Event/Timer/Counter input connector is located on the mainframe. For exact layout and pinning see mainframe data sheet.Digital input events 16 per card
Levels TTL input level, user programmable invert levelInputs 1 pin per input, some pins are shared with Timer/Counter inputs
Overvoltage protection ± 30 V DC continuouslyMinimum pulse width 100 nsMaximum frequency 5 MHz
Digital output events 2 per cardLevels TTL output levels, short circuit protected
Output event 1 User selectable: Trigger, Alarm, set High or LowOutput event 2 User selectable: Recording active, set High or Low
Digital output event user selectionsTrigger 1 high pulse per trigger (on every channel trigger of this card only)
12.8 µs minimum pulse width200 µs ± 1 µs ± 1 sample period pulse delay
Alarm High when alarm condition is activated, low when not activated (alarm conditions of thiscard only)200 µs ± 1 µs ± 1 sample period alarm event delay
Recording active High when recording, low when in idle or pause modeRecording active output delay of 450 ns
Set High or Low Output set High or Low; can be controlled by Custom Software Interface (CSI) extensions;delay depends on specific software implementation
Timer/Counter 2 per cardLevels TTL input levelsInputs 3 pins: signal, reset and direction
All pins are shared with digital event inputsInput coupling Uni-directional, Bi-directional and ABZ incremental encoder (Quadrature)
Measurement modes Count, Angle, Frequency and RPM
HBM: public 14 B4375-1.2 en
Input Coupling Uni- and Bi-directionalUni- and bi-directional input coupling is used when the direction signal is a stable signal.
Signal
∆w ∆w
∆s ∆h
∆s ∆hReset
Direction
Figure 1.20: Uni- and Bi-directional timingInputs 3 pins: signal, reset and direction (only used in bi-directional count)Maximum input signal frequency 5 MHz
Minimum pulse width (Δw) 100 nsReset input
Level sensitivity User selectable invert levelMinimum setup time prior to signal edge (Δs) 100 ns
Minimum hold time after signal edge (Δh) 100 nsReset options
Manual Upon user request by software commandStart recording Count value set to 0 at Start of recording
First reset pulse After the recording is started, the first reset pulse sets the counter value to 0. The next resetpulses are ignored.
Each reset pulse On each external reset pulse, the counter value is reset to 0.Direction input
Input Level sensitivity Only used when in bi-directional modeLow: increment counter/positive frequencyHigh: decrement counter/negative frequency
Minimum setup time prior to signal edge (Δs) 100 nsMinimum hold time after signal edge (Δh) 100 ns
B4375-1.2 en 15 HBM: public
Input Coupling ABZ Incremental Encoder (Quadrature)Typically used for tracking rotating/moving devices using a decoder with two signals that are always 90 degree phase shifted. E.g. allow fordirect interfacing to HBM torque and speed transducers.
∆t ∆t∆t ∆t
Signal
1
1
1 12 23 34 45 56 67 78 89 910 1011 11
12 23 34 45 56 6
12 23 34 4
Direction
Single precision counting
Signal
Direction
Double precision counting
Signal
Direction
Quad precision counting
Wheel rotates clock wise Wheel rotates counter clock wise
Wheel rotates clock wise Wheel rotates counter clock wise
Wheel rotates clock wise Wheel rotates counter clock wise
SignalReset
Quadrature disk
Signal
DirectionDirection
∆t: Must be > 100 ns
Figure 1.21: Bi-directional quadrature count modesInputs 3 pins: signal, direction and resetMaximum input signal frequency 2 MHz
Minimum pulse width 200 ns (2 * Δt)Minimum setup time 100 ns (Δt)Minimum hold time 100 ns (Δt)
Accuracy Single (X1), dual (X2) or quad (X4) precisionInput coupling ABZ incremental encoder (Quadrature)Reset input
Level sensitivity User selectable invert levelMinimum setup time prior to signal edge (Δt) 100 ns
Minimum hold time after signal edge (Δt) 100 nsReset options
Manual Upon user request by software commandStart recording Count value set to 0 at Start of recording
First reset pulse After the recording is started, the first reset pulse sets the counter value to 0. The next resetpulses are ignored.
Each reset pulse On each external reset pulse, the counter value is reset to 0.
HBM: public 16 B4375-1.2 en
Measurement Mode AngleIn angle measurement mode the counter will use a user defined maximum angle and revert back to zero when this count value is reached.Using the reset input the measured angle can be synchronized to the mechanical angle. The real-time calculators can extract the RPM fromthe measured angle independent from the mechanical synchronization.Angle options
Reference User selectable. Enables the use of the reset pin to reference the mechanical angle to themeasured angle
Angle at reference point User defined to specify mechanical reference pointReset pulse Angle value is reset to user defined "angle at reference point" value
Pulses per rotation User defined to specify the encoder/count resolutionMaximum pulses per rotation 32767
Maximum RPM 30 * sample rate (Example: Sample rate 10 kS/s means maximum 300 k RPM)
Measurement Mode Frequency/RPMUsed to measure any kind of frequency like engine RPM, or active sensors with proportional frequency output signal.Accuracy 0.1%, when using a gate measuring time of 40 μs or more.
With lower gate measuring times, the real-time calculators or Perception formula databasecan be used to enlarge the measuring time and improve the accuracy more dynamicallye.g. based on measured cycles.
Gate measuring time Sample period (1 / sample rate) to 50 s. Minimum gate measuring time is 50 ns.Can be selected by user to control update rate independent of sample rate
Measurement Mode Uni- and Bi-directional CountCounter mode is typically used for tracking movement of device under test. When possible use the quadrature modes as these are less sensitiveto counting errors.Counter range 0 to 231; uni-directional count
-231 to +231 - 1; bi-directional count
B4375-1.2 en 17 HBM: public
TriggeringChannel trigger/qualifier 1 per channel; fully independent per channel, software selectable either trigger or qualifierPre- and post-trigger length 0 to full memoryMaximum trigger rate 400 triggers per secondMaximum delayed trigger 1000 seconds after a trigger occurredManual trigger (Software) SupportedExternal Trigger In
Selection per card User selectable On/OffTrigger In edge Rising/Falling mainframe selectable, identical for all cards
Minimum pulse width 500 nsTrigger In delay ± 1 µs + maximum 1 sample period (identical for decimal and binary time base)
Send to External Trigger Out User can select to forward External Trigger In to the External Trigger Out BNCExternal Trigger Out
Selection per card User selectable On/OffTrigger Out level High/Low/Hold High; mainframe selectable, identical for all cards
Trigger Out pulse width High/Low: 12.8 µsHold High: Active from first mainframe trigger to end of recordingPulse width created by mainframe; For details, please refer to the mainframe datasheet
Trigger Out delay Selectable (10 μs to 516 μs) ± 1 µs + maximum 1 sample period using decimal time baseSelectable (9.76 μs to 504 μs) ± 1 µs + maximum 1 sample period using binary time baseDefault 516 (504) μs for decimal (binary) time base, compatible with standard behavior.Minimum selectable delay is the smallest delay available for all acquisition cards used withinthe mainframe
Cross channel triggeringMeasurement channels Logical OR of triggers from all measured signals
Logical AND of qualifiers from all measured signalsCalculated channels Logical OR of triggers from all calculated signals (RTC and RT-FDB)
Logical AND of qualifiers from all calculated signals (RTC and RT-FDB)Analog channel trigger levels
Levels Maximum 2 level detectorsResolution 16 bit (0.0015%) for each level
Direction Rising/Falling; single direction control for both levels based on selected modeHysteresis 0.1 to 100% of Full Scale; defines the trigger sensitivity
Analog channel trigger modesBasic POS or NEG crossing; single level
Dual level One POS and one NEG crossing; two individual levels, logical ORAnalog channel qualifier modes
Basic Above or below level check. Enable/Disable trigger with single levelDual (level) Outside or within bounds check. Enable/Disable trigger with dual level
Event channel triggerEvent channels Individual event trigger per event channel
Levels Trigger on rising edge or trigger on falling edgeQualifiers Active High or Active Low for every event channel
HBM: public 18 B4375-1.2 en
Alarm OutputSelection per card User selectable On/OffAlarm modes Basic or Dual
Basic Above or below level checkDual (level) Outside or within bounds check
Alarm levelsLevels Maximum 2 level detectors
Resolution 16 bit (0.0015%) for each levelAlarm output Active during valid alarm condition, output supported through mainframeAlarm output delay 515 µs ± 1 µs + maximum 1 sample period using decimal time base
503 µs ± 1 µs + maximum 1 sample period using binary time base
Real-time Statstream®
Patent Number : 7,868,886Real-time extraction of basic signal parameters.Supports real-time live scrolling and scoping waveform displays as well as real-time meters while recording.During recording reviews, it enhances speed for displaying and zooming extremely large recordings and it reduces the calculation time forstatistical values on large data sets.Analog channels Real-time extraction of Maximum, Minimum, Mean, Peak to Peak, Standard Deviation and
RMS valuesEvent/Timer/Counter channels Real-time extraction of Maximum, Minimum and Peak to Peak values
B4375-1.2 en 19 HBM: public
Real-Time Cycle Based Calculators
N
L H
L2
L1
Cycle Based Calculator
Cycle
Timer Cycle Source
Cycle CountLevel Crossing
Cycle Detect
F(x)Trigger Detector
Source
Selectable
Math To Channel &
Card Trigger
Measured Channels
Calculated
Channels
Figure 1.22: Real-time cycle based calculatorsCycle Source Determines the periodic real-time calculation speed by either setting a timer or using a real-
time cycle detectCycle Source: Timer
Timer duration 1.0 ms (1 kHz) to 60 s (0.0167 Hz)Cycle Source: Cycle detect
Level crossing Real-time monitors one input channel using a signal level, hysteresis and direction todetermine the cyclic nature of the signal
Cycle count Sets the counted number of cycles used for periodic calculation output
Cycle period(1) Maximum Cycle period that can be detected: 0.25 s (4 Hz)Minimum Cycle period that can be detected: 0.91 ms (1.1 kHz)Calculations are stopped when the Cycle period exceeds its maximum Cycle period(0.25 s).Cycle count is temporarily increased when Cycle period becomes shorter than minimumCycle period (0.91 ms).Time event notifications in the channel data indicate when the Cycle period has beenexceeded or when the automatic Cycle count is increased
Cycle based calculatorNumber of calculators 32
DSP load Each calculator can perform 1 calculation. Not every calculation uses the same DSP power.Selecting a calculation with the highest computation power could result in a reduction in thetotal number of calculators. Different combinations require different computation power. Theeffects of selected combinations is reflected in Perception software
Cycle Source calculations Cycle and FrequencyAnalog channel calculations RMS, Minimum, Maximum, Mean, Peak-to-Peak, Area and Energy
Timer/Counter channel calculations Frequency (to enable triggering), RPM of AngleCycle Square wave signal, 50% duty cycle.
Represent Cycle Source; rising edge indicates start of new calculation periodFrequency Detected cycle interval is converted to a frequency (1/cycle time of input signal)
Trigger detectorNumber of detectors 32; One per real-time calculator
Trigger level Defined by the user for each detector. Generates trigger when the calculated signal crossesthe level
Trigger output delay Triggers are delayed by 100 ms on calculated signals. The trigger time is corrected internallyso that the sweep triggering is correct. An additional pre-trigger length of 100 ms is addedto enable the trigger time correction. This reduces the maximum sweep length by 100 ms
(1) Cycle period range depends on signal wave shape and hysteresis setting. Specified for Sine wave with 25% Full Scale hysteresis.
HBM: public 20 B4375-1.2 en
Real-time Formula Database Calculators (Option to be ordered separately)The real-time formula database (RT-FDB) option offers an extensive set of math routines to enable almost any real-time mathematical challenge.The database structure enables the user to define a list of mathematical equations similar to the Perception review formula database.
N
L H
F
n-1
n-1
z-1
z-1
Σ
a1
a2 b2
b1
x(n) y(n)
A
Cycle Based Calculator
Cycle
Calculated Channel
Calculated Channel
CycleInterval Cycle Source
Cycle CountLevel Crossing
CycleDetect / CycleEvent
Recording
Memory
Channel ADC Data
F(x)
Source 1
Source N
Programmable
Math
Sample Based Calculator
F(x)
Source 1
Source N
Programmable
Math
Source HW Filter
Figure 1.23: Real-time formula database (RT-FDB) calculatorsThe real-time formula database supports the following list of calculations (Details of each calculation are described in the manual).Operation Sample based results
synchronousCycle based results
asynchronousStorage in PNRF
recordingPublished onEtherCAT®
Basic calculations+ (add) (1) (2)
- (subtract) (1) (2)
* (multiply) (1) (2)
/ (divide) (1) (2)
Enhanced calculationsAbs (1) (2)
Atan (1) (2)
Atan2 (1) (2)
Cosine (1) (2)
DegreesToRadians (1) (2)
Min (1) (2)
Max (1) (2)
Modulo (1) (2)
RadiansToDegrees (1) (2)
Sine (1) (2)
Sqrt (1) (2)
Tan (1) (2)
B4375-1.2 en 21 HBM: public
Real-time Formula Database Calculators (Option to be ordered separately)Operation Sample based results
synchronousCycle based results
asynchronousStorage in PNRF
recordingPublished onEtherCAT®
Cycle based calculationsCycleArea
CycleBusDelay
CycleCount
CycleCrestFactor
CycleEnergy
CycleFundamentalPhase (3)
CycleFundamentalRMS
CycleFrequency
CycleMax
CycleMean
CycleMin
CyclePeak2Peak
CyclePhase
CycleRMS
CycleRPM
CycleSampleCount
CycleTHD(3) (3)
Cycle sourceCycleDetect(5)
CycleEvent
CycleInterval
Signal filteringHWFilter(5) (1)
Signal transformationDQZeroTransformation(Park)(4)
(1) (2)
SpaceVectorTransformation(4) (1)
SpaceVectorInverseTransformation(4)
(1)
Signal generationSineWave (1)
Ramp (1)
(1) Perception V7.0 to V7.14 are able to display the sample based results in the "Live" display. The sample based results are stored in thePNRF file. During recording use a "review" while recording display to monitor the results.In user mode "Dual" Perception V7.0 to V7.14 calculate and store the sample based results for the sweep (high rate) data, samplebased results for the continuous (low rate) data are not calculate/stored.
(2) GEN DAQ systems can only publish cycle based results to EtherCAT®. To publish recorded channel data or sample based results, usee.g. CycleRMS or CycleMean to extract cycle based results that can be published to EtherCAT®.
(3) The time required to calculate the output depends on maximum cycle length and sample rate. Depending on the selected settings theEtherCAT® output latency will increase. HBM refers to these calculations as not deterministic. All EtherCAT® published values (deter-ministic and/or not deterministic) will always have the same latency.
(4) This formula is only available if the eDrive license is added to Perception.(5) The output of HWFilter is used for CycleDetect.
HBM: public 22 B4375-1.2 en
Acquisition ModesSingle sweep Triggered acquisition to on-board memory without sample rate limitations; for single
transients or intermittent phenomena. No aggregate sample rate limitations.Multiple sweeps Triggered acquisition to on-board memory without sample rate limitations; for repetitive
transients or intermittent phenomena. No aggregate sample rate limitations.Continuous Direct storage to PC or mainframe controlled hard disk without file size limitations; triggered
or un-triggered; for long duration recorder type applications. Aggregate sample ratelimitations depend on Ethernet speed, PC used and data storage media used.
Dual Combination of Multiple sweeps and Continuous; recorder type streaming to hard disk withsimultaneously triggered sweeps in on-board memory. Aggregate sample rate limitationsdepend on Ethernet speed, PC used and data storage media used.In Dual mode the RT-FDB calculators sample based results are only calculated for thesweep sections of the recorded data. Due to the asynchronous nature of cycle based results,all cycle based results are continuously stored and used in both the sweep as well as thecontinuous sections of the recording.
Acquisition Mode Details16 Bit Resolution
Recording ModeSingle Sweep
Multiple Sweeps Continuous Dual Rate
Enabled channels Enabled channels Enabled channels
1 Ch 6 Ch6 Ch &events 1 Ch 6 Ch
6 Ch &events 1 Ch 6 Ch
6 Ch &events
Max. sweep memory 100 MS 16 MS 14 MS not used 80 MS 13 MS 11 MSMax. sweep sample rate 200 kS/s not used 200 kS/sMax. continuous FIFO not used 100 MS 16 MS 14 MS 18 MS 3 MS 2.5 MSMax. continuous samplerate not used 200 kS/s Sweep sample rate / 2
Max. aggregatecontinuous streaming rate not used
0.2 MS/s 1.2 MS/s 1.4 MS/s 0.2 MS/s 1.2 MS/s 1.4 MS/s0.4 MB/s 2.4 MB/s 2.8 MB/s 0.4 MB/s 2.4 MB/s 2.8 MB/s
18 Bit Resolution
Recording ModeSingle Sweep
Multiple Sweeps Continuous Dual Rate
Enabled channels Enabled channels Enabled channels
1 Ch 6 Ch
6 Ch &events &Timer/
Counter 1 Ch 6 Ch
6 Ch &events &Timer/
Counter 1 Ch 6 Ch
6 Ch &events &Timer/
CounterMax. sweep memory 50 MS 8 MS 5 MS not used 40 MS 6.5 MS 4 MSMax. sweep sample rate 200 kS/s not used 200 kS/sMax. continuous FIFO not used 50 MS 8 MS 5 MS 9 MS 1.5 MS 1 MSMax. continuous samplerate not used 200 kS/s Sweep sample rate / 2
Max. aggregatecontinuous streaming rate not used
0.2 MS/s 1.2 MS/s 1.8 MS/s 0.2 MS/s 1.2 MS/s 1.8 MS/s0.8 MB/s 4.8 MB/s 7.2 MB/s 0.8 MB/s 4.8 MB/s 7.2 MB/s
B4375-1.2 en 23 HBM: public
Single SweepPre-trigger segment 0% to 100% of selected sweep length
If trigger occurs before the pre-trigger segment is recorded, the pre-trigger segment istruncated to recorded data only.
Delayed trigger Maximum 1000 seconds after a trigger occurred. The sweep is recorded immediately aftera delayed trigger time with 100% post-trigger after this time point.
Sweep stretch User selectable On/OffWhen enabled, any new trigger event occurring in the post-trigger segment of the sweeprestarts the post-trigger length. If, upon the detection of a new trigger, the extended post-trigger does not fit within the sweep memory, sweep stretch does not happen. The maximumsweep stretch rate is 1 sweep stretch per 2.5 ms.
Multiple SweepsPre-trigger segment 0% to 100% of selected sweep length
If trigger occurs before the pre-trigger segment is recorded, the pre-trigger segment istruncated to recorded data only.
Delayed trigger Maximum 1000 seconds after a trigger occurred. The sweep is recorded immediately aftera delayed trigger time with 100% post-trigger after this time point.
Maximum number of sweeps 200 000 per recordingMaximum sweep rate 400 sweeps per secondSweep re-arm time Zero re-arm time, sweep rate limited to 1 sweep per 2.5 msSweep stretch User selectable On/Off
When enabled, any new trigger event occurring in the post-trigger segment of the sweeprestarts the post-trigger length. If, upon the detection of a new trigger, the extended post-trigger does not fit within the sweep memory, sweep stretch does not happen. The maximumsweep stretch rate is 1 sweep stretch per 2.5 ms.
Sweep storage Sweep storage is started immediately after the trigger for this sweep has been detected.Sweep memory becomes available for reuse as soon as storage of the entire sweep for allenabled channels of this card has been completed. Sweeps are stored one by one, startingwith the first recorded sweep.
Sweep storage rate Determined by the total number of selected channels and mainframes, mainframe type,Ethernet speed, PC storage medium and other PC parameters. For details, please refer tothe mainframe datasheet.
Exceeding sweep storage rate Trigger event markers are stored in a recording. No sweep data is stored. New sweep datais recorded as soon as enough internal memory is available to capture a full sweep whena trigger occurs.
ContinuousContinuous modes supported Standard, Circular recording, Specified time and Stop on trigger
Standard User starts and stops recording. Recording is stopped when the storage media is fullCircular recording User specified recording history on storage media. All recorded data is stored on the storage
media as quickly as possible. As soon as the selected history time is reached, older recordeddata is overwritten. Recording can be stopped by the user or any system trigger.
Specified time Recording is stopped after the time specified or when the storage media is fullStop on trigger Recording is stopped after any system trigger or when the storage media is full
Continuous FIFO memory Used by enabled channels to optimize the continuous streaming rateMaximum recording time Until storage media filled or user selected time or unlimited when using circular recordingMaximum aggregate streaming rate permainframe
Determined by mainframe, Ethernet speed, PC storage medium and other PC parameters.For details, please refer to the mainframe datasheet
Exceeding aggregate streaming rate When a streaming rate higher than the aggregate streaming rate of the system is selected,the continuous memory acts as a FIFO. As soon as this FIFO fills up, the recording issuspended (no data is recorded temporarily). During this period, the internal FIFO memoryis transferred to a storage medium. When internal memory is completely empty again, therecording is automatically resumed. User notifications are added to the recording file forpost recording identification of storage overrun.
HBM: public 24 B4375-1.2 en
DualDual Sweep SpecificationPre-trigger segment 0% to 100% of selected sweep length
If trigger occurs before the pre-trigger segment is recorded, the pre-trigger segment istruncated to recorded data only.
Delayed trigger Maximum 1000 seconds after a trigger occurred. The sweep is recorded immediately aftera delayed trigger time with 100% post-trigger after this time point.
Maximum number of sweeps 200 000 per recordingMaximum sweep rate 400 sweeps per secondSweep re-arm time Zero re-arm time, sweep rate limited to 1 sweep per 2.5 msSweep stretch User selectable On/Off
When enabled, any new trigger event occurring in the post-trigger segment of the sweeprestarts the post-trigger length. If, upon the detection of a new trigger, the extended post-trigger does not fit within the sweep memory, sweep stretch does not happen. The maximumsweepstretch rate is 1 sweep stretch per 2.5 ms.
Sweep storage In dual mode, the storage of the continuous data is prioritized above the storage of thesweep data. If enough storage rate is available, the sweep storage is started immediatelyafter the trigger for this sweep has been detected. Sweep memory becomes available forreuse as soon as storage of the entire sweep for all enabled channels of this card has beencompleted. Sweeps are stored one by one, starting with the first recorded sweep.
Sweep storage rate Determined by the continuous sample rate, total number of channels and mainframes,mainframe type, Ethernet speed, PC storage medium and other PC parameters. For details,please refer to mainframe datasheet.
Exceeding sweep storage rate Continuous recorded data is not stopped, trigger event markers are stored in recording andno new sweep data is stored. A new sweep is recorded as soon as enough internal memoryis available to capture a full sweep when a trigger occurs.
Dual Continuous SpecificationsContinuous FIFO memory Used by enabled channels to optimize the continuous streaming rateMaximum recording time Until storage media filled or user selected timeMaximum aggregate streaming rate permainframe
Determined by mainframe, Ethernet speed, PC storage medium and other PC parameters.For details, please refer to the mainframe datasheet.When the average aggregate streaming rate is exceeded, the sweep storage speed isautomatically reduced to increase the aggregate streaming rate until the sweep storage isstopped completely.
Exceeding aggregate storage rate When a streaming rate higher than the aggregate streaming rate of the system is selected,the continuous memory acts as a FIFO. As soon as this FIFO fills up, the recording issuspended (no data is recorded temporarily). During this period, the internal FIFO memoryis transferred to the storage medium. When the internal memory (Continuous and Sweepmemory) is completely empty, the recording is automatically resumed. User notificationsare added to the recording file for post recording identification of storage overrun.
B4375-1.2 en 25 HBM: public
Environmental SpecificationsTemperature Range
Operational 0 °C to +40 °C (+32 °F to +104 °F)Non-operational (Storage) -25 °C to +70 °C (-13 °F to +158 °F)
Thermal protection Automatic thermal shutdown at 85 °C (+185 °F) internal temperatureUser warning notifications at 75 °C (+167 °F)
Relative humidity 0% to 80%; non-condensing; operationalProtection class IP20Altitude Maximum 2000 m (6562 ft) above sea level; operationalShock: IEC 60068-2-27
Operational Half-sine 10 g/11 ms; 3-axis, 1000 shocks in positive and negative directionNon-operational Half-sine 25 g/6 ms; 3-axis, 3 shocks in positive and negative direction
Vibration: IEC 60068-2-64Operational 1 g RMS, ½ h; 3-axis, random 5 to 500 Hz
Non-operational 2 g RMS, 1 h; 3-axis, random 5 to 500 HzOperational Environmental Tests
Cold test IEC 60068-2-1 Test Ad -5 °C (+23 °F) for 2 hoursDry heat test IEC 60068-2-2 Test Bd +40 °C (+104 °F) for 2 hours
Damp heat test IEC 60068-2-3 Test Ca +40 °C (+104 °F), humidity > 93% RH for 4 daysNon-Operational (Storage) Environmental Tests
Cold test IEC 60068-2-1 Test Ab -25 °C (-13 °F) for 72 hoursDry heat test IEC 60068-2-2 Test Bb +70 °C (+158 °F) humidity < 50% RH for 96 hours
Change of temperature testIEC 60068-2-14 Test Na
-25 °C to +70 °C (-13 °F to +158 °F)5 cycles, rate 2 to 3 minutes, dwell time 3 hours
Damp heat cyclic testIEC 60068-2-30 Test Db variant 1
+25 °C/+40 °C (+77 °F/+104 °F), humidity > 95/90% RH6 cycles, cycle duration 24 hours
Harmonized Standards for CE Compliance, According to the Following DirectivesLow Voltage Directive (LVD): 2014/35/EUElectroMagnetic Compatibility Directive (EMC): 2014/30/EUElectrical SafetyEN 61010-1 (2010) Safety requirements for electrical equipment for measurement, control, and laboratory use - General requirementsEN 61010-2-030 (2010) Particular requirements for testing and measuring circuitsElectromagnetic CompatibilityEN 61326-1 (2013) Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirementsEmissionEN 55011 Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of
measurementConducted disturbance: class B; Radiated disturbance: class A
EN 61000-3-2 Limits for harmonic current emissions: class DEN 61000-3-3 Limitation of voltage changes, voltage fluctuations and flicker in public low voltage supply systemsImmunityEN 61000-4-2 Electrostatic discharge immunity test (ESD);
contact discharge ± 4 kV/air discharge ± 8 kV: performance criteria BEN 61000-4-3 Radiated, radio-frequency, electromagnetic field immunity test;
80 MHz to 2.7 GHz using 10 V/m, 1000 Hz AM: performance criteria AEN 61000-4-4 Electrical fast transient/burst immunity test
Mains ± 2 kV using coupling network. Channel ± 2 kV using capacitive clamp: performance criteria BEN 61000-4-5 Surge immunity test
Mains ± 0.5 kV/± 1 kV Line-Line and ± 0.5 kV/± 1 kV/± 2 kV Line-earth Channel ± 0.5 kV/± 1 kV using couplingnetwork: performance criteria B
EN 61000-4-6 Immunity to conducted disturbances, induced by radio-frequency fields150 kHz to 80 MHz, 1000 Hz AM; 10 V RMS @ mains, 3 V RMS @ channel, both using clamp: performancecriteria A
EN 61000-4-11 Voltage dips, short interruptions and voltage variations immunity testsDips: performance criteria A; Interruptions: performance criteria C
HBM: public 26 B4375-1.2 en
KAB290: Shielded 2 Wire 600 V RMS CAT II Cable (Option, to be ordered separately)This cable is specially designed to be used with the GN610, GN611 and GN610B, GN611B cards. Significantly reduces signal disturbancepickup by using two identical signal wires with earthed shield. This cable must not be used for three wire measurements. The shield is not astandard signal wire.Cable setup 2 wires with shield and isolation.
Signal wires terminated on both sides using (red, black) shrouded banana plugs.Shield connected on one side using (yellow) shrouded banana plug.Note: First cable shipments have shield connected on both sides.
Isolation
Wire to wire
Shield to wire
Measurement
wires
Shield
Shield wire
Cable
isolation
Figure 1.24: Two wire shielded cable setupMaximum current 10 A RMSWire thickness 2 * 0.75 mm2 (0.00116 in2)Maximum wire resistance 25 mΩ / m (7.6 mΩ / ft) ± 5%Weight Approximately 143 g/m (1.54 oz/foot)Outside cable diameter Approximately 9 mm (0.354 inch)Minimum bend radius 10 times that of the cable diameterIsolation
Resistance 20 MΩ / km (32.19 MΩ/ mile)Voltage 600 V RMS CAT II; wire to wire; wire to shield; shield to outside
CapacitanceWire to wire 130 pF/m (39.6 pF/ft) ± 10%
Wire to shield 200 pF/m (61 pF/ft) ± 10%Temperature range
Operational -15 °C (+5 °F) to +80 °C (+176 °F)Non-operational (storage) -40 °C (-40 °F) to +80 °C (+176 °F)
Available lengths 1.5 m (4.92 ft), 3.0 m (9.84 ft), 6.0 m (19.7 ft)
Frequency [MHz]
Mag
nit
ud
e [
% o
f F
S]
Typical conducted immunity disturbance (10 V RMS, Clamp)
0
0.5
1
1.5
2
2.5
3
3.5
0.1 1 10 100
Frequency [MHz]
Mag
nit
ud
e [
dB
]
Typical conducted immunity disturbance (10 V RMS, Clamp)
-70
-60
-50
-40
-30
-20
-10
0
0.1 1 10 100
2 Lead Wire
Coax Cable
1-KAB290-xx
2 Lead Wire
Coax Cable
1-KAB290-xx
Figure 1.25: Typical conducted immunity, tested using ± 10 V range
B4375-1.2 en 27 HBM: public
KAB2128: Shielded 3 Wire 600 V RMS CAT II Cable (Option, to be ordered separately)This cable is specially designed to be used with the GN610B and GN611B cards. Significantly reduces signal disturbance pickup by using threeidentical signal wires with earthed shield.Cable setup 3 wires with shield and isolation.
Signal wires terminated on both sides using (brown, grey, black) shrouded banana plugs.Shield connected on one side using (yellow) shrouded banana plug.
Isolation
Wire to wire
Shield to wire
Measurement
wires
Shield
Shield wireCable
isolation
Figure 1.26: Three wire shielded cable setupMaximum current 1 A RMSWire thickness AWG19 0.65 mm2 (0.001 in2)Maximum wire resistance 25.4 mΩ / m (8.0 mΩ / ft) ± 5%Weight Approximately 155 g/m (1.67 oz/foot)Outside cable diameter Approximately 9.1 mm (0.36 inch)Minimum bend radius 10 times that of the cable diameterIsolation
Resistance 20 MΩ / km (32.19 MΩ/ mile)Voltage 600 V RMS CAT II; wire to wire; wire to shield; shield to outside
CapacitanceWire to wire 110 pF/m (39.6 pF/ft) ± 10%
Wire to shield 140 pF/m (61 pF/ft) ± 10%Temperature range
Operational -15 °C (+5 °F) to +80 °C (+176 °F)Non-operational (storage) -40 °C (-40 °F) to +80 °C (+176 °F)
Available lengths 1.5 m (4.92 ft), 3.0 m (9.84 ft), 6.0 m (19.7 ft),12 m (39.37 ft), 20 m (65.62 ft)
Frequency [MHz]
Mag
nit
ud
e [
% o
f F
S]
Typical conducted immunity disturbance (10 V RMS, Clamp)
0
0.5
1
1.5
2
2.5
3
3.5
0.1 1 10 100
Frequency [MHz]
Mag
nit
ud
e [
dB
]
Typical conducted immunity disturbance (10 V RMS, Clamp)
-70
-60
-50
-40
-30
-20
-10
0
0.1 1 10 100
2 Lead Wire
Coax Cable
1-KAB2128-xx
2 Lead Wire
Coax Cable
1-KAB2128-xx
Figure 1.27: Typical conducted immunity, tested using ± 10 V range
HBM: public 28 B4375-1.2 en
G068: Artificial Star Adapter (Option, to be ordered separately)The artificial star adapter creates an artificial star point to measure 3 phase signalsMaximum input voltage 1000 V DC (707 V RMS) between each of the phasesInputs 3; 4 mm safety banana plugsOutputs 6; 4 mm safety banana pins; plugs straight into GN610/GN611/GN610B/GN611B cardsArtificial star N Reference plug only. Not to be used as inputSafety Compliant with IEC61010-1 600 V RMS CAT IIApplication use The 3 phase signals L1, L2 and L3 can be connected with inputs L1, L2, L3 of the artificial
star adapter. The connection N* is the voltage present on the artificial “star point”.
L1
L1
out
L2
out
L3
out
L2 L3
N*
300kΩ
300kΩ
300kΩ
250pF
250pF
250pF
Figure 1.28: Electrical schematicWeight 170 g (6 oz)Material housing Polyurethane, vacuum resin castingSetup Two boxes can be plugged into a single GN610/GN611/GN610B/GN611B card
Two or more GN610/GN611/GN610B/GN611B cards with Artificial star adapters fit next toeach other
Temperature rangeOperational temperature 0 °C to +40 °C (+32 °F to +104 °F)
Non-operational (storage) -25 °C to +70 °C (-13 °F to +158 °F)
1-G068-2
Artificial Star
R=300kΩ
L1 L2 L3 IN
L1
Umax L1L2 = L2L3 = Umax L3L 1≤ 1000V DC (707V RMS)
L2 L3 OUT
N*
C=250pFIEC61010CAT II 600V
CAT II 600V
1kV DC
L1
L2
L3
N*
19.0 mm
(0.74”)
24.0 mm
(0.94”)
17.0
mm
(0.6
6”)
16.0 mm
(0.62”)
106.7
mm
(4.2
0”)
101.9
mm
(4.0
1”)
Figure 1.29: Artificial star adapter
B4375-1.2 en 29 HBM: public
Artifical Star Adapter Wiring Diagram
DC-AC
Power
Converter
Electric
Motor
Phase to
Artificial Star
Voltages
L1 Out
L2 Out
L3 Out
N*
L1
L1
L2
L3
L2
L3
Warning : N* (Artificial Star Point)
is not ground or earth.
This pin should be left unconnected.
Figure 1.30: Three phase representative use of artificial star adapter
DC-AC
Power
Converter
Electric
Motor
Phase to
Artificial Star
Voltages
Phase to
Artificial Star
Voltages
L3 Out
L2 Out
L1 Out
L4 Out
L5 Out
L1
L1
L2
L3
L4
L5
L2
L3
L4
L5
N*
N*Warning : N* (Artificial Star Point)
is not ground or earth.
Only connect the both N* pins together
to create one artificial star point.
Connect two N* Star Points together
Figure 1.31: Five or more phase representative use of dual star adapter
HBM: public 30 B4375-1.2 en
Gxxx: 1 kV RMS and DC CAT IV Differential Probe (Option, to be ordered separately)1 kV RMS/DC CAT IV, 4 MΩ, 10:1, typical 0.1% high precision, differential probe to be used in combination with GN610B and GN611B acquisitioncards.The probe is optimized to match the 33 pF input capacitance when using the ranges ± 10 V, ± 20 V, ± 50 V, ± 100 V, ± 200 V, ± 500 V,± 1000 V. In the lower acquisition card ranges passband attenuation exceeds the specified amplitude response.
-
+
ADCUoutUin
Isolated ChannelDifferential Probe
Figure 1.32: Block diagram and imageMaximum input to earth voltageIEC 61010-031:2015
1000 V RMS or DC CAT IV1500 V DC CAT III
Divider ratio 10:1Measurement ranges ± 100 V up to ± 1500 V (1 kV RMS) (common mode < 1000 V RMS)In-accuracy including GN61xB card ± 0.2% @ 1000 V, 50 Hz, 25 °C (0.1% typical)Input impedance 4 MΩ ± 0.2%Temperature coefficient ± 25 ppm / oC (± 45 ppm / oF)Bandwidth -0.3 dB @ 100 kHz, phase match 1o
Input cable/pins High voltage insulated 4 mm safety connector (similar to Stäubli Electrical Cable MCXHM-5000)
Output pins 4 mm safety banana, 19 mm (0.75") spacingIsolation (terminals – earth) 1000 V RMSResistor technology Metal foilOriginal manufacturers part number HVD10Weight Approximately 53 g (1.87 oz)Operating temperature range +5 oC to +40 oC (41 °F to 104 °F)
stw
EL
AS
HD
V10
23.0 mm
(0,91")
32.0 mm
(1,26")
70.0 mm (2,71") Approximately 0.8 m (31,5")
Figure 1.33: Dimensions
B4375-1.2 en 31 HBM: public
Gxxx: 5 kV RMS High Precision Differential Probe (Option, to be ordered separately)5 kV RMS, 20 MΩ, 50:1, typical 0.1 % high precision, differential probe to be used in combination with GN610B and GN611B acquisition cards.The built-in earthing monitor system increases safety of the user and protects the GEN series inputs for isolation overloads.The probe and output cable are optimized to match the 33 pF input capacitance when using the ranges ± 10 V, ± 20 V, ± 50 V, ± 100 V, ± 200 V,± 500 V, ± 1000 V. In the lower ranges passband attenuation exceeds the HDP specified amplitude response.
+
-
C1
C1
LN PE
Protective earth
Functional earthOvervoltage protection
Earth
monitoring
C2
C2
R1
R2
R2
R1
Uin
Divider Probe
-
+ADC
Isolated Channel
IEC 61010-031:2002
certified
Figure 1.34: Block diagram and imageMaximum input to earth voltageIEC 61010-031:2002, CB Test certificate CA/20325/CSA
1000 V RMS or DC CAT IV5 kV RMS (No measurement category defined for nominal voltage above 1 kV RMS)
Maximum DC input to earth voltageIEC 61010-031:2015
1500 V DC CAT IV
Divider ratio 50:1Measurement ranges ± 500 V up to ± 1500 V (1 kV RMS) CAT IV
± 1 kV RMS up to ± 5k V RMS (Full safety rated)In-accuracy (divider, cable and GN610B) ± 0.2% @ 1000 V, 50 Hz, 25 °C (0.1% typical)Input impedance 20 MΩ ± 0.2%Temperature coefficient ± 10 ppm / oC (± 18 ppm / oF)Bandwidth -0.5 dB @ 100 kHz, phase match 0.1o
Input cable/pins High voltage insulated 4 mm safety connector (similar to Stäubli Electrical Cable MCXHM-5000)
Output pins Cable with 4 mm banana plugsEarth monitoring If functional earth is not attached divider is floating and inputs are disconnected from the
outputOriginal manufacturers part number HVD50ROperating temperature range 5 oC to +40 oC (41 °F to 104 °F)
Approximately 0.7 m (27,5") Approximately 3.4 m (134")
188 mm (7,40")
140 mm
(5,51")
76 mm
(3,00")
Figure 1.35: Dimensions
HBM: public 32 B4375-1.2 en
Gxxx: 5 kV RMS High Voltage Cable Extension (Option, to be ordered separately)The High Voltage Cable (HVC) is an extension for measurement cables with voltages up to 5 kV RMS. This device is designed to be connectedwith a cable on the input terminal of the high precision differential probe HVD10 and HVD50R. The HVC is designed according IEC61010-031:2015 compliant to 1000 V RMS CAT IV and 1500 V DC CAT IV.
Figure 1.36: CableMaximum Current 5AMaximum Input to earth voltageIEC 61010-031:2015
1 kV RMS CAT IV1.5 kV DC CAT IV5 kV RMS (No measurement category defined for nominal voltage above 1 kV RMS /1.5 kV DC)
Impedance Input --> Output < 0,1 ΩOriginal manufacturers part number HVCOperating temperature range + 5 °C to + 40 °C (+ 41 °F to + 104 °F)Input cable connector High voltage insulated 4 mm safety connector (similar to Stäubli Electrical Cable MC
XHM-5000)Output cable connector Female connector compatible with:
l HVD50Rl HVD10l Stäubli Electrical Cable XHM-5000l Stäubli Electrical Cable XHL-5000l Schützinger HSPL 7576 / 1
Available cable lengths 1,5 m (59") / 3m (118") / 4,5m (177") / 6m (236")
Approximately 1.5 m (59,00") / 3 m (118,00") /
4.5 m (177,00") / 6 m (236,00")
19 mm (0,75")
134 mm (5.28")
ø
Figure 1.37: Dimensions
B4375-1.2 en 33 HBM: public
Gxxx: High Precision Burden Resistor (Option, to be ordered separately)Low ohmic, 1 W, 0.02% high precision, low thermal drift burden resistor. Uses 4 wire connection to reduce inaccuracy caused by the currentsrunning to the burden resistor. Using banana input connectors and banana output pins. Directly compatible with GN610, GN611, GN610B andGN611B acquisition cards.
-
+
ADCUout
Iin
Iout
Isolated ChannelBurden Resistor
Figure 1.38: Block diagram and imageIn-accuracy ± 0.02%Temperature coefficient ± 5 ppm / oC (± 9 ppm / oF)Bandwidth -0.5 dB @ 300 kHzInput pins 4 mm safety banana, 13 mm (0.51") spacingOutput pins 4 mm safety banana, 19 mm (0.75") spacingIsolation (terminals – earth) 50 V RMSResistor technology Metal foilMaximum power dissipation 1 WOriginal manufacturers part number HBR1.0 HBR2.5 HBR10Impedance 1 Ω 2.5 Ω 10 ΩMaximum input current 1 A 0.63 A 0.31 AWeight 60 g (2.12 oz)Operating temperature range 0 oC to +40 oC (32 °F to 104 °F)
st w
ELA
S
HB
R2 . 5
99.5 mm (3,91")
23.0
mm
(0,9
0")
32.0
mm
(1,2
5")
13
mm
(0,0
,51")
19.0
mm
(0,7
4")
Figure 1.39: Dimensions
HBM: public 34 B4375-1.2 en
Gxxx: Current Transducers (Option, to be ordered separately)Current loop (compensated) current transducers (CT) using an extremely accurate zero flux detector with excellent linearity and low temperaturedrift. Electrostatic shield between primary and secondary circuit, with low insertion loss and high immunity to electrostatic and magnetic fields.To be used with CT power supply and burden resistor to measure high frequency currents.
y x
SD
- +
Power
Amplifier
Output
Amplifier
Standard
Resistor
IoutIp
Ip
IS
N=
Flux
Detector
Figure 1.40: Simplified block diagram and imageIsolation 150 V RMS reinforced isolation
300 V RMS single isolationLinearity ≤ 3 ppmTemperature coefficient of Ioe ≤ 0.5 ppm/oC (± 0.9 ppm / oF)Supply voltages + 15 V (± 5%) and - 15 V (± 5%), ≤ 80 mA + IsOriginal manufacturers partnumber
LEM ULTRASTAB series
Operating temperature range 10 oC to +50 oC (50 °F to 122 °F)Device IT 1000-S/SP1 IT 605-S IT405-S IT 205-S IT 65-SPrimary nominal current DC 1000 A DC 600 A DC 400 A DC 200 A DC 60 A DCPrimary nominal current AC 707 A RMS 600 A RMS 400 A RMS 200 A RMS 60 A RMSSecondary nominal current DC 1000 mA DC 266 mA RMS 200 mA RMS 200 mA RMS 100 mA DCConversion ratio (N) 1000:1 1500:1 1500:1 1000:1 600:1Burden recommendation HBR1.0 HBR2.5 HBR2.5 HBR2.5 HBR10.0Small signal bandwidth (-3 dB) 500 kHz 300 kHz 300 kHz 1000 kHz 800 kHzWeight (typical) 1.0 kg (35.3 oz) 1.08 kg (38.0 oz) 1.08 kg (38.0 oz) 0.35 kg (12.34 oz) 0.33 kg (11.64 oz)Dimensions
Cable diameter 30 mm (1.18") 30 mm (1.18") 30 mm (1.18") 26 mm (1.02") 26 mm (1.02")Width 128 mm (5,03”) 128 mm (5,03”) 128 mm (5,03”) 93 mm (3,66”) 93 mm (3,66”)
Height 106 mm (4,17”) 106 mm (4,17”) 106 mm (4.17”) 77.7 mm (3,05”) 77.7 mm (3,05”)Depth 85 mm (3,34”) 67 mm (2,63”) 67 mm (2,63”) 47 mm (1.85”) 47 mm (1.85”)
Top view
Heig
ht
Width
DepthCable
diameter
Front view
Figure 1.41: Dimensions
B4375-1.2 en 35 HBM: public
Gxxx: Power Supply for Current Transducer (Option, to be ordered separately)Empty modular 19” rack with 1 to maximum 6 channel CT support. Housing for the modular power supply and control for LEM currenttransducers.
Mains
Out
Sensor
6 5 4 3 2 1ReadoutMCTS
Power1 2 3 4 5 6
Status
-15V/GND/+15V
47-63Hz
110-264VACSIGNALTEC
Figure 1.42: Front side (left) and rear side (right)Maximum number of CTs 6Input connectors 9 pin SUBDOutput connectors 4 mm banana plugsSignal LEDS CT Power, CT StatusPower supply 100 to 240 V AC, 47 to 63 Hz @ 650 mAWeight Typical 6.5 kg (14.33 lb) configured with 6 channelsOriginal manufacturers part number MCTSOperating temperature range 0 °C to +50 °C (32 °F to 122 °F)Dimensions
Height 87.2 mm (3,43”)Width / Width including mounting ears 442 mm (17,40”) / 466 mm (18,34”)
Depth including handles 415 mm (16,33”)
Mains
Out
Sensor
6 5 4 3 2 1Readout
-15V/GND/+15V
47-63Hz
466 mm (18,34")
415 mm (16,33")442 mm (17,40")
87
.2 m
m (
3,4
3")
Figure 1.43: Dimensions
HBM: public 36 B4375-1.2 en
Current Transducer (CT) Wire Diagram
tA7PTP2
PTP1CAL
M/S
I/O
Power
HDD
M/S
0
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
1
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
2
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
3
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
4
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
5
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
6
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
1
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
2
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
3
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
4
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
5
2M
Ω 1
kV
DC
CA
T I
I 6
00
V
6
Mains
Out
Sensor
6 5 4 3 2 1Readout
-15V/GND/+15V
47-63Hz
110-264VAC
DC AC Power
Converter
I1
I2I3
Current
Transducer
mA Signals, use
KABXXX Cables
KAB290 Cable KAB290 Cable
KABXXX
CT Cable
KABXXX
CT Cable
mA Signals, use
KAB290 Cables
Burden Resistor HBR x
Burden
Resistor
HBR x
Cable Connector for
Current Transducer
GEN7tA
Electric
Motor
Current Transducer
Power Supply
Current Transducer
Power Supply
Connect
Yellow
to GEN7tA
Protective
Ground
GN610B/GN611B
Figure 1.44: Current transducer connection diagram
Note For more information, please refer to "A4775-1.0 en GN610_GN611B to Current Transducer cabling" Installation Guide.
B4375-1.2 en 37 HBM: public
Ordering Information (1)
Article Description Order No.Basic 1 kV ISO200 kS/s
6 channels, 18 bit, 200 kS/s, ± 10 mV to ± 1000 V input range,200 MB RAM, 1 kV isolated balanced differential input(600 V RMS CAT II isolation), 4 mm fully isolated banana plugs.Real-time cycle based calculations with triggering oncalculated results. Supported by Perception V6.72 and higher.This card is not supported by GEN2i, GEN5i, GEN7t andGEN16t mainframes.
1-GN611B-2
(1) All GEN series systems are intended for exclusive professional and industrial use.
Option, to be ordered separatelyArticle Description Order No.GEN DAQ real-time formuladatabasecalculators
Option to enable enhanced real-time calculators. Setup uses auser configurable formula database similar to the Perceptionformula database. All calculations are performed by the DSP ofthe acquisition card. Triggering possible on many of the resultsof the calculations. Calculated results can be transferred to theGEN DAQ EtherCAT® option with a 1 ms latency.
1-GEN-OP-RT-FDB-2
HBM: public 38 B4375-1.2 en
Special Voltage Probes, to be ordered separatelyArticle Description Order No.10 MΩ impedanceprobe
1 kV DC, 10 MΩ, 10:1, 0.2% high precision differential probe tobe used in combination with GN610B and GN611B acquisitioncards. Reduces the resistive/current load on the device undertest by increasing the input impedance to 10 MΩ. The highestinput range remains at ± 1000 V due to the maximum voltagerating of the probe.
Ordered fromcustomsystems(1)
1 kV RMS, 4 MΩ,10:1 differentialprobe
1 kV RMS/DC CAT IV, 4 MΩ, 10:1, 0.2% high precision,differential probe to be used in combination with GN610B andGN611B acquisition cards.The probe is optimized to match the 33 pF input capacitancewhen using the GN61xB ranges ± 10 V to ± 1000 V.
Ordered fromcustomsystems(1)
5 kV RMS, 20 MΩ,50:1 differentialprobe
5 kV RMS, 20 MΩ, 50:1, 0.2% high precision, differential probeto be used in combination with GN610B and GN611Bacquisition cards. The built-in earthing monitor systemincreases safety of the user and protects the GEN series inputsfor isolation overloads.
Ordered fromcustomsystems(1)
5 kV RMS HighVoltage Cable
The High Voltage Cable (HVC) is an extension formeasurement cables with voltages up to 5 kV RMS. This deviceis designed to be connected with a cable on the input terminalof the high precision differential probe HVD10 and HVD50R.The HVC is designed according IEC 61010-031:2015compliant to 1000 V RMS CAT IV and 1500 V DC CAT IV.
Ordered fromcustomsystems(1)
(1) Contact custom systems at: [email protected] quote/information for special products for GEN series.
B4375-1.2 en 39 HBM: public
Accessories, to be ordered separatelyArticle Description Order No.600V RMS 2-wireIsolated shieldedtest leads
Black/Red lead set combined within shielded housing (Yellow).600 V RMS CAT II,10 A RMS safety-shrouded stackablebanana plugs. Significantly reduces signal disturbance pickupon GN610/GN611/GN610B/GN611B cards by using twoidentical signal wires with earthed shield.Note: Do not use for 3 wire connections!Available lengths: 1.5 m (4.92 ft), 3.0 m (9.84 ft) and 6.0 m(19.69 ft)
1-KAB290-1.51-KAB290-31-KAB290-6
600 V RMS 3-wireIsolated shieldedtest leads
Brown/Grey/Black lead set combined within shielded housing(Yellow). 600 V RMS CAT II, 1 A RMS safety-shroudedstackable banana plugs. Typically used for 3 phase voltagemeasurements using the GN610B/GN611B cards. The onesided earthed shield reduces high frequency emissions.Available lengths: 1.5 m (4.92 ft), 3.0 m (9.84 ft) and 6.0 m(19.69 ft)
1-KAB2128-1.51-KAB2128-31-KAB2128-61-KAB2128-121-KAB2128-20
BNC to bananaadapter
Set of six pieces, safety isolated female BNC to dual 4 mmprotected banana adapter.1000 V RMS CAT II, 600 V RMS CAT III and 1 A current safetyratings. Can be used with GN610/GN611/GN610B/GN611Binput cards.
1-G067-2
Artificial staradapter
The artificial star adapter is a plug-on interface card to measure3 phase signals with the GN610/GN611/GN610B/GN611Bcards. This adapter is intended for measuring 3 phase signalswhile creating a virtual/artificial star point.
1-G068-2
HBM: public 40 B4375-1.2 en
Current Sensors, to be ordered separatelyArticle Description Order No.Ultrastab 60 ARMS currenttransducer
LEM IT 65-S Ultrastab. 60 A DC, 60 A RMS current transducerwith 800 kHz bandwidth. Recommended burden resistorHBR10.0. Resulting ratio 60 A/V (16.6667 mV/A).
Ordered fromcustomsystems(1)
Ultrastab 200 ARMS currenttransducer
LEM IT 205-S Ultrastab. 200 A DC, 200 A RMS currenttransducer with 1000 kHz bandwidth. Recommended burdenresistor HBR2.5. Resulting ratio 400 A/V (2.5 mV/A).
Ordered fromcustomsystems(1)
Ultrastab 400 ARMS currenttransducer
LEM IT 405-S Ultrastab. 400 A DC, 400 A RMS currenttransducer with 300 kHz bandwidth. Recommended burdenresistor HBR2.5. Resulting ratio 666.6 A/V (1.5 mV/A).
Ordered fromcustomsystems(1)
Ultrastab 600 ARMS currenttransducer
LEM IT 605-S Ultrastab. 600 A DC, 600 A RMS currenttransducer with 300 kHz bandwidth. Recommended burdenresistor HBR2.5. Resulting ratio 600 A/V (1.6667 mV/A).
Ordered fromcustomsystems(1)
Ultrastab 1000 ARMS currenttransducer
LEM IT 1000-S Ultrastab. 1000 A DC, 707 A RMS currenttransducer with 500 kHz bandwidth. Recommended burdenresistor HBR1.0. Resulting ratio 1000 A/V (1.0 mV/A).
Ordered fromcustomsystems(1)
HBR 1 Ω, 1 Wprecision burdenresistor
1 Ω, 1 W, 0.02% high precision, low thermal drift burdenresistor. Internally uses 4 wire connection to reduce inaccuracycaused by the currents running to the burden resistor. Usingbanana input connectors and banana output pins. Directlycompatible with GN610, GN611, GN610B and GN611Bacquisition cards.
Ordered fromcustomsystems(1)
HBR 2.5 Ω, 1 Wprecision burdenresistor
2.5 Ω, 1 W, 0.02% high precision, low thermal drift burdenresistor. Internally uses 4 wire connection to reduce inaccuracycaused by the currents running to the burden resistor. Usingbanana input connectors and banana output pins. Directlycompatible with GN610, GN611, GN610B and GN611Bacquisition cards.
Ordered fromcustomsystems(1)
HBR 10 Ω, 1 Wprecision burdenresistor
10 Ω, 1 W, 0.02% high precision, low thermal drift burdenresistor. Internally uses 4 wire connection to reduce inaccuracycaused by the currents running to the burden resistor. Usingbanana input connectors and banana output pins. Directlycompatible with GN610, GN611, GN610B and GN611Bacquisition cards.
Ordered fromcustomsystems(1)
(1) Contact custom systems at: [email protected] quote/information for special products for GEN series.
B4375-1.2 en 41 HBM: public
Current Measurement Solutions, to be ordered separately3 phase 60 A RMScurrent solution
Package consisting of 3 * LEM IT 65-S current transducers,1 * MCTS 19" rack with built-in 3 * power supply and CT control,3 * MCTS to CT connection cable 10 m (32.81 ft), 3 * KAB2903.0 m (9.84 ft) and 3 * HBR10.0 burden resistor
Ordered fromcustomsystems(1)
1 phase 60 A RMScurrent extension
Package consisting of 1 * LEM IT 65-S current transducers,1 * power supply and CT control, 1 * MCTS to CT connectioncable 10 m (32.81 ft), 1 * KAB290 3.0 m (9.84 ft) and 1 * HBR10.0 burden resistor. One extension channel for the 3 phase60 A RMS current solution. Factory installed and must beordered together with the 3 phase 60 A RMS current solution.
Ordered fromcustomsystems(1)
3 phase 200 ARMS currentsolution
Package consisting of 3 * LEM IT 205-S current transducers,1 * MCTS 19" rack with built-in 3 * power supply and CT control,3 * MCTS to CT connection cable 10 m (32.81 ft), 3 * KAB2903.0 m (9.84 ft) and 3 * HBR2.5 burden resistor
Ordered fromcustomsystems(1)
1 phase 200 ARMS currentextension
Package consisting of 1 * LEM IT 205-S current transducers,1 * power supply and CT control, 1 * MCTS to CT connectioncable 10 m (32.81 ft), 1 * KAB290 3.0 m (9.84 ft) and1 * HBR2.5 burden resistor. One extension channel for the 3phase 200 A RMS current solution. Factory installed and mustbe ordered together with the 3 phase 200 A RMS currentsolution.
Ordered fromcustomsystems(1)
3 phase 400 ARMS currentsolution
Package consisting of 3 * LEM IT 405-S current transducers,1 * MCTS 19" rack with built-in 3 * power supply and CT control,3 * MCTS to CT connection cable 10 m (32.81 ft), 3 * KAB2903.0 m (9.84 ft) and 3 * HBR2.5 burden resistor
Ordered fromcustomsystems(1)
1 phase 400 ARMS currentextension
Package consisting of 1 * LEM IT 405-S current transducers,1 * power supply and CT control, 1 * MCTS to CT connectioncable 10 m (32.81 ft), 1 * KAB290 3.0 m (9.84 ft) and1 * HBR2.5 burden resistor. One extension channel for the 3phase 400 A RMS current solution. Factory installed and mustbe ordered together with the 3 phase 400 A RMS currentsolution.
Ordered fromcustomsystems(1)
HBM: public 42 B4375-1.2 en
Current Measurement Solutions, to be ordered separately3 phase 600 ARMS currentsolution
Package consisting of 3 * LEM IT 605-S current transducers,1 * MCTS 19" rack with built-in 3 * power supply and CT control,3 * MCTS to CT connection cable 10 m (32.81 ft), 3 * KAB2903.0 m 9.84 ft and 3 * HBR2.5 burden resistor
Ordered fromcustomsystems(1)
1 phase 600 ARMS currentextension
Package consisting of 1 * LEM IT 605-S current transducers,1 * power supply and CT control, 1 * MCTS to CT connectioncable 10 m (32.81 ft), 1 * KAB290 3.0 m (9.84 ft) and 1 * HBR2.5burden resistor. One extension channel for the 3 phase 600 ARMS current solution. Factory installed and must be orderedtogether with the 3 phase 600 A RMS current solution.
Ordered fromcustomsystems(1)
3 phase 1000 ARMS currentsolution
Package consisting of 3 * LEM IT 1000-S current transducers,1 * MCTS 19" rack with built-in 3 * power supply and CT control,3 * MCTS to CT connection cable 10 m (32.81 ft), 3 * KAB2903.0 m (9.84 ft) and 3 * HBR1.0 burden resistor
Ordered fromcustomsystems(1)
1 phase 1000 ARMS currentextension
Package consisting of 1 * LEM IT 1000-S current transducers,1 * power supply and CT control, 1 * MCTS to CT connectioncable 10 m (32.81 ft), 1 * KAB290 3.0 m (9.84 ft) and1 * HBR2.5 burden resistor. One extension channel for the 3phase 1000 A RMS current solution. Factory installed and mustbe ordered together with the 3 phase 1000 A RMS currentsolution.
Ordered fromcustomsystems(1)
(1) Contact custom systems at: [email protected] quote/information for special products for GEN series.
B437
5-1.
2 en
HBM
: pub
lic
©Hottinger Baldwin Messtechnik GmbH. All rights reserved.All details describe our products in general form only.They are not to be understood as express warranty and donot constitute any liability whatsoever.
Hottinger Baldwin Messtechnik GmbHIm Tiefen See 45 ∙ 64293 Darmstadt ∙ GermanyTel. +49 6151 803-0 ∙ Fax: +49 6151 803-9100E-mail: [email protected] ∙ www.hbm.com
measure and predict with confidence