58
GENERAL INFORMATION USERS MANUAL Colormetry CMU-324HE Hardness For Europe Manual Code No. S855-080-5813 This document was formulated in Japan. Comply with the regulations and standards of the country of use regarding installation and usage. The specifications of products and components may vary with country of use and the site situation. INFORMATION IN THIS MANUAL MAY BE CHANGED WITHOUT NOTICE.

GENERAL INFORMATION USER S MANUAL€¦ · GENERAL INFORMATION USER’S MANUAL Colormetry CMU-324HE Hardness For Europe Manual Code No. S855-080-5813 This document was formulated in

  • Upload
    others

  • View
    7

  • Download
    0

Embed Size (px)

Citation preview

GENERAL INFORMATION USER’S MANUAL

Colormetry

CMU-324HE

Hardness

For Europe

Manual Code No. S855-080-5813

This document was formulated in Japan. Comply with the regulations and standards of the country of use regarding installation and usage. The specifications of products and components may vary with country of use and the site situation. INFORMATION IN THIS MANUAL MAY BE CHANGED WITHOUT NOTICE.

Release date: March 2017

Important Safety Information

i

Important Safety Information

Before using the Colormetry CMU-324HE, ensure that this User’s Manual is read and fully

understood; furthermore, the instructions given herein should be strictly observed during

operation.

Failure to use the Colormetry CMU-324HE in the appropriate manner can, therefore, result in

death, injury, the outbreak of fire, and other serious accidents.

In order to ensure that tasks undertaken during operation and maintenance of the Colormetry

CMU-324HE can be completed safely, it is critical that all dangerous and hazardous parts of this

equipment be confirmed in advance. Here at MIURA, it is practically impossible to anticipate all

potentially dangerous situations, and therefore, this User’s Manual deals only with known

hazards. A higher level of safety can be achieved by careful observation of the warnings and

instructions set forth herein.

Safety-related warnings are classified according to the following three levels, and notification of

these warnings is provided within this User’s Manual and using warning labels on the equipment

itself.

Indicates an imminently dangerous situation which leads to serious

injury or death to the user, if the product is mishandled.

Indicates a situation which might cause serious injury or death to

the user, if the product is mishandled.

Indicates a situation which might cause minor injury to the user or

the occurrence of physical damage only, if the product is

mishandled.

If any part used in the product fails, contact your dealer or MIURA sales office to have it replaced

with a genuine part. Using parts with other model numbers may reduce the safety.

It should be noted that unauthorized alteration of the Colormetry CMU-324HE and repair

procedures other than those described herein can pose a serious safety risk. Never customize or

repair the equipment without permission from MIURA.

DANGER

WARNING

CAUTION

Introduction

ii

Introduction This document describes the methods of handling the Colormetry CMU-324HE (hereinafter

called “the equipment”) to ensure its proper use.

Not only those who use the equipment for the first time but also those who know the handling

methods should read this document carefully and understand the proper handling methods

before use of the equipment.

Furthermore, we also recommend that this User’s Manual be safely kept in close proximity to the

equipment so that it can be referred to at any time to confirm the correct usage.

Follow the INSTALLATION AND START UP USER’S MANUAL (separate volume) for installing

the equipment.

[Explanation of Notation]

The meanings of the symbols used in this manual are as shown below.

Indicates a prohibited action (what you must not do).

Indicates a possibility of injury when a specific location of the product is

touched.

Indicates a possibility of electric shock when the product is handled with

wet hands.

Indicates an action to be taken according to an instruction (what to be

executed).

Indicates an action connecting a grounding wire.

Indicates a caution

Indicates a possibility of injury due to high temperatures under specific

conditions.

NOTE Used to highlight important information to prevent equipment malfunction

as well as tips for efficient work and other useful information.

Indicates the page of related information.

Table of Contents

iii

Table of Contents

Important Safety Information ................................................................. i

Introduction .......................................................................................... ii

SECTION 1 SAFETY .................................................................. 1

1 Safety Regulations ......................................................................... 1

2 Safety-related Knowledge and Expertise ........................................ 2

3 Protective Equipment...................................................................... 2

4 Prohibition of Unapproved Modification .......................................... 2

5 Warning Labels on the Equipment .................................................. 3

6 Safety Precautions ......................................................................... 4

6.1 Installation ................................................................................................ 4

6.2 Safety Precautions ................................................................................... 5

6.3 Pre-operation Inspection (Excerpt from page 21 onwards) ...................... 5

6.4 Inspection and Maintenance (Excerpt from page 37 onwards) ................. 6

6.5 Storage (Excerpt from page 46 onwards) ................................................. 7

SECTION 2 OVERVIEW ............................................................. 8

1 Outline ............................................................................................ 8

1.1 Usage ....................................................................................................... 8

1.2 Purpose .................................................................................................... 8

1.3 System Layout .......................................................................................... 9

1.4 Principle .................................................................................................. 10

1.5 Applicable Water Condition .................................................................... 11

1.6 Highlights of Features ............................................................................. 12

2 Name of Parts............................................................................... 15

2.1 Overall View ........................................................................................... 15

2.2 Reagent Cartridge .................................................................................. 17

2.3 Filter Casing Assembly ........................................................................... 18

3 Specifications ............................................................................... 19

3.1 General Specifications ............................................................................ 19

3.2 Physical Data .......................................................................................... 19

3.3 Monitoring Capabilities ........................................................................... 19

3.4 Feed-water and Drain-water Connection ................................................ 20

3.5 Accessories ............................................................................................ 20

3.6 Consumable parts (Sold separately) ...................................................... 20

Table of Contents

iv

SECTION 3 PRE-OPARATION INSPECTION .......................... 21

1 Inspections and Preparations before Start up .............................. 21

1.1 Inspection and Preparation of the Surrounding Area Peripheral

Equipment .............................................................................................. 21

2 Preparations before Start up ........................................................ 22

SECTION 4 OPERATION .......................................................... 23

1 Operation ..................................................................................... 23

1.1 Self Check Mode .................................................................................... 23

1.2 Automatic Monitoring .............................................................................. 24

1.3 Manual Monitoring .................................................................................. 24

2 FUNCTIONS ................................................................................ 25

2.1 Remote Signal Input ............................................................................... 25

2.1.1 Method .............................................................................................................. 25

2.1.2 Detailed descriptions ........................................................................................ 25

2.1.3 Delay time of remote signals ............................................................................ 27

2.2 Colormetry Monitor Timing ...................................................................... 28

2.2.1 Automatic monitoring ........................................................................................ 28

2.2.2 Manual monitoring ............................................................................................ 30

2.3 Concentration Evaluation ........................................................................ 31

2.3.1 Manual monitor ................................................................................................. 31

2.3.2 Concentration anomaly .................................................................................... 32

SECTION 5 INSPECTION AND MAINTENANCE .................... 37

1 Reagent Cartridge ........................................................................ 37

1.1 Reagent Cartridge Replacement Timing ................................................. 37

1.2 Cautions for Handling ............................................................................. 37

1.3 Reagent Cartridge Replacement Method ............................................... 38

2 Fiber Filter and Constant Flow Valve Replacement ...................... 39

2.1 Fiber Filter Replacement Timing ............................................................. 39

2.2 Fiber Filter Replacement ........................................................................ 39

2.2.1 Remove filter casing assembly ........................................................................ 39

2.2.2 Fiber filter replacement..................................................................................... 40

2.2.3 Filter case assembly installation ...................................................................... 41

3 Daily Inspection ............................................................................ 42

3.1 Inspection Item ....................................................................................... 42

3.1.1 Monitor operation ............................................................................................. 42

3.1.2 Feed-water and drain-water tube ..................................................................... 42

3.1.3 Filter case assembly ......................................................................................... 43

3.2 Others ..................................................................................................... 43

Table of Contents

v

SECTION 6 BREAKDOWN AND COUNTERMEASURES ...... 44

1 Self Check Function ..................................................................... 44

1.1 Display .................................................................................................... 44

1.2 Alarm ...................................................................................................... 44

1.3 Alarm Reset ............................................................................................ 44

2 Self Check Items .......................................................................... 45

SECTION 7 STORAGE ............................................................. 46

1 Extended Inactivity ....................................................................... 46

2 Transfer or Resale ........................................................................ 46

3 Export ........................................................................................... 46

SECTION 8 DISPOSAL ............................................................ 47

1 Disposal ........................................................................................ 47

SECTION 9 WARRANTY .......................................................... 48

1 Note Regarding Warranty ............................................................. 48

2 Questions Regarding the Product and User’s Manual .................. 48

3 If the Operation Manual is Lost ..................................................... 48

SECTION 1 SAFETY

1

SECTION 1 SAFETY

1 Safety Regulations Handling of the equipment is comprised of operations associated with running of the equipment and operations regarding maintenance. Only those who have received a handling explanation with this User’s Manual should operate the equipment.

Applicable to All Water Treatment Equipment (Related Laws and Regulations) Document No.: 98-002-03

<Precautions for installation and use> — Related laws and regulations 1) The table below shows part of the laws and regulations on equipment installation and use. Refer to the separate document for more

information on other relevant laws and regulations. (There are also prefectural and municipal ordinances; therefore, contact the supervisory authority or competent authority for more information.)

2) The information listed in the table was created based on the laws and regulations in effect when the information was published. 3) The documents to be submitted and their destination vary depending on your equipment and facilities.

If you have any questions, contact your nearest dealer, MIURA sales office, or the competent supervisory authority. 4) The laws and regulations listed below may not apply to all water treatment equipment. Refer only to those laws and regulations that

apply to your equipment and facilities.

Related laws and regulations

Form(s) to be submitted

Destination Submission

timing Remarks

Water Quality Pollution Control Act, River Act, Sewerage Service Act, etc.

Verification required from local towns and cities in which the water treatment equipment or other equipment is installed. (The laws and regulations in some areas may be defined even more specifically in accordance with regional regulations.)

Drain water may be produced depending on the type of water treatment equipment and equipment operation. Drain water must be treated in other ways to prevent it from damaging waterways and the like. In cases where the total amount of drain water exceeds the legal amount on a per-plant basis or where the equipment is used at specified facilities, an application is necessary. However, the conditions may vary from region to region. Accordingly, you should contact your local authorities for more information.

Regional Pollution Prevention Ordinance

Notification as required by the applicable ordinance

As set forth by the relevant regional authorities

As set forth by the relevant regional authorities

Applicability is defined independently by regional pollution prevention ordinances based; accordingly, you should contact your local authorities for more information. (Certain regional ordinances concerning air pollution, noise, vibration, drain water, and the like may supersede the corresponding national regulations.)

Water Supply Act

None — —

(Prohibition of direct waterworks contact) Water treatment equipment (excluding certified equipment) cannot be connected directly to waterworks. Instead, it is necessary to either set up a feed water tank or to isolate the systems using, for example, a float valve. Details may vary from region to region; accordingly, you should contact your local authorities for more information.

Design specifications for dedicated waterworks, Notification of the start of feeding water

As set forth by the relevant regional authorities

Business operators are to submit notification to local governments for drinking water wells for industrial use in facilities where the volume of water intake exceeds the allowed quantity as prescribed by law which require the use of dedicated waterworks systems regardless of the form of usage. Contact your local authorities for details.

Act on Maintenance of Sanitation in Buildings

None — — Buildings are to be preserved and maintained (through cleaning, water quality analysis, and other activities) in accordance with Hygiene Control Standards.

Food Sanitation Act

No notification made for equipment and related matters

— — Measures are to be implemented to maintain hygiene and safety and prevent sanitation hazards in accordance with applicable laws.

Industrial Safety and Health Act

None — —

When using substances specified in applicable laws (such as specified chemical substances), a qualified operator must be selected, all such substances are to be handled and managed in accordance with applicable laws, and the operator’s healthcare is necessary.

Poisonous and Deleterious Substances Control Act

None — —

All such substances are to be stored and managed in accordance with applicable laws. Chemical substances are to be handled properly in accordance with the guidance and information listed on the SDS for the relevant material.

Fire Service Act Notification as set forth by local fire departments

The nearest fire department

Prior to the start of work

Local fire chiefs or fire department chiefs must be notified in advance of any plans for any party to use or store certain substances (chemical substances and other materials) of a designated quantity as specified by applicable laws and regulations.

SECTION 1 SAFETY

2

2 Safety-related Knowledge and Expertise Usage of the equipment requires knowledge and experience of the operation and maintenance

of mechanical equipment.

In addition, only the following persons should be permitted to perform work on the equipment.

・Operators who have read and fully understood this User’s Manual.

3 Protective Equipment

Wear protective items such as a helmet, safety glasses, a safety mask, safety footwear, and

leather gloves as necessary for the task in hand.

4 Prohibition of Unapproved Modification

Customization or modification not recommended by MIURA may present safety problems and is

therefore prohibited. If you wish to modify your water treatment equipment, contact your dealer

or MIURA sales office in advance. It should be noted that MIURA will accept no responsibility for

the outcome of unapproved customization.

SECTION 1 SAFETY

3

5 Warning Labels on the Equipment

Warning labels are pasted at all locations that require attention to be paid during operation.

If any warning label is peeled off or torn, replace it with a new one.

For further details on warning labels contact your dealer or MIURA sales office.

Packing of reagent cartridge

Reagent cartridge left side / back side

SECTION 1 SAFETY

4

Ensure that drain piping is fixed securely in place to prevent it from moving in reaction

to drain discharge and that the end of the piping opens in a safe location such as a pit.

Failure to observe this precaution may lead to electric shock or other problems due to

splashing of water. Touching the drain pipe may result in burns or other injury.

High concentration water may be discharged from the water treatment equipment

temporarily in some cases. Additional water treatment may be necessary to prevent

lakes and other bodies of water from being adversely affected depending on the

volume of drain and the water quality.

When connecting a power supply to the product, use wire of the specified rating and

ensure that the equipment are correctly grounded (Japanese D-type grounding work

or better).

Furthermore, a dedicated power supply should be used, and an earth leakage circuit

breaker (with overcurrent protection) should be installed. If this precaution is not

observed, serious accidents such as electric shock and the outbreak of fire may

result. In case of emergency, the power supply has to cut immediately. Install an earth

leakage circuit breaker at the easy accessible location.

6 Safety Precautions

6.1 Installation

During installation, observe the following safety information.

NOTE

Water Leakage Countermeasures

Factors such as loose connectors and aging of pipe materials can lead to

water leakage. It is important that drain ditches be setup for the water

treatment equipment and all peripheral equipment.

(In order to prevent water from leaking outside the room or downstairs, set

up a drain ditch that completely encloses the whole equipment, ensure that

the floor itself is waterproof, make the floor slope downward to the drain

port, and take any other necessary steps.)

Drain ditch

WARNING

Drain piping

Pit Drain piping

WARNING

Ground terminal

Breaker capacity

Earth leakage circuit breaker (with overcurrent protection)

Wiring

Earth leakage circuit breaker (with overcurrent protection)

SECTION 1 SAFETY

5

Do not place anything which might cause electric leakage underneath the

equipment.

The equipment is designed to drain water from the bottom of the equipment if

there is inner water leakage. Water might splash to the item under the

equipment.

Do not insert any objects into the reagent cartridge receptacle other items.

It will cause an evaluation and system errors. Even if the equipment is not

to use for a while and disconnect from the power supply, do not remove

the reagent cartridge to prevent a contamination.

Do not insert fingers into the regent cartridge receptacle. There is possibility of

injury by chemical injection roller inside the equipment.

Do not operate the equipment if water is leaking.

Turning switches ON and OFF while water is leaking may lead to fire or

electric shocks.

Make sure the main power circuit breaker is turned OFF and contact your

dealer or MIURA sales office.

6.2 Safety Precautions

During operation, observe the following safety information.

NOTE

Please wipe with dry soft cloth when cleaning.

Rubbing with hard cloth may make scratches.

6.3 Pre-operation Inspection (Excerpt from page 21 onwards)

During pre-operation Inspection, observe the following safety information.

WARNING

WARNING

CAUTION

CAUTION

SECTION 1 SAFETY

6

Replace a tube if a defect such as deformation (a bend or fold), discoloration,

hardening, or cracking is detected on the tube surface. If a deteriorated tube is

continuously used, it can rip and cause water leakage.

Please use specified tubes when exchanging tubes. Using tubes other than

specified tubes may cause leakage.

Do not apply grease or lube when install the filter casing assembly. It might

damage the material of the equipment and cause water leakage.

Never disassemble the reagent cartridge.

Reagent may splatter onto the skin or the eyes.

If the reagent gets on the skin or in the eyes, rinse immediately with water.

Do not place anything which might cause electric leakage underneath the

equipment.

The equipment is designed to drain water from the bottom of the equipment if

there is inner water leakage. Water might splash to the item under the

equipment.

Do not use the equipment if there is smoke, abnormal odor or noise,

excessive overheating, or other abnormalities. If an abnormality occurs, do

not operate the equipment. Also, make sure an earth leakage circuit breaker

is turned OFF and a main feed-water valve is closed and contact your dealer

or MIURA sales office.

Failure to observe this precaution may lead to electric shock, fire, or a fault.

6.4 Inspection and Maintenance (Excerpt from page 37 onwards)

During inspection and maintenence, observe the following safety information.

CAUTION

WARNING

WARNING

CAUTION

CAUTION

SECTION 1 SAFETY

7

Do not perform any unapproved installation work or connect any peripheral

equipment.

Failure to observe this precaution may lead to personal injuries.

If there is a defect or cracking on the filter casing assembly, turn OFF an earth

leakage circuit breaker and close a main feed-water valve. Then, replace the

If water leakage is not stopped even though tightening the filter casing

assembly and tube joint, replace the filter casing assembly.

6.5 Storage (Excerpt from page 46 onwards)

During storage, observe the following safety information.

WARNING

CAUTION

SECTION 2 OVERVIEW

8

SECTION 2 OVERVIEW

1 Outline

1.1 Usage

Basically, the Colormetry CMU-324HE has been developed as part of a processing system for

boiler water.

The Colormetry automatically and regularly implements the process of sampling the water,

injecting the reagent, stirring and evaluating the result, thereby obviating the conventional

manual procedure.

The equipment has been designed with emphasis on the reliability of the monitoring results.

The manually selectable conditions—for instance, to verify monitoring upon the detection of

hardness leakage—prevent temporary fluctuations from triggering alarms.

1.2 Purpose

The Colormetry CMU-324HE is a device for monitoring water quality. It is not a device for

measuring. The result data cannot be used for inspection or analysis of the water treatment

system.

Usage example:

Monitoring hardness leakage of the water softener

SECTION 2 OVERVIEW

9

1.3 System Layout

The Colormetry consists of the reagent cartridge (the indicator), reagent indicator mechanism,

monitoring system and feed-water and drain systems.

Colormetry layout

Photoreceptor

Monitor system

Constant flow valve

Solenoid valve

Reagent

cartridge

Light source

Drain

Regent injection system

Drain system

Feed-water system

Tube

Electromagnetic

induction coil

Agitator

Filter

Nozzle

Sample water

Feed water

Reagent

injection pump

SECTION 2 OVERVIEW

10

1.4 Principle

The Colormetry is a device to monitor ionic concentrations in water by applying colorimetric

analysis. In a colorimetric analysis process the ionic and other concentration in water is

monitored by allowing a reagent to react against the target ions and others, and monitoring the

transmissivity of the resultant coloration for light of a specific wavelength. The Colormetry

replaces this process by an electrical monitoring system and automates the entire process.

Other features include the remote alarm output, self check function, and display, as well as the

monitor standby while the water softener is regeneration or the sample water stopped (remote

signal input).

Monitor process is indicated as follows.

[1] Washing and filling the monitor chamber with sample water

The solenoid valve opens, filling the monitor chamber with sample water. The old sample

water remaining in the chamber is pushed out, and the chamber is washed at the same time.

[2] Injecting reagent and stirring sample water

The solenoid valve closes, and the injection pump inject the reagents into the sample water.

The electromagnetic induction coil, activated at the same time, mixes the water and reagent.

The reagent causes coloration of the sample water in accordance with the hardness

concentration of the water.

[3] Monitoring the sample water

The light source is activated to project light into the sample water. The photereceptor

electronically monitors the light transmitted through the sample water. The intensity of the

transmitted light varies, depending on the color of the sample water. These variations are

caused by light absorption in the water, which varies with the color of the water.

[4] Evaluating the concentration and displaying the result

The concentration is evaluated from the monitoring value, and the result is indicated on the

LCD.

.

Displays the result

Washes and fills a monitor chamber with sample water

Injects reagent and stirs the sample

Monitors light transmitted

through the sample

Evaluates the concentration

SECTION 2 OVERVIEW

11

1.5 Applicable Water Condition

The Colormetry CMU-324HE monitors the hardness. Use CMU-H2 reagent cartridge.

The reagent may be affected by substances other than hardness.

The equipment cannot measure water of the condensate return tank.

Keep to the following items in an allowable range. If the water condition is not followed, the

equipment may fail to monitor correctly.

Items Allowable range

Fe (Iron) 1.0mg/L or less (*1)

Free chlorine 2.0mg/L or less

M-Alkalinity+ Silica(*3) 500mg/L or less

pH 5.0 ~ 8.6

Zn (Zinc) (*2)

Cu (Copper) (*2)

*1: It is the value measured by adding hydrochloric acid.

*2: The metal ion such as zinc and copper are determined as hardness.

Therefore, the system judges hardness higher.

Pay close attention in case the galvanizing pipe or copper pipes are used.

The metal ion (zinc and copper) is evaluated as 1.5 times the calcium and

magnesium hardness.

(Ex.) hardness: 0.1mg/L

Zinc: 1.0mg/L

Evaluate value: 0.1+1.0×1.5=1.6mg/L

*3: Measurement results of this unit is affected by the M-Alkali and dissolved carbon

dioxide gas and silica in a complex manner. Therefore, you must calculate the result

of the following equation is within the acceptable range.

Silica + M-Alkalinity + M-Alkalinity × 11,700,000/10^pH

SECTION 2 OVERVIEW

12

1.6 Highlights of Features

The Colormetry has the following features:

[1] Monitors hardness leakage automatically

The monitoring process is fully automated, saving a significant amount of work by

eliminating the need for complicated manual procedures.

[2] Requires no periodic calibration

The equipment needs no cumbersome periodic calibrations.

[3] Includes built-in timer

Monitoring period may be set as desired. (e.g. daily between 9 a.m. and 5 p.m.)

Interval for each monitoring may set as desired. (The interval is selectable in 30-minute

increments between 30 and 240 minutes.)

[4] Alarm set point for the hardness leakage

The evaluation ranges are 0mg/L, 1mg/L, 2mg/L, 3mg/L, 5mg/L and up.

[5] Evaluates hardness leakage at higher accuracy

When a hardness leakage is detected, monitoring is repeated a number of times

(concentration anomaly retry; selectable between 1 and 3 times) to prevent a temporary

fluctuation from triggering an alarm. In addition, such an abnormality must be repeated in a

series of monitoring at a preset interval for a number of times (response alarm cycle; also

selectable between 1 and 8 times). When all of these monitoring results indicate a hardness

leakage, it is evaluated that hardness leakage exists and the alarm is issued.

[6] Indicates data on the display screen

The display indicates the evaluated result, and the major cause of system error occurs in

the equipment.

[7] Offers an alarm function

When it is evaluated that there is a hardness leakage, the buzzer sounds. By the remote

alarm output, the alarm signal can be transmitted to a remote location.

Offers a self check function. If a system error occurs in the equipment, a typical cause will

be displayed in the same manner as the hardness leakage

[8] Stores records of hardness leakage

The equipment stores the records of occurrence date and time, evaluated result and

recovery date and time for each of the 5 latest incidents of hardness leakage. The records

may be utilized to analyze the causes of hardness leakage.

SECTION 2 OVERVIEW

13

[9] Requires minimal maintenance

The reagent cartridge may be replaced using a one-touch action. The reagent cartridge

needs no replacement for approximately 4 months in typical applications. (Note that more

frequent replacement may be necessary, depending on the application.)

[10] Compact in design, easy to install

The main equipment is installed easily on a wall.

Installation is a simple process.

[11] About advanced features

・Remote signal input function

The external signal controls whether to perform monitoring or not. This enables to

remotely stop monitoring while the system, which the water softener is regenerating.

・Remote alarm output

The remote alarm output, the alarm signal can be transmitted to a remote location as a

contact output in case that hardness leakage or system error occurs in the equipment.

・Regent cartridge exchange output

This output may be used to transmit to a remote location when a reagent cartridge

needs to be replaced.

SECTION 2 OVERVIEW

14

[12] Data capture tool

Up to 10,000(*1) monitor records (monitored date and results) can be stored in the

Colormetry. Also, by using a dedicated data capture tool installed on the PC (*2), the

monitor record data can be transferred to the PC. However, the Colormetry and PC must

be connected by an RS-232C cable (*3).

*1: Data for almost 1 year can be stored when monitoring at an interval of 60 minutes.

However, if manual monitoring is repeated, or if the monitor interval is set to less than

60 minutes, the data will be stored for less than 1 year. When the number of records

exceeds 10,000, the oldest monitor record data is deleted.

The data is in CSV format. The year-round trend can be managed through graph

creation software such as Microsoft Excel.

*2: Compatible PC OS is Windows® XP 32bit and Windows® 7 32bit.

*3: Use a female 9-pin D-Sub serial straight RS-232C cable (sold separately).

The cable length should be 1 m or less.

NOTE

If power is not supplied to the Colormetry for long periods of time, the internal memory

will be lost, and the record data will be initialized. If you do not intend to use the

Colormetry for a long time, capture the monitor data before cutting off the power. The

standard backup period is a maximum of 7 days in a 25°C environment.

Data capture connection

** Microsoft® Excel®, Windows® XP, and Windows® 7 are trademarks or registered trademarks

of Microsoft® Corporation in the United States and other countries.

RS-232C cable

Data transfer

SECTION 2 OVERVIEW

15

2 Name of Parts

2.1 Overall View

Manual monitor switch

(Buzzer reset switch)

LCD

displa

y

Cartridge lever

Reagent cartridge

Front cover

Filter casing assembly

Front cover fixing screws

System problem

or hardness

leakage alarm

LED

Reagent cartridge

replacement

notification LED

SECTION 2 OVERVIEW

16

5”(125)

33/4”(95)

4”(100) 1”(25) 1

3/4”(45)

3”(75)dia

1/2”(15)

2”(50)

43/ 4

”(120)

12”(

305)

31/ 2

”(90)

2

01/ 4

”(515)

83/ 4

”(225)

41/ 4

”(105)

3/ 4

”(20)

SECTION 2 OVERVIEW

17

2.2 Reagent Cartridge

Reagent injection nozzle

D-Ring

Tube

Check tube

SECTION 2 OVERVIEW

18

2.3 Filter Casing Assembly

Filter casing assembly

Filter casing assembly

Tube joint Constant flow valve

(the black rubber plate

installed on the top)

O-Ring

Filter case

(top)

O-Ring Fiber filter

Filter case

(bottom)

SECTION 2 OVERVIEW

19

3 Specifications

3.1 General Specifications

Power supply DC 24V (10%) (*1)

Power consumption 15W (on operation)

Ambient operating temperature 41ºF ~ 122ºF (5ºC ~ 50ºC)

Raw water temperature 41ºF ~ 104ºF (5ºC ~ 40ºC)

Humidity 20%RH ~ 90%RH

(without ice or dew condensation)

Raw water pressure 7.1~ 71psi

(0.05MPa ~ 0.49MPa)

Altitude Up to 2000m

Overvoltage category I (not directory connected to mains)

Pollution Degree 2

Installation method Indoor wall mount (*2)

EU Harmonized Standards EN61326-1:2013

This product is out of the scope of Low Voltage Directive.

**Usage Environment

Industrial area, Laboratory

(Note) This product may cause interference if used in residential areas. Such use must be

avoided unless the user takes special measures to reduce electromagnetic emissions to

prevent interference to the reception of radio and television broadcasts.

Do not install the product near a device that generates a large amount of electromagnetic noise.

Separate the power supply of this product from the power supply of devices that generate

electromagnetic noise.

There is a possibility that equipment does not perform normal operation.

3.2 Physical Data

External dimensions 5”[W] 4”[L] 201/4”[H]

(125[W] 100[L] 515[H] mm)

Mass 4.2lb (1.9kg)

3.3 Monitoring Capabilities

Evaluation method Via the Colorimetric method

Evaluation ranges Hardness: 0 ~ 5mg/L (*3)

Alarm ranges 1mg/L and up/2mg/L and up/3mg/L and up

5mg/L and up (Choose one)

Reagent cartridge replacement cycle Every 4 months (*4)

Remote alarm output (*5) Open collector output

Maximum rating: DC 24V 70mA Water passing output

Cartridge exchange output (*5)

Remote signal input No voltage contact input

(Contact A or Contact B)

Drainage Approx. 1000mL/monitor (*6)

**10,000 measurement results can be stored as record.

SECTION 2 OVERVIEW

20

3.4 Feed-water and Drain-water Connection

Feed-water connection diameter For connecting 6mm external dia. tube (*7)

Drain-water connection diameter For connecting 8mm external dia. tube (*7)

*1: The DC24V input should be supplied by a power supply (AC adaptor) as a LPS (Limited

Power Source) under the EN60950-1.

*2: The mounting bracket is attached to the equipment.

*3: Indicated value: 0,1,2,3,5mg/L or more (5 types)

*4: The warranty period of a reagent cartridge is 1 year for unopened product or 4 months after

opening the pack. The reagent will last approximately 4 months on the basis of an hourly

monitor. If monitors are taken every 30 minutes or repeated frequently, then the reagent may

not last for 4 months.

*5: The LEDs on the front cover turn on when an external alarm or when the reagent cartridge

needs replacement. Once the system error and hardness leakage are resolved, and the

reagent cartridge is replaced, each LED lights off.

*6: The drainage volume with a constant flow valve is listed. The drainage volume may vary

depending on the water temperature or degradation level of a constant flow valve.

*7: The feed-water and drain-water tubes are attached to the equipment.

Replace a tube if a defect such as deformation (a bend or fold), discoloration, hardening, or

cracking is detected on the tube surface. If a deteriorated tube is continuously used, it can rip

and cause water leakage. Make a daily check on whether the tubes are in normal condition.

Please use specified tubes when exchanging tubes. Using tubes other than specified tubes

may cause leakage.

Please contact your dealer or MIURA sales office.

3.5 Accessories

6mm dia. feed-water tube

8mm dia. drain-water tube

Fiber filter (cartridge type)

Filter casing assembly

Ball valve and other plumbing parts

Mounting bracket and other miscellaneous parts

User’s Manual

3.6 Consumable parts (Sold separately)

Reagent cartridge(CMU-H2)

Fiber filter (cartridge type)

SECTION 3 PRE-OPARATION INSPECTION

21

Do not operate the equipment if water is leaking.

Turning switches ON and OFF while water is leaking may lead to fire or

electric shocks.

Make sure the main power circuit breaker is turned OFF and contact your

dealer or MIURA sales office.

SECTION 3 PRE-OPARATION INSPECTION

1 Inspections and Preparations before Start up

1.1 Inspection and Preparation of the Surrounding Area Peripheral Equipment

Check the following points before starting operation of the Colormetry.

Make sure no water is leaking from the equipment and from around the piping.

・ Do not start the operation if a leakage is found. Instead, ensure that the main power

breaker is set to the OFF position, and then contact your dealer or MIURA sales office.

NOTE

Water Leakage Countermeasures

Factors such as loose connectors and aging of pipe materials can lead

to water leakage. It is important that drain ditches be setup for the

water treatment equipment and all peripheral equipment.

(In order to prevent water from leaking outside the room or downstairs,

set up a drain ditch that completely encloses the whole equipment,

ensure that the floor itself is waterproof, make the floor slope

downward to the drain port, and take any other necessary steps.)

WARNING

Drain ditch

SECTION 3 PRE-OPARATION INSPECTION

22

Do not place anything which might cause electric leakage underneath the

equipment.

The equipment is designed to drain water from the bottom of the equipment if

there is inner water leakage. Water might splash to the item under the

equipment.

2 Preparations before Start up

*Check the following items before start up.

Raw water pressure should be 7.1~ 71psi (0.05MPa ~ 0.49MPa).

The sample water collecting point (branching from main piping) to the filter casing assembly

should be within 1m.

If the Colormetry is installed much higher place, it might cause an evaluation error because of the

lack of sample water.

Open the ball valve at the sample water collecting point and let water run into the feed-water tube

(See Fig.3-1).

To the sample water collecting point

(branching from main piping)

Ball valve

6 mm dia. Feed-water tube

Filter casing assembly

8 mm dia. Drain-water tube

Figure 3-1 Feed-water system

CAUTION

SECTION 4 OPERATION

23

SECTION 4 OPERATION

1 Operation

1.1 Self Check Mode

When supply the power to the equipment, the self check starts automatically.

・If the equipment is working correctly, the self check completes in about 2 minutes.

・At its completion of the self check, the buzzer sounds 4 times and monitoring starts.

●Remote signal

When the remote signal is utilized, the equipment keeps in standby until the remote signal

permits monitoring.

●System error

If the equipment is not normal condition, the self check is repeated.

During a repeated self check, “Self Check Retry” is displayed.

If the equipment fails for the self check all 5 times, the buzzer sounds and he mode changes

to the system error standby mode and an error message displays in the LCD.

P44 Section 6 1 Self Check Function

NOTE

“Wash F :F265” or Wash Flow F:F086” is displayed and buzzer sounds during start up or

first operation after replacing the fiber filter in spite that the ball valve is open and pressure is

applied.

This is an initial phenomenon caused by bubbles in the filter casing assembly. It is not a

system error.

If the same alarm occurs, repeat this process several times.

P24 Section 4 1.3 Manual Monitoring

SECTION 4 OPERATION

24

1.2 Automatic Monitoring

When the present time of the monitor interval has passed, monitoring starts.

The monitoring start timing is also controlled by the remote signal.

*While monitoring, the monitoring stop time passed, the Colormetry back to standby as soon as

the monitor process completed.

1. Monitoring will take 2 to 5 minutes. The washing period will change from 40 to 200 sec.

according to the sample water volume.

2. If the system error occurs, the equipment will enter the system error standby mode.

P44 Section 6 1 Self Check Function

1.3 Manual Monitoring

Press the manual monitor switch to monitor

regardless of the monitor interval, remote signal,

monitor start or stop time settings.

If the Colormetry is under the high temperature

control mode and the sample water is hot, monitoring

does not start.

NOTE

The monitoring does not start, if the reagent cartridge is not installed.

Manual monitor switch

(Buzzer reset switch)

SECTION 4 OPERATION

25

2 FUNCTIONS

2.1 Remote Signal Input

2.1.1 Method

Monitoring the stagnant water in the piping may result in an evaluation error or a system

error of the equipment due to the lack of flow. To avoid such errors, the Colormetry provides

the following two methods, which may be used simultaneously.

Method No.1: Monitor start and stop times

The monitor start time [S Start] and monitor stop time [S Stop] settings limit the period

during which monitoring is performed. The provision helps avoid monitoring while water is

not used.

Example: Operation period: 8:00 through 17:00

Set the monitor start time [S Start] at 8:00 and stop time [S Stop] at 17:00

Method No.2: Remote signal input

Connecting the external contact with no-voltage enables to control the Colormetry for

monitoring only while the water softener is turned on, or to suspend monitoring while the

water softener is regeneration.

The purpose of remote signal is to prevent a monitoring while the system, which the

Colormetry is monitoring for, is stopped, such case as the water softener is at the

regenerating period.

2.1.2 Detailed descriptions

The remote signal input may be processed to enable monitoring in either two methods.

[1] The OFF state

[2] The ON state

The two methods are selectable in the setting mode [Set Mode]. Both methods achieve the

same objective (of preventing an evaluation error), though they process the signal

differently.

[1] Monitoring is enabled by remote being turned OFF [S Rte Sgl OFF] (the factory setting)

The OFF state of the remote signal input (the external contact is open) permits

scheduled monitoring after the monitor interval [S Intvl] as preset in the setting mode

[Set Mode].

The monitor interval [S Intvl] may be set in 30 minutes increments up to 240 minutes.

The reagent will last approximately 4 months on the basis of every 60 minutes

monitoring.)

SECTION 4 OPERATION

26

Example: Input regeneration signal from water softer.

When there a water softener contact has the contact which turns ON (closes) during

regeneration, and connecting this contact to the Colormetry, this remote signal

disables scheduled monitoring. When the softener contact turns OFF (open) at the

completion of regeneration, scheduled monitoring is again enabled.

[2] Monitoring is enabled by remote signal being turned ON [S Rte Sgl ON]

The ON state of the remote signal (the external contact closes) permits scheduled

monitoring after the monitor interval [S Intvl].

Example [2]-1: Input feeding water signal of water softener or other filtration equipment.

When there is a water softener or water treatment equipment has the contact which

turns ON (close) while feeding water, and connecting this contact to the Colormetry,

this remote signal enables scheduled monitoring only while feeding water. When water

stops, and the contact of a water softener or other filtration equipment turns OFF

(open), scheduled monitoring is not performed.

Example [2]-2: Input control signal of motor valve or solenoid valve

When a signal from a motor valve or solenoid valve that controls feeding water, is used

as a remote signal to the Colormetry, it enables monitoring only while the valve is open

and while feeding water.

Example [2]-3: Input supply tank water-level control signal

When a “requesting feed water” signal from the equipment that controls a water level of

a supply tank is used as a remote signal to the Colormetry, monitoring is performed

only for a while the request is made, and while feeding water.

NOTE

It is recommended to input a remote signal if it is possible, to prevent an evaluation error

or a system error. The feeding water signal, if available, should be utilized as the first

priority. If the regeneration signal is the only remote operation signal available from the

water softer, an evaluation error or temperature error may occur from monitoring the

stagnant water while the water is stopped, or a system error may occur due to the lack

of flow. A remote signal input arrangement, as shown in examples [2]-1, [2]-2 and [2]-3 is

highly recommended.

NOTE

If no remote signal is input, be sure to set to the OFF state [S Rte Sgl OFF] (the factory

setting).

SECTION 4 OPERATION

27

2.1.3 Delay time of remote signals

The purpose of remote signal delay time [S DelayTime]:

The setting determines the number of seconds the monitoring (filling with the sample

water) is to be delayed after receiving the remote signal. It is not commonly used

function, however it is effective to prevent an evaluation error in the example case

shown below.

Example of delay:

It is recommended that the sample water for the Colormetry should be taken off as

close to the main piping. However, in a case as illustrated below, where the sample

water is taken off from the secondary line of a motor valve or the valve required a

certain amount of time to open fully, the valve might not open at the time the

Colormetry received the remote signal from the water softer. A wash flow error [Wash

Flow F] may occur due to insufficient sample water pressure.

Raw water

Motor valve

Sample water for

Colormetry system

Remote signal

Water softener

equipment

Drain water from

Colormetry

Soft water

Colormetry

To prevent such a problem, select the remote signal delay time [S DelayTime] setting and

set the required time needed the valve to open fully and ensure the required raw water

pressure at the Colormetry inlet.

SECTION 4 OPERATION

28

2.2 Colormetry Monitor Timing

2.2.1 Automatic monitoring

The interval between monitoring is set in the monitor interval [S Intvl] setting at the setting

mode [Set Mode] (The interval is selectable in 30minutes increments between the 30 to 240

minutes range).

The initial time point of monitor interval [S Intvl] is the first monitoring which carried out soon

after the Colormetry is supplied the power or is reset. The next monitoring will start after the

preset time of the monitor interval [S Intvl] from this initial time point.

If the monitor interval [S Intvl] setting is changed, the initial time point is also changed to the

monitoring soon before the change is made. And the monitoring will start after the new

setting time from the last monitoring.

[1] Without remote signal input

Monitoring is performed periodically at the monitor interval [S Intvl].

[2] With remote signal input

a. If the remote signal setting is for “monitoring is enabled by remote signal being

turned OFF” [S Rte Sgl OFF]:

(Example of remote signal: Regeneration signal from water softener equipment)

When the preset time of the monitor interval[S Intvl] has passed, if the remote signal

is OFF, the remote signal delay time [S Rte Sgl dl] will start. If the remote signal is

continuously OFF while preset remote signal delay time, monitoring will start soon

after the remote signal delay time is finished.

When the preset time of the monitor interval [S Intvl] has passed, if the remote signal

is ON, the Colormetry keeps in standby until the remote signal is change back to OFF.

When the remote signal is back to OFF, the preset remote signal delay time [S Rte

Sgl dl] will start. If the remote signal is continuously OFF while preset remote signal

delay time, monitoring will start soon after the remote signal delay time is finished.

Remote signal delay time [S DelayTime]: 10 sec.

ON

OFF

Remote signal

Monitoring

10 sec 10 sec

No monitoring Monitor interval [S Intvl]: 60 min.

Monitor

interval

60 min

Monitor

interval

60 min

Monitoring

10 sec

Monitor time Monitor time

Water Softer

equipment is

regenerating

SECTION 4 OPERATION

29

b. If the remote signal setting is for “monitoring is enabled by remote signal being

turned ON ” [S Rte Sgl ON]:

(Example of remote signal: Feeding water signal from water softener equipment)

When the preset time of the monitor interval [S Intvl] has passed, if the remote signal

is ON, the remote signal delay time S Rte Sgl dl[] will start. If the remote signal is

continuously ON while preset remote signal delay time, monitoring will start soon

after the remote signal delay time is finished.

When the preset time of the monitor interval [S Intvl] has passed, if the remote signal

is OFF, the Colormetry keeps in standby until the remote signal is change back to

ON. When the remote signal is back to ON, the preset remote signal delay time [S

Rte Sgl dl] will start. If the remote signal is continuously ON while preset remote

signal delay time, monitoring will start soon after the remote signal delay time is

finished

Remote signal delay time [S DelayTime]: 10 sec.

ON

OFF

Remote signal

Monitoring

10 sec or

more

10 sec

No monitoring Monitor interval [S Intvl]: 60 min.

Monitor

interval

60 min

Monitor

interval

60 min

Monitoring

10 sec

Monitor time Monitor time

Precautions on operation with remote signal input

If the remote signal turns to disabling monitoring in middle of monitoring process, the

monitoring process will still be carried out to completion. However, the verification of

monitoring result is different by which state of monitoring when the remote signal is

received by the Colormetry

In case the remote signal turns to disable monitoring after water discharging from the

drain-water tube is complete and drainage is stopped by the solenoid valve in the

Colormetry is closed, the monitoring result will remain valid.

In case the remote signal turns before water discharge from the drain-water tube

completed and the solenoid valve is still open, the monitoring result will be processed as

follow:

a. If the concentration is evaluated as being lower than the alarm set point:

The result will remain valid and will be displayed as usual.

b. If the concentration is evaluated as being the alarm set point and up:

The evaluation will be invalidated and the previous result will be displayed.

Monitoring is disabled under the following conditions:

When the remote signal is turned OFF while the remote signal setting is [S Rte Sgl ON].

When the remote signal is turned ON while the remote signal setting is [S Rte Sgl OFF].

SECTION 4 OPERATION

30

[3] With monitor start time [S Start] and stop time [S Stop] are set up

a. When the preset time of the monitor interval [S Intvl] has passed, but not monitor start

time [S Start] yet, the Colormetry keeps in standby until the monitor start time. When

the monitor start time came, the preset remote signal delay time will start and

monitoring will start soon after the remote signal delay time is finished.

b. Without a remote signal input, monitoring will be performed at Monitor interval [S Intvl]

only during the period between Monitor start time [S Start] and stop time [S Stop].

c. With a remote signal input, monitoring will be performed in the same manner as in

paragraph [2] with remote signal input, only during the period between the monitor start

time [S Start] and stop time [S Stop].

Monitor Start time [S Start] 8:00

Monitor Stop time [S Stop] 18:00

Monitor interval [S Intvl] 60 min.

Remote signal delay time [S DelayTime] 10 sec

ON

OFF

Remote signal

Monitoring

10 sec 10 sec

No monitoring

7:00

No monitoring

8:00 9:00

Monitor time Monitor time

d. If it becomes Monitor stop time [S Stop] during the monitoring process, the system will

enter Monitor standby mode upon the completion of that monitoring.

NOTE

If the remote signal setting is for “monitoring is enabled by remote signal being turned

ON” [S Rte Sgl ON], periodically check the remote signal.

If no remote signal is available due to a malfunction, the monitoring process will never

start.

2.2.2 Manual monitoring

Press the manual monitor switch to monitor regardless of the monitor interval, remote signal,

monitor start or stop time settings.

(The monitoring will not start, if the reagent cartridge is not installed. The Colormetry may

start the self check on a system error.)

SECTION 4 OPERATION

31

2.3 Concentration Evaluation

The Colormetry evaluates ionic concentrations in 5 levels: 0mg/L, 1mg/L, 2mg/L, 3mg/L, and

5mg/l.

Alarm set point may be set to trigger at 1mg/L and up, 2mg/L and up, 3mg/L and up and 5mg/L

and up.

NOTE

For boiler use, select 1mg/L or 2mg/L as the alarm set point or follow the equipment in the

boiler.

2.3.1 Manual monitor

If a monitoring result is below the Alarm set point [S AlarmSet], as set in Setting mode [Set

Mode], it will be evaluated as normal. The result of such evaluation will be displayed and

monitoring is complete.

Mode Sample LCD

*1: The sample display, shown in the table, represents an evaluation within the 0mg/L range.

If the alarm set point is set at 2mg/L and the evaluated result is 1mg/L range, the display

on LCD will be [Result: 1mg/L].

(*1)

(*1)

Monitor

Monitor standby

Displaying evaluation result Monitor standby

Monitor On

Result:0mg/L

If monitoring result < Alarm set point [S AlarmSet]

Result:0mg/L

SECTION 4 OPERATION

32

2.3.2 Concentration anomaly

[1] If an evaluation result is the alarm set point [S AlarmSet] and up, monitoring is repeated

the number of times as set (between 1 to 3) in the concentration anomaly retry [S Alarm

Inc No] setting. If all retry results have been the alarm set point [S AlarmSet] and up, only

then, the monitor result will be determined as concentration anomaly.

If, on the other hand, a retry results become to lower than the alarm set point [S

AlarmSet] setting, the condition will be determined to be normal and monitoring will be

finished.

[2] However, an anomaly evaluation made in step [1] alone would not trigger a high

concentration anomaly alarm (that is, to sound buzzer and close the remote alarm output

contact).

In addition, such an anomaly evaluation must be repeated in a series of monitoring at a

preset interval [S Intvl] for a number of times. The response alarm cycle [S Alarm Det No]

is set the number of times(selectable between 1 to 8 times). When all of these monitoring

results indicate a concentration anomaly, it is finally evaluated that a concentration

anomaly exists and the alarm is issued.

YES

Result:0mg/L

N1: Concentration anomaly retry

[S Alarm Inc No]

N2: Response alarm cycle

[S Alarm Det No]

Displays evaluation

result

Monitor standby

Monitor

Result:0mg/L

Monitor On

Num. of monitoring

performed < N1

Displays evaluation

result

Hardness Leakage

Alarm is issued

YES

NO

NO

Num. of concentration anomaly < N2

SECTION 4 OPERATION

33

[3] If a concentration anomaly occurs repeatedly in a series of automatic monitoring at the

monitor interval [S Intvl], or in manual monitoring, the concentration anomaly alarm will

stay on continuously.

[4] The concentration anomaly alarm is automatically terminated (the buzzer stops and the

remote alarm output contact opens) upon that the condition is determined normal in

automatic monitoring at the monitor interval [S Intvl], or in manual monitoring.

Example of Monitor 1:

Abnormal condition retry [S Alarm Inc No] setting: 2

Response alarm cycle [S Alarm Det No] setting:2

Monitor interval

[S Intvl]

Monitor interval

[S Intvl]

Monitor A Monitor B

Monitor interval

[S Intvl]

Monitor C

Monitor D

Concentration anomaly

alarm is issued

Condition is evaluated

to be abnormal

Normal condition Concentration anomaly

Concentration anomaly Concentration anomaly Concentration anomaly

Concentration anomaly Concentration anomaly

Condition is evaluated

to be abnormal

Condition is evaluated

to be abnormal

Concentration anomaly

alarm is continued

Monitor A: The result is below the alarm set point. The condition is normal.

Monitor B: The result is higher than the alarm set point. Since the concentration anomaly retry

setting is for two times, another monitoring is performed, which again is higher.

The condition in monitor B is evaluated as a concentration anomaly.

Monitor C: As monitor B, the both results are higher than the alarm set point. The condition in

monitor C, is also evaluated as a concentration anomaly.

A concentration anomaly has been verified twice, in monitor B and C, so that the

concentration anomaly alarm is now issued.

Monitor D: As monitor C, the results are higher than the alarm set point.

The condition in monitor D is evaluated as a concentration anomaly, following an

already issued concentration anomaly alarm. The alarm will continue.

SECTION 4 OPERATION

34

Example of Monitor 2:

Abnormal condition retry [S Alarm Inc No] setting: 3

Response alarm cycle [S Alarm Det No] setting: 1

Monitor interval

[S Intvl]

Monitor interval

[S Intvl]

Monitor A Monitor B

Monitor interval

[S Intvl]

Monitor C

Monitor D

Concentration anomaly

alarm is issued

Normal condition Concentration anomaly

Concentration anomaly Concentration anomaly

Concentration anomaly Normal condition

Condition is evaluated

to be abnormal

Concentration anomaly

alarm is terminated

Normal condition Concentration anomaly

Monitor A: The result is below the alarm set point. The condition is normal.

Monitor B: The result is higher than the alarm set point. Since the concentration anomaly retry

setting is for three times, another monitoring is performed, which again is higher.

The third retry result, however, is below the alarm set point. The condition in

Monitor B is normal.

Monitor C: The first three results are higher than the alarm set point. The condition in monitor

C is also evaluated as a concentration anomaly.

A concentration anomaly verified once, since the response alarm cycle is set as

once, a concentration anomaly alarm is now issued.

Monitor D: The result is below the alarm set point. The condition in monitor D is normal. The

concentration anomaly alarm is now automatically terminated.

NOTE

If a result is evaluated as a concentration anomaly and carrying out the series of

monitoring as the concentration anomaly retry, if one of the monitoring result is

invalidated so that the series of monitoring fails to satisfy the concentration anomaly retry

[S Alarm Inc No] requirement, the evaluation of the session will still evaluated as a

concentration anomaly.

SECTION 4 OPERATION

35

2.3.2.1 Display on evaluation result (*1)

Switched automatically

*1: The sample display represents an evaluation exceeding 2mg/L

2.3.2.2 Output on concentration anomaly

a. When a concentration anomaly alarm is issued, the buzzer sounds. In case that the remote

alarm output [S Rte Alarm] is NC, the remote alarm output contact opens. In case that the

remote alarm output [S Rte Alarm] is NO, the remote alarm output contact closes.

b. When a concentration anomaly alarm is issued, pressing the manual monitor (buzzer reset)

switch on the front cover of the equipment will stop the buzzer.

The remote alarm output contact, however, will remain closed until the condition is

evaluated as normal in an automatic monitoring at the monitor interval [S Intvl], or manual

monitoring.

c. Once the buzzer is stopped by pressing the manual monitor (buzzer reset) switch, it will

remain disabled if carry out the manual monitoring immediately and the result was

evaluated as a concentration anomaly.

If, however, a condition is once evaluated as normal in automatic monitoring at the monitor

interval [S Intvl] or manual monitoring, then the buzzer will activate for another result of a

concentration anomaly next time.

Result: 2mg/L

Hardness Leakage

SECTION 4 OPERATION

36

2.3.2.3 Workings of hardness leakage alarm and action to take

Hardness leakage alarm issued

Buzzer starts sounding

Remote alarm output contact closes

Press Manual

monitor switch?

Buzzer sounds

Display evaluation result

NO

Buzzer stops

YES

[1] The buzzer sounds for a concentration anomaly.

[2] The buzzer stops by pressing the manual monitor (buzzer reset) switch.

(The Colormetry will not automatically start manual monitoring. To start manual monitoring,

press the manual monitor (buzzer reset) switch again.)

(The LCD will remain the same.)

NOTE

The remote alarm output contact will not be cancelled until the condition is evaluated as

normal.

SECTION 5 INSPECTION AND MAINTENANCE

37

Never disassemble the reagent cartridge.

Reagent may splatter onto the skin or the eyes.

If the reagent gets on the skin or in the eyes, rinse immediately with water.

SECTION 5 INSPECTION AND MAINTENANCE

1 Reagent Cartridge

1.1 Reagent Cartridge Replacement Timing

While the self check, if “New Cartridge” is displayed, the remained volume of reagent is low

and the evaluation error might occurs. For the new reagent cartridge, please contact your

dealer or MIURA sales office.

1.2 Cautions for Handling

The cartridge life is about 1 year from the date of manufacture, and is used up within

approximately 4 months of installation. Check the manufacturing date on the package and

replace the reagent cartridge over the life span.

Do not store the reagent cartridge long time. Store the reagent cartridge in a cool, dark place.

Do not open the bag until it actually use. If it is open, the reagent degrading might be

accelerated.

When installing a reagent cartridge, insert it slowly, keeping the end of the tube from hitting

the body of the equipment. It might affect to the chemical feeding volume and cases

evaluation error or damage completely.

Do not use the reagent cartridge for the use other than the Colormetry.

Do not disassemble the reagent cartridge and dispose it as chemical waste in accordance

with local regulations.

For the detail information of reagent, refer to the safety data sheet.

WARNING

SECTION 5 INSPECTION AND MAINTENANCE

38

1.3 Reagent Cartridge Replacement Method

NOTE

Replace the reagent cartridge while the power is supplied to the equipment. Close the sample

water collecting ball valve and replace while monitoring standby period.

When install the reagent cartridge, attach the D-Ring first. To install D-Ring to the reagent

cartridge, dip the D-Ring into water.

If the reagent cartridge is removed from the equipment other than replacement and put it back,

do not push the manual monitor (buzzer stop) switch for about 60 sec. after install the old

reagent cartridge. If you press the manual monitor (buzzer stop) switch soon after installation,

the reagent cartridge replacement timer counter will be reset.

Reagent cartridge replacement

Cartridge OFF

①Pull out the cartridge lever to unlock

②Hold the regent cartridge by both hand and pull up to

remove from the equipment.(Buzzer starts sounding)

③Press the manual monitor

switch to stop the buzzer.

④Insert the new reagent cartridge on top of the

equipment firmly.(Buzzer starts sounding again)

⑤Push the cartridge lever

back in to lockSW ON Cfm

⑥Press the manual monitor

switch to stop the buzzer.

The self check starts. Self Check

*Press the manual monitor

switch within 60 sec. from

buzzer starts.

Figure 5-1 Reagent cartridge replacement

①⑤

③⑥

②④

D-Ring

SECTION 5 INSPECTION AND MAINTENANCE

39

Do not place anything which might cause electric leakage underneath the

equipment.

The equipment is designed to drain water from the bottom of the equipment if

there is inner water leakage. Water might splash to the item under the

equipment.

2 Fiber Filter and Constant Flow Valve Replacement

2.1 Fiber Filter Replacement Timing

If “Wash F : F265”, “Injection F : F281” and “Wash Flow: F0866” displayed even though the

raw water pressure is the normal range and there are no other cause, the fiber filter might

clogged.

The filter life is about an year, but it might be shorten depend on the water quality.

If there are no alarm displayed, replace the fiber filter every year.

2.2 Fiber Filter Replacement

2.2.1 Remove filter casing assembly

1) Close the sample water collecting ball valve.

2) Press the manual monitor (buzzer stop) switch to carry out manual monitoring to reduce the

internal pressure.

3) 15 sec. after pressing the manual monitor (buzzer stop) switch; cut the power supply off from

the equipment.

NOTE

When you take off the feed-water tube, push the release bush on the tube joint and pull the tube

out (refer to Fig. 5-3). If you pull the tube without pushing the bush, end of the tube clogged in the

tube joint and small pieces of the tube might be remained. When you push the new tube in, it

might not able to push it through.

4) Pull off the feed-water tube from the filter casing assembly.

5) Take out the filter casing assembly from the Colormetry.

6) When you take out the filter casing assembly, the constant flow valve is on top of the filter

casing assembly. If it is not there, it might remain in the Colormetry. If so take it out gently.

CAUTION

SECTION 5 INSPECTION AND MAINTENANCE

40

2.2.2 Fiber filter replacement

1) The filter casing assembly comes apart into two sections. To take them apart, rotate the top and

bottom sections in a counterclockwise direction.

2) Remove the fiber filter cartridge from the box (or bag). Insert the tip end of the fiber filter into the

center of the filter case (top). Then, firmly tighten the filter case (top) and filter case (bottom) back

together.

Filter casing assembly

Constant flow valve (the black rubber plate)

Fiber filter

Tube joint

Filter case (bottom)

Filter case (top)

6 mm dia.Feed-water tube

Figure 5-2 Fiber filter replacement

SECTION 5 INSPECTION AND MAINTENANCE

41

Do not apply grease or lube when install the filter casing assembly. It might

damage the material of the equipment and cause water leakage.

2.2.3 Filter case assembly installation

NOTE

When you install the filter casing assembly, simply screw it in by hand. Do not use any kind of

fitting tool. If you use tools to screw in with excess power, the filter casing assembly and the

bottom plate of the Colormetry might be broken.

1) Attach the constant flow valve on top of the filter casing assembly, then screw in the filter

casing assembly into the Colormetry.

NOTE

If using the removed tube once again, cut the end of the tube clean and even by the tube

cutter. Exercise with care to avoid injury when using the tube cutter.

2) Push in the feed-water tube to the filter case assembly

3) Open the sample water collecting ball valve and make sure there is no leakage.

4) Supply the power to the Colormetry.

Screw the filter casing assembly into

the bottom of the equipment and insert

the 6mm dia. feed-water tube into the

joint firmly.

※If the filter casing assembly is  

 screwed with the tube, the tube or the

tube joint may turn around and

damage itself, then the connection

becomes loose.

Before screwing in the filter

casing assembly, be sure that

the constant flow valve is fitted.

Bundle the tube together with a

cable tie.

Be sure that no undue force is

exerted on the tube.

※If the tube is tighten too hard

until it changes its shape,

it can crack and cause water

leakage.

Attach the warning label "Don't apply

grease or lube" to the filter casing

assembly, after that, place it to be

seen from the front.

Insert the 8mm dia.Drain-water tube properly into the Cylindrical part.

Figure 5-3 Filter case assembly installation

CAUTION

SECTION 5 INSPECTION AND MAINTENANCE

42

Do not use the equipment if there is smoke, abnormal odor or noise,

excessive overheating, or other abnormalities. If an abnormality occurs, do

not operate the equipment. Also, make sure an earth leakage circuit breaker

is turned OFF and a main feed-water valve is closed and contact your dealer

or MIURA sales office.

Failure to observe this precaution may lead to electric shock, fire, or a fault.

Replace a tube if a defect such as deformation (a bend or fold), discoloration,

hardening, or cracking is detected on the tube surface. If a deteriorated tube is

continuously used, it can rip and cause water leakage.

Please use specified tubes when exchanging tubes. Using tubes other than

specified tubes may cause leakage.

3 Daily Inspection

3.1 Inspection Item

NOTE

Please wipe with dry soft cloth when cleaning.

Rubbing with hard cloth may make scratches.

3.1.1 Monitor operation

Make sure it is monitoring automatically. If the remote signal input is used, but the remote

signal has some problem, the automatic monitoring does not start.

P24 Section 4 1.2 Automatic Monitoring

3.1.2 Feed-water and drain-water tube

Exercise with care to keeping the tubes free from bents or holds.

If the feed-water tube is bent, it might cause insufficient feed water volume.

If the drain-water tube is clogged, the internal pressure builds up in the equipment and

causes water leakage. Use specified tubes when exchanging tubes. Using tubes other than

specified tubes may cause leakage.

CAUTION

WARNING

SECTION 5 INSPECTION AND MAINTENANCE

43

If there is a defect or cracking on the filter casing assembly, turn OFF an earth

leakage circuit breaker and close a main feed-water valve. Then, replace the

filter casing assembly.

If water leakage is not stopped even though tightening the filter casing

assembly and tube joint, replace the filter casing assembly.

3.1.3 Filter case assembly

3.2 Others

・When the hardness leakage occurred (CMU-324HE)

The hardness leaking is very low concentration such as 1.0 mg/L or 2.0 mg/L, it cannot

determined by the hardness indicator. To determine, carry out the water quality analysis.

Collect the sample water from raw water into the Colormetry or utilize drain water which does

not colored by reagent used in the Colormetry.

CAUTION

SECTION 6 BREAKDOWN AND COUNTERMEASURES

44

SECTION 6 BREAKDOWN AND COUNTERMEASURES

1 Self Check Function

1.1 Display

If the Colormetry fail for the self check, the major cause is displayed in the LCD.

If you cannot handle, please contact your dealer or MIURA sales office as soon as possible.

1.2 Alarm

Concentration anomaly

alarm is issued

Buzzer starts sounding

remote alarm output contact close

Press Manual

monitor switch?

Buzzer sounds Buzzer stops

Displays evaluation result

Monitor interval

YES

NO

Figure 6-1 Flowchart when alarm occurs

NOTE

It is possible to cut the buzzer from the beginning.

Check the Installation and Start Up User’s Manual.

1.3 Alarm Reset

・The buzzer stops by pressing the manual monitor (buzzer reset) switch.

The Colormetry will not automatically starts manual monitoring and error message displays

at LCD for 3 second. If you press the manual monitor (buzzer reset) switch once again,

another self check starts.

・The automatic monitoring will carry out even though the alarm condition.

If the error caused an alarm is resolved, the alarm is reset automatically.

If alarm is caused because the reagent cartridge is not installed, monitoring will not start until

the equipment confirms the reagent cartridge is installed.

SECTION 6 BREAKDOWN AND COUNTERMEASURES

45

2 Self Check Items

NOTE

If the problem did not solve by the procedure on below table, contact your dealer or MIURA

sales office.

Table 6-1 Display and Solution of a Self check items

Display Cause Solution

Cartridge_OFF_F

Reagent cartridge is not installed or not

installed well

Firmly installed the reagent

cartridge and lock the

cartridge lever

Wash Flow F:F086

Sample water volume is reduced.

Raw water pressure is low.

Raw water pressure should be

on or above 0.05MPa

Feed-water and/or drain-water tube is

clogged.

Fix the tube from holds or

bends

Filter is clogged Replace the fiber filter

Wash F :F265

Insufficient wash

Ball valve at the water collecting point is

closed

Open the ball valve

Feed-water and/or drain-water tube is

clogged

Fix the tube from holds and

bends

Filter is clogged Replace the fiber filter

High Temperature

Sample water is too hot

Should not feed water its temperature

above 50 dig. C to the Colormetry

Check the sample water

temperature

New Cartridge Rplacment timing of the regent cartridge Replace the reagent cartridge

Injection F F281 Insufficient reagent injection Replace the reagent cartridge

Thmsta F :C131 Thermista is broken or disconnected Contact your dealer or MIURA

sales office

Pump F :C444 Chemical feed pump motor or sensor is

malfunction

Contact your dealer or MIURA

sales office

Photo Rpt F:C365 Light source or photoreceptor is

malfunction

Contact your dealer or MIURA

sales office

SelfCheck F:A000

Data F :C500

Control board is malfunction Contact your dealer or MIURA

sales office

Comm F :F090

Communication system does not match.

Power failure in the communication board

Recover from power failure

Short circuit or malfunction on the

communication board.

Contact your dealer or MIURA

sales office

SECTION 7 STORAGE

46

Do not perform any unapproved installation work or connect any peripheral

equipment.

Failure to observe this precaution may lead to personal injuries.

SECTION 7 STORAGE

1 Extended Inactivity

If the equipment is to be inactive for an extended period of time, Contact your dealer or MIURA

sales office.

NOTE

If the equipment will not be used for an extended period of time (more than a week), it should

be stored properly. If the equipment is not used, it is often overlooked and is not kept

properly.

If not properly maintained while idle, the equipment may not operate correctly when

subsequently started up.

2 Transfer or Resale

If you intend to relocate or sell your MIURA water treatment equipment, contact your dealer or

MIURA sales office in order that installation work can be carried out correctly and that the

peripheral equipment may be set up. In addition, ensure that User’s Manual and the like are kept

in order that they may also be provided to the new owner in the event of resale.

3 Export

This product was manufactured for use in the country where MIURA judges that the export of

the product is allowable.

Therefore, the full performance may be inhibited due to differences in operating environment if

the product is used outside the current country of use.

Regarding export, you are to comply with the laws and regulations.

You will also come under control in many cases, with regard to export, on-site import, and

on-site use according to the laws and regulations of the destination for the export.

When exporting the product, contact your dealer or MIURA sales office.

WARNING

SECTION 8 DISPOSAL

47

SECTION 8 DISPOSAL

1 Disposal

When disposing of the product, comply with the laws and regulations.

SECTION 9 WARRANTY

48

SECTION 9 WARRANTY

1 Note Regarding Warranty

Regarding the contents of the warranty, contact your dealer or MIURA sales office.

2 Questions Regarding the Product and User’s Manual

For any questions about your purchased product or the content of this User’s Manual, contact

your dealer or MIURA sales office.

3 If the User’s Manual is Lost

The User’s Manual contains the information for the safe usage.

In the event of loss of the User’s Manual, contact your dealer or MIURA sales office.

AUTHORIZED REPRESENTATIVE in EU

Wolfgang Kiebert

Industrie- und Verfahrenstechnik

Dipl.-Ing. Wolfgang Kiebert

Kapellenstr. 19

D-54597 Auw bei Prüm

Tel.: +49 (0)6552 92 99 000

Fax: +49 (0)6552 92 99 004

Mobil: +49 (0)172 42 22 416

email:[email protected]

GERMANY, Mar 2017