GTE Industrieelektronik GmbH Measuring Technology Product Division GTE Industrieelektronik GmbH Tel. +49-(0)2162-3703-0 Helmholtzstr. 38-40 Fax +49-(0)2162-3703-25 41747 Viersen E-mail: [email protected] Germany Internet: www.gte.de FPM Vision I Quick Start Guide for KMG 300/500 System with Fuji-Prescale Films 325-2810-002 EN02

GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

  • Upload
    others

  • View
    14

  • Download
    0

Embed Size (px)

Citation preview

Page 1: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH Measuring Technology Product Division

GTE Industrieelektronik GmbH Tel. +49-(0)2162-3703-0 Helmholtzstr. 38-40 Fax +49-(0)2162-3703-25 41747 Viersen E-mail: [email protected] Germany Internet: www.gte.de

FPM Vision I

Quick Start Guide

for KMG 300/500 System with Fuji-Prescale Films

325-2810-002 EN02

Page 2: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 2 of 15

Safety information Read this documentation carefully and familiarize yourself with the product prior to its use. Pay particular attention to the safety information and warnings to avoid injuries and product damage. Keep this documentation readily available for later use as needed. Pass on this document to future users of the product. To avoid physical injury or significant damage to property, only qualified persons should be allowed to work with the measuring equipment. Qualified persons are deemed to be those who are familiar with the setup, installation, initial operation and running of collaborative robots and who have the requisite qualifications for this kind of activity. They must be able to diagnose the tasks they are asked to conduct, recognise potential sources of risk and take adequate safety measures.

Purpose The software is designed for the visualization, analysis, editing, and archiving of force measurements Recorded using the KMG 300-75 and KMG 500-75 measuring systems and of pressure images generated using Fuji-Prescale films. The measurement method is used for determining impact and clamping forces at collaborative robots based on ISO/TS 15066. It may only be used for this purpose. The software specifies the configuration of compression elements on basis of the selected body region. It offers an import function for printed images with automatic filtering and color calibration. It features an integrated report generation for data archiving purposes.

Scope of the system

• KMG 300 and/or KMG 500 force gauge • Power supply: AC100-240V to DC5V • Voltage converter DC12-24V to DC5V • USB-A to USB-B Mini cable • Transport case for force gauge • Compression elements K1 • USB stick with PC software FPM Vision • Mounting base with adapter • EPSON scanner V370 • Scanner power supply • USB-A to USB-B cable • Calibration element for scanner • Fuji-Prescale film, 2-piece

FPM VISION software updates Due to possible normative changes further updates may be necessary. You can find the current version of the software on our homepage www.gte.de/downloads. We recommend visiting our webpage regularly and after normative changes and checking the software for available updates.

Page 3: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 3 of 15

Software license

The software is available in several configurations. The function modules are enabled using a license key. You will receive a respective license key when purchasing the measuring gauge and software in your desired configuration.

The license key is entered at the factory, when the device is shipped. After the installation, the software is ready according to the purchased license.

An update, e.g., from “force analysis” to “force and pressure” analysis, is possible without returning the measuring device. When purchasing the extended license, you will receive a new license key you can enter via the “KMG\License” menu. For this purpose, the measuring device must be connected to the PC. Next, the device can also be used at all other stations.

The PC software may be installed on 10 computers/stations. As soon as the KMG 300 or KMG 500 is connected, the enabled functions can be used.

Installation of the FPM Vision software System requirements: WIN7 or WIN 10, MS Word 7 or higher

• Unzip the software zip file (FPM Vision X.X) on your data carrier.

• Start file .\ Setup.exe

• Follow the installation instructions and install the software completely.

• You may be asked to install additional drivers.

• If asked, connect a KMG device with PC using the enclosed USB cable.

• Restart the computer.

• You can find shortcuts to the software on the desktop and in the start menu

• Install the driver of the EPSONV370 scanner (EPSON CD).

Support: For a fast and effective response to your questions, access our website www.kolrobot.com/support via the “Support” menu item under the “Help” menu. Here you can ask your question and send it to the correct contact. Here you can also find a FAQ area with information to the most important questions. Tip: The software screen contains a “Tip” button. Click it for tips on handling the software and for explanations of diverse fields.

Page 4: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 4 of 15

USB connection of the KMG The force gauge features an USB connection for transferring the measured data of the KMG 300/500. In the case of a complete installation, all necessary drivers are already on the PC. You can find further device information in the respective operating instructions.

EPSON scanner with calibration element The EPSON scanner of type V370 is required for the pressure analysis. The scanner properties are decisive for a scan followed by pressure force analysis. An alternative scanner may only be used after previous consultation. The enclosed calibration element must be used for every scan. It is necessary for the correct conversion of the color intensities into pressure values. Make sure the alignment on the scanner is correct (details, see page 13). The calibration element can fade due to light exposure or become damaged due to other circumstances. The software checks the functionality of the calibration element and displays the status after scanning using 6 elements. The 6 elements must all be illuminated. If this is not the case:

- Check the calibration element and QR code for damage. - If damage is present, the calibration element must be replaced. - Check the scanner area for residues and contaminations. - Clean the calibration element. - Check the scanner functionality by scanning a full-area image as needed. Check for irregularities in the

color gradient. Irregularities indicate a defective scanner. The calibration element is available as spare part from GTE GmbH.

Page 5: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 5 of 15

Starting the software The following shortcuts were installed during installation to start the software:

• 'FPM Vision' desktop icon • Start menu\ All programs\ FPM Vision \ FPM Vision

Menu overview File menu:

� New: New data report is opened. � Open: An existing analysis is opened. � Template from file: Loads the data from an existing analysis into a new data report � Storage path: Defines the superordinate folder for measured and report data. � Change language: Selection of the desired language. � Settings: The Settings menu is opened. � Exit: FPM Vision is exited.

Evaluation menu:

� Generate report: Generates a new report from the completed measurement using the current template.

� Edit template: Opens the template file. Edits and saves the new template. � Rest template: Resets the current template to the standard. � csv Export: Compilation of several measurement data into a table in csv format.

KMG menu:

� Export measurement data: Exporting the measurements from the KMG 300/500 into a csv file. � License: Inputting or changing the license key.

Help menu:

� Manual: Download link for the operating instructions � Support: Opens our online support area. � Version information: Program version and information on sources. � GTE: Street address of GTE GmbH

Default settings – system configuration After installation completion, the system is configured and ready for operation. The calculation basis for the configuration of the measuring equipment is preset to the calculation basis according to the BGHM – Employer's Liability Insurance Associations. The calculation basis can be changed as needed via the File\Settings menu item. Select the File\Settings menu to access the settings. The window displays the standard configuration settings for the components. We recommend not changing the standard settings.

Page 6: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

Subject to technical changes

Typical measuring setup To ensure a correct force and pressureaccording to the configuration specifications. For recording the force profile, the KMG 300/500 must be firmly secured, placed and aligned. The force application should be perpendicular to t The required compression element is also secured on the measuring surface. Adhesive stripes attached to the outside can be used for this purpose. Furthermore, the Fuji-Prescale film is cut to a suitable size and secured on the compr This measuring setup is used for recording force and pressure. Also write down the measuring number on the pressure image.

Information on the functionality of the Fujiwith the rolls.

GTE Industrieelektronik GmbH

FPM Vision – EN02

Typical measuring setup

To ensure a correct force and pressure measurement, the measuring equipment must be selected and placed according to the configuration specifications. For recording the force profile, the KMG 300/500 must be firmly secured, placed and aligned. The force application should be perpendicular to the measuring surface.

The required compression element is also secured on the measuring surface. Adhesive stripes attached to the outside can be used for this purpose.

Prescale film is cut to a suitable size and secured on the compr

This measuring setup is used for recording force and pressure. Also write down the measuring number on the

Information on the functionality of the Fuji-Prescale films can be found in the original Fuji instructions enclosed

GTE Industrieelektronik GmbH

Page 6 of 15

measurement, the measuring equipment must be selected and placed according to the configuration specifications. For recording the force profile, the KMG 300/500 must be firmly

he measuring surface.

The required compression element is also secured on the measuring surface. Adhesive stripes attached to the

Prescale film is cut to a suitable size and secured on the compression element.

This measuring setup is used for recording force and pressure. Also write down the measuring number on the

Prescale films can be found in the original Fuji instructions enclosed

Page 7: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 7 of 15

Step 1: Configuring the measuring equipment

I. Create a new data report via the File\New menu. Connect the force gauge KMG 300/500 to the PC. Select the force sensor and the pressure sensor in the top right area. The available sensors are displayed for selection depending on the license. The serial number of the device is displayed indicating the successful connection to the KMG 300/500. Important: Select the type of film used. An incorrect selection will impact the measuring result!

II. Switch to the “Biomechanical configuration” tab and select the body region using the drop-down menu.

All body locations corresponding to the body region are shown on the body model. The numbers correspond to the numbering of the standard.

III. Select the suitable measuring point using the “Specific location” drop-down menu. The selected point is marked highlighted in green. Tip: You can also select specific locations per mouse click on the body model.

IV. Next, the currently applicable thresholds for this collision point are displayed underneath the menu.

V. The compression element K1 to be used and the required spring rate are shown on the right side of the window. For the KMG 300/500 force sensors, this rate is fixed. The compression elements are selected automatically based on the preset calculation basis (see “Default settings – system configuration”).

VI. If – deviating from this – an alternative compression element is used, the setting can be changed from “Automatic” to “Manual”. Different compression elements can then be selected using the then available drop-down menus. Using the menus, the desired and available compression elements and springs can be selected (only in connection with the “Piezo” force sensor type).

Sensor selection

Compression element and spring constant

Select body location Thresholds for body location

Update after change

Page 8: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 8 of 15

Step 2: Report data of the measurement Additional fields for report data and parameters of your measurement are available in the top left area of the window.

• Use descriptive names for the “Test series” and “Test name” fields. Directories with the same name will be created in the storage path, when the measu rement is saved!

• Enter the name of the tester.

• You can indicate time and location.

• Fields “ Temperature” and “ Relative humidity” are parameters and influence the presented measurement result of the Fuji films! For this purp ose, these values must be determined on site using a suitable measuring device.

• The “Edit me” fields are used for further documentation, e.g., to specify diverse parameters, such as the collision form or the name of the test object. Field headers can be edited and renamed by clicking on the “Edit me” fields.

• Using the “File\Template from file” menu function, the content of an already stored measurement can be imported. For this purpose, select the respective measurement zip file.

Input options for measurement conditions. Tip: The headers can be renamed by clicking on them.

Page 9: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 9 of 15

Step 3: Importing measurement data The KMG 300 and KMG 500 feature an integrated analysis and memory. After a measurement, the force profile is displayed graphically. Furthermore, the results for static and dynamic/transient force can be read out directly on the device's display. This way, measurements can be performed and analyzed on site independently from a laptop/PC. The measurements are numbered and saved. After all pressure measurements are completed, the data can be imported. Tip: Film can be numbered using a pen or similar. Note: if possible, the pressure measuring film should be imported immediately. Preparation:

• Connect the KMG 300 or KMG 500 with the laptop/PC.

• Connect the scanner with the laptop/PC and switch it on.

• Place the pressure measuring film together with the calibration sheet onto the scanner (see notes page 13).

• Prior to importing the pressure data, check once mo re that the correct film type is set and that the temperature and air humidity data is correct!

Import:

Switch to the “Measurement data” tab for importing. The memory data of the force gauge is automatically loaded.

By operating the “Force measurement data import” drop-down field, you can load the desired measurement from the KMG. For this purpose, the device must remain connected.

After selection of the force measurement, the corresponding pressure measurement can be imported. For this purpose, press the “Scan” button. Select scanner V370. Scanning starts.

Page 10: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 10 of 15

Import:

Next, the scan is displayed.

Press and hold the left mouse button and select the relevant area. Confirm with “OK”.

• The pressure image is imported and filtered. The evaluated result is displayed in the right window.

• Measuring points can be added to the pressure image as needed. For this purpose, right-click on the desired point. Select “Add note” in the menu. Left-click on the text and drag the text field onto the desired position.

• Force measurement and pressure image can be newly loaded as needed. At this point in time, the measurement has not been saved yet.

Once all data is correct, click on the “Complete” button. The measurement data is now saved in the folders. Next, the “Results and thresholds” window opens.

After scanning, check that all color fields of the calibration element were recognized. If this is not the case, check the calibration element!

Page 11: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 11 of 15

Step 4: Measurement results and thresholds After completion or after opening an existing measurement file, a separate window opens displaying the stored results.

After a measurement is “completed”, directories are created based on the names in the “Test series” and “Test name” fields (section “Step 2”). You can access the directory via the “Open measurement directory” button. 5 files are generated in the destination folder:

• A file with the raw data

• An image file of the force measurement

• An image file of the pressure measurement

• A Word report. The report is created based on a Word template (see below) and contains the configured data fields and/or bookmarks. The Word document can be opened an edited. This way, a measurement can be documented comprehensively.

• An image file across the scanned calibration element with the pressure measurement. This file can be imported later as needed via the “From file” button in the configuration window.

Measurement results

Thresholds of the specific locations of the group and measurement parameters. The selection of the body region shows the respective thresholds in the force diagram and the 3D view of the pressure image.

Open the destination folder with the generated files. Open the report (Word file).

Page 12: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

Subject to technical changes

Graph palette

Select the cursor function in the graph palette for a selective value determination. In the diagram for force analysis, you can move the cursor on the left edge of the table using drag and drop. The values at the

cursor position are displayed when releasin In the pressure distribution presentation, you can display the measurement value of a point. For this purpose, left-click on the desired point in the diagram.

Click on the “Zoom” menu and select one of the following functions

• Zoom into an area of the graph

• Zoom into an X

• Zoom into a Y

• Full view

• Zoom in

• Zoom out

csv Export Batch analysis can be used to compile the parameters and results of measurements into a .csv file. Startprocess via the “Evaluation\csv Exportconsider and compile all measurement data in this folder and its subfolders. Save the file. Next, the files can be imported into MS Excel.

GTE Industrieelektronik GmbH

FPM Vision – EN02

Select the cursor function in the graph palette for a selective value determination. In the diagram for force analysis, you can move the cursor on the left edge of the table using drag and drop. The values at the

cursor position are displayed when releasing the mouse button.

In the pressure distribution presentation, you can display the measurement value of a point. For this purpose, click on the desired point in the diagram.

Click on the “Zoom” menu and select one of the following functions

Zoom into an area of the graph

Zoom into an X-range of the graph

Zoom into a Y-range of the graph

Full view

Zoom in

Zoom out

Batch analysis can be used to compile the parameters and results of measurements into a .csv file. Startcsv Export” menu. You will be prompted to select a folder. The batch analysis will

consider and compile all measurement data in this folder and its subfolders.

Save the file. Next, the files can be imported into MS Excel.

GTE Industrieelektronik GmbH

Page 12 of 15

Select the cursor function in the graph palette for a selective value determination. In the diagram for force analysis, you can move the cursor on the left edge of the table using drag and drop. The values at the

In the pressure distribution presentation, you can display the measurement value of a point. For this purpose,

Batch analysis can be used to compile the parameters and results of measurements into a .csv file. Start the ” menu. You will be prompted to select a folder. The batch analysis will

Page 13: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 13 of 15

Notes regarding the scanning process: Place the Fuji Prescale pressure sample with the calibration sheet onto the scanner as shown below. Accurate placement of the calibration sheet is important.

Sample centered on the scanner, coarse side downward

Placement of the calibration sheet – the sample must be centered within the framed area

The calibration sheet must touch the marked corner. The edges of the calibration sheet must not cross the scan window

Page 14: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 14 of 15

Creating a report template Open the template via the “Evaluation\Create template” menu. You can control, which measurement data will be displayed, by inserting or deleting bookmarks. Place your logo and address data. Insert text, such as measurement execution instructions, and format the template. An overview of all possible bookmarks and their names can be found below.

Bookmarks

Bookmark Function

Parameter1 ... Parameter6 Content of field (Edit Me) 1 … Content of field (Edit Me) 6

Parameterlabel1 ... Parameterlabel6 Header of field (Edit Me) 1 … Header of field (Edit Me) 6

Date Date

Humidity Rel. Humidity

Location Location

PressuredistributionTransient Pressure distribution

Forcediagramm Force diagram

PressuredistributionStatic Pressure distribution

Executor Tester

Series Series

PressureMaxValue Pressure maximum, transient

PressureMaxTime Pressure maximum time, transient

ForceMaxValue Force maximum, transient

ForceMaxTime Force maximum time, transient

Temperature Temperature

ForceStaticValue Force maximum, static

ForceStaticTime Force maximum time, static

PressureStaticValue Pressure maximum, static

PressureStaticTime Pressure maximum time, static

Time Time

SoftwareVersion Software version

Measurement Measurement

K1Material K1 Material

K1Layer K1 Layer

K2SpringConstant K2 Spring const

K2Label K2 label

BodyRegion Body Region

ThresholdForceTransient Threshold force transient

ThresholdPressureTransient Threshold pressure transient

ThresholdForceStatic Threshold force static

ThresholdPressureStatic Threshold pressure static

Page 15: GTE Industrieelektronik GmbH Vision... · 2017-04-06 · GTE Industrieelektronik GmbH Subject to technical changes FPM Vision – EN02 Page 3 of 15 Software license The software is

GTE Industrieelektronik GmbH

Subject to technical changes FPM Vision – EN02 Page 15 of 15

KMGSerialnumber KMG Serial number

KMGType KMG Type

KMGNumber KMG Measurement number

FUJICalSN Calibration element S/N

FUJIType Film

SpecificBodyArea Specific Body area

SpecificBodyAreaNumber Specific Body area number

BodyRegionNumber Body Region number

AllThresholdForceTransient Threshold force transient

AllThresholdPressureTransien Threshold pressure transient

AllThresholdForceStatic Threshold force static

AllThresholdPressureStatic Threshold pressure static

SourceThresholds Source of the thresholds

AllSpecificBodyArea Specific Body area

AllSpecificBodyAreaNumbers Specific Body area number

FUJIQual Found Color fields

FUJIFilter Filter

TSType Film Type

TSSN Serial number of the film

TSEquilibrationfile Equilibration file

TSSensitivity Sensitivity

TSHandleSN Serial number of the Handle

KMGLowerforce Lower threshold of the force values

PiezoType Sensor Type

PiezoFrequency Measuring frequency

PiezoTriggerlevel Trigger level in N

PiezoPretrigger Data acquisition before trigger level

PiesoSN Serial number of the Logger

PiezoFilter Filter frequency of the force in Hz

Pressure3DMax Print image with transient limit value and maximum pressure

Pressure3Dstatic Print image with static limit value and static pressure

Warnings Warning message

PressureError Deviation of Pressure

ForceError Deviation of force