USES:• Hipot/Dielectric Withstand
- AC 5kV/100mA- DC 6kV/20mA
• Insulation Resistance- 50 to 5kVDC- 100kΩ~50GΩ
FEATURES:• 500VA output rating• Programmable AC Hipot Frequency
50Hz - 600Hz• Total and Real Current Measurement• Floating output complies with
EN50191• Open Short Check (OSC)• High Frequency Contact Check
(HFCC)• Ground Fault Interrupt• Standard RS232 and USB Interfaces• Standard HANDLER interface• Optional GPIB Interface• Key Lock on fail• Programmable voltage & test limit• Front and Rear Output Connectors• LabView Drivers are available• CE Mark
IntroductionQuadTech 500VA Plus Hipot Analyzer is designed for applications requiring a 500VA output transformer capacity with 100mA AC current output. The analyzer has both hipottest and analysis modes making the unit flexible for both production and R&D applications. The tests of AC/DC/IR can be programmed to 5000VAC, 6000VDC and5000VDC for IR. The full floating output complies with the EN50191 requirements foroperator safety.
DescriptionThe Guardian 500VA Plus features a large LCD screen for programming and measurement. Ground Fault Interruption and Floating Output features can also be used foroperator protection. The hipot tester is equipped with an output to monitor the arc signalwith an oscilloscope.
The user interface and overall operation is similar to the Guardian 6000 Plus. The remotecommand set features SCPI commands for both control and measurement. The commandsyntax is also compatible with the Guardian 6000 Plus.
The Guardian 500VA Plus is equipped with a new Function called High FrequencyContact Check (HFCC) besides the Open Short Check (OSC). These two features aredesigned to determine if a connection to the product has been made during testing or priorto testing. HFCC operates during testing and OSC is used as a step prior to hipot or IR test-ing. OSC operates by measuring the capacitive load (Cx) of the DUT from tens to thou-sands pF and looking for a change in capacitance from the nominalcapacitance of the DUT to determine connection.
HFCC (High Frequency Contact Check) is a new measurement technology for contactcheck. HFCC is able to be performed during the hipot test using a high frequency, around1MHz, for improved test time on the production line.
HIGH FREQUENCY CONTACT CHECK (HFCC, PATENT PENDING) & OPENSHORT CHECK (OSC, PATENT:254135)
Guardian 500VA Plus Hipot Tester
For more detailed information onspecifications, pricing and specialpurchase, rent and lease options,contact us at:
www.quadtech.com 800-253-1230
Hipot Testers • LCR Meters • Cable Testers • AC/DC Programmable Power Sources • Megohmmeters • Milliohmmeters
Guardian 500VA
AC Output Voltage: Range: 50V to 5000V AC, 2V resolution
Frequency: 50 to 600 Hz Programmable
Waveform: Sinusoidal
Regulation: <(1% +0.1% FS) at Rated Load
Voltage Display: Accuracy: ±(1% of reading + 0.1% FS)
Resolution: 2Volt
AC Current Display: Total currentRange: 1µA to 100mA AC
Resolution: 1µA
Accuracy: ±(1% of reading + 0.5% of range)
AC Real currentRange: 1µA to 100mA AC
Resolution: 1µA
Accuracy: ±(1% of reading + 5% of
Total Current)
High/Low Limit Test: 1µA to 100mA AC
Low limit can be turned OFF
DC Output Voltage: Range: 50V to 6000V DC, 2V resolution
Regulation: <(1% of setting +0.1% of FS) at
Rated Load
Voltage Display: Accuracy: ±(1% of reading + 0.1% FS)
Resolution: 2Volt
DC Current Display: Range: 0.1µA to 20mA DC
Resolution: 0.1µA
Accuracy: ±(1% or reading + 0.5% of range)
High/Low Limit Test: 0.1µA to 20mA DC
Low limit can be turned OFF
Insulation Resistance: Range: 100kΩ - 50GΩ
Accuracy: ±3% to ±15% depending upon
V and R, see manual for details
Resolution: 0.1MΩ
Voltage Range: 50V to 5000V DC
Voltage Accuracy: ± (5% of setting + 5V)
High/Low Limit Test: 100kΩ - 50GΩ
Low limit can be turned OFF
Common Features:
AC/DC Test Time: Ramp: 0.1 to 999s and Off
Test: 0.1 to 999s and Continuous
Dwell: 0.1 to 999s and Off (WDC only)
Fall: 0.1 to 999s and Off
Arc Detection: Programmable Level and OFF, 0.001mA/step
AC: 1mA to 20mA
DC: 1mA to 10mA
HFCC: Capacitance Range: 1pF to 100pF
Display Range: 1pF to 100pF
Accuracy: ± (50% of reading + 3pF)
Remote Control: Inputs: Start,Stop
Outputs: Pass/Fail/Under Test
Connector: 24 Pin Centronics
Standard Interfaces: Remote I/O, RS-232, USB
Optional Interfaces: GPIB
Test Setups: 100 Test Setups with 50 Steps each, Max. 500
Connectors: Front and Rear HV and GND Connections
Front Panel Lockout: Password
Safety Features: Fast Cutoff (<0.4ms) and Fast Discharge
<0.2s @5.1kV
GFI Ground Fault Interrupt
Floating Output porENS0191
Miscellaneous: Continuous Voltage on Fail
Indication: Pass/Fail lights, audible sound
Buzzer Level: 1, 2, 3, and Off
Dimensions: (w x h x d):17x6.8x17.7in (430x175x450mm)
Weight: 54 lbs (25kg) - Net, 61 lbs (27kg) Shipping
Environmental: Operating: 0°C to + 45°C,
Humidity: <95%
Storage: -10°C to +50°C
Warm-up Time: 1minute
Power: • 100 - 240V AC • 47 to 66Hz
• 1200W max
Guardian 500VA Plus Hipot TesterIncludes:150942 Instruction Manual700108 HV LeadW3800044 Return Lead320241 BNC to Binding Post Adapter700070 AC Power CableCalibration Certificate Traceable to NIST
Optional AccessoriesCalibration Data
S04 HV Lead Set 2m
S06* High Voltage ProbeS07* Power Entry Adapter CableS08* Gun ProbeS09 HV Lead, 1 meter, unterminatedS10 HV Lead, 2 meters, unterminatedS11* Gun Probe with Remote StartG13* Corded Product Adapter (115V)G16* International Power StripG25* Corded Product Adapter (240V)A190355 Rack Mount Flanges
*Maximum Voltage Rating: 5kV AC & 6kV DC A190356 GPIB Interface
Ordering InformationP
roduct Specifications
For more detailed information on specifications, pricing andspecial purchase, rent and lease options, contact us at: www.quadtech.com or 800-253-1230
P/N: 030210/A1